Rotapro: Rotational Atherectomy System Console
Rotapro: Rotational Atherectomy System Console
Rotapro: Rotational Atherectomy System Console
ROTAPRO™
Description.......................................................................................................................................................................................................... 3
Intended Use....................................................................................................................................................................................................... 3
How Supplied...................................................................................................................................................................................................... 3
Handling and Storage...................................................................................................................................................................................... 4
References........................................................................................................................................................................................................... 8
APPENDIX D - TROUBLESHOOTING................................................................................................................................................................. 12
Warranty........................................................................................................................................................................................................... 13
Intended Use
The ROTAPRO Console is intended for use with the ROTAPRO Rotational Atherectomy System. Refer to guidewire and advancer
package inserts for specific information on the use of these components.
Restrictions
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician. Rotational atherectomy should be
performed only by physicians trained in percutaneous interventional procedures.
Warnings
• Never use oxygen as the propellant for the ROTAPRO Rotational Atherectomy System. Never connect the regulator to an
oxygen cylinder. Oxygen combined with oil or other combustibles in the system can result in an explosion.
• The use of accessories, other than those specified, with the exception of those sold by the manufacturer of the ROTAPRO System
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
as replacement parts for internal components, may result in increased emissions or decreased immunity of the ROTAPRO System.
• The ROTAPRO System should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, the ROTAPRO System should be observed to verify normal operation in the configuration in which it will be used.
• This device is not to be used in the presence of flammable anesthetics.
• If a hissing noise is detected from the console, check to make sure that the pressure of the gas supplied to the air or
nitrogen connector does not exceed 758.4 kPa (110 psi). The console is equipped with a pressure relief valve to protect
against excessive inlet pressure. Do NOT operate the ROTAPRO Console with gas pressures in excess of 758.4 kPa (110 psi),
as a gas hose may burst.
• If patient defibrillation becomes necessary, the physician should take the appropriate measures to protect themselves
against electrocution from the defibrillator.
• Do not modify or repair beyond replacement of fuses as described in Service Information. Modification or repair of the
instrument by a person other than an authorized Boston Scientific Representative may compromise the integrity of the
device and/or lead to device failure which, in turn, may result in patient injury or death. Boston Scientific assumes no
liability with respect to any instrument which has been modified or repaired by a person other than an authorized Boston
Scientific Representative and makes no warranties, express or implied, including but not limited to merchantability or fitness
for a particular purpose, with respect to such instruments. If repair is needed, call your Boston Scientific Representative.
• Following any change in console set-up or configuration, the ROTAPRO System should be observed to verify normal
operation in the configuration in which it will be used.
• If the ROTAPRO System shows evidence of mechanical failure at any time prior to or during the Atherectomy procedure,
immediately discontinue use of the device and contact your Boston Scientific Representative for evaluation. Do NOT
attempt to use a damaged ROTAPRO System; use may result in device malfunction and/or patient injury.
Precautions
• Care must be taken not to spill saline or other fluids on the console. Saline spilled in the console may result in corrosion or
electrical hazard.
• User should take precautions when using the console in conjunction with other medical electrical equipment, as
electromagnetic interference between the equipment may affect the performance of the console or other devices. The
console complies with IEC 60601-1-2 regulations for radiation of and immunity from electromagnetic energy.
• The ROTAPRO Console requires special precautions regarding Electromagnetic Compatibility (EMC) and needs to be
installed and put into service according to the EMC information provided in Appendix D.
• Portable and mobile RF (Radio Frequency) communications equipment can affect the ROTAPRO Console.
How Supplied
Do not use if package is opened or damaged.
Do not use if labeling is incomplete or illegible.
Event Timer
Procedure Timer
The symbols used on the console are listed in Appendix A. The main features and functions of the console shown in Figure 1 are
described below.
FRONT PANEL/SCREEN DISPLAY
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
• Timer Reset Button: Pushing and releasing the reset button, located next to the Speed Control Knob, resets the event and
procedure timers both to 0:00.
• Advancer Gas Line Connector: The gas line connector receives the Advancer hybrid harness gas hose and supplies filtered,
regulated compressed gas to the advancer when rotation is activated via the advancer knob button or momentary switch.
• Advancer Fiber Optic Connector: The female connector receives the mating male duplex connector from the advancer hybrid
harness fiber optic tachometer cable. The orientation of the cable to the female connector is not important. The fiber optic
tachometer cable carries light pulses which the console uses to determine the rotational speed of the gas turbine and burr.
• Advancer Electrical Connector: The female connector supplies electrical power to the Advancer through the hybrid
harness multi-pin male connector. To connect to the console, orient the Advancer Power Connector with white chevrons
facing toward the front panel LCD display, or upward when console is vertically pole mounted.
• RPM Adjustment Knob: The knob located between the display and advancer connection ports is used to adjust the gas
pressure to the turbine and consequently, the rotational speed. Turning the knob clockwise increases the turbine pressure
(speed), and counterclockwise rotation decreases the turbine pressure (speed).
• Rotational Speed Display (Tachometer): The rotational speed is displayed on the screen in units of kRPM. See Table 2 for
additional information on operating range, accuracy, and precision.
• Event Timer: The event timer records the duration of rotation from activation to deactivation for each run. The event timer
automatically resets to 0:00 each time rotation is activated. See Table 2 for additional information on operating range,
accuracy, and precision.
• Procedure Timer: The procedure time is the sum of the individual event times and indicates the total cumulative time the
burr has been spinning in Normal Mode during the procedure. See Table 2 for additional information on operating range,
accuracy, and precision.
10+ kRPM A solid yellow triangle will appear when rotational speed drops more
Deceleration than 10 kRPM below the free lumen platform speed. Free lumen
platform speed is achieved approximately 1 second after each
instance rotation is activated or the speed control knob is adjusted.
Stall A red STALL indication appears and delivery of compressed gas to the
advancer discontinues when the rotational speed falls to 15 kRPM
or below for approximately 1/2 second. A STALL condition may also
be detected if the fiber optic connection is not properly engaged.
Stall detection is a safety feature designed to discontinue delivery of
compressed gas to the advancer in the event of excessive mechanical
loading or incorrect connection of the fiber optic. Releasing the
momentary button in Dynaglide™ Mode or activating rotation in Normal
Mode via the knob button will extinguish the STALL indication. Note:
When a STALL condition occurs, dashes will appear in the Rotational
Speed Display.
Note: Operating ranges apply to the display, refer to the Advancer Directions for Use for recommended treatment ranges.
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
Event Timer 00:00-59:59 ± 3.6 seconds 1 second
REAR PANEL
• Power Switch: Power is supplied to the console when the rocker-style switch is depressed in the (“line” symbol) position.
The rocker switch is located adjacent to the power line cord entry module. The FRONT PANEL display illuminates to indicate
that the power has been turned on.
• Power Cord: This cable plugs into a conventional 100-120 V~ or 220-240 V~ receptacle (as indicated on the name plate
located on the rear of the console) and provides power to the console. The ground wire of the power cord is internally
connected to the console chassis.
• Fuses: The fuses protect the console’s electrical components in the event of a serious electrical fault. If a fuse should fail,
refer to Appendix B for replacement instructions.
• Potential Equalization Connector: Located adjacent to the compressed gas inlet, the potential equalization connector is
provided to allow potential equalization between various hospital electrical instruments.
• Compressed Gas Inlet: This male connector mates to the corresponding connector on the supply line from the compressed
gas source. Pressure at this inlet should always be between 620.5 kPa and 758.4 kPa (90 psi and 110 psi) with a minimum
flow capacity of 140 l/min (5 standard cubic feet per minute [scfm]). Pressure will be reduced by the console to operating
limits. An internal pressure-relief valve protects against input pressures in excess of 792.9 kPa (115 psi) and creates a loud
hissing noise in the console when the pressure exceeds 792.9 kPa (115 psi).
Note: It may also be possible to operate this system from a hospital (house) gas system, as discussed in Appendix C.
Console Placement
The console attaches vertically to a standard five-wheel rolling intravenous (IV) pole or rests horizontally on a table or hard, flat
surface. To avoid risk of tipping, do not use the console in the vertical orientation without being secured to an IV pole.
Use the IV pole screw clamp to attach the console to a standard IV pole with five wheels and a 20 inch diameter base making sure
to attach the console to the IV pole at a distance not greater than 60 in (153.0 cm) from the floor to the top edge of the console.
ROTAPRO
Pole Mount Clamp
Console
Advancer Fiber
Optic Connector
Power
Cord
Saline Infusion
Catheter
Port
WARNING
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
NEVER use oxygen as the propellant for the ROTAPRO Rotational Atherectomy System. NEVER connect the regulator to an oxygen
cylinder. Oxygen combined with oil or other combustibles in the system can result in an explosion.
WARNING
This device is not to be used in the presence of flammable anesthetics.
7. Connect the power cord to a properly rated hospital grade grounded receptacle (as indicated on the label located on the
rear of the console). Do not position console to make it difficult to disconnect the console from the supply mains.
WARNING
To avoid risk of electric shock, this equipment must only be connected to a supply mains with a protective earth ground.
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
8. Open the compressed gas cylinder valve, or line valve if running on house air (see Appendix C), to supply compressed gas
to the console.
Note that the cylinder regulator gauge indicates the pressure of the gas remaining in the cylinder. The regulator should be
adjusted so that it never supplies more than 758.4 kPa (110 psi) to the console, and no less than 620.5 kPa (90 psi). Do not
initiate the procedure if less than 3,447 kPa (500 psi) of gas remains in the tank.
WARNING
If a hissing noise is detected from the console, check to make sure that the gas pressure supplied to the air or nitrogen connector
does not exceed 758.4 kPaz (110 psi). The console is equipped with a pressure relief valve to protect against excessive inlet
pressure. Do NOT operate the ROTAPRO™ Console with gas pressures in excess of 758.4 kPa (110 psi), as a compressed gas
hose may burst.
9. Splash
The ROTAPRO Console is designed to be placed outside of the sterile field. Care should be taken to protect the console from
splash and ingress of liquids which may cause damage to internal components.
10. Ventilation
The ROTAPRO Console uses natural convection cooling to maintain the proper operating temperature for internal
components. Cooling vent is located on the rear of the console enclosure. In order to ensure proper ventilation of the
console, it must be mounted on an IV pole (per instructions), or placed on a hard, flat surface with a minimum of 2.5 cm (1 in)
clearance maintained at the rear of the enclosure. Do not set on drapes or bedding.
11. Push the console power switch and confirm that the screen illuminates indicating power is on.
The ROTAPRO Console is now ready for use.
12. Disconnecting cables.
Do not use excessive force when disconnecting the advancer harness or power cord.
Caution:
• To avoid the potential for electric shock, turn the power switch off and disconnect the power cord from the hospital’s
electrical outlet before cleaning and disinfecting the console. The power cord and gas supply hose may remain connected
to the back of the console.
• To avoid damage to the console:
- Do not allow fluids to enter the enclosure or any connectors.
- Do not immerse the console enclosure into fluids.
- Do not use solvents or abrasive cleaners.
- Do not spray directly on to the console.
The console has been tested and demonstrated to be compatible with the following cleaning and disinfection solutions:
• 70% Isopropyl Alcohol (IPA)
• Bleach (0.5% to 1% Sodium Hypochlorite)
• Super Sani-Cloth™ Germicidal Wipes
Cleaning Instructions
1. Use Super Sani-Cloth Germicidal Wipes, or a pre-saturated wipe containing one of the following:
a. 70% IPA
b. Bleach (0.5% to 1% Sodium Hypochlorite)
Or, prepare a clean wipe by soaking with solution (a or b above). Wring excess solution ensuring the wipe remains
saturated but not dripping.
2. Wipe each panel (front, sides, top and bottom) of the console for a minimum of one minute until all visible soil is removed.
• While wiping, give particular attention to crevices and hard to clean areas.
• Replace soiled wipes as needed and use additional wipes to ensure that all surfaces are uniformly cleaned.
3. Use a clean, lint-free cloth/wipe dampened with water to wipe all accessible surfaces of the device a minimum of two times
to remove residual solution. Use additional clean, dampened cloths/wipes as needed.
4. Thoroughly dry all surfaces of the console with a clean, lint-free cloth/wipe.
5. Repeat steps 1 through 4 once. Use a soft-bristled brush, as necessary.
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
6. Visually inspect the console for the absence or presence of remaining soil in a well-lit area with an unaided eye. While
inspecting, give particular attention to verifying soil has been removed from the
hard-to-clean areas. If soil is present, then repeat the cleaning steps until all visible soil is removed.
Disinfection Instructions
1. Use Super Sani-Cloth Germicidal Wipes, or a pre-saturated wipe containing one of the following:
a. 70% IPA
b. Bleach (0.5% to 1% Sodium Hypochlorite)
Or, prepare a clean wipe by soaking with solution (a or b above). Wring excess solution ensuring the gauze/wipe remains
saturated but not dripping.
2. Thoroughly wipe each panel (front, sides, top and bottom) of the console and allow the surfaces to remain visibly wet for a
minimum of two minutes if using Super Sani-Cloth Germicidal Wipes (minimum of ten minutes for other wipes).
• Thoroughly wipe crevices and hard to disinfect areas.
• Use additional wipes as necessary to ensure surfaces remain wet.
3. Use a clean, lint-free cloth/wipe dampened with water to wipe all accessible surfaces of the device a minimum of two times
to remove residual solution. Use additional clean, dampened, lint-free cloths/wipes as needed.
4. Thoroughly dry all surfaces of the console with a clean, lint-free cloth/wipe.
References
For coronary use, please refer to the ROTAPRO Rotational Atherectomy System physician training program course materials for
a listing of publications, or contact your Boston Scientific Representative.
ROTAPRO Rotational
Device Identification Contents
Atherectomy System
00 %
Type CF Applied Part Humidity Limitation
00 %
0000 °C
Power Off 0000 °C
Temperature Limitation
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
2X Fuse Do not use if package is damaged.
T 1A 250X
Follow Instructions
ARG Argentina Local Contact
for Use
UL Certification
TECHNICAL SPECIFICATIONS
Frequency: 50 Hz - 60 Hz
Operating Conditions: Ventilation: Minimum 2.5 cm (1 in) clearance on all sides, with console
on a hard flat surface.
QUARTERLY INSPECTIONS
These inspections should be performed once per quarter by an operator familiar with the ROTAPRO™ System. If console fails any
of the inspections listed below, contact your Boston Scientific Representative.
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
1. Check the physical condition of the power cable, strain relief, and plug.
Make sure that there are no frayed ends on the power cable connector, the strain relief is attached, the plug has no
bent prongs, and the ground prong in the plug is present and secure. Verify that there are no unacceptable scratches or
punctures on the exterior of the AC cord set.
2. Check for loose or missing screws, sharp edges, or loose connectors.
By hand, check the console pneumatic connectors and verify that they are not loose. Verify that all screws are in place on
pole clamp and seam between front and rear panels. With console disconnected from power cord, gas supply hose, and IV
pole, hold console in both hands and gently shake in two different directions to verify that there is no loose hardware inside
the console.
3. Perform an external visual inspection for mechanical damage.
Verify that there are no unacceptable chips, dents, scratches, or marks on the console enclosure. Verify that the knob is
secured properly and not loose. Make sure that the pneumatic connectors are in working order, not stuck open, and are
tight against the console.
4. Check the operation of the CHECK PRESSURE indication feature.
With the gas supply hose disconnected, verify the CHECK PRESSURE indication appears on the screen. Supply gas pressure
in the range of 620.5 to 758.4 kPa (90 to 110 psi), and verify the CHECK PRESSURE indicator disappears from the screen.
IEC 62353 RECURRENT TESTING
National regulations may require that the user, manufacturer, or manufacturer representative periodically perform and document
testing of the device. If such testing is required in your country, follow the testing interval and extent of testing as regulated in
your country. If you do not know the national regulations in your country, please contact your Boston Scientific Representative.
If IEC 62353 is a required standard by national regulations but no specific testing or interval is specified, it is recommended to
perform testing using the direct method as specified in IEC 62353 at an interval of every 24 months or as per local regulations.
IEC 62353 Test Limit
Protective Earth Resistance < 300mΩ (when measured with power cord attached)
Earth Leakage Current < 5 mA
Touch Current < 100 μA
If the ROTAPRO Console exceeds the limits described in table above, immediately discontinue use of the console and contact
your Boston Scientific Representative.
SERVICE INFORMATION
Fuse: In the event of a fuse failure, replacement fuses should be ordered from Boston Scientific. Turn off the power and unplug
the line cord from the power outlet. The power cord may also be unplugged from the rear of the console to improve fuse access.
Using a tool such as a screwdriver, remove the fuse drawer by depressing both locking tabs. See Figure 4.
10
Fuse Drawer
Power Cord
Receptacle
Replace both fuses at the same time by inserting replacement fuses into the drawer. Reinsert the fuse drawer until the locking
tabs snap into place. Reconnect the power cord and restore power to the console. If a fuse fails again, disconnect all power to
the console and contact your Boston Scientific Representative.
All other service must be performed by Boston Scientific Corporation personnel. Field repair, other than the console’s external
fuse replacement, voids all warranties and may not be performed without express authorization from Customer Service.
CONTACT INFORMATION
The Boston Scientific phone numbers listed below are not intended to assist with any medical emergencies. Use these contacts
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
only for assistance for technical issues involving the operation of the ROTAPRO™ Console.
Before contacting the Boston Scientific Technical Service Department:
• Obtain the ROTAPRO Console serial number
• Obtain details of the event or problem
Next, call, email, or fax the Technical Assistance Center according to your geographic region.
Region Telephone Number Email
AMEA (Asia, Middle East, Africa) +65-6418-8878 [email protected]
Europe (see below for individual countries) +31-45-546-7707 [email protected]
France +33-139-304-971 [email protected]
Italy +39-022-698-3218 [email protected]
Spain +34-917-619-999 [email protected]
Germany +49-815-126-86118 [email protected]
United Kingdom +44-1442-411-686 [email protected]
Japan 81-44-287-7660 [email protected]
United States 1-800-949-6708 [email protected]
Argentina (+54)1157772692 [email protected]
Brazil (+55)1155459063 [email protected]
Colombia (+57)16295045 [email protected]
México (+52)15559924100 [email protected]
11
The ROTAPRO System is intended for use in the electromagnetic environment specified below. The customer or the user of
™
RF emissions Class A The ROTAPRO System is suitable for use in all establishments
CISPR 11 other than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used
Harmonic emissions Class A for domestic purposes.
IEC 61000-3-2
IEC 61000-3-3
The ROTAPRO System is intended for use in the electromagnetic environment specified below. The customer or the user of
the ROTAPRO System should assure that it is used in such an environment.
12
Radiated RF 3 V/m
IEC 61000-4-3 80 MHz to 2.7 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
EMC WARNINGS
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper
operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are
operating normally.
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
WARNING
Use only those accessories intended to be used with the ROTAPRO System.
WARNING
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
WARNING
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used
no closer than 30 cm (12 inches) to any part of the ROTAPRO System, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Warranty
Boston Scientific Corporation (BSC) warrants that reasonable care has been used in the design and manufacture of this
instrument. This warranty is in lieu of and excludes all other warranties not expressly set forth herein, whether express or
implied by operation of law or otherwise, including, but not limited to, any implied warranties of merchantability or fitness
for a particular purpose. Handling, storage, cleaning and sterilization of this instrument as well as other factors relating to the
patient, diagnosis, treatment, surgical procedures and other matters beyond BSC’s control directly affect the instrument and
the results obtained from its use. BSC’s obligation under this warranty is limited to the repair or replacement of this instrument
and BSC shall not be liable for any incidental or consequential loss, damage or expense directly or indirectly arising from the
use of this instrument. BSC neither assumes, nor authorizes any other person to assume for it, any other or additional liability or
responsibility in connection with this instrument.
13
Turkey
TUR
Local Contact
Boston Scientific Tıp Gereçleri Ltd. Şti.
Küçük Çamlıca Mahallesi
Oymak Çıkmazı Sok. No:1 Üsküdar 34696, İstanbul, Türkiye
Tel +90 216 544 47 00
Faks +90 216 544 47 01
EU Authorized
EC REP
Representative
Boston Scientific Limited
Ballybrit Business Park
Galway
IRELAND
Australian
AUS
Sponsor Address
Boston Scientific (Spiral Bound Manual Template, 5.5in x 8.5in Global), DFU, MB, ROTAPRO Console, Global, 50480266-01B_pretrans
Argentina
ARG
Local Contact
Para obtener información de
contacto de Boston Scientific
Argentina SA, por favor, acceda al
link www.bostonscientific.com/arg
Legal
Manufacturer
Boston Scientific Corporation
300 Boston Scientific Way
Marlborough, MA 01752
USA
USA Customer Service 888-272-1001
Recyclable
Package
0344
© 2018 Boston Scientific Corporation or 2018-01
its affiliates. All rights reserved. 50480266-01