Philips: ® BV 25 Mobile Image Intensifier

The Philips BV 25 Mobile Image Intensifier is a mobile C-arm x-ray system with the following key features: - It has a 15cm image intensifier and dual x-ray focal spots of 0.6mm and 1.8mm. - The C-arm has movements including height adjustment, horizontal movement, scanning, rotation and orbital movement. - It has capabilities for both fluoroscopy and radiography, with controls for kV, mA, and exposure time during radiography.

Uploaded by

ayham Hamama
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
100 views2 pages

Philips: ® BV 25 Mobile Image Intensifier

The Philips BV 25 Mobile Image Intensifier is a mobile C-arm x-ray system with the following key features: - It has a 15cm image intensifier and dual x-ray focal spots of 0.6mm and 1.8mm. - The C-arm has movements including height adjustment, horizontal movement, scanning, rotation and orbital movement. - It has capabilities for both fluoroscopy and radiography, with controls for kV, mA, and exposure time during radiography.

Uploaded by

ayham Hamama
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

PHILIPS® BV 25 MOBILE IMAGE INTENSIFIER

Specifications

X-ray Source
X-ray generator: DC converter-generator. X-ray tubes – focal spots: dual-focus stationary anode 0.6 and
1.8.

Fluoroscopy System
Image intensifier: 15 cm (6 in) high quantum absorption model with fibre output. TV camera tube:
VIDICON with fibre input. Optical coupling between image intensifier and TV camera tube: glass-fibre
optics. Source of image distance: 90 cm. Dose rate control: automatic, simultaneous kV/mA control with
lock and manual override facilities. KV/mA range: from 0.1 mA at 40 kV to 3 mA at 100 kV. Focal spot: 0.6.
Grid: fixed circular grid 44 lines/cm, ratio 8, SID = 90 cm.

Collimator
Iris diaphragm: electronically controlled iris diaphragm, automatically limited to size of image intensifier
input field (15 cm 0). Remotely controlled and steplessly adjustable to a field size of 5 cm diameter at input
field image intensifier. Shutters: two parallel shutters of 0.5 mm Cu, remotely controlled and steplessly
adjustable for a field of 1 to 16 cm wide at image intensifier. The shutters can be rotated ± 90°; in the mid
position, the slit aperture is perpendicular to the length of the generator housing.

TV System
XTV 5 50 Hz: 625 lines, 60 Hz: 525 lines. Bandwith: 15 MHz. Control: AGC feedback to dose rate control
system. Video monitor(s): 51 cm (20”) screen motorized, remotely controlled image rotation and reversal.
Output signal for recording system: output level of the composite signal always 1.0 Vp-p positive across 75
ohms. Measuring field selection for adapting control mode to the objects to be examined: small measuring
field: 0.3 x diameter of TV image, e.g. extremities. Intermediate measuring field: 0.5 x diameter of TV
image, e.g. bile duct and gall bladder. The measuring filed is selected automatically as a function of the kV
used.

C-Arm Movements
Height adjustment: 50 cm, motorized. Horizontal movement: 20 cm. Scanning in horizontal plane: 12 1⁄2°
to either side. Rotation of c-arm: 205° in either direction. Orbital movement of c-arm: 115°. Distance
between x-ray tube space and image intensifier with 30 cm spacer: 60 cm.

Radiography
Tube current: fixed, 20 mA. Tube voltage: adjustable in steps of 5 kV from 40 to 100 kV. Exposure time:
adjustable in steps of 10 ms to 4.0 s (mAs is displayed). Focus: 1.8. Radiography parameter display:
digital display of tube voltage (kV) and exposure time/current product (mAs). Collimator: if iris diaphragm
was set to less than 15 cm during fluoroscopy, this remains so unless cancelled. Circular fields of 15 or 24
cm can be selected by remote control. Rotatable cassette holder: suitable for cassette and grid cassettes
of 24 x 24 cm. Proffered grid: ratio 10.44 lines/cm, SID = 90 cm. Inherent filtration: 3 mm Al at 75 kV.
Leakage radiation factor: 100 kVdc, 0.06 mA (2160 mAs/hr)

Accuracy of Display
Fluoroscopy: voltage deviation ± 6% ± 2kV, current deviation ± 7% ± 0.02 mA. Radiography: voltage
deviation ± 10%, current deviation ± 10%, time deviation ± 2 ms ± 3%, mAs product deviation ± 12% ±
0.04 mAs

Weights
C-arm stand: 210 kg. Monitor trolley excluding image memory single monitor: 131 kg, dual monitors: 164
kg.

Mains Supply
Voltage: 110/120/127/190/208/220/240 V single phase, ±10%. Frequency: 50/60 Hz. Earth leakage
current: less than 100 ìA