Calibration of Thermal Manikin
Calibration of Thermal Manikin
Calibration of Thermal Manikin
Calibration of the manikin is essential for the value of measuring results. The manikin has no
precision from beginning, but has to “learn” which temperatures the incoming digital values are
related to.
The principle in calibration is to bring the manikin body to two well known temperatures with a
minimum difference of 10 dgr. and insert these known temperatures in the manikin software.
Since we want to know the temperature of the body, we must bring the body to be equal to the
ambient temperature, to use the measured air temperature as known body temperature.
In order to obtain T. amb = T. body, it is important to minimize the thermal resistance from air
to body.
To obtain that, the body must be nude and placed in an open chair, that does not insulate or
stop air movement. Air velocity must be high by using fans. Feet should not rest on the flour,
but placed on eg.a box.
Place a good quality mercury thermometer near the manikin for double checking chamber
temperature.
Set the manikin in no heat mode.
Press body section – calibrate – A and type chamber temperature (can be done more than once)
Set chamber temperature B to 30 to 37 Dgr. C. ( B – A must be more than 10 dgr. C.)
Wait for steady state using graf range one dgr (eg.33 and 34 dgr.)
Press body section – calibrate – B and type air temperature.
The digital number for each body section for the two measured temperatures has now been stored in
memory in the controllers inside the body.
In the controllers the two known measure points will shape an equation, which is the temperature
calibration.
This is useful in case the calibration is accidentally damaged by user, or if a user wants to send the
copy of manikin memory to PT-Teknik for analyzing.
A stored calibration can be uploaded to the manikin by firmware.exe by using import – browse –
controller – transfer to all controllers.