Compressor Restart After Power Failure or Cooling Problem
Compressor Restart After Power Failure or Cooling Problem
Products
1.) Customer Mains Power goes off and on (following
Biograph mMR
steps describe what happens when Siemens SIRIUS Aera
relay is used) ESSENZA
• Compressor restarts immediately and runs for about 2-3 Skyra
min. Then it stops due to overheating because the cooling Verio
pump starts just after 5 min. Symphony, A Tim System
• MSUP recognizes that compressor is off and starts a Trio, A Tim System
timeout of 20 min. Avanto
• After 20 min MSUP restarts the compressor (Compressor Espree
Drive signal gets active). Aera24
Avanto fit
2.) Cooling problem at Non DaVinci-Systems ESSENZA Dot
(compressor was overheating) Avanto Dot
• MSUP recognizes that compressor is off. Verio Dot
• MSUP sets Compressor Drive signal inactive. Prisma
• MSUP starts a timeout of 20 min. Prisma fit
• After 20 min MSUP tries to restart the compressor Skyra fit
(Compressor Drive signal gets active). This is only Spectra
successful when cooling problem is gone.
• MSUP will try this restart 3 times. If compressor does not
restart then MSUP stops trying to restart compressor.
• After this the only way to restart the compressor is to switch
it manually off/on. Of course this will only work if the cooling
problem is solved!