Installation & Maintenance Manual
Installation & Maintenance Manual
Installation & Maintenance Manual
MAINTENANCE MANUAL
MANUFACTURER
Via Roma, 45
21057 OLGIATE OLONA (VA) – ITALY
Tel. +39 0331 376760
Fax +39 0331 376763
www.degotzen.com - www.acteongroup.com
E-mail: [email protected]
[email protected]
www.acteongroup.com
ITALY
WORLDWIDE
Page 2 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 3 of 61
ENGLISH
de Götzen® S.r.l. – ACTEON Group RESERVES ITSELF THE RIGHT TO MODIFY THE PRODUCTION
AND THE MANUAL WITHOUT NOTICE.
THE MANUAL MUST ALWAYS BE KEPT NEAR THE MEDICAL DEVICE FOR FUTURE REFERENCE.
de Götzen® S.r.l. – ACTEON Group SHALL NOT BE LIABLE FOR AN INCORRECT USE OF THE
INFORMATION CONTAINED IN THIS MANUAL.
Page 4 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
CONTENTS
CHAPTER 1.......................................................................................................................................... 7
1.1. PRELIMINARY INFORMATION............................................................................................................ 7
1.2. INFORMATION FOR THE INSTALLER................................................................................................. 7
1.3. WARRANTY CONDITIONS................................................................................................................... 8
1.4. TRANSPORT CONDITIONS.................................................................................................................. 8
1.5. SAFETY WARNINGS.............................................................................................................................. 9
CHAPTER 2........................................................................................................................................ 13
2.1. RADIOGRAPHIC SYSTEM...................................................................................................................13
2.2. SYSTEM COMPONENTS.....................................................................................................................14
2.3. OVERALL DIMENSIONS.....................................................................................................................15
2.4. IDENTIFICATION TAGS......................................................................................................................18
CHAPTER 3........................................................................................................................................ 21
3.1. INSTALLATION SPECIFICATIONS......................................................................................................21
CHAPTER 4........................................................................................................................................ 23
4.1. INSTALLATION....................................................................................................................................23
4.2. UNPACKING........................................................................................................................................23
4.3. ASSEMBLING THE WALL PLATE........................................................................................................24
4.4. ASSEMBLING THE BRACKET.............................................................................................................25
4.5. ASSEMBLING THE PANTOGRAPH ARM...........................................................................................26
4.6. ASSEMBLING THE TIMER..................................................................................................................27
4.7. ASSEMBLING THE TUBEHEAD..........................................................................................................28
4.8. BALANCING THE PANTOGRAPH ARM.............................................................................................30
CHAPTER 5........................................................................................................................................ 33
5.1. ELECTRIC CONNECTION...................................................................................................................33
CHAPTER 6........................................................................................................................................ 36
6.1. CONFIGURATION...............................................................................................................................36
6.2. CHANGING THE CONFIGURATION..................................................................................................37
CHAPTER 7........................................................................................................................................ 40
7.1. START UP.............................................................................................................................................40
CHAPTER 8........................................................................................................................................ 41
8.1. CHECKING THE INSTALLATION........................................................................................................41
8.2. CHECKING THE EXPOSURE FACTORS..............................................................................................42
CHAPTER 9........................................................................................................................................ 44
9.1. DIAGNOSTIC.......................................................................................................................................44
CHAPTER 10...................................................................................................................................... 46
10.1. CALIBRATION OF THE TUBEHEAD.................................................................................................46
CHAPTER 11...................................................................................................................................... 47
11.1. ERROR MESSAGES...........................................................................................................................47
CHAPTER 12...................................................................................................................................... 49
12.1. SUGGESTED MAINTENANCE..........................................................................................................49
12.2. CLEANING THE OUTER SURFACE...................................................................................................49
CHAPTER 13...................................................................................................................................... 51
13.1. REPLACEMENT OF FUSES................................................................................................................51
CHAPTER 14...................................................................................................................................... 53
14.1. REPAIR...............................................................................................................................................53
13.2. DISPOSAL..........................................................................................................................................53
CHAPTER 15...................................................................................................................................... 54
15.1. SPARE PARTS....................................................................................................................................54
ANNEXES.......................................................................................................................................... 65
A1. TECHNICAL SPECIFICATIONS............................................................................................................65
ANNEX 2........................................................................................................................................... 68
A2. INTENDED ENVIRONMENT................................................................................................................68
ANNEX 3........................................................................................................................................... 69
A3. LIST OF INTERNATIONAL STANDARDS AND DIRECTIVES..............................................................69
ANNEX 4........................................................................................................................................... 70
A4. DOSIMETRIC INDICATIONS...............................................................................................................70
ANNEX 5........................................................................................................................................... 73
A5. ELECTROMAGNETIC COMPATIBILITY...............................................................................................73
ANNEX 6........................................................................................................................................... 77
A6. DRAWINGS AND DIMENSIONS.........................................................................................................77
ANNEX 7........................................................................................................................................... 79
A7. INSTALLATION ELECTRICAL SCHEME...............................................................................................79
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 5 of 61
ENGLISH
CONTROL PANEL
MAIN SWITCH
CONTROL BUTTON
RADIOGRAPHIC DISTANCE
INDICATOR
MEMORY
RADIOGRAPHIC CURRENT
MAXILLARY LOWER
INDICATOR
TEETH
BITE-WING EXAM
X-RAY OUTPUT SIGNAL
MALFUNCTIONING
INDICATOR KEY SWITCH CONVENTIONAL
RADIOGRAPHIC TECHNIQUE
Page 6 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
1
CHAPTER 1
1.1. PRELIMINARY INFORMATION
Before beginning to use the “Xmind®dc” radiographic system, it is mandatory to carefully read and follow the
instructions contained herein, so as to obtain the best possible performance and to assure the safety of the
patient, operator, device and environment.
CAUTION
The word CAUTION identifies those occurrences which might compromise the operator’s personal safety or cause injuries to people.
! WARNING
The word WARNING identifies those occurrences which might compromise the radiographic system’s performance.
PLEASE NOTE
PLEASE NOTE serve to give special indications so as to facilitate maintenance or make important information clearer.
CAUTION
The installer is responsible for the installation, with regards to the system safety and operation.
For safe and reliable installation of the “Xmind®dc” radiographic system it is advisable to:
- check that the voltage mentioned in the rating plate matches the line voltage;
- install the radiographic system according to the procedures described in this manual;
- provide the user with any information regarding the use of the radiographic system according to what stated
in the manual;
- certify the work done by a “Declaration of Conformity;
- return to “de Götzen® S.r.l.” the warranty certificate duly filled in: if this is not done, the warranty is not valid.
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 7 of 61
ENGLISH
Inappropriate use or any arbitrary tampering with the equipment exempt de Götzen® S.r.l. – ACTEON Group,
as manufacturer of the “Xmind®dc” radiographic system, from any service under warranty or from any other
liability.
- any repairs, modifications, adjustments, re-calibrations must be performed only by de Götzen® S.r.l. – ACTEON
Group;
- the installation must be made by professionally qualified technicians according to the regulations in force;
- the system must be installed and used in compliance with the instructions given in this Manual and for the
purposes and applications for which it was designed;
- the power supply must be adequate to supply the required power indicated in the radiographic system’s
nameplate data;
- in order to safeguard one’s warranty rights, please fill in the enclosed Warranty Document, immediately after
the installation is completed, together with the technician;
- the system must be checked completely at least each 12 months by professionally qualified technicians according
to the regulation in force. Use the manuals provided with the device “Xmind®dc” for reference.
- in case of repair, please use only spare parts from the manufacturer of the “Xmind®dc”. Otherwise basic safety
and essential performances of the device will not be guaranteed.
de Götzen® S.r.l. – ACTEON Group is not responsible for any damages caused by any person or thing as
a consequence of non-compliance of any or all guidelines contained in all the manuals provided with the
“Xmind®dc” device.
CAUTION
Non compliance of any of the above mentioned rules and all the indications provided by the manufacturer in the documentation, or successively
in written paper or electronic format, will result in losing the warranty of the product and the manufacturer will be discharged from any obliga-
tion, including consequential damages, direct or indirect that may derive to people, things or environment. Furthermore, the facility represen-
tative, customer or employees of the facility, will be liable for any damage and/or incident and/or degeneration of the health status of a patient,
operator, involved people and the surrounding environment.
All claims for damage or miscarriage regarding the shipment must be pointed out in the presence of the shipping
agent.
In case of miscarriages, or actual or suspected damage, the receiver shall indicate the proper reserves on the
way-bill or on the consignment note.
Page 8 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
A few safety recommendations which should be followed when using the “Xmind®dc” radiographic system are
listed here below.
CAUTION
GENERAL REQUIREMENTS
GENERAL REQUIREMENTS
RESPONSIBLE ORGANIZATION is the authority that has the responsibility for the USE and MAINTENANCE of the “Xmind®dc” radiographic system.
The training and preparation of personnel is the responsibility of THE RESPONSIBLE ORGANIZATION.
“Xmind®dc” radiographic system is an X-ray generator and must be used and handled only by specialised surgeons, dentists and authorized
personnel, who meet the requirements provided by the national laws in force in the country of installation.
It is mandatory for the RESPONSIBLE ORGANIZATION to provide a routine and special maintenance schedule for medical equipment; this sche-
dule must be documented for every device and transmitted to the various operating levels (*). The preventive maintenance (that must be perfor-
med at least every 12 months), which includes functional, performance and safety tests of the device, must be carried out by qualified, autho-
rized professional technicians. It is mandatory to ensure patients’ health and safety and proper “Xmind®dc” radiographic system operation (IEC
60601-1 etc.). These operations must be carried out according to the methods and frequency indicated in this manual and in the installation and
maintenance manual. Failure to comply with this requirement or with the messages concerning anomalies will release the manufacturer from
any liability for direct and indirect injuries to persons and/or damage to property or the environment. Furthermore, the managers of the facility,
customers or collaborators shall be held liable for any damage and/or accidents and/or degeneration of patients’ or operators’ health or of the
surrounding environment.
The RESPONSIBLE ORGANIZATION must also provide for the safe and proper use of the equipment.
(*) For Italy refer to Presidential Decree 14/01/1997, Legislative Decree No. 81/2008 (as subsequently amended and modified).
Operators must know the environmental and operating specifications of the device, as well as the procedures to follow in the event of hazards or
emergency stops.
“Xmind®dc” radiographic system has been designed to acquire radiography images for dental intraoral X-ray imaging. The “Xmind®dc” medical
device must not be used for X-ray imaging of other body parts.
Carefully follow the instructions in this manual to install, operate and maintain the “Xmind®dc” radiographic system. In the event that local laws
and standards are more restrictive than the manufacturer’s indications, the former supersede the latter.
The RESPONSIBLE ORGANIZATION must comply with the standards and regulations in force concerning the installation of the medical device in
consideration of the place of installation.
The operator is cautioned to monitor the patient and the parameters of the “Xmind®dc” radiographic system throughout the entire duration of
the X-ray examination.
de Götzen S.r.l. – ACTEON Group and its authorized technicians are not required to verify compliance of the installation site with local standards
concerning electrical safety and X-ray protection and with any other directive concerning safety in force in the country of installation.
The RESPONSIBLE ORGANIZATIONS of the facility must ensure compliance of the installation site with the local laws in force.
Before each examination, it is mandatory to apply a disposable protection sheath to the collimator cone (Beam Limiting Device) designed to
cover the end part of the X-ray unit, which is more susceptible of being directly contaminated during the X-ray exposure (class I Medical Device
Directive 93/42/EEC and subsequent amendments). It can come into contact with the patient’s skin: verify biocompatibility according to the prin-
ciples given in the ISO 10993 series of standards, refer for details to the disposable use protection’s instructions for use.
Before operating the “Xmind®dc” radiographic system the operator must assure that the device has no visible signs of damage.
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 9 of 61
ENGLISH
CAUTION
“The general principles regarding safety and protection of workers and people” must always be applied when using the unit:
1. Justification of the practice
2. Protection Optimisation
3. Reduction of the limits of individual dose and risks
The “Xmind®dc” is a medical device that generates X-rays; therefore, both the patients and the operator are exposed to risks due to ionizing
radiation. The physician must assess the actual need for X-ray exposure.
All personnel present during an X-ray examination must comply with safety regulations concerning protection against radiation. For their own
safety, the operator must always keep a distance of more than 2 meters (6 ft.) and out of the path of the X-ray beam, in order to avoid the exposi-
tion to the stray radiation and residual radiation.
The “Xmind®dc” medical device must be used in compliance with the local standards in force and with the international directives concerning
radiation protection.
The device must comply with the guidelines and indications provided by an accredited specialist in radiation protection, who will recommend, if
necessary, the additional shields or precautions for every specific case.
The device installation site must be shielded in compliance with the local standards in force to protect the operator, patient and other people
against X-rays.
The “Xmind®dc” device is intended to be used solely by surgeons, dentists and qualified and authorized physicians. The operator must:
- determine, when appropriate, the possible need for sedation and the related operating methods and precautions best suited appropriate for
the patient
- supervise the entire X-ray examination procedure, paying attention to the indications and information from the unit.
The device must be used only for diagnostic purposes by qualified and authorized dentists and/or physicians.
The operator and other personnel must keep clear from the patient during the scan.
The personnel involved in the radiographic examination must take all the safety measures concerning radiation protection.
It is the operator’s responsibility to protect the patient against unnecessary or excessive radiation doses.
Additional protection devices (aprons, collars, etc.) are required to protect the patient from radiation.
Before exposing patients with pacemakers, contact the manufacturer of the latter to ensure that the X-rays generated by “Xmind®dc” do not
interfere with its functionality.
.
“Xmind®dc” generates X-rays: Before using this X-ray system please refer to the regulation in force in your area concerning paediatric patients,
pregnant women and anyone with health issues that contraindicate the use of X-rays. Investigate and make sure of this condition before starting
the exposure.
Page 10 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
CAUTION
MECHANICAL RISK
Before removing the tubehead from the positioning arm, RELEASE THE SPRING.
The sudden opening of the joint may cause damage to people and/or things.
Check that the installation of the unit complies with the mechanical specifications of the support (walls, ceiling, etc..) where it is installed.
Adjustments or any kind of attempt of repairing or disassembling must only be performed by qualified and authorized service personnel.
The “Xmind®dc” must not be used in environments or close to environments subjected to mechanical vibrations or mechanical shocks.
CAUTION
ELECTRIC SAFETY
The radiographic system contains high voltage. It is prohibited to inspect internal parts of the system.
The unit must be used only in environments that are in compliance with all electrical safety standards set forth for medical environments.
To avoid the risk of electric shock, this device must only be connected to a supply mains with protective earth.
The unit is NOT equipped with protections against penetration of liquids; it will therefore be necessary to make sure that no water or other
liquids penetrate inside in order to avoid short circuits or corrosion.
Always disconnect the radiographic system from the power supply and wait for 2 minutes before beginning to clean, disinfect and maintenance
operations.
Do not connect the X-ray system to a multiple portable socket outlet (MPSO) nor to any type of extension cord.
External equipment intended for connection to signal input, signal output or other connectors shall comply with the relevant product standard
e.g. IEC 60950-1 for IT equipment and the IEC 60601-series for medical electrical equipment. In addition, all such combinations – systems – shall
comply with the safety requirements stated in the general standard CEI EN 60601-1, edition 3, clause 16. Any equipment not complying with the
leakage current requirements in CEI EN 60601-1 shall be kept outside the patient environment i.e. at least 1,5 m from the patient support.
It is mandatory to use an isolation device (Separation Device) to isolate the equipment located outside the patient environment from the equip-
ment located inside the patient environment. In particular such a Separation Device is required when a network or data connection is made. The
requirements on the Separation Device is defined in CEI EN 60601-1, edition 3, clause 16.
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 11 of 61
ENGLISH
CAUTION
EMC COMPATIBILITY
EMC requirements must be considered and the “Xmind®dc” must be installed and used accordingly to the specific EMC information provided in
the accompanying documents.
The device complies with the EMC (Electromagnetic Compatibility) requirements, according to CEI EN 60601-1-2. Radio transmitting equipment,
cellular phones etc. shall not be used in close proximity of the unit as they could influence the performance of the system.
Carefully read the indications relevant to the EMC in the dedicated appendix A5. EMC compatibility of this manual.
CAUTION
The radiographic system MUST NOT be used in the presence of disinfectant, flammable or potentially explosive gases or vapours that might
catch fire and cause damage.
In case these disinfectants have to be used, let the vapour completely disperse before turning on the radiographic system.
CAUTION
Modifications or upgrades of the system can be carried out only if advised by de Götzen® S.r.l. – ACTEON Group and performed by authorized
and qualified personnel, using ONLY genuine original spare parts of de Götzen® S.r.l. – ACTEON Group.
de Götzen® S.r.l. – ACTEON Group proscribes improper, unauthorized modifications or upgrades of the device, in order to avoid malfunctions
resulting in breakdowns and/or accident for patient, operator and equipment. de Götzen® S.r.l. – ACTEON Group assumes no responsibility and,
consequently, declines all responsibility with respect to direct or indirect damages to people, the device or environment due to these reasons.
Disregarding this warning can result in irreversibly compromising the overall safety of the system and can be dangerous for operators, patients
and environment.
Page 12 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
2
CHAPTER 2
2.1. RADIOGRAPHIC SYSTEM
The “Xmind®dc” radiographic system guarantees the maximum safety both for the operator and the patient.
The following protective measures were adopted in the design and construction of the unit:
- protection against the risk of electric injuries, ensured by a grounded protection conductor
- protection against leakage radiation, made negligible by the shielded casing
- protection against excessive radiations, thanks to the immediate activation of the safety device
- protection against continuous service, since the system is designed, according to standards, not to allow use
in radioscopy
- protection for the patient against dangerous radiations, obtained by means of the high frequency technology
capable of producing a constant and hard radiation
- protection against exposure mistakes obtained with the high frequency technology which is unaffected by
voltage fluctuation and consequently capable to guarantee extremely accurate exposure parameters
- protection for the operator against irradiation ensured by the extensible cable of the hand control which
allows for a safety distance of more than 2 meters (6 ft.)
- protection against involuntarily selection of radiographic technique (FILM or DIGIT) obtained, according to
standards, by means of the confirming on the selection key
“ELECTRO-MEDICAL” CLASSIFICATION
According to paragraph §6 of the general safety regulations CEI EN 60601-1: 2007 on safety of medical equip-
ment, the system is classified as: Class I - Type B.
“E.M.C.” CLASSIFICATION
According to paragraph §4 of the CEI EN 55011, the system is classified as: Group 1 – Class B
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 13 of 61
ENGLISH
1. “Xmind®dc” TIMER
The timer is the control panel used to manage the exposure times and to safely use the tubehead.
To make the exposure, the control button with safety key is available.
The timer can be connected to n° 2 dc tubeheads.
2. WALL PLATE
3. BRACKET
The horizontal bracket is available in 3 different lengths (110 cm, 80 cm, 40 cm) and represents the support for
the pantograph arm. Its shaft is fixed in a dedicated section of the timer (top or bottom) and allows for 180°
movement.
4. PANTOGRAPH ARM
Thanks to the new shape and new mechanisms of the positioning arm, it can be adjusted in height and depth
in order to precisely explore any spot in its reach.
It is made of light alloy with an ABS coating.
5. TUBEHEAD “Xmind®dc”
The intra-oral “Xmind®dc” is a tubehead type and its light alloy housing is divided into two compartments.
The high voltage transformer, the X-ray tube and the expansion chamber are submerged in highly dielectric
insulating oil inside a light alloy container.
The expansion chamber guarantees an adequate compensation to oil expansion for the entire temperature
range.
The X-ray tube is located in the back part of the container, allowing a source-skin distance 50% higher than
traditional structures.
In the second compartment the main electronic board and the control electronic board are placed.
6. CONE
The collimator cone or Beam Limiting Device represents the applied part of the device. Made of the transparent
polycarbonate, it allows:
- the correct distance between focal spot and skin
- dimension, direction and centering of X-ray beam
- the realization of different radiographic technique (bisecting and parallel technique).
During X-ray exposition, the collimator cone comes in contact with the skin of the patient. Before each exam, it is
Page 14 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
necessary to apply to the cone a disposable protective cover designed to cover the end part of the X-ray generator.
Such protection has two functions: avoid cross-contamination (from patient to patient) and prevent the possibi-
lity of inflammations or other types of reactions of the skin caused by contact with the material that constitutes
the cone.
OPTIONAL ACCESSORIES
PLEASE NOTE
For the installation of the optional accessories, refer to the specific installation manuals supplied.
Fig. 2A, 2B, 2C give the overall dimensions of the possible supply conditions:
BRACKET 400: L = 16” 9/64
BRACKET 800: L = 82,5cm – 32” 31/64
BRACKET 1100: L = 110cm – 43” 5/16
BRACKET 400
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 15 of 61
ENGLISH
BRACKET 800
BRACKET 1100
Page 16 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 17 of 61
ENGLISH
The identification tags on the tubehead, on the timer and on the cone indicate the model number, the serial
number, the manufacturing date and the symbols of the main technical characteristics.
2.3.1. TUBEHEAD
Model 230 V
Model 115 V
2.3.2. TIMER
Model 230 V
Model 115 V
Page 18 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
CONE
GRADUATED SCALE
PICTOGRAMS USED
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 19 of 61
ENGLISH
THIS SYMBOL GUARANTEES THAT THE RADIOGRAPHIC SYSTEM COMPLIES WITH THE
REGULATIONS CONTAINED IN THE EUROPEAN DIRECTIVE EEC 93/42 REGARDING MEDICAL
DEVICES
WEEE (Waste Electrical and Electronic Equipment) SYMBOL, IN CONFORMITY WITH 2012/19/
CE DIRECTIVE AND EN 50419 STANDARD.
Page 20 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
3
CHAPTER 3
3.1. INSTALLATION SPECIFICATIONS
! WARNING
Prior to installing the radiographic system the Office Owner must ascertain that:
the environment, the electric system and the power supply comply with the requirements needed, otherwise he must provide the required
adjustment.
ENVIRONMENT REQUIREMENTS
- The installation environment must be wide enough: check that the size and overall dimension of the environment
don’t present any obstacle while positioning the radiographic system.
- The environment must not be exposed to explosion hazards and must not be pressurized.
- While operating, the ambient temperature must range within +5°C and +40°C.
- The storage temperature must range within -15°C and +50°C.
- The relative humidity must range within 25% and 75%.
PLEASE NOTE
It is mandatory to check the nature and consistency of the wall; if required, ask the support of a brickwork expert.
The walls of uncertain consistency must be provided with a buried counter plate or with a sandwich type system.
Chart A
MANUFACTURER’S RATING PLATE
230 V ± 10% 115 V ± 10%
ELECTRIC SYSTEM
NOMINAL VOLTAGE 230 V 115 V
MINIMUM LINE VOLTAGE 207 V 103,5 V
MAXIMUM LINE VOLTAGE 253 V 126,5 V
FREQUENCY 50 / 60 Hz 50 / 60 Hz
ABSORBED POWER 1,2 kVA 1,1 kVA
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 21 of 61
ENGLISH
Chart B
MANUFACTURER’S RATING PLATE 230 V ± 10% 115 V ± 10%
POWER SUPPLY VOLTAGE 198 ≤ V ≤ 253 103,5 ≤ V ≤ 126,5
MINIMUM CONDUCTOR SECTION 1,5 mm2
MAXIMUM LINE LENGHT 10 m
MAXIMUM CONDUCTOR SECTION 2,5 mm2
MAXIMUM LINE LEGHT 20 m
PLEASE NOTE
- The communication cables (C11, C12 – C21, C22) between the timer and the tubehead must be two-pole, twisted
and shielded with 0,25mm2 (es. tipo Belden 9501).
- The cables (S11, S12 – S21, S22) connecting the timer and the Rx signalling lamp for external use must be two-
pole type, of section 0,5mm2.
- The electric line characteristics must (Chart C).
Chart C
MANUFACTURER’S RATING PLATE
230 V ± 10% 115 V ± 10%
ELECTRICAL LINE
MAXIMUM VOLTAGE DROP 3%
APPARENT LINE RESISTANCE 0,5 Ω 0,2 Ω
ELECTRIC CONNECTIONS
! WARNING
Prior to installing the radiographic system, it is advisable that all the electric connections are arranged.
TIMER
On the timer installation wall, suitable runs for the following electric cables must be provided, according to the
installation electric diagram (refer to Chapter 15):
} timer electric cables;
} cables for the connection between timer and tubehead;
} cables connecting the timer and the X-ray signalling lamp for external use Xmind® LIGHT (OPTIONAL);
} cables connecting the timer and the remote control button Xmind® ECB (OPTIONAL).
CAUTION
According to the relevant standard, the timer must be installed in a position that allows the operator to permanently control the radiographic
exposure.
TUBEHEAD
On the wall plate installation wall, a suitable run for the cable connecting the timer and tubehead must be
provided.
Page 22 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
4
CHAPTER 4
4.1. INSTALLATION
CAUTION
The “Xmind®dc” radiographic system must be installed by professionally trained technicians, who must be able to certify their work by
“Declaration of Conformity”.
! WARNING
Prior to installing the radiographic system, it is advisable to verify that all needed requirements have been met (refer to Chapter 3)
4.2. UNPACKING
The components of the “Xmind®dc” radiographic system are duly packed within a carton box, as shown in the
Fig. 5:
1. TUBEHEAD PACKAGING
2. TIMER + OPTIONAL PACKAGING
3. PANTOGRAPH TYPE ARM PACKAGING
4. WALL PLATE and BRACKET PACKAGING
D MANUALS and WARRANTY CARD
PLEASE NOTE
Prior to installation, duly check all components.
PLEASE NOTE
The carton board and the polystyrene foam can be completely recycled and can be disposed by authorized recycling companies.
PLEASE NOTE
It is advisable to store the original packaging to return the goods for repairs.
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 23 of 61
ENGLISH
! WARNING
When the timer is installed aside the wall plate, please consider what follows:
} the timer must be mounted on the left side of the wall plate;
} the distance between timer and wall plate must be 2,5 mm.
! WARNING
To fix the wall plate DO NOT use plastic or rubber anchor screws.
For cement walls, or built with solid or hollow bricks, use metal anchor screws Ø12 (NOT included in the supply).
Fig. 6
Page 24 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
1. Take out from the packaging the wall plate and the drilling template 1.
2. Position the drilling template 1 on the radiographic system installation wall, at the required height (130cm
from the base in the suggested height).
3. Fix the template 1 with adhesive tape.
4. Check the holes for verticality and alignment with the floor, using a plumb line.
5. Mark the wall plate 8 fixing holes.
6. If required, mark the holes for the electric cables connecting the timer to the tubehead.
PLEASE NOTE
To prevent any flaking in the white coat and to control the centre distances between the holes, it is advisable to start drilling with a tip Ø7,
increasing this measure gradually.
PLEASE NOTE
If the wall is not perfectly levelled, put a suitable shim between the wall and the wall plate, so as to prevent any possible deformations.
PLEASE NOTE
The 82,5cm and 110cm brackets are provided with a stop key 1 (Fig. 7A and 7B) to prevent the electric cable from twisting.
PLEASE NOTE
Generally, the stop key is installed so that the equipment position at rest is on the right side of a possible watcher standing in front of the
wall plate (Fig. 7A). Should the position at rest be on the left side, the stop key must be rotated by 180° (Fig. 7B).
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 25 of 61
ENGLISH
Fig. 8
PLEASE NOTE
Prevent all foreign matters (ground, dust, cement, etc.) from settling on the pin seat.
The pin must slide freely in its seat.
If required, thoroughly clean and lubricate with grease Molikote D.
PLEASE NOTE
Check accurately with a line level instrument, the exact concurrency between the bracket and the ground floor.
Page 26 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
CAUTION
Check that the cable runs are arranged in the timer installation wall; check the compliance of the power supply with the installation specifica-
tions (refer to Chapter 3).
! WARNING
Check that the rating data match the power supply voltage.
! WARNING
When the timer is installed aside the wall plate, please do as follows:
} the timer must be mounted on the left side of the wall plate;
} the distance between timer and wall plate must be 2,5mm;;
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 27 of 61
ENGLISH
1. Take out the timer out of the packaging and take out the drilling template 1.
2. Position the drilling template 1 on the radiographic system installation wall, at the required height.
3. Fix the template 1 with adhesive tape.
4. Check the holes for uprightness and alignment with the floor, using a plumb line.
5. Mark the timer fixing holes on the wall using the drilling template.
6. If required, mark the holes for the electric cables connecting the timer to tubehead.
7. Drill using a Ø3 tip, then drill again with a Ø6 tip to prevent any flaking of the white coat.
8. Remove the template 1 and insert the suitable anchor screws provided 2.
9. Open the timer by unscrewing the three screws 8.
10. Withdraw the 26-pole connector from its seat to release both timer guards.
11. Approach the timer 8 to the wall and insert the electric feeding cables into the slot A or the slot B.
12. Insert the connection cables coming from the tubeheads into the slot A or the slot B: when the timer is ins-
talled aside the wall plate insert in between the rubber cover for the electrical cable.
13. Insert the cables of the X-ray signaling lamp for external use (OPTIONAL) and the cables of the remote
control button (OPTIONAL) into the slot A or the slot B.
14. Approach the timer base 8 to the wall, matching the three anchor screws with the holes screw the screws
with the relevant washers.
PLEASE NOTE
The use of the slots A or B for the cables is arbitrary; use the slots which adapt better to the specific configuration of installation.
ASSEMBLY INSTRUCTION
3. Remove both guards from the pantograph type arm by loosening the relevant
screws.
Page 28 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
8. Check that during insertion, the pin of the turning preventing device correctly
fits the seat located on the pantograph head.
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 29 of 61
ENGLISH
9. Couple the pantograph and tubehead connections and fit the into their seats.
CAUTION
The pantograph arm must be adjusted when the tubehead is assembled only (Fig. 12).
! WARNING
To prevent damages to the internal mechanism while performing adjustment and the balancing tests, the adjustment key must not be left in
place.
! WARNING
PLEASE NOTE
To reach the adjustment screw X the arm A must be put in vertical position.
To reach the adjustment screw Y the arm B must be put in horizontal position.
The adjustment key provided can be inserted only under the above conditions.
BRACCIO B – ARM B
BRACCIO A – ARM A
Page 30 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
PLEASE NOTE
The pantograph type arm is supplied with arm A already preloaded.
The arm B is supplied unloaded for safety reasons.
} arm A vertical
} arm B horizontal
} insert the adjustment key in Y
} load the spring by n° 22 turns
} withdraw the key
PLEASE NOTE
Repeat the test and adjustment until the arm B is steady and stable in all positions, even when the arm A is completely extended.
4. READJUSTMENT OF ARM A
PLEASE NOTE
Repeat the test and adjustment until the arm A is steady and stable in all positions, even when the arm B is completely extended.
5. Insert the movable finish F between the pantograph type arm guard and metal
frame.
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 31 of 61
ENGLISH
6. Insert the pins of the guards into the relevant seats; then position them and
check that the movable finish is coupled to the guards.
Page 32 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
5
CHAPTER 5
5.1. ELECTRIC CONNECTION
CAUTION
CAUTION
For electric safety, it is essential that the ground conductors are correctly connected.
! WARNING
While performing the connection, always respect the polarity PHASE/ NEUTRAL.
! WARNING
While stripping the cables, pay attention to the small copper wires that may fall on the printed circuit and cause short circuits or malfunc-
tioning.
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 33 of 61
ENGLISH
1 BROWN L = PHASE
2 YELLOW GREEN T = GROUND
3 BLUE N - NEUTRAL
4 BLACK COMMUNICATION
5 RED COMMUNICATION
6 YELLOW GREEN T = GROUND
7 YELLOW GREEN T = GROUND
THE COMMUNICATION CABLES ARE NOT POLARIZAED
Page 34 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 35 of 61
ENGLISH
6
CHAPTER 6
6.1. CONFIGURATION
On the control panel the LED relevant to the following exposure parameters will light up
supplied cone
LED 8” = SHORT CONE
LED 12” = LONG CONE
type tubehead
LED DC = DIRECT CURRENT
radiographic voltage
LED 70kV
radiographic current
LED 8mA
type of patient
LED ADULT
radiographic technique
CONVENTIONAL LED D
The above configuration depends on the position of dip-switch on the timer electronic card:
LEGEND
ON = INSTALLED
OFF = NOT INSTALLED
Page 36 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
ON OFF
1 ■ TUBEHEAD 1
2 ■ TUBEHEAD DC
3 ■ TUBEHEAD 2
4 ■ TUBEHEAD DC
5 ■ II° CONTROL BUTTON
6 ■ LONG CONE 12’’ (31cm)
7 ■ NOT AVAILABLE
8 ■ NOT AVAILABLE
ON OFF
1 ■ TUBEHEAD 1
2 ■ TUBEHEAD DC
3 ■ TUBEHEAD 2
4 ■ TUBEHEAD DC
5 ■ II° CONTROL BUTTON
6 ■ SHORT CONE 8’’ (20cm)
7 ■ NOT AVAILABLE
8 ■ NOT AVAILABLE
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 37 of 61
ENGLISH
INSTRUCTION
To change the number of tubehead installed, move the dip-switch n° 1 or the dip-switch n° 3:
DIP SWITCH 1 3
ON ON
OFF OFF
►if the tubehead is connected to the XRAY1 terminal board, put the dip-switch n° 1 in the ON position; otherwise
in OFF position
►if the tubehead is connected to the XRAY2 terminal board, put the dip-switch n° 3 in the ON position; otherwise
in OFF position
To change the type of tubehead installed, move the dip-switch n° 2 or the dip-switch n° 4:
DIP SWITCH 2 4
ON ON
OFF OFF
►if the tubehead connected to the XRAY1 is of ac, put the dip-switch n° 2 in the OFF position; otherwise put it in
the ON position
►if the tubehead connected to the XRAY2 is of ac, put the dip-switch n° 4 in the OFF position; otherwise put it in
the ON position
PLEASE NOTE
DIP SWITCH 5
ON
OFF
Page 38 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
PLEASE NOTE
DIP SWITCH 6
ON
OFF
►with short cone (8”=20cm (SSD)) move the dip-switch 6 to the OFF position
►with long cone (12”=31cm (SSD)) move the dip-switch 6 to the ON position
PLEASE NOTE
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 39 of 61
ENGLISH
7
CHAPTER 7
7.1. START UP
CAUTION
When all connections are completed, the installer must check the electric safety and functions of the system.
INSTRUCTION
Bring the main switch located on the upper part of the timer to the “I” position (ON)
1. the green light turns on, indicating that the system is powered
2. the LEDs of the set parameters automatically light up
3. the exposure time is shown on the display
CAUTION
If an error is detected when the system is turned on, the anomaly is indicated as follows:
- an intermittent sounds off (beep);
- MALFUNCTIONING INDICATOR LED
intermittently turns on
- the error code (E ….) appears on the display (refer to Chapter 11)
- all control panel functions are inhibited;
In this case turn off the timer and then turn it back on.
If the error should repeat itself, call the “Assistance Service”.
PLEASE NOTE
The exposure time and parameters which appear on the display are the last that were set before the timer was turned off.
If the timer remains inactive for a few minutes, it switches to the stand-by mode.
Press any key on the control panel to restore it to the operative mode.
Page 40 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
8
CHAPTER 8
8.1. CHECKING THE INSTALLATION
Check on the control panel that all LEDs corresponding to the required configuration are lit; otherwise, change
them.
1. Check the correct operation of the control panel by selecting different exposure times.
2. Check the time on the display.
3. Check that, when changing the selected tubehead, the corresponding X-ray signalling lamp for external use
turns on (OPTIONAL).
3. On the control button press the X-RAY key and keep it pressed until the acoustic signal (beep) stops and
the yellow LED turns off.
PLEASE NOTE
If the “X-RAY” key is released early, the exposure is immediately interrupted and the E12 error message appears on the display.
4. At the end of the exposure the green LED PAUSE intermittently turns on.
5. The display indicates the actual exposure time.
6. All the timer functions are inhibited.
To check the power absorbed by the radiographic system a tester must be used, in the mode ammeter in AC:
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 41 of 61
ENGLISH
PLEASE NOTE
To check the electric system, a tester must be used, in the mode voltmeter in AC:
PLEASE NOTE
In order to guarantee the safety of the radiographic system, it is necessary to set up a periodic control schedule
of the exposure factors.
The responsible organization is required to organise and observe a control schedule of the exposure factors as
detailed in this chapter.
All the following controls must be performed every 12 months.
CAUTION
During these procedures X-rays will be emitted. Please take every necessary precautions in order to avoid accidental exposure to ionizing
radiations.
Page 42 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
The dose in air is measured with a “non invasive” instrument, by positioning the detector at a source-skin distance
= 31cm (12”) (SSD) or 20cm (8”) (SSD)
SET TECHNICAL FACTORS
NOMINAL VOLTAGE Vn ± 10%
MAXIMUM VOLTAGE DROP 3%
NOMINAL HIGH VOLTAGE 60 / 70 kV
NOMINAL CURRENT 4 / 8 mA
SET EXPOSURE TIME 1s
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 43 of 61
ENGLISH
9
CHAPTER 9
9.1. DIAGNOSTIC
With “Xmind®dc” radiographic system it is possible to set and visualise certain functional parameters.
4. when the message INST is displayed, the installer can set the minimum exposure time in this way:
5. to exit this mode, turn the timer off and then on again.
! WARNING
Both in digital and in conventional mode, the minimum exposure time must not be less than 0.02s.
Page 44 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
RADIOGRAPHIC SYSTEM
NOMINAL VOLTAGE
LINE VOLTAGE
SOFTWARE VERSION
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 45 of 61
ENGLISH
10
CHAPTER 10
10.1. CALIBRATION OF THE TUBEHEAD
CAUTION
INSTRUCTION
5. take the exposure: on the control button press the X-RAY key and keep it pressed until the acoustic signal
(beep) stops and the yellow LED turns off.
6. once the exposure has been taken, if the display does not show errors, the calibration has been successfully
done.
Page 46 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
11
CHAPTER 11
11.1. ERROR MESSAGES
The following chart gives a list of error messages that may appear while “Xmind®dc” radiographic system is
working.
The chart also includes the causes of the error messages and what to do to solve them.
ERROR
CAUSE SOLUTION
MESSAGES
RX1 TUBEHEAD
E00 CALL THE “ASSISTANCE SERVICE”
IS NOT CONNECTED OR IS OUT OF ORDER
RX2 TUBEHEAD
E01 CALL THE “ASSISTANCE SERVICE”
IS NOT CONNECTED OR IS OUT OF ORDER
E02 CORRUPTED EEPROM DATA CALL THE “ASSISTANCE SERVICE”
E03 EEPROM DATA NOT SAVED PROPERLY CALL THE “ASSISTANCE SERVICE”
LINE VOLTAGE VALUE NOT INCLUDED
E07 CALL THE “ASSISTANCE SERVICE”
WITHIN THE NOMINAL VALUE
THE X-ray KEY
E08 MAKE SURE IT IS NOT JAMMED
ALWAYS SEEMS TO BE PRESSED
E09 ANOMALY IN THE CONTROL PANEL CALL THE “ASSISTANCE SERVICE”
THE EXPOSURE KEEP THE X-ray KEY
E12
HAS BEEN PREMATURELY INTERRUPTED PRESSED TILL THE END OF THE EXPOSURE
E20 ANOMALY IN THE TRIAC/RELAY CALL THE “ASSISTANCE SERVICE”
E21 ANOMALY IN THE ELECTRONIC CIRCUIT CALL THE “ASSISTANCE SERVICE”
E22 ANOMALY IN THE CONTROL CIRCUIT CALL THE “ASSISTANCE SERVICE”
E23 INCORRECT DIP-SWITCH CONFIGURATION CALL THE “ASSISTANCE SERVICE”
THE CONTROL BUTTON DOES NOT CORRESPOND
E24 CALL THE “ASSISTANCE SERVICE”
TO THE SELECTED TUBEHEAD
THE TUBEHEAD
E30 CALL THE “ASSISTANCE SERVICE”
DOES NOT WORK PROPERLY
THE TUBEHEAD
E32 CALL THE “ASSISTANCE SERVICE”
IS NOT IN THE CORRECT MODE
THE TUBEHEAD REPEAT THE EXPOSURE
E33
HAS NOT COMPLETED THE EXPOSURE CALL THE “ASSISTANCE SERVICE”
PROBLEM IN THE FREQUENCY
E40 CALL THE “ASSISTANCE SERVICE”
OR REGULATION
THE TUBEHEAD CARRY OUT WITH THE CALIBRATION
E41
IS NOT CALIBRATED CALL THE “ASSISTANCE SERVICE”
E42 EEPROM DATA NOT SAVED PROPERLY CALL THE “ASSISTANCE SERVICE”
E43 CORRUPTED EEPROM DATA CALL THE “ASSISTANCE SERVICE”
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 47 of 61
ENGLISH
Page 48 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
12
CHAPTER 12
12.1. SUGGESTED MAINTENANCE
In order to guarantee the safety of the radiographic system, it is necessary to set up a maintenance schedule.
The owner is responsible for organising and observing a maintenance schedule which must be executed by
qualified technicians who must be able to certify their work with a “Conformity Declaration”.
CAUTION
Run an inspection on the radiographic system and on its operation when it is installed and every twelve months.
Once a year, lubricate the pins and bushes of the wall plate and the positioning arm, as specified.
! WARNING
Do not lose the adjustment key that comes with the system, since, in time, it could become necessary to make readjustments.
! WARNING
If the parts should become hard to move or should squeak, call the “Assistance Service”.
Clean the external surface using a damp cloth and non-corrosive non oil-based detergent and disinfect it using a
non-aggressive medical detergent.
Do not spray detergent or disinfectant directly on the device.
The spacer cone may be cleaned with cotton wool soaked with surgical alcohol.
INSTRUCTION
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 49 of 61
ENGLISH
Page 50 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
13
CHAPTER 13
13.1. REPLACEMENT OF FUSES
The timer electronic equipment is protected by n° 4 fuses located on the electronic board.
INSTRUCTION
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 51 of 61
ENGLISH
Page 52 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
14
CHAPTER 14
14.1. REPAIR
In case of a malfunction, send the defective part using the original packaging to:
CAUTION
13.2. DISPOSAL
The use of the WEEE symbol indicates that, at the end of its lifespan, this product may not be treated as
household waste, but must be treated separately, in conformity to the Directive 2012/19/EC.
EU Council Directive 2012/19/EC (WEEE) imposes the disposal or recycling of electric and electronic equip-
ment. The product is marked with the indicated icon. This product must not be disposed of as domestic waste.
The crossed-out wheelie bin identifies a product placed on the market after the 13th of August 2005 (see EN
50419:2006). This product is subjected to Council Directive 2012/19/EU (WEEE) and implementation standards
in force in your country.
CAUTION
To avoid any risk of environmental contamination, don’t dispose the device and its accessories together with the domestic waste.
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 53 of 61
ENGLISH
15
CHAPTER 15
15.1. SPARE PARTS
The manufacturer undertakes to supply, upon request, drawings, circuit diagrams, component parts lists,
instructions or other information needed by qualified technical personnel to perform repairs on those parts of
the “Xmind®dc” radiographic system which may be repaired.
REF SPARE PARTS TUBEHEAD Xmind® CODE OLD COLOR Q.TY PCS
(RAL 9010 color)
1 DC XG (1) FORK COVER 29700979.C 29700979 8
2 AC/DC XG (2) FORK RING SET 29700980 5
4 AC/DC XG (4) ROTATION STOPPER r.B 29701298 1
5 IX AC/DC XG CONE DISK COIL 21100162 1
A Ø60 DEG ROUND LONG CONE 31cm (12”) 29700397 1
A Ø60 SAT ROUND LONG CONE 31cm (12”) 29701197 1
B Ø60 DEG ROUND SHORT CONE 20cm (8”) 29700396 1
B Ø60 SAT ROUND SHORT CONE 20cm (8”) 29701195 1
C 44x35 DEG SQUARE CONE 31cm (12”) 29700615.C 29700615 1
C 44x35 SAT SQUARE CONE 31cm (12”) 29701199.C 29701199 1
Page 54 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 55 of 61
ENGLISH
Page 56 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 57 of 61
ENGLISH
CODE
REF SPARE PARTS TIMER Xmind® OLD COLOR Q.TY PCS
(RAL 9010 color)
1 AC/DC TIMER (1) DRILL TEMPLATE 24200883 1
2 AC7DC XG TIMER (2) CONTROL BOARD 39200306 7
2 AC/DC XG TIMER (2) CONTROL BOARD US 39200307 7
3 AC/DC XG TIMER (3) SPIRAL CABLE 32000325.C 32000325 1
KIT SPIRAL CABLE + LOCAL BUTTON 29701028.C COMPLETE
AC/DC XG TIMER (4) DOUBLE LENGHT
4 32000359 1
SPIRAL CABLE
KIT DOUBLE LENGHT SPIRAL CABLE +
29701029.C COMPLETE
EXPOSURE BUTTON
AC/DC XG TIMER (5) EXPOSURE BUTTON
5 29700974.C 4
CASE
AC/DC XG TIMER (6) ESPOURE BUTTON
6 39700303 1
BOARD
7 AC/DC XG TIMER (7) LED BOARD 39200313 7
8 AC/DC XG TIMER (8) KEYBOARD DEG 39200510 39200308 1
8 AC/DC XG TIMER (8) KEYBOARD SAT 39200309 1
9 AC/DC XG TIMER (9) BOTTOM TIMER CASE 29700975.C 29700975 1
AC/DC XG TIMER (10) TOP TIMER CASE
10 29700304.C 29700304 4
DEG
10 AC/DC XG TIMER (10) TOP TIMER CASE SAT 29700976.C 29700976 4
11 AC/DC XG TIMER (11) SECURITY KEY SET 39700305 2
12 DC XG TIMER (12) EMI FILTER 36200382 3
13 BIPOLAR GREEN SWITCH 35000044 1
14 BIPOLAR SWITCH COVER 33500228 1
Page 58 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 59 of 61
ENGLISH
Page 60 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
CODE
REF SPARE PARTS WALL PLATE 350 OLD COLOR Q.TY PCS
(RAL 9010 color)
1 PLATE (1) COVER DEG 29700985.C 29700985 5
1 PLATE (1) COVER SAT 29700986.C 29700986 5
2 AC/DC XG B & HALF-MOON LOCKER 22000176 1
3 PLATE (3) DRILL TEMPLATE 24200984 1
PIASTRA (4) PLASTIC JOINT + PLASTIC
4 29701036.C 29701036 3
CAPS
WALL PLATE 350 DEG 29700383.C 29700383 COMPLETE
WALL PLATE 350 SAT 29700384.C 29700384 COMPLETE
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 61 of 61
ENGLISH
CODE
REF SPARE PARTS WALL PLATE 350 OLD COLOR Q.TY PCS
(RAL 9010 color)
1 PLATE (1) COVER DEG 29700985.C 29700985 5
1 PLATE (1) COVER SAT 29700986.C 29700986 5
2 AC/DC XG B & HALF-MOON LOCKER 22000176 1
3 PLATE (3) TOP MOUNT DRILL TEMPLATE 24201002 1
PIASTRA (4) PLASTIC JOINT + PLASTIC
4 29701036.C 29701036 3
CAPS
5 PIASTRA (5) TOP MOUNT COLLAR 22000597
WALL PLATE 350 TOP DEG 29700988.C 29700988 COMPLETE
WALL PLATE 350 TOP SAT 29700385.C 29700385 COMPLETE
Page 62 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
CODE
REF SPARE PARTS BRACKET OLD COLOR Q.TY PCS
(RAL 9010 color)
BRACKET 400 29700380.C 29700380 COMPLETE
BRACKET 800 29700378.C 29700378 COMPLETE
BRACKET 1100 29700379.C 29700379 COMPLETE
1 BRACKET CAP COVER + SCREW RAL 9010 29701281.C 2
BRACKET SLIDE COVER AC/DC 14X1.4cm
2 27502305 3
RAL 9010
3 BRACKET ROTATION STOPPER 29701008 5
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 63 of 61
ENGLISH
CODE
REF SPARE PARTS BRACKET OLD COLOR Q.TY PCS
(RAL 9010 color)
CIELING BRACKET 29700382.C 29700382 COMPLETE
CODE
REF SPARE PARTS BRACKET OLD COLOR Q.TY PCS
(RAL 9010 color)
DENTAL CHAIR BRACKET 29700381.C 29700381 COMPLETE
CODE
REF SPARE PARTS BRACKET OLD COLOR Q.TY PCS
(RAL 9010 color)
TOP MOUNT BRACKET 400 29700620.C 29700620 COMPLETE
TOP MOUNT BRACKET 800 29700621.C 29700621 COMPLETE
TOP MOUNT BRACKET 1100 29700622.C 29700622 COMPLETE
Page 64 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
A1
ANNEXES
A1. TECHNICAL SPECIFICATIONS
PLEASE NOTE
The measurements criteria are based on the requirements stated by the applicable standards
listed in the annex A.3 of this manual.
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 65 of 61
ENGLISH
X-ray TUBE
Page 66 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
MECHANICAL DATA
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 67 of 61
ENGLISH
A2
ANNEX 2
A2. INTENDED ENVIRONMENT
CAUTION
Page 68 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
A3
ANNEX 3
A3. LIST OF INTERNATIONAL STANDARDS AND DIRECTIVES
“Xmind®dc”: X-ray equipment for dental intraoral radiography, is in compliance with the following:
• In compliance with the classification indicated in the Medical Device Directive 93/42/EEC, Annex IX, article 10:
“Active devices intended to emit ionizing radiation and intended for diagnostic radiology”, the system is classified
as:
CLASS IIb
Certifications
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 69 of 61
ENGLISH
A4
ANNEX 4
A4. DOSIMETRIC INDICATIONS
The radiation exposure is reported in terms of Dose Area Product (DAP), which takes into account the entire
area of the X-ray beam and the total amount of X-ray radiation incident on the patient. The DAP is obtained
by multiplying the Air Kerma by the corresponding X-ray beam area, which is dependent by the typology of
beam limiting device installed. It is independent by the measured location, because increases in beam area are
compensated by the reduction of beam intensity (inverse square law).
The dosimetric values reported here are relevant to the following measured values of Total Filtration and Half
Value Layer (HVL):
In the following tables the radiation exposure is indicated in terms of DAP [mGy cm2] for each setting of kV, beam
limiting device length (SSD) and Beam Limiting Device type (circular or rectangular).
As per paragraph 203.6.4.5 of the IEC 60601-2-65, the overall deviation from the estimated air kerma is within 50%.
Page 70 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 71 of 61
ENGLISH
Page 72 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
A5
ANNEX 5
A5. ELECTROMAGNETIC COMPATIBILITY
EMISSION
IMMUNITY
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 73 of 61
ENGLISH
Page 74 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
Notes:
• At 80 MHz and 800 MHz the higher frequency range applies.
• These guidelines may not apply in every situation. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a - Field strength from fixed RF transmitters, such as base stations for radio (cellular/wireless) telephones and
land mobile radios, amateur radio, AM and FM radio and TV broadcast cannot be predicted with accuracy
on a theoretical basis. To assess the electromagnetic environment created by fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the place where the
equipment is used exceeds the corresponding RF compliance level (see above), it is important to ensure regular
equipment operation. In the event of abnormal operation, additional measures may be required, such as
redirecting or relocating “Xmind®dc”.
b - Over the frequency range between 150 kHz and 80 MHz, the field strength must be below 10 V/m.
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 75 of 61
ENGLISH
Recommended separation distances between portable and mobile RF communication equipment and
“Xmind®dc” medical device
These devices are intended to be used in environments where radiated RF interference is controlled. The
customer or “Xmind®dc” operator can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communication equipment (transmitters) and “Xmind®dc” , as
indicated below, according to the maximum output power of the communication equipment.
Separation distance according to transmitter frequency
Rated maximum output [m]
power of the transmitter
[W] 150 kHz - 80 MHz 80 MHz - 800 MHz 800 MHz - 2.5 GHz
d = 1,2 √P d = 1,2 √P d = 2,3 √P
0.01 0.12 0.12 0.24
0.1 0.38 0.38 0.73
1 1,2 1.2 2.3
10 3,8 3.8 7.3
100 12 12 23
In the event of transmitters whose maximum nominal output power coefficient does not fall within the indicated
parameters, the recommended separation distance in metres (m) can be determined by means of the equation
corresponding to the frequency of the transmitter, where P is the maximum output power coefficient of the
transmitter in watts (W) according to the information provided by the manufacturer.
Note 1: At 80 MHz and 800 MHz apply the separation distance corresponding to the highest frequency range.
Note 2: These guidelines may not apply in every situation. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Page 76 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
A6
ANNEX 6
A6. DRAWINGS AND DIMENSIONS
WALL INSTALLATION
A
40 cm (16’’) bracket 63 cm
80 cm (31’’) bracket 104 cm
110 cm (43’’) bracket 132 cm
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 77 of 61
ENGLISH
B
40 cm (16’’) bracket 178 cm
80 cm (31’’) bracket 220 cm
110 cm (43’’) bracket 247 cm
The system can also be mounted with the timer on the top. For details, refer to the Installation and Maintenance
Manual.
MOBILE INSTALLATION
“Xmind®dc” exists also in the mobile version and it is sustained by the stand shown in the following figure:
For details, refer to the Mobile Unit Technical Note, supplied with this structure.
Page 78 of 61 INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B
ENGLISH
A7
ANNEX 7
A7. INSTALLATION ELECTRICAL SCHEME
INSTALLATION & MAINTENANCE MANUAL • XMIND DC • W1100100 • V1 • (15) • 12/2015 • NXDCEN020B Page 79 of 61
3
Tel +39 0331 376 760 • Fax +39 0331 376 763
de Götzen S.r.l • A company of ACTEON Group
Via Roma 45 • 21057 OLGIATE OLONA (VARESE) • ITALY