Lecture 3.3 - Introduction To The Synthesis of Nanomaterials - SEM and TEM
Lecture 3.3 - Introduction To The Synthesis of Nanomaterials - SEM and TEM
Lecture 3.2
Characterisation of Nanomaterials:
Introduction to Electron Microscopy
(Scanning Electron Microscopy and
Transmission Electron Microscopy)
• Light Microscopes
• 500 X to 1500 X magnification
• Resolution of ~0.2 µm
• Limits reached by early 1930’s
Li, F., Yao, C., Zheng, Y., & Hou, S. (2018). Facile synthesis of wavy carbon
nanowires via activation-enabled reconstruction and their applications towards
nanoparticles separation and catalysis. RSC Advances, 8(37), 20593-20602.
Resolution
• Ability to reveal closely adjacent structural
details as separate and distinct.
Limit of Resolution
• The main distance between two visible bodies at which
they can be seen as separate and not in contact with
each other
LR = (0.61 x lambda) / NA
Resolution
allows us
to see
Objects as
separate
From one
another
What is Depth of Field?
• The distance of the nearest and farthest object
from the detector if it appears sharp and with
enough details.
Understanding Depth of Field
https://photographylife.com/what-is-depth-of-field
Optical Microscopy vs Scanning Electron
Microscopy
25mm
radiolarian
OM SEM
Small depth of field Large depth of field
Low resolution High resolution
http://www.mse.iastate.edu/microscopy/
Electron Microscopes
• Wavelength of the electron dependent on:
• Electron mass (m)
• Electron charge (q)
• Potential difference to accelerate electrons (V)
h
=
2mqV
Comparison of OM,TEM and SEM
Light source Source of
electrons
Condenser
Magnetic
lenses
Specimen
Objective
Specimen
Projector CRT
Eyepiece Cathode
Ray Tube
detector
OM TEM SEM
• The SEM has a large depth of field, which allows a large amount
of the sample to be in focus at one time and produces an image
that is a good representation of the three-dimensional sample.
The SEM also produces images of high resolution, which means
that closely features can be examined at a high magnification.
W or LaB6 Filament
Thermionic or Field Emission Gun
2- Electromagnetic lenses
All modern SEMs use electromagnetic lenses. These
consist of a coil of copper wires inside iron pole pieces.
solenoid of wire together with a magnetic pole piece that
creates and concentrates a magnetic field.
● Secondary electrons
These originate in the specimen itself,
and have a much lower energy than
the backscattered electrons (typically
< 50 V).
● X-rays
These give information about the
elemental composition of the sample.
Information given by secondary and
backscattered electrons
Objective
lens
Sample stage
Transmission Electron Microscope
1. e-beam strikes
sample and is
transmitted through
the sample
2. Scattering occurs
3. Un-scattered
electrons pass
through sample and
are detected
Transmission Electron Microscope
TEM Copper Grid Sample Holder
Transmission Electron Microscope