0% found this document useful (0 votes)
316 views2 pages

Endress & Hauser Pressure Transmitters: Deltabar S PMD235

235

Uploaded by

Morteza alizadeh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
316 views2 pages

Endress & Hauser Pressure Transmitters: Deltabar S PMD235

235

Uploaded by

Morteza alizadeh
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Endress & Hauser Pressure Transmitters

As distributors for Endress & Hauser, the world's largest private instrumentation company, PCi
are able to offer a selection from their range of top quality pressure transmitters.

Deltabar S PMD235
The Deltabar S PMD235 differential pressure transmitter with
metallic sensor can be used for measuring the flow-rate
(volumetric or mass) of gases, vapours and liquids, level, volume
or mass flow in liquids and the differential pressure monitoring of
filters and pumps.

Zero and span both freely adjustable, with or without referential


pressure, communication with all standard protocols, an optional
four-character pressure display and self-monitoring from sensor to
electronics all ensure the PMD235 is simple to operate.

With standard system pressures of 160 and 420 bar and measuring differential pressure from 10 mbar to
40 bar, the PMD235's excellent long-term stability, accuracy, guaranteed resistance to single-side overload
and Hastelloy C276 diaphragm as standard make it a robust, reliable and versatile DP solution.

Metal Sensor Metal Sensor


10 & 40 mbar From 100 mbar
1. Measuring element The separating
2. Silicon diaphragm 1 diaphragm is
3. Separating deflected on both
diaphragm as nap 2 sides by the acting
diaphragm extended pressure with a
4. Fill fluid 7 filling fluid
5. Integrated overload transmitting the
protection 8 pressure to a
3 resistance bridge
6. Measuring element 4 6 whose output
7. Overload diaphragm 9 voltage, which is
8. Filling fluid proportional to
9. Separating 5 d i f f e r e n t i a l
diaphragm as nap pressure, is then
diaphragm extended measured.

Feature Standard Option


Diaphragm Material C276 Hastelloy 316L St. Steel, Monel, Tantalum
Housing 316L St. Steel Epoxy-coated aluminium
Mounting Configuration Horizontal (T5) - top display Vertical (T4) -side display
Display Blind - no display LCD 4-digit
Operating Temperature -40 to + 120°C
Ambient Temperature -40 to + 85°C
Safety Certification EEx ia EEx d
Material Certification - NACE MR0175 / 3.1.B
Power Supply 11.5 to 45V DC
Process Connection ¼" NPT(F) & 7/16" UNF oval flange Others on request
Cable Entry M20 Others on request
Output 4 - 20mA (Smart)
Communication Protocol HART® Profibus-PA, Foundation Fieldbus
Inaccuracy <0.1% of adjusted measuring range to TD 10:1

Overload 1-sided PN, 2-sided 1.5xPN


Fill Fluid Silicone oil
Pressure
Min Range 0.5 mbar
Transmitters
Endress & Hauser Pressure Transmitters
Instrument Limitations
Nominal
Operating Limits Rec. Span
Range
mbar LRL (mbar) URL (mbar) Min (mbar) Max. (mbar)

10 -10 10 0.5 10

40 -40 40 2 40

100 -100 100 5 100

160 -160 160 8 160

500 -500 500 25 500

1,000 -1,000 1,000 50 1,000

3,000 -3,000 3,000 150 3,000

6,000 -6,000 6,000 300 6,000

16,000 -16,000 16,000 800 16,000

40,000 -40,000 40,000 2,000 40,000

Turn down (TD) =


Nominal value / set span
E.g. Nominal value = 3000 mbar
Set span = 1000 mbar
TD = 3:1

The PMD235 can be fitted to any of PCi's range of


hygienic, chemical and flanged diaphragm seals

Ordering Matrix
Locate the order code for your specification and enter it into the appropriate box. See the pressure
range codes page for further information.
Nominal Range Static Housing Mounting Certification
A: 10mbar F: 1bar Pressure A: St. Steel 316L T5: Horizontal A: None
B: 40mbar G: 3bar 1. 160bar B: Epoxy-coated T4: Vertical B: NACE MR0175 / 3.1.B
C: 100mbar H: 6bar 2: 420bar
D: 160mbar I: 16bar Calibration Display
Units A: No display Process
E: 500mbar J: 40bar
1: mbar/bar B: LCD Display Connection
Diaphragm 2: kPa/MPa 1: 316L St. Steel
Material Communication 2: Hastelloy C276
1: Hastelloy C276 3: mmH2O Protocol
2: St. Steel 316L 4: inch H2O A: HART ® Safety
Quantity 5: kgf/cm2
3: Monel B: Profibus-PA A: EEx ia
Model No 4: Tantalum 6: psi C: Fieldbus B: EEx d

PMD
235

Select the diaphragm seal system you want to have connected to this
transmitter and write in the complete order number in the empty matrix
Can be fitted with DS FS above. Together with the transmitter order number this completes the
Pressure specifications for your instrument.
Transmitters www.pci-instruments.com..... +44 (0) 1704 509090..... [email protected].....