MIC-5010 / MIC-5005: Insulation Resistance Meter

The document describes two insulation resistance meters, the MIC-5010 and MIC-5005. [1] The meters can measure insulation resistance from 50 kΩ to 15 TΩ, capacitance from 1 nF to 49.99 μF, and DC/AC voltage from 0.1 V to 600 V. [2] Both meters feature 990 memory cells, automatic discharge, and filters to reduce noise. The MIC-5010 also measures continuity and has adjustable limits. [3] The meters are powered by the main power line or internal battery and include standard test leads, clips, carrying case, and software for analysis. Additional optional accessories are also listed.

Uploaded by

Jaaffer Ali
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
125 views2 pages

MIC-5010 / MIC-5005: Insulation Resistance Meter

The document describes two insulation resistance meters, the MIC-5010 and MIC-5005. [1] The meters can measure insulation resistance from 50 kΩ to 15 TΩ, capacitance from 1 nF to 49.99 μF, and DC/AC voltage from 0.1 V to 600 V. [2] Both meters feature 990 memory cells, automatic discharge, and filters to reduce noise. The MIC-5010 also measures continuity and has adjustable limits. [3] The meters are powered by the main power line or internal battery and include standard test leads, clips, carrying case, and software for analysis. Additional optional accessories are also listed.

Uploaded by

Jaaffer Ali
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Index: WMGBMIC5010

WMGBMIC5005

INSULATION RESISTANCE METER


MIC-5010 / MIC-5005

Capability of wireless transmission


of the data from memory to a PC
ORDER ADDITIONALLY
THE WIRELESS USB INTERFACE

· Insulation resistance measurement:


- measurement voltage any in the range of 50…1000 V at 10 V and 1000...5000 V at 25 V resolution,
- continuous indication of measured insulation resistance or leakage current,
- automatic discharge of measured object capacitance voltage after the end of insulation resistance measurement,
- acoustic signaling of 5 seconds intervals to facilitate capturing time characteristics,
- metered T1, T2 and T3 test times for measuring one or two absorption coefficients from the range of 1…600 s,
- adjustable measuring time to 99'59'’,
- polarization index (PI) and dielectric absorption ratio (DAR) measurement,
- indication of actual test voltage during measurement,
- 1.2 mA and 3 mA test current,
- step voltage insulation resistance measurement (SV),
- Dielectric Discharge calculation (DD),
- protection against measuring live objects,
- measurements with test leads up to 20 m

· Digital filters function for measurements in high noise enviroment (10 s, 30 s, 60 s).
· MIC-5010: Continuity measurement of protective connections and equipotential bonding in accordance with
Sonel S.A. EN 61557-4 with current > 200 mA.
ul. Wokulskiego 11 · MIC-5010: Adjustable limits for measured resistance RISO and RCONT.
58-100 Świdnica, PL · Measurement of leakage current during insulation resistance testing.
· Measurement of capacitance during the measurement of RISO.
tel. +48 74 85 83 860
· DC and AC voltage measurement in the range of 0…600 V.
fax +48 74 85 83 809 · 990 cells of memory (11880 records) with the capability of wireless data transmission to a PC (with the USB-OR
adapter) or through a USB cable.
· Power supply from main power line or battery packs, low battery warning indicator, built-in fast charger.
[email protected] · Keyboard and display backlit (MIC-5005 - only display).
The instruments meet the requirements of the EN 61557 standard.
www.sonel.pl
MIC-5010/MIC-5005
Insulation resistance measurement
Measurement range acc. to IEC 61557-2: 50 kΩ...15,0 TΩ (IISOnom = 1,2 mA or 3 mA) Measurement of capacitance

Range Resolution Accuracy Display range Resolution Accuracy

1...999 nF 1 nF
0...999 kΩ 1 kΩ ±(5% m.v. + 5digits)
1.00...49.99 µF 0.01 µF
1.00..9.99 MΩ 0.01 MΩ
10.0...99.9 MΩ 0.1 MΩ · Capacity measurement result is displayed after the RISO measurement
±(3% m.v. + 10 digits)
100..999 MΩ 1 MΩ
1.00...9.99 GΩ 0.01 GΩ DC and AC voltage measurement
10.0...99.9 GΩ 0.1 GΩ
Range Resolution Accuracy
100...999 GΩ 1 GΩ ±(3.5% m.v. + 10 digits)
1.00...9.99 TΩ 0.01 TΩ ±(7.5% m.v. + 10 digits) 0.0...29.9 V 0.1 V ±(2% m.v. + 20 digits)

10.0...15.0 TΩ 0.1 TΩ ±(10% m.v. + 10 digits) 30.0...299.9 V 0.1 V ±(2% m.v. + 6 digits)
300...600 V 1V ±(2% m.v. + 2 digits)
Values of measured resistance depending on measurement voltage · Frequency range: 45...65Hz

Voltage UISO Measurement range


Standard accessories:
250 V 500 GΩ - test lead banana plug; 1,8 m; 10kV; red WAPRZ1X8REBB10K
500 V 1.00 TΩ - test lead banana plug; 1,8 m; 10 kV; blue WAPRZ1X8BUBB10K
- test lead banana plug; 1,8 m; 10 kV; black; shielded WAPRZ1X8BLBBE10K
1000 V 2.00 TΩ - USB cable WAPRZUSB
- "crocodile" clip 5,5 kV; black WAKROBL32K07
2500 V 5.00 TΩ - "crocodile" clip 5,5 kV; red WAKRORE32K07
5000 V 15.0 TΩ - "crocodile" clip 5,5 kV; blue WAKROBU32K07
- pin probe 5,5 kV with banana connector; red WASONREOGB5X5
- pin probe 5,5 kV with banana connector; black WASONBLOGB5X5
Measurement of leakage current - carrying case L4 for accesories WAFUTL4
- power cord WAPRZ1X8BLIEC
Range Resolution Accuracy - battery pack (built-in)
resolution and units result - “SONEL Reader” software
0…1,2 mA* from the measurement range Calculated basing on - calibration certificate
0…3 mA*
of individual insulation resistance measurements
resistance.
Additional accesories:
* - depending on the setting
- test lead banana plug; 3 m; 10kV; red WAPRZ003REBB10K
- test lead banana plug; 3 m; 10kV; blue WAPRZ003BUBB10K
- test lead banana plug; 3 m; 10kV; black; shielded WAPRZ003BLBBE10K
Step voltage insulation resistance measurement - test lead banana plug; 5 m; 10kV; red WAPRZ005REBB10K
- test lead banana plug; 5 m; 10kV; blue WAPRZ005BUBB10K
Measurement - test lead banana plug; 5 m; 10kV; black; shielded WAPRZ005BLBBE10K
Target voltage - test lead banana plug; 10 m; 10kV; red WAPRZ010REBB10K
voltage sequence
- test lead banana plug; 10 m; 10kV; blue WAPRZ010BUBB10K
1 kV 200, 400, 600, 800, 1000 V - test lead banana plug; 10 m; 10kV; black; shielded WAPRZ010BLBBE10K
- test lead banana plug; 20 m; 10kV; red WAPRZ020REBB10K
2.5 kV 0.5, 1, 1.5, 2, 2.5 kV - test lead banana plug; 20 m; 10kV; blue WAPRZ020BUBB10K
5 kV 1, 2, 3, 4, 5 kV - test lead banana plug; 20 m; 10kV; black; shielded WAPRZ020BLBBE10K
- carring backpack L-7 WAFUTL7
· duration of each "step" adjustable from 30 s to 5 mins - OR-1 radio receiver for data transmission WAADAUSBOR1
· measurement result for each voltage step is stored in memory

Continuity measurement of protective connections and equipotential bonding Electrical safety:


with 200 mA current (MIC-5010 only) - type of insulation double, in acc. with EN 61010-1 and IEC 61557
Measurement range acc. to IEC 61557-4: 0.12...999 Ω - measurement category IV 600 V (III 1000 V) in acc. with EN 61010-1
- case protection rating in acc. with EN 60529 IP54 (IP67 - with lid closed)
Range Resolution Accuracy
Other technical specifications:
0.00...19.99 Ω 0.01 Ω - power supply of the meter built-in battery pack
±(2% w.m. + 3 digits) - weight of the meter approx. 7 kg
20.0...199.9 Ω 0.1 Ω - dimensions 390 x 310 x 170 mm
200...999 Ω 1Ω ±(4% w.m. + 3 digits) - display LCD segment display
- measurement results memory 990 cells of memory (11880 records)
· Voltage on open terminals: 4...24 V
- transmission of measurement results USB or wireless interface
· Output current at R < 15 Ω: min. 200 mA (ISC: 200...250 mA)
· Compensation of test lead resistance
· Current flowing in both directions, mean value of resistance is displayed
The acronym “m.v.” stands for a “measured reference value”.