OMC-116 Installation Manual: Observator
OMC-116 Installation Manual: Observator
instruments
TDC-HB
OBSERVATOR OMC-116
instruments Installation manual
Revision History
Introduction
This manual describes the installation of the OMC-116 ultrasonic wind sensor.
Installation must be carried out by a qualified engineer. This manual assumes the engineer
has basic knowledge of electronics. We recommend reading this manual before installation
of the sensor.
The OMC-116 can’t be directly powered by an Obsermet display (like OMC-138/139) or data
collection unit (like the OMC-183/184). A separate power supply is required:
Recommended cabling:
With heating (24Vdc, 10A): 2x2.5mm2 ( power, max 30m)* + 2x0.75mm2 shielded
*The heating will not be operated if the voltage is below 20V. Especially
when long cables are used check that the losses over the cable are
acceptable and the voltage at the sensor is well above 20V. Alternatively
the power supply unit must be placed closer to the sensor and/or thicker
cable or a power supply with a higher voltage must be used (voltage at
the sensor may not exceed 28V).
Terminals
1. Ground (power)
2. Power (+12 .. 28 VDC)
3. NMEA 0183 B (RS422/485)
4. NMEA 0183 A (RS422/485)
5. Analogue ground
6. Analogue Wind Speed (optional)
7. Analogue ground
8. Analogue Wind direction (optional)
1. Ground (power)
2. Power (+12 .. 28 VDC)
3. NMEA 0183 B (RS422/485)
4. NMEA 0183 A (RS422/485)
5. Analogue ground Wind Speed & Direction
6. Analogue Wind Speed (optional)
7. Control
8. Analogue Wind direction (optional)
Sensor installation
Bracket & U-bolt must make contact with metal of the mast to lead any static
discharge to ground. If this is not possible due to coating etc. a proper earth lead
must be connected to 1 of the U-bolts.
Cable connection
The shield of the cable(s) may not be laid to ground on the receiving (display /
signal conditioning) side of the cable but must be connected on sensor side.
The junction box does have an extra (plugged) M20 hole in case separate cables are used
for signal and power.
Alignment
The sensor must be lined up to North (fixed applications) or for ship applications to the bow
of the vessel.
1. Arrow on top
2. Small arrow on sensor deck
3. Drilling hole visible from below
4. Front junction box
NMEA output
The OMC-116 sensor outputs a MWV message once a second at 4800 baud:
$IIMWV,xxx.x,R,xxx.x,M,A*xx(CR)(LF)
R, Relative
M, m/s
CR Carriage Return
LF Line Feed
Specifications
Storage conditions
Permissible storage temperature: -55°C ... +80°C
Permissible relative humidity: 0 ... 95% RH Non-condensing
Operating conditions
Permissible operating temperature: -40°C ... +60°C (with heating)
Permissible operating temperature: -20°C ... +60°C (without heating)
Permissible relative humidity: 0 ... 100% RH
Dimensions
OMC-116-M
Disposal
Within the EC
The device must be disposed of in accordance with European Directives 2002/96/EC
and 2003/108/EC (waste electrical and electronic equipment). Waste equipment must
not be disposed of as household waste! For environmentally sound recycling and the
disposal of your waste equipment please contact a certified electronic waste disposal
company.
Outside the EC
Please comply with the applicable regulations for the proper disposal of waste
electrical and electronic equipment in your respective country.
Please arrange for any faulty equipment to be checked and, if necessary, repaired by the
manufacturer exclusively. Do not open the equipment and do not under any circumstances
attempt to carry out your own repairs.
Contact your local dealer or Observator instruments: