F2400 User Manual - L21F2400ENR1 PDF
F2400 User Manual - L21F2400ENR1 PDF
F2400 User Manual - L21F2400ENR1 PDF
Medical Device
Manufacturer :
Manufacturing Site: 74, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu,
Cheongju-si, Chungcheongbuk-do
+82-31-300-0400 +82-31-300-0499 [email protected] www.sdbiosensor.com
Thank you for purchasing the STANDARD F2400 Analyzer.
This User manual contains all information about the Analyzer. Pease read this User
manual carefully and familiarize yourself with the required preparations and test
procedure before performing a first test.
Please read the instructions carefully which are included in each test device
package before using the Analyzer.
If you have any questions about the analyzer, please contact your local distributor.
You can also visit www.sdbiosensor.com for product demonstrations.
4
Chapter 07. Cleaning & Maintenance.............................................................................. 52
Cleaning of Analyzer..................................................................................................... 52
Maintenance & Transportation................................................................................... 52
5
Chapter 01. General Information
Log On
Main Menu Structure
QC
Run Test Result Run Calibration Res
Run QC
Test Status Review Calibration Review Revi
Start Prepare
(Input CAL
Operator ID) QC with
Device External
Control
QC
solution
Insert
Insert
Device
Device
CAL-1 Input Insert
Control ID Device
Device Device
Checking Checking
Insert Device
Device Checking
Insert
Apply
Device
Sample Select
CAL-2 Device
Control
Checking
Information
Sample Device
Checking Checking
Apply Apply
Sample Sample
Insert
Mounting Device
CAL-3 Sample Sample
Checking Checking
Device
Checking
Mounting Mounting
Analyzing
Result
6
QC Utility
QC
Result Monthly
System Maintenance Statistics Operator Network
Review review
Language Delete
vice
cking
Update
ect
ntrol
mation
Brightness
/Volume
ply
mple
mple
cking
nting
7
Symbols and Abbreviation
The symbols and abbreviation presented below are indicated on the User
Manual, labels and external package.
Symbols
Symbols Description
Manufacturer
Reference Number
Date of Manufacture
To indicate the date of manufacture
Serial Number
Note
Batch code
Indicate the lot number of the system
Caution
Indicate a situation, which if not avoided could result in damage to
the device or incorrect results
8
Abbreviations
Abbreviation Description
Comm Communication
LIS Laboratory Information System
HIS Hospital Information System
GUI Graphic User Interface
S/W Software
F/W Firmware
9
Precautions
10
To reduce the risk of biological hazards
·Discard
· the used specimens in accordance with federal, state
Biohazard! and local requirements.
·Treat
· specimens as potentially biohazardous material.
·If
· you have no experience with collection and handling of
specimens, appropriate education or guidance should be
received.
·Nitrile
· or latex gloves are recommended when handling
patient specimens.
11
Chapter 2. Overview
Intended Use
STANDARD F2400 Analyzer is an in vitro diagnostic medical device that measures
quantitative or qualitative biomarkers of body fluids such as blood, urine, nasal
discharge, etc., in a laboratory or POCT environment. The analyzer is indicated
for monitoring and diagnosing from the body fluid parameter in clinical settings
by healthcare professionals. In all cases, the Analyzer should be used with
designated test devices produced by SD Biosensor, Inc.
For details regarding to specific tests, refer to the user manual included with
each test device package.
Product overview
STANDARD F2400 Analyzer scans the 2D barcode of inserted test device. Based
on the scanned result of the inserted test device, the analyzer determines the
appropriate LED (UV, RGB) settings for the test device.
An antigen-antibody reaction takes a certain period of time after samples are
dispensed.
Once a reaction is completed, fluorescence dye on the control line and test line
will reflect the light from LED and this reflected value will be measured.
Then the measured value will be calculated based on the formula that is input
to the 2D barcode.
12
Precautions before perform a test
·Read
· and follow the instructions carefully in the user manual
and Instructions for use of the test device and control. It is
Note very important to follow the instructions in order to prevent an
inaccurate result or improper treatment.
·Be
· sure to check with your distributor and SD Biosensor for the
most up-to-date software and to update before use.
·Fatal
· consequences, such as fire or serious system damage,
may occur when the system is installed at an inappropriate
location, since the Analyzer cannot be replaced or removed
from power supply.
·The
· Analyzer must be installed in such a way that easy access
to power supply and power switch is ensured.
Specimens
The STANDARD F2400 Analyzer should be only used the specific test devices for
the analyzer. Because specimens are quite different for each parameter, follow
the instructions from the each test device instructions.
Safety Information
There is a potential risk of infection. Professional medical personnel should
always wear gloves when handling patient samples with the STANDARD F2400
Analyzer and test device. It is recommended that you follow all other locally
applicable health and safety regulations.
13
Operating Conditions
To operate the STANDARD F2400 Analyzer correctly, the following guidelines
should be observed :
·The
· operating conditions of the Analyzer are: 0℃-50℃ / 32℉-122℉, 10%-
93% RH.
·The
· operating temperature suitable for analyses differs in each test. For
each test, refer to the instruction manual of the test device that you will
use for analyses.
·To
· perform a test with the Analyzer, place the Analyzer on a flat surface.
·Do
· not use the Analyzer near strong electromagnetic radiation sources.
Otherwise, strong electromagnetic fields may affect the function of the
Analyzer adversely.
·If
· the Analyzer malfunctions suddenly, unplug the power cable from the
outlet.
14
Specification
15
Appearance of STANDARD F2400 Analyzer
C
B
D
E
F
G
H
I
J
K
16
A. Color TFT LCD
Used for test screen display and interaction with graphical user interface
D. Printer cover
Used to cover and protect printer paper
F. Mini USB
Used to upgrade firmware by connecting with PC
G. LAN
Used for communication using LAN Cable
H. AC power connection
Used for AC power cable connection
I. Power switch
Used to turn the Analyzer on/off
J. USB x 4
Used to connect the Analyzer to keyboard, barcode scanner and USB
memory
17
Accessories of the STANDARD F2400 Analyzer
CAL-1
CAL-2
STANDARD F
Calibration Set
CAL-3
Accessories
AC Power cable
STANDARD F
Test Device
(Example : HbA1c,
Influenza A/B)
Barcode scanner
Optional
Printer paper
18
Chapter 3. Log On and Setting
Operation of Analyzer
Step 1. Connecting the power cable
1-1. Place the Analyzer in a proper environment where it can be
plugged into outlet and barcode scanner (optional) can be
positioned together.
1-2. Connect the power cable to the power connector on the back of
the Analyzer and to the outlet.
1-3. Once the connection is complete, the analyzer is ready for use.
19
Step 2: Setting Mode
1. Touch ‘Utility’ on the main menu screen to enter the setting mode.
① ④
② ③
⑤ ⑥ ⑦
⑧ ⑨
System Menu
1) Date/Time Setting
① Press an Up( ) or Down( ) button to change the year, month,
date, hour, and minute.
② The display type of year, month and date and be changed.
Change the date style setting.
- Year-Month-Day
- Month-Day-Year
- Day-Month-Year
③ Time unit is changeable between 12 hours (12h) and 24 hours (24h).
20
2) Print Setting
You can set the number of result sheet to 1 sheet or 2 sheets. Once press
“Auto Print”, the test result will be printed automatically after all.
3) Language Setting
Press a down arrow( ) to select the desired language from the drop-
down list.
4 ) Update
① S/W Update
Items needed for S/W update
·· STANDARD F2400 Analyzer
·· Software update file
·· USB(Not provided)
·· Personal computer(Not provided)
How to update the software
a. Prepare a USB memory containing the F2400's new software file.
ii. Format USB as FAT32 file system.
iii. Copy the software file to USB memory.
b. Connect the USB(that F2400 Software is copied) to the USB port on
the bottom left of the F2400 Analyzer.
c. If the USB is connected normally, “USB connected successfully” will
be displayed.
d. Press the S/W Update button. Then, “S/W Update?” will pop up.
Press “ OK “ to start the S/W update.
21
e. If the S/W Update is completed, “S/W Update OK” will be displayed
on the screen. Press “OK” to reboot the analyzer.
`
f. Make sure that the latest S/W version is displayed in the lower left
corner of the screen.
② F/W Update
Items needed for F/W update
·· STANDARD F2400 Analyzer
·· Firmware update file
·· Mini-5pin USB cable
·· Personal computer(Not provided)
How to update the F/W
a. Connect your PC and device with a Mini USB cable. If the driver is
not recognized on PC, install the CP2101 driver.
b. Press FW update ( )on the screen.
22
c. Press “OK”.
23
g. When the update is completed on the PC, press the OK button
5) Brightness/Volume setting
① Use , buttons or the slider to adjust screen brightness.
② Use , buttons or the slider to adjust the volume.
24
Maintenance Menu
1) On the Maintenance tab, you can check the maintenance status and log
of each item.
① ② ③ ④ ⑤
25
Statistics Menu
1) On the Statistics tab, you can check the number of tests, calibration, and
QC tests for each item.
④ ⑤
②
①
③
① The number of times that each test has been performed is displayed.
② The type and number of times of errors that occurred are displayed.
③ The number of times of calibrations and Q.C. for each test is displayed.
④ Period setting: Press button to set the period that you want to
check.
·Top
· Clear: Delete selecte.
·Bottom
· Clear: Delete selected Cal & QC Log.
Caution
26
Operator Menu
1) An Operator item can be registered, edited, and deleted.
·· Password in Operator is the same as the password used for log-on.
27
·· To modify the Operator ID, select the ID,, press “Edit”, type ID you wish
to modify, and then press “OK”
28
·· To delete the Operator ID, select the ID, select the ID, press “Delete”,
then press “ OK”
Network Menu
1) On the Network tab, you can set the HIS/LIS and Analyzer network.
For the methods of setting, consult with your interface specialist.
29
Chapter 4. Work Place
Performing a Test & Reviewing data
2. The “Work Place” consists of 3 items: “Run Test”, “Test Status”, and “Result
Review”.
30
1) Run test Menu
“Run test” is a mode for testing the samples from patients.
① After entering patient information in “Patient ID”, press to move
to the next step.
Patient ID can be entered by using the touch keyboard. You can
also enter Patient ID by using a barcode scanner or by an external
keyboard connected by USB.
** You can also press button to proceed with the test without
enteringthe Patient ID.
② Insert the test device in the “Insert Device” step and the Analyzer will
move to the next step automatically.
31
③ Analyzer checks the conditions of the inserted test device. If the test
device has already been used, E-1 error pops up.
④ After confirming that the test device can be used with STANDARD
F2400, the samples should be prepared according to the instruction for
each test and applied to the sample well of the test device.
·Press
· the “Test Start” immediately after applying
the sample to the test device.
Caution
32
⑤ Analyzer checks the applied sample’s conditions and flow.
⑦ When the mounting is completed, you can start a new test procedure.
33
2) Test Status Menu
“Test Status” menu shows the progress of test in the Analyzer.
Blue section is shown in the following cases:
·If
· the test is completed correctly.
·If
· the result of qualitative test is Negative
·If
· the result of quantitative test is within the normal
range.
Red section is shown in the following cases:
·If
· the test is completed correctly.
· the result of qualitative test is positive
·If
· the result of quantitative test is out of normal range
·If
Black section is shown in the following cases:
·If
· the test result is Invalid
34
① If you select a completed item, you can check details of the test(test
result, name label, procedural control, test date and time, Operator ID,
Patient ID, manufacturing date of the test deivce, software version). In
this window, you can print the test result by pressing on “Print”. Press
“OK” to close the window.
35
① Select from the drop-down list on the right. And
select a test from list shown in [Area A]. All of the test results
belonging to the selected Patient ID will be shown in [Area B]. If you
select a particular test from the drop-down list, you can see the tests
in order from the most recent ones.
A B
② If you select a test from the list in [Area B], you can see the test
details.
·· Test details window for qualitative test – Example) Influenza A/B
36
·· Test details window for quantitative test – Example) HbA1c
37
Chapter 05. Calibration
Calibration Set Test
Calibration Set Test is an essential function that ensures optimal performance
of the Analyzer by checking the methods specified in the Manual.
Time for use of Calibration Set
·Whenever
· the Analyzer is turned on
·When
· Analyzer is dropped
·When
· the result does not match the desired result
·When
· you want to check the performance of the Analyzer and test
device
·Calibration
· Set test is independent from QC test and cannot
replace QC test.
Caution
38
① Insert the Cal-1 Device into the test slot of the Analyzer.
② After checking the Cal-1 Device, insert the Cal-2 Device into the test slot
of the Analyzer.
③ After checking the Cal-2 Device, insert the Cal-3 Device into the test slot
of the Analyzer.
39
④ After inserting all the calibration devices, the analyzers performs the
calibration analysis.
⑤ When the analysis is completed, you can check the result screen.
·· Calibration check- OK screen
40
2. Calibration Review Menu
The result of calibration can be checked in “Calibration Review”.
① You can check the details by selecting the Result from the list.
·· Calibration Review - OK screen
41
② Select a list and press “Delete Record”( ) to clear the selected
items. Press the “Delete All”( ) to clear all results.
42
Chapter 6. Quality Control
Quality Control Test
To verify whether the Analyzer’s system is functioning properly and whether
the test procedure is correct, perform QC with one or more levels of quality
control materials.
Time for use of Calibration Set
·When
· Analyzer is used for the first time
·When
· usinga tes device of newly opened package.
·When
· Analyzer is dropped
·When
· the result does not match the expected result
·When
· the result is lower or higher than expected even after repeated
tests.
·When
· you want to check the performance of the Analyzer and test
device
Before using the Control
·Check
· the expiration date of quality control material. Do not use after
expiration or disposal date (3 months after the package of quality control
material is opened), Consider the earlier date between those two.
QC Mode consists of “Run QC”, “QC Result Review” and “Monthly Review”.
1. Run QC Menu
Run QC consists of “QC with Control Solution” and “External QC”.
43
1) QC with Control solution
① Select “QC with Control solution” and the next screen will be shown.
44
④ The Analyzer checks the conditions of inserted test device. If the test
device has already been used, an E-1 error pops up.
⑥ The F2400 Analyzer will check the conditions and flow of applied
sample.
45
⑦ After the check is completed, the test device is mounted inside the
Analyzer.
2) External QC
46
③ Analyzer checks the conditions of inserted test device. If the test
device has already been used, an E-1 error pops up.
47
·· Example of Quantitative test
⑥ After applying the prepared quality control material, press “QC Start”
immediately.
⑦ The F2400 Analyzer checks the conditions and flow of applied sample.
48
⑧ Once the verification is completed, the test device is mounted inside
the Analyzer.
ⓒ
ⓔ ⓓ
ⓐ
49
d. If you select the test you want to view from the drop-down list,
you can check the QC result of selected test.
Example) The results obtained when result retrieval period is set
to 2018.1.17-2018.4.19 and then HbA1c is selected from the drop-
down list
50
② If you select from the list, you can check the test details.
51
Chapter 07. Cleaning & Maintenance
Cleaning of Analyzer
To prevent any malfunctioning of Analyzer, keep the test slot free from
specimen moisture or dust. Clean the Analyzer with a lint-free cloth and
suitable cleaning solution (example: mild soap, 70% ethanol or isopropyl alcohl).
If the Analyzer is in a special environment (example: operating room), it
is recommended to clean it with a mixture of 1-propanol, 2-propanol and
glutaraldehyde (trade name: Bacillol plus).
·Do
· not use abrasive cleaners or liquid preservatives.
Otherwise, the display screen may be damaged. Before
Caution cleaning, always turn off the Analyzer and disconnect the
power cables.
52
Chapter 08. Message & Troubleshooting
Warning Message
53
Warning: Incorrect IP address
Incorrect IP address entered.
Solution
Confirm the entered IP address, and enter
the correct IP address.
Error Message
54
E04: Temperature error
The ambient temperature deviates from
the operating temperature range of the test
device.
Solution
Move the Analyzer and test device into
proper temperature range and perform
the test. Do not artificially heat or cool the
Analyzer.
55
E12: Calibration overdue error
Time limit for calibration has been exceeded.
Solution
If same error persists after calibrating,
contact the SD Biosensor.
56
ANNEX 01. Information for Professional
Medical Personnel
Electromagnetic compatibility
The electrostatic discharge immunity test shall comply with the international
standard IEC 61000-4-2. In addition, this Analyzer meets electromagnetic
compatibility requirements in accordance with EN 61326 and there is no
interference with other electronic equipment.
Disposal
Dispose of the STANDARD F2400 Analyzer in accordance with local regulations
for disposal of electrical and electronic equipment. Waste Electrical Electronic
Equipment (WEEE) regulation corresponds to the European Directive
2002/96/EC for reduction of wastes from the Analyzer. Before any disposal of
STANDARD F2400 Analyzer, please contact the concerned authorities or staff.
57
MEMO
58
MEMO
59
F2400
Manufacturer
SD BIOSENSOR, Inc.
Headoffice
C-4th&5th, 16, Deogyeong-daero 1556beon-gil,
Yeongtong-gu, Suwon-si, Gyeonggi-do, 16690,
REPUBLIC OF KOREA
Manufacturing site
74, Osongsaengmyeong 4-ro, Osong-eup,
Heungdeok-gu, Cheongju-si, Chungcheongbuk-
do, 28161, REPUBLIC OF KOREA
www.sdbiosensor.com
60