DF50 Service Manual-1
DF50 Service Manual-1
Service Manual
1
Declaration
Dymind Biotech reserves the right of final interpretation of this service manual.
Dymind Biotech shall be responsible for the safety, security, and performance of the product only
when all of the following conditions are met:
The assembly, re-commissioning, extension, modification, and repair of the product are
performed by authorized personnel of Dymind Biotech.
The installation of the relevant electrical devices complies with applicable national standards.
The product is operated in accordance with this service manual.
Maintenance Services
Scope of free services:
All products complying with Dymind Biotech’s Product Warranty specifications may qualify for free
services.
Scope of fee-based services:
Free services are available to all products exceeding Dymind Biotech's Product Warranty
specifications.
Products within the warranty period require maintenance under the following circumstances:
Artificial damage
Improper use
When grid voltage exceeds the specified operational range
Unavoidable natural disasters
When parts and supplies are replaced with no prior consent from Dymind Biotech, or machine
maintenance is performed by personnel with no prior authorization from Dymind Biotech.
Any failure by hospitals or organizations responsible for using this instrument to implement a
competent repair/maintenance plan is likely to result in abnormal instrument failure or even a
health hazard.
Make sure the analyzer is only operated under the conditions of use as specified in the operator's
manual. If operated outside of the specified conditions of use, the analyzer may not work properly,
which could lead to unreliable measurement results and damaged instrument components or
even bodily injury.
This analyzer must be operated by trained medical testing professionals, doctors, nurses, or
laboratory technicians.
2
Contents
Contents
1 Manual Overview ............................................................................................................................ 1
1.1 Who Should Read This Manual ............................................................................................... 1
1.2 Symbols and Legends.............................................................................................................. 1
1.3 Safety Instructions.................................................................................................................... 4
2 Instrument Configuration .............................................................................................................. 6
2.1 Mechanical Components.......................................................................................................... 6
2.2 Hydraulics System ................................................................................................................... 9
2.2.1 Schematic Diagram of the Hydraulic System ................................................................ 10
2.2.2 Hydraulics Diagram for the Overall Unit ........................................................................ 10
2.2.3 DIFF/WBC Measurement Channel .................................................................................11
2.2.4 HGB Measurement Channel.......................................................................................... 12
2.2.5 RBC/PLT Measurement Module .................................................................................... 13
2.2.6 Sampling and Blood-dispensing Module ....................................................................... 13
2.2.7 Power Supply and Waste Discharge Module ................................................................ 14
2.2.8 Status Monitoring Module .............................................................................................. 14
2.2.9 Hydraulic Components................................................................................................... 15
2.2.10 Main Measurement Modes .......................................................................................... 20
2.2.11 Hydraulics Maintenance............................................................................................... 27
2.2.12 Troubleshooting Common Hydraulics Problems.......................................................... 28
2.3 Hardware System................................................................................................................... 40
2.3.1 Main Control Panel ........................................................................................................ 41
2.3.2 Other Panels .................................................................................................................. 56
2.4 Optical System ....................................................................................................................... 59
2.4.1 Integral Replacement of the Optical Assembly .............................................................. 59
2.4.2 Replacing the Optical Preamplifier Panel ...................................................................... 61
2.4.3 Replacing the Laser Driver panel................................................................................... 62
2.4.4 Fine Tuning of the Flow Chamber.................................................................................. 64
2.4.5 Replacement of High-angle Receiving Plate ................................................................. 65
2.4.6 Common Problems and Solutions ................................................................................. 66
3 Repairs .......................................................................................................................................... 68
3.1 Overview ................................................................................................................................ 68
3.2 Preparatory Work Before Repairs .......................................................................................... 68
3.2.1 Disassembling the Left Side Panel ................................................................................ 68
3
Contents
4
Contents
5
1 Manual Overview
1 Manual Overview
This chapter explains the procedures for maintaining Auto Hematology Analyzer. Please read this
manual carefully to guarantee proper device maintenance and operator safety.
This manual is complemented by the operator's manual; there is no overlap between the two.
Be sure to operate the device in strict accordance with the instructions in the maintenance and
operator's manuals.
Symbol It means
The operator should follow the instruction below each symbol to avoid
personal injury.
1
1 Manual Overview
Symbol It means
The operator should follow the instructions below each symbol and
pay special attention to any pertinent information when following the
procedures.
Laser Warning:
This sign serves as a reminder of laser radiation.
The following symbols can be found on the analyzer, reagents, QC materials, or calibrator:
Caution
Biohazard
Network interface
Protective grounding
Lot No.
Exp. date
Serial No.
2
1 Manual Overview
Date of manufacture
Manufacturer
Storage temperature
Avoid sunlight
Keep dry
No rolling
No Stacking.
Recyclable materials
3
1 Manual Overview
The analyzer, after being scrapped, should not be disposed with other
household garbage, instead, it should be collected and recycled
following the disposal instructions for scrapped electronic and electrical
equipment.
Any failure by hospitals or organizations responsible for using this instrument to implement a
competent repair/maintenance plan is likely to result in abnormal instrument failure or even a
health hazard.
To prevent risk of explosion, do not use combustible gases (e.g. anesthetics) or liquids (e.g.
ethanol) near this product.
The machine should be powered off while troubleshooting. Any maintenance operation while the
power is on can lead to electric shock or damage to its electrical components.
Please connect the device to the socket using a separate fuse and surge protection switch.If the
device shares a fuse and surge protection switch with other equipment, e.g. life support
equipment, any malfunction may cause an electric surge to occur when the instrument is
powered on, which can trip the circuit breaker.
Maintenance personnel must keep their clothes, hair, and hands away from moving parts such
as the sample probe, jaws, and puncture needle to avoid being stabbed or pinched during
maintenance.
Parts with special warnings may be subject to mechanical movements and thus lead to injuries
from pinching or stabbing during normal operation or disassembly and repair.
Operators are obligated to comply with local and national regulations with regard to the disposal
and emission of expired reagents, waste, waste samples, consumables, and so on.
The reagents are irritating to eyes, skin and mucosa. When handling reagents and their related
items in the laboratory, the operator should comply with laboratory safety regulations and wear
personal protective equipment (such as a protective lab suit, gloves, etc.).
If the reagent accidentally comes in contact with your skin, wash it off immediately with plenty of
water and see a doctor if necessary.Do the same if you accidentally get any of the reagent in
your eyes.
4
1 Manual Overview
Improper maintenance may damage the analyzer. Maintenance personnel should maintain the
device in accordance with the instructions contained in the service manual, and inspect the
device properly after each maintenance.
If you encounter a problem not specified in the service manual, please contact Dymind's
aftersales service department. A professional will be assigned to offer you maintenance advice.
When repairing the electronic components of the device, please take off any metal accessories
you might be wearing to prevent personal injury or damage to the device.
Electrostatic discharge may cause damage to the device’s electronic components. If the repair
process can result in any electrostatic damage, please wear an antistatic wrist strap or maintain
the device on an antistatic workstation.
This analyzer must be operated by trained medical testing professionals, doctors, nurses, or
laboratory technicians.
Samples, controls, calibrators, and waste may pose a potential risk of bio-contamination. When
handling reagents and related items in the laboratory, the operator should comply with laboratory
safety regulations and wear personal protective equipment (such as a protective lab suit, gloves,
etc.).
Both the parts and surface of the analyzer are potentially infectious, so please take safety
precautions during operation and maintenance.
The pointed sample probe, any blood stains left on it, QC materials, and calibrators may be
potentially bio-infectious, so please avoid contact with the sample probe.
This sign warns of laser radiation.Do not look directly at the laser beams or see through the optical
system.
5
2 Instrument Configuration
2 Instrument Configuration
6
2 Instrument Configuration
7
2 Instrument Configuration
Figure 2-4 Front view of the analyzer (panel cover, rear shell removed)
8
2 Instrument Configuration
2. Hydraulics System
The hydraulics system consists of the DIFF/WBC measurement module, the HGB measurement
module, the RBC/PLT measurement module, the sampling and blood dispensing module, the power
supply and waste discharge module, and the status monitoring module. Detailed description:
DIFF/WBC measurement module
Consists of the dosing syringe, WBC reaction bath, flow chamber, tubing, and valve.
HGB measurement module
Consists of the dosing syringe, WBC reaction bath, tubing, valve, HGB emission light, and HGB
receiving tube.
RBC/PLT measurement module:
Consists of the dosing syringe, RBC/PLT counting bath assembly, tubing, and valve.
Sampling and blood dispensing module
Consists of the sample probe, sample syringe, swabs, and sampling assembly.
Power supply and waste discharge module
Consists of the vacuum chamber, liquid pump, valve, and tubing.
Status monitoring module
Consists of the optocoupler and sensor.
9
2 Instrument Configuration
Whole-blood CBC+DIFF
Mode Sampling
Dispensing
DIL-C
DIL-C
T70 Sheath
pressure
T155 sensor
T152
J12 T150 T80 T80 T207
T70
T70
T29 T156
LV8 T209 LV6 LV9 J3
T70 T29
J11 T29 T29
T29 T70 T130 T40 LV11
T33 J3 T70 J14 J14 T72 J2 T195
T192 J1 DILU
T153 J3 DILU
T207 T80 T208 T70 T194 J2
T80 T207 LH
LV10 LV7 LV5 LV1 LH
J6 T207
DIFF I
T80 T165 T213 T206 T63 DIFF I T151 T14
T158 T79
J13 T70
T72 J2
T204
T204
J2 T199
T42 T70 T70
T5 T15
T89 T171 T115
T70 T205 T195 T13
T20 T29 J10 T195 J10
T134 T29
T29 T165 T80 T80 T80 T70
T72 T72 T72 J4 LYC-2 LYC-1 DIL-C
T9 J2 J5 J5 J5 T197 T195 T208 T186
T186
T80
T69
J2
LV4 LV2 LV3
WBC RBC T210 T157 T157 Pressure T166
J7 J2 T29 T83 sensor
T28
T29 T197 J6 J6 J6 J6 T212 T14
J6
LV12 T163
T70 T70
J9
J6 J8
T29 T80 T172
T211 T10 T48 T48 T197
Check T10 T29
LV18 J3 J3 T80
LV16 LV15
value J6 T80 T177 T80 T80
J3 T170
T108 T70 T11 T11
T29 T80 T19
T65 J3
T65 T80
J6 T171 T59 J4 J3 J3
T29 T80
J3 T29
T172 T172
LV17 LV14 T80
LV19 LV13
J8 J8
T29
T195 waste
T162
P1
T164 J10 T80 T167 T168
T80 T168
10
2 Instrument Configuration
2. Measurement Process
The measurement process for the DIFF measurement module is as follows:
1. Dosing: The diluent syringe is first applied to add the diluent into the WBC bath, and then the
sample probe is used to add the blood sample into the WBC bath, where they are mixed with
LYC-2 evenly for incubation.
2. Mixing: Open Valve LV19 to generate air bubbles through the intermittent valve opening to mix
the sample well.
3. Measurement: Aspirate the incubated sample from the WBC bath using the sample syringe, then
close LV12 and open LV01, LV05 and LV08 so that the sheath flow syringe can push the sheath
fluid into the flow chamber, thereby forming a stable sheath flow. Then push the sample into the
flow chamber using the sample syringe. This is how the sample, bundled in the sheath flow,
enters the measurement section for measurement.
11
2 Instrument Configuration
2. Measurement Process
The measurement process for the HGB measurement module is as follows:
1. Dosing: The diluent syringe is first applied to add the diluent into the WBC bath, and then the
sample probe is used to add the blood sample into the WBC bath, where they are mixed evenly.
After aspirating the diluted sample for the first time, LYC-1 lyse is added into the WBC bath for
incubation.
2. Mixing: Open Valve LV19 to generate air bubbles through the intermittent valve opening to mix
the sample well.
3. Measurement: The blood sample and lyse are mixed well for a complete reaction so that the
parameter voltage of the sample can be measured. HGB can then be calculated based on the
local voltage and parameter voltage according to Lambert-Beer's Law.
12
2 Instrument Configuration
2. Measurement Process
The measurement process for the RBC/PLT measurement module is as follows:
1. Dosing: The diluent syringe is first applied to add the diluent into the RBC/PLT bath as the base
liquid, before using the sample probe to aspirate the sample diluted for the first time from the
WBC bath into the RBC/PLT bath.
2. Mixing: Open Valve LV13 to generate air bubbles through the intermittent valve opening to mix
the sample well.
3. Measurement: Open Valve LV18 to aspirate the sample out of the RBC/PLT bath through the
aperture by means of the negative-pressure chamber (with a negative pressure of -30Kpa). The
RBC and PLT particles generate electric pulses of various amplitudes when traveling through the
aperture, allowing the RBC and PLT cells to be counted according to the number of pulses
emitted.
4. Cleaning: To clean, open Valve LV02 and add the diluent into the RBC/PLT bath using the diluent
syringe.
5. Waste discharge: Waste is discharged by opening Valve LV13, LV15 and Pump P1.
13
2 Instrument Configuration
14
2 Instrument Configuration
15
2 Instrument Configuration
16
2 Instrument Configuration
17
2 Instrument Configuration
18
2 Instrument Configuration
19
2 Instrument Configuration
20
2 Instrument Configuration
Whole-blood CBC+DIFF
Mode Sampling
Dispensing
DIL-C
DIL-C
Predilute-CBC+DIFF
Mode Sampling
Dispensing
DIL-C
DIL-C
21
2 Instrument Configuration
Whole-blood CBC
Mode Sampling
Dispensing
DIL-C
Diluted sample
DIL-C
RBC/PLT WBC
Measurement module measurement module
Predilute-CBC Mode
Sampling
Dispensing
DIL-C
Diluted sample
DIL-C
RBC/PLT WBC
Measurement module measurement module
22
2 Instrument Configuration
T70 Sheath
pressure
T155 sensor
T152
J12 T150 T80 T80 T207
T70
T70
T29 T156
LV8 T209 LV6 LV9 J3
T70 T29
J11 T29 T29
T29 T70 T130 T40 LV11
T33 J3 T70 J14 J14 T72 J2 T195
T192 J1 DILU
T153 J3 DILU
T207 T80 T208 T70 T194 J2
T80 T207 LH
LV10 LV7 LV5 LV1 LH
J6 T207
DIFF I
T80 T165 T213 T206 T63 DIFF I T151 T14
T158 T79
J13 T70
T72 J2
T204
T204
J2 T199
T42 T70 T70
T5 T15
T89 T171 T115
T70 T205 T195 T13
T20 T29 J10 T195 J10
T134 T29
T29 T165 T80 T80 T80 T70
T72 T72 T72 J4 LYC-2 LYC-1 DIL-C
T9 J2 J5 J5 J5 T197 T195 T208 T186
T186
T80
T69
J2
LV4 LV2 LV3
WBC RBC T210 T157 T157 Pressure T166
J7 J2 T29 T83 sensor
T28
T29 T197 J6 J6 J6 J6 T212 T14
J6
LV12 T163
T70 T70
J9
J6 J8
T29 T80 T172
T211 T10 T48 T48 T197
Check T10 T29
LV18 J3 J3 T80
LV16 LV15
value J6 T80 T177 T80 T80
J3 T170
T108 T70 T11 T11
T29 T80 T19
T65 J3
T65 T80
J6 T171 T59 J4 J3 J3
T29 T80
J3 T29
T172 T172
LV17 LV14 T80
LV19 LV13
J8 J8
T29
T195 waste
T162
P1
T164 J10 T80 T167 T168
T80 T168
23
2 Instrument Configuration
T70 Sheath
pressure
T155 sensor
T152
J12 T150 T80 T80 T207
T70
T70
T29 T156
LV8 T209 LV6 LV9 J3
T70 T29
J11 T29 T29
T29 T70 T130 T40 LV11
T33 J3 T70 J14 J14 T72 J2 T195
T192 J1 DILU
T153 J3 DILU
T207 T80 T208 T70 T194 J2
T80 T207 LH
LV10 LV7 LV5 LV1 LH
J6 T207
DIFF I
T80 T165 T213 T206 T63 DIFF I T151 T14
T158 T79
J13 T70
T72 J2
T204
T204
J2 T199
T42 T70 T70
T5 T15
T89 T171 T115
T70 T205 T195 T13
T20 T29 J10 T195 J10
T134 T29
T29 T165 T80 T80 T80 T70
T72 T72 T72 J4 LYC-2 LYC-1 DIL-C
T9 J2 J5 J5 J5 T197 T195 T208 T186
T186
T80
T69
J2
LV4 LV2 LV3
WBC RBC T210 T157 T157 Pressure T166
J7 J2 T29 T83 sensor
T28
T29 T197 J6 J6 J6 J6 T212 T14
J6
LV12 T163
T70 T70
J9
J6 J8
T29 T80 T48 T172
T211 T10 T48 T29 T197
Check T10
LV18 J3 J3 T80
LV16 LV15
value J6 T80 T177 T80 T80
J3 T170
T108 T70 T11 T11
T29 T80 T19
T65 J3
T65 T80
J6 T171 T59 J4 J3 J3
T29 T80
J3 T29
T172 T172
LV17 LV14 T80
LV19 LV13
J8 J8
T29
T195 waste
T162
P1
T164 J10 T80 T167 T168
T80 T168
24
2 Instrument Configuration
T70 Sheath
pressure
T155 sensor
T152
J12 T150 T80 T80 T207
T70
T70
T29 T156
LV8 T209 LV6 LV9 J3
T70 T29
J11 T29 T29
T29 T70 T130 T40 LV11
T33 J3 T70 J14 J14 T72 J2 T195
T192 J1 DILU
T153 J3 DILU
T207 T80 T208 T70 T194 J2
T80 T207 LH
LV10 LV7 LV5 LV1 LH
J6 T207
DIFF I
T80 T165 T213 T206 T63 DIFF I T151 T14
T158 T79
J13 T70
T72 J2
T204
T204
J2 T199
T42 T70 T70
T5 T15
T89 T171 T115
T70 T205 T195 T13
T20 T29 J10 T195 J10
T134 T29
T29 T165 T80 T80 T80 T70
T72 T72 T72 J4 LYC-2 LYC-1 DIL-C
T9 J2 J5 J5 J5 T197 T195 T208 T186
T186
T80
T69
J2
LV4 LV2 LV3
WBC RBC T210 T157 T157 Pressure T166
J7 J2 T29 T83 sensor
T28
T29 T197 J6 J6 J6 J6 T212 T14
J6
LV12 T163
T70 T70
J9
J6 J8
T29 T80 T48 T172
T211 T10 T48 T29 T197
Check T10
LV18 J3 J3 T80
LV16 LV15
value J6 T80 T177 T80 T80
J3 T170
T108 T70 T11 T11
T29 T80 T19
T65 J3
T65 T80
J6 T171 T59 J4 J3 J3
T29 T80
J3 T29
T172 T172
LV17 LV14 T80
LV19 LV13
J8 J8
T29
T195 waste
T162
P1
T164 J10 T80 T167 T168
T80 T168
25
2 Instrument Configuration
T70 Sheath
pressure
T155 sensor
T152
J12 T150 T80 T80 T207
T70
T70
T29 T156
LV8 T209 LV6 LV9 J3
T70 T29
J11 T29 T29
T29 T70 T130 T40 LV11
T33 J3 T70 J14 J14 T72 J2 T195
T192 J1 DILU
T153 J3 DILU
T207 T80 T208 T70 T194 J2
T80 T207 LH
LV10 LV7 LV5 LV1 LH
J6 T207
DIFF I
T80 T165 T213 T206 T63 DIFF I T151 T14
T158 T79
J13 T70
T72 J2
T204
T204
J2 T199
T42 T70 T70
T5 T15
T89 T171 T115
T70 T205 T195 T13
T20 T29 J10 T195 J10
T134 T29
T29 T165 T80 T80 T80 T70
T72 T72 T72 J4 LYC-2 LYC-1 DIL-C
T9 J2 J5 J5 J5 T197 T195 T208 T186
T186
T80
T69
J2
LV4 LV2 LV3
WBC RBC T210 T157 T157 Pressure T166
J7 J2 T29 T83 sensor
T28
T29 T197 J6 J6 J6 J6 T212 T14
J6
LV12 T163
T70 T70
J9
J6 J8
T29 T80 T172
T211 T10 T48 T48 T197
Check T10 T29 J3
value J6 T80 T177
LV18 T80
J3
T80 T80 LV15
J3 T170
T108 T70
T29
T11 T11 LV16 T80 T19
T65 J3
T65 T80
J6 T171 T59 J4 J3 J3
T29 T80
J3 T29
T172 T172
LV17 LV14 T80
LV19 LV13
J8 J8
T29
T195 waste
T162
P1
T164 J10 T80 T167 T168
T80 T168
26
2 Instrument Configuration
T70 Sheath
pressure
T155 sensor
T152
J12 T150 T80 T80 T207
T70
T70
T29 T156
LV8 T209 LV6 LV9 J3
T70 T29
J11 T29 T29
T29 T70 T130 T40 LV11
T33 J3 T70 J14 J14 T72 J2 T195
T192 J1 DILU
T153 J3 DILU
T207 T80 T208 T70 T194 J2
T80 T207 LH
LV10 LV7 LV5 LV1 LH
J6 T207
DIFF I
T80 T165 T213 T206 T63 DIFF I T151 T14
T158 T79
J13 T70
T72 J2
T204
T204
J2 T199
T42 T70 T70
T5 T15
T89 T171 T115
T70 T205 T195 T13
T20 T29 J10 T195 J10
T134 T29
T29 T165 T80 T80 T80 T70
T72 T72 T72 J4 LYC-2 LYC-1 DIL-C
T9 J2 J5 J5 J5 T197 T195 T208 T186
T186
T80
T69
J2
LV4 LV2 LV3
WBC RBC T210 T157 T157 Pressure T166
J7 J2 T29 T83 sensor
T28
T29 T197 J6 J6 J6 J6 T212 T14
J6
LV12 T163
T70 T70
J9
J6 J8
T29 T80 T172
T211 T10 T48 T48 T197
Check T10 T29
value J6 T80 T177
LV18 T80
J3
T80
J3
T80 LV15
J3 T170
T108 T70
T29
T11 T11 LV16 T80 T19
T65 T80 J3
J6 T65 T59 J4
T171 J3 J3
T29 T80
J3 T29
T172 T172
LV17 LV14 T80
LV19 LV13
J8 J8
T29
T195 waste
T162
P1
T164 J10 T80 T167 T168
T80 T168
27
2 Instrument Configuration
4. After cleaning and reassembling the swab, check and make sure the inflow/outflow interface
tubing is connected correctly.
5. Turn on the device.
6. go to Service > Maintenance > Clean. Click the Clean Sample Probe icon to perform the
operation before performing cleaning and maintenance on the swab.
28
2 Instrument Configuration
Figure 2-20 3-way valve (L) Figure 2-21 2-way valve (L)
4. Place the components in their original positions. (note that the parts with instructions are kept on
the same side during installation. Then remove the tissue paper and scrub and clean the
bottom.)
29
2 Instrument Configuration
Figure 2-22 Installation of the large Figure 2-23 Installation of the small
valve valve
3. Remove the impurities and place the pump in its original position.
Make sure to align the two marked lines correctly while installing, as shown in Figure 2-25.
Figure 2-25 Installing the marked lines
4. Make sure the liquid inlet and outlet are connected to the correct tubing (See Figure 2-26).
30
2 Instrument Configuration
2. Press the pinch valve of Vavle 12 with the fingers of one hand, and pull the plastic syringe in step
1 with the other hand, to absorb the probe solution into flow chamber. If the absorption resistance
is high, the plastic syringe shall be pulled with hand to hold for a certain time, and release to
enter into the probe cleanser stage after the level of the diluted probe cleanser in small sample
31
2 Instrument Configuration
When pushing the syringe, the pinch valve must be pressed tightly by hand, otherwise the cleaning
effects cannot be reached.
3. After soaking it for five minutes or so, restore the original tubing.
4. Click the error message area, and then click Remove Error in the pop-up dialog box to see if the
problem has been resolved. If the problem persists, keep taking the aforementioned steps until
the problem has been resolved.
32
2 Instrument Configuration
Apply proper force while pushing to prevent the tube from dropping and spilling (you can hold the
tube with your hand to keep it in place).
3. Flush back and forth ten times, and then let it settle for 3 minutes. Afterward, flush it three more
times with a syringe containing the pure diluent using the aforementioned method.
4. After the diluent cleaning, drain the liquid inside the WBC bath and reconnect the original tubing.
(To drain the RBC bath by interface operation, click the OK button in the prompt confirmation box
after draining.)
5. Click the error message area, and then click Remove Error in the pop-up dialog box to see if the
clogging problem is resolved.
6. If aperture clogging persists, redo the above procedures until the problem has been resolved.
33
2 Instrument Configuration
Name Quantity Note ECN No. Rev Design Check Approval Date
Three-way value Tow-way value
Value 19 3 1 2:COM 1
1:NC 2:IN
Pump 1
2 3:NO 2 1:OUT
Syring 4
Note:LV-liquid value;P-pump;T-pipe;J-connector
If the problem has been resolved, end the procedure; if the problem persists, continue to perform
the following steps:
4. Check whether or not the isolation chamber is airtight.
a. Empty the WBC bath. There are two ways of emptying the WBC bath:
Method I: after obtaining permission from qualified service personnel, go to the Service >
Maintenance > Maintain screen, click the Empty WBC Bath button (note that a prompt box
will pop up for confirmation after draining; do not press the OK button, otherwise the WBC
bath will be refilled with liquid).
Method II: the liquid can be aspirated dry manually using a plastic syringe or other tools.
b. Disassemble the isolation chamber below the WBC bath and pull out the corresponding
tubing. Use the thick No. 50 tube or 1.6mm (i.d.) silicone tube to attach the two connectors to
the side of the isolation chamber, plugging both apertures. Then attach the plastic syringe to
the connector above the isolation chamber using a 1.6mm (i.d.) silicone tube. Push the
plastic syringe to pressurize the isolation chamber, and hold the syringe steady for 30
seconds to check for any appearance of air bubbles. If any air bubbles occur, the isolation
chamber needs to be replaced. See Figure 2-30.
34
2 Instrument Configuration
5. After inspection, the components and tubing need to be restored to their original locations. Click
the OK button in the confirmation box that pops up after the counting bath is drained by opening
the user interface.
If the liquid level in the WBC bath is found to be too high, immediately shut down the device to
prevent any further overflowing.
35
2 Instrument Configuration
If the problem has been resolved, end the procedure; if the problem persists, continue to perform
the following steps:
4. Check if there is perforation at the bottom of the sample probe.
a. Set the small sample cup below the probe. When pulling out the tube above the probe (push
the tube end outward to pull it out; otherwise there will be crimping in the sampling tube
connected to the end of the probe), disassemble the probe, and connect it to the plastic
syringe filled with the diluent using a 1.6mm (i.d.) silicone tube. Manually push the syringe to
36
2 Instrument Configuration
see if any liquid flows from the bottom end of the probe.If such a flow is detected, this
indicates that perforation exists at the bottom of the probe (the liquid normally flows from the
sides. Be sure to collect the waste when pushing the liquid to the correct location, as shown
in the picture below).
Figure 2-33 Check the liquid pushing of the sample probe
b. Restore the tubing to its original status (cut off the crimped part at the front end of the
sampling tube, and readjust the tubing for installation. Make sure to check if the sampling
tubing gets into the way of other assemblies whenever the sampling assembly moves up and
down, or if it feels too tight. If this is the case, continue to adjust the sampling tubing until it
looks like it fits properly.)
5. Check for any wear and tear in the top aperture of the swab.
a. To disassemble, take off the swab's fixed jump ring.Pull out the swab's inflow and outflow
tubes.
b. Measure the upper-section aperture with a caliper; an aperture with a diameter of more than
1.71mm indicates that the swab needs to be replaced.
c. When reinstalling or replacing the swab, first put the inflow tube and outflow tube in place ,
then put on the swab (insert the probe into the swab aperture) and tap the swab's fixed jump
ring to complete the procedure.
The inflow tube is the Thin 50 tube, connected to the bottom connector; the outflow tube is the Thick
50 tube, connected to the top connector.
37
2 Instrument Configuration
T70 Sheath
pressure
T155 sensor
T152
J12 T150 T80 T80 T207
T70
T70
T29 T156
LV8 T209 LV6 LV9 J3
T70 T29
J11 T29 T29
T29 T70 T130 T40 LV11
T33 J3 T70J14 J14 T72 J2 T195 T192 J1 DILU DILU
T153 J3
T207 T80 T208 T70 T194 J2
T80 T207 LH
LV10 LV7 LV5 LV1 LH
J6 T207
DIFF I
T80 T165 T213 T206 T63 DIFF I T14
T79 T151
T158 J13 T70
T72 J2
T204
T204
J2 T199
T42 T70 T70
T5 T15
T89 T171 T115
T70 T205 T195 T13
T20 T29 J10 T195 J10
T134 T29
T29 T165 T80 T80 T80 T70
T72 T72 T72 J4 LYC-2 LYC-1 DIL-C
T9 J2 J5 J5 J5 T197 T195 T208 T186
T186
T80
T69
J2
LV4 LV2 LV3
WBC RBC T210 T157 T157 Pressure T166
J7 J2 T29 T83 sensor
T28
T29 T197 J6 J6 J6 J6 T212 T14
J6
LV12 T163
T70 T70
J9
J6 J8
T29 T80 T48 T172
T211 T10 T48 T29 T197
Check T10
LV18 J3 J3 T80
LV16 LV15
value J6 T80 T177 T80 T80
J3 T170
T108 T70 T11 T11
T29 T65 T80 T19
T65 T80 J3
J6 T171 T59 J4 J3 J3
T29 T80
J3 T29
T172 T172
LV17 LV14 T80
LV19 LV13
T195 J8 J8
T29
waste
T162
P1
T164 J10 T80 T167 T168
T80 T168
38
2 Instrument Configuration
T70 Sheath
pressure
T155 sensor
T152
J12 T150 T80 T80 T207
T70
T70
T29 T156
LV8 T209 LV6 LV9 J3
T70 T29
J11 T29 T29
T29 T70 T130 T40 LV11
T33 J3 T70J14 J14 T72 J2 T195 T192 J1 DILU DILU
T153 J3
T207 T80 T208 T70 T194 J2
T80 T207 LH
LV10 LV7 LV5 LV1 LH
J6 T207
DIFF I
T80 T165 T213 T206 T63 DIFF I T14
T79 T151
T158 J13 T70
T72 J2
T204
T204
J2 T199
T42 T70 T70
T5 T15
T89 T171 T115
T70 T205 T195 T13
T20 T29 J10 T195 J10
T134 T29
T29 T165 T80 T80 T80 T70
T72 T72 T72 J4 LYC-2 LYC-1 DIL-C
T9 J2 J5 J5 J5 T197 T195 T208 T186
T186
T80
T69
J2
LV4 LV2 LV3
WBC RBC T210 T157 T157 Pressure T166
J7 J2 T29 T83 sensor
T28
T29 T197 J6 J6 J6 J6 T212 T14
J6
LV12 T163
T70 T70
J9
J6 J8
T29 T80 T172
T211 T10 T48 T48 T29 T197
Check T10
LV18 J3 J3 T80
LV16 LV15
value J6 T80 T177 T80 T80
J3 T170
T108 T70 T11 T11
T29 T65 T80 T19
T65 T80 J3
J6 T171 T59 J4 J3 J3
T29 T80
J3 T29
T172 T172
LV17 LV14 T80
LV19 LV13
T195 J8 J8
T29
waste
T162
P1
T164 J10 T80 T167 T168
T80 T168
11. No Scattergram
Troubleshooting steps are as follows:
1. Check if the two tubes are connected correctly in pinch valve LV22. First confirm that they are
correctly positioned; then determine whether both soft tubes are tucked into the bottom of the
tank (as shown in Figure 2-36). (Note that the position of tubes marked in red in the following
figure shall not be installed reversely.)
39
2 Instrument Configuration
2. If the problem persists after the inspection, please check if the optical system is working properly.
For the common problems of the optical system and their solutions, see 2.4.6 Common
Problems and Solutions.
Hardware maintenance principles apply to all the panels. All the panels have passed hardware tests
before leaving the factory, and any hardware problem found later may be mainly caused by power
problems.Thus, hardware maintenance primarily involves the power supply. Hardware problems and
maintenance methods for each panel are described in the subsequent sections.
40
2 Instrument Configuration
Impedance
monitoring
Channel
channel
Optical
Other
External digital interface
Data
communication
ARM unit AD data acquisition part
PFGA unit
Motor
External interface
Valve, pump
Constant current
source control
Burning control
Laser control
HGB control
Pressure signal
Driver
Reagent testing
module
control
Micro-switch
Temperature control
The block diagram above gives a brief overview of the composition of main control panel and the
function of each module. Further description can be found below based on the pictures for each
panel. Picture of main control panel is shown in Figure 2-39.
41
2 Instrument Configuration
3 9
8
4
7
5 6
1 - External wiring interface 2 - Network interface and USB interface
The main control panel can be clearly divided into digital part, analog part and driver part. Among
them:
The left half of the panel is the digital part, including ARM unit, FPGA unit, external interface,
debugging interface, power supply, etc.
The right half of the panel is the analog part, including impedance channel, DIFF channel,
hardware monitoring channel, external interface, power supply, etc.
The lower left corner of the panel is the driver part, including motor driver, valve & pump driver,
temperature control, etc.
In the middle of the panel there are AD and interface chips, where AD is responsible for
analog-to-digital conversion and the interface chip is responsible for on-off control for relevant
functions of the analog part.
42
2 Instrument Configuration
10 J7 MPU_JTAG interface
21 J4 Serial port
22 J5 Serial port
43
2 Instrument Configuration
The debugging Interfaces designated for the main control panel are serial interface J7 (RS232
signals), FPGA JTAG interface J3, FPGA AS interface J4, and SD interface for programming before
shipping. Use of these debugging Interfaces is restricted to developers only.
2.3.1.3 Power input and indicator lights on the Main Control Panel
44
2 Instrument Configuration
NOTE:
1- D14 usually indicates the working status of the ARM power management chip,
TPS65910A31A1, but this indicator light only indicates the working status of one LDO signal
channel in TPS65910; the multimeter can be used to test the corresponding testing points for
other LDO and DC-DC channels. For details, see Digital Testing Points in the section of
2.3.1.4 Testing Points on the Main Control Panel.
+5V and -5V originate from the linear stabilizer. The corresponding lights will not be activated by the
+5V or -5V overload, and the linear stabilizer will be very hot. If either the +5V or the -5V power
indicator light is not on, immediately power off the device and troubleshoot the corresponding
electrical circuits. We recommend replacing the main control panel.
45
2 Instrument Configuration
D79 It indicates the ARM working status and flashes after Flashing
proper system loading.
D90 It indicates the FPGA working status and flashes after Flashing
proper system loading.
D52 It indicates the MPU working status and flashes after Flashing
proper system loading.
Indicator lights are very useful in practice; hardware problems can be identified by the status of the
indicator lights. For details, see 2.3.1.5 Identification of Main Control Panel Problems.
TP15 the testing point for 5V digital input, with the expected voltage of
5V
TP20 the testing point for 3.3V digital power supply, with the expected
voltage of 3.3V
TP19 the testing point for 2.5V digital power supply, with the expected
voltage of 2.5V
46
2 Instrument Configuration
TP17 the testing point for 1.2V digital power supply, with the expected
voltage of 1.2V
TP18 the testing point for 1.8V digital power supply, with the expected
voltage of 1.8V
TP31 SELECT_WBC_CTL testing point, the switch control signal for WBC
bath zap and CC source; 0 for zap and 1 for CC source
47
2 Instrument Configuration
TP29 SELECT_RBC_CTL testing point, the switch control signal for RBC
bath zap and CC source; 0 for zap and 1 for CC source
TP33 VCONST_CTL testing point, the switch control signal for CC source; 0
for on and 1 for off
48
2 Instrument Configuration
Figure 2-40 Digital power input, testing point and indicator light
49
2 Instrument Configuration
Functional Problems with the Main Control Panel and their Indicators
Functional problems with the main control panel are typically related to one of two causes. One
cause may be that the counting function of the main control panel cannot be realized. For example,
the digital part’s operating system on the main control panel is not operational, the FPGA program
does not load, or there is abnormal counting. The other cause refers to control failure of the motor,
50
2 Instrument Configuration
51
2 Instrument Configuration
Counting abnormalities
There are many causes that lead to counting abnormalities, and they can include problems with
main control panel problem. Counting abnormalities caused by main control panel problems
usually involve counting results of zero or overly high counting results on multiple occasions. A
counting result of zero usually results from a failure to apply the CC source to the counting bath,
while an overly high counting result is usually caused by too much noise in the analog-channel
circuit.
52
2 Instrument Configuration
normal.
2. Smell. Power problems may burn out some components; an acrid smell can suggest a burnout.
Do not power on in this case.
3. Touch. Power on and touch the corresponding component on the main control panel by hand.If
the temperature feels abnormal, the component could be broken.
4. Test. The first three steps can be used to identify power overload problems on the main control
panel. Step 4 is designed for determining what has malfunctioned. Testing can also identify the
problem of circuit breakage. This step is also part of maintenance. A multimeter is used to test for
power problems.
Common Problems Found in Main Control Panel Maintenance and their Indicators
See the following table for details.
Analogy power The power indicator light D19 is not This issue is mostly caused by the
input of -12V is not activated. fact that the analog power input of
loaded, and the -12V has shorted out. Another
device power possible reason is the breakdown
supply is switched of the polar capacitors (mostly
into auto protection tantalum capacitors here). Due to
mode when the abundance of -12V networks,
powered on. users are recommended to
replace the motherboard. The
damaged components can be
located by PCB for replacement
by qualified maintenance
personnel.
Analogy power The power indicator light D18 is not This issue is mostly caused by the
input of +12V is not activated. fact that the analog power input of
loaded, and the +12V has shorted out. Another
device power possible reason is the breakdown
supply is switched of the polar capacitors (mostly
into auto protection tantalum capacitors here). Due to
mode when the abundance of +12V networks,
powered on. users are recommended to
replace the motherboard. The
damaged components can be
located by PCB for replacement
by qualified maintenance
personnel.
53
2 Instrument Configuration
Analogy power The power indicator light D23 is not lit This issue is mostly caused by the
input of -5V is not up or looks dim, and the power chip U3 fact that the analog power input of
loaded feels very hot. -5V has shorted out. Another
possible reason is the breakdown
of the polar capacitors (mostly
tantalum capacitors here). Due to
the abundance of -5V networks,
users are recommended to
replace the motherboard. The
damaged components can be
located by PCB for replacement
by qualified maintenance
personnel.
Analogy power The power indicator light D25 is not lit This issue is mostly caused by the
input of +5V is not up or looks dim, and the power chip U2 fact that the analog power input of
loaded feels very hot. +5V has shorted out. Another
possible reason is the breakdown
of the polar capacitors (mostly
tantalum capacitors here). Due to
the abundance of +5V networks,
users are recommended to
replace the motherboard. The
damaged components can be
located by PCB for replacement
by qualified maintenance
personnel.
The ARM system The powered-on device fails to do The power management chip U11
on the main control self-test; the ARM status indicator light may be damaged; it is
panel is not running D79 is not flashing and the indicator recommended that a qualified
light for ARM power management chip maintenance person replace U11
D14 is lit up.
The ARM system The powered-on device fails to do Some circuit in the power
on the main control self-test; the ARM status indicator light management chip U11 may have
panel is not running D79 is not flashing and the indicator shorted out. Qualified
light for ARM power management chip maintenance personnel can use a
D14 is on. multimeter to test the power output
testing points throughout U11 to
locate the problematic circuit. Start
from the capacitor and move along
until locating the short circuit.
54
2 Instrument Configuration
The ARM system The powered-on device fails to do The boot media for the main
on the main control self-test; the ARM status indicator light control panel, NAND Flash U107,
panel is not running D79 is not flashing and the indicator may be damaged; replace the
light for ARM power management chip main control panel as soon as
D14 is on. A multimeter is used to test possible.
the power management chip only to
find that the power output is normal
throughout the chip and the device
lifetime can last for more than three
years.
The ARM system The powered-on device fails to do The ARM of the main control panel
on the main control self-test; the ARM status indicator light is damaged; replace the main
panel is not running D79 is not flashing and the indicator control panel immediately.
light for ARM power management chip
D14 is on. A multimeter is used to test
the power management chip only to
find that the power output is normal
throughout the chip and the ARM chip
U50 feels hot.
The FPGA program The FPGA running status indicator light, The FPGA on the main control
fails to load D90, is not flashing, the FPGA 3.3V panel is damaged; replace the
power indicator light D17 is in normal main control panel immediately.
ON status, and so is the FPGA 2.5V
power indicator light D16. The FPGA
1.8V power indicator light, D15, is in
normal ON status; test the FPGA 1.2V
with a multimeter and the voltage tested
at TP17 is normal.
The FPGA program The FPGA running status indicator light, The FGPA 3.3V power circuit is
fails to load D90, is not flashing, and the FPGA problematic and the issue can be
3.3V power indicator light D17 is not lit located by multimeter testing.
up.
The FPGA program The FPGA running status indicator light, The FGPA 2.5V power circuit is
fails to load D90, is not flashing, and the FPGA problematic and the issue can be
2.5V power indicator light D16 is not lit located by multimeter testing.
up.
The FPGA program The FPGA running status indicator light, The FGPA 1.8V power circuit is
fails to load D90, is not flashing, and the FPGA problematic and the issue can be
1.8V power indicator light D15 is not lit located by multimeter testing.
up.
The FPGA program FPGA running status indicator light, The FGPA 1.2V power circuit is
fails to load D90, is not flashing; test the FPGA problematic and the issue can be
1.2V power with a multimeter, and the located by multimeter testing.
voltage tested at testing point TP17 is
not 1.2V.
Liquid has been The main control panel is eroded. Replace the main control panel
spilled onto the
main control panel.
55
2 Instrument Configuration
The driver unit is not The hydraulic tubes have collapsed. Some valves' drivers may have
working properly. malfunctioned; it is suggested that
the panel be replaced as soon as
possible. If the problem persists,
check the driver path for valves,
since this problem is usually
caused by wiring issues or a
broken valve.
The driver unit is not The pump is not working. First confirm the reliability of the
working properly. connection between the driver
panel and the rear panel. If the
connection is reliable, then the
pump may have a problematic
driver; it is suggested that the
panel be replaced right away. If
the problem persists, check the
driver path for pumps, since this
problem is mostly caused by
wiring issues or a broken pump.
The driver unit is not The motor is not working. First confirm the reliability of the
working properly. connection between the driver
panel and the rear panel. If the
connection is reliable, then the
motor may have a problematic
driver; it is suggested that the
panel be replaced right away. If
the problem persists, check the
driver path for motors, since this
problem is mostly caused by
wiring issues.
The driver unit is not The driver panel has no heating. Some heating systems may have
working properly. a problematic driver; it is
suggested that the panel be
replaced right away. If the problem
persists, check the driver path for
heating, since this problem is
mostly caused by wiring issues.
56
2 Instrument Configuration
driver panel) will not be repaired in principle. If problems with any small panel are found on site,
please replace it directly.
Functions of interfaces on other panels are briefly introduced below.
1 2
57
2 Instrument Configuration
Figure 2-48 Physical picture of LS, MS Signal Figure 2-49 Physical picture of HS Signal
Amplifier Panel Amplifier Panel
58
2 Instrument Configuration
Purpose
The optical assembly can be disassembled and replaced by following the procedures specified in
this section; however, in each case, Preliminary Steps need to be taken before you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
Medical syringe attached to a silicone tube
Optical components that have passed tuning tests
Preliminary Steps
1. Turn off the power switch on the left side of the device.
2. Pull out the power cord plug from the rear panel of the device.
3. Open the right-side door and disassemble the top cover panel.
Disassembly
1. Use a #2 (Ph2) Phillips screwdriver to disassemble the four M3×8 cross-recessed pan-head
combination screws which affix the cover panel of the optical system, and carefully detach it.
2. Pull out the tubing. First pull out Tube P1 of Connector 2 in Figure 2-51 and connect the medical
59
2 Instrument Configuration
syringe with the silicone tube to Connector 2. Pull out Tube P6 from the L-type connector at the
top end of Connector 1. Drain the water from the sheath flow device with a syringe and pull out
Tube P21 and P4c from the Connector 3 and 4 at the bottom of the flow chamber.
Figure 2-51 Pulling out the tubing
1 - Tube P6 2 - Tube P1
60
2 Instrument Configuration
Installation
Follow the corresponding disassembly steps in reverse order.
Put on antistatic gloves and an electrostatic-shielding ring while operating to avoid damaging the
LED.
All the wirings need to be set up according to the original positions of the machine components
to prevent them from being crushed or otherwise damaged.
During transportation and installation, the flow chamber exposed at the bottom of the optical
system needs to be protected from being crushed or pressed.When pulling out the tubes, hold
the flow chamber by hand to prevent it from being dislocated by external force.
Purpose
The optical preamp panel can be disassembled and replaced by following the procedures specified
in this section; however, in each case, Preliminary Steps need to be taken before you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
New optical preamplifier panel
Preliminary Steps
1. Turn off the power switch on the left side of the device.
2. Pull out the power cord plug from the rear panel of the device.
3. Open the right-side door and disassemble the top cover panel.
4. Take down the cover panel of the optical system box.
Disassembly
1. Pull out the three wires from the optical preamp panel.
2. Use a #2 (Ph2) Phillips screwdriver to disassemble the four M3×8 cross-recessed pan-head
combination screws which affix the optical preamp panel, and carefully detach the cover panel of
the optical system box.
61
2 Instrument Configuration
Figure 2-53 Taking down the cover panel of the optical system box
Installation
Follow the corresponding disassembly steps in reverse order.
Put on antistatic gloves and an electrostatic-shielding ring while operating to avoid damaging the
LED.
All the wirings need to be set up according to the original positions of the machine components
to prevent them from being crushed or otherwise damaged.
Purpose
The laser driver panel can be disassembled and replaced by following the procedures specified in
this section; however, in each case, Preliminary Steps need to be taken before you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
2.5mm hex-socket screwdriver
New laser driver panel
Preliminary Steps
1. Turn off the power switch on the left side of the device.
62
2 Instrument Configuration
2. Pull out the power cord plug from the rear panel of the device.
3. Open the right-side door and disassemble the top cover panel.
Disassembly
1. Use a 2.5mm hex-socket screwdriver to disassemble the four M3×6 stainless-steel inner hex
screws that affix the baffle plate of the laser driver panel.
2. Pull out the two wires from the laser driver panel.
3. Pull out the two wires from the laser driver panel. Use a #2 (Ph2) Phillips screwdriver to
disassemble the four M3×8 cross-recessed pan-head combination screws, and carefully detach
the laser driver panel.
Figure 2-54 Replacing the Laser Driver Panel
1 - M3×6 stainless-steel inner hex 2 - Baffle plate of the laser driver panel
screw (×4)
Installation
Follow the corresponding disassembly steps in reverse order.
63
2 Instrument Configuration
Put on antistatic gloves and an electrostatic-shielding ring while operating to avoid damaging the
LED.
All the wirings need to be set up according to the original positions of the machine components
to prevent them from being crushed or otherwise damaged.
Purpose
The fine tuning of the flow chamber can be performed by following the procedures specified in this
section; however, in each case, Preliminary Steps need to be taken before you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
Preliminary Steps
Open the right-side door and disassemble the top cover panel.
Take down the cover panel of the optical system box.
Turn on the device and launch the software application; then enter the Optical Tuning screen.
Tuning Procedures
1. Use a #2 (Ph2) Phillips screwdriver to loosen the two M3×8 cross-recessed pan-head
combination screws that affix the locking plates of the flow chamber.
Figure 22-55 Fine tuning of the flow chamber
64
2 Instrument Configuration
testing.
If the resultant scattergram shows a square or diamond with the upper and lower sides
parallel to each other, the device does not need fine-tuning.
If the scattergram shows that the upper and lower sides are not parallel, then take the next
step.
3. If the slanting directional arrow shown in the scattergram is ↗, then rotate the adjustment knob
of the flow chamber counter-clockwise; if the slanting direction shown in the scattergram is ↘,
then rotate the adjustment knob of the flow chamber clockwise (the adjustment knob should be
lightly rotated).
4. Use the 7um nominal particles as a sample for DIFF testing.
If the resultant scattergram shows a square or diamond with the upper and lower sides
parallel to each other, the device does not need fine-tuning.
If the scattergram shows that the upper and lower sides are not parallel, then repeat Step 3.
5. After tuning, tighten the two locking screws of the flow chamber, and reinstall the cover panel of
the optical system box as well as the four affixing screws; then lock them in place.
Purpose
The high-angle receiving plate can be disassembled and replaced by following the procedures
specified in this section; however, in each case, Preliminary Steps need to be taken before you
begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
2.5mm hex-socket screwdriver
New high-angle receiving plate
Preliminary Steps
1. Turn off the power switch on the left side of the device.
2. Pull out the power cord plug from the rear panel of the device.
3. Open the right-side door and disassemble the top cover panel.
Disassembly
1. Disassemble 2 M3 ×8 stainless steel internal hexagon screws used for fixing the fixed seat of the
side panel card by 2.5mm internal hexagonal wrench, and then remove the fixed seat from the
side basal plate.
2. Pull out 2 connecting wires on the high-angle receiving plate.
3. Disassemble 2 M2.5 ×6 cross recess head screws used for fixing the high-angle receiving plate
by #2 (Ph2) Philips screwdriver, and remove the laser driver plate carefully.
65
2 Instrument Configuration
1 - M3×6 inner hex screw (×2) 2 - Fixed seat of side panel card
66
2 Instrument Configuration
There are only sporadic The flow chamber Check if the collecting corner plate has two
background scattergram is dislocated. bilaterally symmetric light spots in the shape of red
signals when testing the The hydraulics vertical stripes.
DIFF signal of venous have no sheath If there is only one red vertical stripe or the two
blood. flow. stripes are significantly different in brightness,
make small adjustment to the location of the
flow chamber and return the nominal-particle
scattergram to normal.
If there are two red vertical stripes which are
not significantly different in brightness, check if
the pinch valve is working; if it is, check if Valve
10, 11, 12, 13, and 18 are also working.
The neutral-particle signal The flow chamber Make a small adjustment to the location of the flow
(the long horizontal stripe is dislocated. chamber and return the nominal-particle
at the top right corner) in scattergram to normal.
the scattergram is
considerably slanted and
the three signals are
mostly overlapping with
inaccurate classification
when testing the DIFF
signal of fresh venous
blood.
There is a lot of noise The reagent is If the background noise has a certain shape
when testing the contaminated. and is concentrated at the lower half of the
background signal. Tiny air bubbles scattergram, keep the DIFF1 and DIFF2 fluids
get into the at room temperature and perform three rounds
hydraulics. of the reagent replacement sequence each for
the diluent, DIFF1 and DIFF2.
If the background signal has no shape and
randomly spreads all over the scattergram,
replace the sheath flow syringe.
If the problem persists, replace the optical assembly.
67
3 Repairs
3 Repairs
3.1 Overview
The repair engineer can repair the analyzer using standard tools. See the following section for
repairing procedures (including the tools needed). If any repairing step requires a validating step, the
repair engineer should strictly follow the procedure and take the validating step.
When following the repairing procedure as specified in this section, the operator should put on
rubber gloves and clean both hands with disinfectant after performing any repair work.
Purpose
While performing regular machine maintenance and inspection, the left-side door needs to be
opened to disassemble the sample syringe assembly, reagent testing panel, or electromagnetic
valve assembly, which are all in the left section of the machine.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
Opening
1. As shown in Figure 3-1, use #2 (Ph2) cross screw driver to take out four M3×8 cross recessed
pan-head combination screws on the back panel, and take off the back panel.
68
3 Repairs
2. As shown in Figure 3-2, use #2 (Ph2) cross screw driver to take out three M3×8 cross recessed
pan-head combination screws on the left side panel, and take off the left side panel.
Figure 3-2 Disassembling the left side panel
69
3 Repairs
Purpose
While performing machine maintenance and inspection, the right side door needs to be opened to
disassemble the preheating bath assembly, pinch valve, sampling assembly, liquid pump assembly,
impedance counting reaction bath assembly (WBC & RBC), negative-pressure chamber, or
electromagnetic valve & pump assembly, which are all in the right section of the machine.
Tools/Spare Parts
Slot-type screwdriver
Opening
As shown in Figure 3-3, insert the slot-type screwdriver into the slot in the right side door lock and
rotate 90° counterclockwise, then manually open the right side door.
Figure 3-3 Opening the right side door
Purpose
While performing regular machine maintenance and inspection, the panel cover needs to be taken
down for disassembling the sample probe, cleaning swab, sampling assembly, or indicator light
panel.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
70
3 Repairs
Preliminary Steps
1. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
2. The left side panel has been disassembled. See 3.2.1 Disassembling the Left Side Panel.
3. The right-side door is opened. See 3.2.2 Open the Right Side Door.
Disassembly
1. Disassembly the back panel: use a #2 (Ph2) Phillips screwdriver to disassemble the three M3×6
cross-recessed pan-head combination screw on the back panel, and take off the back panel, as
shown in Figure 3-4.
Figure 3-4 Disassembling the back panel
2. Diassembling top cover: Open the right side door and disassemble 4 M3×8 cross-recessed
countersunk-head screws on both sides of top cover by #2 (Ph2) Philips screwdriver, the top
cover can be removed, as shown in Figure 3-5.
71
3 Repairs
When removing the panel cover, make sure to pull out the wires connected to the display circuit
panel on the panel cover.
72
3 Repairs
Purpose
The display screen can be disassembled and replaced by following the procedures specified in this
section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before you
begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
Display screen
Disassembly
1. Pull out both interfaces at the shield panel of the LCD screen interface panel, use #2 (Ph2) cross
screw driver to remove four M3×8 cross recessed pan-head combination screws on the shield
panel, take off the shield panel, then detach the black buckles on both ends, gently pull out the
golden ribbon cable, as shown in Figure 3-7.
Figure 3-7 Replacement of display screen (1)
73
3 Repairs
2. Use #2 (Ph2) cross screw driver to remove eight M3×8 cross recessed pan-head combination
screws on the screen support, take off the screen support, display screen and touch screen
respectively, as shown in Figure 3-8.
Figure 3-8 Replacement of display screen (2)
Purpose
The sample probe can be disassembled and replaced by following the procedures specified in this
section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before you
begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
Sample probe
Disassembly
1. Pull out the swab's snap ring to detach it from the lower bracket panel. See Figure 3-9.
74
3 Repairs
75
3 Repairs
Purpose
The optocoupler can be disassembled and replaced by following the procedures specified in this
section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before you
begin.
Tools/Spare Parts
2.5mm hex-socket screwdriver
Optocoupler
Disassembly
Use the 2.5mm hex wrench to remove the M3 hex screw affixing the optocoupler, then pull out the
plug attached to the optocoupler wire to disassemble the optocoupler to be replaced. See Figure 3-
11.
Figure 3-11 Replacement of optocoupler
Purpose
The sampling assembly can be disassembled and replaced by following the procedures specified in
this section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before
you begin.
76
3 Repairs
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
diagonal cutting pliers
Disassembly
1. Use the diagonal cutting pliers to cut off all the nylon binding tapes affixing the fluid tubes
(aspiration tubes of the sample probe and swab tubes), then pull out the fluid tubes of the sample
probe and the swab to detach them from the sampling assembly.
2. Pull out all the motor wires and optocoupler plug from the horizontal moving module of the
sample assembly, and use a #2 (Ph2) Phillips screwdriver to take out the three M3×6 cross
recessed countersunk head screws affixing the bracket for drag chains to detach the sampling-
specific drag chain from the sampling assembly. See Figure 3-12.
Figure 3-12 Replacement of the sampling assembly in X- or Y-direction (1)
77
3 Repairs
Tools/Spare Parts
Phillips screwdriver
Power assembly with the same specifications
Disassembly
1. Use Phillips screwdriver to take out four M3×8 cross recessed pan-head combination screws on
the back of the device power supply assembly, then remove one M3×8 cross recessed pan-head
combination screw on the front, as shown in Figure 3-14.
78
3 Repairs
2. Unplug all the connectors attached to the lead wire of the power assembly.
3. Take off the power supply required to be replaced from the power supply assembly, and replace
with a new power supply, fix it with screws.
Purpose
The valve assembly can be disassembled and replaced by following the procedures specified in this
section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before you
begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
The replacement valve with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device.
2. Pull out the power cord plug from the rear panel of the device.
3. If it is only the electromagnetic valve on the left or right clapboard that needs to be replaced,
79
3 Repairs
open the corresponding left- or right-side door; if the one on the front panel needs to be replaced,
remove the front panel cover by following the instructions in 3.2.3 Disassembling the Panel
Cover.
Disassembly
1. Disassemble the peripheral fluid tubes connected to the valve assembly.
2. Use a #2 (Ph2) Phillips screwdriver to disassemble the two M3×8 cross-recessed pan-head
combination screws which affix the valve assembly, and carefully take out the valve assembly
while making sure to unplug the attached wires. See Figure 3-15.
Figure 3-15 Replacement of valve assembly
Installation
Make sure to the use correct model of valve and create a reliable connection
All the wirings need to be set up according to the original positions of the machine components
to prevent them from being crushed or otherwise damaged.
This machine involves a great number of valves. Only one valve is discussed as an example
here, and all the other valves follow the same maintenance procedures.
Purpose
The liquid pump assembly can be disassembled and replaced by following the procedures specified
80
3 Repairs
in this section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken
before you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
2.5mm hex-socket screwdriver
The replacement liquid pump with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device.
2. Pull out the power cord plug from the rear panel of the device.
3. Open the right side door.
Disassembly
1. Pull out the peripheral fluid tubes and wire connectors attached to the liquid pump.
2. Use a #2 (Ph2) Phillips screwdriver to loosen the two M4×10 cross-recessed pan-head
combination screws which affix the liquid pump, and move the liquid pump assembly to the left till
the big hole on the affixing panel of the liquid pump is aligned to the affixing screw, then carefully
hold up the liquid pump assembly to take it out. See Figure 3-16.
Figure 3-16 Replacement of liquid pump assembly (1)
3. The further dismantling of the liquid pump assembly is shown in Figure 3-17.
81
3 Repairs
Installation
All the wirings need to be set up according to their original positions to prevent them from being
crushed or otherwise damaged, and to prevent the working liquid pump from being shaken,
which affects the fluid tubes.
Pay special attention to the tubing joints and ensure that the connections are sound.
Purpose
The negative-pressure chamber assembly can be disassembled and replaced by following the
procedures specified in this section; however, in each case, 3.2 Preparatory Work Before Repairs
need to be taken before you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
2.5mm hex-socket screwdriver
The replacement parts or assembly of the negative-pressure chamber with the corresponding
specifications
82
3 Repairs
Preliminary Steps
Open the right side door.
Disassembly
1. Disassemble the peripheral fluid tubes connected to the negative-pressure chamber.
2. Use a #2 (Ph2) Phillips screwdriver to disassemble the two M4×10 cross-recessed pan-head
combination screws which affix the negative-pressure chamber, and carefully take out the
negative-pressure chamber assembly. See Figure 3-18.
Figure 3-18 Replacement of negative-pressure chamber assembly
Installation
Make sure to place the seal ring into the seal tank to keep the pressure chamber airtight.
All the tubing needs to be set up according to the relevant standards or the original positions of
the machine components to prevent them from being crushed or otherwise damaged.
83
3 Repairs
Purpose
The syringe can be disassembled and replaced by following the procedures specified in this section;
however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before you begin.
Tools/Spare Parts
2.5mm hex-socket screwdriver
The replacement syringe assembly with the same specifications
Preliminary Steps
Open the left side door.
Disassembly
1. Pull out the peripheral fluid tubes connected to the syringe assembly to be replaced.
2. Use the 2.5mm hex-socket screwdriver to disassemble the two M3×8 stainless-steel inner hex
screws affixing the syringe assembly. Then take out the syringe assembly. See Figure 3-19.
Figure 3-19 Replacement of syringe
2
1
Installation
Follow the corresponding disassembly steps in reverse order.
84
3 Repairs
Tools/Spare Parts
2.5mm hex-socket screwdriver
#2 (Ph2) Phillips screwdriver
The replacement syringe assembly with the same specifications
Preliminary Steps
Open the left side door.
Disassembly
1. Pull out the peripheral fluid tubes connected to the syringe module to be replaced.
2. Use #2 (Ph2) cross screw driver to remove four M3×8 cross recessed pan-head combination
screws, take off the syringe assembly. See Figure 3-20.
Figure 3-20 Replacement of motor (1)
1
2
3. Use the 2.5mm hex-socket screwdriver to disassemble the two M3×8 stainless-steel inner hex
screws affixing the syringe module. Then take out the syringe. See Figure 3-19.
4. Take out the two M3×8 inner hex screws on the shield cover, then remove the white bolt from the
motor. Next, remove the four M3×8 inner hex screws from the back of the motor for the
replacement. See Figure 3-21.
85
3 Repairs
Installation
Follow the corresponding disassembly steps in reverse order.
Purpose
The WBC counting bath assembly can be disassembled and replaced by following the procedures
specified in this section; however, in each case, Preliminary Steps needs to be taken before you
begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
2.5mm hex-socket screwdriver
The replacement WBC counting bath assembly with the same specifications
Preliminary Steps
1. Run the draining sequence of the software to drain any residual liquid inside the machine.
2. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
3. Open the right side door.
86
3 Repairs
Disassembly
1. Use a #2 (Ph2) Phillips screwdriver to disassemble the M3×8 cross-recessed pan-head
combination screw in the lower section of the shield cover to remove the cover; to take out the
screw, move slightly upwards for 3mm or so and apply force in a direction perpendicular to the
right clapboard. See Figure 3-22.
Figure 3-22 Dismantling and replacing the WBC bath assembly (1)
87
3 Repairs
4. Use 2.5mm hex wrench to remove the two M3×8 stainless-steel hex-socket screws affixing the
upper section of WBC counting bath assembly. See Figure 3-24.
Figure 3-24 Dismantling and replacing the WBC bath assembly (3)
1 - WBC counting chamber assembly 2 - M3×8 stainless-steel inner hex screw (×2)
5. Carefully and slowly take out the WBC counting bath assembly.Pay attention to the wiring in the
process of moving to avoid breaking the wires.
6. Remove the protective ring from the shield case and open the wire clip to take out the inner wire
along the stretching direction of the wire.Then take out the WBC counting bath assembly as a
whole.
7. The further dismantling of the WBC counting chamber assembly. See Figure 3-25.
Figure 3-25 Dismantling and replacing the WBC bath assembly (4)
88
3 Repairs
Installation
All the wirings need to be set up according to the working procedures or the original positions of
the machine components to prevent them from being crushed or otherwise damaged.
Pay special attention to the tubing joints and ensure that the connections are sound.
The aperture, seal gasket, and seal ring need to be correctly positioned and oriented in the right
direction.
Purpose
The RBC counting bath assembly can be disassembled and replaced by following the procedures
specified in this section; however, in each case, 3.2 Preparatory Work Before Repairs need to be
taken before you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
2.5mm hex-socket screwdriver
The replacement WBC counting bath assembly with the same specifications
Preliminary Steps
1. Run the draining sequence of the software to drain any residual liquid inside the machine.
2. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
3. Open the right side door.
Disassembly
1. Use a #2 (Ph2) Phillips screwdriver to disassemble the M3×8 cross-recessed pan-head
combination screw in the lower section of the shield cover to remove the cover; to take out the
screw, move slightly upwards for 3mm or so and apply force in a direction perpendicular to the
right clapboard.
Refer to the corresponding section in 3.8.1 Dismantling and replacing the WBC bath assembly.
2. Pull out the peripheral fluid tubes attached to the RBC counting bath assembly.
89
3 Repairs
Pay attention to handling the residual fluid; try to prevent it from flowing onto the other
components in the machine, and wipe clean if necessary.
3. Use a #2 (Ph2) Phillips screwdriver to disassemble the M3×8 cross-recessed pan-head
combination screw for grounding on the middle clapboard of the shield case cover and pull out
the grounding wire. See Figure 3-26.
Figure 3-26 Dismantling and replacing the RBC bath assembly (1)
1 - RBC counting chamber assembly 2- M3×8 stainless-steel inner hex screw (×2)
90
3 Repairs
5. Carefully and slowly take out the RBC counting bath assembly. Pay attention to the wiring during
the process of moving to avoid breaking the wires.
6. Remove the protective ring from the shield case and open the wire clip to take out the inner wire
along the stretching direction of the wire. Then take out the RBC counting bath assembly as a
whole.
7. The further dismantling of the RBC counting chamber assembly. See Figure 3-28.
Figure 3-28 Dismantling and replacing the RBC bath assembly (4)
Installation
All the wirings need to be set up according to the working procedures or the original positions of
the machine components to prevent them from being crushed or otherwise damaged.
Pay special attention to the tubing joints and ensure that the connections are sound.
The aperture, seal gasket, and seal ring need to be correctly positioned and oriented in the right
direction.
91
3 Repairs
Tools/Spare Parts
Phillips screwdriver
Corresponding specifications needed to replace the main control panel
Preliminary Steps
1. Pull out the power cord plug from the rear panel of the analyzer, and turn off the power switch.
2. Disassemble the screws on the back panel of housing and remove the back panel of the
housing.
Disassembly
1. Pull out the peripheral wires and connectors attached to the main control panel.
2. As shown in Figure 3-29, disassemble four M4×8 cross recessed pan-head combination screws
fixing the main control panel, and remove the main control panel.
Figure 3-29 Replacement of the main control panel
1
2
92
3 Repairs
Installation
Follow the corresponding disassembly steps in reverse order.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
The replacement reagent testing panel with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
2. Open the left side door.
Disassembly
1. Remove the four M3×8 cross recessed pan-head combination screws affixing the protective
cover of the reagent testing panel and remove the protective cover. See Figure 3-30.
Figure 3-30 Replacement of reagent testing panel (1)
1 2
93
3 Repairs
2. Pull out all the exposed peripheral wires connected to the reagent testing panel.
3. Remove the four M3×8 cross-recessed pan-head combination screws affixing the reagent testing
panel and carefully take out the reagent testing panel. Make sure that the metal parts will not
scratch the wiring at the rear of the panel. Pull out the rear wiring to remove the reagent testing
panel. See Figure 3-31.
Figure 3-31 Replacement of reagent testing panel (2)
Installation
All the wirings need to be set up according to the working procedures or the original positions of
the machine components to prevent them from being crushed or otherwise damaged.
Pay attention to the position of wires when installing the sensor panel so as to prevent the wires
from breaking.
94
3 Repairs
Purpose
The temperature sensor can be disassembled and replaced by following the procedures specified in
this section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before
you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
2 hex-socket screwdriver
The replacement temperature sensor with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
2. Open the left side door and the front panel cover.
Disassembly
Loosen the screws in the bracket for temperature sensor to take the sensor out. Then replace it with
a new sensor and tighten the screws. See Figure 3-32.
Figure 3-32 Disassembling the temperature sensor
95
3 Repairs
Installation
Follow the corresponding disassembly steps in reverse order.
Purpose
The temperature sensor can be disassembled and replaced by following the procedures specified in
this section; however, in each case, 3.2 Preparatory Work Before Repairs need to be taken before
you begin.
Tools/Spare Parts
#2 (Ph2) Phillips screwdriver
2 hex-socket screwdriver
The replacement temperature sensor with the same specifications
Preliminary Steps
1. Turn off the power switch on the left side of the device and pull out the power cord plug from the
back panel of the device.
2. Open the left side door and the front panel cover.
Disassembly
Unplug the wire stock and tubing connected to the pressure sensor, and take the pressure sensor
out. Then replace it with a new sensor. See Figure 3-33.
96
3 Repairs
2
2
1
3
3 - Sheath-flow pressure
sensor connecting wire
Installation
Follow the corresponding disassembly steps in reverse order.
97
4 Software Upgrade
4 Software Upgrade
4.1 Preparation
1. Get officially provided upgrade package, and prepare a USB flash disk for storing the upgrade
package.
2. Copy the upgrade package to the root directory of the USB flash disk (Do not paste it into any
newly-created folder).
3. Insert the USB flash disk with upgrade package into the USB interface of the host.
4. Use service authority to log in to the host.
2. Upgrading Steps
1. Enter the Service interface, and click the Upgrade button, as shown in the picture below.
98
4 Software Upgrade
The host will detect the USB flash disk. 2~5 seconds later, if the USB flash disk cannot be
recognized, there will be corresponding prompt in the interface, please insert the USB flash disk
again; if the USB flash disk can be recognized, there will be dialog box in the interface as follows.
99
4 Software Upgrade
There will be prompt with the upgrading progress on the screen. If the upgrade is successful, the
progress bar will display 100%. See the picture below.
4. Turn off the power supply of the analyzer, and restart the analyzer.
5. Use a hard object, such as a toothpick or pencil to click the focal points successively appearing
on the screen (5 points in total).
After clicking 5 focal points, the touch screen calibration is complete.
6. Turn off the power supply of the analyzer, and restart the analyzer.
100
5 Comprehensive Device Tuning
Since the replacement and maintenance of some components can lead to changes in relevant tuning
parameters, this section introduces the tuning procedures for the parameters that may be affected.
1. Position Adjustment
NOTE: The swab height needs to be reset after the replacement of the sampling assembly and
relevant parts of the sampling swab.
After the replacement of the sampling assembly or swab, perform the following tuning procedures.
1. Enter the Service screen and click Position Adjustment.
2. Click Start and check the distance between the bottom of the sample probe and the bottom of
the swab. If the sample probe is lower than the swab, click Up; if higher, click Down. Each click
will move the sample probe slightly.In the end, the bottom of the sample probe will be aligned
with the bottom of the swab. Then click "OK".
3. Next, click Check to run the initialization of the sample probe. Check if the bottom of the sample
probe is still aligned with the bottom of the swab (see Figure 5-1). If it is, click OK to complete the
tuning; if not, repeat Steps 1~2.
101
5 Comprehensive Device Tuning
102
5 Comprehensive Device Tuning
3. Gain Calibration
NOTE: Gain calibration needs to be performed for the replacement of RBC bath and/or main control
panel.
1. Click Service > Gain cal. to access the gain calibration screen. See Figure 5-3.
Figure 5-3 Gain calibration
2. Fill in the cell corresponding to MCV with the MCV reference value for quality control.
3. Perform the QC test for three times in a row for the first run. The results for each time will be
automatically displayed.
If the CV falls within reasonable parameters, the screen will show the CV and Gain values for
the first run. Go to step 4.
If it does not fall within reasonable parameters, you’ll be prompted to redo the calibration.
Please click OK to close the message box. Then click Clear to delete the data, and repeat
the step 3.
4. Perform the QC test for three times in a row for the second run. The results for each time will be
automatically displayed.
If the CV falls within reasonable parameters, the screen will show the CV and Gain values for
the second run, and show the final results. Please click "OK" to complete gain calibration.
If it does not fall within reasonable parameters, you’ll be prompted to redo the calibration.
Please click OK to close the message box. Then click Clear to delete the data, and repeat
the step 4.
4. Calibration of Calibrators
NOTE: The calibrators need calibration for the replacement of the WBC bath, the RBC bath, and/or
the main control panel.
103
5 Comprehensive Device Tuning
3. Input the calibrator lot No. JZQX-01 in the text box for Lot No., and click the Exp. Date control to
set the expiry date of the calibrator.
4. Input the target value of the current calibrator parameter.
The calibration mode is Whole Blood by default, so there is no need to set the mode.
5. Set the well-shaken calibrator under the sample probe, then press the aspiration key on the
analyzer to start the calibration counting.
6. Repeat step 5 for a total of 12 times to get 12 results of calibration counting.
After the counting is complete, a dialog box will pop up indicating that the test is complete.
Click the Save button to save the calibration result.
If significant differences are found across results, a dialog box for data abnormality will pop
up.Please redo the calibration.
104
5 Comprehensive Device Tuning
5. LIS Connection
If the analyzer needs to be connected to laboratory information system (hereinafter referred to as
LIS), you can complete the connection by following the steps in this section.
105
5 Comprehensive Device Tuning
4. Set the IP address and other network information of the analyzer according to the actual
situation.
If the network is accessed through a router on the site, please select Obtain an IP address
automatically and Obtain DNS server address automatically.
If the network is accessed through a network switch, or the analyzer is directly connected to
the LIS on the site, please select Use the following address, so as to manually set the IP
address and subnet mask of the analyzer. The IP addresses of the analyzer and LIS must be
in the same network segment. Furthermore, their subnet masks shall be the same, while
other parameters can maintain null.
5. Click OK to save the settings and close the dialog box.
3. Input the IP address and port of LIS workstation in Network Settings area.
Find the IP address and port of LIS in the network setup interface in the LIS workstation; if IP
address can’t be found, try the method below:
106
5 Comprehensive Device Tuning
The IPv4 address in the red box is the IP address of LIS workstation.
The IP address 192.168.8.44 of the LIS workstation shown as above is used as an example, real IP
should be in the same network segment with LIS server.
The LIS icon in the upper right side on the analyzer screen turns from gray to black ,
which indicates auto hematology analyzer software is connected to LIS successfully.
If the icon stays gray, the connection fails. Please check if the IP address and port of LIS is
correct and reconnect as the steps above; if the problem still exists, please contact the hospital
network administrator to handle it.
107
6 Alarms and Solutions
This section introduces error messages that can appear in the analyzer, possible causes, and
troubleshooting steps to be taken by the operator. If the problem persists after troubleshooting,
please take hardware issues into account and consider replacing the relevant parts or panels.
For the following issues, please click the error message box at the bottom right corner of the
software interface, then click "Remove Error" in the popup dialog box. Usually the problem will be
automatically resolved; if it persists, refer to the "Solution" column for further maintenance.
The syringe fails to This syringe problem can occur while it is being
leave its initial moved.Please refer to the following solution:
position. 1. Follow the instructions in 3.7 Sheath Flow Syringe
Syringe fails to return Assembly Replacement to disassemble the syringe,
to its initial position. then remove the dust cover and optocoupler. Plug
the optocoupler into the connector of the syringe
The syringe takes too optocoupler. The user enters the Status > Sensor
many steps to return screen and covers the center of the optocoupler with
Syringe to its initial position. a piece of paper. Check if the optocoupler status
1
problem shown in the screen is blocked; if yes, then the
The sample syringe is optocoupler is working properly.
busy.
2. Follow the instructions in 3.7.2 Replacing the Motor
to disassemble the syringe motor and replace it with
a new motor. Then go to syringe self-test under
Sample syringe Service > Self-test > Syringe; if the syringe is
timeout working, then the maintenance has been successful.
The horizontal motor 1. The user enters the Status > Sensor screen and
fails to leave its initial covers the center of the optocoupler with a piece of
position. paper. Check if the optocoupler status shown in the
screen is blocked. If it is, then the optocoupler is
The horizontal motor working properly; if not, then refer to 3.4.2 Replacing
fails to return to its the Optocoupler on how to replace the optocoupler.
Horizontal initial position.
2. Refer to 3.4.3 Replacing the Sampling Assembly
2 motor
The motor fails to in X- or Y-direction to disassemble the sampling
problem
move to the WBC assembly and make sure the motor wiring is secure.
position. If the wiring is OK, then disassemble the motor and
replace it with a new one. After the installation, make
The motor fails to proper adjustments to the belt's tension, then go to
move to the RBC the syringe self-test under Service > Self-test >
position.
108
6 Alarms and Solutions
Horizontal motor
timeout
The optocoupler of
the horizontal motor is
not working properly.
The vertical motor 1. The user enters Status > Sensor screen and covers
fails to leave its initialthe center of the vertical optocoupler with a piece of
position. paper. Check if the corresponding optocoupler status
shown in the screen is blocked. If it is, then the
The vertical motor optocoupler is working properly; if not, then refer to
fails to return to its 3.4.2 Replacing the Optocoupler on how to replace
initial position. the optocoupler.
The motor fails to 2. Refer to 3.4.3 Replacing the Sampling Assembly
move into position to in X- or Y-direction to disassemble the motor from
isolate the air bubbles. the sampling assembly and make sure motor wiring
is secure. If the wiring is OK, then disassemble the
motor and replace it with a new one. After the
The motor fails to installation, make the proper adjustments to the belt
Vertical move to the DIFF tension, then go to the syringe self-test under
3 motor bath position. Service > Self-test > Syringe. If the syringe is
problem working, then it has been successfully fixed.
The motor fails to
move to the counting
bath position.
109
6 Alarms and Solutions
The 12V +12V power is not Refer to 3.9 Main Control Panel Replacement to
power is not working properly. replace the main control panel.
6 working -12V power is not
properly working properly.
110
6 Alarms and Solutions
Abnormal pressure of
positive-pressure
chamber (lower than
normal)
Abnormal pressure of
positive-pressure
chamber (higher than
P r e s s u r e normal)
9 c h a m b e r The Refer to 2.2.12.10 Problems in Creating the
problem negative-pressure Negative Pressure for troubleshooting.
chamber fails to
create pressure
Abnormal pressure of
negative-pressure
chamber (lower than
normal)
Abnormal pressure of
negative-pressure
chamber (higher than
normal)
111
6 Alarms and Solutions
LYC-2 expired.
Insufficient LYC-1.
DIL-C running out. Check for the availability of the relevant reagent. If the
reagent is available, then check if the optocoupler of
15 No reagent LYC-1 running out or the reagent testing panel is on. If not, adjust the screw
air bubbles in inlet of the potentiometer and rotate counterclockwise to
tubing.
112
6 Alarms and Solutions
A B C
A1 B1 C1
113
7 Maintenance Inventory
7 Maintenance Inventory
114
7 Maintenance Inventory
33 32010006A Power
115
7 Maintenance Inventory
116
P/N: 65.02.0120A (5.0)