Guidelines For Measuring Turbidity
Guidelines For Measuring Turbidity
Guidelines For Measuring Turbidity
1. Background
1.1 Introduction
Turbidity estimates the impact of light scattering due to particles on the level of light attenuation in the
water column. It is affected by total suspended solids (i.e., organic and inorganic particles) in the water, and
thus differs from water transparency, which is also dependent on the amount of chromophoric dissolved
organic matter (CDOM).
Organic particles are mainly autochthonous (phytoplankton, zooplankton, bacterial cells) and thus are
related to the trophic state whereas inorganic particles are largely allochthonous (e.g., resuspension, fluvial
sediment load). This source of error has to be taken into consideration whenever eutrophication status is
assessed using turbidity in the Baltic Sea that is optically classified as a Case II water body (Morel and Prieur
1977), i.e., the body where concentrations of colour producing substances (e.g. phytoplankton, inorganic
particles and CDOM) vary independently from each other. For turbidity, this feature is emphasized in
coastal areas subject to fluvial impact. The representativeness of turbidity as an eutrophication metrics
increases towards offshore areas; here, an increasing share of particles causing turbidity is of
autochthonous origin. This has been utilized in the ship-of-opportunity (SOOP) approach.
The scope of this guideline is turbidity, measured either with turbidity meters in a laboratory or with
turbidity sensors on board research vessels or in the SOOP and other autonomous systems.
Turbidity provides information of total suspended solids that can serve as a diagnostic of eutrophication
(with certain limitations mentioned above). It is thus an element of eutrophication monitoring, although
only as a supporting parameter.
2. Monitoring methods
2.1 Monitoring features
Turbidity is expressed in formazine nephelometric units (FNU; measurement of diffuse radiation, applicable
to water of low turbidity). It measures the radiation back scatter introduced by the total suspended solid
content of the sample water against the scatter caused by the formazine polymer suspension of known
concentration. Turbidity in FNU is measured with an infrared light source.
Turbidity monitoring is required for the projects including land reclamation or building of transportation /
communication infrastructures close to marine protected areas or sensitive habitats when it is important to
limit the extent of relocation of suspended matter from not harming the marine life.
The monitored area covers, in principle, the entire Baltic Sea, both the open sea and coastal areas, but only
some of the HELCOM Contracting Parties are monitoring turbidity.
Page 1 of 3
Guidelines for monitoring of turbidity
In situ measurements
Turbidity is measured in situ either in the flow-through water within the SOOP systems or in the water
column by the CTD systems. The instrument has to be calibrated according to the manufacturer's
calibration instructions. The turbidity value is to be read directly from the instrument scale, if the scale has
been verified by calibration
Depending on the magnitude of a turbidity value, the results are reported as folIows:
4. Quality control
4.1 Quality control of methods
Laboratories carrying out turbidity analyses should have established a quality management system
according to EN ISO/IEC 17025.
A turbidity meter and a turbidity sensor are calibrated at constant intervals against a dilution series of a
traceable commercial standard.
In a laboratory premises, the daily quality control is carried out by commercial gel control samples
producing data for X-bar charts. Alternatively, a secondary standard suspension can be used as a daily
calibration check, being monitored periodically for deterioration using one of the primary standards.
Repeatability of the measurement is inspected by replicate measurements amongst the sample series.
Contracting Parties should follow the HELCOM monitoring guideline but minor deviations from this are
acceptable if the method achieves comparable results. Validation of the adopted method needs to be
Page 2 of 3
Guidelines for monitoring of turbidity
performed on the relevant matrix and concentration range e.g. by taking part regularly at inter-comparison
studies or proficiency testing schemes.
The correlation between the measured turbidity and the amount of suspended matter in situ in the water is
not constant due to qualitative variations in the particle content, and hence, in the scattering properties of
the sampled water mass. Thus, this correlation can only be fixed for an individual region / location / time
period.
Measurement uncertainty should be estimated using ISO 11352 standard. Estimation should be based on
within-laboratory reproducibility, data from proficiency testings, and internal / commercial reference
material.
Data must be flagged if normal QA routines or recommended storage conditions cannot be followed.
5.2 References
EN ISO/IEC 17025*: General requirements for the competence of testing and calibration laboratories
ISO 11352*: Water quality – Estimation of measurement uncertainty based on validation and quality control
data
Morel A, Prieur L (1977). Analysis of variations in ocean color. Limnology and oceanography 22: 709-722.
* For undated references, the latest edition of the referenced document (including any amendments) applies