Software Update Procedure For S40 - Based On 3083 Step by Step PDF
Software Update Procedure For S40 - Based On 3083 Step by Step PDF
4. Press the “Keyboard” key and hold it, wait the “System Update” appears.
1
Software upgrade procedure
5. Choose the “System Update” and wait it finish , then shut down the S40, remove the USB disk.
6. Reboot. The update for touch panel is finished.
Before Update:
Costumer should provide SonoScape Service Engineer the SN of the S40,
if the S40 can boot up, then the costumer should also provide SonoScape Service
Engineer the CN (Control Number), the CN could be checked in the “system information”
After you send the information to the engineers of SonoScape , they will give the link to
download the software you need.
Software upgrade procedure
Costumer should also make sure that the date, the year set on the S40 is correct, if not,
costumer should change the date and the year in the Bios setting, then save the setting
(Reference the Bios Setting chapter in the “kernel update” part).
After the preparation work, below are the steps of updating the S40:
There are two ways that costumer can update the S40:
1. Normal Update without kernel 2. Update with kernel
Normal Update without kernel: Kernel version won’t change, don’t need a bootable USB
storage, the patient data won’t deleted.
Update with kernel: Kernel version will change, totally reinstall the software, the HDD will
be formatted.
Normal Update
1. Format your USB storage on a windows based PC, and copy the upgrade software like
asecureXXXXXX_3.0.8.3.tgz and adata_3.0.8.3.tgz to the USB storage.
Note: It is important to make sure that the USB storage only has one
asecureXXXXXX_3.0.8.3.tgz file.
If the USB storage has other files, the upgrade may fail.
2. Power up the system, plug the USB disk with software to S40 ; while in exam screen,
press shift, then e, then menu key one by one, the screen will display:
Software upgrade procedure
7. Press “SET”.
8. The system will show: “Files successfully copied. Please exit and reboot system to
complete update”
9. Reboot the system, and remove the USB storage, wait for the install complete (There
will be a blue screen appears, don’t worry, it’s the system installing the software.)
10. After the exam interface appears, copy the “aconfigXXXXXX” to the root directory of
an empty USB disk.
11. Insert the USB disk in the USB ports on the side of the S40.
12. press shift, then e, then menu key one by one, the screen will display:
Software upgrade procedure
(The configuration already installed, costumer can remove the USB disk now)
15. Reboot the system.
16. Copy the keypass file to the root directory of an empty USB disk.
17. Insert the USB disk in the USB ports on the side panel of the S40.
18. press shift, then e, then menu key one by one, the screen will display:
19. Input the Monthly password :JM912 (Dec., 2017) and choose “Probe Table In”
Software upgrade procedure
20. The Keypass already imported to the S40, then we need to active it.
21. Get into the engineer mode(Press “Menu” button in the EXAM interface then input
the yearly password, press enter)
PS: For this step no dialog box shows on the screen, just input the yearly password.
22. Choose “dealer config ”
Kernel Update:
Essentials:
Bootable USB drive (Appendix A)
Update files(adata_3.0.8.3.tgz, asecureXXXXXX_3.0.8.3.tgz, aconfigXXXXXX, Keypass)
External USB keyboard
Below are the steps:
1. Format your USB storage on windows based PC, and copy the upgrade software like
asecureXXXXXX_3.0.8.3.tgz and adata_3.0.8.3.tgz to “Update” folder of the USB drive.
2. Insert the USB drive into the USB port on the left side of the S40 after step1.
3. Connect the external USB keyboard to the USB port on the left side of the S40.
4. Boot up the system and set the Bios(Keep pushing “Delete” key on the keyboard
many times when the system boot up)
5. Input the password for the BIOS setting: “service”
6. Get into the Bios, choose “Hard Drive BBS Priorities” then choose the Kingston USB
drive as the 1st drive like the picture below:
Software upgrade procedure
9. The system will reboot and boot from the Bootable USB drive.
10. Wait a while until a prompt message appears on the screen
11. Input the”forcekill” or “Y” command, (forcekill means that HDD will be formatted
on this step)
then press Enter.
Software upgrade procedure
12. When the system prompt: “Recovery Complete! Press any key to shut down” press
“Enter” key on the external keyboard.
13. After the system shut down, remove the bootable USB and the external keyboard,
boot the system and wait until the Exam screen appears (There will be a blue screen
appears, don’t worry, it’s the system installing the software.).
14. After the exam screen appears, that means the software has already install , the aconfig
Software upgrade procedure
file and keypass are not installed yet, there will be no inform about the license.
The software already installed, next we need to update the configuration and Keypass file.
15. Copy the “aconfigXXXXXX” to the root directory of an empty USB disk.
16. Insert the USB disk in the USB ports in the front of the S40.
17. press shift, then e, then menu key one by one, the screen will display:
Software upgrade procedure
(The configuration already installed, costumer can remove the USB disk now)
20. Reboot the system.
21. Copy the keypass file to the root directory of an empty USB disk.
22. Insert the USB disk in the USB ports on the side panel of the S40.
Software upgrade procedure
23. press shift, then e, then menu key one by one, the screen will display:
24. Input the Monthly password :JM912(for Dec,2017) and choose “Probe Table In”
25. The Keypass already imported to the S40, then we need to active it.
26. Get into the engineer mode(Press “Menu” button in the EXAM interface then input the
yearly password, press enter)
PS: For this step no dialog box shows on the screen, just input the yearly password, and
Software upgrade procedure
enter.
27. Choose “dealer config ”
Appendix A:
How to make a bootable USB disk for S40 via Linux live creator.
1. Get the 13.6 iso file of the kernel ;
2. Install “Linux live creator”, the exe file you can get on the internet;
3. Choose the language you need during the installation;
4. Choose the USB disk you will use for the bootable USB disk.
Software upgrade procedure
9. Copy the adata and asecure files into the folder “update”