S1-1054/2 10kV Digital Insulation Tester: User Manual
S1-1054/2 10kV Digital Insulation Tester: User Manual
S1-1054/2
10kV Digital Insulation Tester
USER MANUAL
NOTE
THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS.
Users of this equipment and/or their employers are reminded that National Health and Safety Legislation requires them to carry out valid risk
assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as inadvertent short circuits.
Where the assessments show that the risk is significant then the use of fused test leads may be appropriate.
General Description A comprehensive LCD display shows resistance, current, filter setting,
The S1-1054/2 is a microprocessor controlled 10 kV insulation tester capacitance, time constant, voltage, timer measurements, and figure of
offering measurement capability of up to 35 TΩ. The instrument merit measurements such as polarisation index.
performs automatic tests and has data storage and data retrieval facilities.
Features
Polarization index and dielectric discharge tests are performed n Continuous resistance or current measurement on main display.
automatically, and test duration and voltages can be adjusted according n Optional resistance, current, or figure of merit measurements on
to user preference for these tests. A step voltage test can be performed secondary display.
automatically, with a default voltage of 1 kV, and test duration of 5
n Standard test voltages – 250 V, 500 V, 1 kV, 2.5 kV, 5 kV, 10 kV
minutes. It is possible for the user to set different voltages and test
n Non-standard test voltages – selectable between 50 V and 1 kV in 10
durations for the step voltage test.
V steps, and selectable between 1 kV and 10 kV in
The S1-1054/2 incorporates a hardware filter designed to tolerate at 25 V steps.
least 4 mA rms of noise current at 50Hz and above. This filter is enabled
n Test modes include insulation resistance, polarisation index, step
by default, but may be switched off in order to speed up the settling
voltage and dielectric discharge.
time when there is little noise current. The Insulation Resistance mode
n 4 mA rms hardware noise current filter (selectable)
offers additional firmware filtering to average out slow variations during
testing. n Firmware filter 10, 30, and 100 seconds (optional)
n The dielectric absorption ratio is automatically calculated if the
The S1-1054/2 can be powered from the mains supply or by its own
internal rechargeable battery, which provides for at least 4 hours of corresponding timers are set.
continuous testing with a 100 MΩ load. A battery level indicator on n Either burn or breakdown selectable in insulation resistance mode.
the LCD display indicates battery capacity. Connecting power to the n Insulation ‘Alarm Limit’ available in insulation resistance mode.
mains supply connector will automatically charge the battery whether
n Programmable timers include a main test duration timer, plus T1 and
the instrument is switched ‘on’ or ‘off’, except during testing. A high
T2 timers for time resistance method type tests.
level of internal isolation allows the instrument to be used while the
n Load capacitance and time constant measurements – displayed at end
unit is supplied from the mains. An internal battery management system
switches the instrument off after ten minutes of inactivity. If the battery of test.
approaches a very low state of charge the instrument turns itself off, and n LCD backlight.
mains power must be applied before the instrument can be used again. n Data storage, data retrieval and real time data output.
Recorded test results and settings will not be lost when the instrument is
n USB or RS232 communications.
switched off.
n Battery level / charge level indicator.
4
Cleaning If the power lead supplied is not suitable for your mains connection,
Disconnect the instrument and wipe it with a clean cloth slightly do not use an adaptor. Always use a power lead fitted with the correct
damped with soapy water or Isopropyl alcohol (IPA). plug.
The instrument can be powered from 85 - 265 Vrms a.c. at 50/60 Hz.
The battery will charge as long as the mains supply is connected, except
when a test is in progress. The power On/Off button has a green LED,
which illuminates when mains power is present. For optimum battery
life, charge the battery after each use. A completely exhausted battery
will take 14 hours to recharge.
Line USB RS232 Measurement terminals Voltage at Test modes Burn indicator Alarm Data recording
input -ve guard +ve terminals
A non standard voltage between 50 V and 10 kV can be selected by Testing will stop if the test start / stop button is pressed again, the
holding down the ‘Fn’ function button whilst operating the buttons. The preset test time is reached, or, if the unit is not in burn mode, insulation
selectable voltage is adjustable in 10 V steps between 50 V and 1 kV, and breakdown is detected.
adjustable in 25 V steps between 1 kV and 10 kV. An auto repeat facility
is enabled when the button is held down, allowing faster travel through When a test has finished, the instrument will discharge the load, which
the range. may take some time. The operator must always check the load has been
discharged before touching the test leads.
If there is an external voltage greater than 50 V on the test leads, the
high voltage warning indicators are flashed, and the display shows this Ω/I button
voltage instead. The instrument will not perform a test if this voltage is Pressing this button toggles the digital and secondary displays. The
greater than 160 V.
7
In the ‘IR’, ‘PI’, and ‘DD’ modes the secondary display initially shows the
PI (polarity index), DAR (dielectric absorption ratio), and, on completion Press Ω/I
of the test, the TC (time constant) and capacitance measurements.
Further presses of the Ω/I button cycles the display through insulation
resistances, currents, and the filter setting. Press Ω/I
In the ‘SV’ (step voltage) mode, the secondary display cycles through
insulation resistances, currents, and the filter setting.
Press Ω/I
Figures 1 to 4 illustrate the display status for sequential ‘Ω/I’ key presses
in the different modes.
Press Ω/I
Press Ω/I
Press Ω/I
Press Ω/I
Press Ω/I
test voltage buttons. ‘Ts and ‘Tt represent the timer buttons.
Figure 4 - The result of a step voltage (SV) test Fn + Download mode See ‘Downloading
Test condition: timer defaults to 5 minutes and test voltage defaults to results’
1000 V. Fn + Clear stored data The ‘record’ button is
pressed twice
B button Mode button
Pressing this button toggles the display backlight on and off. Press down on the ‘MODE’ button to cycle through and select the test
mode. Modes of test to be chosen from include an insulation resistance
Fn button ‘IR’ test, a polarization index ‘PI’ test, a step voltage ‘SV’ test, and a
The function button when pressed allows other keys to perform another dielectric discharge ‘DD’ test.
10
Record button Results will not be erased during this operation and so may be
This button is used to start and stop recording. Recording can only be downloaded repeatedly.
activated before testing. When data recording is enabled the ‘record’
symbol flashes. Deleting test results
Switch the instrument on and wait until initialisation is complete. Hold
Data is stored in solid-state memory and under normal circumstances down the function key and press the record key twice. The instrument
will maintain its integrity for in excess of ten years, but may rarely be now displays ‘clr’ to indicate clear mode. Press and hold down test
corrupted or lost by external influences such as transients and static button until clear process begins, shown by analogue display lighting anti
discharge. Megger Limited cannot accept responsibility for any losses of clockwise. Press the MODE button to exit without deleting the results.
data. Regular downloading to a PC using software such as Download
11
Test terminals
There are three test terminals marked +, - and G. These terminals are The guard terminal is at the same potential as the negative terminal.
designed to accept only the test leads supplied. Shutters across the Since the leakage resistance is effectively in parallel with the resistance
terminals prevent accidental ingress of dirt and other objects. Test lead to be measured, the use of the guard causes the current flowing through
plugs interlock with the shutters and are released by rotating the test lead surface leakage to be diverted from the measuring circuit. The instrument
plug a quarter turn. therefore reads the leakage of the insulator, ignoring leakage across its
surface.
The Guard terminal is explained below and is only used in cases where
surface leakage currents need to be eliminated. Most measurements use The display will show ‘FUS’ if the internal guard terminal fuse is found to
just the + and – terminals. The instrument’s internal voltage generator have blown. The instrument must be switched off to clear the message
drives the + terminal with respect to the – terminal, current being before further testing is permitted. The fuse should be replaced by an
measured in the – terminal. authorised repairer. The instrument may be used in the mean time if the
guard terminal not used. Refer to notes regarding measurements above
Guard terminal 100 GΩ on page 17.
For basic insulation tests and where there is little possibility of surface
leakage affecting the measurement, it is unnecessary to use the guard
12
Fully charged battery After testing, the voltage displayed is the voltage present on the
terminals. To see what the voltage was immediately prior to the end on
the test, press either test voltage s or t buttons.
50% charged battery
Timer indicator
The timer shows minutes and seconds. At the start of a test the timer
Tests cannot be started, and the battery will start from zero and at the end of a test it will stop. The duration of
may fail at any time the last test remains on the display until another test is started.
Digital display
Symbol flashes when there is not The digital display shows the resistance or current being measured
enough charge for a test. The during a test. The Ω/I button toggles between the two. After a test the
instrument then turns itself off 13
Secondary display Ts/ Tt Increase / Display set time Select main Display time
This part of the display shows the results of ‘time resistance’ method decrease the on main timer timer T1 or T2 set for main
tests. time of the briefly timer, T1 or
selected timer T2 briefly
14
The effects of noise can be reduces significantly with the use of high
quality Megger screened test leads.
The picture displayed on the right shows the expected screen result of
an IR test on with a 5000V test voltage after 2 minutes 9 seconds. The
15
Press the ‘MODE’ button to cycle through and select the test mode. On test completion, insulation capacitance and the time constant
Modes of test to be chosen from include an insulation resistance associated with it is calculated and displayed.
‘IR’ test, a polarization index ‘PI’ test, a step voltage ‘SV’ test, and a
Time Constant (TC) = Rins x Cins
dielectric discharge ‘DD’ test. If timers T1 and T2 are set, the instrument
will automatically calculate the dielectric absorption ratio ‘DAR’ of If timers T1 and T2 are set, the instrument will calculate and display
the insulation. PI, DAR, and IR values are automatically stored and the DAR value under the segment ‘DAR’. On completion of the test,
displayed if the data is available, irrespective of the selected test mode. the instrument will display the insulation resistances, measured at these
At the end of a test the instrument can be made to display either the times under the segment symbols ‘T1’ and ‘T2’. If the test runs for longer
insulation resistances, insulation currents, the filter setting, or ratios plus than 10 minutes, the instrument will calculate the ‘polarisation index’.
capacitance measurements by using the ‘Ω/I’ toggle button. See section This value will be displayed under the display segment symbol ‘PI’.
‘Ω/I button’ for cycling through results and toggling units. On completion of the test, the instrument will display the insulation
Hardware filter resistance recorded at 1 minute and 10 minutes under the segment
The hardware noise current filter may be switched on or off before, but symbols ‘1m’ and ‘10m’.
not during, testing. Turning the hardware filter off may be beneficial for
short test durations and higher load capacitances providing there is little Pressing the ‘Ω/I’ button toggles the display to show the insulation
noise current. resistances, insulation currents, the filter setting, DAR and PI ratios, and
capacitance. The resistances and currents will be displayed under the
This feature may be set in any mode, i.e. IR, PI, SV, or DD. Press and ‘T1’, ‘T2’, ‘1m’ and ‘10m’ segment headings, the ratios under the ‘DAR’
hold the Fn button while repeatedly pressing the Ω/I button to scroll and ‘PI’ segment headings. The capacitance reading is indicated by its
through the filter options. Select FIL OFF for the fastest settling time, or units of either ‘nF’ or ‘µF’. Refer to section ‘Ω/I button’.
FIL ON for the default 4mA noise filter. Note that if FIL OFF is selected,
the firmware filtering option (refer to section “Insulation Resistance “IR” Firmware filtering
test”) will be disabled. This setting may not be adjusted during a test. There is an option (in Insulation Resistance mode only) to switch in
firmware filtering to average out slow variations. The firmware filter time
Insulation Resistance ‘IR’ test constants are 10, 30, and 100 seconds. The additional firmware filter
This is the default mode in which the instrument powers up. This setting may be adjusted before or during a test, unless the filter was set
test mode measures insulation resistance continuously at the selected to FIL OFF before the test started, in which case the additional firmware
voltage. The test voltage may be varied during an ‘IR’ test by pressing filtering option will be unavailable during testing. If the additional
the test voltage s or t buttons. The test duration can be set using the firmware filter is switched in during a test, then any subsequent PI, DAR,
‘main timer’. The test will finish automatically after this time has elapsed. or TC calculated results are regarded as invalid, and are consequently
not displayed. The resistances and currents at these times are still
16
Ground
19
EMC
Meets the requirements of EN61326-1 Optional Accessories
21
The instrument contains static sensitive devices, and care must be taken in handling the printed circuit board. If an instrument’s protection has been
impaired it should not be used, but sent for repair by suitably trained and qualified personnel. The protection is likely to be impaired if for example, it
shows visible damage, fails to perform the intended measurements, has been subjected to prolonged storage under unfavourable conditions, or has been
subjected to severe transport stresses.
NEW INSTRUMENTS ARE GUARANTEED FOR 1 YEAR FROM THE DATE OF PURCHASE BY THE USER.
Note: Any unauthorized prior repair or adjustment will automatically invalidate the Warranty.