849 Level Control: Manual
849 Level Control: Manual
Manual
8.849.8001EN / 2019-09-30
Metrohm AG
CH-9100 Herisau
Switzerland
Phone +41 71 353 85 85
Fax +41 71 353 89 01
[email protected]
www.metrohm.com
Manual
8.849.8001EN / 2019-09-30
Technical Communication
Metrohm AG
CH-9100 Herisau
[email protected]
Table of contents
1 Introduction 1
1.1 Instrument description ......................................................... 1
1.2 Model versions ...................................................................... 1
1.3 Intended use ......................................................................... 2
1.4 About the documentation ................................................... 2
1.4.1 Symbols and conventions ........................................................ 2
1.5 Safety instructions ................................................................ 3
1.5.1 General notes on safety ........................................................... 3
1.5.2 Electrical safety ........................................................................ 3
1.5.3 Flammable solvents and chemicals ........................................... 4
1.5.4 Recycling and disposal ............................................................. 4
3 Installation 8
3.1 Configuration ........................................................................ 8
3.1.1 Basic setting ............................................................................ 8
3.1.2 Overall status ........................................................................... 9
3.1.3 Selective operating mode ........................................................ 9
3.2 Connecting the sensor ....................................................... 10
3.3 Connecting the remote cable ............................................ 11
5 Technical specifications 15
5.1 Measuring inputs ................................................................ 15
5.2 Remote socket .................................................................... 15
5.3 Supply .................................................................................. 16
5.4 Ambient temperature ......................................................... 16
6 Accessories 17
Index 18
Table of figures
Figure 1 Front 849 Level Control ..................................................................... 5
Figure 2 Rear 849 Level Control ...................................................................... 5
Figure 3 6.1113.000 sensor ............................................................................ 6
Figure 4 6.1113.100 sensor ............................................................................ 6
Figure 5 6.1113.110 sensor ............................................................................ 7
Figure 6 Printed circuit board .......................................................................... 8
Figure 7 DIP switch strip ................................................................................. 8
Figure 8 Wiring the sensor ............................................................................ 10
Figure 9 Bottle adapter ................................................................................. 10
Figure 10 Remote connection 849 Level Control - Sample Processor with remote
interface .......................................................................................... 11
Figure 11 Remote connection 849 Level Control - Metrohm instrument with
remote box ..................................................................................... 12
1 Introduction
2.849.0010
849 Level Control with sensor, comprised of three steel rods (stainless
steel, non-acid-resistant, 35 cm long).
For aqueous media.
2.849.0020
849 Level Control with sensor, comprised of three metal-free rods (mate-
rial: carbon fiber compound, 35 cm long, unbreakable)
For eluents in chromatography. Can be used in 6.1621.000 20 L canisters.
Also integrated into the screw cap are two FEP tubes for aspirating or fill-
ing the container.
2.849.0030
849 Level Control with sensor, comprised of three metal-free rods (mate-
rial: high-purity graphite).
For eluents and ultra pure water in chromatography.
Standard length 25 cm for 2 L bottles. The sensor rods can be shortened
to the required lengths suitable for 1 L or 0.25 L bottles (predetermined
breaking points).
Also integrated into the screw cap are two FEP tubes for aspirating or fill-
ing the container.
CAUTION
Warning
This symbol draws attention to a possible hazard due
to electrical current.
Warning
This symbol draws attention to a possible hazard due
to heat or hot instrument parts.
Warning
This symbol draws attention to a possible biological
hazard.
Caution
This symbol draws attention to a possible damage of
instruments or instrument parts.
Note
This symbol marks additional information and tips.
WARNING
WARNING
WARNING
WARNING
1 2 3 4
Figure 1 Front 849 Level Control
1 Sensor connector 1 for overflow moni- 2 Sensor connector 2 for overflow moni-
toring (Level Full) toring (Level Full)
Signal at contact Signal at contact
3 Sensor connector 1 for empty level 4 Sensor connector 2 for empty level
monitoring (Level Empty) monitoring (Level Empty)
Signal at contact loss Signal at contact loss
The connectors for the sensor cables (6.2151.060) are located on the
front of the 849 Level Control.
1
Figure 2 Rear 849 Level Control
1 Remote connector
For the 6.2125.090 cable
The 6.2125.090 connection cable providing power supply for the conduc-
tivity measurement is connected on the rear. It is connected to the remote
socket of a suitable Metrohm instrument in order to transmit the activat-
ing signals of the sensors
2.1 Sensors
2.1.1 6.1113.000 sensor for aqueous media
The 6.1113.000 sensor can be used for monitoring fluid levels in rinse or
waste containers (e.g. 20 L canisters). It is suitable exclusively for conduc-
tivity measurements in aqueous media.
R
E F
E F
E F
3 Installation
3.1 Configuration
3.1.1 Basic setting
The logic of the 849 Level Control must be configured before it is used for
the first time. The operating mode can be set with DIP switches.
Remove the two fastening screws on the rear side of the housing and
then carefully pull the electronic print located there out of the housing to
the rear.
ON
OFF
7 Level Empty 2
Example: Example:
Monitoring overflow and empty level (incl. Monitoring overflow in two containers (incl.
overall status) overall status)
F R
F Full level
Sensor rod for detecting the maximum fill
level
One sensor cable has two lines and two narrow test plugs at one end.
Always connect the red plug to the connector R (= Reference, see
below). The reference connection is always linked with a long sensor rod.
R E F
E F
6.1113.000 6.1113.100/6.1113.110
Figure 9 Bottle adapter
NOTICE
The extra opening in the bottle adapter is used for ventilating the con-
tainer. This opening may not be sealed. The pressure equilibrium in the
container must always be ensured.
Overflow monitoring
In the case of overflow monitoring, connect the black plug of the sensor
cable to the connector F (Level Full). The Full level connector must be
linked with a short sensor rod (activation upon contact).
Ext. Ext.
Pump 1 Pump 2
Swing
Head
Made by Metrohm
Herisau Switzerland
Remote
1.838.0010
01107
Type:
Nr.:
6.2125.09
0
Figure 10 Remote connection 849 Level Control - Sample Processor
with remote interface
If the remote lines are to continue to be available for the purpose of con-
trolling a different device, then a 6.2125.000 stackable plug and an addi-
tional 6.2125.090 connecting cable are to be used.
6.2125.09
0
Figure 11 Remote connection 849 Level Control - Metrohm instrument
with remote box
The 6.2148.010 remote box is also equipped with a fully functioning MSB
connector, which means that the complete number of MSB connectors
continues to be available for use.
Input 76543210
Bit 76543210
(The bits are always numbered from right to left)
Example:
*****0** expects an inactive input line 2 (0=not activated). This
line is set by the 849 Level Control, in the event that a
sensor signal has been set.
Input lines that are of no interest or for which no defined condition can be
predicted should also be masked with an asterisk (*).
4.2 Maintenance
The 849 Level Control is maintenance-free to a large extent.
If one of the stainless steel sensors exhibits traces of corrosion, then the
sensor rod can be rubbed clean with a cloth and a conventional cleaning
agent.
5 Technical specifications
5.3 Supply
Voltage 5.0 V
Power consump- < 15 mA
tion
6 Accessories
Up-to-date information on the scope of delivery and optional accessories
for your product can be found on the Internet. You can download this
information using the article number as follows:
NOTICE
Index
A G Remote cable ........................... 11
Aqueous media .......................... 6 Graphite ..................................... 7 Remote connector ...................... 5
Remote interface ...................... 13
B I Remote socket ......................... 15
Basic setting ............................... 8 Input line .................................. 13
Bit sample ................................ 13 S
M Safety instructions ...................... 3
C Measuring inputs ...................... 15 Sample Processor ............... 11, 12
Carbon fiber compound ............. 6 Model versions ........................... 1 SCAN command ....................... 13
Conductivity measurement ......... 6 Monitoring ................................. 8 Screw cap adapter .................... 10
Configuration ............................. 8 MSB connector ......................... 12 Selective operating mode ........... 9
Corrosion ................................. 14 Sensor ........................................ 6
N Sensor connection ...................... 8
D Non-aqueous media ................... 6 Sensor connector ....................... 5
Dimensions ................................ 6 Sensor rod ................................ 10
DIP switches ............................... 8 O Service ....................................... 3
Overall status .............................. 9 Stackable plug .......................... 12
E Overflow .................................... 9 Stainless steel ............................. 6
Electronic print ........................... 8 Overflow monitoring ................ 11 Supply ...................................... 16
Electrostatic charge .................... 4
Empty level ............................... 10 P T
Empty level monitoring ............. 11 Professional IC .......................... 12 Test plugs ................................. 10
Titrando ................................... 12
F R
Fill level .................................... 10 Reference ................................. 10 U
Full level connector ................... 10 Remote box .............................. 12 Ultra pure water ......................... 7