TES - 3060/3063
INSTRUCTION MANUAL
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
MRC.7.14
CONTENTS
I. SAFETY INFORMATION .........................................................1
Safety symbols
Dangerous voltages.
Meter is protected throughout by double insulation or
reinforced insulation.
When servicing, use only specified replacement
parts.
1
II. TECHNICAL SPECIFICATIONS
2-2 Maintenance
Repairs or servicing not covered in this manual should
only be performed by qualified personnel.
Periodically wipe the case with a dry cloth. Do not use
abrasives or solvents on this instruments.
2-3 Features•
RS-232 interface to talk with PC
4000 continuity point data logger , 25 point manual
data logger
True Power / Apparent Power / Power Factor
measurement
True RMS. V, A, W, lead, Iag, PF indicator
4 digits with dual display LCD
MIN / MAX Measurement
Amp / Voltage frequency measurement
PEAK HOLD detector
Dual display KW + PF , KVA + PF , V+A
Dual display A + Hz , V + Hz
3
2-4 General Specifications
Maximum voltage between any terminal and earth ground
:600Vrms.
Numerical dual display:4 digit liquid crystal display
(LCD) maximum reading 9999.
Bargraph display :40 segments.
Battery life :30hr approx. (Alkaline).
Low battery indication :The is displayed when the
battery voltage drops below the
operating voltage.
Auto power off time :30 minutes approx.
Sampling rate :5 times / sec ( Barograph ),
5 times / sec ( Digital display )
1 times /3sec ( on kw function )
Jaw opening diameter :Cables φ 46mm.
Operating temperature :0℃to 50℃ (32℉ to 122℉)
and humidity R.H. < 80% non-condensing.
Storage temperature :-10℃to 60℃ (14℉to 140℉)
and humidity R.H. < 70% non-condensing.
Dimensions :260(L) x 93(W) x 45mm(H).
Weight :Approx. 450 g.
Accessories :Carrying case, Test leads,
Battery (one 006p 9V) &
Instruction manual.
Optional Accessory : RS-232 interface & computer
P.C. software.
4
2-5 Measurement Specifications
Accuracy are•±(• of reading • number of digits) at 18• to
28• ( 64• to 82•) with relative humidity to 80•.
TRUE power & Apparent power measurement
Frequency Overload
Input Resolution Accuracy
range protection
V<130V , A<150A
V>130V , A<150A 0.01 ±(2%+5) 45Hz〜
600V/1100A
V<130V , A>150A 500Hz
V>130V , A>150A 0.1 ±(2%+1)
DCA
Range Resolution Accuracy Overload protection
1000A 0.1A ± ( 1.5% + 5) 1100A
ACA
Frequency Overload
Range Resolution Accuracy
range protection
1000A 0.1A ± ( 1.5% + 5) 45Hz〜500Hz 1100A
※ Crest factor : < 3 for stated accuracy
DCV
Input Overload
Range Resolution Accuracy
impedance protection
600V 0.1V ± (0.5% + 5) 1MΩ 600V
ACV
Input Frequency Overload
Range Resolution Accuracy
impedance range protection
45Hz〜
600V 0.1V ± ( 0.5%+5) 1MΩ 600Vrms
500Hz
※ Crest factor : < 3 for stated accuracy
5
Peak indication
Range Resolution Accuracy Overload protection
20A~80A 0.1A ± (10% + 10) 1100A
80A~1000A 0.1A ± ( 6% + 10) 1100A
※Peak detect acquisitions time ≦1ms
Peak indication
Range Resolution Accuracy Overload protection
20V~80V 0.1V ± (10% + 10) 600Vrms
80V~600V 0.1V ± ( 6% + 10) 600Vrms
※Peak detect acquisitions time ≦1ms
Audible continuity
Continuity Overload
Range Open circuit voltage
beeper protection
< 0.050 ≦3.2V 600Vrms
Resistance ( Ω )
Open circuit Overload
Range Resolution Accuracy
voltage protection
10KΩ 1Ω ± ( 1 ﹪ + 5) ≦ 3.2 V 600Vrms
Diode test
Range Resolution Overload protection
0.001V 600Vrms
Frequency ( Hz )
Voltage Overload
Range Resolution Accuracy
sensitivity protection
1KHz 0.1Hz
± (0.5% + 5) 10V or 10A 600V / 1100A
5KHz 1Hz
Optional accessories:
RS-232 interface adaptor / software
6
III. PARTS & CONTROLS
HOLD
Hz
V OFF
1000A
KW
3 3W
3 4W
W L 1 23 ME M R E A D P P- H O L D MA X - MIN
3 3W 3 4W Hz
KV A
PF lead lag
RE C O RD
KW
KV A
K
(Figure-1)
7
(1). Transformer jaws
Pick up the AC and DC current flowing through the conductor.
(2). Jaw opening trigger.
(3). Data Hold button
Press it once to hold the measured value and store the
value in memory. Press again to release hold function.
(4). Function selector
For selection of desired function, and awakening from auto-
power-off mode.
(5). Yellow button (P+ or P- selection)
a. Press “Yellow” button once, LCD will display “P+P-”
symbol, then press “PEAK” button to measure peak
value of a transient signal.
b. Press “Yellow” button until the READ symbol is shown on
LCD and then Press “•” button to read the previous data.
Press “•” button to read the next data. Press “Yellow”
button again to exit.
c. When the rotary switch is set to KW function, press this
button to select KW/PF, A/V or KVA/PF dual displayed.
d. When the rotary switch is set to the 3•3W or 3•4W
function, press on the yellow button to select the phase
to be measured.
(6). MIN/MAX/PEAK function button
a. Press it once to select MIN or MAX. Keep pressing it for
2 seconds to exit MIN/MAX mode.
b. Press “Yellow” button once then press this button to
measure peak value P+ or P- of a transient signal, keep
pressing it for 2 seconds to exit peak mode. It can be
used in ACV and ACA. One of the most common
applications is the measurement of the current value
when starting electrical motors. When measuring
alternating signals, LCD shows the peak value.
9
W L 1 23 ME MR E A D P P- H O L D MA X - MIN
3 3W 3 4W Hz
KV A
PF lead lag
RE C O RD
KW
KV A
K
(Figure-2)
10
IV. AC+DC POWER MEASUREMENT
WARNING
Wait until “- - - -” symbol is shown on LCD before clamping
on to any conductor and pressing on the ZERO button to
zero any residual magnetic field in the jaws.
HOLD
Hz
V OFF
1000A
KW
3 3W
3 4W
PF lag
KW
PF lag
KW
(Figure-3)
NOTE
The "+" sign printed on jaw must face the power source for
accurate measurement.
12
4-2 3φ3W AC+DC Power Measurement
HOLD
Hz
V OFF
1000A
KW
3 3W
3 4W
W L1 2
3 3W
PF lag
KW
W L1 2
W L1 2
3 3W
PF lag
(Figure-4)
13
HOLD
Hz
V OFF
1000A
KW
3 3W
W L 23 3 4W
W L 23
3 3W
PF lag
3 3W KW
W L 23
3 3W
PF lag
KW
(Figure-5)
W L 1 23 H O LD
3 3W
PF lag
KW
(Figure-6)
14
First, measure WRS(L1L2) (refer to figure 4).
a. Turn the power on without clamping to any wire.
b. Set the rotary switch at 3 φ 3W, and WL12 symbol will
appear to instruct users to take measurement of WRS(L2L1).
c. Insert the test leads into the input jack.
d. Select one phase (eg. S or L2) as COM and connect the
test probe of the COM (black) terminal to that phase (eg.
S or L2).
e. Connect the test probe of V (red) terminal to the second
phase (eg. R or L1).
f. Clamp the same phase as step e. (eg. R or L1).
g. The power clamp will automatically select proper range.
h. Wait until the reading is stable (about 3~6 seconds),
press the YELLOW button and WL23 symbol will appear
to instruct users to take measurement of WTS (WL3L2).
15
The power clamp will process those two sets of data (WL12,
WL23), and show the result on the LCD. WL123 symbol will
be shown to indicate the watt of 3φ3W power.
At this moment, the watt of 3φ3W power is stored into the
memory.
If willing to read the details of that singly data record,
please refer to 5-9.
W3φ3W = WRS(L1L2) + WTS(L3L2)
KVA 3φ 3W = K W 2 3φ 3W + KVAR 2
3 φ 3W
KW 3φ 3W
PF 3φ 3W =
KVA 3φ 3W
If willing to record data (WL12 or WL23), press “RECORD”
button to complete.
NOTE
Once a phase is selected as COM, users can not change this
selection in the subsequent measurement. For example, if S (or
L2) phase is selected, S (or L2) phase is always connected to the
COM during measurement of WRS (or WL1L2) and WTS (or WL3L2)
in 3φ 3W unbalanced power.
NOTE
The "+" sign printed on jaw must face the power source, and
make sure all the connections and clamping are correct for correct
measurement.
NOTE
In the 3 φ 3W unbalanced power measurement, one of WRS or
WTS could be negative. Users must make sure all the
connections and clamping are correct to obtain correct power.
16
4-3 3φ4W AC+DC Power Measurement
HOLD
Hz
V OFF
1000A
KW
3 3W
3 4W
W L1
3 4W
W L1
3 4W PF lag
W L1 PF lag KW
3 4W
KW
(Figure-7)
HOLD
Hz
V OFF
1000A
KW
3 3W
3 4W
WL 2
3 4W
PF lag
WL 2 KW
WL 2
PF
3 4W
lag
3 4W KW
(Figure-8)
17
HOLD
Hz
V OFF
1000A
KW
WL 3 3 3W
3 4W
WL 3
3 4W
PF lag
3 4W
KW
WL 3
3 4W
PF lag
KW
(Figure-9)
W L 1 23 H O LD
3 4W
PF lag
KW
(Figure-10)
18
First, measure WR(L1) /PFR(L1) (refer to figure 7).
a. Turn the power on without clamping on to any wire.
b. Set the rotary switch at 3φ4W.
c. Insert the test leads into the input jack.
d. Connect the neutral line to the COM (black) terminal.
e. Connect the test probe of the V (red) terminal to the first
phase (eg. R or L1).
f. Clamp on to the same phase (eg. R or L1).
g. The power clamp meter will automatically select proper
range.
h. Wait until the reading is stable (about 3~6 seconds),
press the YELLOW button, and WL1 symbol will disappear.
and WL2 symbol appears to instruct users to take
measurement of WS(L2)/PFS(L2).
KVA 3φ 4W = K W 2 3φ 4W + KVAR 2
3φ 4W
KW 3φ 4W
PF 3φ 4W =
KVA 3φ 4W
NOTE
The "+" sign printed on jaw must face the power source. Make
sure all the connections and clamping are correct for correct
measurement.
NOTE
In the 3φ4W power measurement, all three WR or WS and WT
must be positive. If users find one negative power, check the
connection of test leads and clamping of jaw. Make sure all the
connections and clamping are correct to obtain correct power.
4-4 1φ3W Power Measurement
20
HOLD
Hz
V OFF
1000A
KW
3 3W
3 4W
W L1 2
3 3W
PF lag
KW
W L1 2
W L1 2
3 3W
PF lag
3 3W KW
(Figure-11)
21
HOLD
Hz
V OFF
1000A
KW
W L 23
W L 23
3 3W 3 3W
PF lag
3 4W
3 3W KW
W L 23
3 3W
PF lag
KW
(Figure-12)
W L 1 23 H O LD
3 3W
PF lag
KW
(Figure-13)
22
1φ3W power measurement is similar to 3φ3W unbalanced
power measurement except the nomenclature is different.
Two measurements of WRS(L1G) and WTS(L2G) are required.
First, measure WRS(L1G) (refer to figure 11).
a. Turn the power on without clamping to any wire.
b. Set the rotary switch at 3φ3W.
c. Insert the test leads into the input jack.
d. Connect the test probe of the COM (black) terminal to
ground.
e. Connect the test probe of V (red) terminal to the second
phase (eg. L1).
f. Clamp the same phase as step e. (eg. L1).
g. The power clamp will automatically select proper range.
h. Wait until the reading is stable (about 3~6 seconds),
press the YELLOW button, and WL23 symbol will appear
to instruct users to take measurement of WTS(L2G).
Second, measure WTS (or WL2G) (refer to figure 12).
a. Disconnect the test probe from the phase where jaws
clamped in the previous measurement.
b. Connect the test probe to the L2 line.
c. Clamp the L2 line where test probe is connected to.
d. The power clamp will automatically select proper range.
e. Wait until the reading is stable (about 3~6 seconds), press
the YELLOW button.
The power clamp will add the two values together and show
the result on the LCD. At this moment the watt of 1φ3W
power is stored in the memory.
W1φ3W = WRST = WRS(L1G) + WTS(L2G)
If willing to read the details of that singly data record,
please refer to 5-9.
If willing to record data (WL1G or WL2G), press “RECORD”
button to complete.
23
V. OPERATING INSTRUCTION
HOLD
Hz
V OFF
1000A
KW
3 3W
3 4W
Hz
(Figure-14)
WARNING: Maximum input is 600V. Do not attempt to take
any voltage measurement that exceeds these limits.
Exceeding these limits could cause electrical shock and
damage to the clamp meter.
Set the rotary switch to V (refer to figure 14).
Insert the test leads into the input terminal.
Connect the test probes of the test leads in PARALLEL to
the circuit to be measured.
The clamp will automatically select the appropriate range.
Read the voltage and frequency values displayed on the LCD.
NOTE
The sensitivity for voltage frequency measurement is 10V,
and the frequency range is 45 - 500Hz. If the frequency is
less than 45 Hz, the LCD will show -.- Hz.
24
5-2 AC+DC Current Measurement
HOLD
Hz Hz
V OFF
1000A
KW
A
3 3W
3 4W
Hz
600V CAT
(Figure-15)
25
5-3 To Improve Power Factor of a 3φ4W Power System
Calculate KVARR(L1), KVARS(L2), and KVART(L3) values of
each phase.
Based upon the calculated values, users can purchase
required 3φ or 1φ capacitor at rated voltage and frequency
to improve power factor.
If value of capacitance is needed, users can obtain the
value by the following equation.
Where
KVAR *1000
Capacitance (Farad) =
2 πfV 2
KVAR *1000
Capacitance (Farad) =
2 πfV 2
where
f : frequency in Hz, V: line voltage
26
5-5 To Improve Power Factor of a 1φ2W Power System
Calculate KVAR value of a 1φ2W power system.
Based upon the calculated value, users can purchase
required capacitor at rated voltage and frequency to improve
power factor.
If value of capacitance is needed, users can obtain the
value by the following equation.
KVAR *1000
Capacitance (Farad) =
2 πfV 2
where
f : frequency in Hz, V: line voltage
27
HOLD
Hz
V OFF
1000A
KW
3 3W
3 4W
(Figure-16)
Note : Continuity test is available to check open/short of the
circuit.
Connect the red test lead to the anode side and black test
lead to the cathode side of the diode being tested.
Read forward voltage ( Vf ) value on LCD.
If the test leads is connected rather than procedure(4), the
digital reading should nearly equal to the reading in the
open circuit condition. This can be used for distinguishing
anode and cathode poles of a diode.
When the reading is below 0.050V, it will be indicated by a
continuous beeping.
5-8 Peak Detection Measurement
28
Set Function / Range Switch to desired •A or •V range.
Set the clamp Meter into the " PEAK HOLD " mode by
pushing the "YELLOW" button. The LCD will display " P+ P- ".
Press the “PEAK” button to enter PEAK mode. LCD
will display “P+ MAX” or “P- MIN”. The “PEAK” button
is toggle selector the P+ and P- value.
Press the “PEAK” button for two seconds to exit the PEAK
mode.
Follow this procedure for AC Voltage and current measurement.
The displayed reading is the positive peak value or negative
peak value of a surge in current or a voltage transient pulse.
Because the surge transient pulse is usually asymmetric.
WARNING
To prevent electrical hazard or shock, turn off clamp meter
and disconnect test leads before removing back cover.
30