Ecg Circuit
Ecg Circuit
Circuit
Electrocardiogram (ECG) Circuit
Many electrical devices are used to measure and record biological activity in
the human body. One such device is the electrocardiogram, which measures
the electrical signals produced by the heart. These signals give objective
information about the structure and function of the heart. The ECG was first
developed in 1887 and gave physicians a new way to diagnose heart
complications. ECGs can detect heart rhythm, heart rate, heart attacks,
inadequate blood and oxygen supply to the heart, and structural
abnormalities. Using simple circuit design, an ECG can be made that could
include:
1- Breadboard
3- Resistors
4- Capacitors
5- Wires
6- diodes
7- Stick-on electrodes
Design Steps
1- protection circuit
the proposed circuit successfully limits the high voltage surges
and current spikes to 1.5 TO 2.5 Volts and 50 to 150
microamperes respectively and generates negligible leakage
current.
The low pass filter at input stage with cut-off frequency of 3.48
KHz provides electrosurgical RF unit noise cancellation and also
avoids phase distortion of ECG signal
2-Instrumentation Amplifier :
We will use the next values for sitting up the Instrumentation
op amp parameters
R1= 1.8 kΩ, R2=8.2 kΩ, R3 = 1.5 kΩ, and R4 =15 kΩ.
K1 is the gain of the first stage (OA1 and OA2), and K2 is the
gain of the second stage (OA3). Equal capacitance bypass
capacitors are used on the power supplies of the operational
amplifiers to remove noise.
3- Isolation stage:
[
5- High-pass Filter
The high-pass filter is used so that frequencies below a certain
cut-off value are not recorded, allowing a clean signal to be
passed through. The cut-off frequency is chosen to be 0.5 Hz (a
standard value for ECG monitors).
How to build it:
The resistor and capacitor values needed to achieve this are
seen below. Our actual resistance used was 318.2 kΩ
R = 1/(2*pi*f*C)
set f = 0.5 Hz, and C = 1 uF
Calculate R. Our value is 318.310 kΩ
6- low-pass Filter:
The last stage of the device is an active low-pass filter. The ECG
signal is made of many different waveforms, which each have
their own frequency. We want to capture all these, without any
high-frequency noise. The standard cutoff frequency for ECG
monitors of 150 Hz is selected. (Higher cutoffs are sometimes
chosen to monitor for specific heart problems, but for our
project, we will use a normal cutoff.)
If you would like to make a simpler circuit, you could also use a
passive low-pass filter. This will not include an op amp, and will
consist of just a resistor in series with a capacitor. The output
voltage will be measured across the capacitor.
7-Driven Right Leg
The DRL is an inverter operational amplifier with a gain of -10
that is connected to the patient via the 3rd electrode. As
commented before, this stage reduces the common mode
effect and the value of its resistances was obtained using the
following equation:
G = −R2\ R1 = −10