Photon Counting White Paper
Photon Counting White Paper
Photon Counting White Paper
+100V
holes
Index
Requirements for detectors. . . . . . . . . . . . . . . . . . . . . . . . . 3
Detector design. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Readout electronics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Noise rejection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Anti-scatter grids. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
System DQE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2
Requirements
for detectors
Of all medical X-ray applications, mammography puts
the highest demands on the imaging system. As a
result, mammography is the last X-ray-based imaging
technology to convert from film-screen to digital
detector techniques.
3
Available detector
technologies
Conceptually, what differentiates the different detector
technologies is the number of intermediate conversion
steps before a digital signal is formed. This is shown
schematically in Figure 1, which indicates the stages at
which the signals from individual photons are added
(enclosed in green boxes). Each conversion step has
the potential to deteriorate image quality. For systems
that use techniques other than photon counting, the
initial digital signal from the X-rays is converted into an
analog signal, and after the exposure one attempts to
retrieve the original digital signal through an analog-to-
digital converter.
CR plates 2002
• X-ray excitation of electron
states a-Si / CCD 2003
• Accumulation of trapped charge • X-ray conversion to light
• Conversion to light with laser • Light conversion to electrical a-Se
beam charge • X-ray conversion to electrical
• Light conversion into electrical • Charge accumulation in pixel charge Photon counting
analog signal
• Readout of total charge in pixel • Charge accumulation in pixel • X-ray conversion to electrical
• Readout of total charge in pixel pulse
• A/D conversion of signal • A/D conversion of signal • A/D conversion of signal • Incrementation of counter if
charge above threshold
4
Photon counting
technology
X-rays are often perceived as a continuous fluid of
energy. In reality this fluid consists of individual particles,
X-ray photon
so called X-ray photons, which react one-by-one in an
X-ray detector. The X-rays are thus digital by nature.
The Philips MicroDose detector is built to match
this digital nature of X-rays. The Philips MicroDose
detector increments the photon counter directly as
Photon
each individual X-ray is detected. This means that there
counter
are no intermediate steps that can add noise from
the detection of the photon to the pixel value in the
mammogram as indicated in Figure 2. Since there is a
risk in every conversion step of adding noise and losing
signal, significant effort by several companies has been
MicroDose detector
focused on reducing the number of conversion steps.
5
The X-ray interaction in the Si transfers energy to
either a photo-electron or Compton electron that X-ray photons
subsequently excites electron hole pairs. Since only 3.6 +100V
eV is required to create an electron hole pair, several
thousand pairs are created for one incident X-ray
photon and there is no need for further amplification.
A bias voltage is applied across the Si-detector, which electrons
generates a bias field. This electrical field causes the holes
charge cloud to be swept towards electrodes with
positive voltage and the charge cloud of holes to be
swept to electrodes on the opposite side. This drift Figure 4 X-ray generates a cloud of electron-hole pairs that
induces a short electrical signal on the electrodes. The create a short electrical signal.
signal is transferred from the aluminum strip to the
pulse-shaping electronics through wire-bonds.
Readout Electronics
The custom-designed ASIC (Application Specific
Integrated Circuit) contains pre-amplifier, shaper,
comparator and a digital counter. All channels in the
ASIC work in parallel. The shaper creates a signal
around a few hundred nanoseconds in duration. The
comparator discriminates pulses above a certain
threshold level from the detector noise and counts
these pulses. The counter has 15 bits to enable wide
dynamic range. Each ASIC read-out channel can handle
count rates up to 2 MHz, allowing acquisition of images
with sufficient statistics in a few seconds. The number
of counted photons in each channel is accumulated
during a single sampling interval before being sent to the
workstation for image reconstruction and display.
6
Noise rejection
The pulses created in the Philips MicroDose detector by For film-screen systems and scintillator-based digital
the incident photons are easily detected, since the pulse systems, Swank noise can be a serious cause of image
height (5000 electrons) is an order of magnitude higher degradation. This noise is introduced by the statistical
than the electronic noise in the detector (200 electrons variation of released light after the conversion of the
rms). Thus the detector counts the number of peak photon.2 Hence, two photons of the same energy
pulses and nothing else. With no conversion steps from can have different responses in the detector, which
X-ray photons to a digital signal; Swank noise, electronic translates to an overall uncertainty in the integrated
noise and noise from the quantification of the electrical signal. This effect is to some extent present in
signal from the detector can be removed almost entirely. integrating detectors such as amorphous silicon
detectors.
If all external noise sources are eliminated, the system
is said to work at the quantum limit. This is the The absence of Swank noise and the significant removal
ultimate limit for information provided by the laws of of electronic noise is a major advantage of the photon
physics. Ideally the X-ray photons in all pixels should counting technology.
be distributed according to the mathematical Poisson
distribution. These fluctuations are sometimes referred
to as quantum mottle or (wrongly) as quantum noise.
Energy
Threshold
Electronic noise
7
Weighting of X-ray
photons with
different energies
To date, all mammography systems available on the 5 cm thick breast
market, except for the Philips MicroDose system, have 0.8
been integrating systems, i.e. the signal is formed by
a sum of liberated electric charge or light from X-ray 0.7
photon interactions. The amount of charge, or light, is
0.6
proportional to the photon energy. Therefore, with an
integrating technique a higher energy photon will gain
0.5
a higher statistical weight compared to less energetic
Contrast
photons. However, the attenuation in the breast, as 0.4
well as the contrast, is strongly dependent on energy. 0.1 mm micro-calcification
It is the low-energy photons that carry more contrast 0.3
information and should be given a higher statistical
0.2
weight. This subject was first discussed two decades
ago3 but has not generated much attention. The contrast 0.1 5 mm mass
is proportional to the linear attenuation coefficient of
the object being imaged. 0
10 15 20 25 30 35
E (keV)
The subject contrast is roughly proportional to E-3
(photon energy), practically independent of tissue type Figure 6 Low energy photons carry more contrast information.
and the size of the abnormality being detected. For
integrating systems the signal is proportional to energy,
and hence the statistical weight function is incorrect by
four powers.
8
Anti-scatter grids
The conventional way of reducing the scatter-to-primary
Scattered radiation
(S/P) ratio is to insert a Bucky grid between the breast
and the detector. The anti-scatter grids are comprised of
high aspect-ratio lamellae that absorb the rays that have
Radiation scattered in the breast is a common
been scattered at angles beyond acceptance. Depending
source of image degradation in mammography.
on the geometry, discrimination between scattered
Scattered radiation will reduce the contrast and
photons and primary photons is more or less efficient.
increase the noise, which must be compensated
However, for mammography energies, the grids will also
by increasing the dose to the patient. This effect
remove a significant fraction of the direct photons and
is particularly pronounced for thick breasts. It is
are mechanically very difficult to manufacture.
quantified by the (S/P), which is ideally zero but
can be of the order of unity in practice.
The Bucky factor expresses the increase in patient
dosage that is needed to maintain the exposure to the
detector when a grid is inserted. It is typically 2-3, and
contrast improvements exceeding 80% can be achieved. 5
The grid usually moves during the exposure to prevent it
from appearing in the image, but there are also high strip
density, stationary grids. Grids are usually linear, unable
to reject photons scattered along the absorbing strips.
In recent years, high transmission cellular grids with
air interspaces have exhibited improved performance
and become more widespread. However, studies show
that the use of a grid in conventional mammography
provides far from ideal scatter rejection. To quantify, we
introduce the scatter DQE as the ratio of squared SNR’s
between actual scatter rejection and perfect scatter
rejection. This can be calculated through:
Tp
SRE =
1 +S / P
9
Scanning technology
and scatter rejection
Achieving efficient scatter rejection is much easier for
a scanning system in which a small detector is moved
across the object to acquire the full-field image. The
main principle is, the smaller the detector used, the X-ray tube
lower the risk that a scattered photon will interact. That
means no grid is needed in a scanning system and thus
a major mechanical simplification is attained since the
mechanical construction to achieve a linear translation
Collimator 1
of a slit is much simpler compared to the complicated
mechanics of a Bucky.
Object
Systems are divided into slot and slit scanning systems.
For the former, a 2D detector is used, while the latter Collimator 2
has a linear detector covering only one pixel row. A
multi-slit system has several of these linear slits to
Detector
create better photon economy, i.e. using a larger portion
of the output from the X-ray tube. In this way, the image
acquisition time and/or tube load can be reduced. Moving parts
Scan motion
The load for the X-ray tube will be higher for a scanning
system and this is compensated for by increasing the
diameter of the anode disc from 3 inches (standard Figure 7 Multi-slit scanning technology used in Philips MicroDose.
in mammography) to 4 inches, which is a minor
modification of the relatively simple X-ray tubes
used in mammography.
10
System DQE
Scatter DQE Detector DQE
Scatter DQE shows how well the system rejects Detector DQE is the conventional measure of how a
scattered radiation from the breast. Detection of regular pattern is imaged by the system and especially
scattered X-rays decreases the contrast and degrades shows how this is degraded by noise in the system.
image quality. Conventional Bucky grids are far from Degradation by scattered radiation is not included
perfect and have a scatter DQE of roughly 50%. The here. Neither does this measure show how contrasts
scanning multi-slit geometry of Philips MicroDose in actual breast tissue are imaged.9
provides an excellent scatter rejection and a scatter
DQE of 96%.8
Energy weighting
Energy weighting describes how the information in
the X-ray beam is captured by the detector. The low
energy photons contain more information than the
high energy ones, and should thus be weighted higher
by the detector. Energy-integrating systems do the
opposite, while photon counting detectors put equal
weight on each photon. The energy weighting factor
shows how much efficiency is lost compared to a
photon counting system.4
11
Summary of
characteristics for
Philips MicroDose
• Matches the digital nature of X-rays; each X-ray is
processed individually and external noise is almost
entirely eliminated High Quantum Efficiency
• 50 µm pixel size
• Field of view 24 x 26 cm
• No dead pixels
References
1 Abe, F. et al., 1994. Evidence for Top Quark Production in pp Collisions 7 Dance, D.R., Persliden, J., Carlson, G.A., 1992. Calculation of dose and
at sqrt(s) = 1.8 TeV. Fermilab-PUB-94/116-E. May 16, pp.1-19 contrast for two mammography grids. Physics in Medicine and Biology,
2 Swank, R., 1973. Absorption and noise in X-ray phosphors. Journal of 37(1), pp.235-48.
Applied Physics, 44, pp.4199– 203. 8 Åslund, M., Cederström, B., Lundqvist, M., Danielsson, M., 2006.
3 Tapiovaara, M., Wagner, R., 1985. SNR and DQE analysis of broad Scatter rejection in multi-slit digital Mammography. Medical Physics,
spectrum X-ray imaging. Physics in Medicine and Biology, 30, pp.519–29. 33(4), pp.933-40.
4 Cahn, R.N., Cederström, B., Danielsson, M., Hall, A., Lundqvist, M., 9 Dobbins III, J.T., 2000. Image Quality Metrics for Digital Systems. In:
Nygrenn, D., 1999. Detective Quantum Efficiency dependence on X-ray J. Beutel, H.L. Kundel, R.L. van Metter, ed. 2000. Handbook of Medical
energy weighting in Mammography. Medical Physics, 26(12), pp.2680-3. Imaging, Volume 1. Physics and Psychophysics. Bellingham WA, USA: SPIE
5 Rezentes, P., de Almeida, A., Barnes, G., 1999. Mammography grid Press. pp.161-219
performance. Radiology, 210(1), pp. 227–32.
6 Wagner, R., Barnes, G.T., Askins, B.S., 1980. Effect of reduced
scatter on radiographic information content and patient exposure: A
quantitative demonstration. Medical Physics, 7(1), pp.13-8.
© 2011 Koninklijke Philips Electronics N.V. Philips Healthcare is part of Royal Philips Electronics
All rights are reserved.
www.philips.com/healthcare
[email protected]
Philips Healthcare reserves the right to make changes in specifications and/
or to discontinue any product at any time without notice or obligation
and will not be liable for any consequences resulting from the use of this Printed in The Netherlands
publication. 4522 962 79741 * OCT 2011