Vision Turbine Meters

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Badger Meter Europa GmbH

Nürtinger Str. 76
72639 Neuffen (Germany)
Tel. +49-7025-9208-0
Fax +49-7025-9208-15
www.badgermeter.de
[email protected]

Turbine meters of the Vision series


For low viscous and non-aggressive liquids
Features
• Excellent price – performance ratio
• Small size
• Easy and fast installation
• Maintenance free
• High operating pressure
• Installation in any position

Description Applications
The liquid turbine meters of the series Vision are for the For flow measurement of water, demineralized water,
exact measuring of small quantities of liquids. The actual alkaline solutions, oils/edible oils, fuel/fuel consumption,
flow as well as the already passed flow can be measured. fuel oil, beverage, water based liquids, etc.

The large number of pulses gives a good resolution. As the Special for washmaschines, dish washers, coffee
mass of the turbine are very small the response time is machines, laser cooling plants, solar solutions, backery
very short. It is not necessary to install a straight length of machines, steam cooking machines in large kitchen plants,
pipeline at the upstream side. The simple mechanical dialysis equipment, CD cleaning, irrigation plants, etc.
construction of the sensor Vision guarantees a long lifespan
without any loss of accuracy. Pressure surges do not affect
the measuring system. Electrical connections

Measuring principle
The rotor is turned by the fluid force in proportional to the
flow. A hall generator supplies pulses, which can be util-
ized for digital or analogue signal processing. The gener- with 3 flat plugs with AMP Faston connection
ated pulses are specified as the k-factor. 2,8 x 0,5 mm 2,8/6,3 x 0,8 mm

with coupler socket


Micro-EN 60529 with cable connection

TUR_VIS_BR_DB_02_1704.doc
Reprint of texts or text extracts requires prior written authorization of Badger Meter Europa GmbH.
Misuse of texts, pictures or company logo will be prosecuted .
Technical data
Type 1005 2F66 2006 2F66 2006 4F44 2008 2F66
Material Trogamid
Connection* G 1/4" G 3/8"
Order code (with coupler
56547-163-2F66 56510-163-2F66 56510-163-4F44 56500-163-2F66
socket acc. EN 60529)
Order code for AMP Faston - 56510-164-2F66 56510-164-4F44 56500-164-2F66
Order code for 1m-cable 56547-165-2F66-1 56510-165-2F66-1 56510-165-4F44-1 56500-165-2F66-1
Measuring rate l/min 0,1 – 2,5 0,5 - 5 1 – 10 0,5 – 7,5
K-factor PPL 22.000 6900 3300 4700
DN mm 5 6 6 8
Type 2008 4F44 2008 4F22 2008 4F17 3012 4F16
Material Trogamid
Connection* G 3/8" G 3/4"
Order code (with coupler
56500-163-4F44 56500-163-4F22 56500-163-4F17 56422-163-4F16
socket acc. EN 60529)
Order code for AMP Faston 56500-164-4F44 56500-164-4F22 56500-164-4F17 -
Order code for 1m-cable 56500-165-4F44-1 56500-165-4F22-1 56500-165-4F17-1 56422-165-4F16-1
Measuring rate l/min 1 - 15 1 – 25 2 – 35 5 – 65
K-factor PPL 2200 1000 750 210
DN mm 8 8 8 12

Type 1005 2F66 2006 2F66 2006 4F44 2008 2F66


Material Trogamid
Connection* NPT 1/4" NPT 3/8"
Order code (with coupler
56548-163-2F66 56512-163-2F66 56512-163-4F44 56502-163-2F66
socket acc. EN 60529)
Article n° (AMP-Faston) - 56512-164-2F66 56512-164-4F44 56502-164-2F66
Order code for 1m-cable 56548-165-2F66-1 56512-165-2F66-1 56512-165-4F44-1 56502-165-2F66-1
Measuring rate l/min 0,1 – 2,5 0,5 - 5 1 – 10 0,5 – 7,5
K-factor PPL 22.000 6900 3300 4700
DN mm 5 6 6 8
Type 2008 4F44 2008 4F22 2008 4F17 3012 4F16
Material Trogamid
Connection* NPT 3/8" NPT 3/4"
Order code (with coupler
56502-163-4F44 56502-163-4F22 56502-163-4F17 56423-163-4F16
socket acc. EN 60529)
Order code (AMP-Faston) 56502-164-4F44 56502-164-4F22 56502-164-4F17 -
Order code (with 1m 56502-165-4F44-1 56502-165-4F22-1 56502-165-4F17-1 56423-165-4F16
cable)
Measuring rate l/min 1 - 15 1 – 25 2 – 35 5 – 65
K-factor PPL 2200 1000 750 210
DN mm 8 8 8 12
*Other connections: O-ring; Hose nipple for Vision 2000

Type 1005 2F66 2006 2F66 2006 4F44 2008 2F66


Material Brass
Connection - - - G 3/8”
Order code (with coupler - - -
47000-163-2F66
socket acc. EN 60529
Connection - - - NPT 3/8"
Order code (with coupler - - -
47001-163-2F66
socket acc. EN 60529
Measuring rate l/min - - - 2-8
K-factor PPL - - - 4200
DN mm - - - 8
Type 2008 4F44 2008 4F22 2008 4F17 3012 4F16
Material Brass
Connection - G 3/8" - -
Order code (with coupler - -
47000-163-4F22 -
socket acc. EN 60529
Connection - NPT 3/8" - -
Order code (with coupler - -
47001-163-4F22 -
socket acc. EN 60529
Measuring rate l/min - 3 - 25 - -
K-factor PPL - 1000 - -
DN mm - 8 - -

TUR_VIS_BR_DB_02_1704.doc
Reprint of texts or text extracts requires prior written authorization of Badger Meter Europa GmbH.
Misuse of texts, pictures or company logo will be prosecuted .
Type 1005 2F66 2006 2F66 2006 4F44 2008 2F66 2008 4F44 2008 4F22 2008 4F17 3012 4F16
Measuring rate l/min 0.1 – 2.5 0,5 – 5 1 – 10 0,50 – 7,5 1 – 15 1 – 25 2 – 35 5 – 65
K-factor PPL* 22.000 6900 3300 4700 2200 1000 750 210
Size mm 5 6 6 8 8 8 8 12
Viscosity range 0.8 - 16 mm²/sec (mpas)
Accuracy +/- 3% of value
Repeatability Better than 0.5%
Temperature range -20 up to +100 °C
Operating pressure 25 bar max
Burst pressure 200 bar ~100 bar
Electrical connection ** - AMP Faston 2.8/6.3 x 0.8 mmm
- Connection MICRO-EN 60529 and 3 flat plugs 2.8 x 0.5 mm (point plate)
- Cable connection: round cable 3xAWG 24 with free cable end
Filter 20 to 40 micron recommended
Power supply 5 - 24 VDC.
Current consumption Ca. 8 mA.
Output signal Pulses by open collector (NPN sinking).
Output current Max. 20 mA.
Pull-up resistor 1 - 2.2 kOhm
* PPL = Pulses per liter
** DIN connector MICRO EN60529

Pressure drop in bar for water (at 20 °C)


Type 1005 2F 66 2006 2F66 2006 4F44 2008 2F66 2008 4F44 2008 4F22 2008 4F17 3012 4F16
0,5 l / min 0.02 -- -- -- -- -- -- --
1 l /min 0.05 ~0 ~0 ~0 ~0 ~0 ~0 --
1,5 l / min 0.15 -- -- -- -- -- -- --
2 l / min 0.25 * ~0 0.06 ~0 0.05 ~0 ~0 --
5 l / min -- 0.12 0.2 0.05 0.2 0.05 ~0 0,00
10 l / min -- 0.4 0.7 0.2 0.4 0.17 ~0.12 0,01
15 l / min -- 0.9 0.4 --_ 0.27 ~0.25 0,02
20 l / min -- 1.3 0.7 -- 0.48 ~0.45 0,05
25 l / min -- -- -- -- -- 0.65 ~0.60 0,07
30 l / min -- -- -- -- -- - ~0.92 0,11
35 l / min -- -- -- -- -- -- -- 0,14
40 l / min -- -- -- -- -- -- -- 0,18
45 l / min -- -- -- -- -- -- -- 0,23
50 l / min -- -- -- -- -- -- -- 0,28
55 l / min -- -- -- -- -- -- -- 0,34
60 l / min -- -- -- -- -- -- -- 0,40
65 l / min -- -- -- -- -- -- -- 0,47
* Value applies to 2.5 l/min

Display operation

Electr. connection Coil for battery powered display required


Supply 2 – 5,5 VDC
Power consumption Approx. 2,2 mA @ 2V
Output signal Push - Pull
Output current Max. 30 mA

TUR_VIS_BR_DB_02_1704.doc
Reprint of texts or text extracts requires prior written authorization of Badger Meter Europa GmbH.
Misuse of texts, pictures or company logo will be prosecuted .
Dimensions (in mm)

Example: Vision 1000 with DIN plug

(+)Red
5-24 VDC
(-)Black
0 VDC
(Ω)Brown
Output

Example: Vision 2000 with round cable 3 x AWG 24

Example: Vision 2000 with O-Ring

TUR_VIS_BR_DB_02_1704.doc
Reprint of texts or text extracts requires prior written authorization of Badger Meter Europa GmbH.
Misuse of texts, pictures or company logo will be prosecuted .
Example Vision 2000 with round cable 3 x AWG 24 and hose nipple

Example Vision 3000 with DIN plug

TUR_VIS_BR_DB_02_1704.doc
Reprint of texts or text extracts requires prior written authorization of Badger Meter Europa GmbH.
Misuse of texts, pictures or company logo will be prosecuted .
Displays for the VISION series
The ILR 701T, 750T series electronic register

Features
• Large six digit LCD display
• Display in liters, pints, quarts or gallons, freely pro-
grammable
• 11 digits, non-resettable lifetime totalizer and 6 digits,
resettable totalizer
• Temperature range: -20 °C to +80 °C
(-4 °F to +140 °F)
• Calibration factor saved in non-volatile memory
• 9 point linearization (ILR 750T, ILR 701T)
• Scalable pulse output and 4-20 mA output (ILR 750)
• Protection class IP65

Description
the batch totalization or the flow rate of the fluid going
The electronic register module contains a microprocessor board
through the meter on the 6-digit display.
powered by a lithium battery. It can be programmed to batch
The registers batch totalizer is a 6-digit display with three
in liters, pints, quarts, or gallons and will totalize in liters or
places of resolution after the decimal point. If the total
gallons. A calibration factor and unit of measure are
dispensed exceeds 999.999 then the display will shift and
programmed during factory test. Unlike mechanical registers,
only 2 digits will be displayed after the decimal point,
these units can be electronically recalibrated in the field when
9999.99 and will continue to shift to the maximum value of
necessary. A 6-digit LC display, accurate to three decimal
999999. After reaching 999999 the batch totalizer will
places, shows the exact amount of fluid that has passed
rollover to 0.000. The batch totalizer is reset to zero when
through the meter. The entire register module is protected
the reset button is depressed.
from normal wear and tear by a rugged, shock resistant
housing. The register also has a resettable totalizer that requires
that the total and reset button both be depressed to reset
Operation (hold the “Total” button, then press the “Reset” button to
Tubine meters have hall sensors that cause the turbine to send reset this totalizer while resettable totalize is displayed).
pulses to the register as rotates. This would be used for multiple batch totalization purposes.
The register is in a sleep mode until it detects these pulses
caused by fluid going through the meter. The micro-processor The register´s life time totalizer is 11 digits and will either
in the register then measures the flow and will display either be in gallons or liters based on the unit of measure
selected. Pushing and holding the total button while the
life time totalizer is displayed will display the full 11-digit
life time totalizer value.

Register model Register features


ILR 701T standard register • Flow rate and totalizer display selectable in the programming menu
• Selectable unit of measure
• 9 point linearization
ILR 750T pulse output • Scalable pulse output
+ 4-20mA output • Ability to set pulse output length
• Analog 4-20mA output
• Minimum and maximum values can be set for analog output
• 9 point linearization

TUR_VIS_BR_DB_02_1704.doc
Reprint of texts or text extracts requires prior written authorization of Badger Meter Europa GmbH.
Misuse of texts, pictures or company logo will be prosecuted .
The electronic displays are available in two versions

Mounted For Vision 2000, Vision 3000


version as battery-powered, on meter mounted
display providing instant and total flow
rate.
ILR 701T: actual flow, total flow rate
ILR 750T: actual flow, total flow rate
4-20 mA-output,
scalable pulse output

RF-Display For Vision 1000, Vision 2000, Vision 3000


(Remote reading via as a wireless, battery-powered (1 x
radio frequency)) CR123A battery) display, with 2.4 GHz
radio frequency, for remote reading of
total flow rate over a 5 - 6 m range.
Sender on top of turbine meter, receiver
for wall mount.
ILR 701T: total flow rate

Remote For Vision 1000, Vision 2000, Vision 3000


cable connection for remote reading
ILR 701T: actual flow, total flow rate
ILR 750T: actual flow, total flow rate
4-20 mA-output,
scalable pulse output

TUR_VIS_BR_DB_02_1704.doc
Reprint of texts or text extracts requires prior written authorization of Badger Meter Europa GmbH.
Misuse of texts, pictures or company logo will be prosecuted .
Example Vision 3000 with built up display

TRANSMITTER

Wireless system consisting of transmitter and receiver (shown here the transmitter on the Vision 3000)

RECEIVER

Wireless system consisting of transmitter and receiver (shown here the receiver)

TUR_VIS_BR_DB_02_1704.doc
Reprint of texts or text extracts requires prior written authorization of Badger Meter Europa GmbH.
Misuse of texts, pictures or company logo will be prosecuted .