Part Number: RAN456D: Pentra C200 Software V1.11 Release Note
Part Number: RAN456D: Pentra C200 Software V1.11 Release Note
TN
Part Number: RAN456D
Pentra C200
Software V1.11 Release Note
QUAL-QA-TEMP-0220_rev15
Date: 29/1/14
ECR: 5393
Software V1.11 Release Note TN: RAN456D
Intervention
Intervention priority Mandatory on next visit - Immediate and mandatory
information
Tracking number TN-RAN456D Intervention estimated duration: 0h30
on instrument serial #
Applied in factory
since instrument serial #
adjustments
documents
2/8
Software V1.11 Release Note TN: RAN456D
The V1.11 software version must be installed on Pentra C200 that already have the V1.10 software version
and the V1.54 ISE software version (refer to the "RAN456C: Software V1.10 Release Note" technical note).
■ Check the software version on your instrument in the Version screen (Services > Customer Services > Version).
Please note that you can install the V1.11 software version without installing the V1.10 software version
before, on condition that your instrument is not equipped with an ISE module and it already has the V1.07
software version.
However, you must update the customer documentation and train the customer for the V1.10 software
version (refer to the "RAN456C: Software V1.10 Release Note" technical note).
■ Check the OS version and its compatibility with the language to be installed.
- For EN and FR: the V1.11 software version can be installed on an instrument equipped with OS 2550414106 (DOM
1GB) or OS 2550498107 (DOM 1GB) or OS 2550584108 (DOM 2GB).
- For other languages, the V1.11 software version can only be installed on an instrument equipped with OS
2550584108 (DOM 2GB).
■ If the OS version is not compatible with the language to be installed, update your instrument with the DOM Ref.
F10B40431.
3/8
Software V1.11 Release Note TN: RAN456D
4/8
Software V1.11 Release Note TN: RAN456D
■ Reconfigure the ABX Pentra Clean-Chem CP in both DET.W bottles (probe and mixer) in the Reagent Tray screen
(Inventory > Reagent Tray).
5/8
Software V1.11 Release Note TN: RAN456D
■ (Optional) If your instrument is equipped with an ISE module, configure the chloride electrode as follows:
- Go to Services > Customer Services > Sequence.
- Select ON or OFF for Cl Electrode (depending on individual customer needs) and press Exec.
■ Make sure that you have access to the Pentra C200 User Manual on the customer USB key when you press Help
from the generic toolbar.
■ Check the pipetting accuracy and precision of your instrument. Refer to the "RAS595: Check up after intervention"
of your technical manual.
6/8
Software V1.11 Release Note TN: RAN456D
The bug identified in the V1.10 software version has been corrected. Refer to the "RAN731A: V1.10 software bugs list"
service information letter.