TECHNICAL DATA
TECHNICAL DATA – DIGITAL FAST BEAM DENSITOMETER
GENERATOR
X-RAY TUBE
SOURCE COLLIMATOR
Material Lead
Height 5mm
Diameter 1mm & 2 mm
Collimator-patient distance 50 mm
Tube-Patient distance 160 mm
Shutter 5 mm Lead
DETECTOR COLLIMATOR
Material Brass
Height 38 mm
Diameter 30 mm
DETECTOR
MEDILINK PAGE 2 OF 9
TDA_MEDIX90_01
TECHNICAL DATA – DIGITAL FAST BEAM DENSITOMETER
CANNER
SCANNER
ACQUISITION METHOD
ACQUISITION WINDOWS
Site Hip
3D Modeling Cortical Thickness, Volumetric BMD
MEDILINK PAGE 3 OF 9
TDA_MEDIX90_01
TECHNICAL DATA – DIGITAL FAST BEAM DENSITOMETER
SOFTWARE - DIAGNOSTIC
EXAMINATION SITES
In-Row Scan Combined examination of the lumbar spine & left/right Femur
Provides a low dose, lateral image of the spine (to view all the vertebrae of the
Digital Vertebral spine). The deformation or compression is precisely diagnosed, measured and
Assessment (DVA) classified. This analysis can be either automatic using the Genant’s semi- (DVA)
quantitative classification, or manual using the Genant’s visual classification. Single Energy Dual Energy
Operator can use dual (bone image) or single (attenuation) energy images.
Provides a low dose, AP image of the spine (to view all the vertebrae of the spine).
The Cobb Angle invented by Dr. John Cobb (1948) is used as a standard
measurement to determine and track the progression of scoliosis. The angle of
Cobb Angle
curvature can be measured and reported by drawing lines parallel to the upper
Measurement with AP-
border of the upper vertebral body and the lower border of the lowest vertebra of
DVA
the structural curve, then erecting perpendiculars from these lines to cross each
other, the angle between these perpendiculars is ‘angle of curvature’. Operator
can use dual (bone image) or single (attenuation) energy images
Whole Body Fat & Lean Tissue Analysis (total with or without head, android, gynoid, left arm,
Composition * right arm, left ribs, right ribs, t spine, l spine, pelvis, left leg, right leg, head)
DIAGNOSTIC TOOLS
Calculation of Bone Mineral Density (BMD), Bone Mineral Content (BMC), Surface (area), T-score & Z-score,
Advanced Morphometric tools (surface, distance & angle): Measurements can be done on every type of examination (spine, femur, forearm …etc)
Automated Hip Structural Analysis (HAS): Hip Axis Length (HAL), Femoral Neck Axis Length (FNAL), Intertrochanter to Femoral Head Center
Distance (IH) and Femoral axis versus Neck axis Angle (FNA)
Fracture risk information using Hip Structural Analysis (HSA)
FRAX tool from WHO to evaluate the 10 years probability of osteoporotic fractures
Calculation of Body Mass Index (BMI), Fat Mass Index (FMI), Android/Gynoid Ratio, Lean Mass Balance, Appendicular Skeletal Muscle Index
(Sarcopenia), Resting Metabolic Rate (RMR), Basal Metabolic Rate (BMR)
SOFTWARE – TOOLS
SOFTWARE TOOLS
Multi-language software: Chinese simplified, Czech, German, English, Spanish, French, Italian, Polish and Turkish (other languages can be
translated)
Help menu
DVO recommendations (Certified by DachVerband Osteologie)
Telemaintenance software: Maintenance and fault detection software (network connection required)
STORAGE
CONNECTIVITY
MEDILINK PAGE 5 OF 9
TDA_MEDIX90_01
TECHNICAL DATA – DIGITAL FAST BEAM DENSITOMETER
SOFTWARE – ANALYSIS
PARAMETERS AVAILABLE ON FINAL MEDICAL REPORT
Bone Mineral Density (BMD) expressed in g/cm2, stands for the mineral density of the bones
Area expressed in cm2, 2D projection of the bone
T-Score: Difference between the patient’s BMD value and the mean BMD value of a young adult
Multisite population of healthy subjects of the same gender and from the same ethnic background as the patient,
divided by the standard deviation of the young population of healthy subjects.
Z-score: Difference between the patient’s BMD value and the mean BMD value of a same age population
of healthy subjects of the same gender and from the same ethnic background as the patient, divided by
the standard deviation of the same age healthy population.
Total Bone Mineral Density (BMDt), Local Bone Mineral Density (BMDl), Area
Whole Body Fat Mass, Lean Mass, Bone Mineral Mass, Segmentation
(Not available on Compact Version) Body Mass Index (BMI), Fat Mass Index (FMI), Android/Gynoid Ratio, Lean Mass Balance, Appendicular
Skeletal Muscle Index (Sarcopenia), Resting Metabolic Rate (RMR), Basal Metabolic Rate (BMR)
Bone Mineral Density (BMD), Bone Mineral Content (BMC), Area,
Orthopaedic
Automatic ROI selection (ex: Green zone): femur, knee and lateral knee
Paediatric Bone Mineral Density (BMD), Bone Mineral Content (BMC), Area, Skeletal age comparison
Displays BMD according to the age for the examined regions and supplies T-score and Z-score values as
Reference curve diagnosis values.
Body Composition according to NHANES (2009)
Reference Equation Revised Harris-Benedict, Mifflin St Jeor, Katch-McArdle,
BMD Measurement in any site
Manual Analysis Bone Mineral Content (BMC)
Area
Morphometry Quantitative morphometry in any site (areas, lengths, angles). Ex: Automatic Hip Structural Analysis (HSA)
Trending Graphs, Patient Pictures & Parameters Evolution
MEDILINK PAGE 6 OF 9
TDA_MEDIX90_01
TECHNICAL DATA – DIGITAL FAST BEAM DENSITOMETER
EXAMINATION PARAMETERS
Automatic & Manual selection of the Region Of Interest (ROI)
Low Dose adapted to patient Morphology
RADIATION DOSE
Staff Dose at 1m
(Annual Staff Dose 0.678 mSv*) Spine (90s): 0.0264 µSv Proximal Femur(90s): 0.0301µSv Whole Body(300s): 0.0050µSv
*6 patients/h, 1 spine & 1 femur, 2000h/year
PATIENT POSITIONING
Laser Light
PC controlled easy scan repositioning
METHOD OF EXAMINATION
Pre-regulated exam modes: Exam parameters adjusted automatically based on patient’s morphology (Thin, Healthy and Overweight)
Personalized options: Scan speed, scan area and resolution
Automatic and manual selection of the Region Of Interest (ROI) = Image acquisition with Intelligent Scan
IntelliScan: Smart reduction of the scan window and of the examination time
MEDILINK PAGE 7 OF 9
TDA_MEDIX90_01
TECHNICAL DATA – DIGITAL FAST BEAM DENSITOMETER
ACCESSORIES
SUPPLY ACCESSORIES
ENVIRONMENTAL DATA
& ELECTRICAL SPECIFICATIONS
ENVIRONMENTAL DATA
ELECTRICAL SPECIFICATIONS
110 VAC (±10%) 10A
Voltage / Current
220 - 240 VAC (±10%) 5A
Frequency 50 - 60 Hz
Power Consumption 560W
Required UPS
800W/1000VA
(Stratos without PC & Printer)
MEDILINK PAGE 8 OF 9
TDA_MEDIX90_01
TECHNICAL DATA – DIGITAL FAST BEAM DENSITOMETER
PHYSICAL DESCRIPTION
DIMENSION & WEIGHT (L:LENGTH, W:WIDTH, H:HEIGHT)
MEDILINK PAGE 9 OF 9
TDA_MEDIX90_01