The document provides a standard operating procedure for using a lux meter to measure light levels in production areas, sampling booths, and for optical inspection. It details turning on the lux meter, taking measurements, calculating average lux values, acceptance criteria for different areas, and responsibilities for monitoring lux levels.

Uploaded by

anoushia alvi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
238 views2 pages

Winthrox Laboratories Pvt. LTD: Quality Assurance Department

The document provides a standard operating procedure for using a lux meter to measure light levels in production areas, sampling booths, and for optical inspection. It details turning on the lux meter, taking measurements, calculating average lux values, acceptance criteria for different areas, and responsibilities for monitoring lux levels.

Uploaded by

anoushia alvi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 2

WINTHROX LABORATORIES PVT.

LTD
QUALITY ASSURANCE DEPARTMENT

Title:- STANDARD OPERATING PROCEDURE


Document # :
Subject:- SOP for Operation of the Lux Meter and Monitoring the
Lux Level
Issue Date: 27-01- Review Date: Reversion #: 00 Supersedes:
2021
Prepared by: Checked by: Reviewed by: Approved by:
QC Executive Asst. Production Manager Production Manager QA Manager

1.0 OBJECTIVE:

To provide a documented procedure for operation and determination of Lux.

2.0 SCOP:

This SOP is applicable to determine the lux level in production areas, raw material sampling
and dispensing booth.

3.0 RESPONSIBILITY:

Microbiologist, Pharmacist, EHS Member

4.0 ACCOUNTABILITY:

Head of Department

5.0 PROCEDURE:

5.1 Switch on the lux meter.


5.2 Ensure the initial reading should be zero ‘0’ otherwise adjust the zero by using ‘zero
offset’.
5.3 Hold the light sensor facing toward the light and record four reading –the display will
show the lux value.
5.4 Calculate the lux value by using the formula E0= ¼ (Ei1+Ei2+Ei3+Ei4)= ∑Ei.
5.5 Measure the lux at 1 meter height from the ground level.
5.6 Record the readings in lux level monitoring record.

6.0 ACCEPTANCE CRITERIA:

6.1 For production areas the lux value shall not be less than 400 lux.
6.2 For sampling and dispensing booths shall not be less than 300 lux.
6.3 For optical inspection shall not be less then 2000 lux.
6.4 Frequency for monitoring the lux is once in a month.

Page 1 of 2
WINTHROX LABORATORIES PVT. LTD
QUALITY ASSURANCE DEPARTMENT

Title:- STANDARD OPERATING PROCEDURE


Document # :
Subject:- SOP for Operation of the Lux Meter and Monitoring the
Lux Level
Issue Date: 27-01- Review Date: Reversion #: 00 Supersedes:
2021
Prepared by: Checked by: Reviewed by: Approved by:
QC Executive Asst. Production Manager Production Manager QA Manager

Change History:-

Sr. # Revision no. Effective date Reason

01

02

03