Basic CT Training: Reconstruction
Basic CT Training: Reconstruction
36 sheets
(recon.ppt *B)
TUBE
DETECTOR
DAS
CALIBRATION
DATA
Pre-Proc
RAW DATA
Reconstructor
IMAGE DATA
TUBE
DETECTOR
DAS
CALIBRATION
DATA
Pre-Proc
RAW DATA
Reconstructor
IMAGE DATA
Pre-Processing (1/5)
TUBE
ª Input = Pure Raw Data
DETECTOR
CALIBRATION
• Offset correction
Pre-Proc
RAW DATA • Reference correction
• Water Calibration
Reconstructor
• Air Calibration
IMAGE DATA
Pre-Processing (2/5)
TUBE
Pre-Processing (3/5)
TUBE
reference detector.
Pre-Proc
RAW DATA (= very close to tube)
Pre-Processing (4/5)
TUBE Water Calibration
DETECTOR » Calibration data from disk
CALIBRATION
Pre-Proc
RAW DATA
(Calibration data was
taken with water phantom)
Reconstructor
IMAGE DATA
» Is used to set water value
to 0 H (0 Hounsfield).
Pre-Processing (5/5)
TUBE
Air Calibration
DETECTOR
CALIBRATION
Pre-Proc
RAW DATA (Calibration data was
taken during air scan.)
Reconstructor
To HDD
Reconstructor (Raw Data)
Filtering
1 Qn 896
channels
ell
Example orc
ect
1 scan is 900 readings (views) t
de
1
of
e
alu
v 10
© Copyright Toshiba 1998
TSG
1
channels
Qn 896
view n I
1
channels
Qn 896
view 900
I
1
channels
Qn 896
Detector
Array
Detector cell
value of
detector
readout
Value of detector readout
Qn
is ‘smeared’ back on the
reconstruction matrix.
© Copyright Toshiba 1998 12
TSG
Qn
Value of matrix cell + (1* Qn)
Qn ’
Qn’’
4 6
3
7 1
2
5 4 5
3
3 3 7 9 8 14 13 16
7 7 11 13 16 17 19 22
1 2 3 4
QA P1 QB
QB
a
b
α QN
VIEWα
QA P1 QB
a QB
b
QN
α
VIEWα
T
T
T
T
IMPROVEMENT:
IMPROVEMENT:
Filtered
Filtered
Backprojection
Backprojection
Filter function
multiplication Fast Inverse
Fast Fourier Fourier
Transformation Transformation
FFT FFT-1
Input Filter Function Output
Filtered backprojection(FFT)
Amplitude Amplitude of Fourier coefficients
A1 A1
Frequency = f1
t
f1 spatial frequency
Fourier
transform
A2 Frequency = f2 A2
t
f2 spatial frequency
t
f1 f2 spatial frequency
Filtered backprojection(FFT)
Measured
Absorption
value
Real measured profile
in spatial domain
(detector output)
r
(distance in space)
Amplitude
Fourier
coefficients
R
(spatial frequency)
Filtered backprojection
Measurement Filter data Result after
data filtering
part dge
e e
ativ arp
g
e ne a sh
Th l give
wil
Filtered backprojection
T
T
T
T
To HDD
(Raw Data)
After
Afterbackprojection
backprojection
the
theimage
imageis
isin
inthe
the
Reconstructor
Filtering
backprojector
backprojectormemory.
memory.
Reconstructor systems
• FRU
(Fast Reconstruction Unit) Ö8 seconds
• AFRU
(Advanced F.R.U.) Ö7 seconds
• RTRU (Master)
(Real Time R.U.) Ö5 seconds
• RTRU (Master + slave) Ö2 seconds
Viewing
• The values in the backprojector memory are
Hounsfield units (after Sir G.N. Hounsfield).
Viewing
• Some Hounsfield values are :
Bone : 200 - 1000
Dense tumour tissue : 50 - 90
Congealed blood : 55 - 75
Grey matter (brain) : 36 - 46
White matter (brain) : 22 - 32
Blood : 12
Water : 0
Fat : -100
Air : -1000
Viewing
• To view the image on the monitor the H-
values are displayed as gray-values.
Viewing
• Window Width (WW) selects the number of
H-values we see as 256 gray-steps.
(+600)
+500 (+500)
Window Width Ex.: 600H values
Window Width are displayed in
(=400H) (=600H) 256 gray-steps
(+100)
0 (0)
-500
-1000
© Copyright Toshiba 1998 33
TSG
(+600)
+500
Window Level
(= +300H)
0 (0)
(-100)
Window Level
-500 (= -500H)
(-800)
-1000
© Copyright Toshiba 1998 34
TSG
Viewing : WW visual
WW = 100 WW = 1000
WL = 30 WL = 30
Viewing : WL visual
WW = 100 WW = 100
WL = 30 WL = 0