Ch01-Intro PT 2 PDF
Ch01-Intro PT 2 PDF
PRINCIPLES OF
INSTRUMENTATION
AND
MEASUREMENT
WEEK
CHAPTER
1 & 12 – INTRODUCTION (PART 2)
INSTRUMENT PERFORMANCE & STATISTICAL ANALYSIS
LECTURER: KYAIRUL AZMI BAHARIN
LEARNING OUTCOME
◦ By the end of the lesson the student will be able to:
a) explain about the static and dynamic characteristics of
measurement
b) describe about the errors that can occur in measurements
c) calculate error parameters.
d) apply basic statistical analysis principles to verify measurement
results.
Instrument Performance Characteristics
◦ Allow users to select the most suitable instrument for a specific
measuring task.
◦ Two basic characteristics:
1. Static – measuring a constant/steady state process condition
2. Dynamic – measuring a varying process condition
1. Static Characteristics
1.1 Accuracy 1.7 Threshold
1.2 Precision 1.8 Resolution
1.3 Tolerance 1.9 Sensitivity to disturbance/Drift
1.4 Range 1.10 Hysteresis
1.5 Linearity 1.11 Dead space
1.6 Sensitivity of measurement
1.1 Accuracy (and Inaccuracy)
◦ Instrument Accuracy = a measure of how close the output
reading is to the correct value.
Yn − X n
% error = 100%
Yn
Yn = expected value
X n = measured value Yn − X n
relative accuracy, A = 1 −
absolute error, e = Yn − X n Yn
80 − 79
ii. % error =
80
100% = 1.25%
1 79
iii. relative accuracy, A = 1 − = = 0.9875
80 80
where, X =
X n
n
1.2 Precision Meas. No,
n
Meas. Value
Xn
1 98
Example 1.2: 2 101
3 102
From the value given in the table, calculate
4 97
the precision of the 6th measurement.
5 101
X=
X n
=
1005
= 100.5
6 100
n 10 7 103
8 98
X −X 100 − 100.5
P = 1− 6 = 1− = 0.995 9 106
X 100.5
10 99
1005
Accuracy and Precision Relationship
1.3 Tolerance
◦ Closely related to accuracy.
◦ Defines the maximum error that is to be expected in some value.
◦ Usually written with a symbol
Example: 100 5%
means that when you measure the resistance, the reading may have
a value between 95 and 105 .
◦ Sometimes the upper and lower tolerance may be different.
Example: 230 V (+10% −6%)
means the voltage reading may be between 216.2 V to 253 V.
1.4 Range
◦ Defines the minimum and maximum values of a quantity that the
instrument is designed to measure.
◦ The range must match the quantity we want to measure.
1.5 Linearity
◦ Output reading of an instrument should be linearly proportional
to the quantity being measured.
◦ Make it easy to interpret the value given by the instrument.
1.6 Sensitivity of measurement
◦ Sensitivity refers to how much the instrument output changes
when the quantity changes by a given amount.
scale deflection
Sensitivity, s =
value of measurand producing deflection
Example:
The following resistance values of a platinum resistance
thermometer were measured at a range of temperatures.
Determine the measurement sensitivity in ohms/C.
7
s= = 0.233 / C
30C
1.7 Threshold
◦ When an input to a measuring instrument is gradually increased
from zero, the input will have to reach a certain minimum level
before it can be detected by the instrument.
◦ This minimum level is known as threshold of the instrument.
◦ It can be in absolute values or as a percentage of full-scale
reading.
1.8 Resolution
◦ The smallest change that can be measured.
More uncertainty
Better resolution,
More precise.
1.9 Sensitivity to disturbance / Drift
◦ Environmental changes that affect measurement
◦ Two main ways, zero drift and sensitivity drift.
Mistake Error
Mistakes are an accident and is less formal.
Error is due to lack of knowledge and is more formal.
Types of Static Errors
c) Random errors
◦ Errors due to unknown causes
◦ Normally small but unpredictable.
◦ Even after you control all the environment parameters and conduct
the measurement as careful as possible, there will still be some
variation in your reading.
Sources of Errors
1. Insufficient knowledge of process parameters and design
conditions.
2. Poor design.
3. Change in process parameters, irregularities, etc.
4. Poor maintenance
5. Errors caused by person operating the equipment
6. Design limitations.
2. Dynamic Characteristics
◦ When measuring is done, there is usually a time delay between measuring
and obtaining the reading.
◦ Many quantities fluctuate with time or have transient conditions.
Calculate: i) Arithmetic mean ii) Deviation of each value ii) Algebraic sum of
the deviations
No Value Deviation, d
1 49.7 49.7 – 49.86 = − 0.16
2 50.1 50.1 – 49.86 = 0.24
3 50.2 50.2 – 49.86 = 0.34
4 49.6 49.6 – 49.86 = − 0.26
5 49.7 49.7 – 49.86 = − 0.16
249.3 0
249.3
x= = 49.86
5
Average Deviation
◦ Indication of how precise is the instrument used in the
measurement.
◦ Highly precise instruments has a low average deviation between
readings.
Average Deviation
Example 1.4:
Calculate the average deviation for the data given in Example
1.3.
Standard Deviation,
◦ A measure of the amount of variation of dispersion.
◦ A low standard deviation indicate that the values tend to be
close to the mean
◦ A high standard deviation indicate that the values are spread
wider apart.
For N < 30 For N 30
( dn ) ( xi − x ) ( dn ) ( xi − x )
2 2 2 2
= = = =
N −1 N −1 N N
( dn )
2
249.3 0.292
x= = 49.86 = = = 0.073 = 0.27
5 N −1 5 −1
Standard Deviation,
◦ Is also used to measure confidence in statistical conclusion.
Standard Deviation,
Example 1.6:
The normal distribution below is obtained after doing 1000
measurements. If = 10, what is the percentage of data that falls
between value 70 and 90?
Distribution of Measurement
Example 1.7:
Voltage across a resistor is measured 51 times and the results are given in the
following table. Analyze the data using basic statistical analysis.
Distribution of Measurement
Limiting Error
◦ Most measuring instrument manufacturer specify accuracy within
a certain % of full-scale reading. E.g. 2% of full-scale deflection.
◦ This specification is called the limiting error.
◦ It means a full-scale reading, the error is guaranteed to be within
2%.
◦ However, for a reading less than the full scale, the limiting error
increases.
Limiting Error
Example 1.8:
A 600 V voltmeter is specified to be accurate within 2% at full scale.
Calculate the limiting error when the instrument is used to measure a
voltage of 250 V.
Example 1.9:
A 500 mA ammeter is specified to be accurate within 2% at full scale.
Calculate the limiting error when the instrument is used to measure 300
mA.
Summation of Error
◦ Maximum absolute error in sum of the two quantities is equal to
the sum of the absolute errors in the individual quantities.
When, Any mathematic operation that used A
• A and B = the value of instrument A and B and B will produce a summation error as:
• ΔA and ΔB = the limited or absolute error
for A and B Math Operation Summation of error
• %ΔA and %ΔB = average error for A and B A+B
error = (A+B)
A–B
So, the unit and error for A and B can be AxB
written as : % error = (%A+%B)
A/B
• A = A + ΔA = A + % ΔA
• B = B + ΔeB = B + %ΔB
Summation of Error
Example 1.10:
Given for certain measurement, a limiting error for voltmeter at 70V is 2.143%
and a limiting error for ammeter at 80mA is 2.813%. Determine the limiting error
of the power.
Solution:
P = V x I = 70 V x (0.08 A) = 5.6 W
P = 5.6 (%V+% I) = 5.6 W (2.143% + 2.813%) = 5.6 W 4.956%
Summation of Error
Example 1.11:
A 150 V voltmeter and 100 mA ammeter are used to measure voltage and
current in a circuit. The dc source is in series with a resistor. Both of these
instruments has the accuracy of ±1% full scale. If the reading on the voltmeter is
80 V and the reading on the ammeter is 70 mA, calculate the power and its
limiting error.
1% 150V = 1.5V
Solution: 1.5V
100% = 1.875%
80V
1% 100mA = 1mA
1mA
100% = 1.429%
70mA
P =(80 V x 70 mA) (1.875% + 1.429%)
= 5.6 W (3.304%)
Summation of Error
Example 1.12:
A Current of 2.0 ± 0.01 A flows through a resistor of 100 ± 0.2 Ω.
Calculate the power dissipated and the limiting error.
Solution:
1.2
Power dissipated, P = I R
2 Limiting error, P = 400 W = 4.8 W
100
= (2A) 2 100 = 400 W
P = 400W 1.2% (percentage)
0.01 P = 400W 4.8W (absolute)
% I = 100% = 0.5%
2.0
0.2
% R = 100% = 0.2%
100
% P = % I + % I + % R = 1.2%