Manual 772-773
Manual 772-773
Manual 772-773
Page 1
772/773
Milliamp Process Clamp Meter
Instruction Sheet
Introduction
The hand-held, battery-operated Fluke 772 and 773 Milliamp Process
Clamp Meters (the Meter or Product) can be used in troubleshooting
transmitters, valves, PLC and DCS I/O. Unlike conventional clamp meters,
the Meter features a remote jaw that is connected to the main body via
extension cable.
Features
• In-circuit measurement of 0 mA to 24 mA dc and up to 99.9 mA dc
using a remotely connected clamp via extension cable
• 0 mA to 24 mA dc sourcing and simulating
• 0 V to 10 V dc sourcing (773)
• Loop power supply 24 V dc output
• 0 V to 30 V dc measurement (773)
• Scaled mA output (773)
• Simultaneous mA measurement via detachable clamp and mA
sourcing (773)
• 250 Ω HART resistor for mA source
• Electronic zero
• Percentage span (0 % to 100 %)
• Hold
• Auto power off (battery saver)
• Display backlight
• Measurement spotlight LED
PN 3351049
February 2009 Rev. 1, 5/19
© 2009-2019 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies. Specifications are subject to
change without notice.
772/773 English Instruction Sheet
Page 2
Contacting Fluke
To contact Fluke, call one of the following telephone numbers:
• Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
• Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31 402-675-200
• Japan: +81-3-3434-0181
• Singapore: +65-6799-5566
• China: +86-400-921-0835
• Brazil: +55-11-3530-8901
• Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.
Table 1 explains the symbols that are used on the Meter or in this
Instruction Sheet.
Table 1. Symbols
Symbol Explanation
Battery
mA I N / OUT
HA RT
Fjv06.emf
Number Description
Turn on and off the Meter
Measurement spotlight LED button
Turns the display backlight on and off
Switches the Meter to Clamp Measure mode. Zeros the clamp
reading in Clamp mode. The Clamp modes includes clamp
measure, mA scale output, and mA IN/OUT.
Press first to change to mA scale (773).
Cycles through source output ramping and 25 % stepping:
() Slow repeating 0 % - 100 % - 0 % ramp
() Fast repeating 0 % - 100 % - 0 % ramp
() Repeating 0 % - 100 % - 0 % ramp in 25 % steps
Press first to activate mA IN/OUT (773).
Captures and holds the current reading. Pressing first
activates the 250 Ω HART resistor.
activates features listed above some buttons
0 %-100 %- sets voltage or mA sourcing output. Press first to
activate , , , and to adjust the source output. Long press
or to set span range point.
Measure, Source, Simulate button
DC Volts selection (773)
mA selection. Press first to activate Loop Power function.
Figure 1. Buttons
772/773 English Instruction Sheet
Page 6
fjv05.emf
Number Description
Detachable clamp
Tactile Barrier docked and un-docked. Refer to Safety Information
and Symbols.
Display
Measurement spotlight LED
Figure 2. The Milliamp Process Clamp Meter
772/773 English Instruction Sheet
Page 7
MAX
Fjv07.emf
Number Description
Main display values
HOLD is activated
Clamp is active
Test lead jack indicator. Test lead connection is required.
HART 250 Ω resistor is engaged
Shift is active
Reading is scaled
Loop Power is active
Milliamps
Volts dc
Percentage
Secondary display
Ramping is engaged
Low battery symbol
Maximum voltage warning
High voltage is present
Measure, Source, or Simulate is active
Figure 3. Display (773 shown)
772/773 English Instruction Sheet
Page 8
Fjv04.emf
Number Description
Voltage measurement test lead input, also used for voltage
sourcing HI.
Common test lead input, also used for voltage sourcing LO.
-mA test lead input, also used for mA sourcing.
+mA test lead input, also used for mA sourcing.
Common test lead input. -mA test lead input. Also used for mA
sourcing.
Figure 4. Input/Output Jacks
Features
The following sections give more detail about the Meter’s features.
Percentage Span
The Source and Simulate Percentage Span feature displays the span for
4 to 20 mA loops. Use , , , and to adjust the source
or simulated current (772) or dc voltage and current (773).
20 mA 100 % 8 mA 25 %
16 mA 75 % 4 mA 0%
12 mA 50 % 0 mA -25 %
Zero Adjust
Before taking measurements with the clamp, push to zero the
display by removing the offset. Make sure the clamp jaws are closed and
no current is flowing through them before zeroing.
Backlight
Press to turn the backlight on and off. The backlight automatically
turns off after 2 minutes.
772/773 English Instruction Sheet
Page 9
User Options
Several user options can be activated at Meter power up. Hold
when powering on the Meter. While holding down , toggle on/off
each option by repeatedly pressing the following keys:
• toggle on/off backlight auto off. Display shows bLit on or oFF.
• toggle on/off spotlight auto off. Display shows SLit on or oFF.
• toggle on/off auto power off. Display shows PoFF on or oFF.
When all key are released, the software version appears and the Meter
enters Clamp Measure mode.
Measurement Spotlight LED
The Measurement Spotlight LED helps to quickly find mA signal wires.
Press to activate it. To extend battery life, the light automatically turns
off after 2 minutes.
Display HOLD
XW Warning
To prevent possible electrical shock, fire, or personal
injury:
• Be aware of the measurement being taken when using
Display HOLD. When Display HOLD is activated, the
display will not change when different currents are
applied.
• Do not use the HOLD function to measure unknown
potentials. When HOLD is turned on, the display does
not change when a different potential is measured.
Press to activate Display Hold mode. The display shows and
the display freezes. To exit and return to normal operation, press a
second time. When in Auto Ramping mode, stops ramping.
Probe Holder
The Meter is equipped with a probe holder that can either hold a test
probe or can be used to attach the Fluke ToolPak. See Figure 5.
Fjv08.emf
Taking Measurements
XW Warning
To prevent possible electric shock, fire, or personal injury,
do not use the clamp on non-insulated conductors.
Measurements can be taken with the clamp in the docked position,
remotely using the 1 m cable, or via test leads. For accurate
measurements:
• Always zero the Meter prior to taking measurements with the clamp.
• To reduce magnetic influences, zero the Meter as close to the
measurement in the same position or jaw direction that is used for
the measurements as possible.
• Make sure the clamp is free of contamination.
To use the clamp for measurements:
1. Press to enter Clamp Measure mode and to zero the Meter.
Clamp mode includes clamp measure, mA scale output, and mA
IN/OUT. If necessary, press to change to mA scale.
2. Clamp the jaw around the conductor under test. The Meter displays
the measured conductor current. See Figure 6.
• A positive reading indicates current flowing in the direction of
the arrow on the clamp.
• A negative reading indicates current flowing in the opposite
direction of the arrow.
• Do not clamp more than one wire.
The small secondary display shows the reading in mA percentage of
span.
772/773 English Instruction Sheet
Page 11
OK
fjv03.emf
mA I N/ OUT
HA R T
fjv09.emf
Sourcing mA
Use mA Source mode whenever it is necessary to source current into a
passive circuit such as a current loop with no loop supply. Source mode
depletes the battery faster than Simulate mode.
To enter Source mode for the 772, see Figure 4:
1. Insert the test leads into the -mA and +mA jacks.
2. Press .
Fjv10.emf
Simulating mA Output
In Simulate mode, the Meter simulates a current loop transmitter. To
enter Simulate mode, see Figure 9:
1. Insert the test leads into the +mA and -mA input jacks.
2. Press .
Fjv11.emf
Loop Supply
In Loop Supply mode, the Meter powers a transmitter while measuring
the mA signal. To enter Loop Supply mode, see Figure 10:
1. Insert the test leads into the LOOP PWR jacks. See Figure 10.
2. Press .
3. Press .
The Meter is now in Loop Supply mode.
Fjv13.emf
Maintenance
XW Warning
To prevent possible electric shock, fire, or personal injury:
• Remove the input signals before you clean the
Product.
• Repairs or servicing not covered in this manual
should be performed only by qualified personnel.
• Replace all batteries with fresh batteries of the same
manufacturer and type to prevent battery leakage.
Battery Replacement
XW Warning
To avoid false readings, that could lead to possible electric
shock or personal injury, replace the batteries as soon as
the low battery indicator () appears.
To replace the batteries, see Figure 11:
1. Turn the Meter off.
2. Use a flat head screwdriver to loosen the battery compartment door
screw, and remove the door from the case bottom.
3. Remove the batteries.
4. Replace the batteries with four new AA batteries.
5. Reattach the battery compartment door to the case bottom and tighten
the screw.
fjv02.emf
Specifications
Electrical Specifications
DC Current Measurement
With Jaw
Ranges ............................................... 0 mA to 20.99 mA;
21 mA to 100 mA
Resolution........................................... 0.01 mA, 0.1 mA
Accuracy ............................................. 0.2 % + 5 counts, 1 % + 5 counts
In Circuit
Range ................................................. 0 mA to 24 mA
Resolution........................................... 0.01 mA
Accuracy ............................................. 0.2 % + 2 counts
Current Source
Range ................................................. 0 mA to 24 mA
Resolution........................................... 0.01 mA
Accuracy ............................................. 0.2 % + 2 counts
mA Drive ............................................. 24 mA into 1000 Ω
Current Simulate
Range ................................................. 0 mA to 24 mA
Resolution........................................... 0.01 mA
Accuracy ............................................. 0.2 % + 2 counts
Maximum Voltage ............................... 50 V
DC Voltage Measurement (773)
Range ................................................. 0 V to 30 V
Resolution........................................... 0.01 V
Accuracy ............................................. 0.2 % + 2 counts
DC voltage source (773)
Range ................................................. 0 V to 10 V
Resolution........................................... 0.01 V
Accuracy ............................................. 0.2 % + 2 counts
mA Drive ............................................. 2 mA max all conditions
mA IN/OUT (773)
Sourcing range ................................... 0 mA to 24 mA
Sourcing resolution ............................. 0.01 mA
Sourcing accuracy .............................. 0.2 % + 2 counts
Measurement range ........................... 0 mA to 24 mA
Measurement resolution ..................... 0.01 mA
Measurement accuracy ...................... 1 % FS
Scaled mA current output to mA current input from the Jaw (773)
Range ................................................. 0 mA to 24 mA
Resolution........................................... 0.01 mA
Accuracy ............................................. 1 % FS
772/773 English Instruction Sheet
Page 16
Environmental Specifications
Operating Temperature ............................... -10 °C to 50 °C
Storage Temperature .................................. -25 °C to 60 °C
Operating Humidity ..................................... <90 % RH @ <30 °C
<75 % RH @ 30 °C to 50 °C
Operating Altitude ....................................... 0 m to 2000 m
IP Rating ..................................................... IP 40
Temperature Coefficients ............................ 0.1 (/ °C X Specified accuracy for
Temperature <18 °C or >28 °C)
Safety .......................................................... IEC 61010-1, Pollution degree 2
IEC 61010-2-032: O, other circuits
not directly connected to mains.
Electromagnetic Compatibility (EMC)
International ................................................ IEC 61326-1: Portable
Electromagnetic Environment
IEC 61326-2-2
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is necessary for the
internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than
domestic and those directly connected to a low voltage power supply
network that supplies buildings used for domestic purposes. There
may be potential difficulties in ensuring electromagnetic compatibility
in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential
environments and may not provide adequate protection to radio
reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur
when the equipment is connected to a test object.
The equipment may not meet the immunity requirements of this
standard when test leads and/or test probes are connected.
For current measurement with jaw, add 1 mA to specification for EMC
fields from 1 V/m to 3 V/m.
772/773 English Instruction Sheet
Page 17
Miscellaneous Specifications
Power Requirements................................... Four AA batteries, Alkaline,
IEC LR6
Automatic Time-out (Power)........................ 15 minutes ±1 minute
Automatic Time-out (Backlight) ................... 2 minutes ±10 seconds
Automatic Time-out
(Measurement Spotlight) ............................. 2 minutes ±10 seconds
11/99