Hematology SOPs

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 99
At a glance
Powered by AI
The document discusses various hematology analyzers and their operating procedures. It also outlines specimen requirements, quality control procedures and limitations for different hematology tests.

The document discusses ABX Micros ES60, Sysmex KX-21, Cell-Dyn 1800, CELL DYN 3500-3700 and UDI hematology analyzers.

The specimen requirements mentioned are about 2-3 ml of venous blood collected into EDTA tubes. Specimens should be transported at room temperature 18 - 26oC and can be stored in the refrigerator at 2 - 8oC up to 6 hours.

Hematology Standard Operating Procedures (SOPs)

Hematology Department

STANDARD OPERATING PROCEDURES

(SOPs)

1
Hematology Standard Operating Procedures (SOPs)

Contents
S Subje SOP
No ct NO.
SP 01
1 ABX Micros ES60 Hematology Analyzer
SP 02
2 Sysmex KX-21 Hematology Analyzer
Cell-Dyn 1800 Hematology Analyzer SP 03
3
SP 04
4 CELL DYN 3500-3700 Hematology Analyzer
ACL Automated Coagulation Analyzer (Factor SP 05
5 Deficient Plasma VIII, IX, XI, XII Tests- HemosIL®)

ACL Automated Coagulation SP 06


6
Analyzer (PT-FIB/APTT Tests-
HemosIL®)
SP 07
7 CoaLAB 1000 Automated Coagulation Analyzer
Thrombolyzer compact X SP 08
8
Prothrombin Time (MANUAL TEST) SP 09
9
Partial Thromboplastin Time (manual test) SP 10
10
SP 11
11 Bleeding time
SP 12
12 Clotting time
SP 13
13 Blood film (leishman's stain)
(Slide Method) BIOTEC LABORAYORIES Blood Group SP 14
14
Erythrocyte Sedimentation Rate (Westergren technique) SP 15
15
SP 16
16 Osmotic Fragility Test
SP 17
17 Cerebrospinal Fluid Cellular Examination
SP 18
18 Blood Film \ (Giemsa Stain) Coral Clinical system
SP 19
19 Reticulocytes Count \ (Brilliant cresyl blue) RAL.
Diagnostics
SP 20
20 Sickling test (Na-meta-bisulfate method)
Hematology Standard Operating Procedures (SOPs)
UDI Hematology Analyzer

Version: …1………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Purpose & Definition:

The UDI Hematology Analyzer performs automated blood counts and requires no
manual operations for aspirating blood, dilutions, measuring, calculations, print-outs
and computer transfer of data. It measure the following 18 hematologic parameters:
WBC , LYM% , LYM# , MON% , MON# , GRA% , GRA# , PDW* , PCT* , RBC ,
HGB , HCT , MCV , MCH , MCHC , RDW , PLT , MPV.
Responsibilities:

 Hematology department personal are required to be knowledgeable of this procedure.


 New employees are trained and assessed for competence before they can handle
patient sample.
 The head of the department must resolve any problem with the process and difficulties
in using this SOP.
Specimen requirements:

About 2-3 ml of venous blood collected into EDTA tubes.


Specimens should be transported at room temperature 18 - 26ºC and can be store in the
refrigerator of 2 - 8ºC up to 6 hours. If stored in a refrigerator, samples should be returned
to room temperature, for approximately 30 minutes, before analysis.
Specimen reception:

Reception of request and sample should be recorded, and record time of reception. Pay
attention to sample identification and labeling of tubes.
Criteria for rejection haematology specimens
1. When the identification is missing /inadequate.
2. Insufficient quantity.
3. Inappropriate container.
4. Inappropriate transport/storage.
5. Unknown duration of delay.
6. Clotted sample.

Equipment & Items required:

Diluent: ABX Minidil LMG (10L).


Cleaner: ABX Miniclean (1L).
Lyse: ABX Minilyse LMG (1L).

Abbreviations:

CBC: Complete blood count.


EDTA: Ethylene diamine tetra acetic acid
WBC: White Blood Cells
LYM%: Lymphocyte percentage
LYM#: Lymphocyte absolute value
MON%: Monocyte percentage
MON#: Monocyte absolute value
GRA%: Granulocyte percentage
GRA#: Granulocyte absolute value
RBC: Red Blood Cells
HGB: Hemoglobin
HCT: Hematocrit
MCV: Mean Corpuscular Volume
MCH: Mean Corpuscular Hemoglobin
MCHC: Mean Corpuscular Hemoglobin Concentration
RDW: Red cell Distribution Width
PLT: Platelets
MPV: Mean Platelet Volume
PDW*: Platelet Distribution Width
PCT*: Plateletcrit
NRBCs: Nucleated Red Blood Cells
Procedure:

1. Check operation of the machine, ensuring it is clean and that all required supplies
are present in sufficient quantities.
2. Switch the instrument on by pressing the ON/OFF switch, located on the back of
the instrument.
3. The instrument performs an initialization phase for the internal electronics. Please
wait.
4. Once the initialization phase is complete, the ABX Micros ES60 OT/CT will
automatically run a startup cycle.
5. If the ABX does not automatically run a startup cycle after the initialization phase
is completed, press "Startup" button in the "Status" area to initiate a startup cycle.
6. Then, the instrument will perform a blank cycle for a reference blank count (an
analysis cycle based on reagents without any blood sample).
7. Check and verify that the reference blank counts do not exceed the following
parameter limits: WBC < 0.3, RBC < 0.02, HGB < 0.3, PLT < 10 then: Press "OK"
button to validate blank results.
8. Perform quality control analysis on 3 levels of control blood material (low, normal
and high) to verify that the instrument is performing within the specified ranges of
the quality control material.
9. Entering patient ID, sample ID, Patient name, etc
10. Follow the indications displayed in the "Sample analysis" dialog box to run the
analysis.
a. Mix the sample gently and thoroughly.
b. Remove the cap from the sample tube.
c. Place the sample beneath the sampling needle.
d. Raise up the tube so that the sampling needle lowers into the blood and
press the manual sample bar.
e. The analysis cycle will begin.
11. When the analysis is completed, the "Sample analysis" dialog box is closed and
results are displayed in the "Result display" menu for print out.
12. Dilute the sample if White blood cell counts ≥100,000 /mm3 and platelet counts
≥1,000,000 /mm3 are outside the linearity specifications of the instrument.
Quality control procedures:

1. At the beginning of each work shift all parameters are tested with blood control.
2. The 3 levels include: Abnormal Low, Normal, Abnormal High
3. Controls are stored at 2-8ºC and brought to room temperature on a roller mixer before
use .
4. Controls are gently inverted many times according to the manufacturer’s instruction
before use.
5. From the RUN screen, press [SPECIMEN TYPE].
6. Use the arrow key on the keyboard to move the cursor to the appropriate QC file (i.e.,
low, normal or high) and press the [QC SPECIMEN] key.
7. Control values must be within three standard deviation, otherwise the measurement has
to be repeated, if the control still out of range:
a. Check operation of the machine, ensuring it is clean and that all required supplies
are present in sufficient quantities.
b. Check reagents for expiration dates and lot numbers. Ensure that all machine
lines are in appropriate receptacle where applicable. If this does not solve the
problem:
 Prepare new control(s) and try again.
 If the controls are still out, inform your supervisor to check the operator's manual,
or recalibrate instrument and If controls are still out,. Contact Medical Maintenance
where applicable, or servicing engineer.
8. All control data are managed using software that provides graphical reports (Levey-
Jennings graphs, and monthly cumulative histograms).
Linearity:

PARAMETER LINEAR
RANGE
WBC (103/mm3) 0.4 – 106.6
RBC (106/mm3) 0.2 – 8.1
HGB (g/dl) 0.68 - 26
PLT (103/mm3) 13 – 2777
(A)
PLT (103/mm3) 13 - 4856
(B)
HCT (%) 2.0 - 80

11
Limitations/ Interfering substance:

Verification of any "Abnormal" test result (including flagged results or results outside their
normal range) is due to the following listed:
 WBC White Blood Cells (Leukocytes):
NRBC, Non-lysed Red Cells, Multiple myeloma, Hemolysis, Leukemia, Increased
turbidity, Chemotherapy, Cryoglobulins.
 RBC Red Blood Cells (Erythrocytes):
High WBCs, Agglutinated red blood cells, Cold agglutinins.
 HGB (Hemoglobin):
Turbidity of the blood sample which may be due to Elevated WBC or Elevated Lipids,
Fetal bloods mixed with maternal bloods may produce a falsely elevated hemoglobin
value.
 PLT (Platelets):
Very small erythrocytes, Agglutinated red blood cells, Giant platelets in excessive
numbers, Chemotherapy, Hemolysis, RBC inclusions, Platelet agglutination.
Expected values:

Interpretation of the results:


Certain disease states are defined by an absolute increase or decrease in the number of a
particular type of cell in the bloodstream and many types of anemia.
Limitations/ Interfering substance:

The following is a list of possible substances that may interfere with the listed parameters.
1. WBC: platelet aggregation, giant platelets, nucleated RBCs, cryoglobulins, lyse-
resistant RBCs in patients with haemoglobinopathies, severe liver disease or neonates.
2. RBC: Cold agglutinins, severe micryocytosis, fragmented RBCs, large numbers of
giant platelets, in vitro haemolysis.
3. HGB: Lipemia, abnormal proteins in blood plasma, severe leukocytes (above
100,000/µl). The effect of abnormal proteins and Lipemia may be removed by plasma
replacement or plasma blank procedures.
4. HCT: Cold agglutinins, leukocytosis (above 100,000/µl), abnormal red cell fragility.
5. PLT: Pseudothrombocytopenia, platelet aggregation, increased micrcrocytosis,
megalocyttic platelets
6. Low sample volume of <1 mL may dilute patient samples with EDTA in the collection
tube giving falsely low results. If a low sample volume is expected, use a pediatric
EDTA tube; fill to the second line and mix well.
7. Dilute the sample if White blood cell counts ≥100,000 /mm3 and platelet counts
≥1,000,000 /mm3 are outside the linearity specifications of the instrument

Reporting result:

According to lab policy.( automated printing or computerized)

12
13
15
Automated Coagulation Analyzer
(Factor Deficient Plasma VIII, IX, XI, XII Tests- HemosIL®)
SOPs\ HGA \.......H\ Haem \05

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............

Purpose/Definition:

The human clot junior analyzer is a fully automated, microcomputer-controlled,


nephlometric microcentrifugal instrument capable of performing the following tests:
 PT-FIB (Prothrombin Time and Fibrinogen Level).
 APTT (Activated Partial Thromboplastin Time).
 PT-FIB/APTT (all three tests run simultaneously).
 Single Factors (II, V, VII, VIII, IX, X, XI, XII).

Responsibilities:

 Haematology department personal are required to be knowledgeable of this procedure.


 New employees are trained and assessed for competence before they can handle
patient sample
 The head of the department must resolve any problem with the process and difficulties
in using this SOP.
Specimen Requirements:

 Plasma sample, The anticoagulant of choice for coagulation studies is 3.2% Sodium
Citrate (Blue Top Tube).
 The standard ratio for citrated specimens is nine (9) parts of blood + one (1) part of
anticoagulant, (9:1 ratio) that is critical for valid results.
 Platelet poor plasma required, Set the centrifugation speed to 3,000 rpm and centrifuge
a sample for 10 minutes.
 Separate the plasma from the cells within 30 minutes after centrifugation of the sample
31
and place the plasma sample in a plastic tube.

31
 If the specimen will be tested within two to four hours after centrifugation, it may be
kept at room temperature.
 Freeze the specimen if it cannot be processed within four (4) hours after collection.
 Plasma can be frozen at (-20ºC) or lower for at (7) days or at (-80ºC) for six (6)
months without loss of most factors.
 Thaw rapidly at (37ºC) in a water bath. Remove the plasma sample as soon as it is
thawed and perform the test immediately. Samples are viable for a maximum of (2)
hours at room temperature after they have been thawed .
 Plasma samples cannot be re-frozen.

Specimen reception:

1. Prior to performing coagulation testing on patient samples, verify the patient


identification. Any discrepancy must be investigated before processing the
specimen.
2. Check the specimen for clots, visually hemolyzed and lipemic samples.
3. Ensure the specimen is labeled and label a sample cup.
Criteria for rejecting hematology specimen:
 When the identification is missing /inadequate.
 Clotted specimens.
 Lipemic, icteric or hemolyzed plasma samples
 Incomplete filling of tube or over-filled samples.
 Inappropriate container.
 Unknown duration of delay.
 Inappropriate transport/storage

Abbreviations:

 APTT: Activated Partial Thromboplastin Time


 SD: Standard Deviation

Equipment & Items required:

 ACL Analyzer
 Centrifuge
 Sample cups (0.5mL capacity)
 Rotors
 Plastic test tubes
 Reference Solution
 Factor deficient plasma VIII
 Factor deficient plasma IX
 HemosIL APTT reagent
 HemosIL Calcium Chloride reagent
 HemosIL Normal and Abnormal controls
 HemosIL Calibration plasma
 HemosIL Factor Diluent

Reagents Preparation:
Factor deficient plasmas VIII, IX : Reconstitute all deficient plasmas with 1.0ml of
deionized water. Swirl gently and let stand for 30 minutes at room temperature. Maintain at
2-8ºc for no more than four hours.
APTT Reagent (SynthAFax): APTT reagent comes ready for use. Each vial must be
mixed by inversion several times before use to assure homogeneity of the reagent.
Calcium Chloride (0.025M): Calcium Chloride comes ready for use. Mix well by
inversion before use.

Calibration Procedure:
1. Dilute the calibration plasma (or pooled plasma) 1+4 with factor diluent according
to the program selected.
2. Place diluted calibration plasma in the pool position of the special (factor assay)
sample tray.
3. Place factor diluent in the Dil position. Place factor deficient plasmas in appropriate
positions of the sample tray.
4. Place the APTT reagent in the reagent reservoir No.2 and the calcium chloride in
the reagent reservoir No.3.
5. Select the factor VIII or IX program and follow the instructions for calibration
displayed on the ACL analyzer video screen.

32
Assay Procedure:
1. Dilute patient samples (1+4) with factor diluent (100µl sample with 400µl).
2. According to the sensitivity range required, dilute calibration plasma as follows:
a. High curve (1+4) with factor diluent (100µl+400µl factor diluent).
b. Low curve (1+79) with factor diluent (100µl+7.9ml factor diluent).
3. Place pre-diluted samples in the appropriate positions of the sample tray.
4. Place pre-diluted calibration plasma in the pool position of the sample tray.
5. Place factor diluent in the Dil position of the sample tray.
6. Place factor deficient plasmas in the appropriate position of the sample tray.
7. Place the APTT reagent in the reagent reservoir No.2 and the calcium chloride in
the reagent reservoir No.3.
8. Select the single factor program and follow the instructions displayed on the ACL
analyzer video screen.
Quality Control:

 Normal and abnormal control will be run on each tests.


 Control results must be within the specified (±2SD) limits of the quality control
chart.
 If one or more level of control is outside + 3SD, do not report patient results,
Perform troubleshooting action:
 check reagent levels and expiration dates.
 Repeat the control again, If it is still outside + 3SD:,
 reconstitute new controls,
 check instrument maintenance/cleaning, and repeat.
 If results are still outside limits, notify the Hematology Supervisor
Immediately, Corrective action must be taken before reporting patient results.
 Out of range controls will be recorded on the appropriate (normal/abnormal
control) Service Troubleshooting Log along with the corrective action taken.
Limitations/ Interfering Factors:

1. Hemolysis can cause clotting factor activation.


2. Lipemia and icterus interfere with the spectrophotometric measurements resulting
in falsely decreased end point determinations.
Expected values:

 Reference ranges: 50-150%.

Interpretations of results:

 A Factor VIII deficiency indicates the possible presence of Hemophilia A or von


Willebrand’s disease. Hemophilia A is a sex-linked recessive trait. Hemophilia patients
are classified by the amount of Factor VIII activity measured in their plasma, severe
(<1%), moderate (1-5%) and mild (5-30%). Von Willebrand’s disease is an autosomal
dominant trait exhibiting decreased levels of Factor VIII coagulant activity, affecting
both sexes equally. A differential diagnosis is made based on the results of other
specialized coagulation tests, in conjunction with Factor VIII coagulant activity level.
 Factor IX has a decreased activity in a congenital condition known as hemophilia B or
Christmas Disease, which is sex-linked recessive. The most common cause of an
acquired deficiency of blood clotting factors is hepatic dysfunction due to liver cell
damage or non-availability of vitamin K to the liver. Fibrinogen, factors II, VII, IX, X
and possibly factor V are produced in the liver and all except fibrinogen and factor V
require vitamin K for normal synthesis.
Reporting Results:

 Factor VIII or IX (Activity)........................%


ACL Automated Coagulation Analyzer
(PT-FIB/APTT Tests- HemosIL®)
SOPs\ HGA \.......H\ Haem \06

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............

Purpose/Definition:
The ACL analyzer is a fully automated, microcomputer-controlled, nephlometric
microcentrifugal instrument capable of performing the following tests:
 PT-FIB (Prothrombin Time and Fibrinogen Level).
 APTT (Activated Partial Thromboplastin Time).
 PT-FIB/APTT (all three tests run simultaneously).
 Single Factors (II, V, VII, VIII, IX, X, XI, XII).

Responsibilities:

 Haematology department personal are required to be knowledgeable of this procedure.


 New employees are trained and assessed for competence before they can handle
patient sample
 The head of the department must resolve any problem with the process and difficulties
in using this SOP.
Specimen Requirements:

 Plasma sample, The anticoagulant of choice for coagulation studies is 3.2% Sodium
Citrate (Blue Top Tube).
 The standard ratio for citrated specimens is nine (9) parts of blood + one (1) part of
anticoagulant, (9:1 ratio) that is critical for valid results.
 Platelet poor plasma required, Set the centrifugation speed to 3,000 rpm and centrifuge
a sample for 10 minutes.
 Separate the plasma from the cells within 30 minutes after centrifugation of the sample
and place the plasma sample in a plastic tube.
 If the specimen will be tested within two to four hours after centrifugation, it may be
kept at room temperature.
Specimen reception:

1. Prior to performing coagulation testing on patient samples, verify the patient


identification. Any discrepancy must be investigated before processing the specimen.
2. Check the specimen for clots, visually hemolyzed and lipemic samples.
3. Ensure the specimen is labeled and label a sample cup.
Criteria for rejecting hematology specimen:
 When the identification is missing /inadequate.
 Clotted specimens.
 Lipemic, icteric or hemolyzed plasma samples
 Incomplete filling of tube or over-filled samples.
 Inappropriate container.
 Unknown duration of delay.
 Inappropriate transport/storage

Abbreviations:

 PT-FIB: Prothrombin Time and Fibrinogen Level


 APTT: Activated Partial Thromboplastin Time
 INR: International Normalized Ratio
 ISI: International Sensitivity Index
 DIC: disseminated intravascular coagulation
 VWF: Von Willebrand factor
 SD: Standard Deviation

Equipment & Items required:

 ACL Analyzer
 Centrifuge
 Sample cups (0.5mL capacity)
 Rotors
 Pipettes (1mL capacity)
 Plastic test tubes
 Reference Solution
 HemosIL PT-FIB reagent
 HemosIL APTT reagent
 HemosIL Calcium Chloride reagent
 HemosIL Normal and Abnormal controls
 HemosIL Calibration plasma
 HemosIL Sample Diluent

Calibration:

Calibration is necessary for every new reagent lot (thromboplastin), new lot of reference
emulsion, when indicated by QC information, after major maintenance or service.
Calibration Procedure:

1. Reconstitute 2 vials of lyophilized Calibration Plasma with one (1) mL of Reagent


Grade Water each, Invert gently to mix. Do not shake . Maintain the calibration
plasma at room temperature for 30 minutes before use, stable for two hours from
reconstitution.
2. Fill one 1 mL sample cup with HemosIL Sample Diluent and place it in the
"DILUENT" position in the sample tray. Fill another 1 mL sample cup with both
vials of reconstituted Calibration Plasma and place it in the ―POOL‖ position on the
sample tray.
3. Put the Reference Solution, Thromboplastin reagent into the appropriate positions
on the instrument.
4. Load a new rotor into the rotor holder.
5. Press PROG to return to the READY menu. Select PT-FIB using the ―↑‖ or ―↓‖ and
press ENTER.
6. The "Check" frame is displayed. Press ―↑‖ to start the calibration cycle.
7. Enter the reference values of all parameter then press ENTER. Press ―↓‖
key to start the Calibration.
8. At the end of the analysis, the calibration data, to include results and graphics, will
be printed and stored in the memory of the instrument.
Reagents
Preparation: PT-FIB
HS reagent:
 Reconstitute each vial of thromboplastin reagent with one vial of buffer. Mix by gentle
inversion to ensure complete resuspension (DO NOT SHAKE).
 Label the reconstituted reagent vial with the date and initials of tech placing reagent in
use. Maintain at room temperature for (30) minutes before using. A teflon coated
magnetic stir bar must be inserted into the reagent reservoir for continuous mixing
action.
Stability after reconstitution:
 8 Hours at 15ºC (on the ACL with continuous stirring), 3 days at 2 to 8ºC (in the
original bottle).
 Do not freeze, do not use reagents after the expiration date.
APTT Reagent:
APTT reagent comes ready for use. Each vial must be mixed by inversion several times
before use to assure homogeneity of the reagent.
Calcium Chloride:
Calcium Chloride (0.025 M) comes ready for use. Mix well by inversion before use.

Procedure:

1. From the READY status of the instrument, select the desired test (i.e., PT-FIB, PT-
FIB/APTT or APTT) by using ―↑‖ or ―↓‖ keys and press ENTER.
2. The "Check" frame is displayed.
3. Empty the PT-FIB HS (thromboplastin) vial content into reservoir number 1 on the
instrument. Empty the APTT vial content into reservoir number 2 (APTT) on the
instrument, Empty the Calcium Chloride vial content into reservoir number 3
(CaCl2) on the instrument. Ensure that the Reference Solution volume is sufficient
to perform testing.
4. Press ―↓‖ to continue.
5. Load the sample tray with patient plasma. For PT-FIB/APTT testing, only eight (8)
samples can be programmed to run at one time. For PT-FIB testing, 18 samples
can be programmed and loaded on the sample tray. Calibration plasma loaded in
the pool position of the sample tray.
6. Load a new rotor on the instrument.
7. Press the commands key to start analysis.
8. At the end of analysis the "Results" frame is displayed and results will be printed
out for patient samples. Review the results that are to be filed and certified.
Quality Control:

 Normal and abnormal control will be run on each tests.


 Control results must be within the specified (±2SD) limits of the quality control
chart.
 If one or more level of control is outside + 3SD, do not report patient results,
Perform troubleshooting action:
 check reagent levels and expiration dates.
 Repeat the control again, If it is still outside + 3SD:,
 reconstitute new controls,
 check instrument maintenance/cleaning, and repeat.
 If results are still outside limits, notify the Hematology Supervisor
Immediately, Corrective action must be taken before reporting patient
results.
 Out of range controls will be recorded on the appropriate (normal/abnormal
control) Service Troubleshooting Log along with the corrective action taken.
Limitations/ Interfering Factors:

1. Hemolysis can cause clotting factor activation and falsely decrease end point
measurements.
2. Lipemia and icterus interfere with the spectrophotometric measurements
resulting in falsely decreased end point determinations.
3. Contaminated reagents will give inaccurate results.
4. Samples with a fibrinogen levels <25 mg/dl should be suspected of being a
serum sample and rejected.
5. APTT assay results may be affected by many commonly administrated drugs.

Expected values:

PT seconds: 11.5-15.0
PT activity: 70-120%
INR: 0.90-1.15
Therapeutic levels of INR 2.0 – 3.0 target range 2.5
Fibrinogen: (200-400)mg/dl (2.00-4.00) g/l
PTT seconds: lies between (27-35) seconds.
Note: recommendation that each laboratory determines its own normal range according to
the type of reagents.
Interpretation of results:

Interpretation of PT and PTT in patients with a Bleeding or Clotting Syndrome:


Prolonged PT:
Inherited: Factor VII deficiency
Acquired: Vitamin K deficiency, Liver disease, Warfarin use, Factor VII inhibitor
Prolonged APTT:
Inherited: vWF, factor VIII, IX, XI, XII deficiency
Acquired: Heparin use, Inhibitor of vWF, factor VIII, IX, XI, XII,
Antiphospholipid antibodies
Prolonged both PT and APTT:
Inherited: Prothrombin, fibrinogen, factor V, X or combined factor deficiency
Acquired: Severe Liver disease, Disseminated intravascular coagulation (DIC).
Supratherapeutic heparin or warfarin, Combined heparin or warfarin use, Inhibitor
of prothrombin, fibrinogen, factor V, X, Direct thrombin inhibitor.
Reporting Results:

 The PT and PTT tests are reported in seconds along with the reference ranges.
 PT test results reported in:
 Time (Seconds)

 Ratio (PT patient/PT control)


 Activity Percentage
 INR
 The ACL Analyzer will automatically calculate the INR value when the ISI value is
entered in the ACL. This value will appear beside the PT result.

Note: The International Sensitivity Index (ISI) is an experimentally derived measurement,


usually provided by the thromboplastin manufacturer "according to package insert".

41
CoaLAB 1000 Automated Coagulation Analyzer
(PT-FIB/APTT Tests- LABiTec)
SOPs\ HGA \.......H\ Haem \07

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Purpose/Definition:

The CoaLAB 1000 analyzer, is a fully automated photooptical blood plasma


hemostasis instrument capable for performing a wide range of coagulometric,
chromogenic and immunologic coagulation tests such as Prothrombin time, activated
partial Thromboplastin time, Fibrinogen, special tests such as single factor assays,
Anti-Thrombin III, Protein C, Protein S, C-Reactive Protein, D-Dimer and others
based on the wavelength available in different analyzer types.
Responsibilities:

 Haematology department personal are required to be knowledgeable of this procedure.


 New employees are trained and assessed for competence before they can handle
patient sample
 The head of the department must resolve any problem with the process and difficulties
in using this SOP.
Specimen Requirements:

 Plasma sample, The anticoagulant of choice for coagulation studies is 3.2% Sodium
Citrate (Blue Top Tube).
 The standard ratio for citrated specimens is nine (9) parts of blood + one (1) part of
anticoagulant, (9:1 ratio) that is critical for valid results.
 Platelet poor plasma required, Set the centrifugation speed to 3,000 rpm and centrifuge
a sample for 10 minutes.
 Separate the plasma from the cells within 30 minutes after centrifugation of the sample
and place the plasma sample in a plastic tube.
 If the specimen will be tested within two to four hours after centrifugation, it may be
kept at room temperature.
Specimen reception:

1. Prior to performing coagulation testing on patient samples, verify the patient


identification. Any discrepancy must be investigated before processing the
specimen.
2. Check the specimen for clots, visually hemolyzed and lipemic samples.
3. Ensure the specimen is labeled and label a sample cup.
Criteria for rejecting hematology specimen:
 When the identification is missing /inadequate.
 Clotted specimens.
 Lipemic, icteric or hemolyzed plasma samples
 Incomplete filling of tube or over-filled samples.
 Inappropriate container.
 Unknown duration of delay.
 Inappropriate transport/storage

Abbreviations:

 PT-FIB: Prothrombin Time and Fibrinogen Level


 APTT: Activated Partial Thromboplastin Time
 INR: International Normalized Ratio
 ISI: International Sensitivity Index
 DIC: disseminated intravascular coagulation
 VWF: Von Willebrand factor

Equipment & Items required:

 CoaLAB 1000 analyzer.


 Centrifuge
 Sample-cups (0.5mL capacity)
 Cuvette ring contains a 1 x 4 mm stir bar
 Pipettes (1mL capacity)
 Plastic test tubes
42
 Two containers for rinsing system " Distilled water and waste containers"
 Washing solution

43
 Cleaning solution
 LABiTec PT reagent
 LABiTec APTT reagent
 LABiTec Calcium Chloride reagent
 LABiTec Fibrinogen reagent
 LABiTec Normal and Abnormal controls
 Calibration plasma

Calibration:

 Before starting a routine run operation:


 Check the level of the distilled water container,
 Empty the waste water container.
 Fill liquid system and flush system.
 System Start-up:
 Set the power switch on.
 The analyzer now runs through a full initialization procedure checking all
internal modules automatically don’t replace reagents or washing solution or c-
ring during initialization.
 Once the initialization process is finished the MAIN MENU appears in the
display.
Calibration is necessary for every new reagent lot, when indicated by QC information,
after major maintenance or service.
Use this menu to calibrate a test manually or by automated calibration, to
define or change the calibration parameters for a test.
A calibration is required for test to which the raw values need to be converted
into concentration units/activities.
1. From the Main Menu, press the button Setup to access the Setup Menu.

2. From the Setup Menu, press the button Calibration to access the Calibration
Menu.
3. Select a test for calibration by using the arrow keys first to select the test and
then press the OK button to continue. The calibration for test screen appears.
The analyzer offers two different ways to calibrate a test.

 Under Edit Calibration a test can be calibrated manually by editing the


calibration data previously measured. Enter manually the evaluated results
achieved by a normal measurement with dilutions or if available with standard
plasmas or those which have been provided in the package inserts of the
reagent supplier used.
 Under Auto Calibration a test will be automatically diluted from a standard and
the measured calibration data will be automatically added into a calibration
curve and memorized for the test.
Procedure:

 From the Main Menu, press on the Run Preparation the following display
appears.

 Run preparation:
 Load C-ring: load a new cuvette ring.
 Load reagents: load tests, set reagent vials, re-new reagents, define positions.
 Load samples: load patient samples (Routine & STAT), edit patient ID.
 Rinsing modes: flush, intensive wash, cleaning, maintenance.
 From the Main Menu, press on the Measurement to access the Measurement Menu.
 Start the run
 During run: add STAT runs, show status, Immediate STOP, show results.
 Show results (all, by sample, by test), display curve, print results/graphs.
 From the Main Menu, press the button CuvCARD to load C-ring balance.

Quality Control:
 Normal and abnormal control will be run on each tests.
 Control results must be within the specified (±2SD) limits of the quality control
chart.
 If one or more level of control is outside + 3SD, do not report patient results,
Perform troubleshooting action:
 check reagent levels and expiration dates.
 Repeat the control again, If it is still outside + 3SD:,
 reconstitute new controls,
 check instrument maintenance/cleaning, and repeat.
 If results are still outside limits, notify the Hematology Supervisor
Immediately, Corrective action must be taken before reporting patient
results.
 Out of range controls will be recorded on the appropriate (normal/abnormal
control) Service Troubleshooting Log along with the corrective action taken.
Limitations/ Interfering Factors:

1. Hemolysis can cause clotting factor activation and falsely decrease end point
measurements.
2. Lipemia and icterus interfere with the spectrophotometric measurements
resulting in falsely decreased end point determinations.
3. Contaminated reagents will give inaccurate results.
4. Samples with a fibrinogen levels <25 mg/dl should be suspected of being a
serum sample and rejected.
5. APTT assay results may be affected by many commonly administrated drugs.

Expected values:

PT seconds: (11.5-15.0)
PT activity: ( 70-120)%
INR: ( 0.90-1.15)
Therapeutic levels of INR (2.0 – 3.0) , target range (2.5)
Fibrinogen: (200-400)mg/dl (2.00-4.00) g/l
PTT seconds: lies between (27-35) seconds.
Interpretation of results:

Interpretation of PT and PTT in patients with a Bleeding or Clotting


Syndrome:
 Prolonged PT:
Inherited: Factor VII deficiency
Acquired: Vitamin K deficiency, Liver disease, Warfarin use, Factor VII
inhibitor
 Prolonged APTT:
Inherited: vWF, factor VIII, IX, XI, XII deficiency
Acquired: Heparin use, Inhibitor of vWF, factor VIII, IX, XI, XII,
Antiphospholipid antibodies
 Prolonged both PT and APTT:
Inherited: Prothrombin, fibrinogen, factor V, X or combined factor deficiency
Acquired: Severe Liver disease, Disseminated intravascular coagulation
(DIC).
Supratherapeutic heparin or warfarin, Combined heparin or warfarin use,
Inhibitor of prothrombin, fibrinogen, factor V, X, Direct thrombin inhibitor.
Reporting Results:

 The PT and PTT tests are reported in seconds along with the reference ranges.
 PT test results reported in:
 Time (Seconds)

 Ratio (PTpatient/PTcontrol)
 Activity Percentage
 INR
 The ACL Analyzer will automatically calculate the INR value when the ISI value is
entered in the ACL. This value will appear beside the PT result.

Note: The International Sensitivity Index (ISI) is an experimentally derived


measurement, usually provided by the thromboplastin manufacturer "according to
package insert".
Thrombolyzer compact X
SOPs\ HGA \.......H\ Haem \08

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Purpose & Definition

Thrombolyzer compact x is a classic, fast and reliable coagulation analyzer, it


performs up to 160 tests/ hour, capable of performing the following tests
PT,PTT,FIBRONGEN , ATIII, TT.
Responsibilities:

 Hematology department personal are required to be knowledgeable of this


procedure
 The head of the department must resolve any problem with the process and
difficulties in using this sop
 New employees are trained and assessed for competence before they can
handle patient sample
Specimen requirements:

 Plasma sample, The anticoagulant of choice for coagulation studies is (3.2%)


Sodium Citrate (Blue Top Tube).
 The standard ratio for citrated specimens is nine (9) parts of blood + one (1)
part of anticoagulant, (9:1 ratio) that is critical for valid results.
 Platelet poor plasma required, Set the centrifugation speed to (3,000) rpm and
centrifuge a sample for (10) minutes.
 Separate the plasma from the cells within 30 minutes after centrifugation of the
sample and place the plasma sample in a plastic tube.
 If the specimen will be tested within two to four hours after centrifugation, it
may be kept at room temperature.
Specimen reception:

 Prior to performing coagulation testing on patient samples, verify the patient


identification. Any discrepancy must be investigated before processing the
specimen.
 Check the specimen for clots, visually hemolyzed and lipemic samples.
 Ensure the specimen is labeled and label a sample cup.
Criteria for rejecting hematology specimen:
 When the identification is missing /inadequate.
 Clotted specimens.
 Lipemic, icteric or hemolyzed plasma samples
 Incomplete filling of tube or over-filled samples.
 Inappropriate container.
 Unknown duration of delay.
 Inappropriate transport/storage

Abbreviations:

PT : Prothrombin Time.
APTT: activated thromboplastin time.
TT: thrombin time.
ATIII : anti thrombin III.
INR: International Normalized Ratio
ISI: International Sensitivity Index
Equipment & Items required:

 PT reagent .
 PTT reagent.
 Cacl2 reagent .
 Sample cup.
 Cuvette rack segment
 Dropper .
 Centrifuge.
 Probe cleaner solution
Calibration :

1. Form the "calibration" menu's test window choose the test you want to used, then
press (enter).
2. Move the cursor to the "Manual" box, using the (enter) keys, confirming this box
by pressing (enter) the values can be entered.
3. Should you want to use less calibration points for the calibration, change the values
of the last positions to 0 (zero).
4. Having completed the changes to the table, change the Normal, ISI, Min, and Max
values, if necessary.
5. With all changes done, press (Esc). The cursor moves to the "Curve" box (enter).
You can now view the new curve representation. Press (esc) to exit the graph.
6. The cursor moves to the "Valid" box (enter). If you wish to keep the old values,
change the "Valid: Yes" box using the (space) key to "No" before exiting the menu
with (enter. The cursor moves to the "Manual" box (esc).
Procedure :

1. System start-up: set the power switch on.


2. Before starting a routine run operation:
 check the level of the distilled water container,
 empty the waste water container.
3. The analyzer now runs through a full initialization procedure checking all internal
modules automatically.
4. From the main menu, select patient preparation enter the data on keyboard and
select the desired test (a, b, c, or d) by using ―↑‖ or ―↓‖ keys and press enter.
5. Add enough volume of plasma in sample cup .
6. Put sample cup in the rack sample of thrombolyzer according to their position
7. Press ESC then press F2 to start working the test.
8. Wait for result.
9. Record the result in the result book
Quality Control:

 Normal and abnormal control will be run on each tests.


 Control results must be within the specified (±2SD) limits of the quality
control chart.
 If one or more level of control is outside + 3SD, do not report patient
results,
Perform troubleshooting action:
 check reagent levels and expiration dates.
 Repeat the control again, If it is still outside + 3SD:,
 reconstitute new controls,
 check instrument maintenance/cleaning, and repeat.
 If results are still outside limits, notify the Hematology Supervisor
Immediately, Corrective action must be taken before reporting
patient results.
 Out of range controls will be recorded on the appropriate
(normal/abnormal control) Service Troubleshooting Log along with the
corrective action taken.
Limitations/ Interfering Factors:

1. Hemolysis can cause clotting factor activation and falsely decrease end point
measurements.
2. Lipemia and icterus interfere with the spectrophotometric measurements
resulting in falsely decreased end point determinations.
3. Contaminated reagents will give inaccurate results.
4. Samples with a fibrinogen levels <25 mg/dl should be suspected of being a
serum sample and rejected.
5. APTT assay results may be affected by many commonly administrated drugs.

Expected values:

PT seconds: (11.5-15.0)
INR: (0.90-1.15)
Therapeutic levels of INR (2.0 – 3.0) , target range (2.5)
PTT seconds: lies between (27-35) seconds.

51
However, this varies widely between laboratories and is dependent upon a
number of variables including whether the test is automated or manual, the
type of activator and the incubation times employed in the test.
Note: recommendation that each laboratory determines its own normal range

Interpretation of the results:

Interpretation of PT and PTT in patients with a Bleeding or Clotting Syndrome


Prolonged PT:
Inherited: Factor VII deficiency
Acquired: Vitamin K deficiency, Liver disease, Warfarin use, Factor VII inhibitor
Prolonged APTT:
Inherited: vWF, factor VIII, IX, XI, XII deficiency
Acquired: Heparin use, Inhibitor of vWF, factor VIII, IX, XI, XII,
Antiphospholipid antibodies
Prolonged both PT and APTT:
Inherited: Prothrombin, fibrinogen, factor V, X or combined factor deficiency
Acquired: Severe Liver disease, Disseminated intravascular coagulation (DIC).
Supratherapeutic heparin or warfarin, Combined heparin or warfarin use, Inhibitor of
prothrombin, fibrinogen, factor V, X, Direct thrombin inhibitor.
Reporting result:

 The PT and PTT tests are reported in seconds along with the reference ranges.
 PT test results reported in:
 Time (Seconds)
 INR
 The thrombolyzer Analyzer will automatically calculate the INR value when the
ISI value is entered in the thrombolyzer This value will appear beside the PT
result.

Note: The International Sensitivity Index (ISI) is an experimentally derived measurement,


usually provided by the thromboplastin manufacturer "according to package insert"
PROTHROMBIN TIME (MANUAL TEST)
SOPs\ HGA \.......H\ Haem \09

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Definition & Purpose :

 The prothrombin time is the time required for the plasma to clot after an excess of
thromboplastin and an optimal concentration of calcium have been added.
 The PT measures functional activity of the extrinsic and common pathways (VII, V,
and X, prothrombin, and fibrinogen).
 PT is the most widely used method for monitoring patients receiving oral anticoagulant
as warfarin therapy.
Responsibilities:

 Hematology department personal are required to be knowledgeable of this procedure


 The head of the department must resolve any problem with the process and difficulties
in using this sop
 New employees are trained and assessed for competence before they can handle
patient sample
Specimen Requirements:

 Plasma sample, The anticoagulant of choice for coagulation studies is 3.2% Sodium
Citrate (Blue Top Tube).
 The standard ratio for citrated specimens is nine (9) parts of blood + one (1) part of
anticoagulant, (9:1 ratio) that is critical for valid results.
 Platelet poor plasma required, Set the centrifugation speed to 3,000 rpm and centrifuge
a sample for 10 minutes.
 Separate the plasma from the cells within 30 minutes after centrifugation of the sample
and place the plasma sample in a plastic tube.
 If the specimen will be tested within two to four hours after centrifugation, it may be
kept at room temperature.

52
Specimen reception:

1. Prior to performing coagulation testing on patient samples, verify the patient


identification. Any discrepancy must be investigated before processing the
specimen.
2. Check the specimen for clots, visually hemolyzed and lipemic samples.
3. Ensure the specimen is labeled and label a sample cup.
Criteria for rejecting hematology specimen:
 When the identification is missing /inadequate.
 Clotted specimens.
 Hemolyzed plasma samples
 Incomplete filling of tube or over-filled samples.
 Inappropriate container.
 Unknown duration of delay.
 Inappropriate transport/storage

Abbreviations:

 PT : prothrombin time.
 I.N.R : international normalized ratio
 ISI : international sensitivity index
 DIC : disseminated intravascular coagulation
 VWF : Von Willebrand factor
 SD : Standard Deviation

Equipment & Items required:

Calcified thromboplastin reagent, Fresh pooled plasma normal control, abnormal control,
Water bath at 37˚C, Tissue paper, Stop watch, Glass tube, Yellow tips , Micro pipette 50-
200 µl.
Procedure:

1. Reconstitute tissue thromboplastin reagent according to instructions. Label the


thromboplastin with the time, date and initials.
2. Pipet 100 µL of plasma into test tubes.
3. Allow at least 1-3 minute for the control and sample to reach 37 o C in water bath.
4. Pipet 200 µL of PT reagent into the tube containing the plasma. Start the stop
watch simultaneously.
5. Mix the tube and leave in the water bath for a minimum of 7-8 seconds. Then
remove, wipe the exterior, tilt back and forth gently until a visible clot is formed.
As the clot forms, the mixture will gelatinize and may turn cloudy.
6. Stop the stop watch immediately when the clot begins to form and record the time
in seconds.
7. Repeat the procedure for the second run of plasma to confirm result and Record the
time.
8. If results are not within required limits, a third run should be performed and
average the two that match within acceptable limits.
Quality control procedures:

The normal control and the abnormal control will be run on each tests. The control results
must be within the specified (±2SD) limits of the quality control chart.
 If one or more level of control is outside + 3SD, do not report patient results.
Perform troubleshooting action:
 check reagent levels and expiration dates.
 Repeat the control again. If it is still outside + 3SD,
 reconstitute new controls and reagents,
 check water bath temperature, and repeat.
 If results are still outside limits, notify the Hematology Supervisor immediately.
Corrective action must taken before reporting patient results.
 Out of range controls will be recorded on the appropriate (normal/abnormal
control) Service Troubleshooting Log along with the corrective action taken.
Expected values:

The normal values for the prothrombin time range from 11.5 to 15.0 seconds and it is
different from lab to lab depend on the source of thromboplastin used.
PT seconds: 11.5-15.0.
PT activity: 70-120%
INR: 0.90-1.15
Therapeutic levels of INR 2.0 – 3.0 target range 2.5
Note: recommendation that each laboratory determines its own normal range.

Limitations/ Interfering substance:

1. Associated with specimen (Pre-analytical)


 Inappropropriate ratio of anticoagulant to blood
 Delay in testing or processing
 Inappropriate storage
2. Associated with Reagent (Analytical)
 Incorrect preparation of reagents
 Use of reagents beyond reconstituted stability time or expiration date
 Contaminated reagent.
3. Associated with procedure
 Incorrect temperature
 Incorrect incubation times.
 Incorrect volumes of sample, reagents or both
Also test may be affected by :
 Severe diarrhea or vomiting that causes fluid loss and dehydration, this may
increase the INR.
 Getting a lot of vitamin k may decrease the INR.
 Medication: antibiotics, aspirin, cimetidine.

Interpretation of the results:

Interpretation of PT and PTT in patients with a Bleeding or Clotting Syndrome


Prolonged PT:
Inherited: Factor VII deficiency
Acquired: Vitamin K deficiency, Liver disease, Warfarin use, Factor VII inhibitor
Prolonged APTT:
Inherited: vWF, factor VIII, IX, XI, XII deficiency
Acquired: Heparin use, Inhibitor of vWF, factor VIII, IX, XI, XII,
Antiphospholipid antibodies
Prolonged both PT and APTT:
Inherited: Prothrombin, fibrinogen, factor V, X or combined factor deficiency
Acquired: Severe Liver disease, Disseminated intravascular coagulation (DIC).
Supratherapeutic heparin or warfarin, Combined heparin or warfarin use, Inhibitor of
prothrombin, fibrinogen, factor V, X, Direct thrombin inhibitor.
Reporting Results:

 The PT is reported in seconds along with the reference ranges.


 PT test results reported in:
 Time (Seconds)

 Ratio (PT patient/PT control)


 Activity Percentage
 INR

Note: The International Sensitivity Index (ISI) is an experimentally derived measurement,


usually provided by the thromboplastin manufacturer "according to package insert".
PARTIAL THROMBOPLASTIN TIME (MANUAL TEST)
SOPs\ HGA \.......H\ Haem \10

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Definition & Purpose :

The activated partial thromboplastin time is the time required for plasma to clot when
maximal surface contact activation, optimal phospholipid, and calcium concentration are
provided. The PTT measures functional activity of the intrinsic and common pathway as
well as screen for coagulation inhibitors such as the lupus-like anticoagulant. APTT is the
most widely used method for monitoring intravenous heparin anticoagulant therapy.
Responsibilities:

 Haematology department personal are required to be knowledgeable of this procedure.


 New employees are trained and assessed for competence before they can handle
patient sample
 The head of the department must resolve any problem with the process and difficulties
in using this SOP.
Specimen Requirements:

 Plasma sample, The anticoagulant of choice for coagulation studies is 3.2% Sodium
Citrate (Blue Top Tube).
 The standard ratio for citrated specimens is nine (9) parts of blood + one (1) part of
anticoagulant, (9:1 ratio) that is critical for valid results.
 Platelet poor plasma required, Set the centrifugation speed to 3,000 rpm and centrifuge
a sample for 10 minutes.
 Separate the plasma from the cells within 30 minutes after centrifugation of the sample
and place the plasma sample in a plastic tube.
 If the specimen will be tested within two to four hours after centrifugation, it may be
kept at room temperature.
Specimen reception:

1. Prior to performing coagulation testing on patient samples, verify the patient


identification. Any discrepancy must be investigated before processing the
specimen.
2. Check the specimen for clots, visually hemolyzed and lipemic samples.
3. Ensure the specimen is labeled and label a sample cup.
Criteria for rejecting hematology specimen:
 When the identification is missing /inadequate.
 Clotted specimens.
 Hemolyzed plasma samples
 Incomplete filling of tube or over-filled samples.
 Inappropriate container.
 Unknown duration of delay.
 Inappropriate transport/storage

Abbreviations:

 APTT (activated partial thromboplastin time).


 Cacl2 (calcium chloride)
 DIC (disseminated intravascular coagulation).
 VWF (Von Willebrand factor)

Equipment & Items required:

Partial thromboplastin reagent: consists of phospholipids and a contact activator, Cacl 2


(0.025M), Fresh pooled plasma normal control, abnormal control, Water bath at 37˚C,
Tissue paper, Stop watch, Glass tube, Yellow tips , Micro pipette 50-200 µl.
Procedure:

1. Pre-warm a sufficient quantity of 0.025M calcium chloride reagent to 37˚C.


2. Pipette 100 µL of patient plasma into a labeled test tube.
3. Into test tube, add 100 µL of partial thromboplastin reagent.
4. Incubate the patient plasma/partial thromboplastin mixture at 37 oC for a minimum
of three (3) minutes (optimum activation of contact factors).
5. Add 100 µL of calcium chloride into the patient plasma / partial thromboplastin
mixture and start the stop watch immediately.
6. After 20 seconds, remove the tube from the water bath. Wipe off the outside of the
tube. Gently tilt the tube back and forth until a visible clot forms.
7. Immediately record the time in seconds.
8. The result must be run in duplicate.

Quality control procedures:

The normal control and the abnormal control will be run on each tests. The control results
must be within the specified (±2SD) limits of the quality control chart.
 If one or more level of control is outside + 3SD, do not report patient results.
Perform troubleshooting action:
 check reagent levels and expiration dates.
 Repeat the control again. If it is still outside + 3SD,:
 reconstitute new controls and reagents,
 check water bath temperature, and repeat.
 If results are still outside limits, notify the Hematology Supervisor immediately.
Corrective action must taken before reporting patient results.
 Out of range controls will be recorded on the appropriate (normal/abnormal
control) Service Troubleshooting Log along with the corrective action taken.
Expected values:

PTT reported in seconds: lies between 27-35 seconds. However, this varies widely
between laboratories and is dependent upon a number of variables including whether the
test is automated or manual, the type of activator and the incubation times employed in the
test.
Note: recommendation that each laboratory determines its own normal range.
Limitations/ Interfering Factors:

1. Associated with specimen (Pre-analytical)


 Inappropropriate ratio of anticoagulant to blood
 Delay in testing or processing
 Inappropriate storage
2. Associated with Reagent (Analytical)
 Incorrect preparation of reagents
 Use of reagents beyond reconstituted stability time or expiration date
 Contaminated reagent.
3. Associated with procedure
 Incorrect temperature
 Incorrect incubation times (↑ incubation time=↓PTT due to contact
activation and > 5min heating will result in loss of heat-labile factor V)
 Incorrect volumes of sample, reagents or both
4. Some drugs such as heparin, antihistamine ascorbic acid ,chlorpromazine and
salicylates.
Interpretation of the results:

Interpretation of PT and PTT in patients with a Bleeding or Clotting Syndrome


Prolonged PT
Inherited: Factor VII deficiency
Acquired: Vitamin K deficiency, Liver disease, Warfarin use, Factor VII inhibitor
Prolonged APTT
Inherited: vWF, factor VIII, IX, XI, XII deficiency
Acquired: Heparin use, Inhibitor of vWF, factor VIII, IX, XI, XII,
Antiphospholipid antibodies
Prolonged both PT and APTT
Inherited: Prothrombin, fibrinogen, factor V, X or combined factor deficiency
Acquired: Severe Liver disease, Disseminated intravascular coagulation (DIC).
Supratherapeutic heparin or warfarin, Combined heparin or warfarin use, Inhibitor of
prothrombin, fibrinogen, factor V, X, Direct thrombin inhibitor.
Reporting Results: PTT reported in seconds.
61
Bleeding time
SOPs\ HGA \.......H\ Haem \11

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Purpose & Definition:

 Bleeding time is defined as the time taken for a standard skin wound to stop
bleeding upon vessel injury , platelets adhere and form a haemostatic platelet plug.
bleeding time measures the ability of these platelets to arrest bleeding and therefore
measures platelets function as well as the integrity of the vessel wall.
 Bleeding time: it is a test for :
1. Capillary response to injury.
2. Platelet function: stick to each other and form plug (aggregate), and break and
release thromboplastin.
 There are several methods of performing the bleeding time:
1- Duke method
2- IVY method.

Responsibilities:

 Haematology department personal are required to be knowledgeable of this procedure


 The head of the department must resolve any problem with the process and difficulties
in using this sop
 New employees are trained and assessed for competence before they can handle
patient sample
Specimen requirements:

Patient Preparation: Check the patient's history for recent use of drugs that prolong
bleeding time. If the test is being used to identify a suspected bleeding disorder, it should
be postponed and the drugs discontinued. If the test is being used preoperatively to assess
hemostatic function, it should proceed as scheduled.
Specimen reception:

1. Explain to the patient this test is used to measure the time required to form a clot
and stop bleeding.
2. Reassure the patient that, although he may feel some discomfort from the incisions,
the antiseptic, advise the patient that the incisions will leave two small, hairline
scars that should be barely visible when healed.
Equipment & Items required:

Blood pressure cuff, disposable lancet, 70% alcohol, filter paper, bandage, stopwatch.

Abbreviations:

B.T: Bleeding time

Procedures:

Duke method
1. Gently clean the lobe of the ear with cotton wool and alcohol, do not rub. Allow to
dry.
2. Puncture the ear lobe, with the lancet, making the incision 2-4mm deep the blood
should flow freely, without any need to squeeze the ear lobe. Start the stopwatch.
3. After 30 seconds collect the first drop of blood on a corner of the filter-paper
4. Do not touch the skin with the paper.
5. Wait 30 seconds more. Collect the second drop of blood in the same way, a little
further along the strip of paper.
6. Continue to collect one more drop of blood every 30 seconds. The drops become
progressively smaller.
7. When no more blood appears, stop the timer and record the time.
Ivy method
1. Blood pressure cuff is placed on arm and inflated to 40mm Hg
2. With pressure maintained, the arm is cleaned and a lancet is again used to make a
2-4 mm cut on the volar aspect of the forearm, being careful NOT to cut any visible
blood vessels.
3. Filter paper is used to blot blood every 30 seconds until it stops flowing.
4. If bleeding continues for more than 15 minutes, the procedure should be
discontinued.

62
Limitations/ Interfering substance:

1. If the patient has taken aspirin or aspirin-containing compounds 7 to 10 days prior


to the procedure, the bleeding time may be prolonged.
2. Other medications such as dextran, streptokinase, and streptodornase may also
affect the bleeding time.
3. Results may be affected by an improperly performed puncture. A puncture that is
too shallow, too deep, or in an inappropriate location will adversely affect test
results.
4. The alcohol must be completely dried before making the puncture. If residual
alcohol is on a puncture site, the bleeding time will be erroneously prolonged.
5. If the phlebotomist allows the filter paper to touch the wound, the platelet clot may
be dislodged, causing falsely elevated results.
6. Blood pressure cuff should be maintained exactly at (40mm Hg).

Expected values:

Duke method: 1–3 minutes


Ivy method: 3 to 6 minutes
Interpretation of the results:
A. Abnormal increased results could be due to:
1. Thrombocytopenia, with failure to produce a platelet plug.
2. Poor platelet function
3. Failure of vessel constriction on injury
4. Puncture of larger blood vessel (technical error)
5. Disturbing the clot (technical error)
B. Prolonged bleeding time may indicate the presence of disorders associated with
such as Hodgkin's disease, acute leukemia, disseminated intra vascular coagulation,
hemolytic disease of the newborn, liver disease- end stage and in some cases of
Von Willebrand Disease, Hypofibrinogenaemia, Bernard-Soulier disease and
Glanzmann's thrombasthenia.
C. Decreased BT: Not clinically significant.

Reporting result:

Bleeding time: ………minute................sec


Clotting time
SOPs\ HGA \.......H\ Haem \12

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Purpose & Definition:

It is the time required for blood to clot without the presence of any substance .
The whole blood clotting time is a rough measure of all intrinsic clotting factors in the
absence of tissue factors. Variations are wide and the test sensitivity is limited. Whole
blood, when removed from the vascular system and exposed to a foreign surface, will form
a solid clot. Within limits, the time required for the formation of the solid clot is a measure
of the coagulation system.
There are various methods for determining the clotting time, the most common being
the capillary tube method and slide method.
Responsibilities:

 Haematology department personal are required to be knowledgeable of this procedure


 The head of the department must resolve any problem with the process and difficulties
in using this sop
 New employees are trained and assessed for competence before they can handle patient
sample
Specimen requirements:

1- About 3 drops of blood on a glass slide and examining it for clot formation.
2- About 200 ul blood in capillary tube.
Specimen reception:

Explain to the patient this test is used to measure the time required to form a clot and
explain how the test will be performed.
Equipment & Items required:

Alcohol swab, lancet, glass slide, stopwatch and capillary tubes.

Abbreviations:

C.T: Clotting time

Procedures:

Capillary tube method.


1. Clean the patient finger with alcohol swab and allow to dry.
2. Pricked the finger by lancet, remove the first drop of blood.
3. Squeeze the finger to obtain a larger drop of blood and fill two capillary tubes with
blood.
4. After one minute start breaking small pieces of the capillary tube every 30 second
until a fibrin thread is seen between the two broken ends.
Slide method
1. Clean one finger with the alcohol swab then puncture it with the lancet.
2. Place a large drop of blood on the glass slide. Start the stopwatch.
3. After one minute, dip the lancet all the way into the blood drop then slowly lift the
lancet out.
4. If you see a gel-like thread of fibrin (blood clot protein) the blood has clotted.
5. Record the clotting time.
6. If no piece of fibrin is observed, keep repeating the dipping procedure until the
thread is observed.
Limitations/ Interfering substance:

 The test sensitive only in extreme factor deficiencies, and is insensitive to high
doses of heparin.
 The following variables tend to decrease the clotting time: rough handling of the
blood specimen, presence of tissue fluids (traumatic venipuncture).
Expected values:

Slide method (2 to 7 ) minutes


Capillary tube method (2 to 8) minutes
Interpretation of the results:

 In coagulation disorders like hemophilia, clotting time is prolonged but bleeding


time remains normal.
 Clotting time is also prolonged in conditions like vitamin K deficiency, liver
diseases, disseminated intravascular coagulation, overdosage of anticoagulants etc.
Reporting result:

Clotting time ……….minutes : ……… seconds.


Blood film (leishman's stain)
SOPs\ HGA \.......H\ Haem \ 13

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Purpose & Definition:

 Describe the procedure for preparing thin blood smear and staining with leshimans stain
for the purpose of diagnosis of different anemia's, leukemia's, viral and parasitic
infections, in addition associated with inclusion bodies, allergic reactions,
hemoglobinopathies, and platelets disorders
 It is based on a methanolic mixture of "polychromed" methylene blue and eosin.
 The methanolic stock solution is stable and also serves the purpose of directly fixing
the smear eliminating a prefixing step.
Responsibilities:

 Hematology department personal are required to be knowledgeable of this procedure


 The head of the department must resolve any problem with the process and difficulties
in using this sop
 New employees are trained and assessed for competence before they can handle patient
sample
Specimen requirements:

Freshly collected venous blood in an EDTA container is recommended.

Specimen reception:

Reception of samples should be recorded, and record time of reception. Pay attention to
sample identification and labeling of tubes.
Smears of peripheral blood must be made immediately
Criteria for rejection hematology specimens:
1. When The Identification Is Missing /Inadequate.
2. Insufficient Quantity
3. Inappropriate Container
4. Inappropriate transport/storage
5. Unknown duration of delay
6. Haemolysed and/or clotted sample reject

Equipment & Items required:

 Leshimen's stain
 Microscope.
 Microscopic slides.
 Distilled water.
 Methanol 96%.
 Droppers.
 Immersion oil.

Abbreviations:

B. film : Blood film.


RBC : red blood cell.
WBC : white blood cell.
CBC : complete blood count.
EDTA : ethylene diamine tetra acetic acid
RNA : Ribonucleic acid
Procedures:

First preparation of leishman's stock stain :


1. Dissolve 1.5 g of leishman's powder in 1000 ml of methanol.
2. Mix will until the powder dissolve completely & let it to stand for 24 hr.
3. filter the stock stain with filter paper.
Second blood film preparation :
1. Small drop of blood placed at the end of the slide.
2. Using another slide, the blood can be spread to make the smear.
3. Let’s to dry
4. Add one volume of leishman's stain on blood film smear for 1 min
5. Add two volume of distilled water .
6. Mix then wait for 15 min.
7. Wash with tape water.
8. let's to dry and observe under microscope.

Quality control procedures:

 A well-made slide will have three distinct regions: a head, body, and tail.
 Repeat counts of differential WBC's and RBC's morphology on selected slides on
subsequent days will give an indication of the range in the variation of the results.
(include note on reagent quality).
 When a new batch of stain is prepared, decide the best staining time to use, e.g. stain
films made from the same blood at different times, e.g. 5, 7, 10, 12, 15 minutes.
Limitations/ Interfering substance:

Causes of incorrectly stained blood smear (too blue):


1. Buffered water or stain is two alkali.
2. Excessive thickness of the smear.
3. Alkaline residue on the slide.
4. Insufficient washing.
5. Prolonged staining.
Causes of incorrectly stained blood smear (too red):
1. Buffered water or stain is two acid.
2. Acid residue on slide.
Entire smear has a pale stain:
1. Under staining
2. Weak stain.
3. Excessive washing on allowing buffered distilled water to stand on slide.
4. Using worm or hot buffered distilled water for washing slide.
Variations on staining on different areas of the smear:
1. Buffered water was unevenly applied.
2. Acid or alkali residue on slide.
3. Water wasn't properly drained from slide after washing.
Precipitated stain:
1. Lack of through washing.
2. Stain hasn't properly filtered.
Evaporation of alcoholic stain may be due to excessive staining time, slide titled so that
stain runs to one end or to titling stain of slide before washing.
Expected Result:

 Erythrocytes stain buff pink to pale bluish-gray.


 Leukocytes: neutrophils, monocytes and lymphocytes have a pale blue-
gray cytoplasm and purple nucleus.
 Eosinophil have coarse, pink granules in the cytoplasm.
 Basophils have coarse, deep blue granules in the cytoplasm.
 Platelets have pale blue cytoplasm and diffuse red nuclear material.
 Normal RBC’s shape are normocytic, normochromic, no variation in size, no
shape variation , and no inclusion bodies.
 Normal WBC's morphology and differential.
 Normal platelets size, shape and distribution.

Interpretation of the results:

 Red cell according to hemoglobin content : normochromic, hypochromic,


hyperchromic
 Variation of red cell size (anisocytosis): normocytic, macrocytic, and
microcytic
 Variation of red cell shape (poikilocytosis): normal or abnormal; report the
presence of sickle cells, target cells, spherocytes, etc.
 Red cell inclusion: RNA, polychromasia, Howell-Jolly bodies, etc.
 Nucleated red blood cells and all types of erythroblasts are abnormal.
 Elevated white blood cell count may mean infection.
 Decreases in white blood cell count may occur with disease progression or may
indicate bone marrow suppression.
 Total lymphocyte count: decrease in absolute lymphocyte count may reflect
bone marrow suppression.

71
 An increase in neutrophils may be due to an acute bacterial infection or
hematological malignancies such as myeloid leukemia.
 An increase in eosinophils may be due to a parasitic infection or an allergic
reaction.
 An increase in lymphocytes may be due to viral infections or chronic infections
such as tuberculosis or lymphocytic leukemia.
 An increase in monocytes is found in hematological malignancies such as
chronic myelomonocytic leukemia and certain bacterial and parasitic infections
(e.g., typhoid fever, malaria).
Reporting result:

RBC's Morphology:…………………………………………………………
WBC's differential:
Neutrophils:..............%.
Lymphocytes:...........%.
Eosinophil’s:..............%.
Basophils:.................%.
Blast cells:..................%.
Platelets morphology & distribution:………………………………………
Blood Group \ (Slide Method) BIOTEC LABORAYORIES
SOPs\ HGA \.......H\ Haem \14

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Purpose & Definition:

Determine the ABO and Rh group in human blood. ABO and Rh group are determine by
detection the presence or absence of A & B antigen in ABO system and presence or
absence of D antigen in Rh system. Because of its low sensitivity, slide method used only
for preliminary grouping.
Responsibilities:

 Hematology department personal are required to be knowledgeable of this procedure.


 New employees are trained and assessed for competence before they can handle patient
sample
 The Head of the department must resolve any problem with the process and difficulties
in using this SOP.
Specimen requirements:

Blood collected in an EDTA container is recommended.

Specimen reception and rejection:

Reception of samples should be recorded, and record time of reception. It is essential to


pay strict attention to sample identification and labeling of tubes.
Clotted samples, hemolysed samples and samples collected in other anticoagulants than
EDTA are rejected. Samples should be stored at 2-8 Co if not tested immediately.
Equipment & Items required:

 Glass slides, Wooden applicator.


 Anti-A(blue), Anti-B(yellow) and Anti-D reagents.

72
Abbreviations:

 ABO : blood group system. Rh : rhesus factor.

Procedures:

1. Bring the reagents at room temperature.


2. Mix blood tube gently.
3. Transfer one drop of whole blood for each test to a glass slide.
4. Transfer one drop of each reagent to each blood drop.
5. Mix whole blood & reagent by using wooden applicator.

Quality control procedures:

The reactivity of blood grouping reagents should be confirmed by testing with known
blood group blood samples on each day of use.
Limitations/ Interfering substance:

 Drying up of reaction mixture can cause aggregation of cells, giving false positive
results.
 Abnormal plasma proteins, cold auto agglutinins, positive direct anti globulin test
and in some cases bacteremia may interfere.
Expected result:

Agglutination indicates a positive result to the corresponding antigen, no agglutination


indicates a negative result to corresponding antigen.
Interpretation of the results:

Anti- Anti- Anti- Blood


A B D group
- - +
O+
+ - +
A+
- + +
B+
+ + + AB+
- - -
O-
+ - -
A-
- + -
B-
+ + - AB-
Reporting result:

Blood group: ………… Rh:………………


Erythrocyte Sedimentation Rate (Westergren technique)
SOPs\ HGA \.......H\ Haem \15

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Purpose/Definition:

The erythrocyte sedimentation rate (ESR) is a non-specific test. It is raised in a wide range
of infectious, inflammatory, degenerative, and malignant conditions associated with
changes in plasma proteins, particularly increases in fibrinogen, immunoglobulins, and
CRP.
When well-mixed venous blood is placed in a vertical tube, erythrocytes will tend to fall
toward the bottom. The length of fall of the top of the column of erythrocytes in a given
interval of time.
Responsibilities:

 Hematology department personal are required to be knowledgeable of this procedure.


 New employees are trained and assessed for competence before they can handle patient
sample
 The head of the department must resolve any problem with the process and difficulties
in using this SOP.
Specimen requirements:

Either venous blood collected directly into sodium citrate and tested within 2 hours, or
EDTA anticoagulated blood diluted in sodium citrate can be used.
If EDTA blood is used and kept refrigerated at 4-8°C, citrate dilution of the blood and
testing can be delayed for up to 6 hours.
Specimen reception:

Samples must be transported as soon as possible after collection at 18–22ºC, and the tests
samples should be analyzed within 2 hours after collection.
Criteria for Rejecting Hematology Specimens:
1. When the identification is missing /Inadequate.
2. Insufficient quantity.
3. Inappropriate container.
4. Inappropriate transport /storage.
5. Unknown duration of delay.

Equipment & Items required:

1. Tri-Sodium citrate, 32 g/l (3.2 % w/v) anticoagulant.


2. Glass Westergren pipettes or when available, disposable plastic Westergren pipettes
can be used.
3. Westergren pipettes measures 300mm in length (plastic pipettes are often shorter) and
are graduated from 0-200mm. The diameter should not be less than 2.55mm.
4. Timer.

Abbreviations:

ESR: Erythrocyte sedimentation rate


CRP: C-reactive protein.
EDTA: Ethylene diamine tetra acetic acid

Procedures:

1. Pipette 0.4 ml of sodium citrate anticoagulant into a small container.


2. Add 1.6 ml of venous blood or EDTA anticoagulated blood and mix well.
3. Remove the cap of the container and place the sample in the ESR stand (make a note of
the number in the patient’s notes).
4. Insert a Westergren pipette and ensure it is positioned vertically.
5. Using a safe suction method, draw the blood to the 0 mark of the Westergren pipette,
avoiding air bubbles.
6. Check that the ESR stand is level by making sure that the bubble in the spirit level is
central. If required, adjust the screws on the bottom of the stand. Re-check that the
pipette is vertical.
7. Set the timer for 1 hour. Ensure the ESR stand and pipette will not be exposed to direct
sunlight.
8. At the end of the hour read the height of clear plasma above the upper margin of the
column of sedimenting cells to the nearest millimeter.
Quality control procedures:

The most practical way of controlling ESR tests is to follow the test method exactly.

Limitations/ Interfering substance:

1. Using the wrong volume of blood to anticoagulant.


2. Blood not sufficiently mixed with anticoagulant.
3. Clots in the blood.
4. Air bubbles at the top of the column.
5. Testing blood samples at the hottest time of the day, or leaving tests in direct
sunlight. Temperatures over 25°C increase sedimentation.
6. Using a pipette, which is, not clean or not dry.
7. Pipette not positioned vertically. Even slight variations from the upright increase
sedimentation.
8. Not checking whether the ESR stand is level on the bench.
9. Placing an ESR stand on the same bench as a centrifuge where vibration will
interfere with sedimentation.
10. Measuring the ESR when a patient is dehydrated.

Expected values:

 Men…Up to 15 mm/hour
 Women…Up to 20 mm/hour
 Elderly…Up to 20 mm/hour
 Children…Up to 15 mm/hour

Interpretation of the results:

An elevated ESR may be found in:


 Pregnancy (after the third month).
 Acute and chronic infections.
 Rheumatic fever.
 Rheumatoid arthritis.
 Myocardial infection.
 Nephrosis.
 Acute hepatitis.
 Menstruation.
 Tuberculosis.
 Hypothyroidism.
 Hyperthyroidism.
Adults over 60 years of age frequently have a slightly higher ESR value due primarily to
decreased concentrations of plasma albumin.
A decreased ESR will be present in:
 Polycythemia.
 Congestive heart failure.
 Hypofibrinogenemia.
 The presence of red blood cell abnormalities (poikilocytosis, spherocytes,
and sickle cells).
Reporting result:

ESR: ………mm/1st Hour


Osmotic Fragility Test
SOPs\ HGA \.......H\ Haem \16

Version:1……………………………. Head of department: ………………………


Date effective: ……………………... Quality Officer: ……………………………
Copy number: ……………………… Director of : …………………………..........
Definition &Purpose:

Osmotic fragility test is a test to measure red blood cell (RBC) resistance to hemolysis
when exposed to a series of increasingly dilute saline solutions "hypotonic solutions" .
The test is done to evaluate hemolytic anemia especially hereditary spherocytosis and
hemolytic states.
This procedure is employed to help diagnose different types of anemia's, in which the
physical properties of the red blood cell are altered.
Responsibilities:

 Hematology department personal are required to be knowledgeable of this procedure.


 New employees are trained and assessed for competence before they can handle patient
sample
 The head of the department must resolve any problem with the process and difficulties
in using this SOP.
Specimen requirements:

Heparinized venous blood sample (2-3 ml) . The test should be carried out within 2 hours
of collection or 6 hours if kept at 4oC.
Specimen reception:

Reception of samples should be recorded, and record time of reception. Pay attention to
sample identification and labeling of tubes.
Samples must be transported as soon as possible, and the tests samples should be analyzed
within 2 hours after collection.
Criteria for rejection haematology specimens
1. When the identification is missing /inadequate.
2. Insufficient Quantity
3. Inappropriate Container
4. Inappropriate transport/storage
5. Unknown duration of delay
6. Haemolysed and/or clotted sample reject

Equipment & Items required:

Plastic centrifuge tube, Racks, Stopwatch, Parafilm, Automatic pipette 1ml, Automatic
pipette 10 to 50 µL, Stock solutions consist of buffered NaCl 1%, Distilled water,
Centrifuge, Calculator, Yellow tips, Blue tip, Cuvette، Spectrophotometer, Vortex.
Abbreviations:

EDTA: Ethylene diamine tetra acetic acid.


O,D: Optical Density.
HS: (hereditary spherocytosis).
NaCl: (sodium chloride).
Procedures:

Prepare dilutions of buffered NaCl and place in appropriately labeled test tube

Test 1.0%NaCl D.W. (ml) Final conc.


tube (ml) (%)
1 10.0 0. 1.00
0
2 8.5 1. 0.85
5
3 7.5 2. 0.75
5
4 6.5 3. 0.65
5
5 6.0 4. 0.60
0
6 5.5 4. 0.55
5
7 5.0 5. 0.50
0
8 4.5 5. 0.45
50
9 4.0 6. 0.40
0
10 3.5 6. 0.35
5
11 3.0 7. 0.30
0
12 2.0 8. 0.20
0
13 1.0 9. 0.10
0
14 0.0 10 0.00
.0
1. Mix the preceding dilutions well, using Parafilm to cover each test tube while
mixing.
2. Add 0.050 ml of the patients heparinized blood to each of 14 test tubes.
3. Mix each test tube immediately by gentle inversion in vortex.
4. Allow the test tubes to stand at room temperature for 30 minutes.
5. Remix the test tubes gently and centrifuge at 2500 rpm for 5 minutes.
6. Carefully transfer the supernatants to cuvettes and read on Spectrophotometer at
wavelength of 540 nm. Set the optical density at 0.0% hemolysis using the
supernatant in test tube 1,which represent the blank , or 0.0% hemolysis . Test tube
14 represent 100% hemolysis.
7. Calculate the percent of hemolysis for each supernatant and will be reported.
8. The result of the test may be graphed.

Quality control procedures:

Freshly collected normal heparinized sample obtained in the same way as that the patient.

Limitations/ Interfering substance:

 This test should be performed immediately because shape change and osmotic
conditions change with time.
 Presence of hemolytic organisms in the sample gives invalid results "fragility
increased"
 Severe anemia or other conditions with fewer RBCS available for testing.
 Recent blood transfusion .
 If anticoagulants blood is used for test use only heparin as anticoagulant , in order
to avoid adding more salts to blood such as oxalate , EDTA.
 Arise in the temperature at which the tests are carried out "decreased osmotic
fragility".
 The fragility of the red cells is increased by a fall in pH.
 If hemolysis is present in the 0.85% sodium chloride tube or in the normal control,
the buffered sodium chloride stock and working solutions should be discarded and
re-prepared.
Expected values:

Hemolysis begins 0.45% and complete 0.35%.

81
Interpretation of the results:

Osmotic fragility decreased in:


 Thalassemia.
 Iron deficiency anemia.
 Sickle cell anemia
 After splenectomy, chronic liver disease
 Hyponatremia.
 Polycythemia Vera.
Osmotic fragility increased in :
 Hemolytic anemia
 Hereditary spherocytosis.
 Acquired spherocytosis
 Hypernatremia
 And whenever spherocytes are found.
 The older red cells are also more fragile.

Reporting results:

 Calculate the percent of hemolysis , the percent of hemolysis is calculated as

O.D. of supernatant
Percent of hemolysis = x

O.D. supernatant tube #14

 The results of the test may then be graphed, with the percent hemolysis plotted on
the ordinate (vertical axis) and the sodium chloride concentration on the (horizontal
axis).
Cerebrospinal Fluid Cellular Examination
SOPs\ HGA \.......H\ Haem \17

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............
Purpose & definition :
Cerebrospinal Fluid (CSF) is the product of the secretory activity of the choroid plexus. It
is the third major fluid of the body and supplies nutrients to the nervous tissue, removes
metabolic waste, and protects the brain and spinal cord from trauma. The purpose of a CSF
analysis is to diagnose medical disorders that affect the central nervous system.
Responsibilities:

 Haematology department personal are required to be knowledgeable of this procedure.


 The head of the department must resolve any problem with the process and difficulties
in using this SOP.
 New employees are trained and assessed for competence before they can handle patient
sample.
Specimen requirements

 A physician collects CSF by lumbar puncture and always under aseptic conditions.
 It is a routine practice to collect three (3) sterile tubes of CSF (1-5 ml per tube) for
analysis.
 Tubes should be collected and labeled sequentially by the physician at the time of
collection.
 All tubes must be labeled properly and delivered immediately to the following
sections: Tube #1: To Chemistry for protein and glucose or serology study.
Tube #2: To Microbiology for culture and gram stain.
Tube #3: To Hematology for cell count and differential
If only one tube is collected, perform testing in the following order to
preserve specimen and avoid contamination:
1. Microbiology - culture and gram stain.
2. Hematology - cell count and differential.
3. Chemistry

82
Specimen reception:

Reception of samples should be recorded, and record time of reception. Pay attention to
sample identification and labeling of tubes.
Samples must be transported as soon as possible, and the tests samples should be analyzed
within 1 hours after collection.
Criteria for rejection haematology specimens
1. when the identification is missing /inadequate.
2. insufficient quantity
3. inappropriate container
4. inappropriate transport/storage
5. unknown duration of delay
6. Clotted bloody sample.

Equipment & Items required:

1. Microscope
2. Counting chamber
3. Automatic pipet
4. Yellow tips
5. Coverslips (supplied with the counting chamber)
6. Tubes 2–5ml
7. Turks solution
8. Glass slides
9. Giemsa stain
10. Methanol
11. Normal saline

Abbreviations:

CSF: cerebrospinal fluid


Procedures:

For turbid CSF sample


1. Make a 1: 20 dilution using 0.05 ml of the CSF and 0.95 ml of Turks solution.
2. Pipette into a small tube and mix.
3. put one drop from the diluted sample on the counting chamber.
4. Leave the counting chamber on the bench for 5 minutes to allow the cells to settle.
5. Place the chamber on the microscope stage.
6. Count the cells in the 4 WBC squares' using the 10 x objective.
7. the counted WBC in the 4 squares' multiply with 50.
For clear CSF sample:
1. If undiluted CSF is used, no calculation is necessary; count the 9 WBC squares' and
another random square'.
2. Gives the number per mm3 of CSF.
3. If erythrocytes are present: Make a dilution 1: 2 using 0.1 ml of the CSF and 0.1 ml
of Turks solution.
4. Count the cells in the 10 WBC squares' using the 10 x objective.
5. The counted WBC in the 10 squares' multiply with 2.
Erythrocytes counting:
1. Make a dilution 1: 200 using 0.01 ml of the CSF and 1.99 ml of saline solution.
2. Count the cells in the 5 RBC squares' using the 40 x objective.
3. The counted RBC in the 5 squares' multiply with 10.000.

N.b: to use other dilution the equation can be used:


Number of cells counted x dilution = cells/µl
Number of squares counted x vol. of 1 square (0.1)

For bloody sample use the equation as follow:


WBCs (added) = WBC(blood) x RBC (CSF)
RBC (blood)
True CSF WBCs = WBC CSF – WBC added
Determination of the leukocyte differential leukocyte count:
Clear CSF sample:
1. Centrifuge the CSF at 3000 g for 10 minutes. Pour off the supernatant fluid into another
tube (to be used for other tests).
2. Mix the deposit by tapping the end of the tube, spread on a clean slide and leave to dry.
3. Fix with methanol and stain with a Giemsa stain as described in blood film.
Turbid CSF sample:
1. Pipette one drop of uncentrifuged, mixed CSF on to a slide.
2. Make a thin smear and leave to dry.
3. Fix with methanol and stain with a Giemsa stain as described in blood film section.

Limitations/ Interfering substance

1. Specimens must be well mixed. Failure to mix the specimen can cause invalid
results.
2. Leukocytes may begin to lyse within one (1) hour after collection. Cell counts must
be performed promptly.
Interpretation of the results:

An increased number of leukocytes can be found in:


Bacterial meningitis mostly neutrophils.
Viral meningitis: mostly lymphocytes.
Other microbial infection.
Expected values:

Normal RBC count :( 0 ) cell/μl


Normal WBC count (0-5) cell/μl
Reticulocytes Count \ (Brilliant cresyl blue) RAL. Diagnostics
SOPs\ HGA \.......H\ Haem \18

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............

Purpose & Definition:

Describe the procedure for preparing thin blood smear stained with supra-vital dye in order
to visualize reticulocytes.
Reticulocytes are the final immature cells in the red blood cells maturation process, and the
only immature red cells seen normally in peripheral blood, they are released from the bone
marrow and circulate in the peripheral blood to be fully differentiated within 24 hours.
The reticulocyte count is an important diagnostic tool. It is a reflection of the amount of
effective red blood cell production taking place in the bone marrow.

Responsibilities:

 Haematology department personal are required to be knowledgeable of this procedure.


 The head of the department must resolve any problem with the process and difficulties
in using this SOP.
 New employees are trained and assessed for competence before they can handle patient
sample.

Specimen requirements:

Freshly collected venous blood in an EDTA container is recommended.

Specimen reception & rejection :

Reception of samples should be recorded, and record time of reception. Pay attention to
sample identification and labeling of tubes.
Criteria for rejection haematology specimens
1. When the identification is missing /inadequate.
2. Insufficient Quantity
3. Inappropriate Container
4. Inappropriate transport/storage
5. Unknown duration of delay
6. Haemolysed and/or clotted sample reject.

Equipment & Items required:

 Microscope.
 Microscope slides.
 Plastic tube.
 Commercially prepared liquid Brilliant cresyl blue .
 Immersion oil.
 Micropipette .

Abbreviations:

 Retics count : Reticulocyte count.


 RBCs: Red blood cells.
 µl: Microliter.

Procedures:

1. Deliver about 100 µl of stain into tube, add equal volume of patient blood.
2. Mix well & incubate 15 min at 37ºC.
3. At the end of the time, resuspend the red cells by gentle mixing.
4. Dispense one drop of the mixture on slide.
5. Spread the mixture as the blood film, lets dry.
6. Choose the area of field where cells are undistorted and stain is good, using 100X
immersion lens count retics seen per 1000 red cells.
7. Report the result in percentage in the result book.

Result and calculation:

Reticulocyte appears as cell containing dark blue granules or blue network.


Count 1000 RBCs including reticulocytes, the number of reticulocyte is reported as
percentage of the total RBCs .

Retics count
Retics % =-------------------------100
1000 RBCs
Quality control procedures:

Two slides should be done for each retics count, final calculated retics % from the two
slides should agree with 15%, if this agreement is not reached, prepare a third slide and
count.

Limitations/ Interfering substance:

1. High glucose levels and the use of heparin can cause reticulocytes to stain poorly.
2. Falsely decreased reticulocyte counts can result from under staining the blood with
new methylene blue. Be sure the stain/blood mixture incubates the full 15 minutes.
3. Thick and thin spreading are improper for counting.
4. Filtration of the stain is necessary when precipitated material is present which can
resemble a reticulocyte.
5. A retractile appearance of erythrocytes should not be confused with reticulocytes.
Retractile bodies are due to poor drying owing to moisture in the air.
6. If no reticulocytes are observed after scanning at least two slides, report ―none seen‖.

Expected value:

 Up to 1.5% (15/1000) of total red blood cells in all ages.


 Up to 6% (60/1000) of total red blood cells in neonates but return to adult levels in 1-
2 weeks.

Interpretation of the results:

 The reticulocyte count is elevated:


1. In patients with hemolytic anemia including G6PD deficiency attacks and
hemolytic diseases of newborn.
2. In those with hemorrhage (acute and chronic).
3. Following treatment of iron-deficiency anemia and the megaloblastic
anemia.
4. In patients with uremia.
5. thalassemia, sideroblastic anemia.
6. Pregnancy.
7. Medications such as levodopa, malarial medications, corticotrophin,
and fever-reducing medications.
 The reticulocyte count is decreased in cases of:
1) Aplastic anemia.
2) Aplastic crises of hemolytic anemia .
3) Ineffective erythropoiesis pernicious anemia.
4) Untreated pernicious anemia, megaloblastic anemia and iron deficiency
anemia.
5) Exposure to radiation or radiation therapy.
6) Medications such as azathioprine, chloramphenicol, dactinomycin,
methotrexate, and other chemotherapy medications.
 Reticulocytopenia in the presence of a suggested hemolytic anemia may often make
diagnosis difficult. The diagnosis of a hemolytic anemia can be made because the
combination of both hemolysis and reticulocytopenia results in a rapidly falling
hemoglobin and hematocrit.
Reporting result:

Reticulocytes count......................% .
Blood Film \ (Giemsa Stain) Coral Clinical system
SOPs\ HGA \.......H\ Haem \19

Version:1…………………………….. Head of department:


…………………………
Date effective: 01/02/2015………......
Quality Officer:
Copy number:1………………………
……………………………...
Director of :
………………………….............

Purpose & Definition:

Describe the procedure for preparing thin blood smear and staining with Giemsa stain for
the purpose of diagnosis of different anemia's, leukemia's, viral and parasitic infections, in
addition associated with inclusion bodies, allergic reactions, hemoglobinopathies, and
platelets disorders. Blood film is a visible blood picture that confirms CBC reports and
explains them in order to detect and identify abnormal cells.
Responsibilities:

 Haematology department personal are required to be knowledgeable of this procedure.


 The head of the department must resolve any problem with the process and difficulties
in using this SOP.
 New employees are trained and assessed for competence before they can handle patient
sample.
Specimen requirements:

Freshly collected venous blood in an EDTA container is recommended.

Specimen reception and rejection:

Reception of samples should be recorded, and record time of reception. Pay attention to
sample identification and labeling of tubes.
Smears of peripheral blood must be made immediately
Criteria for rejection haematology specimens
1. When the identification is missing /inadequate.
2. Insufficient Quantity
3. Inappropriate Container

91
4. Inappropriate transport/storage
5. Unknown duration of delay

91
6. Haemolysed and/or clotted sample reject.

Equipment & Items required:

 Microscope.
 Microscopic slides.
 Distilled water.
 Methanol 96%.
 Droppers.
 Immersion oil.
 Giemsa stain.

Abbreviations:

B. film : Blood film.


RBC : Red blood cell.
WBC : White blood cell.
CBC : Complete blood count.
Hb : Hemoglobin.
EDTA : Ethylene diamine tetra acetic acid.

Procedures:

1. Small drop of blood placed at the end of the slide.


2. Using another slide, the blood can be spread to make the smear.
3. Let’s dry.
4. Fix blood film with methanol for 2 to 3 min.
5. Wash the slide with distilled water & let the blood film to dry.
6. Dissolve one volume of commercially prepared Giemsa stain + nine volume of
distilled water.
7. Cover the slide with working stain for 15 min. then wash with water.
8. Observe the normal & abnormal morphology of WBC & RBC & record the
result in the result book.
Quality control procedures:

A well-made slide will have three distinct regions: a head, body, and tail.
Repeat counts of differential WBCs and RBCs morphology on selected slides on
subsequent days will give an indication of the range in the variation of the results. (include
note on reagent quality).
Limitations/ Interfering substance:

 A ragged spreader or one that is not smooth can ruin a slide.


 Make sure the drop of blood is not too big or too small.
 Hold the spreader at the correct angle .
 Do not use a glass slide that is dirty or oily.
 Films made from blood that had been standing for more than 6 hours affects
the quality of staining.
 Unfiltered stain gives bad staining.
 Long fixation time may give a wrong view of aberrant RBC’s.

Expected Result:

 Erythrocytes stain buff pink to pale bluish-gray.


 Leukocytes : neutrophils, monocytes and lymphocytes have a pale blue-
gray cytoplasm and purple nucleus. Eosinophil have coarse, pink granules in
the cytoplasm. Basophils have coarse, deep blue granules in the cytoplasm.
 Platelets have pale blue cytoplasm and diffuse red nuclear material.
 Normal RBC’s shape are normocytic, normochromic, no variation in size, no
shape variation , and no inclusion bodies.
 Normal WBCs morphology.
 Normal WBCs differential:
Neutrophils: 45% - 65%.
Lymphocytes: 35 – 55%.
Eosinophil’s: up to 3%.
Basophils: 0 – 1%.
 Normal platelets size, shape and distribution.

92
Interpretation of the results:

 Hb content (red cells color): normochromic, hypochromic, hyperchromic.


 Red cell size: normocytic, macrocytic, and microcytic.
 Anisocytosis: variation in RBCs size.
 Red cell shape: normal or abnormal; report the presence of sickle cells, target cells,
spherocytes, etc.
 Poikilocytosis: variation in RBCs shape.
 Red cell inclusion: RNA, polychromasia, punctate basophilia, Howell-Jolly bodies,
etc.
 Report the presence of parasites such as sporazoa, nematodes, and trypanosomes as
well as bacteria such as spirochetes. .
 Nucleated red blood cells and all types of erythroblasts are abnormal.
 Elevated white blood cell count may mean infection.
 Decreases in white blood cell count may occur with disease progression or may
indicate bone marrow suppression.
 Total lymphocyte count: a decrease in absolute lymphocyte count may reflect bone
marrow suppression.
 An increase in neutrophils may be due to an acute bacterial infection or hematological
malignancies such as myeloid leukemia.
 An increase in eosinophil’s may be due to a parasitic infection or an allergic reaction.
 An increase in lymphocytes may be due to viral infections or chronic infections such
as tuberculosis or lymphocytic leukemia.
 An increase in monocytes is found in hematological malignancies such as chronic
myelomonocytic leukemia and certain bacterial and parasitic infections (e.g., typhoid
fever, malaria).
Reporting result:

RBCs Morphology:…………………………………………………………
WBCs differential: …………………………………………………………
Neutrophils:...............%.
Lymphocytes:...........%.
Eosinophil’s:..............%.
Basophils:..................%.
Blast cells:.................%.
Platelets morphology & distribution:………………………………………
Sickling test (Na-meta-bisulfate method)
SOPs\ HGA \......H\ Haem \20

Version: 1…………………………… Head of department:


…………………………
Date effective: …………… …...…...
Quality Officer:
Copy number: ……………………
……………………………...
Director of :
………………………….............

Purpose& definition :

The purpose of this test is to detect sickle cell disorder (Anemia or Trait). Sickle cell
anemia is caused by an abnormal form of hemoglobin known as hemoglobin-S, which
tends to precipitate in a way that the red cell takes the sickling shape.
Whole blood is mixed with a sodium metabisulfite reagent on a slide. If the red cells
contain an abnormal hemoglobin, they will become sickle-shaped (or half-moon shape).
The reagent sodium metabisulfite removes oxygen from the cells, allowing sickling to take
place.
Responsibilities:

 Hematology department personal are required to be knowledgeable of this procedure


 The head of the department must resolve any problem with the process and
difficulties in using this SOP
 New employees are trained and assessed for competence before they can handle
patient sample
Specimen requirements:

Whole blood using EDTA or heparin as anticoagulant sample.

Specimen reception

Reception of samples should be recorded, and record time of reception. Pay attention to
sample identification and labeling of tubes.
Samples must be transported as soon as possible, and the tests samples should be analyzed
within 2 hours after collection.
Criteria for rejection hematology specimens
1. When The Identification Is Missing /Inadequate.
2. Insufficient Quantity
3. Inappropriate Container
4. Inappropriate transport/storage
5. Unknown duration of delay
6. Haemolysed and/or clotted sample reject

Equipment & Items required:

 Distilled water & plastic tubes


 Na-meta-bisulfate powder
 slide and cover slide
 petroleum jelly
 microscope

Abbreviations:

Na: sodium
D.W: distill water

Procedures:

First preparation of Na-meta-bisulfate


1. prepare 2% of Na-meta-bisulfate solution
2. dissolve 2 gram of Na-meta-bisulfate in 100ml D.W
3. This solution, if stored at 3º or 4ºC remains effective for about one week.
4. Stable for 8 hr at room
temperature Second sickle cell test
1. Place a small drop of capillary blood in the centre of a slide.
2. Add an equal drop of the fresh sodium metabisulfite solution.
3. Mix carefully and cover with a cover slide with petroleum jelly , make sure that no
air bubbles form.
4. Place the slide in a wet chamber.
5. Examine under the microscope after 15 minutes using the 40 x objective.
6. The test is negative if the red cells remain round.
7. The test is positive if the cells become sickle-shaped, or banana-shaped.
Quality control procedures:
Its advantage to run positive control each time this test is performed
Limitations/ Interfering substance

1. This test should not performed on infant less than six months old
2. Sickling of red blood cells is maximum at 37°c and decrease as the temperature
lowers
3. Sever anemia will cause false negative result
4. Low Hb cause false negative : the use of packed cells will overcome this problem
5. Specimen with 25% Hb F present may cause false negative.
6. Blood from patients with polycythemia ,multiple myleoma ,dysglobulinemia
may cause false positives
7. False-negative results may occur if:
— outdated reagents are used.
— concentrations of hemoglobin S are low.
— patients have moderate or severe anemia.

8. In all cases where abnormalities are indicated or suspected electrophoretic


confirmation is recommended
Interpretation of the results:

Negative result
 The erythrocytes remain round
If the test is negative, re-examine the slide after a further 30
minutes, then after 2 hours and after 24 hours.
Positive result
 The erythrocytes become sickle-shaped or banana-shaped
often with spikes
It is important to examine several parts of the preparation,
as sickling can occur more quickly in one part than in another.
 Do not mistake normal erythrocytes lying on their side
created cells for sickle cells.
Reporting result:
Positive for sickle cell test.
Negative for sickle cell test.
References:

1. Alvin H. Schmaier, Hillard M. Lazarus. Concise Guide to Hematology. John Wiley


& Sons. 2011
2. Bernadette F. Rodak, George A. Fritsma, Kathryn Doig. Hematology:
Clinical Principles and Applications. Elsevier Health Sciences. 2015
3. Elaine Keohane, Larry Smith, Jeanine Walenga. Rodak's. Hematology: Clinical
Principles and Applications 5th ed Elsevier Health Sciences. 2016
4. Martha Roper. Blood Cells and Hematology . Hayle Medical. 2015
5. Shirlyn B. McKenzie, J. Lynne Williams, Ph.D., Kristin Landis-Piwowar. Clinical
Laboratory Hematology. Pearson. 2014
6. World Health Organisation. Guidelines on Standard Operating Procedures for
Haematology. WHO 2000
7. World Health Organisation. Manual of Basic Techniques for a Health Laboratory
2nd ed. WHO 2003

You might also like