Sartorius Combics 1 - Combics 2: Service Manual
Sartorius Combics 1 - Combics 2: Service Manual
and OK
REF
WCI5005-e03104
Service Concept
Prerequisites for performing maintenance and repair work on Sartorius Combics
scales requires considerable experience with both indicators and weighing platforms.
In case of defects, repairs are performed on site. Generally, the equipment is not
replaced.
General Information
– Do not connect or disconnect cables to or from the equipment; always disconnect the
power cable from the wall socket (mains supply) first!
– To ensure safety, an isolating transformer must be installed between the indicator and
the power supply before performing work that entails opening the Combics indicator
housing.
(The procedure is the same as for all scale and weighing platform models.)
– Repeatability (standard deviation)
– Off-center load
– Span
– Linearity
Activate the BPI mode to load a modified data record in the controller
of the A/D converter See page 15
Activate the xBPI protocal to perform calibration and adjustment, if necessary See page 16
using Sartocas
Use Sartocas software - to reactivate the SBI protocol by selecting the „Close“
command on the Combics terminal (or by restoring
factory defaults (menu line 9-1)
- to program a new A/D converter PCB with data read from the old PCB
Drift, display
Inf 02 when starting an
adjustment routine to check the A/D converter (quick-test) See page 73
or with the strain-gauge simulator,
check load cell(s) with the load cell tester
Display blank (dark) Disconnect the equipment from the power supply and
disconnect all cables and wires from the subassemblies,
connect equipment to power again, connect subassemblies
in turn and measure voltages See pages 70-72
No communication with
connected equipment Check connections and data transfer parameters;
start internal test program; if necessary, replace
Important: Mechanical and electrical service or repair work on the Combics scale requires
considerable experience, and for this reason should be performed only by
Sartorius technicians trained at the factory. Any attempt to perform repair work
can result in damage to the equipment.
Accompanying Literature
10
11
12
Combics 2
Analog/digital converter
combics_15_klemmen.eps
– Combics indicators
are equipped with either cable
glands or D-Sub 25 Power PCB DC/DC converter
connectors. (Combics2 has a clock chip)
13
B. SBI/BPI Key
closed open To use the Sartocas program (v 1.46 or later) or Psion server (v 4.10 or later).
verriegelungs_schalter.eps
the Combics indicator must be in BPI (binary processor interface) mode (for
example, to load a modified data record from the A/D converter into its
controller). This setting is valid for the COM1 port.
Note: To perform adjustment only with the service software, the COM1
interface is set to the xBPI (eXtended Binary Processor Interface) protocol
and the laptop or Psion server connected to that port.
B
sbi_bpi_taster.jpg
14
Important: Peripheral devices that communicate over the SBI (Sartorius Balance Interface)
protocol cannot be operated while the Combics is in BPI mode.
If you press the p key, only InF 30 is displayed.
To return the indicator to the SBI mode, either use the “Close” function in the
Sartocas service software or a Psion server (and then turn the Combics off and
back on again), or select the Menu Reset option (9 - 1 - 1) in the Combics
operating menu.
C. Boot Key
Activate this switch only if an attempt to program the application memory
with the PPLoader program fails (programming routine stops responding).
Procedure for making the PCB bootable:
– Disconnect the Combics from the main power.
– Press and hold the boot switch (approx. 3 sec.) while reconnecting the Combics
C to power.
– Release the switch. After a brief period, the PCB is bootable again
(load application software again with the PPLoader program).
boot_taster.jpg
15
Activate the xBPI protocol to perform adjustment with the service software
Sartocas software or Psion server without activating the BPI mode
through the SBI/BPI key.
– Via Setup select the Device parameters menu item (for the COM1 port).
– Open the Device parameters menu and scroll down to the Interface
submenu.
– Open the Interface submenu and activate the xBPI menu item.
Display_xbpi.eps
Note: Under xBPI a submenu is opened for assigning addresses, but this function is only
available when using xBPI with RS-485.
16
display.eps
appl.eps
setup.eps
display_wp1.eps
code.eps
17
d_vore12.eps
– Press k to select “2” and then press ) to confirm; the cursor jumps to the
second position; now you can enter the next number. Repeat this procedure
with the appropriate numbers to enter the service password (202122).
202122.eps
Note: If you find that one of the numbers is wrong, press ( to go back to that digit
and change it; then press the ) key repeatedly to return to the sixth digit
(the 2 flashes).
– After entering the last digit, press ) (< 2 sec.) and ( ( (in that order)
to activate the service mode. Setup appears in the display.
A “S” in the upper right-hand corner of the display indicates that the service
code_s.eps mode is active.
Note: The service mode can be deactivated only by turning the off the Combics indicator.
Important: When the service mode is active (password 202122), you can only view the
customer password. The customer password can be deleted only by entering
the general password (40414243) or the customers password.
18
19
20
21
display_wps1_serv.eps
display_serv_01.eps
– Select menu item 1-18-1 (external user-defined weight) by pressing k )
repeatedly.
menu_1181.eps
– Activate input mode for the user-defined weight (press ) briefly).
The first digit flashes.
user1Ib.eps
– Enter the user-defined weight (in this example: 10,000 lb) (key sequence:
) k ) p ) ) ) k )). The last number entered flashes.
user2Ib.eps
– Press the ) key again to store the user-defined weight; the menu jumps back
to 1-18-1.
– To exit the service mode, press and hold the ) key for longer than 2 seconds.
menu_1181.eps
22
– Activate the input mode for the first linearization weight (press the ) key
briefly). The first number flashes.
user1Ib.eps
– 1. Enter the linearization weight (in this example: 2,500 lb) (key sequence:
3 x k, ) p ) 6 x k, ) k ) k). The last number entered
flashes.
userl2Ib.eps
– Press and hold the ) key (> 2 sec.) to store the value for the first
linearization weight. The menu jumps back to item 1-18-2.
menu_1182.eps – Follow the above steps for menu items 1-18-3, 1-18-4, and 1-18-5 to enter
values for the second, third and fourth linearization weights as well.
Enter “0” instead of a weight value for unused linearization points.
– At the end of input, the display must show menu item 1-18-5.
– To exit the service mode, press and hold the ) key for longer than 2 seconds.
menu_1185.eps
23
If a weighing system is used in a location other than that for which it was
originally intended (i.e., the geographical data at the new place of use does not
match the entered at Sartorius AG in Goettingen, Germany), the scale span can be
adjusted by using external weights or by entering the specifications of the load
cell(s). In this case, the geographical data entered for the original place of use is
no longer valid. The same applies for span adjustment performed after any repair
work on the weighing system.
Once you store the geographical data or acceleration of gravity at the place of use
and subsequently perform span adjustment using weights, the scale is correctly
programmed and adjusted.
You can now enter the geographical data of the new place of use to adapt the
scale for use at that location.
24
The following values are entered when the Combics scale is adjusted
in Goettingen, Germany:
– Latitude: 51.53 degrees (= 51° 32’)
– Altitude: 151 m
– or the corresponding acceleration of gravity, 9.811590 m/s-2
The greater the precision of the geographical data entered, the greater the
precision achieved with the weighing instrument; the tolerance range, however,
is restricted accordingly (see above).
25
The area of use valid for this scale is designated “Zone D” (valid only in Germany).
26
display_wps1_serv.eps
display_serv_01.eps
27
5.0lb.eps
display_wps1_serv.eps
display_serv_01.eps
28
– After taring or zeroing the scale, press and hold the ) key until Cal is
displayed.
If menu item 8-12-2 is active, the geographical data is displayed before the
display_cal.eps calibration weight is prompted. Confirm each of the values shown by pressing
the ) key to continue with the calibration routine.
– After approx. 2 seconds the calibration weight (user-defined weight) required
is shown on the display. Select menu item “1-18-1” to enter the user-defined
weight.
10Ib.eps
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last span adjustment is displayed
(calibration).
display_serv_01Ib.eps
– Press the ( key to stop the calibration/adjustment routine.
10Ibw.eps
29
Display_WP1_Serv.eps
display_serv_01.eps
– Select menu item 1-9-7 (External linearization with user-defined weight) (key
sequence: ), then, several times, k )) and confirm by pressing )
briefly.
menu_197.eps
– Press and hold the ) key to store the selected menu items and return the
Combics indicator to the weighing mode.
When this menu setting is already active, you can exit the Service mode by
turning the Combics indicator off and then on again.
– After taring or zeroing the scale, press and hold the ) key until Lin is
displayed.
display_lin.eps
30
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration).
– Press the ( key to stop the calibration/adjustment routine.
display_serv_02Ib.eps
– Press the ) key to store the value for the first linearization weight;
the second linearization weight is prompted.
display_5Ib.eps
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration).
display_serv_03lb.eps
– Press the ) key to store the value for the second linearization weight;
the third linearization weight is prompted.
7.5Ib.eps
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration).
display_serv_04lb.eps
31
10Ib.eps
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration).
display_serv_03lb.eps
– Press the ) key to store the value for the fourth linearization weight.
The scale now prompts a requires the zero point (remove all weights from the
load plate).
display_serv_06Ib.eps
– Once the zero point is stored, the Combics automatically returns to the
weighing mode.
0kg.eps
32
display_wp1_serv.eps
display_serv_01.eps
– Select menu item 1-9-8 (Set preload) (key sequence: ), then, several times,
k )) and confirm by pressing ) briefly.
– Press and hold the ) key to store the selected menu items and return the
menu_198.eps
Combics indicator to the weighing mode.
– Press )) to tare the Combics indicator, or press ( to zero it.
If you tare the indicator, “NET” might be displayed.
0lb_o.eps
vorl1.eps
33
vorl2.eps
– Release the ) key; after a brief pause, the Combics indicator automatically
returns to the weighing mode.
0lb_o.eps
display_wps1_serv.eps
display_serv_01.eps
– Select menu item 1-9-9 (Clear preload) (key sequence: ), then, several times,
k )) and confirm by pressing ) briefly.
– Press and hold the ) key to store the selected menu items and return the
vorl3.eps Combics indicator to the weighing mode.
34
Display_-0335lb.eps
vorl5.eps
– Release the ) key; the preload is cleared and, after a brief pause, the Combics
indicator automatically returns to the weighing mode.
0lb_o.eps
35
display_wps1_serv.eps
display_serv_01.eps
– Press the ) key, then, several times, k repeatedly and then ) again to
open menu item 1-19, “Calibration without weights”; then select item 1-19-1,
“Nominal capacity”.
display_serv_01191.eps
Note: If a weighing platform has more than one load cell, multiply the nominal capacity
accordingly. Example: The weighing platform consists of 4 load cells, each of them
has a capacity of 30 kg.
In this case, the nominal capacity (= maximum load) equals 4 x 30 kg, or 120 kg.
36
– Enter the resolution (lowest scale interval d) (e.g., 0.01 kg). This value has to
match that stored under menu item 11-4-1 (or 11-5-1 or 11-6-1).
display_schrittw_001Ib.eps – To store the setting, press and hold the ) key (> 2 sec).
display_wze19440.eps – To store the setting, press and hold the ) key (> 2 sec).
Note: If a weighing platform consists of more than one load cell, enter the sensitivity
values for the other cells (up to 4) under menu items 1-19-4 (for load cell 2)
through 1-19-6 (for load cell 4) and confirm each value (k ) keys).
If the platform has less than four load cells, enter “0.000” under the unused menu
items.
37
display_serv_01197.eps
display_serv_01.eps
– Exit the operating menu and return to the weighing mode.
38
When you activate menu item 1-20-1, the value 99999.99 is displayed; under
1-20-2, is 9999999 displayed too. Otherwise, if values have already been entered
for geographical latitude and altitude, then 0.000000 is displayed under menu
item 1-20-3.
During calibration, the Combics show will whether and which gravimetric data
(geographical latitude and altitude, or acceleration of gravity) is used, if the menu
item 8-12-2 is activated.
If latitude and altitude are used, the display shows “Altitude” for 2 seconds when
the calibration procedure is started (CAL), followed by the configured elevation
above sea level in meters. Press ) to confirm this information. Next, the display
shows “LATITUDE” for 2 seconds, followed by the value set for the geographical
latitude (in degrees). Press ) to confirm this information. The calibration
weight is now prompted. If the acceleration of gravity is given rather than the
geographical latitude and elevation, the display shows “GRAVITY” for 2 seconds
(after “CAL”), followed by the value entered for the local acceleration of gravity.
Press ) to confirm this information.
However, if the calibration weight is prompted as soon as the “CAL” display goes
out, this means whether the correction of the calibration factor has been
deactivated nor the menu item 8-12-2 is activated; in other words, neither the
geographical latitude and altitude nor the acceleration of gravity have been
entered under menu line 1-20.
39
display_wp1_serv.eps
– Press the ) key to open the numeric menu. The first menu page is displayed.
display_serv_01.eps
display_serv_01201.eps If the geographical latitude and altitude of the place of installation are known,
enter this data under menu items 1-20-1 (latitude) and 1-20-2 (altitude).
If you wish to use the acceleration of gravity at the place of installation for the
basis of adjustment rather than these two parameters, enter this value under
menu item 1-20-3.
Important: A value entered for acceleration of gravity takes precedence over the
combination of geographical latitude and altitude.
40
In the example shown here, the most recently entered parameter was the value for
the acceleration of gravity (menu item 1-20-3). For this reason, the input fields
for menu items 1-20-1 and 1-20-2 are filled with 999…
display_999.eps
Enter the geographical latitude of the place of installation (or the reference
value 51.53) as a positive decimal number (convert angular minutes to decimal
places). Press ) to confirm your input.
41
display_151.eps Note: A negative value may be entered for local elevation; for example,
if the scale is installed in a mine.
Rather than the geographical latitude and local elevation of the place of
installation, you can enter a value for the acceleration of gravity at the place of
installation (menu item 1-20-3).
Example: In the example shown here, the geographical latitude and local
altitude (menu items 1-20-1 and 1-20-2) were the last parameters
entered in menu line 1-20. Thus the input field for acceleration of
display_000.eps gravity is filled with 000…
– Activate menu item 1-20-4 to store new data (key sequence: ) k).
The display shows “busy”, indicating that the parameters entered are being stored.
– Restart the scale: Turn the indicator off and then on again (key sequence: e
display_busy.eps e, or press and hold the ) key (> 2 sec).
display.eps
Note: Once the precise geographical data or acceleration of gravity has been entered, the
tolerance range is restricted (see page 23).
– After data has been entered as described above, the span must
be adjusted.
Example
Settings (changes in the factory settings required for this example):
Menu item 1-9-1: External calibration/adjustment with default weight;
Menu item 1-10-1: Calibration/adjustment in one operation;
Menu item 1-16-1: External calibration/adjustment/linearization
accessible.
43
The altitude at the place of installation is displayed in meters above sea level.
In the example shown here, the altitude of Goettingen, Germany (reference value)
is shown.
display_alt_151.eps
display_latitud.eps
The geographical latitude of the place of installation is shown in degrees north
or degrees south.
In the example shown here, the altitude of Goettingen, Germany (reference value)
display_lat_5153.eps
is shown.
– Place the prompted calibration weight on the scale (see page 25 for details on
10Ib.eps
calibration and adjustment).
or
44
– Place the prompted calibration weight on the scale (see page 25 for details on
calibration and adjustment).
10Ib.eps
45
46
Features With the menu access switch open, you can Important: The A/D converter can be configured only in
configure most of the parameters affecting the service mode.
the following specifications:
– Toggling between Standard and Trade configurations
The maximum load for each weighing range must be
(for use in legal metrology)
entered in an appropriate weight unit. Select the desired
weight unit for operation under menu line 11-7, “Weight
– Verification scale interval e
units”. The maximum load for each range must be a value
– Scale interval d
that can be displayed in any available weight unit.
– Minimum load
Weight units that are accessible (or blocked) for use as
– Maximum load
Weight unit 1 are also available (or blocked) for use as
– Maximum load for a given range
Weight unit 2. The weight unit used for the configuration
– Verification scale interval e for a given range
cannot be blocked.
– User-definable weight units
Checking and Configuring the Equipment for Use in Legal
These parameters are not reset when you restore the
Metrology:
factory defaults in the operating menu.
Parameters not listed above are not affected by your
choice of Standard or Trade configuration; the same
restrictions apply as for Sartorius weighing instruments
which cannot be configured.
47
Open menu line 1-7 to make sure only the permitted When the Trade configuration is active (for verifiable or
weight units are accessible. verified weighing platforms), only menu item 11-1-4 is
shown.
Perform necessary routines for scales used in legal Select the accuracy class appropriate for the weighing
metrology. platform connected. Which class is appropriate depends
on the load cell(s) installed.
Descriptions of the Individual Menu Items
In legal metrology, only classes l
Menu Item 9-1-3 / 9-1-4:
(menu item 11-1-4) can be selected.
Selecting and loading configuration data
Press k to select the accuracy classes and ) to
confirm. A circle (o) after the menu code indicates the
Before selecting the menu item for A/D converter
active setting.
configuration, check whether the weighing platform is
used in the standard operating mode (Standard
configuration) or in legal metrology (Trade configuration):
48
Scale Interval d
The lowest scale interval d indicates the resolution of the
weighing instrument. The scale interval d can be entered
only in increments of 1, 2, 5, 10, 20, 50, etc. When working
on a verifiable or verified scale in accuracy class l, you
are not prompted to enter the lowest scale interval, as this
is equal to the verification scale interval e.
49
50
display_wp1_serv.eps
– Press the ) key to open the numeric menu. The first menu page is displayed.
display_serv_01.eps
– Open menu item 9-1 (Factory setting/Menu reset)
(key sequence: k k )).
display_serv_91.eps
Loading the Trade Configuration
– Select menu item 9-1-4 (Trade configuration) and confirm
(key sequence: ) k k)).
display_serv_914.eps
display_busy.eps
Afterwards, the program returns to the display of menu item 9-1-2
(Restore factory defaults: »Off«; factory setting).
display_serv_912.eps
52
Note: If the Trade configuration is loaded, menu line 11-1 (for selecting
display_serv_11.eps
accuracy class) is displayed at this point.
Otherwise (Standard configuration), that menu line is skipped and 11-2
(weight unit 1) is displayed.
In this case, press k to open menu line 11-3.
display_serv_111.eps
– To use the weighing instrument in legal metrology (i.e., with the Trade
configuration loaded) press ) to enter the accuracy class:
display_serv_1114.eps
– Press ) to confirm.
The configured menu item (in the picture here,
11-1-4 for accuracy class l) is marked by a circle (o).
display_serv_1114w.eps
53
display_serv_113.eps
Note: Enter the scale interval, minimum load, range limits (multiple-range or multi-
interval scales only) and maximum capacity:
The menu line corresponding to the selected weighing range structure is displayed;
i.e.
– for single-range scale (menu item 11-3-1): in relation to menu item 11-4;
– for multiple-range scale (menu item 11-3-2): in relation to menu item 11-5;
– for multi-interval scale (menu item 11-3-3): in relation to menu item 11-6.
In each case, the menu lines for the other two weighing range structures are not
shown.
54
display_serv_114.eps
– If the Trade configuration is active, select menu item 11-4-2 (Verification scale
interval e) (key sequence: ( k k k )).
display_serv_1142.eps
eiche1lb.eps
– Enter the value for e (for example, e = 0.1 lb) by pressing the key sequence )
) k ) ) p ) and store (press and hold ) > 2 sec).
eiche2lb.eps
When menu item 11-4-2 is opened again, the display shows 0.1 lb.
display_schrittw_01Ib.eps – Select menu item 11-4-3 (Minimum load) (key sequence: ( k).
display_serv_1143.eps
– Press the ) key again to open this item; then enter the minimum load (key
sequence: k k k ) p ) k).
Example for accuracy class K: Minimum load = 20 x e = 2.0 lb
display_mnl20Ib.eps – To store the new setting, press and hold the ) key (> 2 sec).
55
– Press ) again to open this menu item; then enter the maximum capacity (for
Display_Serv_1144.eps example 300.0 kg) (key sequence: k k k k ) k ) k )
p ) k). Enter “300” (it is not necessary to enter the decimal point and
the last “0”).
– To store the new setting, press and hold the ) key (> 2 sec).
Display_Mxl30.eps
Each of these menu lines has 3 items to choose from, for setting range limits
1, 2 and 3.
– Store the configuration data: Select menu line 11-10 (key sequence: ( (
k k k )).
The default setting is 11-10-2, “Do not store configuration”.
Display_Serv_11102.eps
Display_Serv_11101.eps
– To store the new settings, press and hold the ) key (> 2 sec).
56
Display_busy.eps
Afterwards, the Combics is re-started. The scale is in the normal weighing
mode.
Then adjust the weighing instrument and, if the Trade configuration is active,
Display.eps
return the menu access switch from the “accessible” to the “closed” position (see
page 12).
Menu Item 9-1: A/D Converter Configuration with Load Cell(s) Connected
(Example: Standard Configuration
– Activate the service mode (see page 17).
– Press ) repeatedly to select weighing platform »WP1«.
Display_WP1_Serv.eps
– Press the ) key to open the numeric menu. The first menu page is displayed.
57
display_serv_91.eps
– Open menu item 9-1 (Standard configuration) and confirm (key sequence: k
k )).
Display_Serv_913.eps
Display_busy.eps
Afterwards, the program returns to the display of menu item 9-1-2 (Restore
defaults: »Off«; factory setting).
Display_Serv_912.eps
Display_Serv_113.eps
58
Note: Entering the scale interval, range limits (multiple-range or multi-interval scales
only) and maximum capacity:
The menu line corresponding to the selected weighing range structure
is displayed, i.e.,
– for single-range scale (menu item 11-3-1): in relation to menu item 11-4;
– for multiple-range scale (menu item 11-3-2): in relation to menu item 11-5;
– for multi-interval scale (menu item 11-3-3): in relation to menu item 11-6.
In each case, the menu lines for the other two weighing range structures are not
shown.
– When you activate menu code 11-3-1 (Single-range scale) (by pressing the key),
menu line 11-4 is opened automatically. In this case, menu lines 11-5 and 11-6
are not shown.
Display_Serv_114.eps
Display_Serv_1141.eps
59
– Otherwise, overwrite the value as desired (in this example, by entering 0.01 lb;
key sequence: k ) p ) k ) k k); then press and hold the
) (> 2 sec) to confirm.
display_schrittw_001Ib.eps
After you store the data, the program jumps back to menu item 11-4-1.
Display_Serv_1141.eps
Display_Serv_1144.eps
standard1Ib.eps
60
After you store the data, the program jumps back to menu item 11-4-4.
Display_Serv_1144.eps
– Store the configuration data: Select menu line 11-10 (key sequence: ( k
kk
display_serv_1110.eps
Display_Serv_11102.eps
Display_Serv_11101.eps
61
– Press the ) key, then seveal times, k ) to open menu line 1-19,
“Calibration without weights”; then select item 1-19-1, “Nominal capacity”.
Display_Serv_01.eps
Display_Serv_01191.eps
– Enter the nominal capacity of the load cell(s) (e.g., 30.00 kg).
The nominal capacity has to match the maximum load set under
menu item 11-4-4.
display_mxl30Ib.eps
62
– To store the setting, press and hold the ) key (> 2 sec).
– Select menu item1-19-2 (key sequence: k, )).
– Enter the resolution (lowest scale interval d) (e.g., 0.01 kg). This value has to
match that stored under menu item 11-4-1 (or 11-5-1 or 11-6-1).
display_schrittw_001Ib.eps
– To store the setting, press and hold the ) key (> 2 sec).
– Open menu item 1-19-3 (key sequence: k )).
Enter the sensitivity of the load cell in mV/V (usually listed in the load cell
specification sheets). If a weighing platform consists of more than one load
cell, enter the sensitivity of load cell 1.
Display_WZE19440.eps
Example: The sensitivity of the load cell is 1.9440 mV/V.
– To store the setting, press and hold the ) key (> 2 sec).
Note: If a weighing platform consists of more than one load cell, enter the sensitivity
values for the other cells (up to 4) under menu items 1-19-4 (for load cell 2)
through 1-19-6 (for load cell 4) and confirm each value (k and ) keys).
If the platform has less than four load cells, enter “0.000” under the unused menu
items.
63
Display_Serv_01197.eps – Press ) ( ( to store the load cell specifications entered; the menu
automatically jumps back to the first menu page.
code_s.eps
64
cursor.eps
s-date.eps
– Press e twice to return the Combics indicator to the weighing mode.
0lb_o.eps
code_s.eps
65
ser-no.eps
Note: After you enter the last digit, the first digit is no longer displayed.
– Press and hold the ) key to store the serial number; the display returns
automatically to Ser-no.
ser-no.eps
0lb_o.eps
66
code_s.eps
– Press k k k k to scroll through the menu until the display shows
ModEL.
model.eps
– Press ) activate the input mode (the cursor flashes in the first position).
cursor.eps
– Press the ) k keys as needed to enter the model designation (for
example, CISL_1U of the weighing system.
Display_CISLIU_.eps
– Press and hold the ) key (with the cursor to the right of the last character)
to store the model designation; the display returns automatically to ModEL.
model.eps
0lb_o.eps
67
code_s.eps
– Press k k to scroll through the menu until the display shows MEM-no.
memory.eps
– Press ) activate the input mode (the cursor flashes in the first position).
cursor.eps
– Press the ) k keys as needed to enter the allocation number in the Alibi
memory.
memory1.eps
– Press and hold the ) key (with the cursor to the right of the last character)
to store the allocation number; the display returns automatically to MEM-no.
memory.eps
0lb_o.eps
68
To open the Combics indicator, remove the four nuts as shown in the illustration
on the left.
Note: After completing maintenance or repair work, check the seal between the
front panel and the housing body for damage and replace if necessary.
If the Combics indicator in question has an IP67 protection rating,
a special test procedure is used to check the IP67 protection after the
housing has been closed.
combics_10.eps
69
– The voltage at the power supply output (connector A; see page 71) is
15V ± 0.3 V direct current.
– The output voltages of the DC/DC converter are listed on page 72.
– After replacing the power cable, use a torque wrench to tighten the cable gland
to 3 Nm.
pg_netz.jpg
70
netzt_kappe.jpg – Remove the protective cap and disconnect the two wires (blue/brown).
– Unplug connector A, remove the 2 screws and replace the power supply.
A After replaceing the PCB, make sure to return the protective cap to its
original position.
netzteil.jpg
71
dc_dc.gif
M1 M2 M3 M4 M5 M6
Note: Use a digital voltmeter (DC setting) to measure voltages against ground (earth).
No particular points of measurement are defined for this procedure; for this
reason, use a thin test pin and exercise extreme caution to avoid short-circuiting
the equipment.
– Because the model designation and serial number are not stored in the A/D
converter’s controller, you do not need to enter this data again after replacing
the A/D converter.
– When connecting the weighing platform (load cell(s)) to the A/D converter,
it is important to install a ferrite ring over the signal lines leading from the
platform, to ensure electromagnetic compatibility.
adu_001.jpg
73
2. Connect the strain-gauge simulator and turn the switch to simulate different
input signals. The weight readout should be stable.
– After replacing the digital PCB, use the PPLoader program to load the
application software, if necessary.
haupt_25.jpg
– Afterwards, enter the model designation and the Serial number of the Combics
scale (see pages 65 and 67).
74
75
If the weighing platform has only one load cell, the off-center load is not adjusted
in case of error; rather, the load cell must be replaced.
1246 lb
The off-center load is adjusted only in weighing platforms that have four load
Load cell 2
cells.
1255 lb
In the example given here, load cell 3 has an off-center loading error of 8 lb,
Load cell 1
1254 lb
1260 lb
Eckl_2_usa.eps
Note: The resistance values detected are based on 350-ohm load cells.
For 700-ohm load cells, double the resistance values.
76
0,39+1,5+3Ω
capacity of the weighing platform)
.
1,5+3Ω
rather than a 12.5 klb test weight,
.
0,39+3Ω
.
adjustment resistance is set as
3Ω
.
when a 12.5 klb test weight is used.
0,39+0,82+1,5Ω
.
0,82+1,5Ω
.
0,39+0,82Ω
.
0,82Ω
.
0,39Ω
0,2 1 2 3 4 5
2 10 20 30 40 50
20 100 200 300 400 500
Corner load error in lb
77
eckl_4.jpg
The allocation of load cells in the weighing platform to contact on the adjustment PCB can be determined by checking
the wiring in the weighing platform.
Note: On stainless steel models, the IP67 protection must be checked after closing the
junction box.
78
79
10 12 12
25 30 30
50 60 60
100 120 120
200 250 250
80
Alternatively, you can use a feeler gauge to adjust the corner stops.
In this case, weights are not required.
81
Alternatively, you can use a feeler gauge to adjust the corner stops.
In this case, weights are not required.
Platform nominal Feeler Gauge
capacity in Lb. in inch.
50 0.12
100 0.12
200 0.12
250 0.15
500 0.12
1000 0.15
2500 0.15
5000 0.20
10000 0.15
82
– After replacing the cable or load cell, use a torque wrench to tighten the cable
gland to 5 Nm.
Note: On stainless steel models, the IP67 protection must be checked after
fct01_23.eps closing the junction box.
83
Important:
– When replacing load cells in a weighing platform with four load cells, lift the
corner using an M16x120 threaded rod to access the cells.
eckl_5.jpg
– When reassembling the weighing platform, make sure to replace any plates that
were under the platform frame.
– When replacing a load cell designated “GWT Type 011275/500 C3”, the three
bushings (two for affixing the load cell; one for the load cell base) must be
removed from the defective load cell and installed in the new load cell. These
bushings are located in bore holes in the load cell.
lastz_1.jpg
84
– After replacing load cells, you need to check the overload stops and adjust
them if necessary (see page 77-80).
85
kabelan1_e.gif
adjustment resistor
Color Codes: Load Cell Cables and Connecting Cable
(Weighing Platform to the Junction Box) (Junction Box to A/D Converter)
Load cell 1-4 Output
Ye (yellow) = Ye (yellow) = Shield = Ye (yellow)
Bl (blue) = Gn (green) = (1) BR_POS (Bridge supply voltage +) = Bl (blue)
Gn (green) = ---- = (2) SENSE_POS (Sense +) = Gn (green)
Blk (black) = Blk (black) = (6) (BR_NEG (Bridge supply voltage -) = Blk (black) or (brown)
Gr (gray) = ---- = (5) SENSE_NEG (Sense -) = Gr (gray)
Rd (red) = Rd (red) = (4) OUT_NEG (Measuring voltage negative) = Rd (red)
Wh (white) = Wh (white) = (3) OUT_POS (Measuring voltage positive) = Wh (white)
(Wiring Bridge )
Note: The color coding of the connecting cable (junction box to A/D converter) might be
different when connecting a non-Sartorius platform to the Combics indicator.
86
(Wiring Bridge )
36 x 36 Green --- White Red --- Black
48 x 48 Green --- White Red --- Black
60 x 48 Green --- White Red --- Black
60 x 60 Green --- White Red --- Black
87
88
89
90
weighing platform.jpg
91
weighing platform_2.jpg
92
Example for the order number of a Combics full range scale: CAPP1U-1000HH-LU (Type Designation see page 91-92)
CAPP1U-10..-LU ---
10 lbs 0,0002 lbs 5 lbs 0,002 lbs 5 lbs 0,004 lbs M1 1 lbs 3 lbs 0,002 lbs lbs 0,7/1,5/2,2/3 lbs 0,002 lbs 10 n
CAPS1U-10..-LU (1)
CAPP1U-25..-LU ---
25 lbs 0,0005 lbs 12,5 lbs 0,005 lbs 12,5 lbs 0,01 lbs M1 2 lbs 6 lbs 0,004 lbs lbs 1,5/3/4/6 lbs 0,005 lbs 10 n
CAPS1U-25..-LU (2)
CAPP1U-50..-LU ---
50 lbs 0,001 lbs 25 lbs 0,01 lbs 25 lbs 0,02 lbs M1 5 lbs 15 lbs 0,01 lbs lbs 3/7/11/15 lbs 0,01 lbs 10 n
CAPS1U-50..-LU (5)
CAPP1U-100..-LU ---
100 lbs 0,002 lbs 50 lbs 0,02 lbs 50 lbs 0,04 lbs M1 10 lbs 30 lbs 0,02 lbs lbs 7/15/22/30 lbs 0,02 lbs 10 n
CAPS1U-100..-LU (10)
CAPP1U-200..-LU ---
200 lbs 0,005 lbs 100 lbs 0,05 lbs 100 lbs 0,05 lbs M1 20 lbs 60 lbs 0,04 lbs lbs 15/30/45/60 lbs 0,05 lbs 10 n
CAPS1U-200..-LU (20)
CAPP1U-250..-LU ---
250 lbs 0,005 lbs 125 lbs 0,05 lbs 125 lbs 0,1 lbs M1 50 lbs 150 lbs 0,1 lbs lbs 30/70/110/150 lbs 0,05 lbs 10 n
CAPS1U-250..-LU (50)
CAPP1U-500..-LU ---
500 lbs 0,01 lbs 250 lbs 0,1 lbs 250 lbs 0,2 lbs M1 100 lbs 300 lbs 0,2 lbs lbs 70/150/220/300 lbs 0,1 lbs 10 n
CAPS1U-500..-LU (100)
CAPP1U-1000..-LU --- 300/700/
1000 lbs 0,02 lbs 500 lbs 0,2 lbs 500 lbs 0,4 lbs M1 500 lbs 1500 lbs 1 lbs lbs lbs 0,2 lbs 10 n
CAPS1U-1000..-LU (500) 1100/1500
just_combics_USA.xls
93
Example for the order number of a Combics full range scale: CAPS4U-5000KK-RR (Type Designation see page 91-92)
just_combics.xls
94