User Manual DMi8
User Manual DMi8
2
Leica DMi8
Instructions for Use
Leica Microsystems CMS GmbH t Instructions 11934056, Revision 1.1 from 2015-07-15
COPYRIGHT/ CONTENTS
Copyright Contents
All rights to this documentation are held by Leica Micro- 1 Important Notes about this Manual ......................... 4
systems CMS GmbH. Reproduction of text or illustrations 1.1 Text symbols and their meanings .............................. 4
(in whole or in part) by print, photocopy, microfilm or other
method (including electronic systems) is not allowed with- 2. Intended Purpose of the Microscope ...................... 6
out express written permission from Leica Microsystems 3. Safety Notes ................................................................. 8
CMS GmbH. 3.1 General safety notes ................................................... 8
3.2 Electrical safety ........................................................... 8
The instructions contained in the following documentation 3.3 Laser safety ................................................................ 11
reflect state-of-the-art technology. We have compiled the 3.3.1 Safety switch (interlock) on the microscope ........ 11
texts and illustrations as accurately as possible. Still, we 3.3.2 Screwed-on filter turret cover................................. 12
are always grateful for comments and suggestions regard- 3.3.3 Laser coupling by the user ....................................... 12
ing potential mistakes within this documentation. 3.3.3.1 Safety hazard due to folded back
Transmitted Light arm ............................................... 13
The information included in this manual may be changed with- 3.3.3.2 Safety hazard at eyepiece........................................ 13
out prior notice. 3.3.3.3 Safety hazard in the condenser area
during operation. ....................................................... 13
3.3.3.4 Safety hazard above the condenser....................... 14
3.3.3.5 Safety hazard in the specimen area ....................... 14
3.3.3.6 Safety hazard due to openings at microscope ..... 15
3.3.3.7 Note regarding the safety hazards in Sections
3.3.3.1 and Section 3.3.3.2 ......................................... 16
3.3.4. Exception for Class 1 lasers ..................................... 17
3.5 Storage and transport ............................................... 18
3.6 Ambient conditions.................................................... 18
3.7 Notes on handling light sources ............................. 19
3.8 Notes on handling the Touch Screen ..................... 20
3.9 Notes on handling immersion oil............................. 20
3.10 Notes on handling acids and bases ....................... 20
3.11 Disposal ....................................................................... 20
4. Leica DMi8 configurations instrument overview 21
4.1 Microscope configurations (examples) ................. 21
4.2 Dimensions (specified in mm).................................. 22
4.3 Specifications+) .......................................................... 24
4.4 Controls ....................................................................... 28
4.4.1 Control panel fields.................................................... 29
4.4.2 Additional control elements ..................................... 31
4.4.3 The Touch Screen ...................................................... 34
4.4.4 The Leica SmartMove control element.................. 36
Revision 1.1, published on 2015-07-15 by: 4.4.5 Leica STP4000/8000 external control panel........... 37
3
1. IMPORTANT NOTES ABOUT THIS MANUAL
Caution!
These Instructions for Use are an essential component
of the product; they must be read carefully before the
product is assembled, put into operation or used, and
must be kept for later reference.
Caution!
In this manual, additional safety notes are indicated with
the triangle symbol shown here, and have a gray back-
ground.
4
1. IMPORTANT NOTES ABOUT THIS MANUAL
Connection to ground
Explanatory note
MM/YYYY
Manufacturing date, e.g. 11 / 2011 for November 2011
5
2. INTENDED PURPOSE OF THE MICROSCOPE
6
2. INTENDED PURPOSE OF THE MICROSCOPE
Caution!
WARNING
The manufacturer assumes no liability for damage Notes on handling laser devices
caused by, or any risks arising from, using the micro- In the standard design without additional laser safety
scope for other purposes than those for which they are precautions, the microscopes are not suitable for cou-
intended or not using them within the specifications of pling laser radiation (such as to the camera ports), since
Leica Microsystems CMS GmbH. this radiation poses a hazard for the user (eye injury in
In such cases, the Declaration of Conformity shall be in- particular).
valid. For use of the microscope with lasers,
Leica Microsystems CMS GmbH offers special micro-
scope variants with additional safety features. Laser
couplings require corresponding safety devices that
have to be inspected and installed by trained personnel.
For further information, please contact your authorized
Leica Microsystems CMS GmbH representative.
7
3. SAFETY NOTES
3. Safety Notes
3.1 General safety notes
These instruments of protection class 1 are built and in- Caution!
spected in accordance with
EN 61010-1 /IEC 61010-1 Safety requirements for The instruments and accessories described in this In-
electrical equipment for measurement, control and labora- structions for Use have been tested for safety and po-
tory. tential hazards.
The responsible Leica Microsystems CMS GmbH af-
They also meet the requirements in filiate or the main plant in Wetzlar must be consulted
EN / IEC 61326-1 Electrical equipment whenever the device is altered, modified or used in con-
for measurement, control and laboratory use – EMC re- junction with non-Leica components that are outside of
quirements. the scope of this manual.
The products were also tested in accordance with EN Unauthorized alterations to the instrument or noncom-
62471 / IEC 62471 Photobiological safety of lamps and lamp pliant use shall void all rights to any warranty claims!
systems.
3.2 Electrical safety
General specifications
Caution! Compact Leica CTR electronics box
In order to maintain this condition and to ensure safe For indoor use only.
operation, the user must follow the instructions and Supply voltage: 100 – 240 VAC
warnings contained in this operating manual. Frequency: 50 / 60 Hz
Power consumption: max. 150 VA
Fuses: 3.15 A, slow-blowing,
Breaking capacity H, 250 VAC
Size: 5x20 mm
Ambient temperature: 15° - 35°C
Relative humidity: 90% up to 30°C,
non-condensing
Protection class: I
Overvoltage category: II
Pollution degree: 2
8
3. SAFETY NOTES
9
3. SAFETY NOTES
Caution!
! Caution!
Only insert or unplug data and control circuits when the
By connecting it to the grounding connection (earth
instrument is switched off; otherwise, the instrument
screw at the back of the electronics box) ancillary
may be damaged.
equipment with its own and/or extra power supply may
be brought to the same ground wire potential. For con-
nections without a ground protection conductor, Leica
Microsystems CMS GmbH Service must be consulted.
WARNING
By definition, the main circuit breaker of this instrument
is the connection between the power cable and device
Caution! port. The user must ensure unobstructed access to the
main circuit breaker at all times.
Never use any fuses as replacements other than those
of the types and the current ratings listed here. Using
other fuses or bridging the fuse holder is not permitted.
The use of incorrect fuses may result in a fire hazard.
WARNING
Only use power cables, or alternative cables with VDE/
HAR codes that minimally fulfill the following conditions:
Caution! 3x0.75 mm2 and 10A/250V.
The microscope's electrical accessory components are
not protected against water. Water can cause electric
shock.
Caution!
Protect the microscope from excessive temperature
fluctuations. Such fluctuations can lead to the accumu-
lation of condensation, which can damage the electrical
and optical components.
Operating temperature: 15-35°C
WARNING
Before exchanging the fuses or lamps, be absolutely
certain to switch off the main power switch and remove
the power cable.
10
3. SAFETY NOTES
3.3 Laser safety Two equivalent interlock connectors, which lead to the
microscope's interlock switch, are attached to the back of
3.3.1 Safety switch (interlock) on the microscope
the microscope. The interlock connector of the microscope
Two safety switches are integrated into the microscope us- must be connected to your laser beam interrupter.
ing a laser safety kit option. These must be integrated into
the laser safety concept so that the light path of the laser
beam is interrupted when the interlock switch is opened.
Compatible coupling:
Manufacturer: binder
Type: 99-0096-102-05
(5-pin cable socket,
2 711 series subminiature round plug connector)
Type: 79-1414-12-05
(5-pin cable socket with 2m cable,
Interlock switch (1) and (2) 702 series subminiature round plug connector)
11
3. SAFETY NOTES
3.3.2 Screwed-on filter turret cover 3.3.3 Laser coupling by the user
As a safety precaution on laser microscope systems, the Because of a danger for the user due to laser radiation, the
fluorescent turret disk cover is secured using 2 Leica DMi8 microscopes in the standard version are not
suitable without additional laser safety measures for cou-
DIN 912 M4x6 lens head Allen screws. pling laser radiation1).
(Order No. 11 703 121 630 000)
If you are, however, modifying a microscope and couple a
laser beam into it, as the user, you assume responsibility
as the manufacturer of the complete system.
WARNING
Danger of serious and irreversible eye damage from la-
ser radiation!
Turn the screws to secure the filter turret cover Radiation of the eye or skin from direct or indirect laser
radiation has to be avoided by all means! The laser light
can cause serious eye damage and skin damage!
12
3. SAFETY NOTES
13
3. SAFETY NOTES
Laser beam path between condenser base and illuminated field dia-
phragm without obstruction
Specimen area
14
3. SAFETY NOTES
a
i
c g
f
d
e
15
3. SAFETY NOTES
3.3.3.7 Note regarding the safety hazards in Sections a) Safety function for the safety hazard mentioned
3.3.3.1 and Section 3.3.3.2 in Section 3.3.3.1.
The special microscope variants have a built-in interlock
switch on the Transmitted Light arm that is activated when
Important safety note
folding back the Transmitted Light arm (interlock switch
The overall responsibility for laser safety of the laser opens).
microscope system assembled by you remains with you The interlock connector of the microscope must be con-
as the responsible body, even if you use the microscopes nected to your laser beam interrupter, which interrupts the
from Leica Microsystems CMS GmbH listed below in laser beam coupled into the microscope when opening the
conjunction. Within the context of your system composi- interlock switch or folding back the illumination arm.
tion and modifications, you have to verify whether the
component selected by you from this chapter is effec-
b) Safety function for the safety hazard mentioned
tive and suitable for your laser protection concept.
in Section 3.3.3.2.
We recommend that for your laser applications you use An additional interlock switch is also installed in the micro-
a microscope variant with the following selection options scope, which is activated (interlock switch opens) when
because the microscope is already equipped with interlock the beam path to the eyepiece is released.
switches at the factory: The interlock switch of the microscope must be connected
to your laser beam interrupter, which interrupts the laser
Leica No. 11889086 beam coupled into the microscope when opening the in-
Laser Safety Kit terlock switch or releasing the beam path towards the eye-
piece.
and also one of the following options: Please note that the laser radiation may not be coupled via
the bottom port!
Leica No. 11889077
Left motorized side port, 100% and interlock
16
3. SAFETY NOTES
When doing so, note the electrical characteristic values of 3.3.4. Exception for Class 1 lasers
the integrated interlock switches in the microscope. You
Coupling a laser of Class 1 in accordance with IEC/EN
must not exceed these values by connecting a laser beam
60825-1 into a microscope does not result in any danger to
interrupter, as otherwise the interlock switches will be
the user.
damaged and no longer function safely and reliably:
The listed safety hazards apply only for lasers of Class 1M,
2, 2M, 3R, 3B and 4.
Maximum permitted electrical characteristic values of the
integrated interlock switch:
Note that the rules and regulations of the United States FDA
21CFR1040.10 Class I (CFR – Code of Federal Regulation Title
Umax= 40 V DC
21, revised as of April 1, 2011) do not distinguish between
Imax= 0.2 A
Class 1 and 1M. Thus the statement that there are no safety
hazards on the microscope do not apply for the laser clas-
The integrated interlock switches in the microscope and
sification in accordance with FDA Class 1.
their wiring with a laser beam interrupter/laser ensure that
the laser radiation you coupled into the microscope cannot
get into the user's eyes through the eyepieces.
17
3. SAFETY NOTES
Transport CAUTION
At a full load, the microscope weighs more than 18
Caution! kg. The user must take corresponding precautions for
Transport and storage at -20 °C – +85 °C and max. 90% transporting the instrument.
humidity.
t 3FNPWFUIFDPOEFOTFS
t 3FNPWFUIFTQFDJNFOTUBHF
Caution!
t 3FNPWFUIF5SBOTNJUUFE-JHIUBSN Do not use this instrument near sources of high electro-
magnetic radiation (for example, unshielded, intention-
t 3FNPWFUIFMBNQIPVTJOHT ally operated ultra-high frequency sources), because
these can disrupt proper operation.
t 4FDVSF UIF NBHOFUJD nVPSFTDFOU UVSSFU DPWFS FH XJUI
adhesive tape. We recommend assessing the electromagnetic environ-
ment before operation of this instrument and then giving
t 3FNPWFBMMNPWJOHPSMPPTFQBSUT corresponding instructions.
t 1SPUFDUUIFNJDSPTDPQFGSPNEVTUBOEEJSUVTJOHBQMBTUJD
bag.
18
3. SAFETY NOTES
Caution!
CAUTION
The microscope must be set up so that the power volt-
age input is not obstructed and the instrument can Make sure to follow the safety notes in the operating
quickly be removed from the power supply if necessary. manual for the light source.
Caution!
CAUTION
Electrical components must be placed at least 10 cm Switch on the light source only if it is firmly connected to
away from the wall and from flammable substances. the microscope.
Uncontrolled light output poses a blinding hazard!
CAUTION
Do not look into the eyepieces when toggling the con-
trast method!
During the toggling procedure, the maximum radiant
power of the light source may be present on the eye-
pieces for a short time and can cause blindness in the
user!
19
3. SAFETY NOTES
3.11 Disposal
CAUTION Once the product has reached the end of its service life,
please contact Leica Microsystems GmbH Service or Sales
Lamps and lamp housing may be hot! about disposal.
They must be placed at least 10 cm away from the wall
and away from flammable substances. Please observe and comply with the national and federal
Supply and data lines in particular may not come into laws and regulations that are equivalent to EC directives
contact with lamp housings! such as WEEE.
Caution!
When using immersion oil, take care to avoid skin con-
tact! Ask the supplier for a safety data sheet!
Caution!
Be absolutely certain to avoid direct contact with these
chemicals.
20
4. OVERVIEW OF THE INSTRUMENT
Leica DMi8 with manual components Leica DMi8 with manual components
With a fixed Transmitted Light arm With a tiltable Transmitted Light arm
Leica DMi8 with encoded components Leica DMi8 with at least one motorized component
21
4. OVERVIEW OF THE INSTRUMENT
162.5
168.1
195 130
204
Weight: approx. 2 kg
310 215
275.6
270
326
Weight: approx. 4 kg
22
4. OVERVIEW OF THE INSTRUMENT
670
500 281
542 310
Weight: approx. 19 kg
23
4. OVERVIEW OF THE INSTRUMENT
4.3 Specifications+)
Contrast methods Leica DMi8 series - depending on Transmitted Light arm
t 5SBOTNJUUFE-JHIU 5-
#'
%'
1)
%*$
1PM
t *OUFSNFEJBUFFZFQPJOU *.$ *OUFHSBUFE.PEVMBUJPO$POUSBTU
or
Fixed Transmitted Light arm
t With integrated LED illumination
t.BOVBMBEKVTUNFOUPGCSJHIUOFTTBOEBQFSUVSFEJBQISBHN
t"VUPNBUJDCSJHIUOFTTBEKVTUNFOUXIFODIBOHJOHUIFDPOUSBTUNFUIPE
t"VUPNBUJDTIVUPGG BEKVTUBCMF
+) Depending on configuration
24
4. OVERVIEW OF THE INSTRUMENT
t 3FDFQUBDMFGPSTUSVDUVSFEJMMVNJOBUJPO 0QUJHSJE
t .BOVBMnVPSFTDFODFJOUFOTJUZNBOBHFS '*.
25
4. OVERVIEW OF THE INSTRUMENT
t Heated stage plate (3°C above room temperature up to 60°C) (248 x 212 mm)
t 5FNQFSBUVSFBEKVTUBCMFTUBHFQMBUF ¡$UP¡$
NNYNN
t 'JYFENJDSPNBOJQVMBUJPOTUBHFT
t 4UBHFQMBUFDPBUFEJODFSBNJD NNYNN
t )FBUFETUBHFQMBUF ¡$BCPWFSPPNUFNQFSBUVSFVQUP¡$
NNYNN
t 5FNQFSBUVSFBEKVTUBCMFTUBHFQMBUF ¡$UP¡$
NNYNN
t .BOVBMBOENPUPSJ[FEQMBUFDSPTTTUBHF
t 5SBWFMSBOHFNNYNN
t 0QUJPOBMJOTFSUT OPSNBM
IFBUBCMFBOEDPPMBCMF
GPSEJWFSTFBQQMJDBUJPOT
Size of inserts:160 mm x 110 mm (compatible with scanning stages)
t 4DBOOJOHTUBHF
t 0QUJPOBMJOTFSUT OPSNBM
IFBUBCMFBOEDPPMBCMF
GPSEJWFSTFBQQMJDBUJPOT
Size of inserts:160 mm x 110 mm
t -.5MJOFBSNPUPSJ[FETUBHF
26
4. OVERVIEW OF THE INSTRUMENT
t 8JUINPUPSJ[FEPSFODPEFE8PMMBTUPOQSJTNEJTL PQUJPOBM
t .BYJNVNUSBWFMTQFFENNT
t GPDVTJODSFNFOUTNNNNN
t &MFDUSPOJDGPDVTSFQPTJUJPOJOH
t "VUPNBUJDMPXFSJOHCFGPSFPCKFDUJWFDIBOHF
t &MFDUSPOJDQBSGPDBMJUZ
t 0QUJPOBM"EBQUJWF'PDVT$POUSPM "'$
t 0QUJPOBM#PUUPNQPSU
t 0QUJPOBM5PQQPSUXJUITXJUDIQPTJUJPOT
t UPUIFFZFQJFDFT
t UPUIFFZFQJFDFTUPUIFQPSU
t 0QUJPOBM5PQQPSUXJUITXJUDIQPTJUJPOT
t UPUIFFZFQJFDFT
t UPUIFFZFQJFDFTUPUIFQPSU
27
4. OVERVIEW OF THE INSTRUMENT
t #VUUPOGPS5-'MVPSUPHHMJOH
t CVUUPOTGPSNPUPSJ[FEBQFSUVSFEJBQISBHNBEKVTUNFOU
t CVUUPOTGPSNPUPSJ[FEJMMVNJOBUFEmFMEEJBQISBHNBEKVTUNFOU
t CVUUPOGPSPQFSBUJOHUIFNPUPSJ[FETIVUUFS
t WBSJBCMFGVODUJPOLFZT
t PQFSBUJOHCVUUPOTGPSGPDVTUISFTIPMET POMZGPSFMFDUSPOJDGPDVTJOH
t IBOEXIFFMTGPSGPDVTJOH
t 3PUBSZLOPCGPSTFUUJOHUIFCSJHIUOFTT
t CVUUPOTGPSnVPSFTDFODFDVCFT
t CVUUPOTGPSPCKFDUJWFT
t -FJDB4NBSU.PWFFSHPOPNJDDPOUSPMFMFNFOUGPSNPOJUPSJOHY
Z
[BOEBEEJUJPOBMWBSJ-
able function keys
t -FJDB451PQFSBUJPOWJBFYUFSOBM5PVDI4DSFFO
t -FJDB451PQFSBUJPOWJB5PVDI4DSFFOVTJOHDPOUSPMFMFNFOUGPSNPOJUPSJOHY
Z
[BOE
additional variable function keys
t 4XJUDISPETGPSPQFSBUJOHUIFQPSU
NBOVBMTIVUUFSBOE.BHOJmDBUJPO$IBOHFS
t 4MJEFSGPSNBOVBM'*.
*.$BOE*1$BEKVTUNFOU
t 4XJUDISPEGPSPQFSBUJOHUIFJOmOJUZQPSU
Electronics box t Separate unit for controlling all motorized and electronic elements of the
microscope
Alternatively:
t -FJDB$53Compact
t -FJDB$53"EWBODFE
t -FJDB$53"EWBODFE XJUIPVUQPXFSTVQQMZ
t -FJDB$53"EWBODFE
Interfaces t 4FSWJDFQPSUGPSTBWJOHMPBEJOHNJDSPTDPQFTFUUJOHTWJBUIF5PVDI4DSFFO
Leica DMi8 motorized
t Y*¤$GPSFYUFSOBMJOUFSOBMQFSJQIFSBMEFWJDFT
t $53CPYFT
t YJOUFSMPDLGPSMBTFSBQQMJDBUJPOT
4.4 Controls
Depending on the configuration of your microscope, vari-
ous controls are available to you. Accordingly, the control
panel fields on the left and right sides of the stand contain
a different number of buttons. However, the functions them-
selves are assigned to specific buttons.
28
4. OVERVIEW OF THE INSTRUMENT
4
4 Adjusting the Transmitted Light brightness of the inten-
sity levels for the automated FIM (fluorescence inten-
sity manager) 3
6 2 1
An LED indicates the illumination method which is cur-
rently affected by the brightness adjustment.
Maximum system design of the left control panel field
5 Opening/closing the motorized shutter The number of buttons may vary depending on the configu-
or during manual shutter light control ration.
An LED displays the state of the shutter:
LED on = shutter opened
LED off = shutter closed.
6 Focusing
29
4. OVERVIEW OF THE INSTRUMENT
30
4. OVERVIEW OF THE INSTRUMENT
1 Filter receptacle 1
2
2 Aperture Diaphragm Adjustment
3 4
3 Stop lever for condenser height adjustment
4 IMC or light ring slide
5 Bertrand lens centering
6 Toggling the top port and Bertrand lens 5
7 Manual objective change 6
7 8 9 10 11 12
2 3
4
5
6
31
4. OVERVIEW OF THE INSTRUMENT
DIC setting
1 2
10 9 8 7 6 5
32
4. OVERVIEW OF THE INSTRUMENT
1 2 3 4 5 6 7
10
9
8
11
5 Manual Shutter
33
4. OVERVIEW OF THE INSTRUMENT
Initialization
Basic microscope settings
Magnification
Optional
34
4. OVERVIEW OF THE INSTRUMENT
Note:
Before working with the touch screen, we recommend cali-
brating it first.
Calibration
t 1SFTTUIF button.
t 1SFTTUIF button.
Saving the microscope settings This key shows you the version number of the Touch Screen
In Microscope Settings, press Save to write the cur- software and that of the master firmware of the connected
rent microscope settings to a USB stick, inserted into microscope.
the left side of the microscope (Service port). Pressing Please have these available for any inquiries.
Load restores all settings.
To prevent unwanted interference from light, you can The meaning of the other menu items of the navigation bar
switch off the display using the Switch off key. To switch it is explained in the respective chapters.
back on, simply touch any location on the display.
You can also change the brightness incrementally using the
Brightness key.
35
4. OVERVIEW OF THE INSTRUMENT
You can adjust the height of the rotary knobs by turning the
wheel 4 to a working height that is convenient for you.
1
3
2 4
5
36
4. OVERVIEW OF THE INSTRUMENT
2 1 8
3 4 5 6
Touch Screen for STP4000/8000
37
5. OPERATION
5. Operation
Caution! Note:
The microscope may only be set up, installed and put Operating the motorized components using the Leica
into operation by authorized Leica personnel! STP4000/8000 control panel is described in separate In-
structions for Use.
Note:
Caution!
For information about using the LAS X software, refer to the
The microscope may only be operated by personnel who corresponding online help.
have been instructed by an authorized Leica employee!
Note:
If at all possible, avoid touching optical surfaces such as
the lens surfaces on the objective. If fingerprints do appear
on the glass surfaces, remove them with a soft leather or
linen cloth. Even small traces of finger perspiration can
damage the surfaces in a short time. For additional instruc-
tions, see the chapter entitled "Care of the Instrument" o S.
89
38
5. OPERATION - SWITCHING ON
t4XJUDIPOUIFFMFDUSPOJDTCPYPOUIFPOPGGTXJUDI5IF
indicator lamp is illuminated in green during operation.
Note:
If you have connected a PC, switch on the electronics box
first, and then the computer.
Note:
At initialization, all of the user's most recent settings are re-
stored.
! Caution:
The focus position and lower threshold are stored when
switching off the microscope.
39
5. OPERATION - SWITCHING ON
For fluorescence:
t 4XJUDIPOUIFLeica EL6000 compact light source on the
on/off switch. The indicator lamp is illuminated during op-
eration.
CAUTION
Do not switch on the light source until the light guide is
connected to the microscope.
Uncontrolled light output from the light guide poses a
blinding hazard!
There is also a burn hazard at the light guide output!
Caution!
When using other light sources, observe the respective
instructions for the light source.
40
5. OPERATION - ILLUMINATION
5.2 Illumination
Caution!
When using other light sources than those described
here, observe the respective instructions for the light
source.
t *GUIFNJDSPTDPQFIBTBNBOVBMTIVUUFSBOEBTIVUUFSCVU-
ton, the light is switched off when pressing the button. To
switch it on again, press the rotary knob for the light in-
tensity.
t 4FUUIFMJHIUJOUFOTJUZVTJOHUIFLOPCPOUIFMFGUTJEFPGUIF
stand.
41
5. OPERATION - ILLUMINATION
t "EKVTUUIFJOUFOTJUZVTJOH+ and –.
The setting can be made in coarse and fine steps in a
range from 0 to 255. Pressing the middle key, which also
displays the current value, switches between coarse and
fine adjustment.
5.2.2 Fluorescence
Leica DMi8 with manual encoded fluorescence
CAUTION
Do not switch on the light source until the light guide is
connected to the microscope.
Uncontrolled light output from the light guide poses a
blinding hazard!
There is also a burn hazard at the light guide output!
42
5. OPERATION - ILLUMINATION
Note:
If the Leica DMi8 is equipped with a 2 port infinity port (oS. 44)
switch over to the current port .oS. 58
CAUTION
Do not switch on the light source until the light guide is
connected to the microscope.
Uncontrolled light output from the light guide poses a
blinding hazard!
There is also a burn hazard at the light guide output!
43
5. OPERATION - ILLUMINATION
t 8IFOVTJOHUIF'MVPSFTDFODF*OUFOTJUZ.BOBHFS '*.
The brightness is adjusted in 5 defined increments on the
brightness adjusting knob:
100% / 55% / 30% / 17% / 10%
44
5. OPERATION - ILLUMINATION
5.2.4 Shutter
Leica DMi8 with manual shutter
t 0QFOBOEDMPTFUIFNBOVBMTIVUUFSVTJOHUIFTXJUDISPE
on the manually integrated fluorescence axis or using the
switch rod on the manual external fluorescence axis.
t 0QFOPSDMPTFUIFTIVUUFSVTJOHUIFTIVUUFSLFZPOUIFMFGU
side of the stand.
This switches the active illumination axis (TL or Fluor) on
or off.
When the shutter is open, the LED is illuminated.
45
5. OPERATION - OBJECTIVES
5.3 Objectives
! Caution:
If necessary, first lower the focus to prevent damaging the
objective front lenses!
t $POmHVSFBOPCKFDUJWFCZQSFTTJOHUIFDPSSFTQPOEJOHPC-
jective changer button on the keypad of the front side of
the microscope.
The LED of the objective changer button is illuminated.
46
5. OPERATION - OBJECTIVES
Note:
The positions of the objectives in the objective turret have
been specified at the factory and must be observed when Function keys on the right side of microscope
installing the objectives.
(See Installation Manual).
t 5IFNPTUSFDFOUMZDPOmHVSFEMJHIUJOUFOTJUZJOUIFSFTQFD-
tive contrast method (with DIC fine adjustment where ap-
plicable)
t 5IF NPTU SFDFOUMZ DPOmHVSFE GPDVT JODMVEJOH QBSGPDBM-
ity)
47
5. OPERATION - OBJECTIVES
Immersion objectives
t 'PSimmersion objectives use the appropriate immersion
medium.
OIL: Use optical immersion oil in accordance with DIN/
ISO standards
only
Cleaning oS. 89
W: Water immersion
IMM: Universal objective for water, glycerin,
Oil immersion
Caution!
Follow safety data sheet for immersion oil!
t 4FMFDUBOJNNFSTJPOPCKFDUJWF
The objective nosepiece is lowered to the lower thres-
hold. This makes it possible to apply the immersion fluid
when switching from the dry objective to the immersion
objective. Conversely, the immersion fluid can be re-
moved.
The current objective remains in the beam path.
t 1SFTTUIFLFZGPSUIFEFTJSFEJNNFSTJPOPCKFDUJWF
XIJDI
is now approached.
Note:
For lockable immersion objectives, press the front part
upwards as far as it will go (approx. 2 mm). Then, after a
gentle turning motion to the right, the objective is locked.
Note:
If objectives are retrofitted, they have to be taught in using
the LAS X software: Afterwards, the parfocality should also
be taught in again.
See the instructions for the LAS X.
48
5. OPERATION - TUBE
5.4 Tube
There are also Leica DMi8 without tube.
Note:
Close any unused tube openings, as otherwise stray light
can interfere with observation.
Observation(VIS) Photo
100 % 0%
0% 100 %
Alternatively: 50% 50%
BL Activating the Bertrand lens*
Switch rod
49
5. OPERATION - EYEPIECES
5.5 Eyepieces
Note:
The eyepiece's aperture protector must be removed or
folded back during microscopy while wearing eyeglasses. Protection screen
We recommend removing bifocals and spectacles with pro-
gressive-addition lenses when using the microscope. Adjusting the eyelens
t 'PDVTPOUIFPCKFDUUISPVHIUIJTFZFQJFDF
t 5IFO
DMPTFUIBUFZFBOEGPDVTPOUIFPCKFDUCZBEKVTUJOHUIF
second ocular only.
t 5IFO
XJUI ZPVS MFGU FZF
WJFX UIF TBNF QPTJUJPO PG UIF
specimen and rotate the left eyepiece tube until this posi-
tion is brought into sharp focus. While doing so, do not
change the adjustment of the z-position!
Note:
We recommend using the LAS X software to teach in eye-
pieces that are not included in the scope of delivery or are
retrofitted. This ensures that the total magnification speci-
fied on the Touch Screen is correct.
See the instructions for the LAS X.
50
5. OPERATION - STAGES AND OBJECT DISPLACEMENT
Note:
When using other external stages, observe the respective
instructions for the external stages.
WARNING
Move in y-direction
In spite of the low speed, there is a risk of pinching and
crushing injuries when moving the scanning stage.
Never put objects or body parts in the travel path!
Stage movement using Leica SmartMove
51
5. OPERATION - STAGES AND OBJECT DISPLACEMENT
Zero point
52
5. OPERATION - STAGES AND OBJECT DISPLACEMENT
t 4BWJOHUIFDVSSFOUTUBHFQPTJUJPOVOEFSUIJTOVNCFSVTJOH
the key
t .PWJOHUPUIFDVSSFOUTUPSFEQPTJUJPOVTJOHUIF key
When driving to the load position, the objective nose-
piece is raised to the lower threshold.
Note:
If a stored position is outside the travel range of the stage,
for example as a result of subsequent changes to the travel
range in the LAS software, you must restore the maximum
53
5. OPERATION - FOCUSING
5.7 Focusing
Note:
If your microscope is equipped with a closed loop, this can
be switched on and off using the LAS X software.
See the instructions for the LAS X.
t 'PDVTUIFJNBHFCZUVSOJOHUIFGPDVTEJBMTPOUIFleft and
right sides of the stand.
Coarse focusing
Fine focusing
t 'PDVTUIFJNBHFVTJOHUIFSFBSSPUBSZLOPCPOUIFLeica Focusing
SmartMove control element.
Note:
For coarse and fine focusing with the handwheel, 2 focus
stages each can be selected using the LAS X software.
54
5. OPERATION - FOCUSING
Note:
4
The parfocality is taught in at the factory. As a result of
screwing in the objectives during assembly, it might be nec-
essary to teach in the parfocality again. 3
We recommend checking the parfocality before setting the
thresholds and, if necessary, teaching them in again using 2
the LAS X software. 1
Note:
Right side of Leica DMi8 with motorized focus
The focus position and stops are stored by the microscope 1 Focusing
and retained from one session to the next when power is 2 Focus stop
switched off. (the LED illuminates when focus threshold is set)
3 Setting the focus position and focus threshold
Setting thresholds 4 Focus position
(the LED illuminates when the focus position is set)
Lower focus threshold:
t 1SFTTUIFSET key and the key
p
— Z
When the threshold is set, the LED is illuminated.
Pressing the same key combination again deletes the
threshold.
Moving to the thresholds
The lower threshold is the same for all objectives and can- Lower focus threshold:
not be exceeded. t 1SFTTBOEIPMEUIFLFZ
Focus position: p
— Z
t 1SFTTUIFSET key and the key
— Focus position:
n Z Press and hold the key
—
When the focus position is set, the LED is illuminated. n Z
Pressing the same key combination again deletes the fo-
cus position.
Note:
The focus position cannot be exceeded. When moving to the thresholds, the keys must be held until
It should be defined for the dry objective with the highest the position has been reached.
magnification level. For all other objectives, it is set auto-
matically, taking into account the parfocality adjustment
and the working distance.
55
5. OPERATION - FOCUSING
t 5PTFUPSDMFBSUIFGPDVTQPTJUJPOPSUIFMPXFSthreshold,
select the corresponding position under Set/Delete, then
press the Set or Delete key.
56
5. OPERATION - FOCUSING
t "'$NPOJUPSJOHDBOCFQFSGPSNFEVTJOHWBSJBCMFGVODUJPO
keys, LAS X software or via the Touch Screen.
t "DUJWBUF UIF "'$ VTJOH UIF On/Off key. The knob behind
AFC turns green.
t 'PDVTUIFTQFDJNFOBOETBWFUIFQPTJUJPOVTJOHUIFHold
key.
57
5. OPERATION - SELECTING THE PORT
The manual switch rod on the right side of the stand en-
ables or disables the left side port.
VIS LEFT
100 % 0%
20% 80 %
Alternatively: 0% 100%
t 4FMFDUUIFSJHIUPSMFGUTJEFQPSUVTJOHUIFDPSSFTQPOEJOH
key or point 100% of the light at the eyepiece.
100% VIS Right side port
Note:
The "Select the bottom port" function can be assigned
to one of the variable function keys on the stand or Leica
SmartMove.
Note:
Side ports not in use must by covered with cover caps.
58
5. OPERATION - MAGNIFICATION CHANGER
t 6TJOHUIFTMJEFCBS
UPHHMFCFUXFFOYBOEUIFNBHOJmDB-
tion factor.
Note:
The mechanical Magnification Changer works on the eye-
piece and the top port.
Note:
The selected factor is included in the calculation of the to-
tal magnification.
Note:
The motorized Magnification Changer works on all ports.
59
5. OPERATION - APERTURE AND ILLUMINATED FIELD DIAPHRAGM
Leica DMi8 with manual illuminated field diaphragm Leica DMi8 with tiltable Transmitted Light arm
Note:
Manual or motorized illumi-
The illuminated field diaphragm is available only in conjunc- nated field diaphragm
tion with ≤ S28 condensers.
For an integrated fluorescence axis: Leica DMi8 with manual integrated fluorescence axis
t 5IFNBOVBMJMMVNJOBUFEmFMEEJBQISBHNJTBEKVTUFEVTJOH
a slide bar on the right side of the microscope.
60
5. OPERATION - APERTURE AND ILLUMINATED FIELD DIAPHRAGM
Note:
The old values will be overwritten by the current ones!
Note:
When using PH or DF, the aperture diaphragm is open all
the way and cannot be closed.
t "EKVTUUIFBQFSUVSFBOEmFMEEJBQISBHNTVTJOHUIF+ and
– keys.
61
5. OPERATION - FILTER
5.11 Filter Leica DMi8 with tiltable Transmitted Light arm: Transmitted
Light filter
5.11.1 Transmitted Light filter
For a tiltable Transmitted Light arm:
t 5IF5SBOTNJUUFE-JHIUmMUFSBOEUIFEBSLTUPQBSFTXVOH
in and out on the right side of the Transmitted Light arm.
Note:
The front position is assigned at the factory using a shutter
(dark stop), which is interchangeable with a filter.
For a fixed Transmitted Light arm: Leica DMi8 with a fixed Transmitted Light arm: Transmitted
t *OTFSUUIFmMUFSJOUPUIFmMUFSSFDFQUBDMFPOUIF5SBOTNJU- Light filter
ted Light arm.
62
5. OPERATION - FILTER
Note:
Observe the Installation Manual when equipping the fluo
turret or replacing individual filter cubes.
t 1VMMUIFmMUFSDVCFGPSXBSEUPSFNPWFJUBOETMJEFUIFOFX
filter cube straight into the receptacle.
t 1VUUIFDPWFSCBDLJOQMBDF
t 6TJOHBNPUPSJ[FEnVPSFTDFODFUVSSFUEJTLXJUI-*% -FJDB
Radio Frequency Identification Device), there is an auto-
matic full rotation of the turret to read out the filter. For
additional instructions, see the chapter entitled “Leica Empty position for mounting a filter cube. The filter cube is pushed in
RFID Module” o S. 65 from the front evenly until it clicks into place.
t $POmHVSFBmMUFSDVCFCZQSFTTJOHUIFDPSSFTQPOEJOHmM-
ter cube changer button on the keypad of the front side of
the microscope.
The LED of the filter cube changer button is illuminated.
63
5. OPERATION - FILTER
t 4FMFDUFLUO.
WARNING
For use of the microscope with lasers, Leica
Microsystems CMS GmbH offers special microscope
variants with additional safety features.Laser couplings
require corresponding safety devices that have to be in-
spected and installed by trained personnel.
WARNING
Store the screws in a safe place. Never operate the
microscope when the cover is not screwed into place!
t $MPTFBOESFGBTUFOUIFDPWFS
64
5. OPERATION - MEMORY FUNCTION ON THE TOUCHSCREEN
Note:
This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15
of the FCC Rules. These limits are designed to provide rea-
sonable protection against harmful interference when the
equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequen-
cy energy and, if not installed and used in accordance with
LID Module 1.0 the instruction manual, may cause harmful interference to
radio communications. Operation of this equipment in a res-
5.11.3.2Specifications LID Module 1.0
idential area is likely to cause harmful interference in which
Supply Voltage 5V case the user will be required to correct the interference at
Dimensions 36 x 25 x 9.8 mm his own expense.
RFID Protocol
Transmit frequncy
ISO 15693
13,56 MHz
! Notice:
This device complies with Part 15 of the FCC Rules [and
Operating temperature -5°C to +65°C
with Industry Canada licence-exempt RSS standard(s)].
Storage temperature -20°C to +85°C Operation is subject to the following two conditions:
Communication interface I2C 1. this device may not cause harmful interference, and
2. this device must accept any interference received, in-
5.11.3.3Specifications RFID Tag D6.7 special cluding interference that may cause undesired opera-
tion.
Modulation Scheme Close coupling
Dimensions D=6,7mm ! Notice:
TH=2,6mm Changes or modifications made to this equipment not ex-
RFID Protocol ISO 15693 pressly approved by Leica Microsystems CMS GmbH may
Transmit frequncy 13,56 MHz void the FCC authorization to operate this equipment.
Read write type 2kbit EEProm
Hereby, Leica Microsystmems CMS GmbH declares
Operating temperature -15°C to +65°C
that the radio equipment type LID Module 1.0 designa-
Storage temperature -25°C to +125°C tion of type of radio equipment is in compliance with Di-
rective 2014/53/EU. The full text of the EU declaration of
conformity is available at the following internet address:
www.leica-microsystems.com
65
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t 0OUIF5PVDI4DSFFO
VTFUIF key to switch to ba- Leica DMi8 with manual Transmitted Light method
sic microscope settings.
The contrast method on the manual or encoded Leica DMi8
can be operated using the manual condenser, the manual
t 4FMFDUUIFMemory Function menu using the key. objective nosepiece, or the rotary knobs and slide bars on
the stand.
t 4FMFDUUIFEFTJSFE5SBOTNJUUFE-JHIUDPOUSBTUNFUIPE
66
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
Note:
For the stands with 3-plate mechanical stages, posi-
tion the arm 25 mm below the marks, as the illumina-
tion axis is installed 25 mm higher on an adapter than
on the fixed stage.
t 4XJOHBMMSFNBJOJOHPQUJDBMDPNQPOFOUT
TVDIBTUIFBOB-
lyzer, polarizer or IC prisms, from the beam path.
t*OTFSUB5SBOTNJUUFE-JHIUTQFDJNFO
t4FMFDUZPVSPCKFDUJWF
67
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t'PDVTUIFJNBHFVTJOHUIFGPDVTEJBMT
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
t 'PSBOPQUJNBMmFMEEJBQISBHNTFUUJOH
DIFDLUIF,PFIMFS
illumination.
t 4FMFDUUIFDPOUSBTUNFUIPEBF (Brightfield).
The Brightfield position is approached for the motorized Condenser setting for tiltable Transmitted Light arm - con-
condenser. figuring BF position
An empty position or the filter cube "DAPI-TL" on the fluo-
rescent turret disk is approached automatically.
t *OTFSUB5SBOTNJUUFE-JHIUTQFDJNFO
t 4FMFDUBTVJUBCMFPCKFDUJWFJOUIFNBHOJmDBUJPONFOV
3
t 'PDVT UIF JNBHF VTJOH UIF SPUBSZ LOPC PO UIF -FJDB
SmartMove or using the focus dial. 2
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
1
Note:
1
You can find additional information regarding the correct
settings for Koehler illumination in Leica Science Lab.
Link: Condenser setting for fixed Transmitted Light arm
http://www.leica-microsystems.com/science-lab/ 1 Marks
2 Stop lever for condenser adjustment
3 Aperture diaphragm
68
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t *G OFDFTTBSZ
TFU UIF 5- #SJHIUmFME QPTJUJPO PS BO FNQUZ
position on the fluo turret, or select the "DAPI-TL" filter
cube.
t 6TFUIFMJHIUSJOHPGUIFMJHIUSJOHTMJEFUIBUDPSSFTQPOET
to the magnification (5, 10/20 or 40).
Note:
The Phase Contrast light ring slide is encoded. As soon as
the slider is taken out of the Brightfield position (available
opening) and engaged in a Phase Contrast position (any
light ring), the brightness increases. Conversely, switch-
ing from Phase Contrast to Brightfield decreases the
brightness. Phase Contrast for fixed Transmitted Light arm
6 1 6 2
4
1 Slider for light rings (S80/0.30) 5
2 Slider for light rings (S40/0.45) 4
3 Click stop 4
4 Light rings
5 Brightfield position 3
6 "TOP LEFT" marking Light ring slide for fixed Transmitted Light arm
69
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t$PNQMFUFMZPQFOUIFBQFSUVSFPOUIFDPOEFOTFS
t 4XJOHBMMSFNBJOJOHPQUJDBMDPNQPOFOUT
TVDIBTUIFBOB-
lyzer, polarizer or IC prisms, from the beam path.
t*OTFSUB1IBTF$POUSBTUTQFDJNFO
t'PDVTUIFJNBHFVTJOHUIFGPDVTEJBMT
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
t *OTFSUB5SBOTNJUUFE-JHIUTQFDJNFO
t 4FMFDUBTVJUBCMFPCKFDUJWFJOUIFNBHOJmDBUJPONFOV
Objectives that are suitable for Phase Contrast are en-
graved with PH.
t 'PDVT UIF JNBHF VTJOH UIF SPUBSZ LOPC PO UIF -FJDB
SmartMove or using the focus dial.
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
Note:
When selecting the Phase Contrast method, the aperture
diaphragm is opened fully and can not be adjusted.
70
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t *G OFDFTTBSZ
TFU UIF 5- #SJHIUmFME QPTJUJPO PS BO FNQUZ
position on the fluo turret, or select the "DAPI-TL" filter
cube.
t4FMFDUB%BSLmFMEPCKFDUJWF
t4FMFDUUIFDPSSFTQPOEJOH%BSLmFMETUPQ MJHIUSJOH
POUIF
condenser.
t$PNQMFUFMZPQFOUIFBQFSUVSFPOUIFDPOEFOTFS
t 4XJOHBMMSFNBJOJOHPQUJDBMDPNQPOFOUT
TVDIBTUIFBOB-
lyzer, polarizer or IC prisms, from the beam path.
t*OTFSUB%BSLmFMETQFDJNFO
t'PDVTUIFJNBHFVTJOHUIFGPDVTEJBMT
Condenser setting for tiltable Transmitted Light arm - con-
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC figuring Darkfield stop (light ring)
t 4FMFDUUIFDPOUSBTUNFUIPEDF (Darkfield).
t *OTFSUB5SBOTNJUUFE-JHIUTQFDJNFO
t 4FMFDUBTVJUBCMFPCKFDUJWFJOUIFNBHOJmDBUJPONFOV
t 'PDVT UIF JNBHF VTJOH UIF SPUBSZ LOPC PO UIF -FJDB
SmartMove or using the focus dial.
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
Note: Note:
The maximum applicable objective aperture for the Dark- When the Darkfield method is selected, the aperture dia-
field is for the condenser S1 0.70 and for the condenser S23/ phragm is opened fully and cannot be adjusted.
S28/S40 0.40.
71
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t *G OFDFTTBSZ
TFU UIF 5- #SJHIUmFME QPTJUJPO PS BO FNQUZ
position on the fluo turret, or select the "DAPI-TL" filter
cube.
t 4XJOHBMM*$QSJTNTPVUPGUIFCFBNQBUI
t4FMFDUZPVSPCKFDUJWF
t *OTFSUBTQFDJNFO
t'PDVTUIFJNBHFVTJOHUIFGPDVTEJBMT
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
Pushing in analyzer
72
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
Motor procedure:
t "GUFS UIF POL contrast method has been selected, the
polarizer is swung in (if the microscope is equipped with
these components). The analyzer cube is also automati-
cally brought into the beam path.
Combined procedure:
t :PVBMTPIBWFUIFPQUJPOPGDPNCJOJOHQVSFMZNFDIBOJDBM
and motorized components, i.e. you can combine a me-
chanical analyzer and a motorized polarizer.
Manual procedure:
For a manual swing polarizer and a mechanical analyzer,
proceed with the manual Transmitted Light method as de-
scribed for the Leica DMi8.
t 4FMFDUBTVJUBCMFPCKFDUJWFJOUIFNBHOJmDBUJPONFOV
t *OTFSUBTQFDJNFO
t 'PDVT UIF JNBHF VTJOH UIF SPUBSZ LOPC PO UIF -FJDB
SmartMove or using the focus dial.
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
73
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t *G OFDFTTBSZ
TFU UIF 5- #SJHIUmFME QPTJUJPO PS BO FNQUZ
position on the fluo turret, or select the "DAPI-TL" filter
cube.
t4FMFDUBOPCKFDUJWF
t *OTFSUBTQFDJNFO
Condenser setting for tiltable Transmitted Light arm - con-
t'PDVTUIFJNBHFVTJOHUIFGPDVTEJBMT
figuring Wollaston prism
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
74
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t 5IFQPMBSJ[FSMPDBUFEJOUIFDPOEFOTFSBOEUIFmUUJOHDPO-
denser prisms are automatically brought into the beam
path. The corresponding objective prism and the analyzer
cube are also approached automatically.
t1MBDFB%*$TQFDJNFOPOUIFTUBHF
t 4FMFDUBTVJUBCMFPCKFDUJWFJOUIFNBHOJmDBUJPONFOV
t 'PDVT UIF JNBHF VTJOH UIF SPUBSZ LOPC PO UIF -FJDB
SmartMove or using the focus dial.
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
t 5IFBEKVTUNFOUJTDBSSJFEPVUWJBUIFCVUUPOT+ and – on
the Touch Screen.
Manual alternative:
t .BOVBMMZ TXJOH UIF QPMBSJ[FS PO UIF DPOEFOTFS JOUP UIF
beam path.
t 5IFO
BMTPNBOVBMMZJOTFSUUIFBOBMZ[FSPOUIFSJHIUTJEF
of the stand until it locks into place.
t 'PSmOFBEKVTUNFOUVTFUIFLOVSMFESJOHCFMPXUIFPCKFD-
tive nosepiece.
75
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
Note:
The Integrated Phase Contrast requires a condenser on the
tiltable Transmitted Light arm.
t *G OFDFTTBSZ
TFU UIF 5- #SJHIUmFME QPTJUJPO PS BO FNQUZ
position on the fluo turret, or select the "DAPI-TL" filter
cube.
t 4FMFDUB#SJHIUmFMEPCKFDUJWFXJUIFZFQPJOU#PS$
Receptacle for Phase Contrast module
t4FMFDUUIFDPSSFTQPOEJOHlight ring on the condenser (see
table). Phase Contrast module for tiltable Transmitted Light illumi-
nation carrier
t$PNQMFUFMZPQFOUIFBQFSUVSFPOUIFDPOEFOTFS 1 Centering screws for light ring 20x - 63x
2 Position for objectives 20x - 63x
t 4XJOHBMMSFNBJOJOHPQUJDBMDPNQPOFOUT
TVDIBTUIFBOB-
lyzer, polarizer or IC prisms, from the beam path. 3 Brightfield position
4 Position for objectives 5x - 10x
t *G OFDFTTBSZ
SFNPWF UIF DPWFS PG UIF IPMEFS BOE JOTFSU
the Phase Contrast module evenly from the front on the 5 Phase ring 20x - 63x
left side of the microscope. 6 Phase ring 5x - 10x
t*OTFSUB1IBTF$POUSBTUTQFDJNFO
t'PDVTUIFJNBHFVTJOHUIFGPDVTEJBMT
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
76
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t *OTFSUB5SBOTNJUUFE-JHIUTQFDJNFO
t *G OFDFTTBSZ
SFNPWF UIF DPWFS PG UIF IPMEFS BOE JOTFSU
the Phase Contrast module evenly from the front on the
left side of the microscope.
t 4MJEFUIF1IBTF$POUSBTUNPEVMFJOUPUIFDPSSFDUFZFQPJOU
B or C.
t 'PDVT UIF JNBHF VTJOH UIF SPUBSZ LOPC PO UIF -FJDB
SmartMove or using the focus dial.
Note:
When selecting the Phase Contrast method, the aperture
diaphragm is opened fully and can not be adjusted.
Note:
For centering the phase ring 20x-63x, refer to the Installa-
tion Manual.
77
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
5.13.7 Integrated Modulation Contrast – IMC (TL) IMC module for tiltable Transmitted Light arm
Leica DMi8 with manual Transmitted Light method 1 Position for objectives with eyepoint C
2 Position for objectives with eyepoint B
t *G OFDFTTBSZ
TFU UIF 5- #SJHIUmFME QPTJUJPO PS BO FNQUZ
position on the fluo turret, or select the "DAPI-TL" filter 3 Brightfield position
cube. 4 Height adjustment
3
t 4XJOHBMMSFNBJOJOHPQUJDBMDPNQPOFOUTTVDIBTUIFBOB- 2
1
lyzer or IC prisms from the beam path.
For a tiltable Transmitted Light arm: IMC module for fixed Transmitted Light arm
t 4MJEFUIF*.$NPEVMFJOUPUIFDPSSFDUFZFQPJOU#PS$
1 IMC module
For a fixed Transmitted Light arm: 2 IMC modulator
t *OTFSUUIF*.$NPEVMBUPSJOUPUIF*.$NPEVMFGSPNUIF
3 Fine adjustment
side according to the eyepoint B, C or D of the objec-
tive. 4 Positioning depending on eyepoint B, C or D
Ensure that the "TOP" marking faces upwards.
1
t &OHBHFUIFTMJEFJOUIF*.$QPTJUJPO 2
(Push in for IMC, pull out for the Brightfield).
3
4
78
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
Note:
The position of the polarizer influences the appearance of
the relief. The best relief images are obtained in the extinc-
tion position; the relief disappears at a 90° position. Condenser setting for tiltable Transmitted Light arm - con-
figuring slit diaphragm
For a fixed Transmitted Light arm:
t *G OFDFTTBSZ
SFNPWF UIF Phase Contrast slider in the IMC slit diaphragm slider
Transmitted Light arm.
t )PMEUIF*.$TMJEFTPUIBUUIFNBSLJOH501-&'5JTBU
the top left and the rest of the marking faces forwards.
The click stops are on the front longitudinal side of the
slide.
(Depending on whether the S40/0.45 or S80/0.30 con-
denser is used, various IMC slit diaphragm sliders ex-
ist).
t &OHBHFUIFTMJEFJOUIF*.$QPTJUJPO5IFSPVOEPQFO-
ing is the Brightfield position.
1 "TOP LEFT" marking on the diaphragm slide
2 Click stops
Note:
3 IMC slit diaphragm slider for S80/0.30
The IMC slit diaphragm slider is encoded. As soon as
the slider is taken out of the Brightfield position (avail- 4 IMC slit diaphragm slider for S40/0.45
able opening) and engaged in the IMC position, the 5 Adjusting screw on the IMC slit diaphragm slider
brightness increases. Conversely, switching from IMC
to Brightfield decreases the brightness. 3
1
t To adjust the slit width, adjust the slider of the IMC slit 2 5
2
diaphragm slider to the position allocated on the ob- 1
jective, e.g. to the 10x marking for the 10x objective.
5
2 2
4
79
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t *OTFSUBTQFDJNFO
t'PDVTUIFJNBHFVTJOHUIFGPDVTEJBMT
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC
Fine adjustment:
For a tiltable Transmitted Light arm: IMC module for tiltable Transmitted Light arm
t 5IFmOFBEKVTUNFOUJTNBEFVTJOHUIFLOVSMFETDSFX
on the IMC modulator and the polarizer.
t 6TJOHUIFTMJEFS
ZPVDBOBMTPPGGTFUUIFIFJHIUUPBE-
just the slit diaphragms to the focal plane. This setting
depends on the specimen being observed.
Note:
The position of the polarizer influences the appearance of
the relief. The best relief images are obtained in the extinc-
tion position; the relief disappears at a 90° position.
Note:
For adjusting the aperture diaphragms, refer to the Installa-
tion Manual.
80
5. OPERATION - FLUORESCENCE
t $BVUJPO'PSUIF*OUFHSBUFE.PEVMBUJPO$POUSBTU
mSTUFO-
able the DIC button!
t *OTFSUBTQFDJNFO
t *G OFDFTTBSZ
SFNPWF UIF DPWFS PG UIF IPMEFS BOE JOTFSU IMC module for tiltable Transmitted Light arm
the IMC module evenly from the front on the left side of
1 Position for objectives with eyepoint C
the microscope.
2 Position for objectives with eyepoint B
t 4MJEFUIF*.$NPEVMFJOUPUIFDPSSFDUeyepoint B or C. 3 Brightfield position
t 'PDVT UIF JNBHF VTJOH UIF SPUBSZ LOPC PO UIF -FJDB 4 Height adjustment
SmartMove or using the focus dial. 5 Fine adjustment
t$POmHVSFUIFCSJHIUOFTTPOUIFJOUFOTJUZSPUBSZLOPC 3
2
1
t 5IFfine adjustment is made using the knurled screw (5)
and the polarizer.
4
Note:
5
The position of the polarizer influences the appearance of
the relief. The best relief images are obtained in the extinc-
tion position; the relief disappears at a 90° position.
t 6TJOHUIFTMJEFS
ZPVDBOBMTPPGGTFUUIFIFJHIUUPBE-
just the slit diaphragms to the focal plane. This setting de-
pends on the specimen being observed.
Note:
For adjusting the aperture diaphragms, refer to the Installa-
tion Manual.
81
5. OPERATION - TRANSMITTED LIGHT CONTRAST METHOD
t *GOFDFTTBSZ
QVMMPVUUIFBOBMZ[FSTMJEFS
DPNQMFUFMZUP 3
open the beam path.
If the Leica DMi8 is equipped with a 2 port infinity port (oS. 44) 4 Manual setting for FIM
switch over to the current port .oS. 58 5 Connection for light guide
6 Diaphragm module
t 8IFOVTJOHUIFFluorescence Intensity Manager (FIM):
The brightness is adjusted in 5 defined increments on the 7 Receptacle for EFW or structured illumination
FIM slider (4).
100% / 55% / 30% / 17% / 10%
t " DPMPS NBSL PO UIF FEHF PG UIF EJTL JOEJDBUFT UIF DVS-
rent filter cube. If the shipped configuration is changed by
inserting another filter cube, note the color of the position
where the filter cube was inserted so you can allocate it 5 4 6 7 3
later.
82
5. OPERATION - FLUORESCENCE
t 'PSJODJEFOUMJHIU TFMFDUUIFDPOUSBTUNFUIPEFLUO.
t *OTFSUBTQFDJNFO
t 4FMFDUBTVJUBCMFPCKFDUJWFJOUIFNBHOJmDBUJPONFOV
83
5. OPERATION - FLUORESCENCE
t :PVNBZQSPUFDUZPVSTQFDJNFOGSPNGBEJOHCZDMPTJOHUIF
incident light shutter.
To do so, push the IL shutter button.
t 'PDVT UIF JNBHF VTJOH UIF SPUBSZ LOPC PO UIF -FJDB
SmartMove or using the focus dial.
84
5. OPERATION - COMBINATION METHOD
t *OTFSUBTQFDJNFO
t 4FMFDUBTVJUBCMFPCKFDUJWFJOUIFNBHOJmDBUJPONFOV
t 'PSUIFDPNCJOBUJPO
TFMFDUUIFDPOUSBTUNFUIPEFLUO-PH
or Fluo-DIC.
FLUO-PH:
t 5IFDVSSFOUnVPSFTDFODFmMUFSDVCFJTEJTQMBZFEPOUIF
Touch Screen. If necessary, push the button for the de-
sired fluorescence filter cube.
t 5IFJMMVNJOBUJPOTFUUJOHTGPSUIFnVPSFTDFODFBOE5SBOT-
mitted Light axis can be adjusted separately.
To do so, toggle between the using the TL/Fluor button.
FLUO-DIC:
t 4XJUDI CFUXFFO UIF UXP DPOUSBTU NFUIPET CZ QSFTTJOH
the FLUO-DIC key.
t *O'-60NPEF
UIFDVSSFOUnVPSFTDFODFmMUFSDVCFJTEJT-
played on the Touch Screen. If necessary, push the but-
ton for the desired fluorescence filter cube.
t *O%*$NPEF
UIF%*$mOFBEKVTUNFOUJTNBEFVTJOHUIF+
and – key.
Note:
The manual analyzer must be used for the FLUO/DIC meth-
od.
85
6. TROUBLESHOOTING
6. Troubleshooting
Problem Cause/remedy
Stand
The microscope does not respond. Ensure that the AC outlet has power.
Make sure that the electronics box or the microscope is
connected to the power supply.
Check the cable connections.
Inform Service and have the supply unit fuse checked.
Illumination
The image is completely dark. Open the shutter.
Check the connection of the lamp housing to the micro-
scope (Transmitted Light).
Ensure that the lamps are not defective.
Check the connection of the EL6000 compact light source
(fluorescence).
Switch over to the active port.
Inform the service department and have them check
whether a fuse is defective.
The image is not uniformly illuminated. Remove all unneeded filters from the light path.
Brightfield
The specimen cannot be brought into focus. Use the correct immersion medium.
Place the specimen with the cover slip downward onto
the stage.
Make sure that the thickness of the cover slip is correct
and corresponds to the specifications on the objective.
Make sure that you are using an objective with cover slip
correction.
Adjust the correction ring if present on the objective.
86
6. TROUBLESHOOTING
Problem Cause/remedy
Darkfield
No definite DF contrast is possible. Be sure that a DF objective is being used.
The objective aperture setting is too high:
Maximum 0.7 for condenser S1
Maximal 0.4 for condenser S23/28/40
If necessary, reduce the objective aperture using the iris
diaphragm on the objective.
Check the condenser centering.
Unwanted stray light. Clean the specimen and neighboring lens surfaces.
Phase Contrast
Phase Contrast cannot be adjusted. The specimen is too thick, too thin or too brightly stained.
Refractive indexes of the mounting medium and specimen
are identical, so that there is no phase jump.
The cover glass is not placed uniformly.
Check that the correct light ring is positioned.
Check that the correct light slider is positioned.
The S40/0.45 condenser has been switched with the
S80/0.30 condenser.
Open the aperture diaphragm completely.
IMC modulator in IMC position.
Check the centering of the light rings.
Polarization
Polarization contrast cannot be adjusted. Bring the polarizer and analyzer into cross position until
they reach maximum darkness (without specimen).
87
6. TROUBLESHOOTING
Problem Cause/remedy
Fluorescence
The image is completely dark (no fluorescence). Open the shutter.
Check the antigen-antibody combination.
Switch over to the active port (or Infinity Port).
The fluorescence is too weak. Unsuitable specimen (stored incorrectly, too old,
bleached).
Unspecific filter combination.
Objective with too low of a numerical aperture.
Eyepiece magnification too high.
Room where microscope is located too bright.
Incorrect beamsplit on the trinocular tube.
Secondary light due to reflection at condenser.
88
7. CARE OF THE MICROSCOPE
and maintenance work! Residual fiber and dust can create unwanted background
Protect electrical components from moisture! fluorescence during fluorescence microscopy.
! Caution:
Note: Thinners containing acetone, xylene or nitrogen can
To protect against dust, cover the microscope and acces- harm the microscope and thus must not be used.
sories with the dust cover after each use.
Test cleaning solutions of unknown composition on a less
visible area of the unit first. Be sure that coated or plastic
surfaces do not become matted or etched.
CAUTION
Cleaning the specimen stage
Let the microscope and lamp housings cool down Remove light-colored spots on the stage by rubbing with
before covering the stand with a dust cover. The dust paraffin oil or acid-free Vaseline.
cover is not heat-resistant. In addition, condensation
may occur.
89
7. CARE OF THE MICROSCOPE
h t t p : / / w w w. l e i c a - m i c r o s y s t e m s . c o m / p r o d u c t s / First, wipe off the immersion oil with a clean cotton cloth,
light-microscopes/life-science-research/inverted-micro- and then re-wipe the surface several times with ethyl alco-
scopes hol.
.
Select the microscope type and go to the "Download" page. 7.3 Handling acids and bases
For examinations using acids or other aggressive chemi-
In case of questions, please contact our Technical Service. cals, particular caution must be taken.
! Caution:
Never allow the optics and mechanical parts to come
into direct contact with these chemicals.
90
8. ESSENTIAL CONSUMABLE AND SPARE PARTS
Additional spare parts such as cover caps, stickers, screws, dust covers etc., on request.
91
9. APPENDIX – ASSIGNING THE VARIABLE FUNCTION KEYS
9. Appendix
9.1 Assigning the variable function keys
The variable function keys can always be defined by the
user via the LAS X software.
Button 2
Button 1
Depending on the configuration, useful functions are al-
ready assigned to the function key upon delivery.
Button 4
Button 3
92
9. APPENDIX – ASSIGNING THE VARIABLE FUNCTION KEYS
Possible assignments of the variable function keys to the stand and SmartMove:
93
9. APPENDIX – ABBREVIATION GLOSSARY
94
9. APPENDIX – OVERVIEW OF TOUCHSCREEN MENUS
Memory function
Left or right port split/top port with further split-
ting
Focus operation
Top port/rear port split
Stage operation
Control
AFC (Adaptive Focus Control)
Saving
Focus position
Delete
Lower threshold For the table: reset working area to the maxi-
mum positioning range of the stage
Objectives
Reload
Valid dry objective For the table: move the current stored position
Valid immersion objective
Various
Shutter
Touch Screen calibration
Incident-light shutter open
95
9. APPENDIX – ADDITIONAL INFORMATION
t 'MVPSFTDFODFUFDIOJRVFT
t 5IFJOTUSVDUJPOTGPSDMFBOJOHUIFNJDSPTDPQFPQUJDT
96