Recovery of Semivolatile Organic Compounds During Sample Preparation: Implications For Characterization of Airborne Particulate Matter
Recovery of Semivolatile Organic Compounds During Sample Preparation: Implications For Characterization of Airborne Particulate Matter
Recovery of Semivolatile Organic Compounds During Sample Preparation: Implications For Characterization of Airborne Particulate Matter
Hex:MeCl2:MeOH solvent and wash the walls of the glassware, recoveries were determined using the n-alkane-PAH mixture
to redissolve the nonpolar organics that have precipitated with 10 times lower concentrations than the original. The
during the reduction of volume step. recoveries were still less than 100% for n-alkanes > nC32. The
For the higher MW n-alkanes with <100% recoveries results suggest that the concentrations of the higher MW
(>nC32), lower concentrations could lead to higher recoveries n-alkanes were low but not enough to redissolve completely
because of less precipitation from the solution. To test this, in the solvent mixture. We are finding that most of the extracts
FIGURE 3. Percent recovery of n-alkanes (a) and PAHs (b) from various reduction of volume procedures. A reduced extract only (0.5 mL)
is shown as open squares; the same 0.5 mL reduced extract with addition of 1 mL of Hex:MeCl2 is shown as open diamonds; a reduced
extract with the Labconco glassware rinsed with 1 mL MeOH is shown as open triangles; a reduced extract with the Labaconco glassware
rinsed with 1 mL of Hex:MeCl2 are shown as closed circles. The open circle markers (at carbon 24.5) show the recovery of the laboratory
surrogate nC24D50. The error bars represent the uncertainty at the 1σ level.
from the LaPorte, TX site have concentrations above those earlier. The extracts were reduced in volume and each extract
used in the low concentration standard mixture. was analyzed separately. Each extraction step removed >85%
The recoveries of the PAHs also showed similar trends, of the remaining spiked material from the denuder. Figure
based on their lability and solubility. The PAH concentrations 4 shows the comparison of the recovery from the denuder
used in the spiking mixture (and found in field samples) (sum of all three extracts) with the optimized procedure for
were lower than the n-alkanes. The lower concentrations, as the standard compounds (above). A t-test found no difference
well the chromatographic properties of the PAHs, led to larger at the 95% confidence interval between the means of
error bars associated with these compounds. However, the recoveries. Therefore, when compound-by-compound nor-
recoveries of phenanthrene (PAH 6) and fluoranthene (PAH malizations were made for losses during the reduction of
8) were 70 and 80%, respectively, consistent with values of volume procedure, the extraction efficiency of the hydro-
70% for deuterated phenanthrene and fluoranthene previ- carbons was determined to be >95 ( 10%, overall, for the
ously observed using rotary evaporation after cleanup on alkanes and PAHs.
silica (2). Recovery of hydrocarbons from filters (quartz and
Because the two classes of hydrocarbons showed similar XAD-4 impregnated quartz) was determined by spiking
patterns of recovery, for convenience they could be grouped 1 mL of the n-alkane-PAH mixture to representative filters
for preparation of future surrogate mixtures, tailored to of each type. Each filter was extracted twice in Hex:MeCl2:
specific detection requirements or availability. For each group MeOH, using sonication, as described earlier. The extracts
the recovery varied similarly with MW or ring size and were reduced in volume and analyzed separately. Each
chromatographic retention time. These groups have been extraction step removed >85% of the remaining spiked
identified in Table 1: (1) hydrocarbons nC14-nC17 showed material from the filters. Figure 4 shows the comparison of
recovery similar to naphthalene-fluorene; (2) nC18 to the recovery from the filters (sum of the two extracts) with
phenanthrene-anthracene; (3) nC19-nC24 to fluoranthene- the recovery of standards. Significantly lower recoveries of
pyrene; (4) nC24-nC32, to benz[a]anthracene-benzo[k]fluor- the lower molecular weight n-alkanes-PAHs (<nC18 and
anthene; and (5) nC34-nC36, to indeno[cd]pyrene-benzo- associated PAHs) were observed for the spiked quartz filters,
[ghi]perylene. but not for the XAD-4 impregnated filters. The results for the
Recovery of Hydrocarbons from XAD-4 Coated Denud- XAD-4 impregnated filter were similar to the XAD-4 coated
ers, Quartz Filters and XAD-4 Impregnated Quartz Filters. annular denuder, as expected for recovery from the same
To test recovery from the denuders, we spiked an XAD-4 sorbent. When compound-by-compound normalizations
coated denuder with the n-alkane-PAH mixture and extracted were applied to account for losses during the reduction
three times using the ‘rolling rinse’ method, as described of volume, the extraction efficiency of the hydrocarbons
FIGURE 5. Average concentration of n-alkanes for LaPorte, TX on 8/31/2000 from 00:00-11:30 (local time), including the contribution from
each HiC IOGAPS component. The error bars represent the uncertainty at the 1σ level.
averaged >95% (10 for all standard compounds for XAD-4 concentrations for the hydrocarbons have been corrected
impregnated quartz filters and >95% ( 10 for >nC18 and only for the losses during the reduction of volume procedure.
associated PAHs for quartz filters. Furthermore, the normalized recovery was 104% ( 8 for
The reduced recovery from quartz filters can be attributed phenanthrene-d10 and 97% ( 6 for fluoranthene-d10; an-
to evaporation of the more volatile hydrocarbons (<nC18 and thracene-d10 was below the detection limit. (While the spiked
associated PAHs) along with the solvent, from the filter amount of anthracene-d10 could not be quantified by GC-MS,
surfaces before extraction. The extraction efficiency for it can be quantified using HPLC with fluorescence detection.
remaining amounts of these compounds from the quartz The amount used was chosen to avoid swamping the native
filters was also assumed to be >95% ( 10. This assumption anthracene during HPLC analysis.)
is supported by observation that (1) the extraction efficiency Figure 5 shows the ambient concentrations of the n-
of the less volatile species on the quartz filter was >95%, and alkanes. Most of the mass of the lighter molecular weight
(2) the extraction efficiency for <nC18 and associated PAHs n-alkanes < nC19 was on the denuders, consistent with their
from the XAD-4 impregnated quartz filters was also >95%. higher volatility. No evidence was seen for particle loss in
Ambient Sample. The optimized procedure was used to the first or second denuders. For n-alkanes g nC19 a significant
analyze an entire HiC IOGAPS sample that was collected fraction of the mass was on the quartz filters, but even more
during the TexAQS-2000 at the LaPorte Airport, Houston, TX was found on the XAD-4 impregnated after-filters. This figure
(21). The mean ambient temperature was ∼28 °C during illustrates the need for back-up sorbent substrates, such as
this period. Since the extraction efficiency was >95% for the XAD-4 impregnated quartz filters, to trap SVOC desorbed
alkanes and aromatic hydrocarbons, the reported ambient from the collected particles. Figure 6 shows concentrations
of the PAHs that could be detected by the GC-MS. Most of semivolatile and nonvolatile organic compounds from air
the mass of the PAHs was on the first denuder, and the samples. The results showed how to minimize and account
concentrations generally decreased as the MW and ring size for losses of both semivolatile hydrocarbons that can
increased and the lability decreased. Naphthalene was the evaporate, and high molecular weight species that can
only PAH found in appreciable concentration on the second precipitate, during concentration steps that are intended to
(downstream) denuder, and it also broke through to the lower limits of detection. Quantitation of polar compounds
XAD-4 impregnated filters. From these data, the volumetric will require meeting these objectives with suitable laboratory
capacity of the two denuders for naphthalene in ambient air and field surrogates, internal standards and detection
at 28 °C was less than the 58 m3 of air that passed through methods, while accounting for additional causes of decreased
them. The apparent volumetric capacity for the other PAHs recovery such as lower extraction efficiency and analyte
was greater than 58 m3 but could not be established more degradation during sample preparation and analysis.
precisely because they did not break through to the second Our results point to the need for inclusion of a range of
denuder, or they were below limits of detection for the standards for recovery. Even for relatively inert compounds
GC-MS. such as hydrocarbons, non-normalized recovery depended
on volatility, MW, ring size and solubility. We recommend
Discussion that a series of deuterated surrogates be used to account
adequately for the preparation-related losses of organic
In other studies involving only one class of nonpolar species that are associated with ambient particulate matter.
compounds, or classes with similar polarity and solubility, These surrogates should span the ranges of both volatility
single solvents led to higher and more uniform non- and polarity that are expected in an environmental matrix
normalized (uncorrected for reduction of volume losses) that could also contain multifunctional and polymeric
recovery of deuterated PAH from polyurethane foam (PUF) species. Investigators should adequately describe the nor-
and filters than we found for the solvent mixture of Hex: malization algorithm(s) for each compound or group of
MeCl2:MeOH. For example, Hawthorne et al. (22) reported compounds. A good example is provided by McDonald et al.
non-normalized recoveries of over 95% for PAHs from (27) who used a suite of deuterated PAH from naphthalene-
naphthalene-d8 to benzo[ghi]perylene-d12 and several deu- d8 to coronene-d12. They reported that the recovery of each
terated methoxyphenols. Fernandez et al. (23) reported better analyte was normalized to the recovery of the deuterated
recoveries of naphthalene-d8-benzo[ghi]perylene-d12 ex- PAH with the closest chromatographic retention time and
tracted with hexane from PUF (82-91%), than of anthracene- stability.
d10-benzo[ghi]perylene-d12 from filters that were extracted Although we found no other reports in the literature that
in a mixture of MeCl2 and methanol (53-126%). Liang et al. included non-normalized recovery data for such a wide range
(24) and Mader and Pankow (25, 26) also found non- of alkanes and PAHs, our results for individual native
normalized recoveries of several deuterated PAH recoveries compounds were similar to published recoveries of the same
close to 100% from PUF and filters that were extracted only individual deuterated compounds added to filters and
in MeCl2. extracted with the solvents used by Mazurek et al. (8, 9):
However, more comprehensive characterization of air- hexane, followed by a benzene-2-propanol mixture. Ad-
borne semivolatile and particulate organics frequently ditionally, the recovery data of Schauer et al. (28-29) for
requires sample preparation procedures for compounds that nC15D32 and nC24D50 fit nicely with the data for our optimized
span wide volatility, molecular weight and polarity ranges. procedure (Figures 3 and 4). Although the Mazurek et al. (8,
In this paper we concentrated on the recovery of nonpolar 9) procedure has been widely cited, there are few reports in
hydrocarbons from a solvent mixture that was chosen to the literature of non-normalized recoveries of deuterated
optimize dissolution of both nonpolar and polar organics, standards. The literature contains even fewer descriptions
without the need for multistep extractions. Table 3 sum- of whether and how analyte concentrations were corrected
marizes our recommendations for optimal recovery of both for variation of losses with MW or ring size.
Our results have several implications for ongoing efforts nonpolar compounds, concerns may remain about inter-
to characterize airborne PM and associated semivolatile pretation of these recent studies (28, 29, 31-34). We found
compounds. First, caution must be applied to generalizing two and a half times lower recovery for nC15H32 than for n
the recoveries of SVOC from clean filters in the laboratory. C24H50. Quantitative comparisons among the results of many
The evaporation of labile species from filters may not investigators will be affected by unclear explanation of
represent the true behavior of the SVOC in PM because of normalized algorithm and inconsistent use of surrogates.
their stronger adsorption to PM than to the filter substrates.
Overestimates of their ambient PM-associated concentrations
would result. Second, caution must also be applied when Acknowledgments
interpreting results based on one surrogate for recovery of
We would like to thank our colleagues Douglas A. Lane and
many compounds. Basing ambient concentrations on re-
Cristian Mihele from Environment Canada and Yanbo Pang
covery of a midrange (volatility and MW) surrogate would
lead to worsening underestimation of ambient concentra- from Washington State University and the University of
tions of compounds the more different in volatility and MW Washington, for their inspiration to use the multiple-sample
they were from the surrogate. Problems could arise concern- programmable evaporation technique, and for their as-
ing mass closure estimates from such studies and they should sistance with the field study. We thank Robert Seila from US
be interpreted with caution. This may have been the case for EPA, Jenny Sprung from the Labconco Corporation, and
the extensive characterization of PM sources by Rogge et al. Raymond L. Dod of Lawrence Berkeley National Laboratory
(10-17) who used only nC24D50 for recovery of all non- for their contributions. We also gratefully acknowledge access
derivatized compounds. Our results suggest that using only to the laboratory facilities of Prof. Jim Lester at the University
nC24D50 could underestimate PM concentrations for both of Houston, Clear Lake. The United States Environmental
lighter and heavier hydrocarbons outside the nC19-nC34 Protection Agency through its Office of Research and
range. Similarly, Fraser et al. (30, 31) used nC24D50 as the Development partially funded and collaborated in the
recovery surrogate for PM, but also added it to a PUF, without research described here under Interagency Agreement DW-
clearly detailing the normalization algorithm for the semi- 898164-01 to E. O. Lawrence Berkeley National Laboratory
volatile compounds. There is good agreement with our data (LBNL), through the U.S. Department of Energy, under
for semivolatile compounds. However, significant bias could Contract DE-AC03-76SF00098. The project was also sup-
exist even when a few surrogates for recovery are used, unless ported by the University of Texas under an intergovernmental
recovery has been measured and normalized through a series personal act and a contract with LBNL. It has been subjected
of related compounds. Despite improved procedures that to Agency Review and approval for publication. Mention of
use solvent mixtures (or sequential solvent extraction) and trade names or commercial products does not constitute an
two deuterated species (nC15D32 and nC24D50) for recovery of endorsement or recommendation for use.