ProComSol HM-USB-ISO User Manual
ProComSol HM-USB-ISO User Manual
ProComSol HM-USB-ISO User Manual
Software Drivers
Before use, a USB Virtual Serial Port must be installed on the PC. Insert the enclosed
CD into your PC. The USB Virtual Serial Port driver should install automatically. Once
the driver is installed, you can connect the HM-USB-ISO to the USB port on your PC.
Windows XP
After the HM-USB-ISO is connected and the New Hardware Found wizard is finished,
the following will appear at the bottom of the screen:
This port assignment can be changed by the user. From the “Device Manager” menu,
select the USB Serial Port you wish to change. From the “Properties” page, select “Port
Settings”. Then select “Advanced”. The Com port can be set in the range 1-256.
HART Connections
The modem can be connected in one of two ways: across the loop load resistor (A – B)
or across the HART transmitter terminals (C – D). See Figure 1.
USB Connection
Connect the integral USB-A connector to any available USB port on your PC. The
module supports both USB 1.1 and USB 2.0.
Power
No external power sources are required for the HM-USB-ISO. Power is derived from the
USB connection to the PC.
Test Software
Program “HM Test” is included on the installation CD to test the operation of the HM-
USB-ISO. Launch the program from the CD or from the installed icon. Enter the com
port that was assigned to the modem by Windows. Then select “Poll HART Network” to
connect to a HART device. The program sends HART Command 0 to determine what
transmitters are connected to the loop. The “Status” box will indicate successful
operation of the HM-USB-ISO in your system. Consult the “Troubleshooting” section of
this manual if test failure.
Troubleshooting
Verify the following:
1. Com port number in application is the HM-USB-ISO com port number.
2. Loop power supply is on.
3. Loop resistance between 250 ohms and 1Kohms.
4. Loop current within HART limits.
5. If multi drop configuration, all transmitters in loop have unique addresses.
6. HM-USB-ISO HART connections across loop resistor or across transmitter terminals.
7. HM-USB-ISO USB connector seated firmly in PC USB connector.
Warranty
The HM-USB-ISO is warranted for 1 year for materials and workmanship. Contact
Support at ProComSol, Ltd if having trouble. An RMA (Return Material Authorization)
number obtained from ProComSol, Ltd is required on all returned items.
Contact Information
ProComSol, Ltd
Process Communications Solutions
13000 Athens Ave.
Suite 104G
Lakewood, OH 44107
USA
Phone: 216.221.1550
Fax: 216.221.1554
Email: sales@procomsol.com