Firmware Update For The BARBIERI Spectrophotometers Spectro 50xy, LFP, Swing, 40xy
Firmware Update For The BARBIERI Spectrophotometers Spectro 50xy, LFP, Swing, 40xy
BARBIERI electronic OHG, I.-Seidner-Straße 35, I-39042 Brixen (BZ), Tel. +39 0472 834024 Fax +39 0472 833845
V.A.T.-Id. No: IT 01402900219 e-mail: [email protected] http://www. BARBIERIelectronic.com
Firmware Update
for the BARBIERI Spectrophotometers
Spectro 50xy, LFP, Swing, 40xy
January 2008
Contents
Scope of this document .............................................................................................. 1
Hardware and software needed.................................................................................. 2
Programming the program memory ............................................................................ 3
Re-setting instrument configuration ............................................................................ 5
Run the Gateway Setup program to install it on the computer. See also the operating
manual of the software.
Warning: Do not under any circumstance power off or restart the system during the
firmware update procedure.
- Run the Gateway software and configure the connection with your instrument.
- When using Gateway version 2, click on the Icon on the left side in the title
bar. With Gateway version 3 choose the menu: Extras
- You will be asked if you are sure to proceed. Confirm this message.
- The computer will start updating the flash memory block by block. The process
will take a few minutes.
Note: if the firmware update process is interrupted or fails, the instrument remains
without valid firmware. It then runs only in a special mode named “bootrom”
mode. When instrument is in this mode, please retry again to write the firmware
into the instrument.
You might see the message “instrument not connected” when starting the
Gateway software, but please go ahead with above firmware update instructions.
Note: use only the specific configuration file of your device supplied by BARBIERI
electronic. Using a different file will result in wrong functionality of your instrument!
- Run the Gateway software and configure the connection with your instrument.
- When unsing Gateway version 2, click on the Icon on the left side in the title
bar. With Gateway version 3 choose the menu: Extras
- A progress bar will now indicate that the configuration takes place. This
process will take a few minutes.