WelchAllyn Propaq Encore Vital Signs Monitor - Reference Guide
WelchAllyn Propaq Encore Vital Signs Monitor - Reference Guide
Reference Guide
0123
Contents
General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Propaq Encore Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
The Documentation Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
About This Reference Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Statement of Expectations of the Reader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Getting Started. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Introducing the Propaq Encore. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Propaq Encore Models and Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Expansion Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Propaq Encore Pulse Oximetry Option (SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Capnography (CO2) Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Impedance Pneumography (RESP) Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Printer Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Propaq-to-Acuity Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Modem-Propaq Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
HP-compatible Side Panel Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Using the Propaq Encore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
System Controls (Right Side Panel) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Alarm Lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Power-up Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Patient Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Propaq Encore Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Propaq Encore Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Propaq Encore Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Monitor Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Selecting Waveforms for Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Setting the Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Changing the Date Format, Filter, and Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Setting the Current, Custom, and Powerup Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Printer Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Learning the Propaq Encore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Using In-Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Patient Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
ECG/RESP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Intended Use–Impedance Pneumography (RESP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Intended Use–ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
ECG Connector and Applicable Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
How ECG/RESP is Displayed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Using the Propaq Encore With Pacemaker Patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Using the Filter to Better Display a Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
ECG Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
RESP Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
Invasive Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
Invasive Pressure Connectors and Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
Rezeroing a Transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
How Invasive Pressure is Displayed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Invasive Pressure Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
Improve NIBP Accuracy with Smartcuf™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
NIBP Connector and Cuffs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
Important Information About Automatic Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
NIBP Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Temperature Connectors and Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
How Temperature is Displayed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Temperature Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Pulse Oximetry (SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
Perform SpO2 Monitoring with Masimo Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Perform SpO2 Monitoring with Nellcor Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Perform SpO2 “Spot-Check” Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Capnography (CO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
Trends. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
The Trend Status Window and Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
How Trends are Accumulated . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
NIBP Trends. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Displaying Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Selecting a Trend (NXT TRND) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Printing Patient Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Printing Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .107
Printing NIBP Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .109
Printing the Apnea Ticket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .110
Printing When a Patient Alarm Occurs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111
OxyCRG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
Printing Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
Printing a Single Trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
Printing Several Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
Automatic Trend Prints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
ECG. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Real-Time ECG Analog/Defib Sync . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Impedance Pneumography (RESP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Invasive Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Pulse Oximetry (SpO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
Capnography (CO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Monitor (Environmental) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
Monitor (Physical) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Power Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
In-Service Simulated Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171
Safety Summary
This Safety Summary should be read by all Propaq Encore users. Specific warnings and
cautions will be found throughout the Propaq Encore documentation where they apply.
Place the Propaq monitor and accessories in locations where they cannot
harm the patient if they fall from their shelf or mount. Lift the monitor only by its
handle; do not lift it by any attached cables.
Do not connect more than one patient to a monitor. Do not connect more than
one monitor to a patient.
Inspect the power adapter cord periodically for fraying or other damage, and
replace the adapter as needed. Do not operate the apparatus from mains
power with a damaged power adapter cord or plug.
When using a power adapter with this monitor, be sure to connect the power
adapter to a three-wire, grounded, hospital-grade receptacle. Do not under
any circumstances attempt to remove the grounding conductor from the power
plug of the power adapter. Do not plug the power adapter into an extension
cord. If there is any doubt about the integrity of the protective earth ground of
the receptacle for the power adapter, do not plug in the power adapter;
operate the monitor only on battery power. Contact your biomedical
engineering department for assistance in identifying the proper power
receptacle and making appropriate power connections.
Warning Make frequent electrical and visual checks on cables and electrode
wires.
During defibrillation, keep the discharge paddles away from ECG and other
electrodes, as well as other conductive parts in contact with the patient. Avoid
contact with any accessories connected to the Propaq’s left side panel.
Warning Safe interconnection between the Propaq monitor and other devices
must comply with applicable medical systems safety standards such as IEC
60601-1-1. Within certain governmental jurisdictions, all interconnected
accessory equipment must be labeled by an approved testing laboratory. After
interconnection with accessory equipment, risk (leakage) current and
grounding requirements must be maintained.
As with all medical equipment, carefully route the patient cabling to reduce the
possibility of patient entanglement or strangulation.
The Propaq Encore should be serviced only by a Welch Allyn service technician while under
warranty. The Propaq Encore Service Manual (P/N 810-0696-XX) is available from Welch Allyn
to assist the biomedical engineer during post-warranty period service.
Symbols
The following symbols may appear on the Propaq Encore monitor or accessories or
documentation. These internationally recognized symbols are defined by the International
Electrotechnical Commission, IEC 878 and IEC 417A.
For continued fire Patient connections are Type For indoor use only (on
protection, use only BF, and protected against power adapter only)
the specified fuse. defibrillation.
Direct current Patient connections are Type Caution: Refer to
B. Reference Guide and
accompanying
documentation.
Battery charging when IPX1 Enclosure Protection Drip The CE Mark signifies
green indicator proof: Classification IPX1 per the device has met all
illuminated IEC Publication 529 essential requirements
The CE Mark and The Canadian of European Medical
0123 Notified Body NRTL/C StandardsAssociation has Device Directive
Registration Number
Evaluated to CSA 601-1
and UL2601-1
evaluated this device 93/42/EEC for a Class
signify the device has according to CSA 601-1 and 1 product. (This symbol
met all essential Underwriters Laboratory is on the Universal
requirements of Standard UL 2601-1. (This Power Adapter.)
European Medical symbol is on the Universal
Device Directive Power Adapter.)
93/42/EEC. Input port Temperature sensor
input
Two way Output port Single-use only (not
communication port 2 reusable).
Apply the NIBP cuff as
shown.
NIBP cuff sizes:
Thigh
Large adult
Adult
Small adult
Child
Infant
This Propaq Encore Reference Guide contains important safety and operating information for
the clinician.
The Propaq Encore Service Manual (P/N 810-0696-XX) contains information on how to
properly maintain the Propaq Encore through routine calibration, inspection, and
maintenance.
The Welch Allyn Products & Accessories booklet (P/N 810-0409-XX) provides a
comprehensive list of accessories recommended for Propaq Encore monitors and options.
Intended Use
Before using the Propaq Encore on a patient, be sure you understand the Safety Summary at
the front of this book. It provides important information about safely using the Propaq
Encore. The Propaq Encore monitor is intended to be used by skilled clinicians for
multiparameter vital signs monitoring of neonatal, pediatric, and adult patients in health
care facility bedside applications; as well as for intra- and interfacility transport.
Expansion Module
The Propaq Encore Expansion Module attaches to the monitor and houses additional
capabilities. The Expansion Module can be fitted with the SpO2, CO2, and Printer options.
ALARM(S)
OFF Light
ALARM Light
SpO2 Option
Mainstream Expansion
CO2 Option Module
Printer Option
Sidestream
CO2 Option
Printer Option
The Expansion Module with Printer (EMP) provides a lightweight 3-channel recorder.
Propaq-to-Acuity Option
This option allows communication between the Propaq Encore and the Acuity Central
Monitoring System by means of an ethernet network system installed in your facility. The
Acuity System operator can view the patient data and control most of the bedside Propaq
functions. The Propaq Encore connects to the Acuity System through an Acuity network
cable that plugs into the Propaq right side panel.
Modem-Propaq Option
This option allows telecommunication between a Propaq Encore and the Acuity System by
means of external modems. This option is configured with the Propaq-to-Acuity option. For
more information refer to the Modem-Propaq Reference Guide.
Speaker
Input Fuse
3A
EKG x 1000
2AG Battery Charging
! Light
12-28V, 3A
Warning Safe interconnection between the Propaq Encore and other devices
must comply with applicable medical systems safety standards such as IEC
601-1-1.
On/Off Switch
This switch turns the monitor on and off. The switch is recessed to prevent accidentally
turning off the monitor, which would result in losing patient data.
Input Fuse
The input fuse, which protects the Propaq Encore against power surges, is a 3 Ampere fuse,
externally replaceable by qualified service personnel. See page 124 for fuse replacement
instructions.
Alarm Lights
Alarms and limits are described in detail beginning on page 91.
ALARM Light
When an alarm limit is violated, the red ALARM light turns on.
Power-up Screen
Warning Before you use a Propaq on a new patient, always turn off the
Propaq for a few seconds, then turn it on again. This clears the prior patient’s
trend values, alarm limit settings, and NIBP cuff inflation target.
1. If the Propaq Encore has been used for a previous patient, switch the monitor off,
then on again. The monitor will turn on in the powerup patient mode with the
associated settings.
Verify that the powerup tone is produced. If the monitor has SpO2, verify two
tones are produced to make sure that both speakers are working.
2. Verify the monitor is in the correct patient mode according to the patient’s age. If the
patient mode is not correct, change it (see page 29 to change the patient mode).
Warning Always check the patient mode when monitoring a new patient. The
patient mode determines default alarm limits, maximum cuff inflation pressure,
and internal algorithm settings.
3. Verify the battery voltage is sufficient for monitoring. If it is less than 7.4 V, connect to
a power adapter (see page 119 for information about the power adapter).
If this occurs, the monitor provides a special sequence of display windows to help you regain
use of your monitor as quickly as possible. Do the following:
1. Connect an ac power adapter to recharge the battery (if the battery is drained).
2. Press any button below the equipment alert screen to acknowledge the alert. The
monitor will display the Mode Setup window (shown on page 33).
3. Press these buttons to select one of the Factory patient modes for use:
After you press YES, the monitor will display the Time/Day window.
4. Press NEXT, UP, and DOWN as needed to set the time and date. Then press ENTER to store
the new time and date.
These display screens are only displayed in this order if the PROGRAM
FAULT equipment alert occurs.
5. Turn off the monitor, then turn it on again so the settings will take effect.
The monitor is ready for use. If you want to store some customized patient mode program
settings, refer to page 33.
If you follow these steps and the equipment alert reappears at powerup, the monitor may
need to be serviced and the battery replaced. Contact a qualified service person.
Patient Connections
The left side panels differ depending on the Propaq Encore model. All models have ECG,
NIBP, and two temperature connectors. The Propaq Encore 204 left side panel includes one
invasive-pressure connector, and the Propaq Encore 206 includes two invasive-pressure
connectors.
ECG / EKG RESP INV. BP P1 ECG / EKG RESP INV. BP P1 ECG / EKG RESP
T2 T2 T2
INV. BP P2
ECG / EKG RESP INV. BP P1 ECG / EKG RESP INV. BP P1 ECG / EKG RESP
!
COMPATIBLE
HP
! P P P
S ! S S
COMPATIBLE N COMPATIBLE N N
HP I I I
HP
NIBP NIBP INV. BP P2 NIBP
On Propaqs with the Hewlett-Packard connector option, all models have only one
temperature connector, the YSI 400 connector.
Option Connectors
CO2
CO2
!
!
SpO2
NELLCOR
SENSORS/CAPTEURS
Warning The Propaq Encore will show + + + for HR numerics between 301-
350 beats per minute. Above 350 beats per minute, it may display incorrectly
low heart rates, due to intermittent picking of R-waves.
Time of day, caution and Temperature values are Bells indicate alarm
status messages. displayed here. limit status
The screenspace is reallocated when vital signs are added or removed. By changing the size
of the numeric windows below the heart rate, the Propaq Encore provides the best possible
view of all numerics for vitals signs being monitored.
You can select up to three waveforms to be shown on the Propaq. When only one waveform
is selected, a trend window automatically appears below the waveform. While changing
Propaq Encore settings, a status window may appear below the waveform.
MAIN MENU Pressing MAIN MENU always returns the monitor to the top level menu.
SUSPEND/ FREEZE/
RESUME UNFREEZE
MAIN MENU
START/
STOP The five labels
(NIBP) above the
associated
buttons
The five buttons below the screen, and their associated labels located on the screen, provide
access to the menus.
Later in this manual, the notation A , B , C is used as a shorter way to say “Press Button A, then
B, then C.”
MAIN MENU
NIBP ECG/RESP INV PRS SpO2/CO2 SETUP
CO2 MENU 1
RANGE mm/s MORE PREV MENU
SpO2 MENU 1
SIZE MORE PREV MENU
SpO2 MENU 2
RESPONSE C-LOCK PREV MENU
ECG/RESP MENU 1
ECG SIZE ECG LEAD RESP SZE MORE
ECG/RESP MENU 2
NEXT CHANGE PREV MENU
NIBP MENU
START AUTO/MAN INTERVAL TURBOCUF PREV MENU
STOP
SETUP MENU 1
STATSCALE ALARMS WAVE SEL TRENDS MORE (MORE button takes you to next Setup Menu)
TRENDS MENU
PRINT NXT TRND OXYCRG
ALARMS MENU
STAT SET SUSPEND ALL ALRM LIMITS PREV MENU
RESUME
TIME/DAY MENU
PREV MENU
SERVICE MENU 1*
SERVICE MENU 2*
TEMP TEST PIXL TST NET TEST MORE
SERVICE MENU 3*
SETTINGS KEY TEST SYSTEM MORE
SETTINGS MENU
Monitor Setup
Setup Menu 1 is accessed by pressing the SETUP button on the Main Menu.
CURRENT SOURCE When the selected HR/PR source is no longer available, the current
source is the active source with highest priority. The RR/BR source
cannot be manually selected. It will always be CO2 if CO2 is active.
Otherwise, it will be ECG/RESP.
SELECTED SOURCE The user-selected HR/PR source is displayed along with the HR/
PR source currently being used by the monitor.
SWEEP (mm/s) The selectable sweep speeds for HR/PR are 12.5, 25, and 50 mm/
sec. The sweep speeds for RR/BR are 3.13, 6.25, and 12.5 mm/sec.
ALARM TONE Sets the Alarm Tone volume to HIGH, MEDIUM, or LOW.
HR/PR TONE Sets the Heart Tone volume to HIGH, MEDIUM, LOW, or OFF.
PATIENT MODE Pressing CHANGE in this selection displays the following Patient Mode
window:
If you change the patient mode, the CO2 alarm limits in the
new mode might vary slightly from the originally-programmed
CO2 limits for the new mode. Check the CO2 alarm limits.
Display Priorities
You can turn on more than three waveforms, but only the first three waveforms listed in the
wave select window that are monitored are displayed. The patient parameters being
monitored are listed in the order they will be displayed if all are turned on.
Because of the critical nature of the ECG waveform, you cannot turn off ECG. However, if
ECG is not monitored, another waveform will occupy its place.
The displayed waveforms are also the ones printed if a printer is attached.
H:MIN:S MO/DA/YR
35 12
mmHg
07:45:32 06/12/97 SpO2
NEXT UP DOWN ENTER
92
PREV MENU
Press NEXT, UP, and DOWN as needed to set the time and date. Then press ENTER to store the new
time and date.
When you change the hour/minute/second setting for the monitor in the Time/Day
window, the monitor deletes any patient trend data that is older than five hours for non-
NIBP trends or older than eight hours for NIBP trends according to the new clock setting.
However, if the monitor has not yet stored the full capacity of trends and you change the
hour/minute/second setting to a time that is within the stored trend period, previously
stored trends are not erased.
Changing the day, month, or year setting does not affect the stored patient trends.
DECIMAL Sets the decimal character as either a period (.) or a comma (,).
HR/PR ALARM Allows or prohibits turning off the HR/PR alarm limits. If CANNOT
LIMITS
TURN OFF is selected, the ON/OFF button is not displayed on the HR/PR
Alarm Limits Menu.
CO2 UNITS Sets the CO2 display units as mmHg, kPa, or percent (%). If you change
the units, the CO2 trends will be cleared and CO2 alarm limit settings
change to the factory default settings for the currently-used patient
mode.
Any time you change the Date, Filter, Temp F/C, Decimal, HR/PR Alarm
Limits (CAN or CANNOT TURN OFF) or CO2 Units setting, the new setting
also becomes the powerup default setting.
• Factory patient modes. The powerup settings and alarm limits for these patient
modes are preset and cannot be changed. They are listed in Appendix B on page 158.
• Custom patient modes. You can customize the powerup settings and alarm limits for
these patient modes. (See SAVE on page 34.)
Warning If any alarms are set to OFF and you select SAVE to store the
settings for that CUSTOM patient mode, those alarms will be OFF whenever
the Propaq powers up in that CUSTOM patient mode or when that CUSTOM
patient mode is selected. Consider carefully before setting CUSTOM patient
mode powerup alarms to OFF.
You can program the Propaq Encore to power up in any of the Factory patient modes or the
Custom patient modes. You can also change the current patient mode during operation.
Whenever you change the patient mode, the alarm limit settings automatically
change to the settings for that mode.
From the Main Menu, press SETUP, MORE, CHANGE, SETUP. The Mode Setup window appears:
The asterisk (*) indicates which patient mode is currently selected for powerup.
1. Make sure the patient mode you want to reprogram (ADULT, PED or NEO) is
currently-used (to change patient modes, highlight the desired mode and
press USE NOW, YES).
2. Exit the Mode Setup window, then use other menus and buttons to set the
monitor settings and alarm limits as desired.
3. Re-enter the Mode Setup window, highlight the desired Custom mode, and
press SAVE, YES.
Printer Functions
Press SETUP, MORE, PRINTER to display the printer menu and setup window.
LOW BATTERY, This message appears when the Propaq’s battery voltage is
PRINTER DISABLED less than 7.6 volts. To continue operation, plug the ac power
adapter into the Propaq.
CHECK DOOR The door on the bottom of the printer is open. Close door to
remove this message.
PAPER OUT To add printer paper, see page 128.
OVERHEATING The printer is overheating. Service may be required.
The front panel of the printer lets you control the basic printer functions.
If you press FREEZE prior to pressing SNAPSHOT, the printer prints the 8 seconds of patient
information obtained prior to when you pressed FREEZE.
To begin practicing with your Propaq, disconnect all patient cables connected to the monitor.
Leave the cuff connected so you can take NIBP measurements. If you have been monitoring a
patient, turn off the Propaq Encore and turn it back on. From the Main Menu, press SETUP,
WAVE SEL, INSERV.
The Propaq Encore has two sets of simulated patient information—an initial set and an
alternate set. To change between them, press the INSERV button again.
If you connect a patient cable or set the NIBP channel to automatically take pressure
measurements, the Propaq Encore stops simulating, goes through its powerup tests, and
erases any simulated trend data it might have stored.
• change from °F to °C
NIBP
For noninvasive pressure measurements, keep the Propaq Encore in manual NIBP operating
mode and take pressure measurements by pressing the START button. You can also press the
NIBP Menu's TURBOCUF button to consecutively take pressure measurements for five minutes.
Printer Message
Simulated data can be printed on the Propaq Encore Printer. All printouts include the
message “SIMULATED DATA” every four inches to prevent simulated data from being
mistaken for actual patient data.
• You cannot set the Propaq Encore to take automatic noninvasive pressure measurements
(except Turbocuf) while using in-service mode.
• You cannot use Defib Sync or Real-time ECG output while using in-service mode.
• You cannot activate in-service mode if you have been monitoring a patient.
ECG/RESP
With any monitor that detects respiratory effort via impedance pneumography,
artifact due to patient motion, apnea mattress shaking, or electrocautery use
may cause apnea episodes to go undetected. Always monitor and set alarms
for SpO2 when using impedance pneumography to monitor respiratory
function.
Don’t place the Propaq Encore monitor with RESP in close proximity with
another respiration monitor because the RESP measurement frequencies
may interfere with one another.
Since RESP is derived from the same leads as the ECG channel, the Propaq
Encore determines which signals are cardiovascular artifact and which signals
are a result of respiratory effort. If the breath rate is within five percent of the
heart rate or a multiple or sub-multiple of the heart rate, the monitor may
ignore breaths and trigger an apnea alarm.
Intended Use–ECG
The Propaq Encore is intended for ECG monitoring of either a five-lead or three-lead
configuration, including the Marriott configuration 1 (MCL1 requires all three electrodes).
The five lead configuration can derive one of seven user-selected signals, Lead I, II, III, aVR,
aVL, aVF, or V.
The monitor will automatically determine if only three lead wires are connected, and will
automatically reduce the number of selectable leads to three (I, II, III). If four-wire ECG
cables are used, they will be handled as if they were three-wire cables.
Warning The Propaq Encore monitor does not have automated arrhythmia
analysis, therefore, some ventricular tachycardias and ventricular fibrillation
may not be interpreted correctly and may display an inaccurate heart rate.
The Propaq Encore 200 series does not have automated ST segment monitoring, although
with ECG set for extended bandwidth, ST segments may be accurately displayed and
printed.
The Propaq Encore ECG’s bandwidth is 0.5-40 Hz in Monitor Mode and 0.05-40 Hz in
Extended Mode. Monitor Mode is useful to minimize baseline wander due to respiration or
other artifact. However, in Monitor Mode, ST segments can be distorted, potentially causing
underestimation of ST elevation and overestimation of ST depression. Always use Extended
Mode when observing ST segment morphology on the display or printer.
The Propaq Encore can be used during procedures using electrosurgical machines and
defibrillators. However, even though the ECG channel contains electrosurgical interference
suppression (ESIS) circuitry, noise artifact may be displayed on the ECG trace while an
electrosurgical device is in use. This will vary depending on ECG electrode placement and
the operative site.
Verify patient mode. Incorrect patient mode may result in inaccurate heart
rates and inappropriate alarm settings.
Even though the Propaq Encore contains fully isolated patient-connected circuitry, it has not
been specially designed for direct cardiac application.
The Propaq Encore can be used on patients with pacemakers. See page 47.
To prevent injury, use the provided garment clips to route the ECG cables away from the
patient’s head.
Warning Use of ECG cables with loose or faulty detachable lead wires may
cause erratic behavior of the ECG waveform, SpO2, C-Lock, and NIBP due to
intermittent ECG lead wire connections.
Caution To protect the Propaq Encore from damage during defibrillation, for
accurate ECG information, and for protection against noise and other
interference, use only ECG electrodes and cables (namely, ones with internal
current-limiting resistors) specified or supplied by Welch Allyn, and follow
recommended application procedures.
Preparation
Warning Use only ECG safety cables that are designed so that they cannot
accidently be plugged into an AC mains outlet or make contact with other
hazardous electrical potentials including earth ground. To prevent damage
during defibrillation, don’t use ECG cables without 1K series resistors.
Preparing for ECG monitoring with the Propaq Encore requires you to prepare the monitor,
prepare the patient, set up the ECG channel, and then set the ECG alarms.
Warning Before you use a Propaq on a new patient, always turn it off for a
few seconds, then turn it on again. This clears the prior patient’s trend values,
alarm limit settings, and NIBP cuff inflation target.
3. Select the patient mode appropriate for the patient (Neonatal, Pediatric, Adult). To
change patient modes, see page 29.
2. If you are using pre-gelled electrodes, use only electrodes that have not expired.
Make sure there is a generous amount of gel in the electrode and that it has not dried.
For best results, use silver/silver chloride electrodes.
3. If you are using non-gelled electrodes, apply a 1/4 to 1/2 inch mound of gel over the
electrode contact area.
RA LA G
Marriott
Standard 3-lead LL Configuration
Configuration MCL 1
(G)=LA (-)=RA (+)=LL
RA LA
V1 V6
RL LL
5-lead Configuration Neonate 3-lead Configuration
5. Support the ECG cable so it does not stress the electrode wires, ECG cable connectors,
or electrodes.
Two RESP leads are available. Choose the one that gives you the best signal.
If neither signal is adequate, it may be necessary to experiment with
nonstandard electrode placement such as placing the RA and LA electrodes
on the respective mid-axillary lines just above the level of the nipples.
6. If an electrosurgical unit is going to be used, place the ECG cable and electrode wires
as far as possible from the surgical site and from the electrosurgical return electrode
and its cables. This will minimize interference.
By now there should be some kind of ECG waveform displayed on the monitor. A
heart rate should be displayed to the right of the waveform. Depending on how the
Propaq Encore is programmed, a beep tone may occur with each detected QRS event.
7. If there is no waveform, check the electrodes, wires, cable, and the monitor for a
possible lead fault.
Motion artifact or other factors can cause false HR/PR alarms. To help minimize false alarms,
the Propaq delays or “holds off” triggering an HR/PR alarm for 3 seconds. During this
holdoff period, if the Propaq detects that the patient’s HR/PR vital sign has returned to
acceptable limits, the Propaq cancels the alarm holdoff. The next time an HR/PR limit is
violated, the Propaq starts a new 3-second HR/PR alarm holdoff period.
Respiration rate numerics are sourced from the CO2 channel and displayed as BR when CO2
is active. Otherwise, respiration rate numerics are sourced from the ECG/RESP channel and
displayed as RR.
Patient Artifact
Patient movement and other artifact might cause the waveform to move on the display. Most
artifact such as this is automatically detected, and the waveform is adjusted so that it always
remains centered in the waveform window.
Severe artifact and interference (such as interference from defibrillation) may cause the
waveform to move off the display. The Propaq Encore will always automatically reposition
the waveform in just a few seconds so you can see it again.
ECG SIZE Selects the ECG waveform size; sizes are shown in millivolts per
centimeter (.2, .5, 1, 2, or 4 mV/cm) to the left of the waveform.
ECG LEAD Selects the ECG lead. The available leads are lead I, II, III, aVR, aVL,
aVF, or V. The Propaq's factory default lead setting is Lead II.
The Size function “increases” and “decreases” the ECG or RESP waveform size. Each time
you press a SIZE button, the waveform approximately doubles in height. When you reach the
largest waveform size, the next press displays the smallest size.
When you press MORE in the first ECG/RESP Menu, a status window appears showing you
the current ECG/RESP settings and additional selections.
Selected ECG
lead 18:45:28 NIBP T1 102.3 °F HR
Sensitivity
III
mV
1cm NIBP
80
factor MON 122
ECG/RESP ( 85) 58
Bandwidth
HR/PR TONE : LOW PA 35/ 18
Selection MCO2 BR
PACER DISPLAY: ON
ECG BANDWIDTH: MONITOR (0.5-40Hz) 35 12
mmHg
RESP LEAD : Ld1 (RA-LA)
SpO2
Available with the
RESP
NEXT
: ON
CHANGE
92
PREV MENU
Impedance
Pneumography
Option
HR/PR TONE Sets the heart tone loudness to LOW, MEDIUM, HIGH, or OFF.
If SpO2 is monitored, the pitch of the tone varies with the SpO2 value.
PACER DISPLAY Turns on and off the pacer indicator in the ECG waveform.
ECG This selection allows you to determine the bandwidth for the data sent
BANDWIDTH to the display and the printer. If the selection is Extended, the
bandwidth is 0.05-40 Hz. If the selection is Monitor, the bandwidth is
0.5-40 Hz.
The Size function “increases” or “decreases” the ECG or RESP waveform size. Each time you
press a SIZE button, the waveform approximately doubles in height. When you reach the
largest waveform size, the next press displays the smallest size.
The QRS detector sensitivity threshold is not affected by changing the ECG
display size. Likewise, the RESP breath detector threshold is not affected by
changing the RESP display size.
RESP LEAD Selects the RESP lead. Choices are RA-LA and RA-LL, and choice is
independent of ECG lead selection. Experiment with placement for best
signals.
The Propaq Encore counts as “breaths” respiratory efforts that are larger than
two times background Cardiovascular Artifact (CVA).
Since RESP is derived from the same leads as the ECG channel, the Propaq
Encore determines which signals are cardiovascular artifact and which signals
are a result of respiratory effort. If the breath rate is within five percent of the
heart rate or a multiple or sub-multiple of the heart rate, the monitor may
ignore breaths and trigger an apnea alarm.
Warning Pacemaker signals can differ from one pacemaker to the next. The
Association for Advancement of Medical Instrumentation (AAMI) cautions that
“in some devices, rate meters may continue to count the pacemaker rate
during occurrences of cardiac arrest or some arrhythmias. Do not rely entirely
upon rate meter alarms. All pacemaker patients should be kept under close or
constant observation.”
Pacemaker Signals
If the patient being monitored has a pacemaker, the Propaq Encore detects and can indicate
the occurrence of pacemaker signals. With the Propaq, pacemaker signals are not counted as
heart beats as long as the pacemaker signal meets the pulse amplitude, pulse width, and
overshoot/undershoot specifications listed in Appendix B.
Pacer Display
On the Propaq Encore display, vertical dashed lines indicate each time a pacemaker signal is
detected when the Propaq Encore PACER function is turned on. The waveform “spike”
produced by the pacemaker will also be displayed if it contains sufficient energy. Whether
the pacer is atrial, ventricular, or both, the indicator and the spike appear. If the PACER
function is turned off, only the pacemaker spike is displayed:
18:45:28 HR
III
mV
1cm
80
MON
Warning The presence of much pacer-like noise can cause the displayed
heart rate to be erratic even though the ECG trace may look clean with the
pacer indicator off. Fix the noise problem by using fresh ECG electrodes and
an ECG cable whose lead wires make good connections.
Set the filter to the ac mains frequency of your hospital (either 60 or 50 Hz).
ECG Messages
If a lead fault occurs, an ECG equipment alert typically shows which lead failed. Otherwise,
the Propaq Encore displays MULTIPLE in place of the failed leads.
ECG LEAD CHANGED. The Propaq Encore has automatically changed an ECG lead
due to a lead wire or electrode problem.
LEAD FAIL: REPLACE ELECTRODES. There is excessive offset on the ECG channel.
RESP Messages
LEAD FAIL. One or more electrodes are making very poor or no contact. Check for
proper connection; replace electrodes if needed.
INAPPROPRIATE ECG CABLE. ECG cable appears not to contain 1K current limiting
resistors. These resistors are required for RESP operation and to protect the Propaq
Encore from damage during defibrillation. Replace cable with proper type.
NOISY SIGNAL, CHECK ELECTRODES. Electrodes are making poor contact and may
be dried out. Replace electrodes.
Invasive Pressure
This section applies only to Propaq Encore Models 204EL and 206EL. If you don't have one of these
models, you can skip this section.
Intended Use
The Propaq Encore invasive pressure channel is intended for measuring arterial, venous, and
intracranial pressures using invasive transducers with 5 µV/V/mmHg sensitivity. The
Propaq Encore can be used with many types of transducers, including nondisposable,
disposable dome, and fully disposable.
Before you use a Propaq on a new patient, always turn it off for a few seconds,
then turn it on again. This clears the prior patient’s trend values, alarm limit
settings, and NIBP cuff inflation target.
Preparation
Preparing for invasive pressure monitoring with the Propaq Encore requires you to prepare
the transducer, zero the transducer, set up the pressure channel, and set the invasive pressure
alarm limits.
2. Apply the transducer according to your hospital's procedures. Always refer to the
transducer manufacturer’s Directions for Use.
3. If the transducer is a disposable unit with separate cable, connect the transducer to
the transducer cable. Plug the transducer cable into an invasive pressure connector on
the left side panel.
The message NOT ZEROED (or NO ZERO, depending on the zone) immediately
appears in the blood pressure numerics window for the invasive pressure channel
being used.
1cm NOT
MON ZEROED
RR
P
1 SpO2
2. If the ZERO menu is not displayed, press the following Propaq Encore buttons: INV
PRS, then ZERO P1 (or ZERO P2). The word ZEROING appears in the numerics window
during zeroing. The button label changes to CANCEL to allow you to cancel the zeroing
process if necessary
3. Wait for a tone to briefly sound and the word ZEROED to appear in the blood
pressure numerics window. You will then see the pressure scale to the left of the
waveform, and the pressure numerics appear.
5. If the transducer will not zero, the words ZERO REJECTED (or NO ZERO, depending
on zone) will appear in the numerics window. Press CANCEL and try zeroing again.
You won’t see the pressure values and the scales until an acceptable zero reference is
established.
6. Check that the transducer is open to atmospheric air and that it is properly connected
to the Propaq Encore, then try zeroing again. The Propaq Encore will not allow
zeroing to occur if the pressure waveform is pulsatile, if there is too much noise in the
signal or if the transducer’s offset is too great. Once the channel is zeroed, the
pressure scale appears next to the waveform.
If the transducer still does not zero, try another transducer or another cable.
Rezeroing a Transducer
You can rezero a transducer at any time, after again opening the transducer stopcock to
atmospheric air. If the transducer has already produced pressure readings, rezeroing
provides a new zero reference for the Propaq.
If the zero value is not accepted, the Propaq Encore continues to use the previous zero
reference and displays the pressure values and waveforms based on that value. If the new
zero value is accepted, the new pressure values based on the new zero value are displayed,
and the waveform is adjusted according to the new scale.
If the invasive pressure channel enters an alarm condition while the ZERO
REJECTED message is overriding the invasive pressure numerics, no
invasive pressure numerics will flash to indicate invasive pressure is in alarm.
To remove the ZERO REJECTED message and to restore the invasive pressure numerics
during an invasive pressure alarm, you must return to the invasive pressure menu and press
CANCEL. This will restore the invasive pressure numerics.
The Propaq Encore allows you to identify the pressure measurement with a selectable label,
and the numerics can be displayed in different formats.
The pressure waveform scales are not displayed until you zero the transducer. Once the zero
reference has been established the scales automatically appear.
RANGE Sets the display to Range Mode. All invasive pressure waveforms monitored
are displayed against the same scale. You can select one of five Propaq Encore
pressure scales. If two waveforms have a great difference in their pressures, the
higher pressure waveform may not be visible if it is out of range of the scale.
Press RANGE until the desired scale appears.
RESCALE Sets the display to Rescale Mode. Each invasive pressure waveform is
displayed against its own scale. Each time you press the button, the scale is
automatically selected based on the highest and lowest pressure levels of each
pressure waveform.
LABEL Selects a label for the pressure channel. The selectable labels are:
ART—arterial, PA—pulmonary artery, CVP—central venous pressure, ICP—
intracranial pressure, UA—umbilical artery, and UV—umbilical vein.
You can still use the generic Propaq Encore pressure label, P1 or P2.
FORMAT The Propaq Encore displays the invasive pressure values in two different
numeric formats in the pressure numerics window. You can select which
pressure value(s) are most prominently displayed.
Range
Mode 180 ( 85)
I CVP
In this mode, there
N
V
15
is one scale and 90
25/ 9
one label for both P SpO2
pressure
waveforms.
R
S 0
RANGE RESCALE ZERO ART ZERO CVP
99
MORE
NOT ZEROED (or NO ZERO). No zero reference has been established. The monitor
displays the pressure waveform, but to protect against erroneous readings, the pressure
waveform scale is not displayed. To remove this message, zero the transducer.
ZEROED. This message appears after the zero value has been accepted. It remains for
eight seconds and is replaced by the current pressure values.
ZERO REJECTED (or REJECT). Unable to establish a zero reference value. The message
remains until the CANCEL button is pressed.
CANCELED. This message appears if CANCEL is pressed while the channel is zeroing.
TRANSDUCER SHORT CIRCUIT. This message appears when the Propaq Encore
senses a short in the transducer. The transducer should be replaced.
NIBP
Intended Use
The Propaq Encore noninvasive blood pressure channel (NIBP) indirectly measures arterial
pressures using an inflatable cuff. If ECG is also monitored, the Propaq Encore synchronizes
the NIBP measurement process to the occurrences of the R-wave, increasing accuracy in
cases of extreme artifact, diminished pulses, or some dysrhythmias.
Warning The patient's limb should be periodically observed to ensure that the
circulation is not impaired for a prolonged period of time.
The Propaq Encore should never be used to monitor NIBP on one patient
while simultaneously monitoring ECG on another patient.
Some or all NIBP safety functions are disabled in the NIBP TEST screen in the
Service Menu. Do not attempt to conduct NIBP TEST when the cuff is
attached to a patient.
The Propaq Encore NIBP channel has been calibrated to agree with a central invasive blood
pressure. Diastolic pressures may be 5 to 10 mmHg lower than the auscultatory equivalent.
Systolic pressures may be lower than radial invasive equivalent.
Neonatal Mode is intended for use on infants of up to about 44 weeks gestational age in
neonatal care settings. The Neonatal Mode provides the lowest cuff pressure and shortest
inflation time limits to ensure patient safety and comfort.
Pediatric Mode is intended to be used on larger infants and small children up to nine years
old in pediatric care settings. This mode supports the widest range of cuff sizes and a higher
range of patient numerics for the hypertensive infant or child while still restricting the cuff
pressure and inflation times to limits lower than those allowed for adults.
Adult Mode provides the full range of patient numerics and cuff pressures but limits the cuff
sizes available to the standard child cuff and larger.
NIBP measurements are affected by normal physiological pressure variations from reading
to reading. Normal respiration may affect pressure by as much as 10 to 20 mmHg. Patient’s
emotional state, body position, and cuff fit may also adversely affect NIBP measurements. In
some individuals, the act of taking blood pressure readings may alter the blood pressure.
Successive readings on the same patient may vary for the above reasons.
The static accuracy of the Propaq’s internal manometer can be verified by a qualified
biomedical engineer using a mercury column manometer (refer to the Propaq Encore Service
Manual). The accuracy of the Propaq’s determination of systolic, diastolic, and mean
pressures in a clinical setting can only be assessed by careful statistical analysis of controlled
clinical trials of representative patient populations.
The patented Smartcuf software filtering technology greatly increases NIBP measurement
accuracy in the presence of motion artifact or diminished pulses. Smartcuf synchronizes the
NIBP reading with the R-wave of the patient’s ECG to eliminate noise created by external
stimuli such as patient motion or vibration. The monitor must perform ECG monitoring
while using Smartcuf.
• Connect the ECG leads to the patient and perform ECG monitoring during NIBP.
• From the Main Menu, press NIBP to display the NIBP Menu and set Smartcuf to ON.
Set SMARTCUF ON
If artifact is so severe while Smartcuf is enabled that it affects the accuracy of an NIBP
measurement, that measurement is marked with a special symbol on the display and on
printouts
There may be some situations where it is desirable to disable Smartcuf. This may include
situations with very extreme motion artifact, certain types of arrhythmias, or other situations
where it is not possible to obtain a good ECG signal. NIBP measurements can still be
performed when Smartcuf is disabled.
To disable Smartcuf, from the Main Menu press NIBP to display the NIBP Menu and set
Smartcuf to OFF.
Warning When monitoring NIBP, match the Propaq patient mode to the style
of the cuff. For neonates, set the monitor to Neonatal Mode unless the
circumference of the limb is too large for the cuff. In that case, use the
Pediatric Mode. In the Pediatric Mode, the maximum cuff inflation pressure
can exceed 150 mmHg, and two retries are allowed.
For information about patient mode specifications, see “NIBP Specifications” on page 142.
Preparation
Setting up for noninvasive blood pressure monitoring requires three steps: place the cuff on
the patient and connect the cuff to the monitor, set up the NIBP channel, and set the NIBP
alarm limits.
Warning Before you use a Propaq on a new patient, always turn it off for a
few seconds, then turn it on again. This clears the prior patient’s NIBP cuff
inflation target, trend values, and alarm limit settings.
At powerup, the Propaq has an NIBP default inflation pressure (cuff inflation target) based
on the patient mode (see page 142 for the values). After each NIBP measurement, the Propaq
adjusts the target inflation pressure to optimize the next NIBP measurement. To avoid
possible patient discomfort, be sure to turn the monitor off and then on between different
patients to reset the cuff inflation target to the default value.
The cuff should fit snugly, but not be uncomfortable. The hose must not be kinked or
pinched.
Ensure that the cuff tubing is centered over the brachial artery.
A cuff that is not properly connected to the patient may result in a false
reading if the patient and cuff are moved by motion artifact or clinical
personnel during the NIBP measurement. Always verify the cuff is properly
placed on the patient.
3. Screw the hose connector onto the NIBP connector on the monitor's left side.
4. If motion artifact such as shivering, coughing, or vehicle motion interferes with NIBP
readings, do the following:
Position the patient’s limb away from the body so the applied cuff is not in contact
with the patient’s body or any other object such as a bed rail. Try to keep the cuff at
the same level as the heart.
Connect the ECG leads to the patient and perform ECG monitoring during NIBP.
NIBP values and units Time of last measurement This symbol indicates the NIBP reading was taken
of measure rotate in in the presence of high motion artifact while
and out of this window monitoring ECG. Artifact can affect accuracy.
for up to one hour Last measurement To help reduce artifact, see page 60.
(International English
only) mmHg
122
Measure-
18:46:00 NIBP S 122 T1 102.3 °F HR
85 58
ment mode
Time
III
mV
1cm ART
80 NIBP values are
displayed in one of
remaining in MON 109 these windows if
vacant. After 16
Turbocuf
Mode or Auto NIBP ADULT 18:45:28 mmHg
122 NIBP ( 87) 72 minutes, the display
Mode interval MODE : TURBOCUF changes to that
TIME : 4:21 85 58 122 shown below for 44
Manometer Bar.
SMARTCUF : ON
85 50 100 150 200
85 58 minutes.
SpO2
Systolic, diastolic,
and mean values
are indicated above STOP AUTO/MAN INTERVAL TURBOCUF SMARTCUF
99 NIBP 122/
the manometer as
small triangles. 18:45 58
( 85)
Alarm Limits Bell
START/STOP Starts and stops NIBP measurements. Any time the Propaq Encore is taking a
noninvasive pressure measurement, the START button changes to STOP so you
can stop the measurement in progress. This button initiates the same action as
the START/STOP button at the left side of the screen. Pressing STOP will
automatically vent the cuff.
AUTO/MAN This button switches the mode between Automatic or Manual Mode. The
Manual Mode is the default unless you change it by reprogramming your
Propaq. Measurements can be taken at intervals of 1, 2, 3, 5, 10, 15, 30, and 60
minutes. Press START to initiate a measurement.
INTERVAL Selects the interval at which NIBP measurements are automatically taken. The
interval you select, ranging from one minute to 60 minutes, is shown on the
display next to the word TIME.
TURBOCUF Automatically starts NIBP measurements and continues to take as many
measurements as possible within five minutes.
SMARTCUF Enables or disables the Smartcuf motion artifact filter. NIBP measurements can
still be taken when Smartcuf is off. Artifact may interfere with the accuracy of
NIBP measurements with Smartcuf off.
To enable the display of a waveform or large NIBP numerics, select ON (ECG cannot be set to
OFF). The monitor displays the first three active waveforms set to ON in the order of priority
listed in the Wave Select window. If NIBP is ON and only one or two other waveforms are
ON and active, the monitor displays large NIBP numerics in a waveform window.
The start time may be delayed if the previous measurement ended within 30 seconds of the
scheduled start time, because the monitor requires that the cuff pressure be below 5 mmHg
for a minimum of 30 seconds between measurements to allow time to restore blood flow to
the limb.
NIBP Messages
The following NIBP messages can appear in the equipment alert window. An NIBP caution
message also appears in the numerics window. If an error number (ERR# x) is listed in an
NIBP trend printout or display, it indicates that the corresponding NIBP equipment alert
occurred.
AIR LEAK, CHECK HOSE (ERR# 1). The Propaq Encore could not properly inflate cuff.
Check the hose and cuff for obvious leaks, such as O-rings in the hose connections.
CUFF NOT DETECTED (ERR# 2). During cuff inflation the detected pressure did not
sufficiently rise. Check that the cuff connection is tight and take the measurement again.
KINKED HOSE, CHECK HOSE (ERR# 3). The Propaq Encore could not properly inflate
cuff. Check for a kinked hose between the monitor and the patient.
WEAK PULSES, CAN'T FIND SYS/DIA (ERR# 5). There are not enough pulses to
determine the systolic or diastolic pressures, but a mean pressure is available. Try
reapplying the cuff after squeezing as much air from it as you can.
ARTIFACT, CAN’T FIND SYS/DIA (ERR# 6) . The systolic or diastolic pressures are
unreliable due to artifact, but a mean pressure is available. May be caused by patient
motion.
NO PULSES DETECTED (ERR# 7). The cuff may not be properly applied to the patient,
or the patient may not have detectable pulses due to shock or arrhythmias.
CONNECT ECG TO REDUCE NIBP ARTIFACT (ERR# 8). NIBP artifact prevents a
valid reading. Connect ECG electrodes to improve NIBP measurements.
NO VALID BLOOD PRESSURE FOUND (ERR# 9) . This message can occur due to
motion artifact, the Propaq Encore being set in the wrong patient mode, or the wrong
hose or cuff being used in relation to the patient mode.
CALIBRATING, PLEASE WAIT (ERR# 10). The Propaq Encore periodically recalibrates
the NIBP channel to ensure it can properly make NIBP determinations. Normal monitor
operation continues while the NIBP channel is calibrating. If the NIBP channel has not
updated its calibration in 15 minutes, the channel will briefly deactivate until a new
calibration has occurred.
LOW BATTERY, NIBP DISABLED (ERR# 11). The battery lacks sufficient voltage to be
able to operate the NIBP channel. Connect the Encore to the ac power adapter.
SERVICE REQUIRED, NIBP DISABLED (ERR# 12). Have the monitor serviced.
CUFF TOO LARGE FOR PATIENT MODE (ERR# 13) . The monitor detects a cuff too
large for the current patient mode. First, verify the patient mode. If the patient mode is
correct, make sure the cuff fits snugly. If this alert occurs in Neonatal Mode, change the
patient mode to Pediatric Mode and check alarm limits. If the alert occurs in Pediatric
Mode, change to Adult Mode and check the alarm limits. Note that different pressures
and retries are used for each mode as stated in “NIBP Specifications” on page 142.
KINKED OR NEONATE HOSE (ERR# 14) . This message occurs when the neonate hose
is detected in adult patient mode. Change the hose or the patient mode selection.
ARTIFACT PRESENT, MINIMIZE ARTIFACT (ERR# 15). The monitor has detected too
much artifact to allow accurate readings. Take steps to reduce artifact. Position the
patient’s limb away from the body so the applied cuff is not in contact with the patient’s
body or any other object such as a bed rail. If the Smartcuf motion artifact filter is on,
make sure that the ECG leads are properly connected to perform ECG monitoring during
NIBP. If the Smartcuf motion artifact filter is off, consider turning it on (and connect ECG
if not already connected).
NIBP DISABLED, SERVICE REQUIRED. See SERVICE REQUIRED, NIBP DISABLED above.
RETRY. Since the Propaq Encore did not receive a valid NIBP reading, it will
automatically attempt to take another reading.
STOP
99
AUTO/MAN INTERVAL TURBOCUF PREV MENU
NIBP IN PROGRESS P2
9
PLEASE WAIT
12/ 7
FILTERING ARTIFACT
SpO2
PRESS ANY KEY BELOW TO ACKN0WLEDGE 99
The monitor displays this message when noise or artifact such as vehicle motion causes a
delay while measuring NIBP. To remove the message, press any button below the screen. To
cancel the NIBP measurement, press STOP.
Temperature
Intended Use
Propaq Encore monitors provide two temperature channels (except for the HP-side panel
option). When both channels are active, the difference temperature (∆T) is also displayed.
You can select °C or °F.
Preparation
1. Place the probe on the patient, and plug it into one of the connectors on the Propaq's
side panel. Within a few seconds, the Propaq Encore will display the temperature.
2. To select the temperature units (°C or °F), press SETUP, MORE, MORE, SERVICE, YES (to
access the Service Menu), MORE, MORE, SETTINGS. Use the NEXT and CHANGE buttons to
select and set the temperature units as desired. The Propaq Encore automatically
updates the temperature display to show the newly selected units. Changing units
does not clear Temperature trends.
Temperature Messages
The following messages can appear in an equipment alert window. A temperature caution
message will also appear in the temperature numeric window when one of these messages
appears (except PROBE NOT DETECTED).
PROBE NOT DETECTED. This message occurs when the Propaq Encore has
successfully measured temperature and a probe is then disconnected. Reconnect the
probe or acknowledge the equipment alert by pressing any menu button.
PROBE SHORT. Verify that the probe is properly inserted in the left side panel. If so,
replace probe.
CALIBRATION ERROR, TEMP DISABLED. This message appears when the Propaq
Encore has detected that it cannot accurately measure the temperature. The monitor
should be serviced.
Malfunction of the temperature probes may result in inaccurate readings. Confirm suspect
readings.
Sensors exposed to ambient light while not applied to a patient can exhibit
semi-normal saturation readings. Be sure the sensor is securely placed on the
patient and check its application often to ensure accurate readings.
The pulse oximetry option can be used during defibrillation, but the readings
may be inaccurate for a short time.
Before you use a Propaq monitor on a new patient, always turn off the monitor
for a few seconds, then turn it on again. This clears the prior patient’s trend
values, alarm limit settings, and NIBP cuff inflation target.
Each SpO2 sensor is designed for application to a specific site on the patient within a certain
size range. To obtain optimal performance, use an appropriate sensor and apply it as
described in the sensor’s directions for use.
If excessive ambient light is present, cover the sensor site with opaque material to block the
light. Failure to do so may result in inaccurate measurements. Light sources that can affect
performance include surgical lights (especially those with a xenon light source), bilirubin
lamps, fluorescent lights, infrared heating lamps, and direct sunlight.
If NIBP will be monitored while using SpO2, place the NIBP cuff on a different limb than the
SpO2 sensor to help reduce unnecessary SpO2 alarms. For optimal measurements, avoid
placing the SpO2 sensor on the same limb as an arterial catheter or intravascular line.
Loss of pulse signal can occur if the sensor is too tight, there is excessive ambient light, an
NIBP cuff is inflated on the same limb as the sensor, there is arterial occlusion proximal to the
sensor, the patient is in cardiac arrest or shock, or the patient has hypotension, severe
vasoconstriction, severe anemia, or hypothermia.
Warning Use only Masimo accessories and sensors with the monitor with
Masimo SpO2 option as listed in the Welch Allyn Products and Accessories
booklet (810-0409-XX).
2. Inspect the Masimo SpO2 cable. Replace it if it shows any signs of wear, breakage, or
fraying. Plug the sensor into the cable and plug the cable into the Propaq monitor.
The monitor displays STANDBY in the SpO2 numeric window until it measures
and displays the SpO2 value.
As oxygen saturation increases and decreases, the pitch of the heart tone
rises and falls.
The monitor self-calibrates the SpO2 channel whenever the monitor is first
turned on or a sensor is first connected to the SpO2 channel.
3. From the Main Menu, press SpO2 (or SpO2/CO2, then SpO2) to display the SpO2 menu
similar to the following:
MON ( 85) 58
S S Oxygen
p Y 1 SpO2 saturation
O N x
2 C
SIZE MORE
99
PREV MENU
percentage
4. Press SIZE to adjust the waveform size for best viewing (1x, 2x, 4x, or 8x).
5. Adjust the placement of the sensor until a good SpO2 waveform is displayed. A
waveform with artifact may cause erroneous oxygen saturation readings.
To help minimize false alarms, the Propaq monitor briefly delays or "holds off"
triggering both audible and visual alarms for limit violations for SpO 2% and
Pulse Rate for 10 seconds. After the alarm hold-off period begins, if the
monitor detects that the patient’s vital sign has returned to acceptable limits,
the monitor cancels the alarm hold-off. The next time a vital sign limit is
violated, the monitor starts a new hold-off period.
Warning Use only Nellcor accessories and sensors with the monitor with
Nellcor SpO2 option as listed in the Welch Allyn Products and Accessories
booklet (810-0409-XX).
2. When using a sensor extension cable, inspect the cable before use. Replace it if it
shows any signs of wear, breakage, or fraying. Plug the sensor into the cable and plug
the cable into the Propaq monitor, or plug the sensor directly into the monitor.
3. If the monitor SpO2 receptacle has a locking ring, lock the connector in place by
turning the locking ring clockwise until it stops. For other connectors, make sure the
plug is all the way in.
Caution If you see the error message DEFECTIVE SpO2 SENSOR, either
the sensor is not compatible with the monitor or the sensor is not working
properly. Check the Welch Allyn Products and Accessories booklet (810-
0409-XX) to be sure the sensor is compatible. If compatibility is not a problem,
try another sensor.
The monitor displays STANDBY in the SpO2 numeric window until it measures
and displays the SpO2 value. As oxygen saturation increases and decreases,
the pitch of the heart tone rises and falls.
4. From the Main Menu, press SpO2 (or SpO2/CO2, then SpO2) to display the first SpO2
menu similar to the following:
MON ( 85) 58
S S Oxygen
p Y 1 SpO2
O N x
2 C
SIZE MORE
99
PREV MENU
saturation
percentage
5. Press SIZE to adjust the waveform size for best viewing (1x, 2x, 4x, or 8x).
6. Adjust the placement of the sensor until a good SpO2 waveform is displayed. A
waveform with artifact may cause erroneous oxygen saturation readings.
If you get false SpO2 alarms with patients with low perfusion states or multiple
arrhythmias, try turning off C-LOCK.
To help minimize false alarms, the Propaq monitor briefly delays or "holds off"
triggering both audible and visual alarms for limit violations for SpO 2% and
Pulse Rate for 10 seconds. After the alarm hold-off period begins, if the
monitor detects that the patient’s vital sign has returned to acceptable limits,
the monitor cancels the alarm hold-off. The next time a vital sign limit is
violated, the monitor starts a new hold-off period.
11. If patient movement interferes with measurements, consider the following possible
solutions:
1. While monitoring SpO2, remove the SpO2 sensor from the patient, but leave it
connected to the monitor. When the monitor detects the lack of a pulsatile waveform,
it sounds a patient alarm and displays this menu:
SUSPEND STANDBY
The monitor suspends the SpO2 alarm tone indefinitely and displays STANDBY
in place of SpO2 numerics. SpO2 remains in the Standby Mode until the SpO2
sensor is reapplied to a patient. Other vital sign monitoring is not restricted. By
contrast, if you press SUSPEND instead of STANDBY, the monitor temporarily
suspends all alarm tones; however, the alarm tone resumes after 90 seconds
if the SpO2 sensor is still disconnected from the patient.
The monitor exits the Standby Mode and resumes SpO2 monitoring
The message STBY on the SpO2 trend display and trend printouts indicates
the monitor was in the SpO2 Standby Mode.
SpO2 Messages
The following status message can appear in the equipment alert window or the SpO2
numeric display area:
NO SENSOR DETECTED appears in the equipment alert window and indicates a probe
has been disconnected from the monitor after being plugged in for more than a few
seconds.
SEARCH: during the search time, the SpO2 channel tries to detect blood pulsing through
the measurement site. Once the measurement has been established, the oxygen saturation
value is displayed in the numeric window.
STANDBY is displayed in the numeric window when the SpO2 sensor is disconnected
from the patient, an alarm occurs, and you press the STANDBY button. STANDBY is also
displayed if you first plug the SpO2 sensor cable into the monitor connector before
attaching the SpO2 sensor to the patient.
DEFECTIVE SpO2 SENSOR. If you see the error message DEFECTIVE SpO2 SENSOR,
either the sensor is not compatible with the monitor or the sensor is not working
properly. Check the Welch Allyn Products and Accessories booklet (810-0409-XX) to be
sure the sensor is compatible. If compatibility is not a problem, try another sensor.
Capnography (CO2)
Intended Use
The Propaq’s Capnography (CO2) option is intended to noninvasively measure the following
vital signs or events: End-tidal CO2 (ETCO2), Inspired CO2 (INCO2), Breath Rate, and Apnea.
The CO2 option is available as Mainstream CO2, Sidestream CO2, or Dualstream CO2.
Although Dualstream CO2 provides both Mainstream CO2 and Sidestream CO2 monitoring,
only one method can be used at a time. The v option is required for any Propaq equipped
with the CO2 option.
CO2 reading accuracy is affected by the presence of interfering gases and vapors. If the CO2
option is used on patients who are being administered oxygen (O2) or nitrous oxide (N2O),
be sure to set the appropriate compensation setting using the GAS COMP button.
Warning Before you use a Propaq on a new patient, always turn it off for a
few seconds, then turn it on again. This clears the prior patient’s trend values,
alarm limit settings, and NIBP cuff inflation target.
Avoid exposing a Propaq with the CO2 option to non-patient sources of CO2
such as vehicle engine exhaust or smoke. When such exposure is possible,
avoid opening the printer door. Exposure to these CO2 sources can
temporarily trap v within the monitor or Mainstream v sensor housing, even
when monitor power is off. This can temporarily cause an erroneous elevated
CO2 measurement baseline until the trapped CO2 leaks out and the baseline
returns to zero (which can require as long as 3-24 hours).
CO2 monitoring outside the specified operating temperature range can cause
inaccurate CO2 readings. The operating temperature range for the CO2 option
is different than the range of 0° to 40° C for other Propaq Encore functions:
Patients using Mainstream CO2 must either be intubated or breathing through a well-fitting
face mask connected to a breathing system such as an anesthesia circle system.
Patients using Sidestream CO2 can either be intubated or non-intubated using a CO2
Sampling cannula or a combination CO2 Sampling/Oxygen Delivery nasal cannula.
If Mainstream CO2 is active, the Propaq displays MCO2. If Sidestream CO2 is active, the
Propaq displays SCO2. The Propaq displays CO2 if either or both the mainstream sensor and
sidestream watertrap are installed in the Propaq but neither are active.
Displayed values of ETCO2 and INCO2 are the highest and lowest values (respectively) of
CO2 measured during the time interval set by the RESPONSE setting on the CO2 Menu.
You can set upper and lower alarm limits for ETCO2, and an upper alarm limit for INCO2.
The monitor only displays the numeric value for inspired CO2 (INCO2) if it is in alarm or if it
is greater than or equal to 7.5 mmHg (or ≥1 kPa or 1%). Refer to the CO2 specifications in the
Propaq Encore Reference Guide for more information.
Apnea Events
In the Adult and Pediatric Mode, you can set the apnea delay to 6, 10, 15, 20, 25, or 30
seconds. In the Neonatal Mode, you can set the apnea delay to 6, 10, 15, or 20 seconds. The
Propaq initiates an alarm in response to each apnea event longer than the apnea delay
setting.
When an apnea event is detected, the BR numeric automatically goes to 0 and an apnea
alarm occurs. After the alarm ceases, the Propaq prints an Apnea Ticket if the Apnea Ticket
setting in the Printer Setup window is set to ON.
UNCAL indicates the monitor has detected a problem such as a lack of calibration, an
obstruction, or a low battery.
WARM UP indicates Mainstream CO2 has been activated and is preparing for operation.
This typically requires 30 seconds at room temperature.
START UP indicates Sidestream CO2 has been activated and is preparing for operation.
This typically requires 30 seconds at room temperature.
mm/s Sets the display sweep speed for CO2 and RESP
CO2 ( 85)
PA 35/ 18
GAS COMPENSATION: OFF MCO2 BR
RESPONSE : NORMAL
CO2 SOURCE : MAINSTREAM 35 12
mmHg
SWEEP SPEED : 6.25 mm/s
SpO2
BAROMETER : 762.0 mmHg
GAS COMP RESPONSE SOURCE
92
PREV MENU
RESPONSE Sets the response time for CO2 measurement (NORMAL, FAST, OR SLOW)
SOURCE Changes between Mainstream CO2 and Sidestream CO2 monitoring (if both
options are installed), or disables CO2 monitoring
Setting CO2 SOURCE to OFF allows you to disable CO2 monitoring without
removing the sensor. When the CO2 SOURCE is set to OFF, the Propaq
displays OFF for CO2 numerics.
SWEEP SPEED The CO2 and RESP display sweep speed (set by the mm/s button)
To change CO2 units, use the Settings Menu under the Service Menus.
Changing these units will clear CO2 trends.
CO2 ( 85)
GAS COMPENSATION: OFF PA 35/ 18
RESPONSE : NORMAL SCO2 BR
CO2 SOURCE
SWEEP SPEED
: SIDESTREAM
: 6.25 mm/s mmHg
35 12
BAROMETER : 762.0 mmHg SpO2
FLOWRATE : 175 ml/min
GAS COMP RESPONSE SOURCE FLOWRATE PREV MENU
92
RESPONSE Sets the response time for CO2 measurement (NORMAL, FAST, OR SLOW)
SOURCE Changes between Mainstream CO2 and Sidestream CO2 monitoring (if both
options are installed), or disables CO2 monitoring
Setting CO2 SOURCE to OFF allows you to disable CO2 monitoring without
removing the sensor. When the CO2 SOURCE is set to OFF, the Propaq
displays OFF for CO2 numerics.
SWEEP SPEED The CO2 and RESP display sweep speed (set by the mm/s button)
FLOWRATE Sets the sampling flow rate to either 90 or 175 ml/min. You can change the
flow rate while Sidestream CO2 is active
To change CO2 units, use the Settings Menu under the Service Menus.
Changing these units will clear CO2 trends.
Use only accessories provided or recommended in the Welch Allyn Products and
Accessories booklet. Always refer to the manufacturer's Directions for Use for
instructions about operation, cleaning, and replacement.
2. Connect the adapter, ventilator circuit, and CO2 sensor according to the
manufacturer’s instructions.
Window Window
Warning Prior to using an airway adapter, always look through the window
lumen and inspect the adapter for inadvertently lodged obstructions and for
window integrity.
If the sensor does not easily slide onto the adapter, do not attempt to force
these components together. They fit together in only one way. Take care not
to damage the glass windows.
After attaching the sensor to the adapter, check the adapter again for proper
placement of the sensor and for window integrity.
Warning When attaching the airway adapter, position the adapter so the
sensor is on top to avoid fluid collection in the sensor airway slot. Any
concentration of fluids here can cause inaccurate CO2 readings.
Always check to make sure there are no leaks in the breathing circuit. Check
all of the connections.
3. Plug in the CO2 sensor cable to the Mainstream CO2 connector on the Propaq left side
panel.
4. Set up the CO2 channel and alarms. Follow the steps described on page 81.
When disconnecting the airway adapter from the ventilator circuit, always
detach the CO2 sensor from the airway adapter before removing the airway
adapter from the ventilator circuit.
After you connect a Mainstream CO2 sensor or Sidestream CO2 watertrap, the
Propaq displays the waveform briefly without a value range. It displays WARM
UP (for Mainstream) or START UP (for Sidestream) in the CO2 numerics
window. After about 20 seconds, the Propaq displays the CO2 measurement
and waveform range.
2. Press RANGE until you see the desired waveform scale range on the Propaq screen. The
range choices are shown below.
3. Press mm/s to select either 3.13, 6.25 or 12.5 mm/sec. The default setting is 6.25 mm/
sec. This sweep speed also applies to RESP.
5. If either O2 or N2O is being administered to a patient, press GAS COMP to set the proper
gas compensation as listed below. If no other gas is being administered, set GAS
COMPENSATION to OFF. (OFF is the default setting.)
A SLOW response will decrease ETCO2 false alarms when breath morphology varies
considerably from one breath to the next.
Warning For patient safety, it is recommended that the Breath Rate alarm
limits always be turned on and set appropriately.
Motion artifact or other factors can cause false RR/BR alarms. To help minimize false
alarms, the Propaq delays or “holds off” triggering a RR/BR alarm for 5 seconds.
After this holdoff period begins, if the Propaq detects that the patient’s RR/BR vital
sign has returned to acceptable limits, the Propaq cancels the alarm holdoff. The next
time a RR/BR limit is violated, the Propaq starts a new 5-second RR/BR alarm
holdoff period.
8. Set the alarm limit for Apnea Delay—this is the maximum time allowed between two
successive breaths before an Apnea alarm occurs.
After the first breath has been detected, the Apnea alarm limit setting will be
automatically turned on for as long as the CO2 channel is active. The RR/BR and
Apnea Alarm Limit window is shown below. STAT SET does not affect the Apnea
limit.
OR
Use only accessories provided or recommended in the Welch Allyn Products and Accessories
booklet. Always refer to the manufacturer's Directions for Use.
Breath rates greater than 50 breaths/minute may reduce the reported ET CO2
values. Select the 175 ml/min flow rate to minimize errors at higher breath
rates.
The 175 ml/min flow rate is recommended for intubated adult patients.
When monitoring a small child with a rapid respiratory rate, Mainstream CO2
can provide a more accurate representation of the expired CO2 waveform.
Warning The cannula is disposable and should only be used for a single
patient. Do not reuse the cannula for another patient.
If oxygen is being delivered while using Sidestream CO2 , be sure to use a CO2
Sampling and O2 Delivery Cannula. Using a different type of cannula could
obstruct oxygen delivery.
The exhaust port for Sidestream CO2 is an output for the expired gases from
the patient and any connected breathing apparatus. The exhaust port is
intended only for connection to gas collection equipment such as gas
scavenger devices. Do not allow any other connection to the exhaust port.
If you use a gas scavenging system with Sidestream CO2, be sure to install it
according to the manufacturer’s instructions. The scavenging system should
comply with ISO 8835-3:1997 (E).
Exhaust Port
WARNING - Do not connect sample
line or patient input to exhaust port.
Elbow
If you use a gas scavenging system with Connector
Sidestream CO2, be sure to install it
according to the manufacturer’s
instructions. The scavenging system
should comply with ISO 8835-3:1997 (E).
2. Connect the sample line to the elbow connector and the watertrap. Make sure the
sample line is firmly connected.
CO2 Messages
If a sensor is damaged, contact Welch Allyn’ Technical Services Department for information
on sensor service options.
ALTIMETER FAILURE - RATE. The altimeter has detected that the ambient pressure is
changing at a rate greater than 100 mmHg/minute. When the rate of change is back
within the 100 mmHg/minute range, disconnect and reconnect the CO2 sensor to the
Propaq.
LOW BATTERY, HEATER DISABLED (UNCAL appears in the numerics area). The
Propaq’s battery voltage is too low. The CO2 waveform is displayed without range
values. To continue operation, plug the ac power adapter into the Propaq.
NON-PROTOCOL SENSOR (UNCAL appears in the numerics area). A CO2 sensor has
been connected that does not match Welch Allyn’s specifications. The CO2 waveform is
displayed without range values. Replace the sensor with a Welch Allyn CO2 sensor.
SENSOR FAILURE - EEPROM. The sensor has failed. Replace the sensor.
SENSOR FAILURE - HEATER. The sensor’s temperature control circuit or the Propaq’s
CO2 circuitry has failed. Try replacing the sensor. If the message reappears, have the
Propaq serviced.
SENSOR FAILURE - MOTOR DRIVE. The sensor’s motor drive (in the sensor head)
has failed. Replace the sensor.
SENSOR TEMPERATURE TOO HIGH. The sensor’s temperature is too high. The
sensor’s ambient operating range is 10° to 46° C. When the ambient temperature returns
to this range, this message is automatically removed and operation is restored.
WARM UP or WARM (appears in the numerics area). The sensor heater is warming up.
Wait 20 to 30 seconds for the sensor to heat. Values should appear in the numerics area
when the sensor is sufficiently warm.
ALTIMETER FAILURE - RATE. The altimeter has detected that the ambient pressure is
changing at a rate greater than 100 mmHg/minute. When the rate of change is back
within the 100 mmHg/minute range, disconnect and reconnect the CO2 sensor to the
Propaq.
ALTIMETER NOT CALIBRATED - EEPROM - The Sidestream CO2 option has not been
calibrated. Refer the Propaq to a Biomedical Engineer for calibration.
AMBIENT TEMPERATURE TOO HIGH. The sensor temperature is too high. The
Sidestream CO2 option is disabled until the ambient temperature is within the operating
range specifications.
AMBIENT TEMPERATURE TOO LOW. The sensor temperature is too low. The
Sidestream CO2 option is disabled until the ambient temperature is within the operating
range specifications.
MOTOR FAILURE, SERVICE REQUIRED. The sensor hardware has failed. Send the
Propaq to a Biomedical Engineer for service.
PUMP FAILURE, SERVICE REQUIRED. The pump is not able to maintain the target
flow rate. Send the Propaq to a Biomedical Engineer for service.
SSP BOARD EEPROM FAILURE. Send the Propaq to a Biomedical Engineer for service.
Bells only appear when at least one limit is turned on and the vital sign parameter is being
monitored. For NIBP, the bell always appears whenever at least one limit is turned on, even
when an NIBP reading is not being taken.
• The full bell shows you that all alarm limits are turned on.
• The half bell indicates that at least one alarm limit is turned off.
• The absence of a bell shows that no alarm limits are turned on.
Alarms Menu
The Alarms Menu shown below the status window lets you access other alarm functions to
individually set alarm limits or automatically set them.
To adjust the alarm tone volume, press SETUP, then MORE. The status window appears. Use the
NEXT button to select ALARM TONE, and the CHANGE button to change the setting.
Warning At the highest volume alarm level, the sound pressure level will not
exceed safe limits (OSHA HSM 73-1101, 1972). However, additional
precautions may be required in patients under treatment with ototoxic
medications.
The Propaq Encore also shows you when any alarm limit is turned off by illuminating the
amber ALARM(S) OFF light.
A steady, high-pitched alarm tone sounds whenever a patient alarm limit is violated.
• you adjust the alarm limit so the vital sign does not violate it
Life-Threatening Alarms
Life-threatening alarms are automatically turned on when the channel is activated, and
cannot be turned off by the user. A life-threatening alarm is the highest priority patient
alarm. You are notified of any life-threatening alarm as soon as it is detected. An apnea alarm
is a life-threatening alarm.
Temporary Alarms
A temporary alarm occurs when a patient's condition changes, causes an alarm to occur,
changes again, and no longer violates a limit.
After a temporary alarm, review each vital sign’s alarm limit window until you find the limit
with an asterisk (*) next to it. This limit caused the alarm. The asterisk(s) are cleared when
you leave that limits window.
1. Press ALARMS and then LIMITS to display the first limits window.
2. Press NEXT PAGE (in the event of multiple alarm violations) until all asterisks are
located.
Alarm Limits
Status ALARM LIMITS ( 85) Asterisk indicates
Window UPPER LOWER PA 35/ 18 a temporary alarm
limit violation
ART S : 141 112 MCO2 BR
D : 95 * 62 35 12
Alarm Limits
mmHg (M) : 102 68 mmHg
SpO2
Menu
NEXT UP
92
NEXT PAGE
DOWN ON/OFF
You can “unsuspend” the alarm before 90 seconds has elapsed by pressing the SUSPEND/
RESUME key to the left of the screen, or the RESUME button in the Alarms Menu. If an alarm
condition still exists, the tone will again sound. For NIBP (both pressure values and pulse
rates derived from NIBP), pressing RESUME will not resume the NIBP alarm because NIBP is
not continuously measured.
If the LIMITS button is available, you can make Propaq Encore alarm limit adjustments by
pressing LIMITS.
Illuminated
alarm light Flashing numerics
indicate the limits that
have been violated.
Patient Alarm
Menu
• Turn off the alarm limits only for the violating alarm limit values by pressing ON/OFF
in the Limits Menu after you have selected the appropriate vital sign limits window
(LIMITS, NEXT PAGE, ON/OFF)
The ON/OFF button is not available for HR/PR alarm limits if HR/PR ALARM
LIMITS in the Settings window is set to CANNOT TURN OFF.
• Manually change the violated alarm limit value by selecting the violated limit value
and adjusting it
• Turn off all alarm limits by pressing MAIN MENU, SETUP, ALARMS, ALL ALRM
Alarm Prints
If you have the Printer option, you can set the Propaq Encore to print when a patient alarm
occurs. See page 107 for more information about printing.
Alarm Holdoffs
After the alarm holdoff period begins, if the monitor detects that the patient’s vital sign has
returned to acceptable limits, the Propaq cancels the alarm holdoff. The next time a vital sign
limit is violated, the Propaq starts a new alarm holdoff period. The following table
summarizes the alarm holdoff periods.
When the Propaq is connected to the Acuity Central Monitoring System, Acuity does not
receive any alarm indication from the monitor until this holdoff period expires.
This audio alarm holdoff does not occur in the Neonatal mode.
Vital Sign If the Patient’s Vital Sign Then Calculated New Lower Then Calculateda New Upper
Value is Limit is Limit is
Heart Rate HR ≤ 99 HR x 0.8 HR x 1.2
100-250 HR - 20 HR + 20
HR ≥ 251 Unchanged 250
Pulse Rate PR ≤ 99 PR x 0.8 PR x 1.2
PR ≥ 100 PR - 20 PR + 20
Invasive Inv Prs ≤ 25 Inv. Pressure - 5 Inv. Pressure + 5
Pressure 26 - 99 Inv. Pressure x 0.8 Inv. Pressure x 1.2
Inv Prs ≥ 100 Inv. Pressure - 20 Inv. Pressure + 20
NIBP NIBP ≤ 25 NIBP - 5 NIBP + 5
26 - 99 NIBP x 0.8 NIBP x 1.2
NIBP ≥ 100 NIBP - 20 NIBP + 20
Respiration RR/BR ≤ 25 RR/BR - 5 RR/BR + 5
Rate/Breath 26 - 99 RR/BR x 0.8 RR/BR x 1.2
Rate RR/BR ≥ 100 RR/BR - 20 RR/BR + 20
Temperature Temp ≥ 0˚C Temp - 0.5 Temp + 0.5
SpO2 SpO2 ≥ 0% SpO2 - 5 100% (adult and pediatric
(min. limit 50%) mode)
SpO2 + 5 (neonate mode)
ETCO2 ETCO2 ≥ 0 mmHg ETCO2 - 5 mmHg ETCO2 + 10 mmHg
(min. 15 mmHg)
ETCO2 ≥ 2.0 (% or kPa) ETCO2 - 0.7 (% or kPa) (min. ETCO2 + 1.4 (% or kPa)
2.0% or 2.0 kPa)
INCO2 INCO2 ≥ 0 mmHg Not affected by STAT SET INCO2 + 5 mmHg
INCO2 ≥ 0 (% or kPa) INCO2 + 0.7 (% or kPa)
Apnea Delay Not affected by STAT SET
a.New alarm limits calculated by STAT SET cannot be outside the allowable alarm limit range. If a new limit
is calculated to be above or below the allowable alarm limit range, it defaults to the maximum or minimum
alarm limit allowed for that vital sign.
Warning If a patient’s vital sign value falls outside of the upper or lower alarm
range limit, STAT SET turns off the alarm and the alarm limit except for the
following:
1. The lower alarm limits for SpO2 and ETCO2 are not turned off by STAT SET.
2. Press LIMITS to display the alarm limits window and the Limits Menu.
5. Press UP or DOWN, or ON/OFF (if ON/OFF is available) to set the limit to the desired limit
value.
6. When the limit is set, select the next limit with the NEXT button. Or, to select another
vital sign, press NEXT PAGE.
Repeat steps 4 through 6 to adjust the desired limits. Continue the process until you've set all
the limits you want.
Equipment Alerts
If an equipment alert condition is detected, a high-pitched alarm tone sounds for one second
at five-second intervals. This alert tone pattern repeats until you respond to the equipment
alert by pressing any button located at the bottom of the screen or until the equipment
condition is corrected. You can also press SUSPEND and suspend the audio alarm tone for 90
seconds.
An equipment alert window also appears on the display identifying the condition. Certain
equipment conditions also result in a caution message. Some of these messages alternate
with the time of day, others appear in corresponding numeric windows.
For some equipment alerts, all you need to do is correct the condition and the Propaq Encore
resumes normal monitoring.
If the equipment condition also caused a patient alarm, you will need to first suspend the
alarm tone by pressing SUSPEND, then take the required action.
The Propaq Encore returns the channel to its settings prior to the equipment alert. If a patient
alarm occurred and you turned off any alarm limits, you will need to turn them back on. You
can see which alarm limits are turned on by pressing SETUP, ALARMS.
If this occurs, the monitor provides a special sequence of display windows to help you regain
use of your monitor as quickly as possible. Refer to page 21 and perform the described steps.
If you follow these steps and the equipment alert reappears at powerup, the monitor may
need to be serviced and the battery replaced. Contact a qualified service person.
Every two minutes, the Propaq Encore collects the monitored vital sign numerics and stores them in
its trend memory, which can save the last five hours of trend information. (See “NIBP Trends “on the
next page for exceptions.) All this information can be printed and viewed as a trend print.
Each trend shows the time of the reading, the HR/PR measurements, the SpO2 value (if
configured), and other values. The three blood pressure trends show systolic, diastolic, and
mean pressures, and all but the TEMP trend show respiration rate, if it’s available. Columns
on the trend table show the word ‘OFF’ for the vital signs not being monitored.
The Trend Menu allows you to select trended data for display and print the displayed trend
if a printer is attached.
The Up Arrow button allows you to scroll up to the most current reading and the
Down Arrow lets you scroll down to the oldest reading, four readings at a time.
NXT TRND Allows you to cycle through the current display of each trend group.
The following is programmable: which trend window comes up first. The NIBP trend
window is the factory default.
To clear trend data before connecting a new patient, the monitor power must be cycled. This
will prevent the trend data of a previous patient from being attributed to the new patient.
NIBP Trends
A maximum of 128 NIBP readings are collected (up to 8 hours). NIBP is not measured
continuously like other vital sign parameters. The numerics on this trend are captured at the
time of the NIBP reading.
When the SEARCH message appears in an This symbol indicates that the NIBP reading was taken
NIBP TREND display or printout, it indicates in the presence of high motion artifact while monitoring
that the monitor was not able to complete an ECG. Artifact can affect accuracy. To help reduce
NIBP measurement during that time period. artifact, see page 59.
If an error number (ERR# x) is listed in an NIBP TREND printout or display, it indicates that
an NIBP equipment alert occurred. See page 62 for NIBP alert error numbers and definitions.
Displaying Trends
Displayed trends show the last five hours of data. Trends are displayed if you are at the main
menu and have only one waveform turned on in the wave selection window. Trends are also
displayed when you press the TRENDS button in the first Setup Menu.
Printing Waveforms
If you have an expansion module set up with your monitor, you can print any waveform by
pressing either the SNAPSHOT or START/STOP button on the expansion module when the
waveform is displayed on the monitor. You can also set the monitor to print automatically.
The number of seconds of data shown on the printout depends on the print speed set in the
Printer Setup window.
On the SNAPSHOT command only, the CO2 and RESP waveforms will be printed with different
sweep speeds than the other waveforms.
1. To turn on NIBP TICKET, press SETUP, MORE, PRINTER. The printer setup window
appears.
2. Use the NEXT and CHANGE buttons to select and turn on the NIBP TICKET.
To set the Propaq to print an Apnea Ticket after an apnea event, follow these steps.
1. From the main menu, press the following buttons: SETUP, MORE, PRINTER.
2. Press the NEXT button until APNEA TICKET is highlighted in the Printer Setup
window.
1. From the Main Menu, press the following buttons: SETUP, MORE, PRINTER.
2. Press the NEXT button to select ALARM PRINT in the printer setup window.
OxyCRG
The OxyCRG is a graphical printout of two minutes of continuous HR/PR and SpO2
numerics, and a condensed respiratory waveform. If any of the parameters have been
completely inactive for the two minutes prior to the initiation of the print, the associated
band will be empty.
OxyCRG on Alarm
When an alarm condition is detected, a print will be initiated if OXYCRG ON ALARM is
turned on. If an SpO2 or HR/PR alarm condition is detected, an OxyCRG will be queued to
print 60 seconds after the alarm is detected. If an Apnea or RR/BR alarm condition is
detected, an OxyCRG will be queued to print 75 seconds after the alarm is detected.
The parameters which have alarmed in the two-minute period are indicated by the highlight
of the corresponding labels.
Printing Trends
Printed trends are useful for reviewing the patient's vital signs over the last several minutes
to the last five hours. The Propaq enables you to print one trend or several trends with a
press of a button, or automatically at 4-hour intervals.
1. From the Main Menu, press SETUP, MORE, PRINTER, MORE. The printer trend select
window appears.
2. Using the NEXT and CHANGE buttons, select each of the trends you want printed and
turn them on. Turn off all other trends.
3. Now, each time you want to print the selected trends, press PRINT TRENDS.
Use the CHANGE button to set the trend print times according to the start time (clock hour) of
each shift or to OFF.
Once the Propaq is set up, it will print all the selected trends at each 4-hour interval. You can
place the printouts in the patient's record at the end of each work shift.
Intended Use
Warning Use of equipment, accessories, and parts not recommended or
supplied by Welch Allyn could result in inaccurate patient information or
damage to the system.
Make sure the Acuity network cable is not damaged. The Acuity network cable
is the sole link between the Propaq Encore and the Acuity Central Monitor.
The Acuity system is used as a central monitoring system for Propaq Encore monitors
configured to interface with Acuity.
The Acuity system may be used to monitor all patients. For neonatal patients, use all Acuity
features except the Welch Allyn Cordless Acuity and the arrhythmia detection option. The
Acuity system is intended to be used in compliance with the instructions in this Propaq Encore
Reference Guide, the Acuity System Reference Guide, and accepted hospital and clinical
protocols.
If the Propaq Encore has not been connected to the patient, clear any prior patient’s
trends and alarm limit settings by turning off the Propaq Encore and after a few
seconds, turning it on again.
2. If the Propaq Encore is not already connected to the patient, attach leads and sensors
to the patient as described in Chapter 2 of this reference guide.
3. Plug in the Acuity network cable to the Acuity network jack on the Propaq Encore
side panel shown in the following figure. Plug in the other end of the cable to the
bedside Acuity network jack.
3A
2AG
EKG x 1000
! Acuity Connector
Power Connector
!
12-28V, 3A
4. Connect the AC adapter to the Propaq Encore and the wall outlet to charge the
battery. Check to see that the battery charging light on the monitor’s right side panel
is on.
5. Confirm the patient identification at the bedside or enter the patient information at
the Acuity Central Monitor using the Patient ID Setup Window.
6. If alarm limits have not been set, do so at the Propaq Encore or at the Acuity Central
Monitor using the Alarms Setup Window.
Warning If you don’t set alarm limits, Acuity uses preset settings (for
arrhythmia limits), and the powerup default settings for the Propaq Encore.
MAIN MENU
NIBP ECG/RESP INV PRS SpO2/CO2 SETUP
SETUP MENU 1
STATSCALE ALARMS WAVE SEL TRENDS MORE
ACUITY
ACUITY MENU
NET OFF SNAPSHOT PREV MENU
FRZ PRNT
NET OFF
When the monitor is connected to Acuity, pressing
HR this button disconnects the monitor from the
ART
80 NET OFF Acuity network.
122 FREEZE/UNFREEZE
( 85) 58 FREEZE/
UNFREEZE
When the monitor is NOT connected to Acuity,
pressing this button freezes or unfreezes the
displayed waveforms.
1. To disconnect the Propaq Encore from the Acuity network, press the front panel NET
OFF button.
2. Within 15 seconds, disconnect the Acuity network cable from either the Propaq
Encore side panel or the bedside jack. If the patient will no longer be monitored with
the Propaq Encore, turn off the monitor to erase trend information.
If you do not disconnect the Acuity network cable within about 30 seconds, the
monitor and Acuity attempt to reconnect and prompt you to confirm the patient ID.
Whenever the monitor is connected to Acuity, the NET OFF function overrides the FREEZE/
UNFREEZE function for the NET OFF front panel button. Therefore, you cannot freeze the
displayed waveforms when the monitor is connected to Acuity. Also, you cannot initiate a
Freeze Print at Acuity when connected to Acuity. When the monitor is not connected to
Acuity, pressing the FREEZE/UNFREEZE button freezes or unfreezes the displayed
waveforms.
Printing at Acuity
You can print various waveforms from the Propaq Encore to the Acuity system printer. To
print a waveform shown on the display screen, press SETUP, ACUITY, SNAPSHOT. If you press the
FREEZE button on the front of the Propaq Encore, the button changes from SNAPSHOT to FRZ
PRNT.
If the Propaq Encore displays this message, check the Acuity network cable to be sure it is
plugged in to the side panel and to the bedside jack. If the cable is damaged, replace the
cable.
If the cable appears undamaged and the Acuity system is operating normally, ask your
service personnel to check the network and the Propaq Encore Acuity connector.
For in-hospital operation and recharging from ac mains, an ac power adapter plugs into the
monitor. Use only a Welch Allyn ac power adapter to ensure protection against risk (leakage)
current hazards.
The Propaq Encore can also be powered and recharged from a dc source (isolated from ac
mains) capable of supplying 12-28 Vdc and continuously supplying 25W.
The power adapter contains symbols on its labeling. For definitions of these symbols, see
page 11.
Warning Place the power adapter where it cannot fall and harm someone.
Caution Use of other than Welch Allyn power adapters with the plug rated for
your ac mains can damage or compromise the safety of the Propaq Encore
monitor and may require fuse replacement in the power adapter. Verify that
the Power Adapter is set for the proper mains voltage prior to plugging it into
the Propaq.
Do not autoclave the power adapter. Do not operate the power adapter with a
damaged case, mains power cord, or plug.
Your biomedical technician can change the voltage setting and fuses on the Power Adapter
and can verify that your facility is using the correct power cord.
Procedure
1. Unplug the power adapter’s removable cord from the ac mains outlet and the power
adapter.
2. Turn the power adapter so you can see the window that indicates the voltage setting.
3. Using a small, flat-blade screwdriver, carefully pry the fuse module from the power
adapter.
Both fuses should be replaced at the same time, even if only one fuse has
opened due to an overcurrent situation. The unopened fuse may have been
stressed and could become unreliable.
Caution Spare fuses are contained in housings next to the fuses in the fuse
module as shown in the following picture. Between the fuses is a small
printed-circuit board (PCB) that sets the power adapter to the desired ac
mains voltage. When handling the fuse module, the PCB may slide out.
If the small PCB between the fuses has slipped out of place, slide it back into
place in the fuse module, and verify that the voltage setting indicated in the
window on the fuse module is correct. If the voltage setting is incorrect, simply
slide the PCB out of the fuse module, rotate it 180° and slide it back into place.
Power Switch
Fuse
Fuse Module
Voltage Selector
and Indicator
Spare Fuse
Caution Make sure the voltage selector indicates the proper ac input voltage.
If you change the adapter voltage setting, you must replace all fuses to match
the appropriate type specified on the bottom of the power adapter. The only
fuses contained in the power adapter when shipped from the factory are fuses
specified for the original adapter input voltage setting.
Battery Care
Caution Leaving the monitor's lead-acid batteries in a completely discharged
state may result in permanent battery damage. The batteries should be kept
fully charged.
Recharging Time
The battery charges to full capacity within eight hours (if the monitor remains off).
For monitors without the Expansion Module or SpO2 option, typical monitor operating time
is about 2 hours at 25° C for a new, fully-charged battery. This is when all patient channels are
active and measurements are taken every 15 minutes.
For monitors without the Expansion Module but with the SpO2 option, operating time is
about 5 hours.
For monitors with the Expansion Module and printer, SpO2 and CO2 options, typical
operating time is about 3 hours under the above conditions when print strips are generated
every 15 minutes.
If the green Battery Charging lamp does not light when an ac adapter is
connected, this fuse may be blown.
Caution Do not autoclave this product or its accessories. Do not immerse the
monitor in liquid when cleaning. Do not immerse accessories in liquid when
cleaning unless the accessory manufacturer’s cleaning instructions instruct
you to do so.
Maintenance
Use the following intervals for a guideline. Service may be needed more often in extreme
environments (heat, cold, dust, etc.).
Product Recycling
You can return a Propaq Encore to Welch Allyn for recycling when the monitor reaches the
end of its life (7 to 10 years). You can also return an Encore’s internal battery to Welch Allyn
for recycling when it reaches the end of its life (3-4 years).
Monitor Care
Warning The monitor may not meet its performance specifications if stored or
used outside the specified temperature and humidity ranges.
Caution Storing the Propaq Encore for extended periods (more than three
months) without being connected to the ac power adapter can cause damage
to the battery. Even when the Propaq Encore is turned off, a very small
amount of current is drawn from the battery. For long-term storage, remove
the battery from the Propaq.
See the Propaq Encore Service Manual for procedures on removing the battery.
Removing the battery will cause programmed settings to be lost, but they can
be reprogrammed when the battery is replaced.
Printer Maintenance
Loading Paper
Paper is loaded through the bottom of the printer.
1. Lay the monitor on its back to gain access to the bottom of the printer.
2. Squeeze the locks on the paper door toward each other and pull the door toward you
to open it.
3. Lift the paper roll from the holder and pull out any paper remaining in the printing
mechanism.
4. Place the new paper roll onto the holder, as shown below, and pull out several inches
of paper.
Load the new paper roll onto the Feed the paper through the printer
spindle on the door. mechanism.
5. Slide the end of the paper into the slot of the printing mechanism until it extends out
of the paper exit slot.
Customer Services
Welch Allyn
8500 SW Creekside Place
Beaverton, OR 97008-7107 USA
Worldwide: (503) 530-7500
In the USA, toll-free: (800) 289-2500
FAX: (503) 526-4200
Technical Service
If you need technical assistance on troubleshooting, are interested in customer technical
training on Welch Allyn products, or help with ordering replacement parts, contact Welch
Allyn’s Technical Services Department at:
Repacking
Before returning the monitor for service, call Welch Allyn for return authorization and
instructions on shipping. Repack the Propaq Encore in its original shipping container, if
possible. The container is designed to protect the monitor from possible damage during
shipment.
These instructions are not intended to replace existing hospital procedures relative to the provision of
cardiac electrical therapy and operation of the specific models of defibrillators. Use all safety standards
and clinical protocols as defined by your institution.
The Defibrillator Synchronization feature is designed to operate only with the Physio-
Control LIFEPAK 5® and LIFEPAK 6s® defibrillators.
Warning Use only the correct Welch Allyn cable with the LIFEPAK 5 or
LIFEPAK 6s. (These cables contain circuitry in addition to their wiring.) The
use of any other cable will result in incorrect operation. The Welch Allyn
Products and Accessories Booklet lists correct cables for use with Welch Allyn
products.
Contact
Points
2. Verify the integrity of the ECG patient electrodes and the fidelity of the ECG
waveform on the Propaq Encore for tall, distinct R-waves and minimal artifact.
Warning The R-wave amplitude must be at least 0.5 mV (5 mm tall when the
Propaq Encore ECG sensitivity—SIZE button—is set to 1 mV/cm) to guarantee
that the defibrillator sync pulse will occur no later than 35 milliseconds after the
peak of an R-wave. Reposition the patient electrodes or change the Propaq
Encore lead selection as necessary to ensure sufficient ECG waveform
amplitude. However, make sure the R-wave amplitude is not so high that it
obscures the displayed sync markers.
As a visual gauge for estimating R-wave amplitude, the ‘V’ of the mV/cm label
to the left of the ECG waveform is about 4 mm in height. With the Propaq
Encore ECG sensitivity set to 1 mV/cm, compare the letter ‘V’ with the height
of the R-wave, which should be at least 5 mm tall.
Encore display for dashed lines above and below each (PUSH)
SYNC
18:45:28 T1 102.3 °F HR
II
NIBP
88
mV
1cm 135
MON (103) 81
MCO2 BR
N
I
SYS DIA MEAN 35
mmHg 12
B
P 135 81
mmHg TIME SINCE READING: 10 min
(103) SpO2
99
NIBP ECG/RESP INV PRS SpO2/CO2 SETUP
You must press the LIFEPAK 5’s SYNC button and check for appropriate
synchronization markers on the Propaq Encore before each attempt at
cardioversion. Welch Allyn cannot guarantee the delay from the sync marker
to the defibrillator discharge.
A fault in the cable between the Propaq Encore and the defibrillator, or
unplugging the cable, will prevent showing markers on the Propaq and can
prevent setting the defibrillator to synchronized mode.
2. Verify the integrity of the ECG patient electrodes and the fidelity of the ECG
waveform for tall, distinct R-waves and minimal artifact.
Warning The R-wave amplitude must be at least 0.5 mV (5 mm tall when the
Propaq Encore ECG sensitivity—SIZE button—is set to 1 mV/cm) to guarantee
that the defibrillator sync pulse will occur no later than 35 milliseconds after the
peak of an R-wave. Reposition the patient electrodes or change the Propaq’s
lead selection as necessary to ensure sufficient ECG waveform amplitude.
However, make sure the R-wave amplitude is not so high that it obscures the
displayed sync markers.
As a visual gauge for estimating R-wave amplitude, the ‘V’ of the mV/cm label
to the left of the ECG waveform is about 4 mm in height. With the Propaq
Encore ECG sensitivity set to 1 mV/cm, compare the letter ‘V’ with the height
of the R-wave, which should be at least 5 mm tall.
3. Connect the Propaq Encore end of the Welch Allyn Defib Sync cable to the Propaq’s
right side panel DEFIB SYNCHRO connector.
20 300
10 360
5
CAUTION
HAZARDOUS ELECTRICAL OUTPUT. THIS EQUIPMENT
IS FOR USE ONLY BY QUALIFIED MEDICAL PERSONNEL.
LIFEPAK6s
PHYSIO-CONTROL DEFIBRILLATOR
You must press the LIFEPAK 6s SYNC button and check for appropriate
synchronization markers on the Propaq Encore before each cardioversion.
Welch Allyn cannot guarantee the delay from the sync marker to the
defibrillator discharge.
A fault in the cable between the Propaq Encore and the defibrillator, or
unplugging the cable, will prevent showing markers on the Propaq and can
prevent setting the defibrillator to synchronized mode.
ECG
The ECG channel meets all the requirements for Cardiac Monitors Heart Rate Meters and
Alarms specified ANSI/AAMI EC13-1992, except for Standardizing Voltage (section 3.2.9.9).
The channel also meets the American National Standard, Safe Current Limits for
Electromedical Apparatus (ANSI/AAMI ES1-1993).
ECG Specifications
Characteristic Specification
Connector AAMI 6 pin or Hewlett-Packard compatible 12-pin style connector
(optional). (See the illustration at the bottom of this table.)
Selectable Leads I, II, III, aVR, aVL, aVF, V
Lead Fault Indicator LA, LL, RA, RL, C, multiple
ECG Size (sensitivity) in mV/cm 4, 2, 1, 0.5, 0.2
Display Sweep Speeds 12.5, 25, and 50 mm/sec
QRS Tone Volume High, Low, Medium, Off
QRS Tone Frequency 900 Hz. for Propaq Encore without Expansion Module, 665 Hertz when
equipped with SpO2 but SpO2 not being monitored; variable pitch with
SpO2 option and SpO2 being monitored
Freeze Buffer 3.9 seconds at 25 mm/sec
Bandwidth 0.5 to 40 Hz in Monitor Mode; 0.05 to 40 Hz in Extended Mode (see
Real-Time ECG Analog/Defib Sync specification).
Sample Rate 364 Hz
Input Protection Electrosurgery and defibrillator protected when used with specified ECG
cables. All models also include electrosurgery interference suppression.
Lead Fail Sense Current 50 nA dc for active leads
100-200 nA dc for driven lead, depending on number of electrodes
attached
Tall T-wave Rejection Meets and exceeds AAMI (USA) EC13-1992, section 3.1.2.1.c, for 1.2
mV T-wave and 1 mV QRS using AAMI test waveform.
Common Mode Rejection <1 mV p-p RTI for 10V rms, 50/60 Hz input, 200 pF source impedance,
input unbalanced, FILTER function OFF
<0.1 mV p-p RTI for 10V rms, 50/60 Hz input, 200 pF source
impedance, input unbalanced, FILTER function ON
Input Impedance >2.5 MΩ differential @ 60 Hz
Input Range (ac) 10 mV peak to peak
Input Range (dc) Up to ±300 mV
System Noise ≤30 µV peak-to-peak, R.T.I., with all inputs = 47K in parallel with .047 µF.
Characteristic Specification
QRS Detector Adult or Pediatric Amplitude Range: 0.22 to 5.0 mV (RTI)
Neonatal Amplitude Range: 0.1 to 5.0 mV (RTI)
Neonatal and Pediatric Width Range (Duration):
40 to 120 msec
Adult Width Range (Duration): 70 to 120 msec
Heart Rate Range 25 to 350 beats per minute (measurement)
25 to 300 beats per minute (display)
Heart Rate Meter Response Time Responds to change in heart rate within 5 to 9 seconds depending on
physiological waveform. (As measured per AAMI standard EC 13-1992
clause 4.1.2.1 (f), including 3.1.2.1 parts f. and g. waveforms.) Includes
1 second readout update interval.
HR Accuracy ±3 beats per minute or 3%, whichever is greater
NOTE: AAMI Test 4.1.4 part f: Accuracy is affected (i.e., rate drops)
when QRS and pacer spikes are nearly simultaneous as occasionally is
the case during this AAMI test.
Heart Rate Averaging Method Heart rate = 60 / latest average interval in seconds
For higher heart rates, latest average interval = 7/8 of previous average
interval + 1/8 of latest interval.
For lower heart rates, latest average interval = 3/4 (previous average
interval) + 1/4 latest interval.
Transition rates for choice of formula include hysteresis and are 70 and
80 beats per minute.
Drift Tolerance (AAMI Specification 80 beats per minute indicated for 80 beats per minute ECG plus drift
EC13-1992, 3.2.6.3) waveform
Pacer Display Pacer indicator shown on screen if PACER function turned on; pacer
spike always shown if of sufficient amplitude.
Pacer Pulse Rejection Pacer detection range (i.e., will show the dashed vertical marker) for 0.1
ms pulses is ±3 mV to ±700 mV, and drops linearly to ±2 mV to ±700 mV
for 0.2 to 2 ms pulses.
Will not count as heartbeats approximately 95% of pacemaker pulses
within pacer detection range, with or without AAMI (EC13 1992) tails of
4, 25, 50, 75, or 100 ms decay time constant, whose tail amplitudes are
2.5% or 25%, 2mV maximum, whether ventricular only, or A-V
sequential pulses, all per AAMI tests 3.1.4.1 and 3.1.4.2
Response to Irregular Rhythm (AAMI specification EC13-1992, 3.1.2.1. Part e.)
Ventricular Bigeminy (VB) 78 to 81 bpm (80 bpm expected)
Slow Alternating VB 57 to 65 bpm (60 bpm expected)
Rapid Alternating VB 118 to 123 bpm (120 bpm expected)
Bidirectional Systole 88 to 93 bpm (90 bpm expected)
1mV Ventricular Tachycardia 197 to 198 bpm (206 bpm expected)
2mV Ventricular Tachycardia 193 to 197 bpm (206 bpm expected)
R SHIELD LA
LL
LA RL C
RL
LL C
SHIELD RA
AAMI 6-pin ECG connector side HP 12-pin ECG connector side
panel view panel view
Signal Specification
Sync Output 0 to 5V pulse, 100 ±5ms wide, starts within 35 ms after peak of R-
wave. 15 mA short circuit current.
Real-time ECG Output Range = ±6V minimum, centered about 0V, Gain = 1000X,
noninverting for lead II, inverting for all other leads, delay <3 msec,
0.05-100 Hz, going to -5.9V ±5% during ECG lead fail. V lead has no
Real-Time analog output.
Marker Input (Defib Sync only) Normally 0V in, a pulse either + or -3 to ±15V for 10-70 ms puts a
marker in ECG trace. ~ 5 kΩ input res.
Shield Common terminal for other signals
The sync and real-time ECG outputs do not operate during In-service mode.
SHIELD SHIELD
N/C N/C
Characteristic Specification
Display Characteristics
Sweep speed 3.13, 6.25, 12.5 mm/sec; user-selectable
Amplitude range 1x, 2x, 4x, 8x, 16x
Performance Characteristics
Excitation signal characteristics 65 µA RMS ±5% at 63.0 kHz pseudo sine wave
Sensing electrodes User selectable RA-LA or RA-LL
Base impedance (in addition to 1K 100 to 1200 ohms is normal monitoring range, approx. 1200-1500
resistors in ECG cables) ohms range produces a “NOISY SIGNAL, CHECK ELECTRODES”
equipment alert. Above approx. 1500 ohms produces a “RESP FAULT,
LEAD FAIL” equipment alert. Thresholds are dependent on ECG cable
type.
Impedance dynamic range 20 ohms
Signal bandwidth after detection 0.06 Hz (single pole) to 3.2 Hz (2 pole)
Breath detection threshold 140 milliohms or 2x CVA, whichever is greater
Respiration rate range Adult/Ped: 0 (apnea), 2 to 150 breaths/min
Neonate: 0 (apnea), 3 to 150 breaths/min
Respiration rate accuracy ±2 breaths/min or ±2%, whichever is greater
Respitation rate source (RR) When CO2 is active, CO2 is the BR source. Otherwise, RESP from
ECG is the RR source.
Apnea alarm delay accuracy +1 second
Resolution 5 seconds
Apnea alarm delay settings Central apnea only - alarm delay is set by the user
Adult/Ped = 6, 10, 15, 20, 25, 30
Neonate = 6, 10, 15, 20 seconds
Cardiovascular artifact rejection Presence of CVA is detected automatically. Breaths will be
(CVA) picked in the presence of CVA unless the Breath Rate is within
5% of the Heart Rate or a sub-multiple of the heart rate.
Motion artifact rejection not rejected
Obstructive apnea not detected
Cardiovascular artifact rejection Presence of CVA is detected automatically. Breaths will be
(CVA) picked in the presence of CVA unless the Breath Rate is within
5% of the Heart Rate or a sub-multiple of the heart rate.
Motion artifact rejection not rejected
Obstructive apnea not detected
Invasive Pressure
Applies only to models 204 and 206.
Characteristic Specification
Transducer Type Strain-gauge resistive bridge, or HP quartz (with HP Option). a
Transducer Excitation Impedance 200 to 2000 Ω
Range
Transducer sensitivity 5 µV/V/mmHg
Excitation Voltage 4.85V Pulsed dc @ 181 Hz b
Connector ITT-Cannon plug MS3106F-14S-6P Std.
Hewlett-Packard compatible 12-pin connector (optional).
Bandwidth Digital filtered, dc to 20 Hz
Zero Drift ±1 mmHg without transducer drift
Zero Adjustment ±200 mmHg including transducer offset
Numeric Accuracy ±2 mmHg or 2% of reading, whichever is greater, plus transducer error
Pressure range -30 to 300 mmHg
Pulse range 25 to 250 beats per minute
Leakage Current Meets ANSI/AAMI risk (leakage) requirements
Electrosurgery interference Included in all models
suppression
a.Transducers with 40 µV/V/mmHg sensitivity are not compatible.
b.Duty factor depends on transducer impedance. For 200 to ~900 Ω, duty factor is ≈ 11%. Above ~900 Ω, the
duty factor increases to ≈ 91%.
+ EXC - SIGNAL
- EXC
+ SIGNAL
NIBP
NIBP Specifications
Characteristic Specification
Method Oscillometric
Control Automatic and manual measurement control
Auto Intervals 1, 2, 3, 5, 10, 15, 30, and 60 minutes
Turbocuf Maximum measurements allowable in a 5-minute period
Displayed Pressures Systolic, Diastolic, and Mean plus on-screen manometer
Systolic Range Adult: 30 to 260 mmHg
Ped: 30 to 160 mmHg
Neonate: 25 to 120 mmHg
Diastolic Range Adult: 20 to 235 mmHg
Ped: 15 to 130 mmHg
Neonate: 10 to 105 mmHg
Mean Range Adult: 20 to 255 mmHg
Ped: 15 to 140 mmHg
Neonate: 10 to 110 mmHg
Static Manometer Accuracy ±3 mmHg
Minimum Inflation Pressure Adult: 100 mmHg
Ped: 80 mmHg
Neonate: 50 mmHg
Maximum Allowable Pressure Adult: 270 mmHg
Ped: 170 mmHg
Neonate: 132 mmHg
Default Inflation Pressure Adult: 160 mmHg
Ped: 120 mmHg
Neonate: 90 mmHg
Normal Overpressure Limit (results in Adult: 280 mmHg
up to 2 retries) Ped: 200 mmHg
Neonate: 141 mmHg
Single Fault Overpressure Limit Adult: 308 mmHg
Ped: 220 mmHg
Neonate: 154 mmHg
Leak Rate After a 1 minute settling period, leak rate is ≤4mm/Hg over a 3-minute
period at 270 mm/Hg.
Pulse Rate Range 30 to 220 beats per minute
Maximum Determination Time (with Adult: 4.5 minutes
retries) Ped: 4 minutes
Neonate: 3 minutes
Maximum Determination Time (no Adult: 3 minutes
retries) Ped: 2 minutes
Neonate: 1.5 minutes
Typical Determination Time without 30 to 45 seconds
Artifact
Characteristic Specification
Minimum Time between automatic 30 seconds (Auto Mode)
measurements 2 seconds (Turbo Mode)
Electrosurgery Interference Included in all models: 202EL, 204EL, 206EL
Suppression
NIBP Performance Per EN 1060-1, EN 1060-3, and ANSI-AAMI SP10-1992
NIBP Safety Per IEC 601-2-30
Temperature
Temperature Specifications
Characteristic Specification
Range 0° to +50°C; 32° to +122°F
Displays T1, T2, and ∆T
Probes Compatible with YSI Series 400 and 700 probes. HP side panel only compatible with
YSI 400 and has HP connector.
Units °C and °F selectable
Channel Accuracy Temperature Range Tolerance
0° to +10°C ±0.2°C
>10° to +50°C ±0.1°C
Characteristic Specification
Saturation (% SpO2)
Range 1% to 100%
Resolution 1%
Alarm Limits 52% to 100% (upper)
50% to 98% (lower)
Probe Accuracy (25° to 41° C)
Adults, Pediatrics: No motion 70% to 100% ±2 counts
0% to 69% unspecified
Characteristic Specificationa
Saturation (% SpO2)
Range 0% to 100%
Resolution 1%
Alarm Limitsb 52% to 100% (upper)
50% to 98% (lower)
Probe Accuracyc (saturation levels
between 70% and 100%, 28° to
42°C)
Adult/Pediatric Digit accuracy: ±2 counts
Neonatal Digit accuracy: ±3 counts
Pulse Rate
Ranged Motion tolerant option: 25 to 249 beats per minute
Option without motion tolerance: 25 to 250 beats per minute
27 to 250 beats per minute (upper)
Alarm Limits 25 to 248 beats per minute (lower)
Pulse Rate Accuracy
No Motion ±3 beats per minute
During Motion ±5 beats per minute
Alarm Hold-Off Time Period 10 seconds; resets if the sensor reports levels within limits before 10
seconds elapses
Circuitry Microprocessor controlled
Automatic self-test of oximeter when powered on
Automatic setting of default parameters
Automatic alarm messages
Electrosurgery interference Yes
suppression
Sensor Compatibility Compatible only with Nellcor sensors listed in the Welch Allyn Products
and Accessories booklet.
Sensor LEDs
RED Wavelength 660 nm (nominal)
INFRARED (IR) Wavelength 890 nm (nominal)
Sensor Energies (Radiant Power)
Electrical Power 52.5 mW max.
Optical Power 15 mW max.
Pulse Rate Accuracy
No Motion ±3 beats per minute
During Motion ±5 beats per minute
a.When performing SpO2 monitoring, the monitor SpO2 channel may not (in rare instances) initially detect a damaged SpO2
sensor or extension cable (as described in EN865:1997, Section 51.109), but may simply display a flat line and no numerics. If
this occurs, try connecting a new SpO2 sensor (and extension cable, if needed) and restart the monitor to resume normal
monitoring.
b.Minimum difference between upper and lower alarm limits is 2%.
c.Refer to the Welch Allyn Products and Accessories guide (810-0409-XX) for accuracy specifications for all Nellcor SpO2
probes recommended for use.
d.When using the pulse oximetry option with motion tolerance, a substantial and rapid (<2 seconds) drop in pulse rate may result
in erroneous pulse rate readings and loss of the audible pulse indicator.
Capnography (CO2)
General CO2 Specifications
Characteristic Specification
CO2 Display
Screen Display CO2 waveform and ETCO2 and INCO2 (when in alarm) numerics
Numeric Display Ranges ETCO2 0-99 mmHg, 0-13.2 kPa, 0-23.1%
INCO2 8a-25 mmHg, 1.1a-5 kPa, 1.1a-5%
Waveform Scale (Maximum) 0-100 mmHg, 0-14 kPa, 0-14%
Units mmHg, kPa,%; user-selectable
Sweep Speed 3.13, 6.25, 12.5 mm/sec; user-selectable
Response Modes Fast 15 sec sampling time period
Normal 30 sec sampling time period
Slow 45 sec sampling time period
Gas Compensation OFF CO2 value = calculated CO2 value;
O2 > 50%, No N2O CO2 value = calculated CO2 value x 1.03;
N2O > 50% CO2 value = calculated CO2 value x 0.952
Alarm Limit Ranges ETCO2 0-99 mmHg, 0-13.2 kPa, 0-13.2%
INCO2 2-25 mmHg, 0.2-5 kPa, % (no lower limit)
Resolution 1 mmHg
Accuracy Mainstreamb 0-30 mmHg, ±3 mmHg
31-99 mmHg, ± 10% of value
Sidestreamc 0-30 mmHg, ±3 mmHg
31-99 mmHg, ± 10% of value
Altitude Error ±0.4%/1,000 ft (304.8 m)
Breath Rate Display
Screen Display Numeric
Breath rate (BR) source When CO2 is active, CO2 is the BR source.
Otherwise, RESP from ECG is the RR source.
Units Breaths/Minute
Range Adult/Ped 0 (apnea), 2 to 150 breaths/min
Neonate 0 (apnea), 3 to 150 breaths/min
Resolution ±1 breaths/min
Accuracy ±1 breaths/min or ±5%, whichever is greaterd
Alarm Limits Range Adult/Ped 2 to 150 breaths/min
Neonate 3 to 150 breaths/min
CO2 Perormance
Specification Per ISO 9918:1993 (E) / EN 864:1996
Apnea Alarms and Tickets
Apnea Ticket Set to auto print after apnea event and after 1 minute continued apnea
Apnea Alarm Accuracy ± 2 sec
Apnea delay setting Adult/Ped = 6, 10, 15, 20, 25, 30 seconds
Neonate = 6, 10, 15, 20 seconds
Characteristic Specification
Barometric Pressure
Pressure Compensation Automatic
Operating Range -2,000 to 15,000 ft (-610 to 4572 m), 817 to 429 mmHg
Screen Display Numeric (CO2 Status Window)
Units mmHg or kPa
Accuracy ±3 mmHg or 2.5% of difference from calibration pressure, whichever is
greater
a.Lower if in alarm.
b.Based on these airway conditions: sensor 42°airway adapter temperature = 33°C, water vapor pressure = 38
mmHg; standard gas mixture = CO2 in balance air, fully hydrated at 33°C; barometric pressure = 760 mmHg
and flow = 60 ml/min.
c.Based on the following additional airway conditions: Sample line = 7 ft, 0.055 in ID (2.13 m, 1.4 mm ID);
Sample flow rate = 175 ml/min; Welch Allyn watertrap (new/unused); Respiratory rate ≤ 50 bpm, stable to ±3
breaths/min; Inspired/Expired time ratio = 1:2. Barometric pressure = 760 mmHg.
d.For Sidestream CO2, this applies only for BR≤50..
Characteristic Specification
Mainstream CO2 Sensor
Sensor Type Mainstream
Principle of Operation Non-dispersive, infrared, single-beam, single path/wavelength,
ratiometric
Warm-up time (CO2 sensor and 45 sec typical, 3 min maximum
monitor)
Response Time 30 ms typical, 60 ms maximum
Waveform Rise Time <120 ms to 90% after step change
Calibration Verify semi-annually, calibrate only as required
Sensor Housing Temperature 42°C nominal
Mainstream CO2 Sensor and Cable Dimensions and Weight
Sensor Height a 1.003 in
Sensor Width a 1.036 in
Sensor Depth a 0.78 in
Sensor Weight a < .39 oz
Cable Length 10 ft (3.05 m) nominal
Mainstream CO2 Airway Adapter
Type Per ISO 3040, single-use
Size 15 mm ID, (meets ISO specifications)
Material clear polycarbonate, with sapphire windows
Added Deadspace < 6cc (.37 cubic inches) for adult model, <0.6 cc (.037 cubic inches) for
low deadspace model
Mainstream CO2 Sensor Environmental Specifications
Operating Ambient Temperature 10° to 40°C
Storage Temperature -20° to 60°C
Operating Altitude -2,000 to 15,000 ft (-610 to 4,572 m), 817 to 429 mmHg
Storage Altitude -2,000 to 40,000 ft (-610 to 12,192 m), 817 to 141 mmHg
Operating and Storage Humidity 0% to 95%, noncondensing
Shock 100 g for 4 msec
Vibration 5-35 Hz, 0.015 in peak-to-peak,
35-100 Hz, 1 g acceleration
Drop 36 inches free fall to floor (tile over concrete, one drop each face, one
drop each edge/corner)
a.not including cable
Characteristic Specification
Sensor Type Sidestream, internal
Principle of Operation Non-dispersive, infrared, single-beam, single path/wavelength,
ratiometric
Operating Ambient Temperature 5° to 40°C
Startup Time 30 seconds typical, 3 minutes maximum
Rise Time 240 ms (10% to 90%) at 175 ml/min
Delay Time 1.12 seconds maximuma
Total System Response Time 1.36 seconds maximum (Rise Time and Delay Time)
Calibration Verify semi-annually, calibrate only as required
Sampling Chamber Internal (replaceable by service technician)
Pneumatic and Exhaust System Integral
Barometric Pressure Compensation Automatic
BTPS, ATPS, STPDb CO2 value = calculated CO2 value x 0.977
Sampling Line 7-foot sampling line, ID 0.055 in (1.4 mm), for use with disposable
single-use cannula (CO2 only or CO2 sampling/O2 delivery)
Watertrap Disposable single-use
Flow Rate 90 or 175 ml/min, user-selectable
a.Based on the following additional airway conditions: Sample line = 7 ft, 0.055 in ID (2.13 m, 1.4 mm ID);
Sample flow rate = 175 ml/min; Welch Allyn watertrap (new/unused).
b.BTPS (Body Temperature and Pressure, Saturated), ATPS (Ambient Temperature and Pressure, Saturated),
STPD (Standard Temperature and Pressure, Dry).
Alarms
Alarms Specifications
Characteristic Specification
Indicators Flashing red ALARM light indicates patient alarm; continuously on
ALARM light indicates patient alarms are suspended. Continuously on
ALARM(S) OFF light indicates one or more alarm limits have been
disabled. A flashing ALARM(S) OFF light indicates an equipment alert.
Tone Frequency 900 Hertz
Tone is steady for a patient alarm and sounds for 1 second every 4
seconds for an equipment alert.
Selectable Tone Volume Low, Medium, High
Limits Settable on all parameters
Control Automatic preset or manual settings
Alarm on Tachycardias Most tachycardias will alarm in less than 8 seconds. These include
AAMI 3.1.2.1 part f. waveforms. Certain multifocal tachycardias may
initially alarm as “low rate.”
Apnea delay setting Adult/Ped = 6, 10, 15, 20, 25, 30 seconds
Neonate = 6, 10, 15, 20 seconds
Alarm Holdoff Time Perioda HR/PR = 3 seconds (except NIBP PR)
SpO2 = 10 seconds
RR/BR = 5 seconds
Audio Alarm Holdoff with Acuity When a Propaq Encore in Adult or Pediatric Mode is connected to an
Acuity System, the audio alarms at the bedside Propaq can be delayed
up to 4 minutes and 15 seconds. The delay time is selected in Acuity
software at the time of Acuity installation. Visual alarm indications are
not delayed.
a.Alarm holdoff time period is reset if the vital sign returns to acceptable limits before an alarm occurs.
Trends
Trends Specifications
Characteristic Specificationa
Model 202 Parameters NIBP, T1, T2, ∆T, HR (heart rate/pulse rate), SpO2, End-tidal CO2,
Inspired CO2, Breath Rate
Model 204 Parameters NIBP, P1, T1, T2, ∆T, HR (heart rate/pulse rate), SpO2,
End-tidal CO2, Inspired CO2, Breath Rate
Model 206 Parameters NIBP, P1, P2, T1, T2, ∆T, HR (heart rate/pulse rate), SpO2, End-tidal
CO2, Inspired CO2, Breath Rate
Duration 5 hours for non-NIBP trends (up to 150 readings)
A maximum of 128 readings (up to 8 hours) for NIBP trends
Resolution All channels except NIBP sample data at 2-minute intervals
For NIBP trends, a new entry is placed in the table each time an NIBP
determination is made.
a.Assumes SpO2 and CO2 functions are present.
Display
Display Specifications
Characteristic Specification
Matrix 552 x 256 pixels EL display
Active Viewing Area 145.75 mm x 67.56 mm
Pixel Size 0.203 mm x 0.203 mm
Pixel Pitch 0.264 mm x 0.264 mm
Character Height Large: 11.03 mm (0.434 in)
Medium: 7.34 mm (0.289 in)
Small: 3.64 mm (0.143 in)
Viewing Angle >160° Horizontal and Vertical
Contrast Ratio >45 (“On” pixel luminance/“Off” pixel luminance)
Display Color Amber
Display Background Color Black
“On” Pixel Luminance >9.0 fL (area of amber pixel; includes protective window)
“Off” Pixel Luminance <0.2 fL (black pixel)
Refresh Rate 109 Hz
Monitor (Environmental)
Monitor Environmental Specifications
Characteristic Specification
Operating Temperature 0° to 40° C
Shipping and Storage Temperature -20° to 60° C
Operating Altitude -2,000 to 15,000 ft (-610 to 4,572 m)
Shipping and Storage Altitude -2,000 to 40,000 ft (-610 to 12,192 m)
Operating Relative Humidity 15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural
Shipping and Storage Relative 15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural
Humidity
Shock 50 g
Vibration, Random 0.02g2/Hz from 10 to 500 Hz, ramping down to 0.002g2/Hz at 2000 Hz.
Operating 1 hour per axis, 3 hours per test. Designed to meet RTCA
DO-160C, Category C.
Vibration, Sinusoidal 0.10 peak to peak inches 5 to 17 Hz, sloping to 0.01 peak to peak
inches at 55 Hz, then sloping to 0.0001 peak to peak inches at 2000
Hz. Operating 1 hour per axis, 3 hours per test. Designed to meet
RTCA DO-160C, Category N.
Electromagnetic Compatibility (EMC) Per IEC 601-1-2, which is a collateral standard of IEC 601-1, for
electromagnetic compatibility. Designed to meet RTCA DO-160C,
Section 21, Category B. Propaq 200 Series SN EA000225 and higher:
MIL-STD 461D:CE102, CS101, CS114, CS115, CS116, RE102,
RS103 (United States Air Force).
Monitor (Physical)
Monitor Physical Specifications
Characteristic Specification
Protection Classifications, all Configurations a
Type of Protection against Electric Power adapter class 1
Shock:—Power Adapter
Type of Protection against Electric Protective earth not available in monitor. Monitor designed and tested
Shock—Monitor (when connected to to meet Double Insulation Requirement.
power adapter or powered by internal
battery)
Degree of Protection Against Electric See monitor labels
Shock, for Parts Applied to Patients
Method of Disinfection Not suitable for autoclaving b
Flammable Anesthetics Not suitable for use with flammable anesthetics
Monitor Only
Height 6.65 in (16.9 cm)
Width 8.25 in (20.9 cm
Depth 5.10 in (12.9 cm)
Weight 6.25 lb (2.8 kg)
Monitor with SpO2 Module
Height 6.65 in (16.9 cm)
Width 8.25 in (20.9 cm)
Depth 7.50 in (19.10 cm)
Weight 9.12 lb (4.10 kg)
Monitor with Expansion Module
(Printer / SpO2 / CO2)
Height 9.65 in (24.5 cm)
Width 8.25 in (20.9 cm)
Depth 7.56 in (19.2 cm)
Weight with Printer, SpO2, and CO2 13.5 lb (6.1 kg)
a.Per EN 60601-1 unless otherwise stated.
b.See Chapter 7 for cleaning instructions.
Printer
Printer Specifications
Characteristic Specification
Operation
Operating Modes Continuous, Snapshot, Freeze Print, Auto Print, Auto Trend, Tabular
Trend, Alarm Print, NIBP Ticket, Apnea Ticket, OxyCRG, OxyCRG on
Alarm
Auto Print Intervals 15 min, 30 min, 1 hour, 2 hours, 4 hours
Auto Trend Shifts Once every 4 hours
Number of Waveforms Up to three: ECG, P1, P2, SpO2, CO2, RESP
Grid 5 mm and 1 mm gradations
Annotation Date, Time, Print Mode, Speed, Heart Rate, Systolic, Diastolic, Mean,
SpO2, Breath Rate, ETCO2, INCO2, Temperature, ∆T, Pacer Status,
Company Logo, ECG Bandwidth, Patient Mode, scale factors for all
traces and, if Acuity is connected, patient name and identification.
Printing Speeds 6.25, 12.5, 25.0 mm/sec,
simulated 6.25 mm/sec for CO2 and RESP in Snapshot mode
Printer Mechanism
Printing Method Thermally sensitive dot method
Dot structure 320 dots per line
Printing width 53 mm
Horizontal Dot Pitch 0.165 mm, 6 dots/mm
Vertical Dot Pitch 0.165 mm
Paper Feed Method Friction Feed
Paper Feed Precision ±2% @ 25° C and 60% Relative Humidity
Paper Width 60 mm
Reliability 30 million pulses/dot
Environmental
Monitor/Expansion Module
Operating Temperature +5° to 40° C
Shipping and Storage Temperature -20° to 60° C
Operating Relative Humidity 35% to 85% noncondensing
Shipping, Storage Relative Humidity 15% to 90% noncondensing
Operating Altitude -2,000 to 15,000 ft (-610 to 4,572 m)
Shipping and Storage Altitude -2,000 to 40,000 ft (-610 to 12,192 m)
Shock 30 g
Vibration, Random 0.02g2/Hz from 10 to 500 Hz, ramping down to 0.002g2/Hz at 2000 Hz.
Operating 1 hour per axis, 3 hours per test.
Characteristic Specification
Electromagnetic Compatibility (EMC) Per IEC 601-1-2, which is a collateral standard of IEC 601-1, for
electromagnetic compatibility.
Paper
Short-term Storage Environment -20 to 40°C; 5% to 80% noncondensing
(up to 7 days)
Long-term Storage Environment 25°C (optimal), 65% noncondensing
(up to 5 years)
Power
Monitor Power Specifications
Characteristic Specification
Mode of Operation Continuous
Battery Pack Type Sealed, gel-type lead acid
Battery Pack Capacity Monitor only: 8 volts, 3 Ampere-Hours; Monitor with Expansion
Modules: 8 volts, 6 Ampere-Hours
Battery Recharger Circuitry Internal, powered by external power adapter
DC Input Power Required 12 to 28 Volts, 25 Watts
Input Fuse Rating 3A/250V, Type 2AG (0.57x 0.177 in)
Operating Times on Battery
Power Adapters
Power Adapter General Specifications
Characteristic Specification
Protection Classifications, all Adapters a
Type of Protection Against Electric Class I, (Protectively Earthed)
Shock
Degree of Protection Against Harmful For ordinary, indoor locations only.
Ingress of Water
Method of Disinfection Not suitable for autoclaving b
Flammable Anesthetics Not suitable for use with flammable anesthetics
Environmental Specifications, All Adapters
Operating Temperature 0° to 50° C
Shipping and Storage Temperature -20° to 60° C
Operating Altitude -2,000 to 15,000 feet (-610 to 4,572 m)
Shipping and Storage Altitude -2,000 to 40,000 feet (-610 to 12,192 m)
Operating Relative Humidity 15% to 95%, noncondensing
Shipping, Storage Relative Humidity 15% to 95%, noncondensing
Shock 50 g
Vibration Random Vibration, 0.02g2/Hz from 10 to 300 Hz, ramping down to
0.002g2/Hz at 500 Hz. Operating 1 hour per axis,
3 hours per test.
Water Resistance For ordinary, indoor locations only.
a.Per EN 60601-1 unless otherwise stated.
b.See Chapter 8 for cleaning instructions.
Characteristic Specification
Universal Power Adapter, Part No. 503-0054-00, 503-0093-XX
Length 5.0 in (12.7 cm)
Width 3.6 in (9.1 cm)
Height 3.1 in (7.9 cm)
Weight 3.1 lb (1.4 kg)
Rated Input 100V-120V ac, 500 mA, 50/60 Hz
Rated Fuses T800 mA/250V, Time-Delay, 5x20mm
Rated Output (Continuous) 16-24V dc, 25 VA
Additional Features Detachable power cord, pilot light
Universal Power Adapter, Part No. 503-0054-01, 503-0092-XX
Length 5.0 in (12.7 cm)
Width 3.6 in (9.1 cm)
Height 3.1 in (7.9 cm)
Weight 3.1 lb (1.4 kg)
Rated Input 200V-240V ac, 250 mA, 50/60 Hz
Rated Fuses T400 mA/250V, Time-Delay, 5 x 20mm
Rated Output (Continuous) 16-24V dc, 25 VA
Additional Features Detachable power cord, pilot light
Power Adapter
+ connector
12-28V, 3A
Characteristic Specification
Datea MO/DA/YR, DA.MO.YR, or YR/MO/DA
Decimala . (Period)
HR/PR Sweep 25 mm/sec
RR/BR Sweep 6.25 mm/sec
Alarm Tone MEDIUM
HR/PR TONE LOW
HR/PR SOURCE ECG
RR/BR Source CO2 if ON or ECG (if CO2 OFF)
Patient Mode Adult
ECG Bandwidth Monitor
ECG Size 1 mV/cm
ECG Lead II
ECG Filtera 60 Hz
ECG Pacer ON
RESP size 2X
RESP lead Ld2
RESP sweep 6.25 mm/sec
RESP ON/OFF ON
RESP window ON
IBP Range 0 to 180 mmHg
IBP Rescale 0 to 140 mmHg
IBP Mode RESCALE
Invasive Pressure Formats Label dependent
NIBP Mode MANUAL
NIBP Auto Time 15 min
SpO2 SIZE 2x
SpO2 C-LOCK OFF
SpO2 Response NORMAL
TEMP F/Ca Celsius
CO2 Range 0 to 60 mmHg
CO2 Sweep 6.25 mm/sec
CO2 Response NORMAL
CO2 Unitsa mmHg
CO2 Gas Compensation OFF
Sidestream CO2 Flow Rate Adult: 90 ml/minute
Ped: 90 ml/minute
Neonate: 90 ml/minute
(The flow rate cannot be programmed to a different value in a Custom
Patient Mode. See Custom Patient Modes.)
Characteristic Specification
Display Wave Select All waves are on except NIBP
Trend Group NIBP
Alarm Limits All are ON except P2
HR Limits Adult: 50, 120 beats per minute
Ped: 50, 150 beats per minute
Neonate: 100, 200 beats per minute
NIBP Limits - Systolic Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg
NIBP Limits - Diastolic Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
NIBP Limits - Mean Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
P1, P2 Limits - Systolic Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg
P1, P2 Limits - Diastolic Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
P1, P2 Limits - Mean Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
SpO2 Limits Adult: 85%, 100%
Ped: 85%, 100%
Neonate: 80%, 95%
RR/BR Adult: 5, 30 BrM
Ped: 10, 45 BrM
Neonate: 10, 60 BrM
TEMP Limits 35.0°, 37.8° C
∆T Limits 0.0°, 2.8° C
ETCO2 Limits 25, 60 mmHg (3, 8 for % and kPa)
INCO2 Limits N/A, 5 mmHg (0.7 for % and kPa)
Apnea Delay Adult/Ped: 20 seconds
Neonate: 15 seconds
PRINTER SETTINGS
Printer Alarm Print OFF
Printer Auto Print OFF
Printer NIBP Ticket OFF
Printer Apnea Ticket ON
Printer Print Speed 25 mm/sec
Printer Auto Trend OFF
Printer Trend Selections NIBP and P1 = ON; all others = OFF
Printer OxyCRG on Alarm OFF
a.Any time you change the Date, Filter, Temp F/C, Decimal, or CO2 Units setting, the new setting also becomes
the powerup default setting.
∆T
Difference temperature. The difference between T1 and T2.
AAMI
Association for the Advancement of Medical Instrumentation (United States of America).
AC Power Adapter
The device that plugs into the 12-28V dc receptacle on the Propaq Encore's side panel to
allow operation and battery charging from ac mains.
Acuity
Welch Allyn’s trade name for its central station patient monitoring system.
Altimeter
A sensor, internal to the Propaq Encore, that measures absolute atmospheric pressure, and is
used to correct CO2 numerics for varying altitudes.
ANSI
American National Standards Institute
Apnea
Condition of no respiration occurring during a prescribed time interval.
ART
Arterial (label for an invasive blood-pressure channel).
Artifact
An unwanted disturbance to or by the patient or attached sensors that adds errors (usually
erratic) to the measured parameters, e.g., muscle motion or shivering, electrical interference,
vibration of the cuff, etc.
Bell
The symbol that appears in a window to indicate alarm limits status. If alarm limits have
been set, a bell appears.
BP
Blood pressure
bpm
Beats per minute
BR
Breath rate, expressed in units per minute or 1/min. BR is derived from CO2. See also RR.
Buttons
The five buttons along the bottom-front of the Propaq Encore. A menu appears above each
button identifying what each button will do when pressed.
Capnogram
Hard copy of the ETCO2 waveform over time.
Capnometer
Analyzer used to measure CO2, specifically ETCO2.
Channel
See Patient Channel.
C-Lock
A processing scheme used in SpO2 that uses QRS timing to improve the noise tolerance of
SpO2 measurements.
Configuration
The patient channels included with each Propaq Encore model. A table in Chapter 1 lists the
configuration of each Propaq Encore model.
Cursor
The highlighted block in a status window that indicates the selection you make by pressing
the NEXT button.
CVA
Cardiovascular artifact.
CVP
Central venous pressure (label for an invasive blood pressure channel).
DC Offset
The DC voltage difference between ECG electrodes. DC offset is caused by using dried out
electrodes or electrodes of dissimilar metal types.
Difference Temperature
The difference between T1 and T2. Also called delta T (∆T).
Digital Filter
A computer program in the Propaq Encore that removes unwanted noise that can be induced
into the ECG signal from ac mains.
EL (Electroluminescent) Display
The display screen used in the Propaq Encore.
EMI
An acronym for Electromagnetic Interference.
Endotracheal Tube
Plastic breathing tube placed into the patient’s windpipe.
Error Message
The message that appears when the monitor detects a malfunction requiring factory service.
Error Number
The number that identifies a problem encountered during operation.
ESD
An acronym for Electrostatic Discharge (from static electricity).
ESIS
An acronym for Electrosurgery Interference Suppression.
ETCO2
An acronym for end-tidal CO2. Amount of CO2 breathed out at the end of an exhalation.
Freeze
The action taken by the FREEZE button to stop the display. If three waveforms are displayed,
all waveforms are frozen. If less than three waveforms are displayed, the current waveforms
are frozen and the top waveform is also shown in real-time. See also Unfreeze.
Highlight
The method of identifying a selected item on the display. Highlighted selections appear as
light characters on a dark background or dark characters on a light background. See also
Cursor.
HR
An acronym for heart rate and expressed in units per minute or 1/min. This is displayed
when the heart rate/pulse rate source is ECG.
ICP
Intracranial pressure (label for an invasive blood pressure channel).
Impedance Pneumography
A method of detecting respiratory effort by measuring the AC impedance between selected
ECG leads.
INCO2
An acronym for inspired CO2. The amount of CO2 measured during inhalation.
Labels
The names appearing above the buttons.
Mainstream
A respiratory CO2 measurement technique which uses a noninvasive sensor located at the
endotracheal tube. This technique avoids signal delays and fluid problems associated with
other techniques.
Menu
A group of labels above the bottom front row of buttons on a Propaq Encore.
Numerics
The numbers that appear along the top and right side of the display for heart rate, blood
pressure, temperature, etc.
OxyCRG
An oxygen cardiorespirogram, a graph showing heart rate, SpO2, and a condensed
respiratory waveform.
P1
A generic label for invasive pressure channel one.
P2
A generic label for invasive pressure channel two.
Parameter
See Vital Sign Parameter.
Patient Alarm
The condition that exists when a vital sign parameter numeric violates an alarm limit.
Patient Channel
ECG, P1, P2, T1, T2, SpO2, CO2, NIBP and RESP.
Patient Mode
Selects Adult, Pediatric, or Neonatal mode settings for the monitor. These settings determine
default alarm limits, maximum cuff inflation pressure, and other internal settings.
Pinout
The signal descriptions for each pin of a connector.
Polarization
The activity that occurs when dissimilar metals between ECG electrodes and leads meet. This
can cause dc offset and other signal problems.
PR
Pulse rate, expressed in units per minute or 1/min. This is displayed when the heart rate/
pulse rate source is from a pressure channel or SpO2.
Pulse Rate
The heart rate determined from either a pressure channel, SpO2, or NIBP; expressed in units
per minute or 1/min.
Pushbutton
See Buttons.
Range Mode
The method used in invasive pressure display to show two waveforms against the same
pressure scale.
Respiration
The exchange of oxygen and carbon dioxide in the lungs and with the cells of the body.
RR
Respiration rate, a measure of the frequency of respiration. See also Impedance
Pneumography.
Sensors
The electrodes, transducers, probes, etc. used to obtain patient information.
Serial Number
The unique number assigned to the monitor. It is located on the rear panel label.
Sidestream
A respiratory CO2 measurement technique which can be used for intubated or non-intubated
patients.
SpO2
The standard term assigned to measuring oxygen saturation using a pulse oximeter. The
SpO2 patient channel noninvasively measures oxygen saturation of arteriolar hemoglobin at
a peripheral measurement site, such as a finger, toe, or the bridge of the nose.
Startup Window
The information window that appears while the monitor performs its powerup test just after
you turn on the Propaq Encore. This information includes the Propaq Encore model number
and software version number.
Status Window
A window that appears and contains information about the Propaq Encore.
• A digital output pulse from the right side panel that starts within 35 msec of the peak
of a QRS complex and is used for cardioversion.
Trend
The accumulation of several hours of data at two-minute intervals.
Trend Parameter
Heart Rate/Pulse Rate, P1, P2, SpO2, INCO2, ETCO2, temperature, NIBP and RR.
Turbocuf Mode
The mode used to acquire as many NIBP measurements as possible in five minutes.
UA
Umbilical artery (label for an invasive blood pressure channel).
Unfreeze
Returns the waveforms to active display. See also Freeze.
UV
Umbilical vein (label for an invasive blood pressure channel).
Waveform Window
The area in which waveforms are displayed.
Waveform/Status Window
See Waveform Window or Status Window.
YSI
An acronym for Yellow Springs Instrument Company.
Zeroing
The process by which an invasive pressure zero reference is obtained so that pressures can be
related to atmospheric pressure. This process also nulls out any residual pressure indicated
by a transducer with zero pressure applied.
SpO2 70, 71
Waveforms
display priorities 31
ECG 31
scales 54, 81
size 45
Windows
Propaq screen 24
Z
ZERO (P1/P2) buttons 54
Zeroing
messages 51
transducers 51