Magnaclave Manual 0
Magnaclave Manual 0
Magnaclave Manual 0
Visual Index........................................................................................................................................................... 3
Section I - Familiarization.................................................................................................................................... 4
I. Safety Features....................................................................................................................................... 4
A. Door Clamp Ring..................................................................................................................... 4
B. Open Door Button.................................................................................................................... 4
C. Door Handle............................................................................................................................. 4
D. Relay Control........................................................................................................................... 4
II. Physical Characteristics......................................................................................................................... 5
A. Exterior.................................................................................................................................... 5
B. Interior...................................................................................................................................... 5
C. Installation Requirements......................................................................................................... 5
III. Electrical Requirements........................................................................................................................ 6
IV.Controls and Indicators......................................................................................................................... 6
A. Indicator Lights........................................................................................................................ 6
B. Gauges.................................................................................................................................... 6
C. Control Knobs.......................................................................................................................... 7
D. Audible Signal.......................................................................................................................... 7
E. Water Reservoir........................................................................................................................ 7
F. Stand........................................................................................................................................ 7
Section 2 - Preparation for Sterilization.............................................................................................................. 8
I. Handling and Cleaning of Instruments.................................................................................................... 8
A. Handling................................................................................................................................... 8
B. Cleaning................................................................................................................................... 9
II. Preparation Guide for Carbon Steel Instruments...................................................................................9
A. Handling................................................................................................................................... 9
B. Cleaning................................................................................................................................... 9
C. Sterilization Preparation........................................................................................................... 9
D. Instrument Wrapping................................................................................................................ 9
III. Tray Preparation and Loading................................................................................................................ 9
A. General Guidelines for Tray Preparation..................................................................................9
B. Unwrapped Trays.................................................................................................................... 10
C. Wrapped Trays and Instruments............................................................................................ 11
D. Packs..................................................................................................................................... 12
Section 3 - Operation.......................................................................................................................................... 15
I. General.................................................................................................................................................. 15
II. Operating Procedures for Normal Sterilization.....................................................................................15
A. Unlock Door............................................................................................................................ 15
B. Fill........................................................................................................................................... 15
C. Load........................................................................................................................................ 15
D.Lock Door................................................................................................................................ 16
E. Set Time Control..................................................................................................................... 16
F. Set Temperature………………………………………………………………………………………….16
G. Vent........................................................................................................................................ 16
H. Unlock Door............................................................................................................................ 16
I. Drying....................................................................................................................................... 16
2
TABLE OF CONTENTS
IV.Optional Accessories........................................................................................................................... 20
A. Trays...................................................................................................................................... 20
B. Temperature Recorder............................................................................................................ 20
V.Inspection............................................................................................................................................. 20
A. General................................................................................................................................... 20
B. Inspecting the Chamber......................................................................................................... 20
C. Door Interlock Check............................................................................................................. 20
VI.Product Disposal................................................................................................................................. 20
Glossary............................................................................................................................................................... 20
Autoclavable Conversion Scale.......................................................................................................................... 22
VISUAL INDEX
3
SECTION 1
FAMILIARIZATION
SAFETY FEATURES
The Magna-Clave design offers several safety
features for your protection.
A. Door Clamp Ring TO OPEN PUSH
BUTTON FIRMLY WHEN
A patented door clamp ring provides maximum DOOR LIGHT IS ON
safety by locking the entire rim of door. When
activated by the door handle, the door clamp
ring expands in diameter, allowing the door to OPEN
open. DOOR
BUTTON
B. Open Door Button
The open door button energizes the door interlock
solenoid. It must be depressed before moving the
door handle up into the open position. The button
is operative only when the Open Door light is
illumi- nated.
C. Door Handle
When door handle is in down position, the door is
locked and the rim is completely sealed. When
door handle is in up position, the door clamp ring
is expanded. To open door, move door handle to DOOR
upward position by the following method: CLAMP
· Ensure that Power On light is illuminated RING
· Open Door light must be illuminated
· Depress Open Door button
34½" 28"
(87.6 cm) (71.1 cm) II. PHYSICAL CHARACTERISTICS
Exterior:
The overall dimensions are 28" (71.1 cm) wide X 34
1/2” (87.6 cm) deep X 53" (134.6 cm) high (with
stand), 23' (58.4 cm) high (without stand).
23"
(58.4 cm) Interior:
The overall chamber shall measure
30" (76.2 cm) deep X 16” (40.6 cm) diameter.
Either side
panel can
be removed
with 6 screws
Outlet can
be rotated
90° in any
direction
Optional Path
Overfill hose
for Drain
can be passed
Tube
through the
hole in back of stand
into a bucket
or a sink outside of unit.
Ill. ELECTRICAL REQUIREMENTS
The Magnaclave is available in 208 VAC at 24 amps or 220/240 Vac, 50 or 60 hz. at 21 amps.
This unit must be installed in accordance to your local electrical codes. An outlet box is provided on the
back of the unit for the electrical connections and requires a dedicated 30 amp circuit. This unit may be
permanently connected to an electrical panel with 30 amp breakers (recommended 10 AWG wire) or by
connecting a 30 amp power cord to the outlet box. If the Magnaclave is wired using a 30 amp power cord,
the receptacle for the power cord should be accessible. The green wire in the outlet box is for the protective
earth ground, care should be taken when wiring the Magnaclave to verify that green ground wire is
connected to the electrical system’s ground.
2. Function Control - All operations of the Magna-Clave are controlled by the Function control as follows:
b. Fill- Water from reservoir is allowed to enter chamber; wall heating elements are energized.
c. Sterilize - Power is supplied to all heating elements as required; door cannot be opened.
d. Vent - Unused water and steam in chamber are returned to reservoir after sterilizing cycle is complete.
The door can be opened when the pressure decreases to a safe level and the Open Door light is
illuminated. The wall heaters remain energized. This reduces pressure build up time for successive
sterilizing cycles and promotes drying of the sterilized material.
3. Time Control - The Time control is manually set at the beginning of each sterilizing cycle. Timing
begins when the preset temperature is reached and time control counts down to “0.’
NOTE: Do not set the time control for an interval of less than five minutes.
D. Audible Signal
1. Buzzer - When the Time control returns to “0”, a buzzer will sound to indicate the end of the cycle. The
buzzer sounds continuously until the Function control is turned to Vent. The Time control must be set
at the beginning of each sterilizing cycle. If it is not set, the buzzer will indicate end of sterilizing cycle
when the preset temperature is reached, and the sterilizing cycle will not be complete. Once buzzer
sounds the main heater will turn off.
Stand
The Magna-Clave stand provides a convenient storage com- partment to house trays or supplies.
7
SECTION 2
PREPARATIONS FOR STERILIZATION
• An internal process indicator strip should be in- • Place all sharps (scissors, knives, skin hooks) so
cluded with each sterilizer load to verify gross they do not touch during autoclaving. Cotton or
heat penetration. gauze may be used to isolate and protect the
• A biological spore test indicator should be used sharp edges and the smaller instruments.
weekly in a representative sterilizer load for • Do not place metal instruments, other than
ster- ilization assurance. stainless steel, directly on stainless steel autoclave
• Date packages and re-sterilize after one trays. The trays should be lined with thin cloth or
month, or according to packaging paper liners.
manufacturers’ speci- fications. • Any item which might hold water should be placed
• Sterilization indicators/monitors should be so the water will drain out.
placed in the front bottom area of the sterilizer. • Use small packs to separate larger ones. See
• DO NOT mix or cause contact of dissimilar instruc- tions for preparing wrapped trays.
met- als (i.e. metals such as carbon steel and • Never stack trays on top of one another. At least
stain- less steel) during sterilization or storage. one inch should separate trays.
• Make sure that all instruments are sterilized in • Wrapped trays and packs must not touch the sides
an open position. (Refer to AORN guidelines.) of the sterilizer.
• Do not overload trays. Overloading may cause
9
steril- ization failure.
10
SECTION 2
PREPARATIONS FOR STERILIZATION
B. Unwrapped Trays
These trays are prepared for sterilization of nonsurgical instruments and canisters to prevent
transmitting infectious disease. Always include a process or spore test indicator with every sterilizer
load.
1. Loose Instruments
a. Distribute a single layer of instruments in
the bottom of a perforated tray. Instru-
ments made of metal other than
stainless steel should not contact the
stainless steel autoclave tray. A very thin
liner of paper or cloth should be used for
separa- tion. Provide adequate space
between instruments for steam
circulation. Do not overload. Be sure to
include a process or spore test indicator.
b. Place tray on rack in sterilizer.
NOTE: If multiple layers of instruments
are to be separated by fabric, see
instructions for preparing wrapped
trays.
c. Refer to Section 3 for operating instructions.
3. Rubber Tubing
a. Clean tubing thoroughly.
b. Rinse with pyrogen free water leave
wet. Leave both ends open. Coil and
wrap without kinks or sharp bends.
c. Place tubing on an autoclave tray. Be
sure to include a process or spore
test indicator.
d. Place tray on rack in sterilizer. Do not
overload or inadequate sterilization
and drying will result.
CAUTION: Tubing which will come in
contact with a surgical wound should
be prepared as outlined above, and
SECTION 2
PREPARATIONS FOR STERILIZATION
wrapped to maintain sterility.
e. Refer to Section 3 for operating instructions.
C. Wrapped Trays and Instruments
There are several ways to prepare wrapped trays. Refer to Table 2-2 for acceptable wrapping materials.
1. Wrapped Instruments
a. Collect a group of similar instruments which have
been cleaned according to Section 2-1.
or
Collect instruments used for a particular procedure (i.e.
amalgam, exam, etc.) which have been cleaned
according to Section 2-1. Do not mix dissimilar metals.
D. Packs
Packs are for sterilizing surgical instruments, gloves and textiles.
1. Instruments and textiles
a. Clean instruments according to Section 2-I. Textiles should be laundered prior to sterilization.
NOTE: Do not use high chlorine or phosphate content detergents chlorine bleach in
laundering items prior to sterilization. Staining of the autoclave and instruments or more
severe damage
may result. The use of chlorides may also result in cracks in the chamber.
b. Loosely package instruments with not more than 10 per pack. Instruments of the same type
which are nested should be separated by a layer of absorbent towels and placed so that
water will run out. Loosely roll or fold textiles.
CAUTION: Density should not exceed 1/2 the capacity of the pack and the packs should not
exceed 1/2 the capacity of the tray. Otherwise inadequate sterilization could result.
Or
Wrap properly cleaned articles in two to four layers of muslin towels or other packaging
material. (Refer to Table 2-2.)
c. Place a process or spore test indicator inside a representative pack.
d. Seal with autoclave tape or heat sealer (3M, St. Paul, MN; Propper, Long Island City, NY).
CAUTION: Do not use staples, pins or other devices that could puncture packaging
mate- rial. Otherwise sterility could be compromised.
e. Place packs on perforated trays. Leave adequate space between packs to allow steam to
circu- late. Load packs upright, side-by-side on the tray. Do not stack.
CAUTION: Adequate drying will not occur unless space is left between packs. Metal
and glass containers should not be used to separate packs as these will inhibit drying.
Packs should not exceed 1/2 the capacity of the tray. Otherwise, sterilization could be
compro- mised.
f. Load trays onto the racks in the sterilizer. Ensure that packs do not touch the sides of the
chamber. Allow adequate space between trays for steam circulation. If packs are large,
some trays may be omitted to allow more clearance.
CAUTION: Do not overload or inadequate sterilization may result.
g. Place pack containing sterilization monitor in the bottom front of the sterilizer.
h. Refer to Section 3 for operating instructions.
SECTION 2
PREPARATIONS FOR STERILIZATION
2. Surgical gloves
NOTE: Disposable gloves should not be sterilized.
SUITABLE UNSUITABLE
K 10 13 1
Material to be Sterilized P 1 20 2
VS. F° 25 26 2
Time in Minutes C° 12 12 1
1 7 3
Fabrics - Loosely woven - 30 20 -
Wrapped in
muslin
Fabrics - Tightly woven 40 30 -
Instruments - Individually 30 15 1
wrapped in muslin 0
Syringes Needles 15 10 7
and
Drums - Loosely woven contents 30 20 -
Drums - Tightly woven contents 40 30 -
Utensils - Loosely woven contents 30 20 1
0
Rubber - In muslin 15 - -
packs Gloves
Rubber In muslin packs 15 - -
Covers -
Wrapped - 15 -
SECTION 3
OPERATIONS
WARNING: Failure to operate or maintain the Magnaclave in accordance to these instructions may
impair the protection provided by the Magnaclave.
WARNING: Do not attempt to bypass any of the safety interlock systems described below. Failure
to do so could result in personal injury.
I. GENERAL
A. When the Function control is in the Power Off position, all electrical power is off and the door cannot
be operated. An additional safety interlock is included to prevent the door from being opened when
there is an unsafe level of pressure in the chamber. Another interlock prevents the main heaters from
becoming energized to generate steam unless the door is fully locked. The Open Door light will
illuminate when it is safe to open the autoclave door. When the Function control is in the Fill
position, water flows from the reservoir into the chamber. When the water level indicator in the
chamber is covered, the Function control should be turned to Sterilize to stop the flow of water. To
generate steam, the door must be closed and locked. The Time control should be set at the
beginning of the sterilizing cycle. Timing will begin when the preset temperature has been reached,
and will continue until the Time control returns to “0.”
B. When the sterilizing procedure is complete, the Sterile light will illuminate and the chime will sound.
When the buzzer sounds, turn the Function control to Vent to cut off the Sterile light and the buzzer.
Steam and unused water will then be returned to the reservoir from the chamber and pressure inside
the chamber will be reduced to near zero. When the chamber pressure reaches a safe level, the
Open Door light will become illuminated.
WARNING: Never attempt to open door unless open door light is illuminated and pressure
gauge indicates “0.’
To aid in absorbing moisture, the door can now be slightly opened. The chamber heaters will remain
energized. Leave materials inside the chamber until dry. Drying time is determined by the size of the
material and manner in which it was packed. Drying times have not been established due to varying
conditions. When the Function control is left in the Vent position, the Magna-Clave will remain in a
warm, standby condition.
A. Unlock door
Turn Function control to Vent or Fill. Depress and hold Open Door
button on front of Magna-Clave and lift door locking handle on right
side of unit.
B. Fill
Turn Function control to Fill. When water in chamber covers water
level indicator, rotate Function control to Sterilize. Note: Ensure
wa- ter level indicator is completely submerged in water. WATER LEVEL
INDICATOR
C. Load
15
SECTION 3
load.)
DOOR
TIME HANDLE
15 Minutes: Unit should be monitored to determine the unit is heating CONTROL
16
IV. OPERATING GUIDELINES INDICATOR
LIGHTS TEMPERATURE
GAUGE
17
SECTION 3
CARE AND MAINTENANCE
1. PREVENTIVE MAINTENANCE SCHEDULE
Clean and inspect Weekly 12 oz.. Omni-Cleaner See Sections 4-Il and
chamber to I gallon distilled water 4-IV. Call authorized
on 20 minute cycle. service representative
cracks or fissures are
found. Do not operate
unit.
Safety valve ring Every 3 months Manually pull ring on back of If valve does not
WARNING:When the unit while chamber is un- open, turn off POWER
ring is pulled on der pressure. and call for service
safety valve with representative.
unit under pressure
steam is discharged
from the chamber
at a high tempera-
ture. Keep clear of
steam discharge
path to avoid burns.
Use of a hot pad or
instrument to pull
ring is suggested.
Door gasket Weekly Inspect and clean using Call authorized service
Omni-Cleaner or mild representative for re-
detergent and distilled placement gaskets.
water. Check for leaks
(have leaking gasket
replaced).
Chamber fill filter Weekly Inspect and clean. See Section 4-Il.
II. CLEANING
It is highly recommended that the autoclave be cleaned a minimum of once a week with Omni-Cleaner (Part
No. WZI-091295) and distilled or demineralized water. The Omni-Cleaner is a mildly alkaline concentrate used
to clean and descale autoclaves. Minerals, especially chlorides, are corrosive to stainless steel. Therefore,
tap water with a high mineral content should not be used. When sterilizing saline solutions, it is imperative
that the autoclave be cleaned after each use. With proper cleaning, the autoclave will provide higher
performance and longer life.
A. Cleaning Procedures
1. Mix 12 ounces of Omni-Cleaner in one gallon of distilled, demineralized water.
2. Drain water from reservoir. Refill reservoir with a solution of Omni-Cleaner and water. (The reservoir
will not be full. However, the level will be adequate.)
3. Run one- 20 minute sterilizing cycle to remove all grease and grime from the system. If the
autoclave is extremely dirty, it may require a second cleaning. Do not sterilize instruments while
cleaning the auto- clave.
4. Drain cleaning solution from reservoir and chamber. Rinse thoroughly with clean, mineral-free water,
and run a rinse cycle for fifteen minutes.
5. Drain rinse solution and wipe inside of boiler thoroughly. If scale or lime deposits remain on inside of
chamber, ensure that autoclave is cool. Then clean with water, plastic or nylon scouring pads and a
nonchlorinated detergent.
NOTE: Detergents containing chlorine are corrosive to stainless steel and should not be used. Do
not use ordinary steel wool or steel brushes on stainless steel. Pads containing metal may damage
cham- ber.
6. Refill reservoir with clean, mineral-free water. The Magna-Clave is now ready for use.
B. Draining Reservoir
1. The reservoir drain hose is located inside the right-hand door of stand, near the top. For Magna-
Claves which do not have a stand, the hose will be located under the front edge and to the left of the
unit. The drain hose may be drained by removing it from its clip and unscrewing the tip end. The
hose should be drained into a 10 quart capacity container. When the reservoir is completely drained,
replace tip and clip drain hose in place.
A. Trays
A basket tray and shallow flat tray are standard equipment. A bedpan tray is available as an option.
B. Temperature Recorder
A recording thermometer which provides a permanent record of the steam temperature may be factory
or field installed.The steam temperature is recorded on a circular chart for a 24 hour period.
V. INSPECTIONS
A. General
1. The Magna-Clave is a pressure vessel that falls under various state and/or local laws which differ in
inspection requirements. Some laws require complete periodic inspection of a pressure vessel. The
inspection period varies according to individual laws. This inspection is usually performed by a qualified
inspector commissioned by the National Board of Boilers and Pressure Vessels. Insurance companies may
also require a similar type of inspec- tion. The governmental agency in your area and/or your insurance
company will determine the inspection require- ments for your Magna-Clave.
2. For additional information concerning the Magna-Clave, contact Pelton & Crane, P. 0.Box 241147,
Charlotte NC, 28224 or your full service dealer. When ordering service or parts, always include the serial
number of your unit.
B. Inspecting the Chamber
1. Inside Chamber Inspection
Make a thorough inspection inside the chamber every six months. If cracks or fissures are found, call a
qualified service technician. Do not operate a unit with cracks in the chamber.
WARNING: Do not perform the following test with any pressure in the chamber.
C. Door Interlock Check
1. With the unit cold, open door and push down on door locking handle.
2. Rotate Function control counterclockwise to Sterilize.
3. Depress and hold Open Door button and pull up on door locking handle as if to open door. Do not force
door handle. Use no more force than it takes to open the door normally.
4. If door locking handle can be pulled all the way up and the clamp ring expanded, turn the Function control
to
Power Off and call a qualified service technician immediately. Never operate a Magna-Clave in this
condition.
5. If door locking ring cannot be opened, rotate the Function control to Vent, depress Open Door button, lift
door locking handle up, and open door locking ring. The Magna-Clave is now ready for use again.
VI. PRODUCT DISPOSAL
Contact your local authorized dealer for proper disposal of the device to ensure compliance with your local
environmental regulations.
GLOSSARY
Cold Sterilant: An agent which sterilizes at room temperature. Usually a chemical such
as glutaraldehyde.
Corrosion Inhibitor: A chemical substance which, when used in small amounts, effectively
reduces the corrosion rate of metals such as carbon steel.
Sterilization: Total destruction of all microbial life including bacteria, viruses and
spores.
Ultrasound: A type of cleaner which uses ultrasonic waves at high frequency to agitate
contaminants and dirt from items.
23
NOTES
We reserve the right to make any alterations which may be due to any technical improvements.