0% found this document useful (0 votes)
44 views2 pages

Cell Sorting Guidelines

The document provides guidelines for cell sorting at the Brown Flow Cytometry Facility. Sorting requires a prior appointment that is scheduled at least one week in advance. Users must provide samples between 5x10^6 and 1x10^7 cells per ml, as well as control samples for calibration. Samples are sorted from 10am to 4:30pm Monday through Friday, and users will be billed for the time reserved.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
44 views2 pages

Cell Sorting Guidelines

The document provides guidelines for cell sorting at the Brown Flow Cytometry Facility. Sorting requires a prior appointment that is scheduled at least one week in advance. Users must provide samples between 5x10^6 and 1x10^7 cells per ml, as well as control samples for calibration. Samples are sorted from 10am to 4:30pm Monday through Friday, and users will be billed for the time reserved.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 2

 

Brown Flow Cytometry Facility


Cell Sorting Guidelines
Cell Sorting:

 Sorting is done by prior appointment only. Due to set up time required, please
explain your needs thoroughly when scheduling an appointment. A completed
Cell Sort Request Form must be submitted before your appointment will be
confirmed. Scheduling generally should be done about one week in advance.

 Sorting hours are from 10:00 A.M. to 4:30 P.M. Monday through Friday.
Extended hours (early or late time) can be arranged a the time of initial
scheduling.

 Any cancellations with less than a 24 hour notice before scheduled start time
will be charged 1 hour. No-shows will be charged for time reserved.

 All users must provide an account for billing at time of scheduling. Users will
be billed monthly.

 Scheduling is on a first come, first served basis.

SORT SAMPLE PREPARATION GUIDELINES:

 Do not fill sample tubes over 2/3 full.

 Suspend cells in PBS at 5 X 106 per ml - 1 X 107 per ml.

 Sample cell concentration affects sort speed.

 Use the highest concentration possible while maintaining a


single cell suspension.
 We can dilute samples at sorting facility but cannot concentrate
them.

 Control Samples: For all sort operations a negative or unstained control and a
separate single stain control sample tube for each stain to be used is required
for instrument pre-sort calibration. (ie: if you are using FITC, PE, PECy7, APC, a
control sample of each is required for calibration and compensation of
instrument. In this case, 5 control samples total including unstained cell
sample)

 GFP sort requires only a negative or not-transfected (not-transduced) control


sample.

 Control samples should be prepared the same as the sort sample from the
same type of cells as experiment samples.

 Each control sample should contain a minimum of around 200,000 cells.

COLLECTION TUBE PREPARATION:

 Pre-coating tubes with serum or media/buffer is recommended. Collection


media/buffer can also be added to the tubes (no more than 1 ml).

 Populations can be simultaneously be sorted into: 12 X 75mm tubes or 17 X


120mm tubes or 15 ml tubes

SINGLE CELL SORTING:

 Single cells can be sorted at 1 – 50,000 cells into 6 to 96 well micro titer plates
or onto a slide.

 Plates must be pre-filled (100 microliters of complete medium in 96 well plate).

For any further questions regarding sorting, please contact us:

[email protected]

You might also like