Standard-Eqp Oq Uv Vis.01.05
Standard-Eqp Oq Uv Vis.01.05
Standard-Eqp Oq Uv Vis.01.05
Table of Contents
Section Page
What is EE 5
RevHistoryUVVisOQ 8
UV-VIS OQ 10
Customer Approval 20
Legal Notice 21
Protocol Details 22
The Equipment Qualification Plan (EQP) documents the activity program that is performed during the qualification services for the
applicable systems. A complete description of the test specifications is provided for the supported services, including setpoints and
acceptance criteria (or limits) for each test. The test specification section of this document is created directly from the EQP file name
listed on the cover.
This document is an abstraction of the EQP file used to perform the service and is generated directly from the electronic Agilent
Equipment Qualification Plan (eEQP) Editor. The purpose of this document is to allow the user to review and record approval of the
EQP that guides the delivery of compliance services provided by the Agilent Automated Compliance Engine.
Statement of Intent
Unless otherwise requested, the qualification is delivered according to the standard test program described in the
Agilent_Recommended EQP. Agilent defines variances as changes to the default recommended values (as stated in the Agilent
Recommended EQP) that fall within a well-defined range. These changes are considered to be within the intended use range of the
system under test.
Customizations are values that (a) subject the system to limits that exceed the typical operational range or (b) additional tests that are
not considered part of the core program required for completion of the selected service. Because custom setpoints and limits may
exceed the operational envelope of the equipment, Agilent reserves the right to warrant conformance only to the closest variance
value. The user is notified of this stipulation at EQP setup time and the qualification report (EQR) will reflect this situation.
A set of ink signature fields, as determined by the creator of this document, can be included at the end of this document. All fields
should be completed or a single set of fields, initialed by an appropriate approver, run through any signature fields that are not to be
used. This is an optional process that allows a paper record of signoff by the appropriate reviewers where a hybrid (electronic/ink)
signature SOP is followed. If this document will be saved electronically and digitally signed in a document management system, it
should be generated without ink signature fields. Signing this document is for the customer's internal documentation purposes, to
assist with the applicable regulatory requirements. It is available for customer review, but it is not a pre-requisite for the delivery. The
delivery of the services is done according to the terms and conditions stated in the corresponding service exhibit. It is recommended
that after approval, this EQP be archived with the electronic EQP file.
For Agilent-recommended setpoints and limits, the range of allowable values (L for low, H for high) are included. As applicable,
variances, customizations, and additional setpoints are listed beneath the Agilent recommended values and marked W (within range)
or O (outside of range) in the left margin; values for added setpoint are also marked W or O and displayed after all configurations
values. Dual limits are marked DW or DO.
Optional tests that are enabled are included and marked as such; required tests that are disabled by the customer are included and
marked as such.
Customer Responsibilities
Agilent Responsibilities
o Deliver the services following the test programs described in the customer EQP
o Provide a locked and e-signed Qualification Report (EQR) upon completion of the service
o If requested, provide an optional ink-signed EQR CD to the customer
The recommended set of hardware OQ tests described in this EQP derives from Agilent's interpretation of authoritative expert
literature issued by the FDA, USP, GAMP5, ASTM 2500, and others. The OQ test design incorporates both modular and holistic
testing, which is a proven approach, acceptable to regulators. As prescribed by the 4Q qualification methodology for Analytical
Instrumentation Qualification (AIQ), the OQ step is separated from the PQ as recommended by the regulatory guidelines.
Enterprise Edition uses a balanced selection of metrology and chemical tests to directly determine the performance of the systems
without unnecessary reliance on inferred or derived results. For example, direct metrology is used to test pump flow rates and
thermal-controlled column compartment and autosampler modules. Holistic chemical testing is used for the evaluation of the following
critical instrument characteristics: linearity, precision, signal to noise, and carry over.
Enterprise Edition is designed to fit the traditional quality systems used by firms and recognized by regulatory agencies
worldwide.
Operational Qualification
Refer to the appropriate Test Definitions document for a detailed description of the testing program, setpoints, and acceptance
limits for each system technique, category, and instrument configuration.
Dual-Acceptance Limits
(Applies to hardware qualifications only)
Within the EQP of the Agilent Enterprise Edition, each of the tests final result can be compared against two different limits if
required. This allows customer-configured OQ to report against a User Limit (Limit 1) and the Agilent Recommended Limit
(Limit 2) simultaneously.
In the standard EQP documents, Limit 1 and 2 values are the same – effectively de-activating this feature. Custom EQPs can
also be prepared on request, making effective use of the two-limit feature of the Agilent Compliance Engine (ACE). In those
cases, Limit 2 will always be the Agilent Recommended limit, and Limit 1 will be the limit requested by the user.
Agilent will not be under any obligation regarding the OQ testing results against user-requested limits that are more stringent
than the Agilent Recommended ones.
The full list of repair and re-test guidance is available for review by customers of the RQ service.
The RQ form in ACE prescribes which tests the operator must perform for each repair circumstance. The test procedure,
setpoints, and limits will be an exact repeat of the previous OQ test (a regression-testing strategy).
www.agilent.com/chem/enterprise
Information, descriptions and specifications in this
publication are subject to change without notice.
Page 1 / 2
• Toluene/Hexane Resolution
• Stray Light - Sodium Iodide
• Stray Light – Sodium Nitrite
• Photometric Accuracy – Potassium Dichromate
• Baseline Flatness
• Photometric Noise
• Photometric Stability
7. Agilent 8453: added/changed optional test suite to include:
• Photometric Accuracy – NIST Glass Filters
• Photometric Reproducibility – Potassium Dichromate
8. Other unspecified models: changed optional test suite to include:
• Wavelength Accuracy – Mercury Line (as applicable)
• Wavelength Reproducibility (deuterium bands)
• Stray Light - Sodium Iodide
• Stray Light – Sodium Nitrite
• Photometric Accuracy – NIST Glass Filters
• Baseline Flatness
• Photometric Noise
• Photometric Stability
• Photometric Reproducibility – Potassium Dichromate
9. Cary models: changed optional test suite to include same as other unspecified plus:
• Maximum Resolution (as applicable)
• Photometric Reproducibility – NIST Glass Filters
10. Added support for Agilent 8454.
New forms: Photometric Reproducibility - Potassium Dichromate, Photometric Reproducibility - NIST Glass Filters
Changed forms: none
Page 2 / 2
UV-VISIBLE SYSTEMS
OPERATIONAL QUALIFICATION
OQ Test Suite
This document describes the test program for qualifying UV-Visible systems, and the following table lists all OQ tests.
In this document, generic models refer to unspecified UV Systems other than the Agilent Cary and 8453/8454. Consult your Agilent
representative for compatibility questions.
Note: Optional tests are NOT INCLUDED in the standard OQ but can be ordered as EXTRA COST TESTS. Select the check boxes
on the right and attach this document to your OQ EQP documentation for a record of qualification conditions.
Test 8453 / 8454 Cary 50/60 Cary 100/300 Cary 4000/ Generic
5000/6000i
Wavelength Accuracy – Source Line
Wavelength Accuracy – Holmium Oxide in Perchloric Acid
Wavelength Accuracy – Didymium
Wavelength Accuracy – Mercury Line N/A N/A N/A
Wavelength Reproducibility
Maximum Resolution N/A N/A N/A N/A
Toluene/Hexane Resolution
Stray Light – Potassium Chloride
Stray Light – Acetone
Stray Light – Sodium Iodide
Stray Light – Sodium Nitrite
Photometric Accuracy – Potassium Dichromate
Photometric Accuracy – NIST Glass Filters
Photometric Reproducibility – Potassium Dichromate
Photometric Reproducibility – NIST Glass Filters N/A N/A
Photometric Noise
Baseline Flatness
Photometric Stability
Description: Wavelength Accuracy is performed as a verification of the installed source lamp (either Xenon or Deuterium)
performance. The source lamp produces peak maxima at various wavelengths. In this test, Wavelength Accuracy is verified by
scanning a spectrum number of nanometers above and below the theoretical maxima for the installed source lamp. The test passes
if the reported wavelength maxima do not differ from the expected values by more than the accuracy limit. This test meets all major
pharmacopoeia and ASTM requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present,
disable the cell changer for testing. Verify there are no samples in the reference and sample cell holders. Using the
spectrophotometer’s controlling data software (CDS), setup the Wavelength Accuracy test for the Source Lamp Emission Line.
Perform the test by following the software prompts. Upon completion, record the values to the test record.
Models/Lamp Setpoints Limits
Default Allowable Range Default Allowable Range
8453, 8454/Deuterium 486.0 nm N/A ≤ 0.20 nm 0.20 – 2.0 nm
656.1 nm
Cary 50, 60/Xenon 541.92 nm ≤ 0.50 nm 0.50 – 2.0 nm
Cary 100, 300/Deuterium Zero order ≤ 0.70 nm 0.70 – 1.0 nm
486.0 nm ≤ 0.20 nm 0.20 – 2.0 nm
656.1 nm ≤ 0.20 nm 0.20 – 2.0 nm
Cary 4000/Deuterium Zero order ≤ 0.08 nm 0.08 – 1.0 nm
486.0 nm 0.08 – 2.0 nm
656.1 nm 0.08 – 2.0 nm
Cary 5000, 6000i/Deuterium Zero order ≤ 0.08 nm 0.08 – 1.0 nm
486.0 nm 0.08 – 2.0 nm
656.1 nm 0.08 – 2.0 nm
1312.2 nm ≤ 0.40 nm 0.40 – 2.0 nm
1968.3 nm 0.40 – 2.0 nm
Generic/Deuterium 486.0 nm ≤ 2 nm 0.1 – 2.0 nm
656.1 nm
Generic/Xenon 541.92 nm ≤ 2 nm 0.1 – 2.0 nm
Description: Wavelength Accuracy is calculated by comparing the reported wavelength values against the expected wavelength
values of a Holmium Oxide in Perchloric Acid test solution (Holmium Perchlorate). The test passes if the reported wavelength
values do not differ from the expected values by more than the accuracy limit. This test meets all major pharmacopoeia and ASTM
requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup the Wavelength
Accuracy test for the Holmium Oxide in Perchloric Acid (Holmium Perchlorate). Perform the test by following the software prompts.
Upon completion, record the values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
8453, 8454 241.13 nm 333.49 nm N/A ≤ 0.6 nm 0.6 – 1.0 nm
249.98 nm 345.38 nm
278.11 nm 361.18 nm
287.40 nm 385.70 nm
416.42 nm 485.26 nm ≤ 0.6 nm 0.6 – 2.0 nm
451.33 nm 536.71 nm
467.86 nm 640.62 nm
Cary 50, 60, 100, 300, 4000, 241.13 nm 333.49 nm N/A ≤ 1 nm 0.1 – 1.0 nm
5000, 6000i; generic 249.98 nm 345.38 nm
278.11 nm 361.18 nm
287.40 nm 385.70 nm
416.42 nm 485.26 nm ≤ 2 nm 0.1 – 2.0 nm
451.33 nm 536.71 nm
467.86 nm 640.62 nm
Wavelength Accuracy—Didymium
Optional for all models
Description: Wavelength Accuracy is calculated by comparing the reported wavelength values against the expected wavelength
values of a Didymium standard test solution. The test passes if the reported wavelength values do not differ from the expected
values by more than the accuracy limit. This test meets the United States Pharmacopoeia requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup a Wavelength Accuracy
test for the Didymium test solution. Perform the test by following the software prompts. Upon completion, record the values to the
test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
All 731.6 nm 799.0 nm N/A ≤ 2 nm 0.1 – 2.0 nm
740.0 nm 864.4 nm
794.1 nm
Description: Wavelength Accuracy is performed as a verification of the mercury lamp performance. The mercury lamp produces
photometric peak maxima at various wavelengths. In this test, Wavelength Accuracy is verified by scanning a wavelength range for
the mercury lamp. The test passes if the reported wavelength maxima do not differ from the expected values by more than the
accuracy limit. This test meets all major pharmacopeia requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Verify there are no samples in the reference and sample cell holders. Using the
spectrophotometer’s controlling data software (CDS), setup the Wavelength Accuracy test for the Mercury Lamp. Perform the test
by following the software prompts. Upon completion, record the values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
Cary 4000, 5000 , 6000i 253.65 nm 334.15 nm N/A ≤ 0.10 nm 0.10 – 1.0 nm
296.73 nm 365.02 nm
302.25 nm 365.48 nm
313.16 nm
404.66 nm 576.96 nm 0.10 –2.0 nm
435.83 nm 579.07 nm
546.08 nm
Generic 253.65 nm 334.15 nm ≤ 1 nm 0.1 – 1.0 nm
296.73 nm 365.02 nm
302.25 nm 365.48 nm
313.16 nm
404.66 nm 576.96 nm ≤ 2 nm 0.1 – 2.0 nm
435.83 nm 579.07 nm
546.08 nm
Wavelength Reproducibility
Standard for all models
Description: The Wavelength Reproducibility test measures the instrument’s reproducibility when making repeated readings of the
same wavelength. The reproducibility is defined as the deviation of 10 consecutive wavelength maximum readings taken using the
peak(s) of the reference–either the deuterium lamp or the Holmium Oxide in Perchloric Acid standard. The test passes if the
maximum deviation of the 10 wavelength maxima values is less than the test limit. This test meets the requirements of the United
States Pharmacopeia, NIST and ASTM. For the 8453 and 8454 models, this test is completed with the Wavelength Accuracy tests.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup the Wavelength
Reproducibility test using either the Holmium Oxide in Perchloric Acid (Holmium Perchlorate) or the deuterium lamp as the
reference. Perform the test by following the software prompts. Upon completion, record the values to the test record.
Models Setpoints (D for Deuterium; H for Holmium Oxide) Limits
Default Allowable Default Allowable Range
Range
8453, 8454 D 486.00 nm H 333.49 nm H 467.86 nm N/A -0.02 – 0.02 nm -0.04 – 0.04 nm
D 656.10 nm H 345.38 nm H 485.26 nm
H 241.13 nm H 361.18 nm H 536.71 nm
H 249.98 nm H 385.70 nm H 640.62 nm
H 278.11 nm H 416.42 nm H 640.62 nm
H 287.40 nm H 451.33 nm
Cary 50, 60 H 536.50 nm -0.1 – 0.1 nm -0.2 – 0.2 nm
Cary 100, 300; generic D 656.10 nm -0.08 – 0.08 nm -0.16 – 0.16 nm
Cary 4000, 5000, 6000i D 656.10 nm -0.025 – 0.025 nm -0.050 – 0.050 nm
Maximum Resolution
Optional for Cary 4000, 5000, 6000i models
Description: Maximum Resolution (often called the ‘limiting’ resolution) of the instrument is determined by calculating the width of
the 656.1 nm peak at half height that the instrument can achieve. The test passes if the calculated maximum resolution is less than
or equal to the test limit.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Verify there are no samples in the cell holders. Using the spectrophotometer’s controlling data
software (CDS), setup the Maximum Resolution. Perform the test by following the software prompts. Upon completion, record the
values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
Cary 4000, 5000 , 6000i N/A N/A 0.000 – 0.048 0.000 – 0.096
Toluene/Hexane Resolution
Standard for all models
Description: Decreased spectral resolution can impact the photo metric sensitivity and selectivity. Toluene/Hexane Resolution is
determined by calculating the peak-to-trough ratio between the reported absorbance maximum and minimum from a scan of 0.02%
v/v Toluene in Hexane. The absorbance maximum should occur at approximately 269 and the minimum at approximately 266 nm.
The test passes if the calculated spectral resolution ratio is greater than or equal to the test limit. This test meets all major
pharmacopeia requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present,
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), set up the Toluene/Hexane
Resolution test. Perform the test by following the software prompts. Upon completion, record the values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
All N/A N/A ≥ 1.6 1.6 – 2.5
Description: The Stray Light Test measures the presence of light outside the bandwidth of a selected wavelength that reaches the
detector. The presence of stray light can decrease the photometric selectivity, increase the photometric response and create non-
linear response of the instrument causing problems with quantitative analysis. Potassium Chloride will completely absorb light
below 200 nm, however is transparent above 200 nm. The test will pass if the Percent Transmittance is less than 1.0% at 198 nm.
This test meets all major pharmacopoeia and ASTM requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present,
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup the Stray Light–
Potassium Chloride test. Perform the test by following the software prompts. When the test is complete record the values to the
test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
All 198 nm 190 – 200 nm ≥ 0.0 and < 1.0%T N/A
Stray Light—Acetone
Optional for all models
Description: The Stray Light Test measures the presence of light outside the bandwidth of a selected wavelength that reaches the
detector. The presence of stray light can decrease the photometric selectivity, increase the photometric response, and create non-
linear response of the instrument causing problems with quantitative analysis. Acetone absorbs light below 320 nm and is
transparent above 326 nm. The test will pass if the Percent Transmittance is less than 1.0% at 300 nm. This test meets United
States Pharmacopoeia requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present,
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup the Stray Light test using
the Acetone standard solution. Perform the test by following the software prompts. When the test is complete record the values to
the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
All 300 nm 250 – 320 nm ≥ 0.0 and < 1.0%T N/A
Description: The Stray Light test measures the presence of light outside the bandwidth of a selected wavelength that reaches the
detector. The presence of stray light can decrease the photometric selectivity, increase the photometric response and create non-
linear response of the instrument causing problems with quantitative analysis.
This Stray Light test uses Sodium Iodide as the filter to measure the stray light. Sodium Iodide will begin to absorb light below 260
nm, however is transparent above 260 nm. The test will pass if the Percent Transmittance is less than the test limit. This test
meets ASTM requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present,
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup the Stray Light–NaI
(Sodium Iodide) at 220 nm test. Perform the test by following the software prompts. Upon completion, record the values to the test
record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
8453, 8454; Cary 50, 60; generic 220 nm 200 – 230 nm ≥ 0.00 and < 0.05 %T 0.05 – 1.0 %T
Cary 100 ≥ 0.00 and < 0.02 %T 0.02 – 1.0 %T
Cary 300 ≥ 0.00 and < 0.0005 %T 0.0005 – 1.0 %T
Cary 4000, 5000, 6000i ≥ 0.00 and < 0.00007 %T 0.00007 1.0 %T
Description: The Stray Light test measures the presence of light outside the bandwidth of a selected wavelength that reaches the
detector. The presence of stray light can decrease the photometric selectivity, increase the photometric response and create non-
linear response of the instrument causing problems with quantitative analysis.
This Stray Light test uses Sodium Nitrite as the filter to measure the stray light. Sodium Nitrite will begin to absorb light below 385
nm, however is transparent above 385 nm. The test will pass if the Percent Transmittance is less than the test limit. This test
meets ASTM requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup the Stray Light – NaNO2
(Sodium Nitrite) at 340 nm test. Perform the test by following the software prompts. Upon completion, record the values to the test
record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
8453, 8454 340 nm 340 – 390 nm ≥ 0.0 and < 0.03 %T 0.03 – 1.0 %T
Cary 50, 60; generic ≥ 0.0 and < 0.05 %T 0.05 – 1.0 %T
Cary 100 ≥ 0.0 and < 0.005 %T 0.005 – 1.0 %T
Cary 300 ≥ 0.0 and < 0.0002 %T 0.0002 – 1.0 %T
Cary 4000, 5000, 6000i ≥ 0.0 and < 0.00007 %T 0.00007 – 1.0 %T
Description: The Photometric Accuracy - Potassium Dichromate Test uses two strengths of the Potassium Dichromate standard.
First, the 60mg/L Potassium Dichromate standard is used to measure the difference between the expected absorbance values
(listed on the Certificate of Analysis) and the actual absorbance values reported by the instrument at four wavelengths: 235 nm,
257 nm, 313 nm, and 350 nm. The test passes if the expected absorbance values are within +/- 0.010 AU of the actual absorbance
values at the four wavelengths.
Then the 600 mg/L Potassium Dichromate standard is used to measure the difference between the expected absorbance value
(listed on the Certificate of Analysis) and the actual absorbance value reported by the instrument at 430 nm. The test passes if the
expected absorbance value is within+/- 0.012 AU of the actual absorbance value at 430 nm. This test meets all major
pharmacopoeia, ASTM, and NIST requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present,
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup the Photometric
Accuracy – Potassium Dichromate test for both the 60 mg/L and 600 mg/L standard concentrations. Perform the test by following
the software prompts. Upon completion, record the values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
All 0.747 Abs at 235 nm N/A ≤ 0.010 Abs N/A
0.867 Abs at 257 nm
0.292 Abs at 313 nm
0.644 Abs at 350 nm
0.954 Abs at 430 nm ≤ 0.012 Abs
Description: The Photometric Accuracy – NIST Glass Filters test uses certified glass filter standard reference materials to measure
the difference between the expected absorbance value (listed on the Certificate of Analysis) and the actual absorbance value
reported by the instrument at various wavelengths. The Test passes if the expected absorbance value is within the test limit of the
expected absorbance value. This test meets all major pharmacopoeia, ASTM, and NIST requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup the Photometric
Accuracy – NIST test using the glass filter standard reference material. Perform the test by following the software prompts. Upon
completion, record the values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
8453, 8454; generic 1 Abs at 440.0 nm 1 Abs at 590.0 nm N/A ≤ 0.0050 Abs 0.0050 – 0.0080
1 Abs at 465.0 nm 1 Abs at 635.0 nm
1 Abs at 546.1 nm
Cary 50, 60, 100, 300, 1 Abs at 440.0 nm 1 Abs at 590.0 nm ≤ 0.0050 Abs 0.0050 – 0.0080
4000, 5000, 6000i 1 Abs at 465.0 nm 1 Abs at 635.0 nm
1 Abs at 546.1 nm
0.7 Abs at 440.0 nm 0.7 Abs at 590.0 nm ≤ 0.0040 Abs 0.0040 – 0.0080
0.7 Abs at 465.0 nm 0.7 Abs at 635.0 nm
0.7 Abs at 546.1 nm
0.5 Abs at 440.0 nm 0.5 Abs at 590.0 nm ≤ 0.0030 Abs 0.0030 – 0.0080
0.5 Abs at 465.0 nm 0.5 Abs at 635.0 nm
0.5 Abs at 546.1 nm
Description: The Photometric Reproducibility - Potassium Dichromate test uses two strengths of the Potassium Dichromate
standard to find the reproducibility of the photometric measurement.
The 60 mg/L Potassium Dichromate standard is used to measure the absorbance at four wavelengths: 235nm, 257nm, 313nm, and
350nm. Ten measurements are taken at each respective wavelength. The standard deviation is then calculated for each respective
wavelength. The test passes if the calculated results of each measurement set are within the test limits.
The 600 mg/L Potassium Dichromate standard is used to measure the absorbance at 430 nm. Ten measurements are taken at each
respective wavelength. The standard deviation is then calculated for each respective wavelength. The test passes if the calculated
results of each measurement set are within the test limits.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup the Photometric
Accuracy – Potassium Dichromate test for both the 60 mg/L and 600 mg/L standard concentrations. Follow the software prompts
to complete the test. Perform the test 10 times to obtain values for reproducibility at each wavelength. Upon completion, record the
values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
All 235 nm 350 nm N/A Standard deviation: 0.000 – 0.005 Abs N/A
257 nm 430 nm
313 nm
Description: The Photometric Reproducibility—NIST Glass Filters test uses certified glass filter standard reference materials to find
the reproducibility of the photometric measurement.
The 30%T glass filter NIST standard is used to take ten measurements at the absorbance at the 546.1 nm wavelength. The
standard deviation is then calculated. The test passes if the calculated standard deviation is within the test limits.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Using the spectrophotometer’s controlling data software (CDS), setup the Photometric
Accuracy – NIST test for the 30%T standard. Follow the software prompts to complete the test. Perform the test 10 times to obtain
values for reproducibility at the 546.1 nm wavelength. Upon completion, record the values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
Cary 50, 60 0.5 Abs at 546.1 nm N/A Standard deviation: 0.000 – 0.005 Abs N/A
Cary 100, 300, 4000, Standard deviation: 0.000 – 0.005 Abs N/A
5000, 6000i
Photometric Noise
Standard for 8453/8454; optional for Cary and generic systems
Description: Photometric noise measures short-term fluctuations in the detector signal caused by inherent optical and electronic
noise. The amount of photometric noise will define the precision and minimum detection limit of the spectrophotometer. The test
passes if the measured photometric noise is within the test limits. This test’s limits conform to manufacturer requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Verify there are no samples in the cell holders. Using the spectrophotometer’s controlling data
software (CDS), setup the Photometric Noise test. Perform the test by following the software prompts. Upon completion, record the
values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
8453, 8454 0 Abs at 500 nm N/A < 0.00020 Abs, rms 0.000020 – 0.00200 Abs, rms
Cary 50, 60; generic < 0.00010 Abs, rms 0.00010 – 0.00100 Abs, rms
Cary 100 < 0.000085 Abs, rms 0.000085–0.000850 Abs,rms
Cary 300 < 0.00006 Abs, rms 0.00006 – 0.00060 Abs, rms
Cary 4000 < 0.00003 Abs, rms 0.00003 – 0.00030 Abs, rms
Cary 5000 0 Abs at 500 nm < 0.00003 Abs, rms 0.00003 – 0.00030 Abs, rms
0 Abs at 1500 nm
Cary 6000i 0 Abs at 500 nm < 0.00003 Abs, rms 0.00003 – 0.00030 Abs, rms
0 Abs at 1500 nm < 0.00002 Abs, rms 0.00002 – 0.00020 Abs, rms
Baseline Flatness
Standard for 8453/8454; optional for Cary and generic models
Description: Baseline Flatness measures the instrument’s ability to compensate for inherent optical and electronic noise across the
entire wavelength range. A baseline correction and test scan are performed across the entire wavelength range of the instrument.
The corrected signal is evaluated for peak-to-peak fluctuations. The test passes if the measured baseline stability is within test
limits. This test’s limits conform to manufacturer requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Verify there are no samples in the cell holders. Using the spectrophotometer’s controlling data
software (CDS), setup the Baseline Flatness test. Perform the test by following the software prompts. Upon completion, record the
values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
8453, 8454 0 Abs at 190 – 1100nm N/A -0.00100 – 0.00100 Abs -0.01000 – 0.01000 Abs
Cary 50, 60, 100, 300, g eneric 0 Abs at 200 – 850 nm -0.001 – 0.001 Abs -0.010 – 0.010 Abs
Cary 4000 0 Abs at 200 – 850 nm -0.0012 – 0.0012 Abs -0.0120 – 0.0120 Abs
N/A
Cary 5000 0 Abs at 200 – 3000 nm
Cary 6000i 0 Abs at 200 – 1700 nm
Photometric Stability
Standard for 8453/8454; optional for Cary and generic models
Description: Photometric Stability measures the photometric drift of the spectrophotometer when monitoring the Zero Absorbance
signal at a single wavelength for an hour. Photometric Stability is an indicator of how stable photometric results will be over an
extended period of time. The test passes if the measured drift of the photometric stability is within test limits. This test’s limits
conform to manufacturer requirements.
Procedure: Power on and warm up the spectrophotometer for the appropriate time prior to testing. If a cell changer is present
disable the cell changer for testing. Verify there are no samples in the cell holders. Using the spectrophotometer’s controlling data
software (CDS), setup the Photometric Stability test. Perform the test by following the software prompts. Upon completion, record
the values to the test record.
Models Setpoints Limits
Default Allowable Range Default Allowable Range
8453, 8454 0 Abs at 340 nm N/A -0.00100 – 0.00100 Abs/hr -0.01000 – 0.01000 Abs/hr
Cary 50, 60 0 Abs at 500 nm -0.0004 – 0.0004 Abs/hr -0.0040 – 0.0040 Abs/hr
Cary 100, 300 0 Abs at 500 nm -0.0003 – 0.0003 Abs/hr -0.0030 – 0.0030 Abs/hr
Cary 4000, 5000, 6000i 0 Abs at 500 nm -0.00018 – 0.00018 Abs/hr -0.00180 – 0.00180 Abs/hr
0 Abs at 200 – 3000 nm
0 Abs at 200 – 1700 nm
Generic 0 Abs at 340 nm -0.00100 – 0.00100 Abs/hr -0.01000 – 0.01000 Abs/hr
Customer Approval
Name:
Title:
Date:
Signature:
Name:
Title:
Date:
Signature:
Name:
Title:
Date:
Signature:
Name:
Title:
Date:
Signature:
Legal Notice
Enterprise Edition and its primary components (ACE software tool, procedures, test design, metrology tools, chemical reference
standards, and operator training materials) have been designed, tested, validated, and released for commercial use following Agilent's
Life-Cycle Development Quality Assurance methodology.
Agilent CrossLabs Group R&D VP and Dir. of Technology: Neil Cook. Santa Clara, California USA.
Agilent CrossLabs Group Quality Manager: Julio Hector. Santa Clara, California USA.
ACE software is patented. Copyright is claimed by this statement for all original work comprising Enterprise Edition. Any unauthorized
use, reproduction, or translation will be prosecuted to the maximum extent possible by law. All customer copies of EQP approval, final
qualification reports, and raw data provided to customer at delivery of the service become the property of the customer.
Protocol Details
Current Revision Current Revision Release Date Previous Revision Previous Revision Release Date