Catalogue Nuve 20litre Autoclave

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

CMYK

OT 020-020V BENCH TOP STEAM STERILIZERS

OT 020 / OT 020V Bench Top Steam Sterilizers are designed for the sterilization
of unwrapped medical or dental instruments and materials
used in laboratories with pressurized steam.
The state of art control unit is based on programmable microprocessor
technology. Easy to use microprocessor control unit allows you to programme,
sterilization temperature and time and drying period for OT 020V.

The program is started by pushing only one key and operates automatically
without user’s interference and follows the steps of water filling, heating,
sterilization, steam discharge and drying.

For OT 020V, the material can be taken


out completely dry by means of the quick
and efficient vacuum drying function and
the air taken to the chamber passes
through 0.2 µm biological filter.

Each step of the sterilization cycle can be


followed by means of the leds
on the control panel.

The control panel includes two digital


displays for temperature and time. At any
time, the operator may know the present
temperature, the remaining sterilization
or drying time. There are also warning
leds on the control panel for open lid,
over pressure and maximum and
minimum water level in the tank.

When safety is concerned, all necessary


steps are taken:

Lid Lock: The lid is locked by starting the program and can not be opened
when the chamber is pressurized. There is no possibility of operating the
program if the lid is open or not closed properly.

Safety Valve: There is a safety valve against over-pressure.

Electrical Protection: If the chamber is over-pressurized or there is


over-temperature, the energy of the heater is cut and the user is warned.

Manometer: The pressure in the chamber can be


always followed on the manometer.

Error Messages: The user is warned by the error messages in case of the
following failures: Broken temperature sensor, electronic failure,
over-temperature, water line failure, heating failure and power failure.

The operation of OT 020 and OT 020V Bench Top Steam Sterilizers is very
simple and there is no need for any water or drain connection.
CMYK

TECHNICAL SPECIFICATIONS
OT 020 OT 020V
Temperature Range 110°C / 140°C
Temperature Sensor Fe-Const
Control System Programmable Microprocessor
Temperature Set and Reading Sensitivity 1°C
Sterilization Timer 1-99 minutes
Vacuumed Drying Timer N/A 10-60 minutes
Useful Volume, liters 20
Internal Material Stainless Steel
External Material Epoxy-polyester painted steel
No of Shelves, (standard/max) 2/5
Power Consumption 1500 W 1550 W
Power Supply 230 V, 50 Hz
Internal Dimensions (QxD) mm 270x390
External Dimensions (WxDxH) mm 535x620x495
Packing Dimensions (WxDxH) mm 690x680x560
Net / Packed Weight kg. 60 / 66

ACCESSORIES

R01 062 S.S Shelf

We reserve the right to modify the design and description

NÜVE SANAYİ MALZEMELERİ


İMALAT VE TİCARET A.Ş.
Esenbo¤a Yolu, 22 km.
Akyurt 06287 ANKARA TURKEY
Tel : (90.312) 399 28 30-31-32 - 399 24 87
03.11/Z15. K34 048

Fax : (90.312) 399 21 97


0473 http ://www.nuve.com.tr
e-mail : [email protected]