Iq5 Installation Guide
Iq5 Installation Guide
iQ 5 ™
Section 1
Safety Information
Important: Read this information carefully before using the iQ5 System.
Grounding
Always connect the iQ5 optics module to a three-prong, grounded AC outlet using the
AC power cord provided with the system. Do not use an adaptor to a two-terminal
outlet. Always ensure power switch is set to the off position when connecting or
disconnecting power cords for the iQ5 system.
Handling
Handle all components of the iQ5 system with care and with clean, dry hands at all
times. The optical system contains mirrors and lenses that may shatter if the unit is
dropped or struck with great force. If the unit is damaged, such that internal
components or wires are exposed, contact your local Bio-Rad office immediately, do
not attempt to repair or power on the instrument.
Service
The only user-serviceable parts of the iQ5 optics module are the lamp and filters. Call
your local Bio-Rad office for all other optics module and thermal cycler related service.
When replacing the lamp or filters, open only the outer casing of the iQ5 optics module.
The camera lamp may get extremely hot during system operation, do not attempt to
remove the lamp without powering off the instrument and allowing the system to cool
for at least 15 minutes. To prevent skin burns and fire hazards, do not attempt to
operate the iQ5 system while the camera case is open. Do not open the casing of
the iQ5 optics module when the instrument is in use.
Temperature
For normal operation, the maximum ambient temperature should not exceed 40°C
(see Appendix A for specifications). To ensure adequate cooling of the system,
maintain a clearance of at least 4 inches around the sides of the iQ5 optics module.
Do not block the fan vents near the lamp, as this may lead to improper operation or
cause physical damage to the iQ5 detector. Do not operate the iQ5 optics module in
extreme humidity (>90%) or where condensation can short internal electrical circuits or
fog optical elements.
1
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 2
Notice
This Bio-Rad instrument is designed and certified to meet EN-61010 safety standards.
EN-61010-certified products are safe to use when operated in accordance with the
instruction manual. This instrument should not be modified in any way. Alteration of
this instrument will:
• Void the manufacturer’s warranty
• Void the EN-61010 safety certification
• Create a potential safety hazard
Bio-Rad is not responsible for any injury or damage caused by the use of this
instrument for purposes other than those for which it is intended, or by modifications
to the instrument not performed by Bio-Rad or an authorized agent.
The iCycler® thermal cycler and iQ5 real-time PCR detection system are intended for
laboratory research applications only.
~
~
!
2
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 3
Section 2
iQ5 System Installation
2.1 Preparing for the Installation Process
The iQ5 system should be installed on a clean, dry, level surface. Identify an
appropriate work area for the installation process prior to unpacking any system
components. The entire installation process should take approximately 15 minutes
to complete.
System Checklist
❑ iQ5 Optics Module (170-9751), 1 box
❑ Power cord
❑ iQ5 optics module
❑ iQ5 software installation disk (170-9753)
❑ Amplification tech notes CD (version 1.0)
❑ iCycler Chassis (170-8701), 1 box
❑ Power cord
❑ iCycler chassis
❑ iCycler thermal cycler operating instructions
❑ iQ5 Optical Kit (170-9752), 1 box
❑ Optical reaction block
❑ Modified sliding rear cover for iCycler thermal cycler
❑ Serial cable
❑ USB cable
❑ Filter extraction tool
❑ Optical tape applicator tools (3)
❑ iQ5 optics support bracket
❑ Support bracket screws
❑ Hex driver
❑ Hex screws
❑ Spare part: halogen lamp
❑ iQ Calibration Kit (170-8792), shipped on dry ice
❑ External well factor solution
❑ 1x calibration dye solutions (9 dyes, 3 tubes of each dye)
Contact your local Bio-Rad office if any system components are missing or damaged.
In addition, the following accessories are required to complete the installation process:
• Scissors • Optical-quality sealing film or tube caps
• #2 Phillips screwdriver • Calibrated micropipet(s)
• Optical-quality PCR plates or tubes • Aerosol barrier pipet tips
3
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 4
Handle the iQ5 optics module and iCycler thermal cycler with care and with clean, dry
hands during unpacking and assembly. Do not handle the iQ5 system components
with wet hands. Refer to Section 1 of this guide for additional safety information.
2.2 Installing the Optics Module on the iCycler Chassis
1. Remove the existing rear
cover from the iCycler chassis
by sliding the cover towards
the front of the iCycler base
(Figure 2.1).
2. Install the modified sliding
rear cover provided with the
optical reaction block,
ensuring that the notch is
oriented towards the rear of
the iCycler thermal cycler.
Fig. 2.1. Slide the rear cover forward to remove it.
3. Push the sliding rear cover on
top of the chassis as far back
as possible.
4. Rotate the green latches
on the optical reaction block
up towards the open lid.
5. Lift the optical reaction block
by the handle and install it
onto the chassis. Lower the
front portion of the reaction
block so that it engages
with the chassis before the
Fig. 2.2. Engage front portion of reaction block with
rear portion. The rear of the chassis.
block lid should fit over the
front lip of the sliding rear cover (Figure 2.2).
6. Secure the optical reaction block in place by rotating the green latches down-
ward.
7. Close the optical lid.
4
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 5
5
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 6
6
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 7
7
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 8
Section 3
iQ5 Optical System Components
The optics module houses the excitation system and the detection system
(Figure 3.1). The excitation system consists of a fan-cooled, 50 W tungsten-halogen
lamp, a heat filter (infrared-absorbing glass), a six-position filter wheel fitted with five
optical filters and one opaque filter “blank”, and a dual mirror arrangement that allows
simultaneous illumination of the entire sample plate. The excitation system is physically
located on the right front corner of the optics module, with the lamp shining from
right to left, perpendicular to the instrument axis. Light originates at the lamp, passes
through the heat filter and a selected color filter, and is then reflected onto the 96-well
plate in the thermal cycler by a set of mirrors. This light source excites the fluorescent
molecules in the wells.
CCD
CCD lens
Excitation
Emitter filter wheel
filter wheel
IR filter
Lamp
Mirror
Lid lens
Heater glass
Sample
Fig. 3.1. Representation of the optical system layout. The detection system occupies the rear portion of
the optics module housing. The primary detection components include a six-position emission filter wheel
and a CCD detector. The emission filter wheel is identical to the wheel in the excitation system and is fitted
with five colored-emission filters and one opaque filter “blank”. Fluorescent light from the wells passes
through the emission filter and is then detected by the CCD.
8
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 9
Appendix A
iQ5 System Specifications
Input power 100–250 VAC ± 10%, 50–60 Hz, 250 W max, 95 W typical
Operating temperature 5–40°C
Storage temperature –20 to 40°C
Humidity 0–90%, noncondensing
Dimensions, optics module 29.3 x 38.74 x 20.3 cm (11.5 x 15.25 x 8.75")
(W x D x H)
Weight, optics module 6.3 kg (13.9 lb)
Data communication USB 2.0 high speed
Regulatory compliance EN-61010, NRTL
Appendix B
Computer Specifications
A computer with the following system specifications is required to run the
iQ5 real-time PCR detection system.
Minimum Computer Specifications
• Pentium or comparable processor, 1.0 GHz or faster
• Windows 2000 or Windows XP operating system
• Internet Explorer browser (version 5.0 or higher)
• 512 MB RAM
• 10 GB hard drive
• CD-ROM drive
• Mouse
• 1,024 x 768 x 256 colors minimum screen resolution
• USB 2.0 high-speed communications port
Recommended Computer Specifications
To ensure optimal performance of the iQ5 system and software, connect the iQ5
instrument to a computer with the following specifications:
• Pentium or comparable processor, 2.0 GHz or faster
• Windows XP operating system
• Internet Explorer browser (version 5.0 or higher)
• 512 MB RAM
• 20 GB hard drive
• CD-RW drive
• Mouse
9
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 10
Appendix C
Ordering Information
Catalog # Description
System Configuration
170-8701 iCycler Chassis, includes iCycler base, power cord, quick reference
cards, instruction manual
170-9750 iQ5 Optical System, includes optics module, software CD-ROM,
5 installed filter sets, 96-well reaction module, calibration solutions,
optical-quality 96-well PCR plates, communication cables, power cord,
instruction manual
Accessories
223-9441 96-Well 0.2 ml Thin-Wall PCR Plates, 25
223-9443 96-Well 0.2 ml PCR Plate Caps, for 223-9441, 300
223-9444 Optical-Quality Sealing Tape, optimized for use with 223-9441,
100 sheets
170-8756 Replacement Halogen Lamp
170-9753 iQ5 Software Installation Disk
170-9761 iQ5 Filter Set, FAM/SYBR dyes
170-9762 iQ5 Filter Set, VIC/HEX/TET dyes
170-9763 iQ5 Filter Set, Cy3/TAMRA dyes
170-9764 iQ5 Filter Set, ROX/Texas Red dyes
170-9765 iQ5 Filter Set, Cy5 dyes
Calibration Solutions
170-8792 iCycler iQ Calibration Kit, shipped on dry ice
170-8794 External Well Factor Solution
170-8793 FAM Calibrator Solution, 1.5 ml tubes, 5
170-8795 HEX Calibrator Solution, 1.5 ml tubes, 5
170-8797 Cy5 Calibrator Solution, 1.5 ml tubes, 5
170-8799 Texas Red Calibrator Solution, 1.5 ml tubes, 5
10
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 11
Appendix D
Care and Maintenance
Cleaning the Unit
Take care not to spill liquids onto or into the iCycler thermal cycler or the iQ5 optics
module.
The outer casing of the instrument may be cleaned using a soft, lint-free cloth
and water.
Replacing the Lamp
Before replacing the lamp, turn the instrument off and allow it to cool for at
least 15 min.
The lamp is located on the right side of the optics module, (Figure D.1).
To replace the lamp:
1. Turn off the power to the optics module.
2. Reach behind the iQ5 optics module and unscrew the short fastener that secures
the lid of the module in place.
3. Using a gentle pressure with both hands, push inward on the rear vents located
on the top half of the iQ5 casing. Lift upward to remove the cover of the optics
module.
4. Push up on the lamp spring
clip to release the lamp from
the bracket.
5. Lift the lamp out of the socket.
6. Install the new lamp using the
reverse of steps 1–5. Hold the
new lamp by the outer
reflector and do not touch the
bulb. Be sure the spring clip is
down before inserting the
lamp into the socket. Push the
lamp firmly into the bracket, Fig. D.1. Replacing the lamp.
then close the case and
secure the lid.
11
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 12
12
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 13
13
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 14
Appendix E
Installing New Versions of Firmware
Bio-Rad is committed to continuous improvement in iCycler features. Towards that
end, the iCycler firmware will be upgraded to offer new features on a regular basis.
We will make a firmware upgrade disk available at the time of each new release
and announce the release on our web site, discover.bio-rad.com. The iCycler
thermal cycler is upgraded via a serial port connection to a PC computer, or via a
serial port pass-through with the iQ5 USB connection. The computer serial port
must support a 57,600 baud rate. The connection may be made with a standard
9-pin serial cable. The iCycler firmware upgrade kit may be ordered from Bio-Rad
using part number 170-8737.
14
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 15
Appendix F
System Warranty
The iQ5 real-time PCR detection system is warranted against defects in materials and
workmanship. For specific warranty information, contact your local Bio-Rad office.
If any defects should occur during the warranty period, Bio-Rad will replace the
defective parts without charge. However, damage or defects resulting from any of the
following causes are specifically excluded:
1. Improper operation.
2. Use of improper solvent or sample.
3. Use with tubes, plates, or sealing materials not specified by Bio-Rad Laboratories
for use with the iQ5 real-time PCR detection system.
4. Deliberate or accidental misuse.
5. Repair or modifications done by anyone other than Bio-Rad Laboratories.
6. Natural disaster of any kind.
The warranty does not apply to fuses.
For inquiry or request for repair service, contact Bio-Rad Laboratories after confirming
the model and serial number of your instrument. For technical support, call your local
Bio-Rad office, or in the US, call 1-800-4BIORAD (1-800-424-6723), or visit us on the
Web at discover.bio-rad.com
15
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 16
Appendix G
iCycler® Thermal Cycler Authorization Statement
This instrument, Serial No. ________________, is an Authorized Thermal Cycler.
Its purchase price includes the up-front fee component of a license under United
States Patent Nos. 4,683,195, 4,683,202 and 4,965,188, owned by Roche Molecular
Systems, Inc., and under corresponding claims in patents outside the United States,
owned by F. Hoffmann-La Roche Ltd, covering the Polymerase Chain Reaction
(“PCR”) process, to practice the PCR process for internal research and development
using this instrument. The running royalty component of that license may be
purchased from Applied Biosystems or obtained by purchasing Authorized Reagents.
This instrument is also an Authorized Thermal Cycler for use with applications licenses
available from Applied Biosystems. Its use with Authorized Reagents also provides a
limited PCR license in accordance with the label rights accompanying such reagents.
Purchase of this product does not itself convey to the purchaser a complete license
or right to perform the PCR process. Further information on purchasing licenses to
practice the PCR process may be obtained by contacting the Director of Licensing,
Applied Biosystems, 850 Lincoln Centre Drive, Foster City, California 94404, USA.
No rights are conveyed expressly, by implication or estoppel to any patents on
real-time PCR.
Applied Biosystems does not guarantee the performance of this instrument.
Cy is a trademark of Amersham Biosciences. Excel, Internet Explorer, Windows, and Windows 2000 and XP
are trademarks of Microsoft Corporation. Pentium is a trademark of Intel Corporation. Phillips is a trademark of
Phillips Screw Company Corporation. SYBR and Texas Red are trademarks of Molecular Probes, Inc.
16
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 17
05-0052 IQ5_GUIDE4-6-05.qxp 4/6/2005 10:31 AM Page 1
Bio-Rad
Laboratories, Inc.
Life Science Web site www.bio-rad.com USA (800) 4BIORAD Australia 02 9914 2800
Austria (01)-877 89 01 Belgium 09-385 55 11 Brazil 55 21 2527 3454
Group Canada (905) 712-2771 China (86 21) 6426 0808
Czech Republic + 420 2 41 43 05 32 Denmark 44 52 10 00
Finland 09 804 22 00 France 01 47 95 69 65 Germany 089 318 84-0
Greece 30 210 777 4396 Hong Kong (852) 2789 3300
Hungary 36 1 455 8800 India (91-124)-2398112/3/4, 5018111, 6450092/93
Israel 03 951 4127 Italy 39 02 216091 Japan 03-5811-6270
Korea 82-2-3473-4460 Latin America 305-894-5950
Mexico 55-52-00-05-20 The Netherlands 0318-540666
New Zealand 64 9 415 2280 Norway 23 38 41 30
Poland + 48 22 331 99 99 Portugal 351-21-472-7700
Russia 7 095 721 1404 Singapore 65-64153188
South Africa 00 27 11 4428508 Spain 34 91 590 52 00
Sweden 08 555 12700 Switzerland 061 717 95 55
Taiwan (886 2) 2578 7189/2578 7241 United Kingdom 020 8328 2000