User'S Manual: DSO-3000 (A) Series
User'S Manual: DSO-3000 (A) Series
DSO-3000(A) Series
V1.0.2
Content
General Safety Summary .............................................................................. 1
Chapter 1 Getting Start ................................................................................. 2
1.1 System Requirements ............................................................................................3
1.2 Install Software .......................................................................................................4
1.3 Install Driver ............................................................................................................7
1.4 Connect Device .................................................................................................... 11
1.5 General Features ..................................................................................................15
1.6 General Check ......................................................................................................16
1.7 Probe Compensation ............................................................................................16
1.8 Function Check .....................................................................................................17
1.9 Self Calibration .....................................................................................................18
Chapter 2 Operating Basics ....................................................................... 19
2.1 The User’s Interface .............................................................................................20
2.2 The Menu System .................................................................................................22
2.3 The Vertical System ..............................................................................................24
2.4 The Horizontal System .........................................................................................25
2.5 The Trigger System ..............................................................................................25
2.6 Input Connectors ..................................................................................................27
Chapter 3 Oscilloscope Functions ............................................................ 28
3.1 Setup the Oscilloscope .........................................................................................29
3.2 Set Vertical System............................................................................................... 30
3.3 Set Horizontal System ..........................................................................................32
3.4 Set Trigger System ............................................................................................... 33
3.5 Set MATH..............................................................................................................34
3.6 Set Reference .......................................................................................................37
3.7 Fast Fourier Transform Function ..........................................................................38
3.8 Save/Load .............................................................................................................39
3.9 Utility Function ......................................................................................................40
3.10 Measure Signal ...................................................................................................42
3.11 The Display System ............................................................................................46
3.12 Zoom In/Out and Move Waveforms....................................................................50
3.13 Acquisition Modes ............................................................................................... 51
3.14 Print And Print Preview .......................................................................................51
Chapter 4 Application Example.................................................................. 52
4.1 Simple Measurement ............................................................................................53
4.2 Capturing a Single-Shot Signal ............................................................................54
4.3 The Application of the X-Y Operation ...................................................................54
4.4 Taking Cursor Measurements...............................................................................56
Chapter 5 Logic Analyzer Function ........................................................... 58
Chapter 6 Waveform Generator ................................................................. 61
6.1 Generate Waveform .............................................................................................62
6.2 Synchronized output .............................................................................................66
6.3 Digital Generator...................................................................................................67
Chapter 7 Appendix .................................................................................... 70
Appendix A: Specifications .........................................................................................71
Appendix B: Accessories ............................................................................................73
Appendix C: General Maintenance ............................................................................73
General Safety Summary
Use Proper Power Cord. Use only the power cord specified for this product and certified for the
country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while
they are connected to a voltage source.
Connect and Disconnect Properly. Connect the probe output to the measurement device
before connecting the probe to the circuit under test. Disconnect the probe input and the probe
reference lead from the circuit under test before disconnecting the probe from the measurement
device.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings
on the product. Consult the product manual for further ratings information before making
connections to the product.
Use Proper Probe. To avoid shock hazard, use a properly rated probe for your measurement.
Avoid Circuit or Wire Exposure. Do not touch exposed connections and components when
power is on.
Do Not Operate With Suspected Failures. If suspected damage occurs with the device, have it
inspected by qualified service personnel before further operations.
Provide Proper Ventilation. Refer to the installation instructions for proper ventilation of the
device.
USER’S MANUAL 1
Getting Start
System Requirements
Install software
Install Driver
Connect Device
General Features
General Check
Probe Compensation
Function Check
Self Calibration
USER’S MANUAL 2
Getting Start
To run the oscilloscope software, the needs of computer configuration are as follows:
Operating System
Window XP/Vista/Win 7/Win 8/Win 10
Processor
Upwards of 1.00G processor
Memory
256M byte
Disk Space
500M disk free space
Screen resolution
800 x 600
Recommended Configuration
Operating System
Windows XP SP3 System
Processor
2.4G Processor
Memory
1G Byte Memory
Disk Space
80G Disk Space
Screen resolution
1024 x 768 or 1280 x 1024 resolution
DPI Setting
Normal Size (96DPI)
USER’S MANUAL 3
Getting Start
2. The installation should start up automatically. Otherwise in Windows Explorer, switch to the
CD-ROM drive and run Setup.exe.
USER’S MANUAL 4
Getting Start
USER’S MANUAL 5
Getting Start
USER’S MANUAL 6
Getting Start
USER’S MANUAL 7
Getting Start
USER’S MANUAL 8
Getting Start
7. Prepare to install.
USER’S MANUAL 9
Getting Start
8. The installation is completed. you will see the running IO program in the lower right corner of the
screen.
USER’S MANUAL 10
Getting Start
USB connection
1.Power on. Connect DSO scope to PC via USB cable. The driver will install by itself. When the
driver is installed successfully, there will have a prompt as follows:
2.After installing successfully, users can see the driver information in the computer management.
3.Click Keysight IO libraries suite icon in bottom right corner, and choice “Connection Expert”. Now
users can see the device information.
4.If installing unsuccessfully, right click to update the driver manually, and choice the correct route:
C:\Program Files\IVI Foundation\VISA\IVI USB Staging\b841aa80\Windows
5.Double click “DSO-xxxx” shortcut on desktop to open the software of DSO device. The device is
connected with PC successfully. You can see “Connected” tip in status column.
USER’S MANUAL 11
Getting Start
LAN connection
1)Connect router with DSO scope via one LAN cable. Connect router with PC via the another
LAN cable.
2)“Open Network and Sharing Center->Local Area Connection-> Properties->Internet Protocol
Version (TCP/IPv4)” . Set the right IP of PC in same network segment with the oscilloscope’s
network name. The default IP is 192.168.1.127.
USER’S MANUAL 12
Getting Start
Click Keysight IO libraries suite icon in bottom right corner, and choice “Connection Expert”. Now
users can see the device information.
USER’S MANUAL 13
Getting Start
2.Connect Device
Double click “DSO-xxxx” shortcut on desktop (Or click “File->Connect” menu) to open the software
of DSO device.
1)Select “LAN” type, the device connect setting dialog is as shown below:
2)The device is connected with PC successfully. You can see “Connected” tip in status column.
USER’S MANUAL 14
Getting Start
Pass/Fail;
20 Automatic measurements;
Cursor measurements;
External trigger;
Waveform average;
USER’S MANUAL 15
Getting Start
■ From the “Probe” menu, select attenuation to 1:10. Set the switch to “X10” on the probe and
connect it to CH1 of the oscilloscope. When using the probe hook-tip, insert the tip onto the probe
firmly to ensure a proper connection.
■ Attach the probe tip to the Probe Compensator and the reference lead to the ground
connector, select CH1, and then press the “AUTOSET” button into the menu or the toolbar.
Correctly Compensated
Over Compensated
USER’S MANUAL 16
Getting Start
Under Compensated
1. If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe for the fattest
square wave being displayed on the oscilloscope.
2. Repeat if necessary.
WARNNING: To avoid electric shock while using the probe, be sure the perfection of the insulated
cable, and do not touch the metallic portions of the probe head while it is connected with a voltage
source.
1. Set the attenuation switch on the probe as 10X and connect the probe on the oscilloscope with
CH1. Aim the slot in the probe connector at the faucet on BNC of CH1 and insert, then, turn right to
lock the probe. Finally, attach the tip of probe and ground nip to the Connector of Probe
compensator.
USER’S MANUAL 17
Getting Start
2. Set the CH1 probe attenuation of the oscilloscope to X10. (The default is X1).
3.Attach the tip of probe and ground nip to the Connector of Probe compensator. Click the
button. A square wave will be displayed within a several seconds. (Approximately 1 kHz, 1V, peak-
to- peak).
4. Inspect CH2, CH3, CH4 with the same method. Repeat steps 2 and 3.
USER’S MANUAL 18
Operation Basics
Input Connectors
USER’S MANUAL 19
Operation Basics
In addition to displaying waveforms, the display area is filled with many details about the waveform
and the oscilloscope control settings.
1.The Main Menu
All settings can be found in the main menu.
2.The Toolbar
3.Change DSO and DDS
4.The Horizontal Panel
The user can change Time/Div, Delay Time in the panel.
5.The Vertical Panel
The user can turn on/off the CH1/CH2/CH3/CH4. Also the user can change the CH1/CH2/CH3
/CH4 volt/div, coupling and probe attenuation.
6.The Trigger Panel
In this panel, the user can change the trigger mode, sweep, source and slope.
7. It shows the system time.
8. Marker shows Edge trigger level.
9. It shows the main time base setting.
10.It shows the CH1/CH2/CH3/CH4 information.
Readouts show the coupling of the channels.
USER’S MANUAL 20
Operation Basics
USER’S MANUAL 21
Operation Basics
USER’S MANUAL 22
Operation Basics
USER’S MANUAL 23
Operation Basics
1.Select channel: User can select the channel by clicking the Combo box.
2.ON/OFF: Turn on or off the selected channel.
3.Volt/Div: Set the selected channel voltage range.
4.Coupling: Set the selected channel to DC/AC/GND.
5.Probe: Set the Select one according to the probe attenuation factor to ensure correct vertical
scale reading
6.Bw Limit: Reject the frequency component higher than 20MHz.
7.Invert: Invert the selected wave.
USER’S MANUAL 24
Operation Basics
The following figure shows the Horizontal System window. It shows the horizontal parameters
settings.
The trigger determines when the oscilloscope begins to acquire data and display a waveform.
Once a trigger is properly set up, the oscilloscope can convert unstable displays or blank screens
to meaningful waveforms. Here introduce some basic concepts about trigger.
USER’S MANUAL 25
Operation Basics
Trigger Source: The trigger can be generated with multiple sources. The most common one is the
input channel. Whether the input signal is displayed or not, it can trigger normal operations. Also
the trigger source can be any signal connected to an external trigger channel or the AC power line
(only for Edge triggers). The source with the AC power line shows the frequency relationship
between the signal and the AC commercial power.
Trigger Mode: The oscilloscope has five types of triggers: Edge, Pulse Width, Slope, Video and
Timeout.
Edge Trigger uses the analog or digital test circuits for triggering. It happens when the
input trigger source crosses a specified level in a specified direction.
Pulse Width Trigger can trigger normal or abnormal pulses that meet trigger conditions.
Slope Trigger uses the rise and fall times on the edge of signal for triggering.
Video Trigger performs a field or line trigger through standard video signals.
Timeout Trigger happens after the edge of signal reaches the set time.
Trigger Sweep: You can select the Auto or Normal mode to define how the oscilloscope acquires
data when it does not detect a trigger condition. Auto Mode performs the acquisition freely in
absence of valid trigger. Normal Mode updates the displayed waveforms only when the
oscilloscope detects a valid trigger condition. Before this update, the oscilloscope still displays the
old waveforms. This mode shall be used when you want to only view the effectively triggered
waveforms. In this mode, the oscilloscope displays waveforms only after the first trigger. To
perform a single sequence acquisition, push the SINGLE SEQ button.
Trigger Position: The horizontal position control establishes the time between the trigger position
and the screen center.
Slope and Level: The Slope and Level controls help to define the trigger. The Slope option
determines whether the trigger point is on the rising or falling edge of a signal. To perform the
trigger slope control, press the TRIG MENU button, select an Edge trigger, and use the Slope
button to select rising or falling. The TRIGGER LEVEL knob controls the trigger point is on which
position of the edge.
USER’S MANUAL 26
Operation Basics
Other Connector:
USER’S MANUAL 27
Oscilloscope Functions
Set MATH
Set Reference
Save/Load
Utility Function
Measure Signal
Display System
Acquisition Modes
USER’S MANUAL 28
Oscilloscope Functions
The oscilloscope will change the current settings to display this signal.
Save Setup
The oscilloscope software saves the current setup before you close the oscilloscope software.
The oscilloscope recalls this setup the next time you run the software. You can use the “Save
Setup” menu to permanently save up to several different setups.
Load Setup
The oscilloscope can recall the last setup before the oscilloscope software was running, any saved
setups, or the factory setup. You can use the “Load Setup” menu to recall a setup.
Factory Setup
The oscilloscope software is set up for normal operation when it is shipped from the factory. This is
the factory setup. To recall this setup, push the “Utility -> Factory Setup” button.
USER’S MANUAL 29
Oscilloscope Functions
The following figure shows the Vertical System window. It shows the vertical parameters settings.
Drag Vertical Position icon to move waveform up and down on the screen. Double
click the icon to reset the waveforms to the vertical center position on the screen.
Change Volt/Div
You can click “Volt/Div” in “Vertical Setting” window to set the voltage.
You can left click and drag the mouse on the knob to change the voltage.
Select the attenuation factor for the probe. To check the probe attenuation setting, toggle the
probe menu to match the attenuation factor of the probe.
This setting remains in effect before you changed again. Click “Probe” in “Vertical Setting”
window to select the probe attenuation.
Note: The attenuation factor changes the vertical scale of the oscilloscope so that the
measurement results refect the actual voltage levels at the probe tip.
USER’S MANUAL 30
Oscilloscope Functions
Bandwidth Limit
The oscilloscope is set to full bandwidth and will pass the high frequency component in the signal if
the “BW Limit” was turned off.
The oscilloscope will reject the frequency component higher than 20MHz if the “BW Limit” was
turned on.
Invert
The invert function turns the displayed waveform 180 degrees, with respect to the ground level.
When the oscilloscope is triggered on the inverted signal, the trigger is also inverted.
USER’S MANUAL 31
Oscilloscope Functions
The following figure shows the Horizontal System window. It shows the horizontal parameters
settings.
Horizontal Setting
Format
Set the waveform display format (Y-T, X-Y, Scan and Roll).
Y -T: Show the relative relation between vertical voltage and horizontal time.
X -Y: Show CH1 value at X axis; CH2 value at Y axis. The XY format is used to analyzer phase
difference, such as those represented by Lissajous patterns. Please refer to 4.3 Section.
X-Y Setting: Users can change X-Coordinate and Y-Coordinate for CH1, CH2, CH3, CH4.
YT Mode
Scan: In Scan Mode, the waveform display updates from left to right.
Roll: In Roll Mode, the waveform display updates from right to left.
In Roll mode, the waveform display rolls from right to left. No trigger or horizontal offset control of
waveforms is available during Roll Mode, and it’s only available when set to 100ms/div or slower.
Drag Horizontal Position icon on screen to adjust the trigger position against screen. Double
click the icon to reset the trigger point back to the screen center.
USER’S MANUAL 32
Oscilloscope Functions
The user can also change the trigger setting on trigger panel in sidebar.
Trigger Setup
The trigger determines when the oscilloscope starts to acquire data and display a waveform. When
a trigger is set up properly, it can convert unstable displays or blank screens into meaningful
waveforms.
If the oscilloscope wants to acquire a waveform, it collects enough data so that it can draw the
waveform to the left of the trigger point. The oscilloscope continues to acquire data while waiting
for the trigger condition to occur. The oscilloscope continues to acquire enough data so that it can
draw the waveform to the right of the trigger point after it detects a trigger.
The Edge trigger determines whether the scope finds the trigger point on the rising or the falling
edge of the signal. Select Edge trigger mode to trigger on Rising edge or Falling edge.
USER’S MANUAL 33
Oscilloscope Functions
Trigger Mode: Select the trigger mode, Edge, Pulse, Slope, Video or Timeout..
Trigger Sweep: Set the sweep mode to Auto, Normal or Single.
Auto: Acquire waveform even no trigger occurred
Normal: Acquire waveform when trigger occurred.
Single: Acquire waveform when trigger occurred then stop.
Trigger Source: You can use the trigger source options to select the signal that the oscilloscope
uses as a trigger. The source can be any signal connected to a channel BNC.
CH1: Select CH1 as trigger signal
CH2: Select CH2 as trigger signal
CH3: Select CH3 as trigger signal
CH4: Select CH4 as trigger signal
EXT/10: Select external signal as trigger signal. Does not display the trigger signal and
attenuates the signal by a factor of 10.
Trigger Slope: Set the slope to Rising (+) or Falling (-).
Rising: Trigger on rising edge
Falling: Trigger on falling edge
USER’S MANUAL 34
Oscilloscope Functions
Operate
Four Types:
A+B Add source A and source B
A-B Subtract source B from source A
AxB Multiply source A by source B
A/B Divide source A by source B
In this function, use the addition, subtraction, multiplication and division function to operate and
analyze the waveform.
Select the operate type in the Operate menu. Select source A and B. Then adjust the vertical scale
and offset to view the math channel clearly. The mathematic result can be measured by the
measure and the cursor.
Invert
The invert function turns the displayed waveform 180 degrees, with respect to the ground level.
When the oscilloscope is triggered on the inverted signal, the trigger is also inverted.
USER’S MANUAL 35
Oscilloscope Functions
USER’S MANUAL 36
Oscilloscope Functions
Note: If you turn on the “Reference” channel, the load file window will appear.
USER’S MANUAL 37
Oscilloscope Functions
To use the FFT mode, you need to click “Setup-> FFT” to open the FFT setting window.
You can select the Source channel, Window type, FFT vertical scale units, vertical scale and FFT
Zoom factor. It displays only one FFT spectrum at a time.
Source: Selects the channel used as the FFT source.
Window: Selects the FFT window type.
The window reduces spectral leakage in the FFT spectrum.
Scale: Selects the vertical scale units.
Vertical: Selects the vertical scale.
Horizontal: Selects the FFT zoom factor.
The FFT display Window:
USER’S MANUAL 38
Oscilloscope Functions
3.8 Save/Load
Save
Click “File” in main menu to save waveform data, setups and image.
1. Save Data
2. Save Setup
3. Save Image
Load
1. Load Data
2. Load Setup
USER’S MANUAL 39
Oscilloscope Functions
This function can record input waveform form CH1, CH2, CH3 and CH4.
“Start” button: Start to record waveforms.
“Stop” button: Stop to record waveforms.
Clicking “Record Scanner” in “Utility” menu will play the recording data.
3.9.2 Pass/Fail
Click “Pass/Fail” in “Utility” menu.
The Pass/Fail window appears:
The Pass/Fail function monitors changes of signals and outputs pass or fail signals by comparing
the input signal with the pre-created mask.
USER’S MANUAL 40
Oscilloscope Functions
Control Setting
Source: Select the Pass/Fail channel
Output: Select the Pass/Fail output condition.
Stop When Output: If it was checked, the Pass/Fail will stop when output.
Mask Setting
Vertical: Set the vertical limit range
Horizontal: Set the horizontal limit range
“Create” button: Click this button to create Pass/Fail area according to the mask
“Save” button: Click this button to save the setups to file
“Load” button: Click this button to load the saved setups file
Information Display
USER’S MANUAL 41
Oscilloscope Functions
When you click the Factory Setup in Utility menu, the oscilloscope displays the CH1/CH2/CH3
/CH4 waveform and removes all other waveforms.
The oscilloscope set up for normal operation when it is shipped from the factory and can be
recalled at anytime by user.
The Factory Setup function does not reset the following settings:
Language option
Date and time
1. Source
The user can set the source to CH1, CH2, CH3 and CH4.
When you use cursors, be sure to set the Source to the waveform on the display that you want to
measure.
2. Type
USER’S MANUAL 42
Oscilloscope Functions
2)Vertical
The Vertical cursors appear as vertical lines on the display and measure the vertical parameters.
The Vertical cursor display window:
USER’S MANUAL 43
Oscilloscope Functions
3)Horizontal
The Horizontal cursors appear as horizontal lines on the display and measure the horizontal
parameters.
Edit Measure
Click “Measure->Edit”.
The oscilloscope provides 20 parametric auto measurements (12 voltage and 8 time
measurements).
USER’S MANUAL 44
Oscilloscope Functions
1.Options
The user can select a measure source CH1, CH2, CH3 or CH4.
2.Vertical
Maximum: Voltage of the absolute maximum level, Measured over the entire waveform
Minimum: Voltage of the absolute minimum level, Measured over the entire waveform
Peak To Peak: Peak-to-peak = Max –Min, Measured over the entire waveform
Top: Voltage of the statistical maximum level, Measured over the entire waveform
Base: Voltage of the statistical minimum level, Measured over the entire waveform
Middle: Voltage of the 50% level from base to top
RMS: The Root Mean Square voltage over the entire waveform
Amplitude: Amp = Base - Top, Measured over the entire waveform
Mean: The arithmetic mean over the entire waveform
Cycle Mean: The arithmetic mean over the first cycle in the waveform
Positive OverShoot: Positive Overshoot = (Max - Top)/Amp x 100 %, Measured over the entire
waveform
Negative OverShoot: Negative Overshoot = (Base - Min)/Amp x 100 %, Measured over the entire
waveform.
3.Horizontal
Period: Time to take for the first signal cycle to complete in the waveform
Frequency: Reciprocal of the period of the first cycle to complete in the waveform
Rise Time: Time taken from lower threshold to upper threshold
Fall Time: Time taken from upper threshold to lower threshold
+Duty Cycle: Positive Duty Cycle = (Positive Pulse Width)/Period x 100%, Measured of the first
cycle in waveform.
-Duty Cycle: Negative Duty Cycle = (Negative Pulse Width)/Period x 100%, Measured of the first
waveform.
+Pulse Width: Measured of the first positive pulse in the waveform. The time between the 50%
amplitude points
-Pulse Width: Measured of the first negative pulse in the waveform. The time between the 50%
amplitude points
4.Clear
Clear all measure items on display screen.
5.Undo
Undo all measure items on display screen.
Clear Measure
Click “Measure->Clear”.
Clear all measure items on display screen.
USER’S MANUAL 45
Oscilloscope Functions
Display Type
Click “Type” in “Display” menu.
USER’S MANUAL 46
Oscilloscope Functions
If the Vectors type mode is selected, the waveform will be displayed as following figure.
If the Dots type mode is selected, the waveform will be displayed as following figure.
Display Grid
Click “Display” in main menu.
USER’S MANUAL 47
Oscilloscope Functions
Intensity
Click “Display->Intensity” in main menu.
The following figure shows the intensity dialog. It shows the display parameters setting.
USER’S MANUAL 48
Oscilloscope Functions
You can change the grid and waveform color intensity in this dialog.
Waveform Color
Click “Display-> Color” in main menu.
The following figure shows the Color Setting dialog. It shows the waveform color setting.
If you click the color rectangle, the color dialog will be showed:
USER’S MANUAL 49
Oscilloscope Functions
Zoom In/Out
The user can click “Zoom In/Out” in “Acquire” menu, then left click the mouse button on display
screen to zoom in/out the waveform. Also the user can change Time/Div in Horizontal menu or
in Horizontal panel to zoom in/out the waveform.
Move Waveforms
The user can change “Delay Time” on the Horizontal Panel to move the waveform.
USER’S MANUAL 50
Oscilloscope Functions
Normal: In this acquisition mode, the oscilloscope samples the signal in evenly spaced intervals to
construct the waveform.
Average: In this acquisition mode, the oscilloscope acquires several waveforms, averages them,
and displays the resulting waveform. You can use this mode to reduce random noise.
1. Click “Print” in “File” menu to set the printer to print the current waveform.
2. Click the “Print Preview” in “File” menu to get into the Preview window.
In “Print Preview” window, use the “Zoom In” button and the “Zoom Out” button to change the
size of the waveform graph. Click the “Close” button to turn this window off and click the “Print”
button to print the report.
USER’S MANUAL 51
Application Example
Simple Measurement
USER’S MANUAL 52
Application Example
The DSO set the vertical, horizontal, and trigger controls at the best status automatically. Also,
you can adjust the controls to meet your measurement to optimize the waveform display.
To measure the frequency and “Vpp”, you can do these steps as follows:
1. Click the “Measure Menu->Horizontal->Frequency” button, the frequency of the signal will
display on the bottom of the waveform interface.
2. Click the “Measure Menu->Vertical->Peak-to-Peak” button, the “Vpp” of the signal will also
display on the bottom of the waveform interface.
To clear the measurement on the waveform interface, click the “Measure->Clear” button.
USER’S MANUAL 53
Application Example
2. Set up the trigger in the Trigger Menu, or in the Trigger Setting window.
4) Set the Trigger Slope to “+” which means you select the rising edge.
5) Adjust the Volts/Div and the time base in a proper range for the signal.
6) Drag the trigger level sign on the waveform display screen to proper position. It ally higher a
little above the normal level.
7) Click button to start capturing. When the trigger conditions are met, data appears on the
display representing the data points that the oscilloscope obtained with one acquisition.
This function helps to capture the signal occurrence easily, such as the noise with large
amplitude; set the trigger level higher a little above the normal level and press and wait. When
noise occurs, the instrument will record the waveform before and after the trigger.
1. Set the probe attenuation to “x10”. Set the switch to “x10” on the probes.
2. Connect the CH1 probe to the input of the circuit, and connect the CH2 probe to the output of
the circuit.
3. Click button.
4. Adjust the vertical scale and offset to display approximately the same amplitude signals on
each channel.
5. Select X-Y format at Horizontal window. The oscilloscope will displays a Lissajous pattern
representing the input and the output characteristics of the circuit.
6. Adjust the scale and offset of the horizontal and vertical to a desirable waveform display.
USER’S MANUAL 54
Application Example
7. Apply the Ellipse Method to observe the phase difference between the two channels.
Sinθ = A/B or C/D, where θ = phase shift (in degrees) between the two signals.
From the formula above:
θ = _arcsine (A/B) or _arcsine (C/D)
θ must be in the range of (0~π/2) or (3π/2~2π) if the main axis of the ellipse is between I and III
quadrant, . If the main axis is at II and IV quadrant, θ must be in the range of (π/2~π) or (π~3π/2).
USER’S MANUAL 55
Application Example
Do these steps:
1. Click “Cursor->Source”, select CH1 (select CH2 if you want measure CH2).
5. Release the left mouse button, the frequency difference and time difference will be shown at
the status bar.
Do these steps:
1. Click “Cursor->Source”, select CH1 (select CH2 if you want measure CH2).
5. Release the left mouse button, the voltage difference will be shown at the status bar.
USER’S MANUAL 56
Application Example
Note: Click “Cursor->Type”, select “Cross”, you can measure time and amplitude at one time.
USER’S MANUAL 57
Appendix
USER’S MANUAL 58
Appendix
As following picture:
2. Set Threshold
Set TTL Threshold Voltage to 0.5V, and click “OK”.
USER’S MANUAL 59
Appendix
Through the LOGIC OUT port output to the LOGIC analyzer LOGIC IN port, measurement result is
shown in figure:
1 2
USER’S MANUAL 60
Appendix
Generate Waveform
Synchronized Output
Digital Generator
USER’S MANUAL 61
Appendix
1.Press the check box "Signal On" to open the wave output function.
1. Press the check box "Signal On" to open the wave output function.
1. Press the check box "Signal On" to open the wave output function.
USER’S MANUAL 62
Appendix
1. Press the check box "Signal On" to open the wave output function.
1. Press the check box "Signal On" to open the wave output function.
1. Press the check box "Signal On" to open the wave output function.
USER’S MANUAL 63
Appendix
1. Press the check box "On/Off" to open the wave output function.
The user can drag the mouse to draw arbitrary waveform. Also you can double click the waveform
display area to open arbitrary wave editor window, and edit the voltage of every points.
After the completion of the set, click “Download” to output Arbitrary Waveform.
1. Press the check box "On/Off" to open the wave output function.
USER’S MANUAL 64
Appendix
Output type:
Single: The waveform output only once. The user selects "Single", and clicks "Burst". Each time
Waveform generator is set to internal and external trigger. The oscilloscope trigger mode is set to
USER’S MANUAL 65
Appendix
The waveform on CH1 is the sine waveform you generated from the “OUTPUT” terminal, and that
on CH2 is the synchronized signal from “SYNC OUT” ports.
USER’S MANUAL 66
Appendix
The user can edit the 00-15 high-low level, and output by LOGIC OUT port.
For example, set low eight to HHHHLLLL to LOGIC analyzer LOGIC IN port. The measurement
result as following figure:
USER’S MANUAL 67
Appendix
For example, set high eight to HHHHLLLL to LOGIC analyzer LOGIC IN port. The measurement
result as following figure.
USER’S MANUAL 68
Appendix
3.Connect Digital Generator LOGIC OUT of 00-11 ports to LOGIC Analyzer LOGIC IN of 00-11
ports. The measurement result as following figure:
USER’S MANUAL 69
Appendix
Chapter 7 Appendix
Appendix A: Specifications
Appendix B: Accessories
USER’S MANUAL 70
Appendix
Appendix A: Specifications
Specifications Table:
USER’S MANUAL 71
Appendix
Max Sample
-- 250MSa/s -- 250MSa/s -- 250MSa/s
Rate
Bandwidth -- 10MHz -- 10MHz -- 10MHz
TTL,LVTTL TTL,LVTTL TTL,LVTTL
-- -- --
CMOS, CMOS, CMOS,
Compatible
LVCMOS, LVCMOS, LVCMOS,
input
-- ECL. -- ECL -- ECL
PECL,EIA PECL,EIA PECL,EIA
Storage depth -- 1.6K~64M -- 1.6K~64M -- 1.6K~64M
Waveform Generator Mode
Waveform
-- DC~25MHz -- DC~25MHz -- DC~25MHz
Frequency
2K~200MHz 2K~200MHz 2K~200MHz
DAC -- adjustable
-- adjustable
-- adjustable
Frequency
-- 0.10% -- 0.10% -- 0.10%
Resolution
Channel -- 1CH -- 1CH -- 1CH
Waveform
-- 4K -- 4K -- 4K
Depth
Vertical
-- 12bits -- 12bits -- 12bits
Resolution
Frequency
-- <30ppm -- <30ppm -- <30ppm
Stability
Wave
-- ±3.5V Max. -- ±3.5V Max. -- ±3.5V Max.
Amplitude
Output
-- 50 Ω -- 50 Ω -- 50 Ω
Impedance
Output Current -- 50mA -- 50mA -- 50mA
System BW -- 25M -- 25M -- 25M
Interface USB 2.0
Power Source 12--36V Wide range of input voltage
Volume 255 x 190 x 45 (mm)
Weight 1Kg
USER’S MANUAL 72
Appendix
Appendix B: Accessories
DSO3104, DSO3204, DSO3254
Probe×4 (1.5m), 1:1, (10:1) Passive Probes
A power adapter
A USB Cable
A PC software CD of the oscilloscope
A power adapter
A USB line
A bag(18) of little test hook Logic analyzer with high-quality test hook
Caution
To avoid damages to the device or probes, do not expose them to sprays, liquids or solvents.
To avoid damages to the surface of the device or probes not use any abrasive or chemical
cleaning agents.
Cleaning
Inspect the device and probes as often as operating conditions require. Make sure the device
disconnect form all power sources.
1. Remove loose dust on the outside of the oscilloscope and probes with a lint-free cloth. Use care
to avoid scratching the clear glass display filter.
USER’S MANUAL 73