Valve Oven: Operating Manual
Valve Oven: Operating Manual
Valve Oven: Operating Manual
Valve Oven
For TRACE GC Ultra
Operating Manual
PN 31709363, Revision May 2010
Home
Valve Oven Operating Manual
May 2010 Edition
Part Number 317 093 63
© 2007-2010 Thermo Fisher Scientific Inc. All rights reserved.
Printed in Italy
Published by Thermo Fisher Scientific S.p.A., Strada Rivoltana, 20090 Rodano - Milan - Italy
Tel: +39 02 95059355 Fax: +39 02 95059388
Disclaimer
Technical Information contained in this publication is for reference purposes only and is subject to change without notice. Every effort has
been made to supply complete and accurate information; however, Thermo Fisher Scientific assumes no responsibility and will not be liable
for any errors, omissions, damage, or loss that might result from any use of this manual or the information contained therein (even if this
information is properly followed and problems still arise).
This publication is not part of the Agreement of Sale between Thermo Fisher Scientific and the purchaser of a TRACE™ GC Ultra system. In
the event of any conflict between the provisions of this document and those contained in Thermo Fisher Scientific’s Terms and Conditions,
the provisions of the Terms and Conditions shall govern.
Reference to System Configurations and Specifications supercede all previous information and are subject to change without notice.
Trademarks
TRACE™ GC Ultra is a trademark of Thermo Fisher Scientific Inc., and its subsidiaries. Other brand and product names may be trademarks
or registered trademarks of their respective companies.
Contents
About This Manual..................................................................................................................................v
Overview .................................................................................................................................... v
Conventions Used in This Manual............................................................................................vi
Instrument Markings and Symbols ..........................................................................................vii
Using the TRACE GC Ultra Document Set.............................................................................. ix
Chapter 1
Valve Oven Overview......................................................................................................................... 11
Introduction .............................................................................................................................. 12
Options ........................................................................................................................ 13
Installation and Power Requirements....................................................................................... 14
Power Requirements and Voltage Selection............................................................... 15
Valve Oven Components ......................................................................................................... 16
Heated Section ............................................................................................................ 17
Unheated Section ........................................................................................................ 17
Valves.......................................................................................................................... 17
Column........................................................................................................................ 17
Chapter 2
Valve Installation................................................................................................................................ 19
Introduction .............................................................................................................................. 20
Opening the Valve Oven.......................................................................................................... 20
Installing Valves in the Heated Oven....................................................................................... 21
Installing Valves in the Unheated Compartment ..................................................................... 28
Installing Valves in the Valve Box .......................................................................................... 31
Valve Box for up to Six Valves .................................................................................. 31
Valve Box for up to Two Valves ................................................................................ 31
Configuring Valves .................................................................................................................. 32
Gas Sampling Valves .................................................................................................. 32
Switching Valves ........................................................................................................ 32
Chapter 3
Operation............................................................................................................................................ 35
Operating Valves Manually ..................................................................................................... 36
Operating Valves Automatically.............................................................................................. 38
Adding a Gas Sampling Valve Event to the Run Table.............................................. 38
Adding a Gas Switching Valve Event to the Run Table............................................. 40
Chapter 4
Troubleshooting .................................................................................................................................43
Leak testing................................................................................................................. 43
Valve Switching ......................................................................................................... 44
Valve Plumbing .......................................................................................................... 45
Appendix A
Customer Communication ...............................................................................................................47
How To Contact Us ................................................................................................................. 47
Reader Survey............................................................................................................. 48
Glossary ................................................................................................................................................ 49
Index...................................................................................................................................................... 55
iv Operating Manual
About This Manual
Overview
This Valve Oven Operating Manual contains information for operating the Valve
Oven, including Troubleshooting and ordering spare parts.
Chapter 4, Troubleshooting, has information to help you find and correct potential
problems when using the TRACE GC Ultra Valve Oven.
The Index presents an alphabetical list of key terms and topics in this guide,
including cross references and the corresponding page numbers.
Operating Manual v
Conventions Used in This Manual
The following symbols and typographical conventions are used throughout this
manual.
vi Operating Manual
Instrument Markings and Symbols
The following table explains the symbols used on Thermo Fisher Scientific
instruments. Not all of them are used on the TRACE GC Ultra gas chromatograph
and Valve Oven.
Symbol Description
Direct Current
Alternating Current
Equipotentiality
On (Supply)
Off (Supply)
Operating Manual ix
Preventive Maintenance Schedule (PN 317 092 80)
This document provides a list of recommended scheduled maintenance and a
year-long log book to record maintenance, observations, supply lists, and service
records.
x Operating Manual
Valve Oven Overview
1
This chapter provides an overview to TRACE GC Ultra Valve Oven features and
components.
Chapter at a Glance...
Introduction ..........................................................................................................12
Installation and Power Requirements...................................................................14
Valve Oven Components ......................................................................................16
Operating Manual 11
Chapter 1
Valve Oven Overview Introduction
Introduction
The TRACE GC Ultra Valve Oven, shown in Figure 1-1, is a temperature
controlled enclosure for mounting up to six (four heated + two unheated)
sampling and/or switching valves, 1/8-inch OD packed columns and up to eight
needle valves and four pressure regulators.
A large number of applications require this additional space for a correct analysis
setup.
HEATED SAMPLE
TRANSFER LINE
HEATED SAMPLE
TRANSFER LINE
D H L
E I M
F J N
G K O
WARNING
USE ONLY THE ORIGINAL
POWER SUPPLY
FOR THIS INSTRUMENTS
AIR
POWER 800 kPa/120 psi Max
12 Operating Manual
Chapter 1
Introduction Valve Oven Overview
Options
The type of valve enclosure can be selected among three different options
depending on the analytical requirements.
Operating Manual 13
Chapter 1
Valve Oven Overview Installation and Power Requirements
NOTE When any Valve box (heated or unheated) is configured with valves, these are factory
tested to verify actuation but are NOT plumbed.
Plumbing can be performed upon availability of plumbing diagrams.
Materials needed:
Shop air or bottled air at a pressure of 60–80 psi
Helium, Nitrogen, or other gas depending on your applications
Power On Indicator
14 Operating Manual
Chapter 1
Installation and Power Requirements Valve Oven Overview
CAUTION AC line voltage is applied to the valve oven when the main GC is turned on. Turn the
main GC power off before removing the valve oven heaters.
NOTE In case of using the TRACE GC Ultra Valve Box for up two valves, any power supply is
necessary since the solenoid valves are activated directly by the TRACE GC Ultra.
Operating Manual 15
Chapter 1
Valve Oven Overview Valve Oven Components
5
1
3 2
16 Operating Manual
Chapter 1
Valve Oven Components Valve Oven Overview
Heated Section
The heated enclosure is a temperature controlled (40 to 175 °C) isothermally
heated zone, able to contain the following:
• up to four (4) valves with associated sample loops
• up to fifty feet of 1/8” OD stainless steel column with associated clamp and
fixed unions
• up to eight (8) heated needle valves (restrictors) accessible from the front
• provisions for sample transfer directly into the heated zone from the front or
rear of the unit
Unheated Section
Unheated enclosures can contain the following:
• two unheated pneumatic Valco liquid or gas sample injection valves
• up to four pressure regulators installed on the front of the unit, and one
mounted internally
• up to six in and out connections for sample loading/vent/flow measurement
Valves
Your TRACE GC Ultra Valve Oven has the following available valves:
• standard pneumatic Valco rotary valves
• standard pneumatic Valco rotary valves with purged housing
Column
Although the type of column you use will vary depending on your application, the
TRACE GC Ultra Valve Oven typically uses packed columns. The column
mounting area can accommodate 1/8” OD columns.
Operating Manual 17
Chapter 1
Valve Oven Overview Valve Oven Components
18 Operating Manual
Valve Installation
2
This chapter provides instructions for installing and configuring valves in the
TRACE GC Ultra Valve Oven.
Chapter at a Glance
Introduction ..........................................................................................................20
Opening the Valve Oven.......................................................................................20
Installing Valves in the Heated Oven ...................................................................21
Installing Valves in the Unheated Compartment ..................................................28
Installing Valves in the Valve Box........................................................................31
Configuring Valves...............................................................................................32
Operating Manual 19
Chapter 2
Valve Installation Introduction
Introduction
NOTE Before starting, cool the Valve Oven to 50°C or cooler. Do not turn the GC off.
The heated section of the Valve Oven accepts standard Valco gas actuated valves
with 3-inch standoff. The unheated section will accommodate Valco gas actuated
valves without standoff.
CAUTION Be careful not to mix actuators with valves that have different numbers of ports, as
the actuator travel for each valve can be different.
1. Remove the top cover by releasing a single screw on the Valve Oven
compartment rear panel.
2. Slide the panel toward the rear about ½ inch and lift away. This exposes the
heated Valve Oven cover and valve pneumatics.
3. To remove the heated Valve Oven cover, locate the two thumb screws on the
lower edge of the heated Valve Oven cover.
4. Turn these screws counter clockwise (ccw) to release the cover. The cover can
be removed by lifting up and pulling out.
20 Operating Manual
Chapter 2
Installing Valves in the Heated Oven Valve Installation
CAUTION Turn the actuator gas off before you start these procedures.
1. Locate the solenoid manifold in the rear floor of the valve compartment
(Figure 2-1). The space closest to the rear of the GC is reserved for
pneumatic valve 1, next is valve two, and so on.
A total of 6 pneumatic valves can be installed (up to 4 in the upper
compartment and up to 2 in the lower).
2. Remove the rectangular shaped cap from the manifold using the #0 phillips
screwdriver.
3. Secure the solenoid valve to the underplate using the two screws supplied
with the valve.
CAUTION Tighten screws evenly, taking care that the solenoid gasket rests on all surfaces to
prevent leaks
Operating Manual 21
Chapter 2
Valve Installation Installing Valves in the Heated Oven
1 2 3 4
1. Time Event Output Jack 2. Solenoid Manifold
3. Valve 1 Position 4. Valve 2 Position
Figure 2-1. Timed Event Output Jacks, Solenoid Manifold, and Valve Positions
4. Connect the cable from the solenoid valve (Figure 2-4) to the proper timed
event output jack (Figure 2-2) located in the lower rear inside corner of the
valve compartment.
Valve 1 is attached to timed event #8 (TE8), Valve 2 is attached to timed
event #7 (TE7), and so forth (Figure 2-2).
22 Operating Manual
Chapter 2
Installing Valves in the Heated Oven Valve Installation
TE #1 TE #2 TE #3 TE #4 TE #5 TE #6 TE #7 TE #8 Sensor
NOTE Valves are usually installed starting from the mounting hole closest to the rear of the Valve
Oven.
1. Remove any insulation found in the valve mounting hole. The valve mounting
hole is located in the lower Valve Oven heater block.
2. Inspect the valve and actuator being sure the valve is in the counter clock wise
(ccw) position. The ccw position is when the mechanical stop is in the
position noted in Figure 2-3. If the valve is not in the ccw position, locate the
square nut on the opposite end of the actuator (Figure 2-5).
Operating Manual 23
Chapter 2
Valve Installation Installing Valves in the Heated Oven
yyyyy
@@
;; @@@
;;;
Clockwise (CW) Position Counterclockwise (CCW) Position
@@yyy
;;
yy @@@
;;;
Mechanical stop
3. Using a 5/16” or 8mm wrench, rotate the nut to the ccw position (Figure 2-5).
4. Once the valve is in the ccw position, separate the actuator from the valve and
standoff by loosening the allen screw in collar B (Figure 2-5).
CAUTION Be careful to pull the actuator away from the valve standoff by gripping the standoff
but without rotating the actuator or the valve.
NOTE Handle the valve and standoff with care so the two pieces do not separate.
5. Remove the collar attached to the valve standoff (collar A, Figure 2-5) by
loosening the screw that secures the collar to the valve standoff. This collar
mounts to the Valve Oven heater block.
1. Locate the desired valve position in the heater block and place the collar over
the hole in the heater block.
2. Secure the collar using two M 4 x10 mm screws. Be sure the 7/64” set screw
in the collar is accessible from the outer edge of the Valve Oven.
24 Operating Manual
Chapter 2
Installing Valves in the Heated Oven Valve Installation
3. Grip the valve standoff by the standoff only, and slide it through the collar
allowing it to extend into the lower valve compartment.
Step 4. Connecting Gas Lines from the Solenoid Valve o the Actuator
NOTE Gas lines are much easier to connect to the actuator if the connection is made before the
actuator is assembled to the valve.
1. Locate the 1/8 in. nylon tubing supplied in the Valve Oven kit and cut it into
two equal 12” pieces.
2. Push one of the tube pieces into the lower fitting on the actuator assembly,
making sure it passes through the ferrule in the actuator fitting. Tighten the
nut and ferrule using a 3/8” open end wrench. Use a backup 3/8” open end
wrench on the mating piece that is screwed into the actuator.
3. Place the other end of the tubing into the valve solenoid fitting located on the
outside edge (Figure 2-4) of the solenoid by pushing the tubing firmly into the
fitting. It should automatically seal in place.
2
1. Inside Edge Solenoids 2. Outside Edge Solenoids
Figure 2-4. Solenoid Manifold Edges
4. Turn the actuator gas supply on. This line should now be pressurized.
Operating Manual 25
Chapter 2
Valve Installation Installing Valves in the Heated Oven
NOTE Applying pressure to this side of the actuator will ensure that the actuator stays in the ccw
position.
6. Take the other 1/8” piece of nylon tubing and place it in the upper fitting on
the actuator. Be sure the tube passes through the ferrule in the fitting. Tighten
the fitting using the 3/8” open end backup wrench.
NOTE The tubing can be released from the solenoid by pushing down on the solenoid fitting and
pulling out on the tubing at the same time.
1. Carefully fit the actuator into the valve standoff being sure not to rotate the
valve.
2. Once the two square fittings on the actuator and valve are coupled, tighten the
collar on the valve. Be sure this fitting is very tight so the actuator will not
rotate on the valve standoff.
3. Slide the valve and actuator assembly up or down to the desired location and
tighten the upper collar to hold the valve and actuator in place.
26 Operating Manual
Chapter 2
Installing Valves in the Heated Oven Valve Installation
1. Valve
1
2. Mechanical Stop Valve
2 3. Valve Standorf
3
4. M4 x 10 mm Screws
4 5
5. Collar “A”
6. Collar Set Screws
6 7. Heater Plate
7
8. Collar “B”
8
9. Actuator
10 Square Nut
10
3. Refer to Figure 2-3 and inspect the valve stop when the valve is in both
positions to ensure that the stop is at the travel limit on each side.
Operating Manual 27
Chapter 2
Valve Installation Installing Valves in the Unheated Compartment
CAUTION Turn the actuator gas off before you start these procedures.
1. Install the bracket to the lower portion of the actuator using two M4 x 10
screws (Figure 2-6).
2. Choose two mounting holes in the bracket that will allow the actuator and the
valve to be oriented in a favorable direction.
NOTE Gas lines are much easier to connect to the actuator if the connection is made before the
valve/actuator and bracket are secured to the floor of the valve enclosure.
1. Locate the 1/8” nylon tubing supplied in the Valve Oven kit and cut it into
two equal 12” pieces.
2. Push one of the tube pieces into the lower fitting on the actuator assembly,
making sure it passes through the ferrule in the actuator fitting. Tighten the
nut and ferrule using a 3/8” open end wrench. Use a backup 3/8” open end
wrench on the mating piece that is screwed into the actuator.
28 Operating Manual
Chapter 2
Installing Valves in the Unheated Compartment Valve Installation
3. Place the other end of the tubing into the valve solenoid fitting located on the
outside edge of the solenoid by pushing the tubing firmly into the fitting. It
should automatically seal in place.
4. Turn the actuator gas supply on. This line should now be pressurized.
NOTE Applying pressure to this side of the actuator will ensure that the valve and actuator rotates
to the ccw position.
6. Take the other 1/8 in piece of nylon tubing and place it in the upper fitting on
the actuator. Be sure the tube passes through the ferrule in the fitting. Tighten
using the 3/8” open end wrench with backup.
1. Locate the three sets of mounting holes in the Valve Oven compartment floor.
2. Select the desired set of holes and secure the valve bracket to the holes using
M 4 x 10 screws (Figure 2-6).
NOTE Liquid sample valves should be mounted such that the C and P ports are next to the feed-
throughs in the Valve Oven.
Operating Manual 29
Chapter 2
Valve Installation Installing Valves in the Unheated Compartment
2
3
30 Operating Manual
Chapter 2
Installing Valves in the Valve Box Valve Installation
Refer to:
• Installing Valves in the Heated Oven
• Installing Valves in the Unheated Compartment
Then, the valves installation follows the same procedure described previously for
the heated oven and unheated compartment.
Refer to:
• Installing Valves in the Heated Oven
• Installing Valves in the Unheated Compartment
Operating Manual 31
Chapter 2
Valve Installation Configuring Valves
Configuring Valves
The valves in the Valve Oven are gas-actuated. The valves are controlled by
solenoid valves that are turned on and off by timed events. Each solenoid valve
uses one timed event. Valve 1 will use timed event # 8, Valve 2 will use timed
event # 7, and so on. Once a valve occupies a timed event, this timed event is
removed from the list of external events that can be added to a run table.
yyyyy
@@
;; @@@
;;;
Load (OFF) Position Inject (ON) Position
@@yyy
;;
yy @@@
;;;
Mechanical stop
Switching Valves
Switching valves have two positions—On and Off. Switching valves are plumbed
so that the counter-clockwise position is Off and the clockwise position is On.
NOTE Typically, the valve nearest to the rear of the Valve Oven will be configured as valve 1.
32 Operating Manual
Chapter 2
Configuring Valves Valve Installation
4. Press ENTER.
6. Repeat as necessary for the remaining valves in the Valve Oven or valve
compartment.
yyyyy
@@
;; @@@
;;;
Clockwise (CW) Position Counterclockwise (CCW) Position
@@yyy
;;
yy @@@
;;;
Mechanical stop
Operating Manual 33
Chapter 2
Valve Installation Configuring Valves
34 Operating Manual
Operation
3
This chapter presents operating instructions for the TRACE GC Ultra Valve
Oven.
The valves can be operated manually using the keypad on the TRACE GC Ultra,
or automatically using the run table accessed on the TRACE GC Ultra keypad.
Chapter at a Glance...
Operating Valves Manually ..................................................................................36
Operating Valves Automatically...........................................................................38
Operating Manual 35
Chapter 3
Operation Operating Valves Manually
Valves Key
Figure 3-1. The TRACE GC Ultra Keypad, Showing the Valves Key
VALVES
Inlet Valves
#1 Gas Sample Load <
#2 Switching Off
• If the valve is a gas sampling valve, press On to place the valve in the
Inject position or Off to place the valve in the Load position.
36 Operating Manual
Chapter 3
Operating Valves Manually Operation
VALVES
#2 Switching Off
#3 Gas Sample Load
Switch vlv defaults Off<
• Scroll to the switching valve you wish to change and press On or Off to
enter its new default condition.
Operating Manual 37
Chapter 3
Operation Operating Valves Automatically
Figure 3-2. The TRACE GC Ultra Keypad, Showing the Run Table Key
38 Operating Manual
Chapter 3
Operating Valves Automatically Operation
4. Scroll to the sampling valve that you want to add and press ENTER.
The Inject at parameter is the time into the run that the sample is to be
injected. The Inject for parameter is the time that the valve remains into the
inject position. After that time, the valve returns to the Load position. If the
Inject for time exceeds the GC run time, the GC will automatically reset the
valve to the Load position at the end of the run.
5. Repeat as necessary for any remaining sampling valves. If you need to edit a
run table entry, select the entry to be edited and press ENTER.
NOTE To delete a run table entry, select it and press CLEAR. The TRACE GC Ultra will ask you if
you really want to delete the entry. Press YES.
Operating Manual 39
Chapter 3
Operation Operating Valves Automatically
NOTE Pressing CLEAR within a run table entry will delete that entry. To use CLEAR to back out
of the menu, you must first be out of the run table entries themselves.
4. Scroll to the Switching valve that you want to add and press ENTER.
The Run time parameter is the time into the run when the valve switches to the
new position. The Setpoint parameter is the position that the valve will go to at
the selected time. The On value usually designates a clockwise switch and the Off
value designates a counterclockwise switch.
If you want the valve to switch back to its original position at the end of run, then
do not need enter a second run table entry.
40 Operating Manual
Chapter 3
Operating Valves Automatically Operation
If you want the valve to switch back before the end of run, press CLEAR to return
to the SELECT PARAM to add menu and select the same valve again. Press
ENTER. Type the desired switching time and press ENTER. Select the position the
valve is to return to by selecting ON or OFF and press ENTER. With no run table
entries selected, press CLEAR multiple times to exit to the run event table.
OPERATING PROCEDURE
Heating the Valve Oven
1. On the TRACE GC Ultra keypad, press AUX.
AUXILIARY ZONE
Valve oven 43 175
Operating Manual 41
Chapter 3
Operation Operating Valves Automatically
42 Operating Manual
Troubleshooting
4
This chapter has information to help you to find and correct potential problems
when using the TRACE GC Ultra Valve Oven.
Chapter at a Glance…
Leak Testing ......................................................................................................... 43
Valve Switching....................................................................................................44
Valve Plumbing ....................................................................................................45
The most common problem associated with valves and plumbing are leaks, valves
not switching properly, and valves not plumbed correctly.
Leak testing
Leak testing requires studying plumbing diagrams and system configurations.
1. Identify all gas sources leading into the flow path to be leak tested. Determine
if the gas sources are either a common source or if they are at the same bottle
pressure.
2. Be sure all components in the sample path will withstand the pressure being
applied to the system.
3. Be sure all possible gas outlets are blocked. This will require knowing how to
identify the various flow paths throughout the valving system to be sure vents
are blocked or bypassed.
NOTE Leak should be detected using an electronic leak detector (Thermo Scientific GLD Pro, or
equivalent).
Operating Manual 43
Chapter 4
Troubleshooting
Refer to the TRACE operations manual, Chapter 4, Gases and Controls, for leak
test suggestions and the necessary hardware required to pressure check a
plumbing system.
Valve Switching
Valco valves contain a rotor that is turned by the actuator. This rotor channels the
gas between the various ports of the valve. If the rotor is not being switched
properly, a flow path may be partially or fully blocked, or leaks may occur
between valve ports.
Each valve contains a mechanical stop that should reach an extreme when the
valve is rotated in the clockwise or counterclockwise direction. Close inspection
of the valve will indicate if the valve is being switched to its limits, Figure 2-4.
Actuator pressure should be 420 Kpa (60psig) to 630 Kpa (90 psig).
44 Operating Manual
Chapter 4
Troubleshooting
Table 4-2. Causes and Remedies for Improper Valve Switching (Continued)
Cause Remedy
Actuator leaking Replace actuator
Valve event not set correctly Check valve event configuration and
event in the run table
Valve Plumbing
Verify the valve and system plumbing conforms to the appropriate plumbing
diagram for your system setup (See Chapter 3). Trace the plumbing from port to
port to insure all connections are made properly
Operating Manual 45
Chapter 4
Troubleshooting
46 Operating Manual
Customer
A
Communication
Thermo Fisher Scientific provides comprehensive technical assistance worldwide
and is dedicated to the quality of our customer relationships and services.
This appendix also contains a one-page Reader Survey. Use this survey to give us
feedback on this manual and help us improve the quality of our documentation.ù
How To Contact Us
Use http://www.thermo.com/com/cda/resources/resource_detail/1,,12512,00.html
address for products information.
Use http://www.gc-gcms-customersupport.com/WebPage/Share/Default.aspx
address to contact your local Thermo Fisher Scientific office or affiliate GC-GC/
MS Customer Support.
Operating Manual 47
Customer Communication How To Contact Us
Reader Survey
Product: Valve Oven for TRACE GC Ultra
Manual: Valve Oven Operating Manual
Part No.: 317 093 63
Please help us improve the quality of our documentation by completing and returning this survey.
Circle one number for each of the statements below.
Strongly Strongly
Agree Agree Neutral Disagree Disagree
If you would like to make additional comments, please do. (Attach additional sheets if necessary.)
________________________________________________________________________________________
________________________________________________________________________________________
________________________________________________________________________________________
Fax or mail this form to:
Thermo Fisher Scientific S.p.A.
Strada Rivoltana km 4
20090 Rodano (MI)
ITALY
Fax: 39 02 95059388
48 Operating Manual
Glossary
This section is an alphabetical descriptive list of terms common to this industry. This also includes
abbreviations, acronyms, metric prefixes, and symbols.
A
A ampere
ac alternating current
ADC analog-to-digital converter
B
b bit
B byte (8 b)
baud rate data transmission speed in events per second
C
ºC Celsius
CIP Carriage and Insurance Paid To
cm centimeter
CPU central processing unit (of a computer)
CSE Customer Service Engineer
D
d depth
DAC digital-to-analog converter
dc direct current
DS data system
Operating Manual 49
Glossary
E
ECD Electron Capture Detector
EMC electromagnetic compatibility
ESD electrostatic discharge
F
ºF Fahrenheit
FID Flame Ionization Detector
FOB Free on Board
FPD Flame Photometric Detector
ft foot
G
g gram
gain A measure of the ability of an electronic circuit or device to
increase the magnitude of an electronic input parameter.
GC gas chromatograph
GND electrical ground
H
h height
h hour
harmonic A high-frequency disturbance that appears as distortion of the
distortion fundamental sine wave.
HOT OC High Oven Temperature Cold On-Column Injector
HV high voltage
50 Operating Manual
Glossary
I
ID inside diameter
IEC International Electrotechnical Commission
impulse See transient
in inch
I/O input/output
K
k kilo (103 or 1024)
K Kelvin
kg kilogram
kPa kilopascal
L
l length
l liter
LAN Local Area Network
lb pound
LED light-emitting diode
LVOCI Large Volume On-Column Injector
LVSL Large Volume Injector
M
m meter (or milli [10-3])
M mega (106)
Operating Manual 51
Glossary
µ micro (10-6)
MBq megabecquerel
mCi millicurie
meniscus The curved upper surface of a column of liquid.
min minute
mL milliliter
mm millimeter
m/z mass-to-charge ratio
N
n nano (10-9)
negative polarity The inverse of a detector signal polarity.
nm nanometer
NPD Nitrogen Phosphorous Detector
O
OCI On-Column Injector
OD outside diameter
ohm
P
p pico (10-12)
Pa pascal
PCB printed circuit board
PDD Pulsed Discharge Detector
PID Photoionization Detector
52 Operating Manual
Glossary
R
RAM random access memory
RF radio frequency
ROM read-only memory
RS-232 industry standard for serial communications
S
s second
S/SL Split/Splitless Injector
sag See surge
slow average A gradual, long-term change in average RMS voltage level,
with typical durations greater than 2 s.
SOP Standard Operating Procedures
source current The current needed to ignite a source, such as a detector
lamp.
surge A sudden change in average RMS voltage level, with typical
duration between 50 µs and 2 s.
T
TCD Thermal Conductivity Detector
Operating Manual 53
transient A brief voltage surge of up to several thousand volts, with a
duration of less than 50 µs.
U
UFM Ultra Fast Module
V
V volt
V ac volts, alternating current
V dc volts, direct current
VGA Video Graphics Array
W
w Width
W Watt
The symbol for a compound unit that is a quotient (for example, degrees Celsius per minute or grams per liter) is
written with a negative exponent with the denominator.
For example:
C min-1 instead of C/min
g L-1 instead of g/L
Index
C Troubleshooting 43
Column 17 Leak testing 43
Components 16 Valve plumbing 45
Configuring Valves 32 Valve switching 44
G U
Gas Sampling Valves 32 Unheated Enclosure 17
Installing valves 28
H
Heated Enclosure 17 V
Installing valves in 21 Valve positions 22
Valves 17
I Voltage Selection 15
Installation 14
Valves in the Heated Oven 21
Valves in the Unheated Compartment 28
Valves in the Valve Box 31
Installing 28
Installing an unheated compartment valve 30
Installing the actuator to the valve standoff 27
M
Mechanical Stop Positions 24
Mechanical Stop Positions for Switching Valves 33
O
Operation 35
Adding a Gas Sampling Valve Event to the Run
Table 38
Adding a Gas Switching Valve Event to the Run
Table 40
Heating the Valve Oven 41
Operating valves automatically 38
Operating valves manually 36
Overview 11
P
Power Requirements 15
S
Solenoid Manifold 22
Switching Valves 32
T
Timed Event Output Jacks 22, 23
Operating Manual 55
Index
56 Operating Manual
Figures
Figure 1-1. The Valve Oven............................................................................................................12
Figure 1-2. The Valve Oven and the TRACE GC Ultra .................................................................14
Figure 1-3. TRACE GC Ultra Valve Oven Components................................................................16
Figure 2-1. Timed Event Output Jacks, Solenoid Manifold, and Valve Positions ......................... 22
Figure 2-2. Timed Event (TE) Output Jacks ...................................................................................23
Figure 2-3. Mechanical Stop Positions for both CW and CCW .....................................................24
Figure 2-4. Solenoid Manifold Edges .............................................................................................25
Figure 2-5. Installing the actuator to the valve standoff .................................................................27
Figure 2-6. Installing an Unheated Compartment Valve ................................................................30
Figure 2-7. Mechanical Stop Positions for Gas Valves .................................................................. 32
Figure 2-8. Mechanical Stop Positions for Switching Valves ........................................................33
Figure 3-1. The TRACE GC Ultra Keypad, Showing the Valves Key...........................................36
Figure 3-2. The TRACE GC Ultra Keypad, Showing the Run Table Key ..................................... 38
Tables
Table 4-1. Leak Testing for Valves ...............................................................................................44
Table 4-2. Causes and Remedies for Improper Valve Switching .................................................. 44