PCS-9830A - X - Instruction Manual - EN - Domestic General - X - R1.00

Download as pdf or txt
Download as pdf or txt
You are on page 1of 87

PCS-9830A

Phase Controlled Switching Device


Instruction Manual

NR Electric Co., Ltd.


Preface

Preface

Introduction
This guide and the relevant operating or service manual documentation for the equipment provide
full information on safe handling, commissioning and testing of this equipment.

Documentation for equipment ordered from NR is dispatched separately from manufactured goods
and may not be received at the same time. Therefore, this guide is provided to ensure that printed
information normally present on equipment is fully understood by the recipient.

Before carrying out any work on the equipment, the user should be familiar with the contents of
this manual, and read relevant chapter carefully.

This chapter describes the safety precautions recommended when using the equipment. Before
installing and using the equipment, this chapter must be thoroughly read and understood.

Health and Safety


The information in this chapter of the equipment documentation is intended to ensure that
equipment is properly installed and handled in order to maintain it in a safe condition.

When electrical equipment is in operation, dangerous voltages will be present in certain parts of
the equipment. Failure to observe warning notices, incorrect use, or improper use may endanger
personnel and equipment and cause personal injury or physical damage.

Before working in the terminal strip area, the equipment must be isolated.

Proper and safe operation of the equipment depends on appropriate shipping and handling,
proper storage, installation and commissioning, and on careful operation, maintenance and
servicing. For this reason, only qualified personnel may work on or operate the equipment.

Qualified personnel are individuals who:

 Are familiar with the installation, commissioning, and operation of the equipment and of the
system to which it is being connected;

 Are able to safely perform switching operations in accordance with accepted safety
engineering practices and are authorized to energize and de-energize equipment and to
isolate, ground, and label it;

 Are trained in the care and use of safety apparatus in accordance with safety engineering
practices;

 Are trained in emergency procedures (first aid).

Instructions and Warnings


The following indicators and standard definitions are used:

PCS-9830A Phase Controlled Switching Device i


Date: 2012-09-20
Preface

DANGER!

It means that death, severe personal injury, or considerable equipment damage will occur if safety
precautions are disregarded.

WARNING!

It means that death, severe personal, or considerable equipment damage could occur if safety
precautions are disregarded.

CAUTION!

It means that light personal injury or equipment damage may occur if safety precautions are
disregarded. This particularly applies to damage to the device and to resulting damage of the
protected equipment.

WARNING!

The firmware may be upgraded to add new features or enhance/modify existing features, please
make sure that the version of this manual is compatible with the product in your hand.

WARNING!

During operation of electrical equipment, certain parts of these devices are under high voltage.
Severe personal injury or significant equipment damage could result from improper behavior.

Only qualified personnel should work on this equipment or in the vicinity of this equipment. These
personnel must be familiar with all warnings and service procedures described in this manual, as
well as safety regulations.

In particular, the general facility and safety regulations for work with high-voltage equipment must
be observed. Noncompliance may result in death, injury, or significant equipment damage.

DANGER!

Never allow the current transformer (CT) secondary circuit connected to this equipment to be
opened while the primary system is live. Opening the CT circuit will produce a dangerously high
voltage.

WARNING!

 Exposed terminals

Do not touch the exposed terminals of this equipment while the power is on, as the high voltage
generated is dangerous

ii PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
Preface

 Residual voltage

Hazardous voltage can be present in the DC circuit just after switching off the DC power supply. It
takes a few seconds for the voltage to discharge.

CAUTION!

 Earth

The earthing terminal of the equipment must be securely earthed

 Operating environment

The equipment must only be used within the range of ambient environment detailed in the
specification and in an environment free of abnormal vibration.

 Ratings

Before applying AC voltage and current or the DC power supply to the equipment, check that they
conform to the equipment ratings.

 Printed circuit board

Do not attach and remove printed circuit boards when DC power to the equipment is on, as this
may cause the equipment to malfunction.

 External circuit

When connecting the output contacts of the equipment to an external circuit, carefully check the
supply voltage used in order to prevent the connected circuit from overheating.

 Connection cable

Carefully handle the connection cable without applying excessive force.

Copyright

Version: R1.00 NR ELECTRIC CO., LTD.

69 Suyuan Avenue. Jiangning, Nanjing 211102, China

P/N: EN_ZNKZ3000.0086.0001 Tel: +86-25-87178185, Fax: +86-25-87178208

Website: www.nrelect.com, www.nari-relays.com

Copyright © NR 2012. All rights reserved Email: [email protected]

We reserve all rights to this document and to the information contained herein. Improper use in particular reproduction and dissemination
to third parties is strictly forbidden except where expressly authorized.

The information in this manual is carefully checked periodically, and necessary corrections will be included in future editions. If
nevertheless any errors are detected, suggestions for correction or improvement are greatly appreciated.

We reserve the rights to make technical improvements without notice.

PCS-9830A Phase Controlled Switching Device iii


Date: 2012-09-20
Preface

Documentation Structure

The manual provides a functional and technical description of this device and a comprehensive
set of instructions for the device’s use and application.

All contents provided by this manual are summarized as below:

1 Introduction
Briefly introduce the application, functions and features about this device.

2 Technical Data
Introduce the technical data about this device, such as electrical specifications, mechanical
specifications, ambient temperature and humidity range, communication port parameters, type
tests, setting ranges and accuracy limits and the certifications that our products have passed.

3 Operation Theory
Introduce a comprehensive and detailed functional description of all protective elements.

4 Supervision
Introduce the automatic self-supervision function of this device.

5 Hardware
Introduce the main function carried out by each plug-in module of this device and providing the
definition of pins of each plug-in module.

6 Human Machine Interface


Introduce the hardware of the human machine interface (HMI) module and a detailed guide for the
user how to use this device through HMI. It also lists all the information which can be view through
HMI, such as settings, measurements, all kinds of reports etc.

7 Installation
Introduce the recommendations on unpacking, handling, inspection and storage of this device. A
guide to the mechanical and electrical installation of this device is also provided, incorporating
earthing recommendations. A typical wiring connection to this device is indicated.

8 Commissioning
Introduce how to commission this device, comprising checks on the calibration and functionality of
this device.

9 Maintenance
A general maintenance policy for this device is outlined.

10 Decommissioning and Disposal


A general decommissioning and disposal policy for this device is outlined.

iv PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
Preface

11 Manual Version History


List the instruction manual version and the modification history records.

Typographic and Graphical Conventions

Deviations may be permitted in drawings and tables when the type of designator can be obviously
derived from the illustration.

The following symbols are used in drawings:

&

AND gate

≥1

OR gate

Comparator

BI
Binary signal via opto-coupler

SET I>
Input signal from comparator with setting

EN
Input signal of logic setting for function enabling

SIG
Input of binary signal except those signals via opto-coupler

XXX
Output signal

Timer
t
t
Timer (optional definite-time or inverse-time characteristic)

10ms 0ms
Timer [delay pickup (10ms), delay dropoff (0ms), non-settable]
[XXX] 0ms
Timer (delay pickup, settable)

PCS-9830A Phase Controlled Switching Device v


Date: 2012-09-20
Preface

0ms [XXX]
Timer (delay dropoff, settable)
[XXX] [XXX]
Timer (delay pickup, delay dropoff, settable)
IDMT
Timer (inverse-time characteristic)

---xxx is the symbol

vi PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
1 Introduction

1 Introduction

Table of Contents
1 Introduction..................................................................................... 1-a
1.1 Application....................................................................................................... 1-1
1.2 Functions ......................................................................................................... 1-1
1.3 Features ........................................................................................................... 1-2
1.4 Ordering Options ............................................................................................ 1-3

List of Figures
Figure 1.1-1 Typical application of PCS-9830A ........................................................................ 1-1

PCS-9830A Phase Controlled Switching Device 1-a


Date: 2012-09-20
1 Introduction

1-b PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
1 Introduction

1.1 Application
Controlled switching is a method for eliminating harmful electrical transients via time controlled
switching operations. Closing or opening commands to the circuit breaker are delayed in such a
way that making or contact separation will occur at the optimum time instant related to the phase
angle.

The phase controlled switching device PCS-9830A supplied by NR have wide applications in
single-pole operated circuit breakers of shunt capacitor banks, shunt reactors, no-load power
transformers and no-load overhead lines. In addition, it have real-time compensation and adaptive
function, which both make it possible to achieve added precision in the timing of the controlled
circuit breaker.

The following figure is a typical application of PCS-9830A.

Busbar phase A

Busbar phase B

Busbar phase C

Feedback analog quantity Reference voltage input

Closing command output


Circuit breakers

PCS-9830A
Opening command output
PHASE CONTROLLED
SWITCHING DEVICE

Closing command input


Opening command input
Load: Shunt capacitor banks, shunt reactors, no-load power
transformers and no-load overhead lines.

Transducer

Figure 1.1-1 Typical application of PCS-9830A

1.2 Functions

Controlled Closing and Opening Functions


Controlled closing function
Controlled opening function
Adaptive function based on current detection
Adaptive function based on position detection

PCS-9830A Phase Controlled Switching Device 1-1


Date: 2012-09-20
1 Introduction

Real-time compensation based on ambient temperature


Real-time compensation based on control voltage
Real-time compensation based on standing time
Automatic calibration of current zero-drift
Binary input via opto-coupler with high voltage
Binary output via potential-free contact and IGBT

Auxiliary Management Functions


Current and voltage metering
Binary input status displaying
Frequency metering
Wave recording
Self-supervision
SOE records, latest records of following elements state changing: operating
abnormality alarm elements, self-supervision alarm elements, controlled
closing and opening operation elements, binary input elements
Dbg2000( A debugging software, in PC )
WaveEv( A waveform analysis software, in PC )

1.3 Features

 On the premise of 96 samples per cycle, all data measurement, calculation and logic
discrimination could be finished within one sampling period.

 Provide controlled closing function and controlled opening function, which can be chosen to
be enabled or disabled.

 Adaptive function and compensation strategy are both provided.

 Provide binary inputs via opto-coupler with high voltage as well as potential-free contact.

 The HMI interface with a 128×64-dot LCD, 8 LED indicators and a 2-multifunction-key keypad
on the front panel is very friendly to the user.

 This device is equipped with two RS-232 serial debugging interfaces for communication with
Dbg2000 as well as WaveEv software, on the convenience of viewing and modifying settings,
displaying metering values and wave recording.

 Strong function optional module to satisfy all kinds of requirement on site.

 Event recording function: 32 latest operation reports, 64 alarm records and 64 binary
input/output state change records.

 This device can store the settings, history reports and the recorded waves in a 2Mbit FRAM,
which makes these information does not be lost even the power supply is switched off.

1-2 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
1 Introduction

1.4 Ordering Options

This device is very customizable to meet the requirements of a practical engineering. The
following table shows the various available customizable options when ordering a PCS-9830A.

 Secondary AC current Input: 1A or 5A.

 Control voltage: 110Vdc or 220Vdc

 The numbers of CT and VT in AI module, 3 CTs, 3 VTs or 6 VTs.

NOTE: In the below chapters, the referred AI module means the one with 3 CTs, 3 VTs.

PCS-9830A Phase Controlled Switching Device 1-3


Date: 2012-09-20
1 Introduction

1-4 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
2 Technical Data

2 Technical Data

Table of Contents
2 Technical Data................................................................................. 2-a
2.1 Electrical Specifications ................................................................................. 2-1
2.1.1 AC Current Input .................................................................................................................. 2-1

2.1.2 AC Voltage Input .................................................................................................................. 2-1

2.1.3 Power Supply....................................................................................................................... 2-1

2.1.4 Binary Input .......................................................................................................................... 2-1

2.1.5 Binary Output ....................................................................................................................... 2-2

2.2 Mechanical Specifications ............................................................................. 2-2


2.3 Ambient Temperature and Humidity Range .................................................. 2-2
2.4 Debuging Port ................................................................................................. 2-3
2.5 Type Tests ........................................................................................................ 2-3
2.5.1 Environmental Tests ............................................................................................................ 2-3

2.5.2 Mechanical Tests ................................................................................................................. 2-3

2.5.3 Electrical Tests ..................................................................................................................... 2-3

2.5.4 Electromagnetic Compatibility ............................................................................................. 2-3

2.6 Certifications ................................................................................................... 2-4


2.7 Protective Functions....................................................................................... 2-4
2.7.1 Metering Scope and Accuracy............................................................................................. 2-4

2.7.2 Telesignal ............................................................................................................................. 2-4

PCS-9830A Phase Controlled Switching Device 2-a


Date: 2012-09-20
2 Technical Data

2-b PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
2 Technical Data

2.1 Electrical Specifications


2.1.1 AC Current Input
Phase rotation ABC
Rated frequency 50±5Hz, 60±5Hz
Rated Current (In) 1A 5A
Linear to 0.05In~2In
continuously 2×In
Thermal for 10s 12×In
withstand for 1s 30×In
capability for half
75×In
a cycle
Burden < 0.2 VA/phase @ In < 0.4 VA/phase @ In

2.1.2 AC Voltage Input


Standard Per IEC 60255-6:1988
Phase rotation ABC
Rated frequency 50±5Hz, 60±5Hz
Rated Voltage (Vn) 100V~130V (Phase)
Linear to 1V~130V
Thermal continuously 130V
withstand 10s 200V
capability 1s 250V
Burden < 0.1VA/phase @ Vn

2.1.3 Power Supply


Rated Voltage (Un) 30Vdc, 110Vdc, 220Vdc, 110Vac(50Hz), 220Vac(50Hz)
Variation (80% ~ 120%)Un
Ripple in the DC
Max 15% of the DC value. Per IEC 60255-11:1979
auxiliary voltage
Per IEC 61000-4-11:1994, IEC 60255-11:1979
Voltage dips and voltage short
20ms for interruption without de-energizing, dips 60% of Un without reset up to
interruptions
100ms
Quiescent
<10W
condition
Burden
Operating
<15W
condition

2.1.4 Binary Input


Rated Voltage (VBI) 110V 220V
Pickup voltage 70% VBI 70% VBI
Dropout voltage 55% VBI 55% VBI

PCS-9830A Phase Controlled Switching Device 2-1


Date: 2012-09-20
2 Technical Data

Maximum permitted voltage 120% VBI 120% VBI


Withstand 2000Vac
Optical isolated binary input with fixed debounce time 20ms

2.1.5 Binary Output


Item Used for tripping contact Used for signal contact
Output model Potential-free contact connecting with IGBT
Max system voltage 250Vac/dc
Test voltage across
1000Vac RMS for 1min
open contact
Making Capacity >1000W at L/R=40ms
Breaking capacity 0.4A resistance 0.2A resistance
at 250Vdc 0.2A inductive(L/R=40ms) 0.05A inductive (L/R=40ms)
Continuous carry 8A 5A
Short duration current 50A for 200ms
Pickup time (Typical/Max) 0.1ms/0.2ms 10ms/15ms
Dropout time
8ms/10ms 4ms/10ms
(Typical/Max)
Bounce time none 1ms
loaded
100,000 operations minimum 100,000 operations minimum
contact
Durability
unloaded
10,000,000 operations minimum 10,000,000 operations minimum
contact

2.2 Mechanical Specifications


Enclosure dimensions 188.92×177.00×165.31 (W×H×D, unit: mm)
Trepanning dimensions 173.00×178.00, M5 screw (W×H, unit: mm)
Mounting Way Flush mounted
Weight per device Approx. 3kg
Small control module: 8 LED indicators, a 2-multifunction-key keypad and a
Local control panel
128×64-dot LCD
Display language Optional: Chinese, English
Housing material Aluminum
Housing color Silver grey
Location of terminals Rear panel of the device
Front side: IP40, up to IP51
Protection class
Rear side, connection terminals: IP20
(per IEC60529:1989)
Other Sides: IP30

2.3 Ambient Temperature and Humidity Range


Operating temperature range -25°C ~ +55°C
Transport and storage temperature range -40°C ~ +70°C

2-2 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
2 Technical Data

Permissible humidity 5% ~ 95%, condensation not permissible

2.4 Debuging Port


Interfaces for debuging through “Dbg2000” software
Port number 1
Front RS-232 (EIA) Baud rate 19200bps
Safety level Isolation to ELV level

2.5 Type Tests


2.5.1 Environmental Tests
Dry heat test IEC60068-2-2:1974, 16 h at +55°C
Dry cold test IEC60068-2-1:1990, 16 h at -10°C
IEC60068-2-30:1980, two (12+12 hours) cycles, 95%RH,
Damp heat test, cyclic
low temperature +25°C, high temperature +55°C

2.5.2 Mechanical Tests


Vibration IEC60255-21-1:1988, Class I
Shock and bump IEC60255-21-2:1988, Class I

2.5.3 Electrical Tests


Dielectric tests IEC60255-5:2000, test voltage: 2kV, 50Hz, 1min
IEC60255-5:2000, test voltage: 5kV, unipolar impulses, waveform 1.2/50μs,
Impulse voltage tests
source energy 0.5J
Insulation
IEC60255-5:2000, insulation resistance >100MΩ, 500Vdc
measurement

2.5.4 Electromagnetic Compatibility


1MHz burst disturbance tests IEC60255-22-1:1988
(idt IEC61000-4-2:1995), Class III
- Common mode 2.5kV
- Differential mode 1.0kV
Electrostatic discharge tests IEC60255-22-2 :1996
(idt IEC 61000-4-2) Class IV
-For contact discharge 8.0kV
-For air discharge 15.0kV
Radio frequency interference tests IEC60255-22-3:2000
(idt IEC 61000-4-3:1995 ) class III
Frequency sweep
-Radiated amplitude-modulated 10V/m(rms), f=80…1000MHz
Spot frequency
-Radiated amplitude-modulated 10Vm(rms), f=80MHz/160MHz
/450MHz/900MHz

PCS-9830A Phase Controlled Switching Device 2-3


Date: 2012-09-20
2 Technical Data

- Radiated pulse-modulated 10Vm(rms), f=900MHz


Fast transient disturbance tests IEC60255-22-4:2002
(idt IEC 61000-4-4)
- Power supply, I/O & Earth terminals Class IV, 4kV, 2.5kHz, 5/50ns
- Communication terminals Class IV, 2kV, 5.0kHz, 5/50ns
Surge immunity tests IEC60255-22-5:2002
(idt IEC 61000-4-5:1995) , Class III
- Power supply, AC inputs, I/O terminals 1.2/50us,
2kV, line to earth;
1kV, line to line
Conducted RF electromagnetic disturbance IEC60255-22-6, Class III
- Power supply, AC, I/O, Comm. Terminal 10V(rms), 150kHz~80MHz
Power frequency magnetic field immunity IEC61000-4-8:1993, Class V
100A/m for 1min
1000A/m for 3s
Pulse magnetic field immunity IEC61000-4-9:1993, Class V
6.4/16 us
1000A/m for 3s
Damped oscillatory magnetic field immunity IEC61000-4-10:1993, Class V
100kHz & 1MHz – 100A/m

2.6 Certifications
 ISO9001:2008

 ISO14001:2004

 OHSAS18001:2007

 ISO10012:2003

 CMMI L4

 EMC: 2004/108/EC, EN50263:1999

 Products safety(PS): 2006/95/EC, EN61010-1:2001

2.7 Protective Functions


2.7.1 Metering Scope and Accuracy
Phase current 0.05 ~ 1.4×In
Accuracy ≤ 0.5% of reading

2.7.2 Telesignal

Resolution of telesignals ≤ 1ms

Input of telesignals Potential-free contact

2-4 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
2 Technical Data

Resolution of SOE ≤ 2ms

PCS-9830A Phase Controlled Switching Device 2-5


Date: 2012-09-20
2 Technical Data

2-6 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
3 Operation Theory

3 Operation Theory

Table of Contents
3 Operation Theory ............................................................................ 3-a
3.1 Closing/Opening Detectors ............................................................................ 3-1
3.2 Controlled Closing/Opening .......................................................................... 3-2
3.3 Real-time Compensation Function ................................................................ 3-3
3.4 Adaptive Function ........................................................................................... 3-4

List of Figures
Figure 3.1-1 Logic diagram of the closing detector ................................................................ 3-1

Figure 3.1-2 Logic diagram of the opening detector ............................................................... 3-1

Figure 3.2-1 Diagram of the controlled closing process ........................................................ 3-2

PCS-9830A Phase Controlled Switching Device 3-a


Date: 2012-09-20
3 Operation Theory

3-b PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
3 Operation Theory

The phase controlled switching device PCS-9830A provides controlled closing function, controlled
opening function, real-time compensation and adaptive function. The following sections will detail
the individual functions in this device.

3.1 Controlled Closing/Opening Detectors

1. Controlled closing detector

The controlled closing detector operates if all of the following conditions are satisfied.

 The logic setting of controlled closing [En_CtrlCls] is enabled;

 The device receives the closing command [BI_ClsCmd];

 The reference voltage sampled is normal.

The following figure shows the logic of closing detector of the PCS-9830A

En_ CtrlCls &


t 0ms
BI _ClsCmd FD_Cls
Alm_Sample_Volt

Figure 3.1-1 Logic diagram of the closing detector

Where:

[En_CtrlCls] is the logic setting of controlled closing;

[BI_ClsCmd] is the binary input of closing command;

[Alm_Sample_Volt] means the sampled voltage is abnormal;

[FD_Cls] means controlled closing detector element operates.

2. Controlled opening detector

The controlled opening detector operates if all of the following conditions are satisfied.

 The logic setting of controlled opening [En_CtrlOpn] is enabled;

 The device receives the opening command [BI_OpnCmd];

 The reference voltage sampled is normal.

The following figure shows the logic of opening detector of the PCS-9830A.

En_ CtrlOpn &


t 0ms
BI_OpnCmd FD_Opn
Alm_Sample_Volt

Figure 3.1-2 Logic diagram of the opening detector

Where:

PCS-9830A Phase Controlled Switching Device 3-1


Date: 2012-09-20
3 Operation Theory

[En_CtrlOpn] is the logic setting of controlled opening;

[BI_OpnCmd] is the binary input of opening command;

[Alm_Sample_Volt] means the sampled voltage is abnormal;

[FD_Opn] means controlled opening detector element operates.

3.2 Controlled Closing/Opening

The purpose of controlled closing/opening is to minimize the energizing/de-energizing transients


(voltage transients as well as inrush currents).

Take the phase A controlled closing processing of a capacitive load for example, in order to avoid
switching transients, the making instant in this case shall be at voltage zero.

After the controlled closing detector operates, the device enters the controlled closing program. It
will determine a reference time instant T0, related to the phase angle of the busbar voltage. When
this has been done, and after an internally created waiting time Tdelay, the controller will then give
an output closing command at time T1 to the circuit breaker. Then, after the delay time of
secondary circuit Tcdly and the make time of the circuit breaker Tclose, meanwhile subtracting the
pre-arcing time Tpres, it reaches the pre-programmed target point T3 (a voltage zero). All of Tclose,
Tcdly and Tpres have been pre-programmed into the controller. The circuit breaker will then make
the closing at the correct time instant and minimize the switching transients.

N*T/2
Tpres
Tdelay Tcdly Tclose

Voltage

T3 T4
T0 T1 T2 Time

Input Output Pre-programmed


Reference
command command target point
point

Figure 3.2-1 Diagram of the controlled closing process

Tdelay=N*T/2-Tclose-Tcdly+Tpres

Where:

Tdelay is the internally created waiting time;

T is the cycle of the voltage;

Tclose is the make time of the circuit breaker, supplied by the manufacturer;

3-2 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
3 Operation Theory

Tcdly is the delay time of secondary circuit, mainly the operation delay of the mid-relay and
IGBT;

Tpres is the pre-arcing time;

N can be 1, 2, 3···

For the controlled opening process, it is similar to the controlled closing process.

Moreover, there are real-time compensation function and adaptive function in controlled
closing/opening, as shown below.

3.3 Real-time Compensation Function

Deviations from the intended targets may be caused by variations in the operating conditions. The
operating conditions that may cause changes of the circuit breaker operating times are, for
example, gradually increasing contact burn-off caused by many switching operations, change of
ambient temperature and variations of the auxiliary voltage.

The real-time compensation mode is especially meaningful for circuit breakers that are not
operated frequently, or the external parameter change influencing circuit breaker behavior can be
considered as major. In the compensation mode, any external parameter variation giving a known
dependency can be compensated for by suitable sensors. The sensors are optional and are to be
ordered separately. The internally created waiting time Tdelay is,

Tdelay=N*20ms-Tclose-Tcdly+Tpres+△Trt

Where:

△Trt is the real-time compensation time.

The following are three examples of real-time compensation, based on ambient temperature,
control voltage, and standing time respectively. In addition, there are other variations in the
operating conditions to real-time compensate for the controlled closing/opening.

1. Real-time control based on ambient temperature

It is studied that the switching time of the circuit breaker relates to the ambient temperature.
Usually, the lower the temperature is, the longer switching time is. Considering the affect of
the ambient temperature, the device detects the ambient temperature and calculates the time
△Trt according to the temperature compensation curve pre-programmed into the device.

2. Real-time control based on control voltage

After a large number of experiments, it is found that the switching time will gradually reduce as
the control voltage gradually increases, but when the control voltage is in the rated voltage of
the range 85% ~ 110%, the time deviation is small. According to the control voltage
compensation curve, the device calculates the time △Trt to compensate for Tdelay when
needed.

3. Real-time control based on standing time

PCS-9830A Phase Controlled Switching Device 3-3


Date: 2012-09-20
3 Operation Theory

The switching time of the circuit breaker is also related to the standing time. The longer the
standing time is, the longer the switching time is. By the consider of the affect, it is to
compensate for the time Tdelay via the time △Trt calculated according to the standing time
compensation curve pre-programmed into the device.

3.4 Adaptive Function

For the adaptive function, it is mainly intended for the circuit breakers that are frequently switched,
or the external parameter change influencing circuit breaker behavior can be considered as minor.
The functioning principle of the compensation control is that a detected error from the target will be
compensated for in the next controlled operation.

There are two strategies of adaptive function, based on analog quantity detection(current/voltage)
or position detection.

1. Adaptive control based on analog quantity detection(current/voltage)

In the mode, adaptation control adjusts the internally created waiting time Tdelay by detection
the signals of current start instants, voltage onset instants, or current change instants from the
CTs or VTs.

2. Adaptive control based on position detection

Alternatively, it also can detect the contact touch instants or the contact separation instants of
the circuit breakers by the use of auxiliary contacts to adjust the internally created waiting time
Tdelay.

NOTE: The above examples are simplied to mainly explain the theories. In actual

controlled closing/opening, other conditions need to be considered, such as the mode of


busbar connection, the load characteristic and the choice of reference voltage etc.

3-4 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
4 Supervision and Metering

4 Supervision and Metering

Table of Contents
4 Supervision and Metering .............................................................. 4-a
4.1 Overview .......................................................................................................... 4-1
4.2 Device Self-supervision .................................................................................. 4-1
4.2.1 Device Hardware Supervision ............................................................................................. 4-1

4.2.2 Setting Supervision .............................................................................................................. 4-1

4.2.3 Supervision Alarms .............................................................................................................. 4-1

4.3 Metering ........................................................................................................... 4-2

List of Tables
Table 4.2-1 Alarm description ...................................................................................................... 4-1

Table 4.2-2 Troubleshooting ....................................................................................................... 4-2

PCS-9830A Phase Controlled Switching Device 4-a


Date: 2012-09-20
4 Supervision and Metering

4-b PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
4 Supervision and Metering

4.1 Overview

Though the device is in non-operating state under normal conditions, it is waiting for a command to
occur at any time and must operate for the command without fail. When the equipment is in
energizing process before the LED “HEALTHY” is on, the equipment needs to be checked to
ensure there are no errors. Therefore, the automatic supervision function, which checks the health
of the device when startup and during normal operation, plays an important role.

The numerical device based on the microprocessor operations is suitable for implementing this
automatic supervision function of the device.

In case a fatal fault is detected during automatic supervision, the equipment will be blocked out. It
means that device is out of service. Therefore you must re-energize the device or even replace a
module to make device back into service.

The device also provides some auxiliary functions. All these make the device meet the demands of
the modern industry requirements.

4.2 Device Self-supervision

4.2.1 Device Hardware Supervision


The RAM, ROM, DC/DC and A/D elements on the CPU module are monitored to ensure whether
they are damaged or have some errors. If any one of them is detected damaged or having error,
the equipment will be blocked and issues a relevant alarm signal. The LED indicator “HEALTHY”
will be extinguished.

4.2.2 Setting Supervision


The settings are checked to ensure them to be correct and not to be modified. If the settings are
checked to have any error in them, the alarm signal [Alm_Settings] will be issued and this
equipment will be blocked at the same time. The LED indicator “HEALTHY” will be extinguished.

4.2.3 Supervision Alarms


If there is any abnormality in the operating or firmware error detected by equipment self-diagnostic,
alarm report will be displayed instantaneously on the LCD. Therefore, the default display will be
replaced by the alarm report. In case there is more than one alarm has occurred, the information
will be displayed alternately on the LCD.

Alarm elements listed below may be displayed.

Table 4.2-1 Alarm description

No. Alarm Information Description


1 Alm_Settings There is a mistake in the settings.
2 Alm_ModSetting The settings have been modified.
3 Alm_CompLogSet The logical setting of adaptive control is abnormal.

PCS-9830A Phase Controlled Switching Device 4-1


Date: 2012-09-20
4 Supervision and Metering

4 Alm_ClsTime_3P Three phases of closing times are abnormal.


5 Alm_Sample_Volt The sampled voltage is abnormal.
6 Alm_DS_BB The position of disconnection switch of busbar is abnormal.
7 Alm_ROM The content of the ROM is broken.
8 Alm_Upgraded The program of the device is upgraded.
2
9 Alm_I2C The I C bus is abnormal.
10 Alm_CTS The current transformer circuit is abnormal.
11 Alm_CCS The closing circuit is broken.
12 Alm_TCS The tripping circuit is broken.
13 Alm_Uncharged The circuit breaker is uncharged.
14 Alm_SPI1 The SPI1 interface is abnormal.
15 Alm_SPI2 The SPI2 interface is abnormal.
16 Alm_AD The A/D converter is abnormal.
17 Alm_Recovered The device is recovered

Table 4.2-2 Troubleshooting

No. Item Handling suggestion


1 Alm_Settings Please check the value of settings, set them again according to the actual ones.
2 Alm_ModSetting After modifying the settings, reset the device to put the settings into operation.
Please check the settings [En_CTMonitor], [En_VTMonitor] and
3
Alm_CompLogSet [En_PosMonitor] and modify them.
Please find the reasons of the relatively large difference in three phases of
4
Alm_ClsTime_3P closing times
5 Alm_Sample_Volt Please check the PT, the connection of AI module.
6 Alm_DS_BB Please check the auxiliary contacts and connections of circuit breakers
7 Alm_ROM Please inform the manufacture or the agent for repair.
8 Alm_Upgraded Please check the program version of the device.
9 Alm_I2C Please inform the manufacture or the agent for repair.
10 Alm_CTS Please check the CT secondary circuit.
11 Alm_CCS Please check the closing circuit.
12 Alm_TCS Please check the tripping circuit.
13 Alm_Uncharged Please check the operating mechanism of the circuit breaker.
14 Alm_SPI1 Please inform the manufacture or the agent for repair.
15 Alm_SPI2 Please inform the manufacture or the agent for repair.
16 Alm_AD Please inform the manufacture or the agent for repair.

4.3 Metering

The device produces a variety of both directly and calculated power system quantities. These
measurement values are updated on a per second basis.

The measurands include Ua, Ub, Uc, Ia, Ib, Ic and f. These measurands are gotten by real-time
calculation and real-time summation locally.

4-2 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
5 Hardware

5 Hardware

Table of Contents
5 Hardware ......................................................................................... 5-a
5.1 Overview .......................................................................................................... 5-1
5.2 AI Module ......................................................................................................... 5-2
5.3 CPU Module ..................................................................................................... 5-3
5.4 HMI Module ...................................................................................................... 5-4
5.5 PWR Module .................................................................................................... 5-4
5.6 BI Module ......................................................................................................... 5-5
5.7 BO Module ....................................................................................................... 5-6

List of Figures
Figure 5.1-1 Hardware block diagram ....................................................................................... 5-1

Figure 5.1-2 Front panel of the PCS-9830A .............................................................................. 5-1

Figure 5.1-3 Rear panel of the PCS-9830A ............................................................................... 5-2

PCS-9830A Phase Controlled Switching Device 5-a


Date: 2012-09-20
5 Hardware

5.1 Overview
The modular design of this device allows it to be easily upgraded or repaired by a qualified service
person. The faceplate is hinged to allow easy access to the configurable modules, and
back-plugging structure design makes it easy to repair or replace any modules.

There are several types of hardware modules in this device; each module takes a different part in
this device. This device mainly consists of PWR module, CPU module, AI module, BI module, BO
module and HMI module. Following figure shows the hardware block diagram of this device.

External BI

BI Module
PWR Module
Transducer
analog input CPU Module

Analog input AI Module

BO Module

LCD Single phase Single phase


Communication
LED Indications closing opening
HMI
binary binary
Keypad and Keys Wave recording outputs outputs

Figure 5.1-1 Hardware block diagram

The following two figures show the front panel and the rear panel of this device.

PCS-9830A
PHASE CONTROLLED SWITCHING DEVICE

HEALTHY
EN
U
M

ALARM ENT

DJU S
A

ESC

RS- 232

Figure 5.1-2 Front panel of the PCS-9830A

PCS-9830A Phase Controlled Switching Device 5-1


Date: 2012-09-20
5 Hardware

AI CPU BI BO PWR

01 02

03 04

05 06

07 08

09 10

11 12

13 14

15 16

17 18

19 20

21 22

23 24

Figure 5.1-3 Rear panel of the PCS-9830A

5.2 AI Module

The AI module is an analog input unit. It contains three current transformers and three voltage
transformer. It can transform these high AC input values to relevant low AC output value, which are
suited to the analog inputs of the CPU module. The transformers are used both to step-down the
currents to levels appropriate to the device’s electronic circuitry and to provide effective isolation
between the device and the power system. A low pass filter circuit is connected to each
transformer secondary circuit for reducing the noise of each analog AC input signal.

NOTE: The rated value of the input current transformer is optional: 1A or 5A. The rated

value of the CT must be definitely declared in the technical scheme and the contract.

NOTE: Because the rated value of the input current transformer is optional, it is

necessary to check that whether the rated values of the current transformer inputs are
accordant to the demand of the practical engineering before putting the device into
operation.

A 24-pin connector is fixed on the front of the AI module. The terminal definition of the connector is
described as below.

Pin connections on the connector of the AI module:

5-2 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
5 Hardware

Pin No. Sign Description


601 Ia
Phase A current input
602 Ian
603 Ib
Phase B current input
604 Ibn
605 Ic
Phase C current input
606 Icn
607 /
608 / Not used
609 ~ 616 /
617 Ua
Phase A busbar voltage input
618 Uan
619 Ub
Phase B busbar voltage input
620 Ubn
621 Uc
Phase C busbar voltage input
622 Ucn
623~ 624 / Not used

NOTE: Never allow the current transformer (CT) secondary circuit connected to this

device to be opened while the primary system is live. Opening the CT circuit will produce
a dangerously high voltage. If this safety precaution is disregarded, personal death,
severe personal injury or considerable equipment damage will occur.

NOTE: Never allow the current transformer (CT) secondary circuit connected to this

device to be opened while the primary system is live. Opening the CT circuit will produce
a dangerously high voltage. If this safety precaution is disregarded, personal death,
severe personal injury or considerable equipment damage will occur.

5.3 CPU Module

The CPU module is the kernel part of this equipment, and contains a powerful microchip processor
and some necessary electronic elements. This powerful processor performs all of the functions for
the device. There are several A/D conversion circuits on this module, which are used to convert
the AC analog signals to corresponding DC signals for fulfilling the demand of the electrical level
standard. A high-accuracy clock chip is contained in this module, it provide accurate current time
for the PCS-9830A.

A 12-pin connector is fixed on the front of the CPU module. The terminal definition of the connector
is described as below.

Pin connections on the connector of the CPU module:

PCS-9830A Phase Controlled Switching Device 5-3


Date: 2012-09-20
5 Hardware

Pin No. Sign Description


401 Opto Optocoupler common terminal
402 BI_ClsCmd Closing command binary input
403 BI_OpnCmd Opening command binary input
404 BI_52a_A
405 BI_52a_B Closing position auxiliary contact binary input
406 BI_52a_C
407 BI_52b_A
408 BI_52b_B Opening position auxiliary contact binary input
409 BI_52b_C
410 /
411 / Not used
412 /

NOTE: Usually, the terminal 402/403 connects to the closing/opening command binary

input, and in BO module the three phases closing/opening contacts connect to the three
phases closing /opening coils respectively.

NOTE: For the occasions only needing the closing functions of single-pole operated

circuit breakers, the terminal 403 and the three phases opening contacts in BO module
are unused, and the opening control circuits are connected to the opening coils directly,
bypassing the device.

5.4 HMI Module

The HMI module is installed on the front panel of this device. It is used to observe the running
status and event information, and device operation mode. It can help the user to know the status
of this device and detailed event information easily, and provide convenient and friendly access
interface for the user. For further details, see Chapter 6.

5.5 PWR Module

The power supply module is a converter module with electrical insulation between input and
output. The power supply module has an input voltage range and relevant parameters as
described in Chapter 2. The standardized output voltages are +3.3Vdc, +5Vdc, ±12Vdc and
+24Vdc. The tolerances of the output voltages are continuously monitored.

The power supply also contains one special binary output, used as a device alarm contact.

A 12-pin connector is fixed on the front of the PWR module. The terminal definition of the
connector is described as below.

Pin connections on the connector of the PWR module:

5-4 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
5 Hardware

Pin No. Sign Description


101 PS+
PWR Power supply input
102 PS-
103 ALM_1
Device failure signal output
104 ALM_2
105 \
\ Not used
106 \
107 \
\ Not used
108 \
109 \
\ Not used
110 \
111 \
\ Not used
112 \

5.6 BI Module

The BI module is mainly used to receive the analog inputs transmitted by the transducers,
transform them into binary inputs and then transmit to CPU module. The analog quantitys sampled
by transducers are to compensate for operating time due to ambient temperature changes and
variations in control voltage, etc.

A 22-pin connector is fixed on the front of the BI module. The terminal definition of the connector is
described as below. Pin connections on the connector of the BI module:

Pin No. Sign Description


301 FGND Ground terminal
Ground terminal of external binary output
302 BI-COM
1~8
303 BI_Resv1 Reserved binary input 1
304 BI_Resv2 Reserved binary input 2
305 BI_Resv3 Reserved binary input 3
306 BI_Resv4 Reserved binary input 4
307 BI_Resv5 Reserved binary input 5
308 BI_Resv6 Reserved binary input 6
209 BI_Resv7 Reserved binary input 7
310 BI_Resv8 Reserved binary input 8
311~313 / Not used
314 AI1+ Transducer 1 input terminal
315 AI1- (Optional, 4-20mA or 0-10V)
316 AI2+ Transducer 2 input terminal
317 AI2- (Optional, 4-20mA or 0-10V)
318 AI3+ Transducer 3 input terminal
319 AI3- (Optional, 4-20mA or 0-10V)
320 AI4+ Transducer 4 input terminal

PCS-9830A Phase Controlled Switching Device 5-5


Date: 2012-09-20
5 Hardware

321 AI4- (Optional, 4-20mA or 0-10V)


322 FGND Ground terminal

NOTE: The transducer input terminal is optional, 4-20mA or 0-10V, which can be

selected by jumper on the BI module. In the real-time compensation mode based on


ambient temperature or control voltage, the device should be de-energized firstly, then
select the jumper.

5.7 BO Module

The BO module consists of three couples of closing command binary outputs and three couples of
opening command binary outputs which are used for closing or opening the three phases circuit
breakers respectively. The binary outputs are output via both potential-free contact and IGBT,
guarantying double-reliability and avoiding the time discreteness influence brought about by the single
potential-free contact mode.

A 22-pin connector is fixed on the front of the BO module. The terminal definition of the connector
is described as below. Pin connections on the connector of the BO module:

Pin No. Sign Description


201 Phase A closing command binary output
Cmd_ClsA
202 (via relay contact and IGBT)
203 /
Not used
204 /
205
Cmd_ClsB Phase B closing command binary output
206
207 /
Not used
208 /
209
Cmd_ClsC Phase C closing command binary output
210
211 /
Not used
212 /
213
Cmd_OpnA Phase A opening command binary output
214
215 /
Not used
216 /
217
Cmd_OpnB Phase B opening command binary output
218
219 /
Not used
220 /
221
Cmd_OpnC Phase C opening command binary output
222

5-6 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
6 Human Machine Interface

6 Human Machine Interface

Table of Contents
6 Human Machine Interface .............................................................. 6-a
6.1 Overview .......................................................................................................... 6-1
6.1.1 Design .................................................................................................................................. 6-1

6.1.2 Functionality......................................................................................................................... 6-1

6.1.3 Keypad and Keys................................................................................................................. 6-1

6.1.4 LED Indications.................................................................................................................... 6-2

6.1.5 Commissioning Port ............................................................................................................ 6-2

6.2 Understand the HMI Menu Tree ...................................................................... 6-3


6.2.1 Overview .............................................................................................................................. 6-3

6.2.2 Submenu of “VALUES” ........................................................................................................ 6-3

6.2.3 Submenu of “REPORT” ....................................................................................................... 6-3

6.2.4 Submenu of “SETTINGS” .................................................................................................... 6-4

6.2.5 Submenu of “CLOCK” ......................................................................................................... 6-4

6.2.6 Submenu of “VERSION”...................................................................................................... 6-5

6.3 Understand the LCD Display .......................................................................... 6-5


6.3.1 Default Display under Normal Operation Condition ............................................................ 6-5

6.3.2 Display Operation Report .................................................................................................... 6-5

6.3.3 Display Alarm Report ........................................................................................................... 6-6

6.3.4 Display Binary Input Report ................................................................................................. 6-7

6.4 View the Settings ............................................................................................ 6-8


6.5 View Device Status.......................................................................................... 6-8
6.5.1 Display Analogue Data ........................................................................................................ 6-8

6.5.2 Display the Command output Status ................................................................................... 6-9

6.5.3 Display the Operation Status ............................................................................................. 6-10

6.5.4 Display the Status of Binary Inputs .....................................................................................6-11

PCS-9830A Phase Controlled Switching Device 6-a


Date: 2012-09-20
6 Human Machine Interface

6.6 View Software Version .................................................................................. 6-12


6.7 View History Reports .................................................................................... 6-12
6.8 Operation through Keypad ........................................................................... 6-13
6.8.1 Password Protection .......................................................................................................... 6-13

6.8.2 Clock Set............................................................................................................................ 6-13

6.8.3 Delete Records .................................................................................................................. 6-14

List of Figures
Figure 6.1-1 Keypad on the front panel..................................................................................... 6-2

Figure 6.1-2 LED indicators on the front panel ........................................................................ 6-2

Figure 6.1-3 Commissioning port on the front panel .............................................................. 6-2

Figure 6.2-1 Menu tree diagram of this device ......................................................................... 6-3

Figure 6.3-1 Default display in normal situation ...................................................................... 6-5

Figure 6.3-2 LCD display of the operation report .................................................................... 6-5

Figure 6.3-3 LCD display of the alarm report ........................................................................... 6-6

Figure 6.3-4 LCD display of the alarm report ........................................................................... 6-7

Figure 6.4-1 LCD display of the settings .................................................................................. 6-8

Figure 6.5-1 LCD display of the measured values ................................................................... 6-9

Figure 6.5-2 LCD display of the device operation state ........................................................ 6-10

Figure 6.5-3 LCD display of the element state ....................................................................... 6-10

Figure 6.5-4 LCD display of the binary input state ................................................................ 6-11

Figure 6.6-1 LCD display of the software version.................................................................. 6-12

Figure 6.8-1 Password input interface .................................................................................... 6-13

Figure 6.8-2 Clock modification interface............................................................................... 6-14

6-b PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
6 Human Machine Interface

Human machine interface (HMI) is an important component of the equipment. It is a convenient


facility to access the device from the front local control panel of the device to view desired
information, such as measurement quantity, binary input state or program version etc. This
function is very helpful during commissioning before putting the equipment into service.

Furthermore, all above functions can be realized through the “Dbg2000” software via the
dedicated RS-232 commissioning port on the front panel of this device.

This chapter will describe human machine interface (HMI), menu tree and LCD display of the
equipment. In addition, how to input settings using keypad is described in detail. For more details
about the “Dbg2000” software, refer the “Dbg2000” software instruction manual.

6.1 Overview

6.1.1 Design
The human-machine interface consists of a human-machine interface (HMI) module which allows
the communication as simple as possible for the user. The HMI module includes:

 A 128×64-dot matrix backlight LCD visible in dim lighting conditions for monitoring status, fault
diagnostics and setting, etc.

 Eight LED indicators on the front panel of this device for denoting the status of this device.

 A 2-multifunction-key keypad on the front panel of the device for full access to the device.

 A dedicated RS-232 commissioning port for the “Dbg2000” software which is applied to
commissioning this device and can show the SOE reports of this device.

The front panel of the device is shown in Figure 5.1-2.

6.1.2 Functionality
 The HMI module helps to draw your attention to something that has occurred which may
activate a LED or a report display on the LCD.

 You as the operator may have own interest to view a certain data.

 Use menus navigate through menu commands and to locate the data of interest.

6.1.3 Keypad and Keys


The keypad and keys on the front panel of the equipment provide convenience to the operator to
view a certain data or change the device’s setting.

The keypad contains two multifunction keys, and each key can be pressed and turned. Pressing
the upper key “ENT” can confirm present operation or enter the next level menu; and pressing the
lower key “ESC” can cancel present operation or back to the upper level menu. Turning the upper
key “MENU” can choose an expected submenu or data; and turning the lower upper key
“ADJUST” can modify the selected data.

PCS-9830A Phase Controlled Switching Device 6-1


Date: 2012-09-20
6 Human Machine Interface

Following figure shows the keypad and keys.

EN

M
U
ENT

DJUS

T
ESC

Figure 6.1-1 Keypad on the front panel

6.1.4 LED Indications


There are eight LED indicators, which can indicate the operation state of the device. Following
figure shows the LED indicators on the front panel of the device.

HEALTHY

ALARM

Figure 6.1-2 LED indicators on the front panel

LED Display Remarks


When the device is not energized or any hardware failure
Off
HEALTHY is detected during supervision.
Steady Green When the equipment is in service.
Off When the equipment is in normal operating situation.
ALARM
Steady Yellow When any abnormality alarm is issued.
Other LEDs / Reserved

6.1.5 Commissioning Port


A dedicated RS-232 standardized commissioning port is on the front of this device. It is used to
uplink to the “DBG2000” software for configuring and debugging this device specially.

2 3 5

TXD RXD GND


RS-232

Figure 6.1-3 Commissioning port on the front panel

6-2 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
6 Human Machine Interface

6.2 Understand the HMI Menu Tree

6.2.1 Overview
In this part, the main layout of the menu which is shown on the LCD of the local human-machine
interface (HMI) will be described in detail.

Under normal operating condition, press the upper key can enter the main menu of this device.
The following figure shows the menu tree structure of this device.

Menu

1. VALUES 2. REPORT 3. SETTINGS 4. CLOCK

1. MEASUREMENT 1. OPR REPORT 1. EQUIP SETUP


2. CMD STATE 2. ALM REPORT 2.SYS SETTINGS 5. VERSION
3. OPR STATE 3. BI REPORT 3. FUN SETTINGS
4. BI STATE 4. CLR REPORT 4. CALBR SETUP
5. DEFAULT SET

Figure 6.2-1 Menu tree diagram of this device

6.2.2 Submenu of “VALUES”

Purpose:

This menu is used to view the measured values, binary input state and operation state.

Access Approach:

Move cursor to the item “VALUES” and press key “ENT” to enter its submenu after entering the
main menu of this device.

“VALUES” has following submenus.

No. Item Description


1 MEASUREMENT To display the measured values.
To display each phase closing and opening command output
2 CMD STATE
states.
3 OPR STATE To display the device operation state.
4 BI STATE To display the states of the binary inputs.

6.2.3 Submenu of “REPORT”

Purpose:

This menu is used to view all kinds of report including operation report, alarm report and so on.

Access Approach:

Move cursor to the item “REPORT” and press key “ENT” to enter its submenu after entering the
main menu of this device.

“REPORT” has following submenus.

PCS-9830A Phase Controlled Switching Device 6-3


Date: 2012-09-20
6 Human Machine Interface

No. Item Description


1 OPR REPORT To display the operation reports.
2 ALM REPORT To display the self-supervision report.
3 BI REPORT To display the binary input reports.
4 CLR REPORT To clear all kinds of history reports.

6.2.4 Submenu of “SETTINGS”

Purpose:

The menu is used to view and modify the settings and parameters of this device.

Access Approach:

Move cursor to the item “SETTINGS” and press key “ENT” to enter its submenu after entering the
main menu of this device.

“SETTINGS” has following submenus.

No. Item Description


1 EQUIP SETUP To view communication settings.
2 SYS SETTINGS To view system settings.
3 FUN SETTINGS To view compensation settings.
4 CALBR SETUP To view measurement accuracy manually.
5 DEFAULT SET To set all parameters and settings to default values.

“SYS SETTINGS” has following submenus.

No. Item Description


1 CFG SETTINGS To view the configuring settings.
2 SYS SETTINGS To view the system settings.

“FUN SETTINGS” has following submenus.

No. Item Description


1 CB SETTINGS To view the circuit breaker settings.
2 TEMP CURVE To view the ambient temperature settings.
3 VOLT CURVE To view the control voltage settings.
4 STAND CURVE To view the standing time settings
5 Resv CURVE To view the reserved settings.

6.2.5 Submenu of “CLOCK”

Purpose:

This menu is used to modify the clock of this device.

Access Approach:

Move cursor to the item “CLOCK” and press key “ENT” to enter the clock modification menu after
entering the main menu of this device.

6-4 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
6 Human Machine Interface

6.2.6 Submenu of “VERSION”

Purpose:

This menu is used to view the program version of this device.

Access Approach:

Move cursor to the item “VERSION” and press key “ENT” to show the program version after
entering the main menu of this device.

6.3 Understand the LCD Display

6.3.1 Default Display under Normal Operation Condition


The default display on the LCD is shown as below when the device is in normal situation.

1 09- 28 08:32:45
2 NR
3 PCS-9830A

Figure 6.3-1 Default display in normal situation

A brief explanation is made in the following table.

No. Display Explanation


1 09-28 08:32:45 The real time of this device: mm-dd hh:mm:ss.
2 NR Manufacturer name
3 PCS-9830A Device model

6.3.2 Display Operation Report


If there is any element operating, a brief operating report will appear on the LCD. The format of the
fault report is shown as below. In case more than one element has operated, the relevant reports
will be displayed alternately on the LCD.

1 12-09-23 NO.001
2 14:25:54:299.7
3 OPR time067.9ms
4 Op_ClsA

Figure 6.3-2 LCD display of the operation report

A brief explanation is made in the following table.

No. Display Explanation

PCS-9830A Phase Controlled Switching Device 6-5


Date: 2012-09-20
6 Human Machine Interface

1 12-09-28 NO.001 The report date: yy-mm-dd; report sequence


2 14:25:54:299.7 The report time: hh-mm-ss-msms
3 OPR time067.9ms The operated time
4 Op_ClsA The operated element.

The tripping report will keep being displayed on LCD until an acknowledgement by pressing the
key “ENT”.

Operated elements listed below may be displayed.

No. Operated Element Description


1 Op_ClsA Phase A closing operates.
2 Op_ClsB Phase B closing operates.
3 Op_ClsC Phase C closing operates.
4 Op_OpnA Phase A opening operates.
5 Op_OpnB Phase B opening operates.
6 Op_OpnC Phase C opening operates.
7 Op_Done_ClsA Phase A closing has done.
8 Op_Done_ClsB Phase B closing has done.
9 Op_Done_ClsC Phase C closing has done.
10 Op_Done_OpnA Phase A opening has done.
11 Op_Done_OpnB Phase B opening has done.
12 Op_Done_OpnC Phase C opening has done.

6.3.3 Display Alarm Report


If there is any abnormality in the operating or firmware error detected by equipment self-diagnostic,
alarm report will be displayed instantaneously on the LCD. Therefore, the default display will be
replaced by the alarm report. In case there is more than one alarm has occurred, the information
will be displayed alternately on the LCD.

1 09- 28 08:38:13
2 Alm_Settings

Figure 6.3-3 LCD display of the alarm report

A brief explanation is made in the following table.

No. Display Explanation


1 09-28 08:33:54 The real time of this device: mm-dd hh:mm:ss.
2 Alm_Settings The abbreviation of the alarm event.

Alarm elements listed below may be displayed.

No. Alarm Information Description

6-6 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
6 Human Machine Interface

1 Alm_Settings There is a mistake in the settings.


2 Alm_ModSetting The settings have been modified.
3 Alm_CompLogSet The logical setting of adaptive control is abnormal.
4 Alm_ClsTime_3P Three phases of closing times are abnormal.
5 Alm_Sample_Volt Sampled voltage is abnormal.
6 Alm_DS_BB The position of disconnection switch of busbar is abnormal.
7 Alm_ROM The content of the ROM is broken.
8 Alm_Upgraded The program of the device is upgraded.
9 Alm_I2C The I2C bus is abnormal.
10 Alm_CTS The current transformer circuit is abnormal.
11 Alm_CCS The closing circuit is broken.
12 Alm_TCS The tripping circuit is broken.
13 Alm_Uncharged The circuit breaker is uncharged.
14 Alm_SPI1 The SPI1 interface is abnormal.
15 Alm_SPI2 The SPI2 interface is abnormal.
16 Alm_AD The A/D converter is abnormal.
17 Alm_Recovered The device is recovered

6.3.4 Display Binary Input Report


If there is any change in the binary inputs detected by equipment, which will be recorded and
flashed instantaneously on the LCD or via the menu command “BI REPORT”. The default display
will be replaced by the BI report. In case there is more than one BI change has occurred, the
information will be displayed alternately on the LCD.

1 09-28 NO.001
2 14:11:29:986.4
3 BI_52a_A 0-1

Figure 6.3-4 LCD display of the alarm report

A brief explanation is made in the following table.

No. Display Explanation


1 09-28 NO.001 The real time of this device: mm-dd; report sequence
2 14:11:29:986.4 The real time of this device: hh:mm:ss:msms
3 BI_52a_A 0-1 The BI change report.

BI reports listed below may be displayed.

No. Alarm Information Description


1 BI_ClsCmd Closing command binary input change report
2 BI_OpnCmd Opening command binary input change report
3 BI_52a_A Phase A closing position binary input change report

PCS-9830A Phase Controlled Switching Device 6-7


Date: 2012-09-20
6 Human Machine Interface

4 BI_52a_B Phase B closing position binary input change report


5 BI_52a_C Phase C closing position binary input change report
6 BI_52b_A Phase A opening position binary input change report
7 BI_52b_B Phase B opening position binary input change report
8 BI_52b_C Phase C opening position binary input change report

6.4 View the Settings

Here take viewing the circuit breaker settings as an example to introduce the operating steps of
viewing setting for operators.

Operating steps:

1. Press key “ENT” to enter the main menu.

2. Turn key “MENU” to locate the cursor on the submenu “SETTINGS” and press key “ENT” to
enter this submenu.

3. Turn key “MENU” to locate the cursor on the submenu “FUN SETTINGS” and press key
“ENT” to enter this submenu.

4. Turn key “MENU” to select one of the enabled elements (take “CB SETTINGS” as an
example), and press key “ENT” to show the detailed settings of the selected element.

5. Turn key “MENU” to view the expected settings.

1 t_ClsCB_A
2 80.000 ms
3 t_ClsCB_B
4 80.00 ms

Figure 6.4-1 LCD display of the settings

A brief description about the setting interface is made in the following table.

No. Display Explanation


1 t_ClsCB_A The setting item name.
2 80.00 ms The setting’s value and unit.
3 t_ClsCB_B The setting item name.
4 80.00 ms The setting’s value and unit.

6.5 View Device Status

6.5.1 Display Analogue Data

Measured values mean AC sampled data in the submenu “MEASUREMENT” Take viewing AC

6-8 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
6 Human Machine Interface

sampled data of “MEASUREMENT” as an example.

Operating steps:

1. Press key “ENT” to enter the main menu.

2. Turn key “MENU” to locate the cursor on the submenu “VALUES” and press key “ENT” to
enter this submenu.

3. Turn key “MENU” to locate the cursor on the submenu “MEASUREMENT” and press key
“ENT” to enter this submenu.

4. Turn key “MENU” to view the expected measured values.

1 Ua = 057.6 V
2 Ub = 057.7 V
3 Uc = 057.7 V
4 Ia = 002.31 A

Figure 6.5-1 LCD display of the measured values

A brief description about the measured value interface is made in the following table.

No. Display Explanation


1 Ua= 057.6 V Measured value name and unit.
2 Ub= 057.7 V Measured value name and unit.
3 Uc= 057.7 V Measured value name and unit.
4 Ia= 002.31A Measured value name and unit.

Measurement data are listed below.

No. Symbol Description


1 Ua The phase A voltage
2 Ub The phase B voltage
3 Uc The phase C voltage
4 Ia The phase A current
5 Ib The phase B current
6 Ic The phase C current
7 f The system frequency

6.5.2 Display the Command output Status

Operating steps:

1. Press key “ENT” to enter the main menu.

2. Turn key “MENU” to locate the cursor on the submenu “VALUES” and press key “ENT” to
enter this submenu.

3. Turn key “MENU” to locate the cursor on the submenu “CMD STATE” and press key “ENT” to

PCS-9830A Phase Controlled Switching Device 6-9


Date: 2012-09-20
6 Human Machine Interface

enter this submenu.

4. Turn key “MENU” to view the expected device operation state.

1 Cmd_ClsA :0
2 Cmd_ClsB :0
3 Cmd_ClsC :0
4 Cmd_OpnA :0

Figure 6.5-2 LCD display of the device operation state

A brief description about the device operation state interface is made in the following table.

No. Display Explanation


1 Cmd_ClsA: 0 Device command output state name and its present state.
2 Cmd_ClsB: 0 Device command output state name and its present state.
3 Cmd_ClsC: 0 Device command output state name and its present state.
4 Cmd_OpnA: 0 Device command output state name and its present state.

Refer to the following table to see more information about the device operation state.

No. Symbol Description


1 Cmd_ClsA Phase A closing command output
2 Cmd_ClsB Phase B closing command output
3 Cmd_ClsC Phase C closing command output
4 Cmd_OpnA Phase A opening command output
5 Cmd_OpnB Phase B opening command output
6 Cmd_OpnC Phase C opening command output

6.5.3 Display the Operation Status


Operating steps:

1. Press key “ENT” to enter the main menu.

2. Turn key “MENU” to locate the cursor on the submenu “VALUES” and press key “ENT” to
enter this submenu.

3. Turn key “MENU” to locate the cursor on the submenu “OPR STATE” and press key “ENT” to
enter this submenu.

4. Turn key “MENU” to view the expected element state.

1 HiPrecAD :1
2 ExtIO :0
3 PwrOn :1

Figure 6.5-3 LCD display of the element state

6-10 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
6 Human Machine Interface

A brief description about the element state interface is made in the following table.

No. Display Explanation


1 HiPrecAD : 1 Device element name and state.
2 ExtIO :0 Device element name and state.
3 PwrOn :0 Device element name and state.

Refer to the following table to see more information about the elements.

No. Symbol Description


1 HiPrecAD AD with high precision
2 ExtIO Whether the external IO is fixed.
3 PwrOn Whether this device is power on.

6.5.4 Display the Status of Binary Inputs

Operating steps:

1. Press key “ENT” to enter the main menu.

2. Turn key “MENU” to locate the cursor on the submenu “VALUES” and press key “ENT” to
enter this submenu.

3. Turn key “MENU” to locate the cursor on the submenu “BI STATE” and press key “ENT” to
enter this submenu.

4. Turn key “MENU” to view the expected binary input state.

1 BI_ClsCmd :0
2 BI_OpnCmd :0
3 BI_52a_A :0
4 BI_52a_B :0

Figure 6.5-4 LCD display of the binary input state

A brief description about the binary input state interface is made in the following table.

No. Display Explanation


1 BI_ClsCmd :1 Binary input name and state.
2 BI_OpnCmd :0 Binary input name and state.
3 BI_52a_A :0 Binary input name and state.
4 BI_52a_B :0 Binary input name and state.

Refer to the following table to see more information about the binary inputs.

No. Symbol Description


1 BI_ClsCmd The binary input of closing command
2 BI_OpnCmd The binary input of opening command
3 BI_52a_A The binary input of phase A closing position

PCS-9830A Phase Controlled Switching Device 6-11


Date: 2012-09-20
6 Human Machine Interface

4 BI_52a_B The binary input of phase B closing position


5 BI_52a_C The binary input of phase C closing position
6 BI_52b_A The binary input of phase A opening position
7 BI_52b_B The binary input of phase B opening position
8 BI_52b_C The binary input of phase C opening position
9~16 BI_Resv1~BI_Resv8 The reserved binary input 1~8

6.6 View Software Version

By viewing the software version, the program version of this device can be known.

Operating steps:

1. Press key “ENT” to enter the main menu.

2. Turn key “MENU” to locate the cursor on the submenu “VERSION”.

3. Press key “ENT” to enter this submenu to view the program version.

1 PCS- 9830A_R1.00
2 CRC : EB4A26D6
3 SUBQ00114826
4 12-09- 20 11:22

Figure 6.6-1 LCD display of the software version

A brief description about the software version interface is made in the following table.

No. Display Explanation


1 PCS-9830A_R1.00 Program name and its version.
2 CRC: EB4A26D6 Program check code.
3 SUBQ00114826 Program development number.
4 12-09-20 11:22 Program creation date and time: yy-mm-dd hh:mm.

NOTE: It is only a sample which is used to explain the indication of the software version

menu. The software version of the device of the practical engineering should be taken
as final and binding.

6.7 View History Reports

This device can record and store the history reports which include the device operation reports,
the self supervision reports and the binary input change reports. It provides convenience for
viewing the history reports at any time.

6-12 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
6 Human Machine Interface

Operating steps

1. Press key “ENT” to enter the main menu.

2. Turn key “MENU” to locate the cursor on the submenu “REPORT” and press key “ENT” to
enter this submenu.

3. Turn key “MENU” to locate the cursor on the expected submenu (“OPR REPORT”, “ALM
REPORT”, etc.) and press key “ENT” to enter this selected submenu.

4. Turn key “MENU” to view the expected history reports in the selected submenu. If there is no
report in the selected submenu, it will show “No Report” on the LCD.

6.8 Operation through Keypad

6.8.1 Password Protection


For the safety purpose, this device provides password protection function for viewing the relevant
menus, deleting the history reports and modifying the settings.

The password is 3-digit number, and its default value is “111”. It can be set through the setting
[Password] in submenu “EQUIP SETUP”. If it is set as “000”, the operation with password does not
need the password, and the password input interface is hidden.

The password input interface is shown as below.

< Password >


0000

Figure 6.8-1 Password input interface

If the inputted password is correct, the relevant operation can be executed; otherwise, it will show
“Invalid Password” on the LCD and prompt the user to input the password again.

6.8.2 Clock Set

Operating steps

1. Press key “ENT” to enter the main menu.

2. Turn key “MENU” to locate the cursor on the submenu “CLOCK” and press key “ENT” to enter
this submenu.

3. Turn key “MENU” to locate the cursor on the expected digital place, and then press key
“ADJUST” to change the number to the expected one.

4. After finishing the clock modification, press key “ENT” and input the Level 1 password to
confirm the modification.

PCS-9830A Phase Controlled Switching Device 6-13


Date: 2012-09-20
6 Human Machine Interface

The clock modification interface is shown as below.

Date: 008-09-28
Time: 14:45:07

Figure 6.8-2 Clock modification interface

The date format is yy-mm-dd, and the time format is hh:mm:ss.

6.8.3 Delete Records

Operating steps

1. Press key “ENT” to enter the main menu.

2. Turn key “MENU” to locate the cursor on the submenu “REPORT” and press key “ENT” to
enter this submenu.

3. Turn key “MENU” to locate the cursor on the submenu “CLR REPORT” press key “ENT” to
enter password input interface.

4. Input the correct password, and then press key “ENT” to delete all the history records.

6-14 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
7 Installation

7 Installation

Table of Contents
7 Installation....................................................................................... 7-a
7.1 Overview .......................................................................................................... 7-1
7.2 Safety Instructions .......................................................................................... 7-1
7.3 Checking the Shipment .................................................................................. 7-2
7.4 Material and Tools Required .......................................................................... 7-2
7.5 Device Location and Ambient Conditions .................................................... 7-2
7.6 Mechanical Installation ................................................................................... 7-3
7.7 Electrical Installation and Wiring ................................................................... 7-4
7.7.1 Grounding Guidelines .......................................................................................................... 7-4

7.7.2 Cubicle Grounding ............................................................................................................... 7-4

7.7.3 Ground Connection on the Device ...................................................................................... 7-5

7.7.4 Grounding Strips and their Installation ................................................................................ 7-6

7.7.5 Guidelines for Wiring ........................................................................................................... 7-6

7.7.6 Wiring for Electrical Cables ................................................................................................. 7-6

7.8 Typical Wiring of the Device........................................................................... 7-7

List of Figures
Figure 7.6-1 Dimensions of PCS-9830A and the cut-out in the cubicle (unit: mm).............. 7-3

Figure 7.6-2 Installation demonstration of this device ............................................................ 7-4

Figure 7.7-1 Cubicle grounding system.................................................................................... 7-5

Figure 7.7-2 Ground terminal of this device ............................................................................. 7-5

Figure 7.7-3 Ground strip and termination ............................................................................... 7-6

Figure 7.7-4 Glancing demo about the wiring for electrical cables ....................................... 7-7

Figure 7.8-1 Typical wiring diagram of this device .................................................................. 7-8

PCS-9830A Phase Controlled Switching Device 7-a


Date: 2012-09-20
7 Installation

7-b PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
7 Installation

7.1 Overview

The equipment must be shipped, stored and installed with the greatest care.

Choose the place of installation such that the communication interface and the controls on the
front of the device are easily accessible.

Air must circulate freely around the equipment. Observe all the requirements regarding place of
installation and ambient conditions given in this instruction manual.

Take care that the external wiring is properly brought into the equipment and terminated correctly
and pay special attention to grounding. Strictly observe the corresponding guidelines contained in
this section.

7.2 Safety Instructions

Modules and units may only be replaced by correspondingly trained personnel. Always observe
the basic precautions to avoid damage due to electrostatic discharge when handling the
equipment.

In certain cases, the settings have to be configured according to the demands of the engineering
configuration after replacement. It is therefore assumed that the personnel who replace modules
and units are familiar with the use of the operator program on the service PC.

DANGER: Only insert or withdraw the PWR module while the power supply is switched

off. To this end, disconnect the power supply cable that connects with the PWR module.

WARNING: Only insert or withdraw the other boards while the power supply is switched

off.

WARNING: The modules of this device may only be inserted in the slots designated in

Chapter 5. Components can be damaged or destroyed by inserting boards in the wrong


slots.

DANGER: Improper handling of the equipment can cause damage or an incorrect

response of the equipment itself or the primary plant.

WARNING: Industry packs and ribbon cables may only be replaced or the positions of

jumpers be changed on a workbench appropriately designed for working on electronic


equipment. The modules and bus backplanes of this device are sensitive to electrostatic
discharge when not in the unit's housing.

The basic precautions to guard against electrostatic discharge are as follows:

PCS-9830A Phase Controlled Switching Device 7-1


Date: 2012-09-20
7 Installation

 Should boards have to be removed from this device installed in a grounded cubicle in an HV
switchgear installation, please discharge yourself by touching station ground (the cubicle)
beforehand.

 Only hold electronic boards at the edges, taking care not to touch the components.

 Only works on boards that have been removed from the cubicle on a workbench designed for
electronic equipment and wear a grounded wristband. Do not wear a grounded wristband,
however, while inserting or withdrawing units.

 Always store and ship the electronic boards in their original packing. Place electronic parts in
electrostatic screened packing materials.

7.3 Checking the Shipment

Check that the consignment is complete immediately upon receipt. Notify the nearest NR
Company or agent, should departures from the delivery note, the shipping papers or the order be
found.

Visually inspect all the material when unpacking it. When there is evidence of transport damage,
lodge a claim immediately in writing with the last carrier and notify the nearest NR Company or
agent.

If the equipment is not going to be installed immediately, store all the parts in their original packing
in a clean dry place at a moderate temperature. The humidity at a maximum temperature and the
permissible storage temperature range in dry air are listed in Chapter 2.

7.4 Material and Tools Required

The necessary mounting kits will be provided, including screws, pincers and assembly
instructions.

A suitable drill and spanners are required to secure the cubicles to the floor using the plugs
provided (if this device is mounted in cubicles).

7.5 Device Location and Ambient Conditions

The place of installation should permit easy access especially to front of the device, i.e. to the
human-machine interface of the equipment.

There should also be free access at the rear of the equipment for additions and replacement of
electronic boards.

Since every piece of technical equipment can be damaged or destroyed by inadmissible ambient
conditions, such as:

1) The location should not be exposed to excessive air pollution (dust, aggressive substances).

7-2 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
7 Installation

2) Severe vibration, extreme changes of temperature, high levels of humidity, surge voltages of
high amplitude and short rise time and strong induced magnetic fields should be avoided as
far as possible.

3) Air must not be allowed to circulate freely around the equipment.

The equipment can in principle be mounted in any attitude, but it is normally mounted vertically
(visibility of markings).

WARNING: Excessively high temperature can appreciably reduce the operating life of

this device.

7.6 Mechanical Installation

This device is made of a single layer 4U height 6” chassis with 4 connectors on its rear panel (See
Figure 5.1-2). Following figure shows the dimensions of this device for reference in mounting.

165.31
188.92 149.90

PCS-9830A
PHASE CONTROLLED SWITCHING DEVICE

HEALTHY
EN
U
M

177.00

ALARM ENT

DJUS
T
A

EXT

RS-232

173.00

AI CPU BI BO PWR

01 02

03 04

05 06

07 08
178. 00

09 10

11 12

13 14

15 16

17 18

19 20

21 22

23 24

Cut-out in the Cubicle

Figure 7.6-1 Dimensions of PCS-9830A and the cut-out in the cubicle (unit: mm)

NOTE: It is necessary to leave enough space top and bottom of the cut-out in the cubicle

for heat emission of this device.

As mentioned former (Chapter 5), up to five modules are installed in the enclosure of this device,
and these modules must be plugged into the proper slots of this device respectively. The safety
instructions must be abided by when installing the boards, see Section 7.2.

PCS-9830A Phase Controlled Switching Device 7-3


Date: 2012-09-20
7 Installation

In the case of equipment supplied in cubicles, place the cubicles on the foundations that have
been prepared. Take care while doing so not to jam or otherwise damage any of the cables that
have already been installed. Secure the cubicles to the foundations.

The following shows the installation demonstration of this device.

Figure 7.6-2 Installation demonstration of this device

7.7 Electrical Installation and Wiring

7.7.1 Grounding Guidelines


Switching operations in HV installations generate transient over voltages on control signal cables.
There is also a background of electromagnetic RF fields in electrical installations that can induce
spurious currents in the devices themselves or the leads connected to them.

All these influences can influence the operation of electronic apparatus.

On the other hand, electronic apparatus can transmit interference that can disrupt the operation of
other apparatus.

In order to minimize these influences as far as possible, certain standards have to be observed
with respect to grounding, wiring and screening.

NOTE: All these precautions can only be effective if the station ground is of good quality.

7.7.2 Cubicle Grounding


The cubicle must be designed and fitted out such that the impedance for RF interference of the
ground path from the electronic device to the cubicle ground terminal is as low as possible.

Metal accessories such as side plates, blanking plates etc., must be effectively connected
surface-to-surface to the grounded frame to ensure a low-impedance path to ground for RF
interference. The contact surfaces must not only conduct well, they must also be non-corroding.

NOTE: If the above conditions are not fulfilled, there is a possibility of the cubicle or parts

of it forming a resonant circuit at certain frequencies that would amplify the transmission

7-4 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
7 Installation

of interference by the devices installed and also reduce their immunity to induced
interference.

Movable parts of the cubicle such as doors (front and back) or hinged equipment frames must be
effectively grounded to the frame by three braided copper strips (see Figure 7.7-1).

The metal parts of the cubicle housing and the ground rail are interconnected electrically
conducting and corrosion proof. The contact surfaces shall be as large as possible.

NOTE: For metallic connections please observe the voltage difference of both materials

according to the electrochemical code.

The cubicle ground rail must be effectively connected to the station ground rail by a grounding strip
(braided copper).

Door or hinged
equipment frame

Cubicle ground
rail close to floor

Braided
copper strip
Station
ground

Conducting
connection

Figure 7.7-1 Cubicle grounding system

7.7.3 Ground Connection on the Device


There is a ground terminal on the rear panel (see Figure 7.7-2), and the ground braided copper
strip can be connected with it. Take care that the grounding strip is always as short as possible.
The main thing is that the device is only grounded at one point. Grounding loops from unit to unit
are not allowed.

There are some ground terminals on some connectors of this device, and the sign is “GND”. All the
ground terminals are connected in the cabinet of this device. So, the ground terminal on the rear
panel (see Figure 7.7-2) is the only ground terminal of this device.

Figure 7.7-2 Ground terminal of this device

PCS-9830A Phase Controlled Switching Device 7-5


Date: 2012-09-20
7 Installation

7.7.4 Grounding Strips and their Installation


High frequency currents are produced by interference in the ground connections and because of
skin effect at these frequencies, only the surface region of the grounding strips is of consequence.

The grounding strips must therefore be of (preferably tinned) braided copper and not round copper
conductors, as the cross-section of round copper would have to be too large.

Data of braided copper strip: threaded M4, 4.0mm2.

Proper terminations must be fitted to both ends (press/pinch fit and tinned) with a hole for bolting
them firmly to the items to be connected.

The surfaces to which the grounding strips are bolted must be electrically conducting and
non-corroding.

The following figure shows the ground strip and termination.

Press/pinch fit
cable terminal

Braided
copper strip Terminal bolt

Contact surface

Figure 7.7-3 Ground strip and termination

7.7.5 Guidelines for Wiring


There are several types of cables that are used in the connection of this device: braided copper
cable, serial communication cable etc. Recommendation of each cable:

 Grounding: braided copper cable, 2.5mm2 ~ 6.0mm2

 Power supply, binary inputs & outputs: brained copper cable, 1.0mm2 ~ 2.5mm2

 AC voltage inputs: brained copper cable, 1.0mm2 ~ 2.5mm2

 AC current inputs: brained copper cable, 1.5mm2 ~ 4.0mm2

7.7.6 Wiring for Electrical Cables


A female connector is used for connecting the wires with it, and then a female connector plugs into
a corresponding male connector that is in the front of one board. For further details about the pin
defines of these connectors, see Chapter 5.

The following figure shows the glancing demo about the wiring for the electrical cables.

7-6 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
7 Installation

01 02

03 04

Tighten 05 06

07 08

09 10

11 12
01

13 14

15 16

17 18

19 20

21 22

23 24

Figure 7.7-4 Glancing demo about the wiring for electrical cables

WARNING: Never allow the current transformer (CT) secondary circuit connected to this

equipment to be opened while the primary system is live. Opening the CT circuit will
produce a dangerously high voltage.

7.8 Typical Wiring of the Device

Relevant information and sections about the modules and the connectors of this device are
described in Chapter 5. Referring the relevant sections can help to wire correctly and effectively.

 The power supply can be DC or AC power supply input. In Figure 7.8-1, if the DC power
supply is applied, the “A~” is the positive input (PS+), and the “N~” is the negative input (PS-).

 The CT currents of circuit breakers connect to the pin 601~606 on the connector of the AI
module. The three phase rated voltages are from the PT of CB near to power supply, then
connect to the pin 617~622 in AI module. And the voltages are used as the criterion of
zero-crossing point in the CPU module.

 The closing command [BI_ClsCmd] and opening command [BI_OpnCmd] are connected to
the pin 402, 403 in CPU module. After the process-time, each phase controlled closing
commands are output via the contacts [Cmd_ClsA], [Cmd_ClsB] and [Cmd_ClsC] to the
closing coils respectively, and each phase controlled opening commands are output via the
contacts [Cmd_OpnA], [Cmd_OpnB] and [Cmd_OpnC] to the opening coils respectively.

 The CPU module gets the closing positions of circuits breakers via the pin 404~406, and the
opening positions of circuit breakers via the pin 407~409.

The typical wiring of this device is shown as below.

PCS-9830A Phase Controlled Switching Device 7-7


Date: 2012-09-20
7 Installation

Metering current Ia, Ib, Ic Metering voltage Ua, Ub, Uc


PS+ PS -

6 6 6 6 6 6 6 6 6 6 6 6
0 0 0 0 0 0 1 1 1 2 2 2
1 2 3 4 5 6 7 8 9 0 1 2

Ia Ib Ic Ua Ub Uc

101 103
PWR
102 Alarm 104
201
Cmd_ClsA
401 BI_COM1 202
402 BI_ClsCmd 205
Cmd_ClsB
403 BI_OpnCmd 206
404 BI_52a_A 209
Cmd_ClsC
405 BI_52a_B 210
406 BI_52a_C
_ 213
Cmd_OpnA
407 BI_52b_A
_ _ 214
408 BI_52b_B_ 217
Cmd_OpnB
409 BI_52b_C 218
221
Cmd_OpnC
222

Figure 7.8-1 Typical wiring diagram of this device

7-8 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
8 Commissioning

8 Commissioning

Table of Contents
8 Commissioning ............................................................................... 8-a
8.1 Overview .......................................................................................................... 8-1
8.2 Safety Instructions .......................................................................................... 8-1
8.3 Commission Tools .......................................................................................... 8-2
8.4 Setting Familiarization .................................................................................... 8-2
8.5 Product Checks ............................................................................................... 8-3
8.5.1 With the Device De-energized ............................................................................................. 8-3

8.5.2 With the Device Energized .................................................................................................. 8-5

8.5.3 On-load Checks ................................................................................................................... 8-7

8.5.4 Final Checks ........................................................................................................................ 8-7

PCS-9830A Phase Controlled Switching Device 8-a


Date: 2012-09-20
8 Commissioning

8-b PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
8 Commissioning

8.1 Overview
This device is fully numerical in their design, implementing all the functions in software. The device
employ a high degree of self-checking and in the unlikely event of a failure, will give an alarm.

To commission numerical devices, it is only necessary to verify that the hardware is functioning
correctly and the application-specific software settings have been applied to the device.

Blank commissioning test and setting records are provided at the end of this manual for
completion as required.

Before carrying out any work on the equipment, the user should be familiar with the contents of the
safety and technical data sections and the ratings on the equipment’s rating label.

8.2 Safety Instructions

WARNING: Hazardous voltages are present in this electrical equipment during operation.

Non-observance of the safety rules can result in severe personal injury or property
damage.

WARNING: Only qualified personnel shall work on and around this equipment after

becoming thoroughly familiar with all warnings and safety notices of this manual as well
as with the applicable safety regulations.

Particular attention must be drawn to the following:

 The earthing screw of the device must be connected solidly to the protective earth conductor
before any other electrical connection is made.

 Hazardous voltages can be present on all circuits and components connected to the supply
voltage or to the measuring and test quantities.

 Hazardous voltages can be present in the device even after disconnection of the supply
voltage (storage capacitors!)

 The limit values stated in the technical data (Chapter 2) must not be exceeded at all, not even
during testing and commissioning.

 When testing the device with secondary test equipment, make sure that no other
measurement quantities are connected. Take also into consideration that the trip circuits and
maybe also close commands to the circuit breakers and other primary switches are
disconnected from the device unless expressly stated.

DANGER: Current transformer secondary circuits must have been short-circuited before

the current leads to the device are disconnected.

PCS-9830A Phase Controlled Switching Device 8-1


Date: 2012-09-20
8 Commissioning

WARNING: Primary test may only be carried out by qualified personnel, who are familiar

with the commissioning of power system, the operation of the plant and safety rules and
regulations (switching, earthing, etc.).

8.3 Commission Tools

Minimum equipment required:

 Multifunctional dynamic current and voltage injection test set with interval timer.

 Multimeter with suitable AC current range and AC/DC voltage ranges of 0 ~ 500V and 0 ~
250V respectively.

 Continuity tester (if not included in the multimeter).

 Phase angle meter.

 Phase rotation meter.

NOTE: Modern test set may contain many of the above features in one unit.

Optional equipment:

 An electronic or brushless insulation tester with a DC output not exceeding 500V (for
insulation resistance test when required).

 A portable PC, with appropriate software Dbg2000 (this enables the rear communications port
to be tested, if this is to be used, and will also save considerable time during commissioning).

8.4 Setting Familiarization

When commissioning this device for the first time, sufficient time should be allowed to become
familiar with the method by which the settings are applied. The Chapter 6 contains a detailed
description of the menu structure of this device.

With the front cover in place all keys are accessible. All menu cells can be read. LED indicators
and alarms can be reset. Fault and event records can be cleared. However, menu cells will require
the appropriate password to be entered before changes can be made.

Alternatively, if a portable PC is available together with suitable setting software “Dbg2000-96.exe”,


the menu can be viewed a page at a time to display a full column of data and text. This PC
software also allows settings to be entered more easily, saved to a file on disk for future reference
or printed to produce a setting record. Refer to the “Dbg2000-96.exe” software user manual for
details. If the software is being used for the first time, allow sufficient time to become familiar with
its operation.

8-2 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
8 Commissioning

8.5 Product Checks

These product checks cover all aspects of the device which should be checked to ensure that it
has not been physically damaged prior to commissioning, is functioning correctly and all input
quantity measurements are within the stated tolerances.

If the application-specific settings have been applied to the device prior to commissioning, it is
advisable to make a copy of the settings so as to allow them restoration later. This could be done
by extracting the settings from the device itself via printer or manually creating a setting record.

8.5.1 With the Device De-energized


This device is fully numerical and the hardware is continuously monitored. Commissioning tests
can be kept to a minimum and need only include hardware tests and conjunctive tests. The
function tests are carried out according to user’s correlative regulations.

The following tests are necessary to ensure the normal operation of the equipment before it is first
put into service.

 Hardware tests

These tests are performed for the following hardware to ensure that there is no hardware
defect. Defects of hardware circuits other than the following can be detected by
self-monitoring when the DC power is supplied.

 User interfaces test

 Binary input circuits and output circuits test

 AC input circuits test

 Function tests

These tests are performed for the following functions that are fully software-based.

 Measuring elements test

 Timers test

 Metering and recording test

 Conjunctive tests

The tests are performed after the device is connected with the primary equipment and other
external equipment.

 On load test.

 Phase sequence check and polarity check.

8.5.1.1 Visual Inspection

After unpacking the product, check for any damage to the device case. If there is any damage, the

PCS-9830A Phase Controlled Switching Device 8-3


Date: 2012-09-20
8 Commissioning

internal module might also have been affected, contact the vendor. Following items listed is
necessary.

 Panel

Carefully examine the panel, equipment inside and other parts inside to see that no physical
damage has occurred since installation.

The rated information of other auxiliary devices should be checked to ensure it is correct for
the particular installation.

 Panel wiring

Check the conducting wire which is used in the panel to assure that their cross section
meeting the requirement.

Carefully examine the wiring to see that they are no connection failure exists.

 Label

Check all the isolator binary inputs, terminal blocks, indicators, switches and push buttons to
make sure that their labels meet the requirements of this project.

 Equipment plug-in modules

Check each plug-in module of the equipments on the panel to make sure that they are well
installed into the equipment without any screw loosened.

 Earthing cable

Check whether the earthing cable from the panel terminal block is safely screwed to the panel
steel sheet.

 Switch, keypad, isolator binary inputs and push button

Check whether all the switches, equipment keypad, isolator binary inputs and push buttons
work normally and smoothly.

8.5.1.2 Insulation Test (if required)

Insulation resistance tests are only necessary during commissioning if it is required for them to be
done and they have not been performed during installation.

Isolate all wiring from the earth and test the isolation with an electronic or brushless insulation
tester at a DC voltage not exceeding 500V, The circuits need to be tested should include:

 Current transformer circuits

 Power supply

 Optic-isolated control inputs

 Output contacts

 Communication ports

8-4 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
8 Commissioning

The insulation resistance should be greater than 100MΩ at 500V.

Test method:

To unplug all the terminals sockets of this device, and do the Insulation resistance test for each
circuit above with an electronic or brushless insulation tester.

On completion of the insulation resistance tests, ensure all external wiring is correctly reconnected
to the device.

8.5.1.3 External Wiring

Check that the external wiring is correct to the relevant device diagram and scheme diagram.
Ensure as far as practical that phasing/phase rotation appears to be as expected.

Check the wiring against the schematic diagram for the installation to ensure compliance with the
customer’s normal practice.

8.5.1.4 Auxiliary Supply

The device can be operated with the power supply voltage 30Vac, 110Vdc, 220Vdc, 110Vac or
220Vac auxiliary supply depending on the device’s nominal supply rating. The incoming voltage
must be within the operating range specified in Chapter 2 before energizing the device, measure
the auxiliary supply to ensure it within the operating range.

8.5.2 With the Device Energized


The following groups of checks verify that the device hardware and software is functioning
correctly and should be carried out with the auxiliary supply applied to the device.

8.5.2.1 Check Program Version

Connect the device to power supply correctly and turn the device on. Check program version and
forming time displayed in command menu to ensure that are corresponding to what ordered.

The liquid crystal display (LCD) is designed to operate in a wide range of substation ambient
temperatures. For this purpose, this device has an automatic “LCD contrast” adjusting feature,
which is capable to adjust LCD contrast automatically according to the ambient temperature.

8.5.2.2 Check Date and Time

If the time and date is not being maintained by substation automation system, the date and time
should be set manually.

Set the date and time to the correct local time and date using menu item “CLOCK”.

In the event of the auxiliary supply failing, with a battery fitted on CPU board, the time and date will
be maintained. Therefore when the auxiliary supply is restored the time and date will be correct
and not need to set again.

To test this, remove the auxiliary supply from the device for approximately 30s. After being
re-energized, the time and date should be correct.

PCS-9830A Phase Controlled Switching Device 8-5


Date: 2012-09-20
8 Commissioning

8.5.2.3 Check the LEDs

 “HEALTHY” LED

In normal operation situation, the “HEALTHY” LED is on. If a fatal error which is described in
Section 6.3.3 occurs in this device, the “HEALTHY” LED will be off.

 “ALARM” LED

In normal operation situation, the “ALARM” LED is off. If an error which is described in
Section 6.3.3 occurs in this device, the “ALARM” LED will be on. So simulate the abnormality
in Section 6.3.3 to check the “ALARM” LED.

 Other LEDs

The LEDs can be defined as various signal output indicators, so define all the LEDs as the
same function, and then simulate the relevant signal to check this LEDs.

8.5.2.4 Testing the AC Current Inputs

This test verified that the accuracy of current measurement is within the acceptable tolerances.

Apply current equal to the current transformer secondary winding rating to each current
transformer input of the corresponding rating in turn, see the following table or external connection
diagram for appropriate terminal numbers, checking its magnitude using a multimeter/test set
readout. The corresponding reading can then be checked in the devices menu.

The current measurement accuracy of the device is ±2.0% or 0.01In, whichever is greater.
However an additional allowance must be made for the accuracy of the test equipment being
used.

Current channel linearity and precision checkout

Item Input Current Measurement (on LCD) Error


Ia
Ib
Ic

8.5.2.5 Testing the Binary Inputs

This test checks that all the binary inputs on the device are functioning correctly.

The binary inputs should be energized one at a time, see external connection diagrams for
terminal numbers.

Ensure that the voltage applied on the binary input must be within the operating range.

The status of each binary input can be viewed using menu item “BI STATE”. Sign “1” denotes an
energized input and sign “0” denotes a de-energized input.

Binary Inputs testing checkout

8-6 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
8 Commissioning

Terminal No. Signal Name BI Status on LCD Correct?

8.5.3 On-load Checks


The objectives of the on-load checks are:

 Confirm the external wiring to the current inputs is correct.

 Measure the magnitude of on-load current (if applicable).

 Check the polarity of each current transformer.

However, these checks can only be carried out if there are no restrictions preventing the
tenderization of the plant being protected.

Remove all test leads, temporary shorting leads, etc. and replace any external wiring that has
been removed to allow testing.

If it has been necessary to disconnect any of the external wiring from the device in order to
perform any of the foregoing tests, it should be ensured that all connections are replaced in
accordance with the relevant external connection or scheme diagram. Confirm the current
transformer wiring.

8.5.4 Final Checks


After the above tests are completed, remove all test or temporary shorting leads, etc. If it has been
necessary to disconnect any of the external wiring from the device in order to perform the wiring
verification tests, it should be ensured that all connections are replaced in accordance with the
relevant external connection or scheme diagram.

Ensure that the device has been restored to service.

If the device is in a new installation or the circuit breaker has just been maintained, the circuit
breaker maintenance and current counters should be zero. If a test block is installed, remove the
test plug and replace the cover so that the device is put into service.

Ensure that all operation reports, BI change reports and alarms have been cleared and LED’s has
been reset before leaving the device.

PCS-9830A Phase Controlled Switching Device 8-7


Date: 2012-09-20
8 Commissioning

8-8 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
9 Maintenance

9 Maintenance

Table of Contents
9 Maintenance .................................................................................... 9-a
9.1 Regular Testing ............................................................................................... 9-1
9.2 Failure Tracing and Repair ............................................................................. 9-1
9.3 Replace Failed Modules ................................................................................. 9-1
9.4 Cleaning ........................................................................................................... 9-3
9.5 Storage ............................................................................................................. 9-3

PCS-9830A Phase Controlled Switching Device 9-a


Date: 2012-09-20
9 Maintenance

9-b PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
9 Maintenance

It is recommended that products supplied by NR receive periodic monitoring after installation. In


view of the critical nature of devices and their infrequent operation, it is desirable to confirm that
they are operating correctly at regular intervals.

The devices are self-supervised and so require less maintenance than earlier designs of device.
Most problems will result in an alarm so that remedial action can be taken. However, some
periodic tests should be done to ensure that the device is functioning correctly and the external
wiring is intact.

9.1 Regular Testing

The device is almost completely self-supervised. The circuits which can not be supervised are
binary input, output circuits and human machine interfaces. Therefore regular testing can be
minimized to checking the unsupervised circuits.

9.2 Failure Tracing and Repair


Failures will be detected by automatic supervision or regular testing.

When a failure is detected by supervision, a remote alarm is issued and the failure is indicated on
the front panel with LED indicators and LCD display. It is also recorded in the event record.
Failures detected by supervision are traced by checking the “Superv Events” screen on the LCD.

When a failure is detected during regular testing, confirm the following:

 Test circuit connections are correct

 Modules are securely inserted in position

 Correct DC power voltage is applied

 Correct AC inputs are applied

 Test procedures comply with those stated in the manual

9.3 Replace Failed Modules


If the failure is identified to be in the device module and the user has spare modules, the user can
recover the device by replacing the failed modules.

Repair at the site should be limited to module replacement. Maintenance at the component level is
not recommended.

Check that the replacement module has an identical module name (AI, PWR, CPU, SIG, BI, BO,
etc.) and hardware type-form as the removed module. Furthermore, the CPU module replaced
should have the same software version. In addition, the AI and PWR module replaced should have
the same ratings.

The module name is indicated on the top front of the module. The software version is indicated in
LCD menu “Version Info”.

PCS-9830A Phase Controlled Switching Device 9-1


Date: 2012-09-20
9 Maintenance

Caution!

When handling a module, take anti-static measures such as wearing an earthed wrist band
and placing modules on an earthed conductive mat. Otherwise, many of the electronic
components could suffer damage. After replacing the CPU module, check the settings.

1) Replacing a module

 Switch off the DC power supply

 Disconnect the trip outputs

 Short circuit all AC current inputs and disconnect all AC voltage inputs

 Unscrew the module.

Warning!

Hazardous voltage can be present in the DC circuit just after switching off the DC power
supply. It takes approximately 30 seconds for the voltage to discharge.

2) Replacing the Human Machine Interface Module (front panel)

 Open the device front panel

 Unplug the ribbon cable on the front panel by pushing the catch outside.

 Detach the HMI module from the device

 Attach the replacement module in the reverse procedure.

3) Replacing the AI, PWR, CPU, BI or BO module

 Unscrew the module connector

 Unplug the connector from the target module.

 Unscrew the module.

 Pull out the module

 Inset the replacement module in the reverser procedure.

 After replacing the CPU module, input the application-specific setting values again.

Warning!

Units and modules may only be replaced while the supply is switched off and only by
appropriately trained and qualified personnel. Strictly observe the basic precautions to
guard against electrostatic discharge.

Warning!

9-2 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
9 Maintenance

When handling a module, take anti-static measures such as wearing an earthed wrist band
and placing modules on an earthed conductive mat. Otherwise, many of the electronic
components could suffer damage. After replacing the CPU module, check the settings.

Danger!

After replacing modules, be sure to check that the same configuration is set as before the
replacement. If this is not the case, there is a danger of the unintended operation taking
place. Persons may also be put in danger.

9.4 Cleaning
Before cleaning the device, ensure that all AC/DC supplies, current transformer connections are
isolated to prevent any chance of an electric shock whilst cleaning. Use a smooth cloth to clean
the front panel. Do not use abrasive material or detergent chemicals.

9.5 Storage
The spare device or module should be stored in a dry and clean room. Based on IEC standard
60255-1 the storage temperature should be from -40oC to +70oC, but the temperature of from
-10oC to +40oC is recommended for long-term storage.

PCS-9830A Phase Controlled Switching Device 9-3


Date: 2012-09-20
9 Maintenance

9-4 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
10 Decommissioning and Disposal

10 Decommissioning and Disposal

Table of Contents
10 Decommissioning and Disposal ................................................ 10-a
10.1 Decommissioning ....................................................................................... 10-1
10.2 Disposal ....................................................................................................... 10-1

PCS-9830A Phase Controlled Switching Device 10-a


Date: 2012-09-20
10 Decommissioning and Disposal

10-b PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
10 Decommissioning and Disposal

10.1 Decommissioning
1. Switching off

To switch off the PCS-9830A, switch off the external miniature circuit breaker of the power supply.

2. Disconnecting Cables

Disconnect the cables in accordance with the rules and recommendations made by relational
department.

Danger!

Before disconnecting the power supply cables that connected with the PWR module of the
PCS-9830A, make sure that the external miniature circuit breaker of the power supply is
switched off.

Danger!

Before disconnecting the cables that are used to connect analog input module with the
primary CTs and VTs, make sure that the circuit breaker for the primary CTs and VTs is
switched off.

3. Dismantling

The PCS-9830A rack may now be removed from the system cubicle, after which the cubicles may
also be removed.

Danger!

When the station is in operation, make sure that there is an adequate safety distance to
live parts, especially as dismantling is often performed by unskilled personnel.

10.2 Disposal
In every country there are companies specialized in the proper disposal of electronic waste.

Note!

Strictly observe all local and national regulations when disposing of the device.

PCS-9830A Phase Controlled Switching Device 10-1


Date: 2012-09-20
10 Decommissioning and Disposal

10-2 PCS-9830A Phase Controlled Switching Device


Date: 2012-09-20
11 Manual Version History

11 Manual Version History


In the current version of the instruction manual, several descriptions on existing features have
been modified.

Manual version and modification history records

Version
Date Description of change
Software Manual
1.00 1.00 2012-09-20 Form the original manual.

You might also like