K550X Sputter Coater Instruction Manual
K550X Sputter Coater Instruction Manual
Instruction Manual
For technical and applications advice plus our on-line shop for
spares and consumable parts visit www.quorumtech.com
Quorum Technologies Ltd
South Stour Avenue
Ashford
Kent
TN23 7RS Tel: +44 (0) 1233 646332
UK Fax: +44 (0) 1233 640744
Email: [email protected]
Website: http:///www.quorumtech.com
For further information regarding any of the other products designed and manufactured by Quorum
Technologies, contact your local representative or directly to Quorum Technologies at the address
above.
Disclaimer
The components and packages described in this document are mutually compatible and
guaranteed to meet or exceed the published performance specifications. No performance
guarantees, however, can be given in circumstances where these component packages are used
in conjunction with equipment supplied by companies other than Quorum Technologies Ltd.
Quorum Emitech is a wholly owned business of Quorum Technologies Limited. Registered offices Unit 19 Charlwoods Road,
East Grinstead, West Sussex. RH19 2HL. Registered in England and Wales, company No. 4273003.
K550X Sputter Coater
1 Contents
2.3 Conformity
This Equipment of this Design and manufacture and marked CE, conforms with the
requirements of the European Directives EMC 89/336/EEC & LVD 73/23/EEC.
3 Description
3.1.1 Features
• Fully Automatic Control
• Low Voltage Sputtering
• High Resolution Fine Coating (Order of 2nm Gold Grain)
• Special Rotating Stage with full Tilt Facility fitted as standard
• Even Thickness Deposition (Typically 20nm or 200 Angstroms for SEM work)
• 165mm Diameter (6 inches) Chamber
• Can be integrated with Film Thickness Monitor
3.1.2 Benefits
• Easy to operate
• No cooling required
• Precise reproducible coatings
• Fully adaptable to a wide range of specimens
• Repeatable film thickness depositions
• Easy loading and unloading of specimens
• Can pre-set deposition thickness
3.2 Specifications
Specifications of the K550X Sputter Coater Unit
3.2.1 Services
Services Argon - Nominal 5 psi
Electrical Supply 230V 50 Hz 10 Amp max Supply
115V 60 Hz 16 Amp max Supply
Vacuum Pump No 5 Pump or similar (approx 6m3/hour) pumping speed. Oil
mist filter, Vacuum Hose, (included with instrument.) is
needed (see Technologies EK3175)
4 Installation
NOTE: - If you are using existing or alternative vacuum pump, and have any difficulty with
connections, please advise.
4.2 Connections
Connections should only be made in accordance with instructions. Refer To: Figure 4-1
and Table 2.
WARNING - OUTPUT
This is for the pump supply only and is the mains voltage at a maximum of
8 Amps
TITLE FUNCTION
Rocker Switch/ Power Inlet/ Main power on to Instrument.
Fuseholder
Note: - A single phase AC supply with Earth is required - selected to the correct voltage
for the country of operation. Either nominal 240V or nominal 120V. The voltage
and frequency range is:
For fuse ratings and voltages refer to: Table 10 and Table 11 Fuse Listings
If required, carry out process gas connections to rear panel (Refer To: Figure 4-1) with
tubing and connectors provided. The connector is push-fit and will 'snap' into a locked
position. It can be released by depressing the metal tongue. Any gas connected should
be regulated at a nominal pressure of 4 p.s.i. ( approx 0.3 bar)
The electrical input to the Instrument is made with the power lead provided. The
Instrument connection is standard and the lead is fitted with the appropriate plug for the
country of operation.
Ensure the plugs are firmly located. Check the voltage is correct voltage for country of
operation, which should correspond to the voltage label on the Instrument. The
appropriate electrical supplies for countries are given in Appendix 7.5.
The vacuum connection is made by 1 Metre length of vacuum hosing. This is a push-on
fit to the Instrument. Ensure that this is firmly in place to the full length of the vacuum
connector.
NOTE: - If you are using existing or alternative vacuum pump, and have any difficulty
with connections, please advise.
An Oil Mist Filter with metal adapter should be fitted to outlet of vacuum pump (See
Section Spare 6.4 ).
Check that the vacuum pump is filled with correct oil (See Section Spares 6.4 for suitable
type). If the vacuum pump is fitted with an ON/OFF switch, ensure that it is left in the
'ON' position as the Instrument will carry out required control.
Ensure that the HT connector to the lid is pushed firmly in place.
Check process gas by pressing the STOP button. The process gas cylinder output gauge
will drop slightly. The K550X lid will lift 'slightly' when chamber fills with gas.
When the display has returned to the initial layout as above press the START key. The
chamber vacuum reading should achieve 5 x 10-1mbar within 1 minute, and 1 x 10-1
mbar (the bleed trip point) within 2 minutes.
Provided that Pump Hold is NOT enabled (see Section 5.5) the gas bleed portion of the
cycle will commence. the screen display should look like that shown below. If necessary,
adjust needle valve at rear of Instrument to achieve a stable vacuum of 1 x 10-1mbar.
When the bleed time has expired the coating will commence. A blue plasma with the
required current level should be observed. The display should look like that below. Allow
the coating to automatically time-out and vent the chamber to atmosphere. The
instrument should complete the automatic cycle, coating at 25mA for 2 minutes (which
are default settings) at a vacuum 1 x 10-1mbar, which may initially need to be adjusted.
.
Coating Sample at 25 milliamps
NOTE: The sputter cycle is rated for a maximum 50mA for 5 minutes, with a duty cycle of
50%. (Off time - 5 minutes.)
5 Operation
5.1 Display
The diagram below shows the layout of the instrument front panel showing the LCD and
the data entry keys.
0-50mA in 5mA
Coating Current Plasma Current. (Default value 25mA)
Steps
0-4 Minutes in
Coating Time one second Time for the coating. (Default 2 Minutes)
steps
This value determines whether the software
monitors the STOP signal from the FTM. If it is
FTM enabled Yes/No
enabled and there is no FTM connected the
coating will stop immediately.
Pump Hold Whether extra pumping is enabled or not.
Yes/No
Enabled (Default - No)
0-8 Hours in 5 How long extra pumping will last. (Default 10
Pump Hold Time
Second Steps Minutes)
Stage Rotate Rotate the specimen stage during the coating.
Yes/No
Enabled (Default - No)
Table 5: Setting Operating Parameters
When options are modified, the software will by default save these so that they are in
force the next time that the instrument is powered up
For most parameters the value is shown along with the variable name. Us the UP or
DOWN arrow keys to amend the value as required. When the required value is shown
press the ENTER key to accept. To exit from the parameter editing menu press the
STOP key.
A typical coating achieved using these settings should be approximately 15nm (150
Angstroms).
NOTE: These conditions for coating thickness can be more readily determined by
referring to the K550X Deposition Chart, (Appendix Error! Reference source not
found.) allowing for any differences in the spacing of any stub and specimen
height when reading final settings.
25
(nm/minute)
20
15
10
5
0
10 20 30 40 50 60 70 80 90 100
Deposition Current (mA)
After the pressing the START key the following will happen:
• The rotary pump will start and commence the pump down
• When the bleed trip point has been reached, if pump hold is not enabled (see
below) the bleed valve will operate, and chamber vacuum will stabilise for 15
seconds at 1 x 10-1mbar (dependent upon needle valve setting)
• Coat will operate at deposition current of 25mA.
• Coating will stop and the vent valve will open an the chamber will vent to
atmosphere
• If further purging is required, the vent-stop can be operated by pressing the
STOP button while the instrument is idle.
NOTE The STOP button can be pressed at any time during a cycle to abort the
process.
2. If it is required to use a K250 for carbon 'flash' evaporation. If this is the case,
then select Pump Hold and select a fairly long time, perhaps longer than 30
minutes. The K550X is now used as a vacuum chamber for carbon head. Using
Carbon String or Cord, outgas vacuum of 1 x 10-1mbar, and evaporation vacuum
-2
of 7 x 10 mbar or better. After completion of the carbon flash Stop on the
K550X.
When Pump Hold is enabled, the instrument will pump to the bleed trip point as normal,
then the pump hold feature becomes active. The display should look like that below.
The pump hold time has a default value of 10 minutes but can be programmed for up to 8
hours. As stated above the instrument will continue pumping until either the time elapses
or, a key is pressed. The cycle will then continue as normal
The K550X is primarily to provide conductive metal coatings for SEM microscopy. In
such applications it would be common to use Aluminium Specimen Stubs. The main
classification of specimen types is between 'bulk' and 'particulate'.
In the case of 'bulk' specimens, a good bonding to the stub is required. In addition,
although an omni-directional coating is achieved, it may be advantageous to use
adhesives, which are electrically conductive. Silver Dag, a Silver Loaded Conductive
Paint is commonly used, but to achieve a somewhat more substantial bonding, Silver
Loaded Epoxy which has good strength and electrical conductivity is advantageous.
In the case of 'particulate' specimens, depending on the nature, again a thin layer of Silver
Dag is suitable, with the specimens 'sprinkled' on it. Alternatively, a Cyanoacrylate or
double-sided tape can be used. In both cases the mounting medium is of low profile.
The coating should be sufficient to make electrical contact with the specimen and stub. If
this is not the case, it may be necessary to bond using one of the previously mentioned
conducting adhesives.
While the standard settings for Sputter Coating, mentioned in Section 5, Operation, may
be satisfactory for most SEM applications (giving typically 15nm (150 Angstroms)
settings: 25mA. 45mm. 2mins), these can be optimised depending on the specimen. The
objective being to obtain as thin and continuous conductive coating as possible to avoid
obstructing detail, while giving specimen stability and avoiding charging.
For very irregular specimens a slightly thicker coating may be required if charging is
observed. While various settings such as time and spacing can be altered, we would
recommend increasing the coating time, while maintaining the sputtering current and
distances. (Giving typically 21nm (210 Angstroms) settings: 20mA. 45mm. 3mins.)
For less irregular specimens and thinner coatings for fine detail, where charging does not
appear as significant, then reducing the sputtering current would be the preferred method.
(Giving typically 10nm (100 Angstroms) settings: 15mA. 45mm. 2 minutes.)
The heat input from sputtering with the K550X is very small. If it is considered the
specimen is heat sensitive, then a low sputtering current, with longer time to achieve the
necessary coating thickness is recommended, while maintaining maximum specimen
spacing. It is not considered necessary, with the low thermal input and low sputtering
voltage, to pre-cool the specimen stage. Assuming the specimen is stable at room
temperature.
The low sputtering voltage is to achieve low thermal input, and high resolution, small grain
size coatings. Typically the grain size of Gold is of the order of 2nm (20 Angstroms).
However, as we normally require somewhat thicker coatings to achieve electrical
conductivity the final resolution will be somewhat less.
Alternative target materials may suggest smaller grain size, Gold/Palladium is the order of
1.7nm (17 Angstroms), there may not necessarily be a recognisable gain in high
resolution, and such coatings may be susceptible to cracking. Careful consideration
should be given when using alternative material, Gold having proved particularly
successful for the majority of SEM work when utilised in Instruments such as the K550X
CAUTION
Ensure mains electrical power is off during any maintenance and service
activities
6.1 Maintenance
PROCEDURE FREQUENCY
Clean the glass chamber and the 'L' gaskets as required using Velin
tissue and foam cleanser (See Section 6.4), or similar. Monthly
Do not use vacuum grease on 'L' gaskets.
Check vacuum pump oil level. Change oil every 6 months using 1
Monthly
litre of Supergrade 'A'. (See Section 6.4).
Check Oil Mist Filter for saturation. Change every 6 months, or more
regularly as required. (See Section 6.4 for suitable part) (This is a Monthly
disposable plastic filter and cannot be reactivated.)
Check the condition of the target material. The wear will depend on
use. This is mainly over an outer annulus, accounting for some 70%
of the surface area.
Monthly
When the backing plate shows at the edges, it may still be used.
When this becomes excessive and sputtering is affected, the target
should be replaced
Regularly inspect electrical power cords and plugs for general
Regularly
condition
Table 6:- Maintenance Tasks
NOTE: A replacement target exchange service is offered for precious metal recovery.
On return of your original target backing plate, a discount made against your
new purchase.
CAUTION
Ensure mains electrical power is off during any maintenance and
service activities
• Check electronic supplies: The LED in the STOP switch should be on at power
up.
• Check fuses: Refer to Section 6.7.
• Check vacuum pump: Local switch should be in 'On' position.
• Check chamber seating for vacuum leaks.
• Check operating conditions of Instrument controls.
• Check Allen screws to target and connections.
• Check correct conditions for sputtering have been set (i.e. vacuum and gas
pressure).
• Check all connections.
• Check that the LCD is showing the correct display.
• Check Pump Hold Enabled is set to NO.
In the event of all checks proving negative, please contact Quorum Technologies, or your
local Distributor.
An Advance Delivery Modular Exchange Service Scheme is operated for the complete
single module control electronics.
NOTE: Spare items can be obtained from Quorum Technologies or approved Distributor.
Only Quorum Technologies recommended items should be used. For technical
assistance and advice - contact Quorum Technologies.
6.4 Spares
The following are available from Quorum Technologies, or your local distributor, and are
featured in more detail in the current Quorum Technologies Consumables Catalogue.
Copies can be sent on request.
SPARES FOR K550X SPUTTER COATER UNIT CATALOGUE QUANTITY
NUMBER
6.5 Accessories
The following are available from Quorum Technologies, or your local distributor, and are
featured in more detail in the current Quorum Technologies Consumables Catalogue.
Copies can be sent on request.
Amray
½” Dia Pin Stub S8620 Pack 10
1" Dia Pin Stub S8621 Pack 10
Cambridge
½” Dia Pin Stub S8622 Pack 10
1" Dia Pin Stub S8623 Pack 10
1¼ Re-entrant Base Stub S8624 Pack 10
3
1¼ Dia x /8" High Stub S8625 Pack 10
Camscan
½” Dia Pin Stub S8622 Pack 10
1½” Dia Pin Stub S8626 Pack 10
Etec
½” Dia pin stub S8622 Pack 10
Hitachi
15mm Dia x 6mm Stub S8627 Pack 10
25mm Dia x 6mm Stub S8628 Pack 10
32mm Dia x 10mm Stub S8629 Pack 10
I.S.I.
15mm Dia x 10mm Stub S8631 Pack 10
15mm Dia x 15mm Stub S8632 Pack 10
Jeol
10mm Dia x 5mm Stub S8633 Pack 10
10mm Dia x 10mm Stub S8634 Pack 10
15mm Dia x 10mm Stub S8631 Pack 10
15mm Dia x 15mm Stub S8632 Pack 10
12.5mm Dia x 10mm Stub S8635 Pack 10
12.5mm Dia x 5mm Stub S8636 Pack 10
Philips
½” Dia Pin Stub S8622 Pack 10
1" Dia Pin Stub S8623 Pack 10
** If an approved distributor is not known - please contact Quorum Technologies direct for
details.
7 Appendices
Pin2
Earth
Pin 1 Pin 3
Live Neutral
7.3.3 Despatch
When returning equipment the procedures set out in the Operating Instructions must be
followed. For example:
- Drain the vacuum pumps.
- Neutralise the flushing with gas.
- Remove filter elements.
- Seal all outlets.
- Pack glass components safely.
- Pack loose attachments securely for example stages.
- Seal in heavy duty polythene or a bag,
- Despatch in suitable transport container.
* We will not accept any Equipment/Accessories which have been radioactively, explosively, or microbiologically
contaminated without written evidence that such Equipment/Accessories have been decontaminated in the
prescribed manner.
Please list all harmful substances, gases and dangerous by-products, which have come into contact with the
Preparation Equipment and Accessories.
Trade name Chemical name and Danger Precautions First aid measures in
Product name symbol class associated with the event of an
Manufacturer substance. accident.
1.
2.
3.
4.
5.
8 Index
A L
Accessories ................................................. 22 List of Figures ................................................4
adhesives .................................................... 18 List of Tables..................................................4
Aluminium Specimen Stubs ........................ 18 M
Appendices.................................................. 25 Maintenance.................................................19
B O
Benefits ......................................................... 8 Oil Mist Filter ................................................12
C Operation .....................................................14
C.O.S.H.H. .................................................... 5 Operation Functions.....................................14
Changing the Target.................................... 19 P
Check or Modify Parameters....................... 15 Preliminary Checks ......................................10
Coating Current ........................................... 15 Pump Hold Feature......................................17
Coating Time ............................................... 15 Pump Plug Wiring ........................................26
Conformity ..................................................... 6
Connections ................................................ 11 R
Contents ........................................................ 3 Return of Goods...........................................27
Cycle Sequence .......................................... 16 Despatch .........................................................27
General Introduction........................................27
D Health and Safety Declaration.........................27
Declaration of Contamination Form ............ 28 Return Address ...............................................27
Description .................................................... 7 S
Display......................................................... 14 Section Contents............................................3
F SEM Specimen Stubs ..................................23
Fail Safe ........................................................ 6 Service and Maintenance.............................19
Features ........................................................ 7 Services .........................................................8
Fuse Listings ............................................... 24 Setting Operating Parameters .....................15
H Spares..........................................................21
Hazard Signal Words .................................... 6 Specifications .................................................8
Health and Safety.......................................... 5 Sputtering Protocols.....................................18
Sputtering Rate ............................................27
I
T
Index............................................................ 29
Initial Operating Checks .............................. 13 Targets ........................................................21
Installation ................................................... 10 Troubleshooting the K550X .........................20
K W
K150X FTM Option........................................ 9 WEEE Compliance ........................................5
K550X Sputter Coater ................................... 7 World Wide Electrical Supplies....................25
For technical and applications advice plus our on-line shop for spares and consumable parts
visit www.quorumtech.com
Quorum Emitech is a wholly owned business of Quorum Technologies Limited. Registered offices Unit 19 Charlwoods
Road, East Grinstead, West Sussex. RH19 2HL. Registered in England and Wales, company No. 4273003.