Technical Document Medonic M-Series M32 Hematology Analyzer: Product Name Manufacturer
Technical Document Medonic M-Series M32 Hematology Analyzer: Product Name Manufacturer
Technical Document Medonic M-Series M32 Hematology Analyzer: Product Name Manufacturer
Product Description The Medonic M-series M32 is a fully automatic, 3-part differential, hematology analyzer intended
for in vitro diagnostic testing of blood specimens under laboratory conditions.
Intended Use
Model configuration
Figure 1 Figure 2
Medonic M-series M32B Medonic M-series M32M
Figure 3 Figure 4
Medonic M-series M32C Medonic M-series M32S
Instructions for use Use in accordance with manufacturer’s instructions for use in Medonic M-series M32 User´s
Manual.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
21869-1 2015-03-24 Page 1 of 5
Technical Document
Medonic M-series M32 Hematology Analyzer
Instrument Specifications
Physical
Size (Instrument versions with sampler) HWD ≤ 395 x 340 x 475
Size (Instrument version without sampler) HWD ≤ 395 x 295 x 475
Weight (Instrument) ≤ 18 kg
Weight (Autoloader additional weight) The additional weight of the Autoloader including two sample
wheels is less than 6 kg
Display Depth: True color (24-bit);
Resolution: 800 x 480 pixels
Keyboard Virtual incorporated keyboard
Communication interface ports 1 USB device/4 USB host/1 LAN port
Barcode reader input Yes (via USB)
Operating Environment
Temperature 18 - 32 °C
Humidity 10% - 90%
Electrical
Main Voltage 100 - 240 V
Frequency 50 - 60 Hz
Maximum power consumptions 100 VA (operating); 50 VA (standby)
Measuring principles
MCV, MPV, RBC, WBC, and PLT Impedance
HGB Photometric
Sampling system Closed shear valve
Floating RBC/PLT discriminator Yes (position printed)
Programmable WBC Discriminator Yes
Mathematical 3-part diff. WBC calculation Yes
Parameters Reported 20-parameters:
RBC, MCV, HCT, PLT, MPV, HGB, MCH, MCHC, WBC,
RDW%, RDW abs, PCT, PDW, LPCR, LYM abs, MID abs,
GRAN abs, LYM%, MID%, GRAN%
Performance
Sample volume (Open Tube) ≤ 110 µL
Sample volume (Sampling Device) ≤ 300 µL
Sample volume (Cap Piercer) ≤ 250 µ L
Sample volume (Micro Pipette Adapter) 20 µL
Pre-diluted mode 1:200 to 1:300 using min. 20 µl
e.g. 20 µl sample to 4.5 mL diluent (1:225)
Dispenser precision (CV) ≤ 0.9%
Number of Samples per hour (Open Tube) ≥ 60 samples
Number of Samples per hour (Cap Piercer) ≥ 45 samples
Number of Samples per hour (Sampling Device) ≥ 43 samples
Built-in test / adjustment programs Yes
QC capabilities Mean, SD, CV, Levey-Jennings and Xb
System Information Indicators on parameter Yes
abnormalities
Memory capacity ≥ 50,000 samples
Reagent Stability 36 months
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
21869-1 2015-03-24 Page 2 of 5
Technical Document
Medonic M-series M32 Hematology Analyzer
98/79/EEC In Vitro Diagnostic Using Annex III as the conformity assessment procedure
Medical Device Directive (IVD)
2004/108/EEC Electro Magnetic Emission EN 61326-1:2006,
Compatibility Directive (EMC) SS-EN55011:2009, +A1:2010, Class B,
with amendment 92/31/EEC EN 61000-3-2:2006
Immunity: EN 61326-1:2006,
EN 61000-4-2, -3, -4, -5, -6, -11
2006/95/EEC Low Voltage IEC 61010-1: 2001
Directive (LVD) IEC 61010-2-081:2001 + A1
IEC 61010-2-101:2002
CAN/CSA-C22.2 No 61010-1: second edition
UL 61010-1: second edition
2012/19/EU Waste of electrical --
and electronic equipment (WEEE)
Quality The instrument has been factory calibrated by Boule prior to shipment. Customers are advised to
Control/Calibration check the daily performance of the Medonic M-series M32 system with a certified blood control
authorized by Boule.
Packaging and Weight ≤ 24 kg (standard version); Double corrugated cardboard outer packaging with shock absorbent
inner packaging.
Warning and As there are no assurances of the absence of HIV, Hepatitis B or C viruses or other infectious
Precautions agents in blood samples, blood controls, calibrators and waste, these products should be handled as
potentially biohazardous. See User manual for biohazard symbol placement and further
instructions.
Customers are advised to be knowledgeable of applicable local, state and federal requirements, and
the content of effluent streams, before disposing of waste in public sewer systems or recycling
decontaminated equipment. The instructions for decontamination can be found on the Medonic
home page, www.medonic.se, under Support.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
21869-1 2015-03-24 Page 3 of 5
Technical Document
Medonic M-series M32 Hematology Analyzer
Histograms The measuring principles of the Medonic M-series M32 analyzers are based on impedance and
spectrophotometry principles. The number of cells for determining RBC and WBC values are counted
from a suspension of 1:40,000 for the RBC and 1:400 for the WBC dilution ratio of whole blood.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
21869-1 2015-03-24 Page 4 of 5
Technical Document
Medonic M-series M32 Hematology Analyzer
For the manufacturing process and the specific process control points for this analyzer see the
Manufacturing Process
flowchart below. The process starts with components for the manufacturing of the analyzer being
checked in by the content of specification (e.g. drawings, specifications for incoming inspection)
against acceptance limits. The process is completed with final QC testing, the release by QA, and the
approved analyzer moved to stock.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
21869-1 2015-03-24 Page 5 of 5
Technical Document
Quintus Hematology Analyzer
Model configuration
Packaging and Weight Instrument 35 kg + autoloader 12 kg. Double corrugated cardboard outer packaging with shock
absorbent inner packaging
Product Conformity Quintus analyzer is in conformance with the 98/79/EC In Vitro Diagnostic Medical Device Directive
(IVD) using Annex III as the conformity assessment procedure.
SS-EN 61010-2-101 (Low Voltage 2014/35/EU)
EN 61326 (2006) (EMC 2014/30/EU)
2012/19/EU WEEE
2011/65/EU RoHS
Quality The instrument has been factory calibrated by Boule prior to shipment. Customers are advised to check
Control/Calibration the daily performance of the Quintus system with a certified blood control authorized by Boule.
Warning and As there are no assurances of the absence of HIV, Hepatitis B or C viruses or other infectious agents in
Precautions blood samples, blood controls, calibrators and waste, these products should be handled as potentially
biohazardous. See User manual for biohazard symbol placement and further instructions.
Customers are advised to be knowledgeable of applicable local, state and federal requirements, and the
content of effluent streams, before disposing of waste in public sewer systems or recycling
decontaminated equipment. The instructions for decontamination can be found on the Boule home
page, www.quintus5part.com, under User Support
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13955-5 2017-01-24 Page 1 of 5
Technical Document
Quintus Hematology Analyzer
Instrument Specifications
Measuring principle, RBC, WBC, PLT Impedance
Measuring principle HGB Light source: green LED with 540 nm wavelength
Detector: light to frequency converter
Measuring principle 5-part Differential Light scattering measurement:
Optical measurement Light source: semiconductor laser diode with 650 nm wavelength and
7mW (Class IIIB laser module if the protective housing is closed)
Quartz flow cell with hydro-dynamic focusing
Detector: fiber optic coupled PIN Si photodiodes
Internal safety interlock
Aperture length and diameter WBC: 80 µm, RBC/PLT: 70 µm
Sampling system Ceramic shear valve with 3 separated primary loops
Aspirated blood volume < 110 µl
Number of Samples per hour > 60 samples
Sample type Human whole blood (K3-EDTA anticoagulant)
Parameters reported WBC, LYM, MONO, NEU, EOS, BASO, LYM%, MONO%, NEU%,
EOS%, BASO%, RBC, HCT, MCV, RDWsd, RDW cv, HGB, MCH,
MCHC, PLT, PCT, MPV, PDWsd, PDWcv, LPCR%, LPCRa
Size distributions printed for RBC, PLT and WBC differential
QC capabilities Mean, SD, CV, Levey-Jennings plots and Xb plots
Carry-over < 0.5%
Warning flags on parameter abnormalities Yes
Setable normal ranges (profile limits) Yes
Reagent and Instrument alerts Yes
Built-in test / adjustment programs Yes
Calibration Manual and SW supported automatic mode
Language available English menu and support for other languages
Software upgrade Via USB
Memory capacity 100,000 records including flags, scatter- and histograms
Data processing Intel Atom 1.6 GHz
Data store Windows® 8.1 Embedded
Display 800 x 600 color graphic LCD, portrait layout
External printing Via USB port, any Windows® 8.1 compatible printer
Virtual incorporated keyboard (External keyboard option via PS/2 or
Keyboard
USB)
External via USB port
Barcode reader
Built-in barcode reader in Autoloader unit
Peripheral ports 4 USB (2.0), Ethernet, PS/2
Barcode reader input Yes
Serial output Yes (Conformed to standard EN 60950)
Main Voltage 100 – 240 V AC
Power consumption (maximum) Max 400VA
Frequency 50 / 60 HZ
Main fuse F 10A H 250V
Operating Temperature 15-30° C (59-86 °F)
Humidity (non-condensing) Up to 80%
Instrument Dimensions HxWxD = 515 x 410 x 465 mm
Autoloader Dimensions HxWxD = 190 x 380 x 322 mm
Instrument weight ≤ 35 kg
Autoloader weight ≤ 12 kg
Diluent Consumption Approximately 51 ml per analysis cycle.
Lyse Consumption Approximately 7 ml per analysis cycle.
Stopper Consumption Approximately 1 ml per analysis cycle.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13955-5 2017-01-24 Page 2 of 5
Technical Document
Quintus Hematology Analyzer
Population Color on
(4-diff diagram) scatter
diagram
Artifact BLACK
Lymphocytes BLUE
Monocytes GREEN
Neutrophilic MAGENTA
granulocytes
Eosinophilic ORANGE
granulocytes Figure 2 – Scatter 4-part diff
Population Color on
(Basophil scatter
diagram) diagram
Basophilic MAGENTA
granulocytes
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13955-5 2017-01-24 Page 3 of 5
Technical Document
Quintus Hematology Analyzer
Histograms Impedance based measurements are represented on histograms. These diagrams show the number of
cells against cell size.
• There are two histograms: RBC and PLT.
• Discriminators (thresholds) are displayed with red color.
• The PLT histogram is the magnified section of the left-hand side of the RBC curve.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13955-5 2017-01-24 Page 4 of 5
Technical Document
Quintus Hematology Analyzer
Manufacturing For the manufacturing process and the specific process control points for this analyzer see the flowchart
Process below. The process starts with components for the manufacturing of the analyzer being checked in by
the content of specification (e.g. drawings, specifications for incoming inspection) against acceptance
limits. The process is completed with final QC testing, the release by QA, and the approved analyzer
moved to stock.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13955-5 2017-01-24 Page 5 of 5
Technical Document
Quintus Autoloader
Product Description The Quintus Autoloader is a compact, easy-to-use option for users who value letting their
instrument to get on with the analyses. It makes Quintus the perfect walk-away system for
laboratories with high workload. Just pre-load up to 10 x 10 samples and let the Autoloader
do the work. It tilts and rotates every tube, to get a proper mixing, and reads its barcode. Most
standard closed EDTA tubes fit and the cap will be detected by a cap recognition sensor. Note
that emergency samples can easily be analyzed in the middle of an automated run.
Model configuration
Packaging and Weight Autoloader 12 kg, Double corrugated cardboard outer packaging with shock absorbent inner
packaging
Product Conformity Quintus Autoloader is in conformance with the 98/79/EC In Vitro Diagnostic Medical Device
Directive (IVD) using Annex III as the conformity assessment procedure.
SS-EN 61010-2-101 (Low Voltage 2014/35/EU)
EN 61326 (2006) (EMC 2014/30/EU)
2012/19/EU WEEE
2011/65/EU RoHS
Quality The autoloader has been factory calibrated by Boule prior to shipment. Customers are advised to check
Control/Calibration the daily performance of the Quintus system with a certified blood control authorized by Boule.
Warning and As there are no assurances of the absence of HIV, Hepatitis B or C viruses or other infectious agents in
Precautions blood samples, blood controls, calibrators and waste, these products should be handled as potentially
biohazardous. See User manual for biohazard symbol placement and further instructions.
Customers are advised to be knowledgeable of applicable local, state and federal requirements, and the
content of effluent streams, before disposing of waste in public sewer systems or recycling
decontaminated equipment. The instructions for decontamination can be found on the Boule home
page, www.quintus5part.com, under User Support
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13955-5 2017-01-24 Page 1 of 5
Technical Document
Quintus Hematology Analyzer
For the manufacturing process and the specific process control points for this analyzer see the flowchart
Manufacturing below. The process starts with components for the manufacturing of the analyzer being checked in by
Process the content of specification (e.g. drawings, specifications for incoming inspection) against acceptance
limits. The process is completed with final QC testing, the release by QA, and the approved analyzer
moved to stock.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13955-5 2017-01-24 Page 2 of 5
Technical Document
Quintus 5-part Diluent
Product Description: Quintus Diluent is a hematology diluent used for cell counting and sizing in the Quintus automated,
hematology analyzers.
Preservatives: The stabilizing solution contains buffer salts with anti-microbial substances to prevent contamination
of bacteria or molds, these substances are biodegradable.
Shelf life: 24 months from day of manufacture. The shelf life of this product is based on accelerated and real-
time stability studies.
Packaging and Weight: 20 liter (21.2 kg) LDPE cubitainer packed in a cardboard box.
Warning and Contains no ingredients at hazardous concentrations. Use good laboratory practice.
Precautions: Avoid contact with skin, eyes, and clothing. Harmful if swallowed.
Instructions for use: Use in accordance with manufacturer’s instructions for use and follow Operator manual for connection
and fill instruction.
Population Color on
(4-diff diagram) scatter diagram
Artifact BLACK
Lymphocytes BLUE
Monocytes GREEN
Neutrophil MAGENTA
granulocytes
Eosinophil ORANGE
granulocytes
Figure 4 – 4-diff diagram
Population Color on
(Basophil diagram) scatter diagram
WBC without BLUE
Basophil
Basophil MAGENTA
granulocytes
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13892-7 2016-06-15 Page 2 of 3
Technical Document
Quintus 5-part Diluent
Histograms Impedance based measurements are represented on histograms. These diagrams show the
number of cells against cell size.
• There are two histograms: RBC and PLT.
• Discriminators (thresholds) are displayed with red color.
• The PLT histogram is the magnified section of the left-hand side of the RBC curve.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13892-7 2016-06-15 Page 3 of 3
Technical Document
Quintus 5-part Lyse
Product Description Quintus 5-Part Lyse is a hematology cyanide free lytic reagent used for cell counting and sizing in the
Quintus Series, automated, hematology analyzers.
Shelf life 18 months from day of manufacture. The shelf life of this product is based on accelerated and real-
time stability studies.
Packaging and Weight 5 liter (5.4 kg) LDPE cubitainer packed in a cardboard box.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13897-8 2016-06-02 Page 1 of 3
Technical Document
Quintus 5-part Lyse
Instructions for use Use in accordance with manufacturer’s instructions for use and follow Operator manual for connection
and fill instruction.
Quintus displays the results of the optical measurements in a scatter diagram representation.
Scatter diagrams
• Scatter diagrams represent data in a two-dimensional plane.
• There are two scatter diagrams in the patient report: the 4-DIFF and the BASO
Population Color on
(4-diff diagram) scatter diagram
Artifact BLACK
Lymphocytes BLUE
Monocytes GREEN
Neutrophil MAGENTA
granulocytes
Eosinophil ORANGE
granulocytes
Population Color on
(Basophil diagram) scatter diagram
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13897-8 2016-06-02 Page 2 of 3
Technical Document
Quintus 5-part Lyse
Histograms Impedance based measurements are represented on histograms. These diagrams show the number of
cells against cell size.
• There are two histograms: RBC and PLT.
• Discriminators (thresholds) are displayed with red color.
• The PLT histogram is the magnified section of the left-hand side of the RBC curve.
For the manufacturing process and the specific process control points for this reagent see the flowchart
Manufacturing Process
below. The process starts with raw materials for the manufacturing of the reagent being checked in by
the content of certificates (e.g. Batch Specific Certificate of Analysis) against acceptance limits and by
follow-up analysis of at least one specific characteristic chemical property (e.g. pH in solution). The
process is completed with final QC parameter, functional, and bacterial testing, the release by QA
accompanied by the Certificate of Analysis, and the approved reagent moved to stock.
QC of parameter testing
Parameter and QC of functional testing
functional testing QA approval for filling
Final parameter,
QC of bacterial, parameter
functional, and and functional testing
bacterial testing
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: [email protected]
www.boule.se
13897-8 2016-06-02 Page 3 of 3