Chapter 4
Chapter 4
Gas
Sterilization
Sulaiman Alqarawneh
1
Outline
1. Ethylene oxide as sterilizing agent
2
Sulaiman Alqarawneh
ETO Sterilization
Sulaiman Alqarawneh 3
ETO Sterilization
Sulaiman Alqarawneh 4
Fact about ETO Gas
ETO gas is carcinogenic, explosive and
mutagenic.
Ethylene Oxide (ETO) is a common gas
used for low temperature sterilization.
It is a colorless, poisonous gas that attacks
the cellular proteins and nucleic acids of
microorganisms.
6
Sulaiman Alqarawneh
ETO Sterilizer
7
Sulaiman Alqarawneh
Type of Ethylene oxide sterilizers
Sulaiman Alqarawneh 8
Pure gas sterilizer.
Pure EtO is inflammable and explosive.
EtO gas is colorless, is heavier than air, has an
ethereal odor, is detectable between 230 to 700
ppm.,
and is soluble in water and in the majority of
solvents. The characteristics of EtO make the
sterilization of materials in special, controlled
conditions possible.
Sulaiman Alqarawneh 9
Mixed gas sterilizer .
Ethylene oxide and its mixtures with carbon dioxide are used
in the biomedical industry for producing plastic-disposable-
sterile equipment for hemodialysis, heart surgery, auto
transfusion, plasmapheresis, blood filtration, etc.
The sterilization rate depends on temperature, length of
treatment, gas concentration and humidity.
For maximum sterilization purposes, it is possible to work in
concentrations from 400 to 1000 mg/l.
For each use with a different purpose, experimental tests
need to be carried out each time in order to establish an
appropriate working cycle.
Sulaiman Alqarawneh 10
Mixed gas sterilizer .
Ethylene oxide is also extremely effective against
microorganisms, spores (typical form five to ten times
stronger than common bacteria), moulds and yeasts and
can inactivate viruses.
Sulaiman Alqarawneh 11
Mixed gas sterilizer .
Sulaiman Alqarawneh 12
ETO Sterilization Cycle
Preconditioning
Sterilization
Aeration (Degassing)
Sulaiman Alqarawneh 13
Preconditioning
prepares the chamber environment to meet the ideal
conditions for temperature, pressure and humidity.
First air is removed from the chamber to allow for gas
penetration.
A leakage test is performed, to ensure that staff and
environment are safe. Next, some steam is injected into
the chamber and humidifies the load, since EtO is only
effective in a humid environment.
The chamber is heated by either steam or hot water which
is present in the jacket.
Normally the jacket is kept at the same temperature 24/7
to minimize temperature fluctuations.
Sulaiman Alqarawneh 14
The second stage is the actual
sterilization process.
Sulaiman Alqarawneh 15
The second stage is the actual
sterilization process.
Sulaiman Alqarawneh 16
Aeration (Degassing)
Aeration is the most important and longest part of the EtO
sterilization cycle.
Sulaiman Alqarawneh 18
Materials Sterilized by Ethylene oxide
❑ Paper
❑ Leather
❑ Wood
❑ Metal
❑ Rubber products as well as plastics
In medicine, Its used to sterilize:
❑ Catheters artificial heart valve
❑ Heart-lung machine components
❑ Multi-lumen tubing products
❑ Stents
❑ Wound care dressings
Sulaiman Alqarawneh 19
Ethylene oxide sterilization
processes monitoring
Monitoring with biological indicators that are
composed of the microorganism Bacillus atrophaeus,
usually inoculated onto a 1.5 X 0.25 inch paper
carrier.
The biological indicators are placed throughout the
sterilizer load and subjected to the sterilization
process. After the process, the biological indicators
are removed from the load and forwarded to the
testing laboratory, where they are placed into a
special growth medium and subjected to ideal growth
conditions for 7 days
Sulaiman Alqarawneh 20
Chemical Indicator
Chemical Indicator
Initial Color: Purple
Signal Color: Green
Chemical Indicator Inks contain no lead or other toxic
heavy metals
22
Sulaiman Alqarawneh
Disadvantages of ETO Sterilization
- Excessively Long cycle
- Safety concerns - carcinogenic to humans
24