Huq CT Simulator
Huq CT Simulator
• What is a CT simulator?
• CT simulation process
• Functional performance characteristics of CT
equipment
• QA of CT simulators
• QA for CT simulation software
• Evaluation of CT simulation process
• Conclusion
CT simulator
• CT scanner with
flat table top
• Laser positioning
and marking
system
(preferable with
external lasers)
• Virtual simulation
software
• Various hardcopy
output devices
CT Simulation Process
3. Treatment setup
2
1 CT Simulator
Virtual Dose
CT Scanner
Simulation Calculation
E. CT performance E. CT performance
parameters parameters
– X-ray system calibration – Image uniformity and noise
– Collimator assessment – Spatial resolution
– Localization laser alignment – Contrast resolution
– Slice width and sensitivity – CT number calibration and
profile linearity
– Radiation exposure and dose – Artifact evaluation
Quality assurance of CT
simulators
Commissioning
– Radiation safety
– CT Dosimetry
– Electromechanical tests
– X-ray Generator
– Image performance tests
QA program
• Purpose
– To assure proper equipment and program operation
– To assure safety of patient, staff, and public
• Should include
– specification of tests
– test frequency
– tolerance limits
– corrective actions
– QA assignments
QA components
• Exposure Measurements
– Two shielding design methods commonly employed
» direct use of isoexposure contours
» NCRP Report No. 49 & 147
• Interlocks (x-ray interruption, dead man)
• Patient Dose (CTDI, MSAD)
Survey Sites
G F
L E K
B
A
C J
I
D
H
Staff and Public NCRP 49:
Brilliance AcQSim
Weekly Weekly
Measured
Exposure Exposure
Instantaneous Occupancy
Location Based on Based on
Exposure Rate Factors
Workload Workload
(mR/hr)
(mR/week) (mR/week)
A. Room 450 Door 0.18 1.00 0.026 0.645
B. Room 450 Window 0.30 1.00 0.043 1.075
C. Room 450 Countertop 0.33 1.00 0.047 1.183
D. Room 450 Chair 0.22 1.00 0.031 0.788
E. Room 451 0.23 0.25 0.008 0.206
F. Corridor 474 Door 0.10 0.25 0.004 0.090
G. Corridor 474 0.17 0.25 0.006 0.152
H. Corridor 362 0.15 0.25 0.005 0.134
I. Room 454 0.10 0.25 0.004 0.090
J. Room 448 0.10 1.00 0.014 0.358
K. Room 447 0.10 0.25 0.004 0.090
L. First floor lobby background
Patient Dose
• CTDI - Computed
Tomography Dose
Index represents
average dose per slice
• MSAD - Multiple Scan
Average Dose
represents average
dose per scan
• Body Dose from CT
Scans ~ 1 cGy
• Need a non-invasive
meter
– kV accuracy
– Timer accuracy
– mA linearity
– HVL
measurements
X-ray Generation
• Quantitative • Qualitative
– Phantom Measurements – Physician Preferences
» High Contrast » Tumor
» Low Contrast » Normal Structures
» Uniformity » DRR/DCR Objects
» Spatial Integrity » Workflow
» Artifacts » Customized protocols
» Slice thickness
» CT # accuracy
Image performance
Catphan 500
CTP401 CTP515
• Ability of the
system to record
separate images
of small objects
that are placed
very close
together
Subject contrast (Low Contrast)
• Ability of a
system to resolve
adjacent objects
with small density
differences
• Noise limited
Field uniformity
• Beam geometry
• Target localization
• Target contour verification (using BEV)
• Isocenter calculation
• Isocenter movement
• Virtual field size
• Virtual gantry, collimator, and table rotation
• Evaluation of digitally reconstructed radiographs (DRRs)
Virtual simulation software QA
• CT scan
• transfer of images
• isocenter localization
• treatment plan design
• multimodality image fusion
• transfer of virtual simulation data for calculation of dose
distributions
• DRR generation
• transfer of the plan to treatment machine
• treatment of the patient
• treatment verification
Quality assurance frequency
• Daily (Therapists)
• Weekly (Therapists)
• Monthly (Clinical Engineering)
• Annual (Medical Physics)
Daily QA
(Sim Therapists)
• Radiographic Uniformity
– Most important - look for noise and
distortions in CT image
• Laser Alignment and Tracking
– Critical for patient setup
– Most critical - Couch in/out 500 mm
tracking
• Audio/Visual System
Weekly Inspection (ok, not ok)
(Sim Therapists)