Hardness, Calcium: Titration Method With Edta Method 8222 0-25,000 MG/L As Caco Buret Titration
Hardness, Calcium: Titration Method With Edta Method 8222 0-25,000 MG/L As Caco Buret Titration
Hardness, Calcium: Titration Method With Edta Method 8222 0-25,000 MG/L As Caco Buret Titration
01157
Test preparation
Before starting
Magnesium is not included in the results but must be in the sample for a sharp endpoint. If the sample does not contain
magnesium, add 1 to 2 drops of Magnesium Standard Solution, 10-g/L as CaCO3, to the sample before the test is started.
As an alternative to the CalVer 2 Calcium Indicator Powder Pillow, use a 0.1-g scoop of CalVer 2 Calcium Indicator Powder.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description Quantity
Sample collection
• Collect samples in clean glass or plastic bottles that have been cleaned with a
detergent and rinsed with 1:1 nitric acid and deionized water.
• To preserve samples for later analysis, adjust the sample pH to 2 or less with
concentrated nitric acid (about 2 mL per liter). No acid addition is necessary if the
sample is tested immediately.
• Keep the preserved samples at room temperature for a maximum of 6 months.
• Before analysis, adjust the pH to 7 with Potassium Hydroxide Standard Solution.
• Correct the test result for the dilution caused by the volume additions.
1
Test procedure
1. Select a sample volume 2. Fill a 25-mL buret to the 3. Use a graduated cylinder 4. Pour the sample into a
and titrant from Table 1 zero mark with the titrant. or a pipet1 to measure the clean, 250-mL Erlenmeyer
on page 3. sample volume from Table 1 flask.
on page 3.
5. If the sample volume is 6. Add 1 mL of 8 N 7. Swirl to mix. 8. Add the contents of one
less than 50 mL, dilute to Potassium Hydroxide CalVer 2 Calcium Indicator
approximately 50 mL with Standard Solution. Powder Pillow.
deionized water.
9. Swirl to mix. 10. Put the flask under the 11. Use the multiplier in
buret. Swirl the flask. Add Table 1 on page 3 to
titrant until the color calculate the concentration.
changes from red to pure mL of titrant × multiplier =
blue. mg/L Ca as CaCO3.
1 Titration accuracy has a direct relation to the accuracy of the sample volume measurement. For smaller
volumes, it is recommended to use a pipet to increase accuracy.
Conversions units
To change the units or chemical form of the test result, multiply the test result by the
factor in Table 2.
Table 2 Conversions
mg/L Ca as CaCO3 to... multiply by... Example
mg/L as Ca 0.40 1000 mg/L as CaCO3 x 0.40 = 400 mg/L Ca
German degrees hardness (Gdh) 0.056 1000 mg/L as CaCO3 × 0.056 = 56 Gdh
Grains per gallon (gpg) 0.058 1000 mg/L as CaCO3 x 0.058 = 58 gpg
Interferences
WARNING
Chemical hazard. Potassium cyanide is toxic. Make sure to add potassium cyanide to the
sample after the 8 N Potassium Hydroxide Standard Solution has been added. Keep
cyanide solutions at more than pH 11 to prevent exposure to hydrogen cyanide gas.
Dispose of reacted solutions according to local, state and federal regulations.
An interfering substance can prevent the color change at the titration endpoint. A smaller
sample volume can often dilute the interfering substance to a level at which the
substance does not interfere. Table 3 shows the substances that can interfere with this
test.
Table 3 Interferences
Interfering substance Interference level
Acidity 10,000 mg/L acidity as CaCO3 does not interfere.
Alkalinity 10,000 mg/L alkalinity as CaCO3 does not interfere.
Aluminum Causes a slow endpoint. The sample can contain a maximum of 200 mg/L aluminum if sufficient
time is given for the color change.
Barium Barium is titrated at the same time with calcium and interferes with this test, but it is unusual to find
high levels of Barium in natural waters.
Chloride The chloride level in seawater does not interfere. Solutions that are saturated with chloride do not
show a sharp endpoint.
Cobalt Interferes directly. Add 0.5 grams of potassium cyanide after the 8 N Potassium Hydroxide
Standard Solution during the test procedure to remove the interference from a maximum of
20 mg/L cobalt.
Accuracy check
Standard additions method (sample spike)—0.020 N titrant
Use the standard additions method to validate the test procedure, reagents, apparatus,
technique and to find if there is an interference in the sample.
Items to collect:
• Calcium Hardness Standard Solution, 10,000 mg/L as CaCO3, 10-mL Voluette
ampule
• Ampule Breaker
• Pipet, TenSette, 0.1–1.0 mL and pipet tips
1. Use the test procedure to measure the concentration of the standard solution. Use
the 0.020 N titrant.
2. Use a TenSette pipet to add 0.1 mL of the standard solution to the titrated sample.
3. Titrate the spiked sample to the endpoint. Record the mL of titrant added.
4. Add one more 0.1-mL addition of the standard solution to the titrated sample.
5. Titrate the spiked sample to the endpoint. Record the mL of titrant added.
6. Add one more 0.1-mL addition of the standard solution to the titrated sample.
7. Titrate the spiked sample to the endpoint. Record the mL of titrant added.
8. Compare the actual result to the correct result. The correct result for this titration is
1.0 mL of titrant for each 0.1-mL addition of the standard solution. If much more or
less titrant was used, there can be a problem with user technique, reagents,
apparatus or an interference.
1. Use the test procedure to measure the concentration of the standard solution. Use
the 0.200 N titrant.
2. Use a TenSette pipet to add 1.0 mL of the standard solution to the titrated sample.
3. Titrate the spiked sample to the endpoint. Record the mL of titrant added.
4. Add one more 1.0-mL addition of the standard solution to the titrated sample.
5. Titrate the spiked sample to the endpoint. Record the mL of titrant added.
6. Add one more 1.0-mL addition of the standard solution to the titrated sample.
7. Titrate the spiked sample to the endpoint. Record the mL of titrant added.
8. Compare the actual result to the correct result. The correct result for this titration is
1.0 mL of titrant for each 1.0-mL addition of the standard solution. If much more or
less titrant was used, there can be a problem with user technique, reagents,
apparatus or an interference.
1. Use the test procedure to measure the concentration of the standard solution. Use
25.0 mL of the standard solution and the 0.020 N titrant.
2. Compare the actual result to the correct result. The correct result for this titration is
25 mL of titrant.
1. Use the test procedure to measure the concentration of the standard solution. Use
10.0 mL of the standard solution and the 0.200 N titrant.
2. Compare the actual result to the correct result. The correct result for this titration is
10 mL of titrant.
Summary of method
Potassium hydroxide is added to the sample to adjust the pH to 12 to 13, which causes a
magnesium hydroxide precipitate to form. CalVer 2 Calcium Indicator is then added,
which reacts with calcium to give a red color. The EDTA titrant is added, which reacts
with all the free calcium. After the EDTA has reacted with all of the free calcium ions, the
Required apparatus
Recommended standards
© Hach Company/Hach Lange GmbH, 2007–2016. All rights reserved. 03/2016, Edition 9