Drägerservice Mode: General Notes 3

Download as pdf or txt
Download as pdf or txt
You are on page 1of 40

Contents

Dräger Medizintechnik

DrägerService Mode

A
B 1 General notes 3
C

1.1 Overview of possible tests ............................................................................................4

1.2 Accessing the DrägerService Mode ...........................................................................5

1.3 Exiting the DrägerService Mode ..................................................................................6

2 DrägerService Mode "Information" 7

2.1 O2 sensor .........................................................................................................................8

2.1.1 Reading out the O2 sensor data ................................................................8

2.2 Software/operating hours .......................................................................................... 10

2.3 Options ........................................................................................................................... 11

2.4 Service interval .............................................................................................................. 13

3 Calibrations 15

3.1 O2 sensor ...................................................................................................................... 16

3.2 Scales ............................................................................................................................. 18

3.3 Height adjustment ........................................................................................................ 20

3.3.1 Adjusting the manual threshold ............................................................... 20

3.3.2 Adjusting the automatic threshold ........................................................... 21

3.4 Tilt adjustment ............................................................................................................... 22

3.4.1 Adjusting the manual threshold ............................................................... 22

3.4.2 Adjusting the automatic threshold ........................................................... 23

4 Logs 24
For internal use only. Copyright reserved.

4.1 Error log .......................................................................................................................... 25

4.2 Error history ................................................................................................................... 26

4.3 Operation history .......................................................................................................... 27


R6150000SM1IVZ.fm

6150.000 Caleo DrägerService Mode Page I


Contents
Dräger Medizintechnik

5 Diagnosis 28

5.1 Operating voltages ....................................................................................................... 29


A
B
C 5.2 Keypad ............................................................................................................................ 30

5.2.1 Control keys test ......................................................................................... 30

5.3 Sensor PCB .................................................................................................................. 31

5.4 Actuator PCB ................................................................................................................ 32

5.5 Output signals ............................................................................................................... 34

5.6 Digital signals ................................................................................................................ 36

5.7 Interface board .............................................................................................................. 37

5.7.1 Nurse call ...................................................................................................... 37

5.7.2 Medibus ........................................................................................................ 39

5.7.3 Modem .......................................................................................................... 40


For internal use only. Copyright reserved.
R6150000SM1IVZ.fm

6150.000 Caleo DrägerService Mode Page II


DrägerService Mode
Dräger Medizintechnik

1 General notes
A
B
C
The DrägerService Mode has four menu levels.

The first menu level contains the following items:


− Information
− Calibrations
− Logs
− Diagnosis
− Back

Service Mode

Information

Calibrations

Logs

Diagnosis

Back

Fig. 1: Start page of DrägerService Mode

Each menu level has submenus.

DrägerService Mode can be used to read out device data and activate/deactivate optional
features.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 3 of 40


1.1 Overview of possible tests
Dräger Medizintechnik

Information
O2 sensor O2 sensor
A Software/operating hours Software/operating hours
B
C Options Options
Service intervals Service interval
Calibrations
O2 sensor O2 sensor
Scales Scales
Height adjustment (optional) Height adjustment (optional)
Tilt adjustment Tilt adjustment
Logs
Error log Error log
Error history Error history
Operation history Operation history
Diagnosis
Operating voltages Operating voltages
Keypad Keypad
Sensor PCB Sensor PCB
WT2 Actuator PCB Actuator PCB
Output signals Output signals
Digital feedback signals Digital signals
Interfaces

Nurse call Nurse call


Medibus Medibus
Modem Modem
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 4 of 40


1.2 Accessing the DrägerService Mode
Dräger Medizintechnik

• Plug the power plug of the Caleo into the mains socket.
• Switch on the Caleo at the power switch.

The Caleo runs through its self-test.


A
B
C
• Press the "menu" and "silence" keys (a) simultaneously for 4 seconds.

Dräger

Fig. 2: Front view of the display housing

The following menu appears on the EL display:

Service Mode

Information

Calibrations

Logs

Diagnosis

Back

Fig. 3: Menu page of DrägerService Mode

• Use the control knob to select the desired menu.


For internal use only. Copyright reserved.

• Press the control knob to activate the desired menu.

The EL display shows the respective submenu.


• Use the control knob to select the desired submenu.
R6150000SM1.fm 06.02.01

• Press the control knob to activate the desired submenu.

6150.000 Caleo Released 01/2001 DrägerService Mode Page 5 of 40


Dräger Medizintechnik

If the control knob has additional functions, the required command will be shown directly on
the EL display.

If additional functions can be selected using the EL display key, it will be shown on the EL
display.
A
B
C
1.3 Exiting the DrägerService Mode

• Exit the DrägerService Mode as follows: Switch off the Caleo or select the „Back“
functions in the respective EL display pages until Caleo is in the operating mode.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 6 of 40


2 DrägerService Mode "Information"
Dräger Medizintechnik

DrägerService Mode "Information" has the following submenus:
− O2 sensor
A
B
C
− Software/operating hours
− Options
− Service interval
− Back

Service Mode: Information

O2 sensor

Software/Operating hours

Options

Service interval

Back

Fig. 4: DrägerService Mode "Information"


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 7 of 40


2.1 O2 sensor
Dräger Medizintechnik

DrägerService Mode "O2 sensor" shows the following O2 measurement data on the EL
display:
− Serial number of the O2 Module PCB
A
B
C
− Serial version of the O2 Module PCB
− Serial numbers of the O2 sensors: sensor 1 and sensor 2
− Date of manufacture of the O2 sensors
− Last calibration date
− Next calibration

2.1.1 Reading out the O2 sensor data

• Switch on the Caleo at the power switch.


• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Information".
• Press the control knob.

The EL display shows the "Information" menu.


• Use the control knob to select the menu item "O2 Sensor".
• Press the control knob.

The EL display shows the serial numbers, the date of manufacture, and the calibration date.

Service Mode: O2 sensor

Sensor 1

PCB serial number:


PCB software version:
Sensor serial number:
Sensor date of manufacture:
Last calibration:
Next calibration:
For internal use only. Copyright reserved.

Sensor 2
R6150000SM1.fm 06.02.01

Fig. 5: DrägerService Mode, O2 sensor 1

6150.000 Caleo Released 01/2001 DrägerService Mode Page 8 of 40


Dräger Medizintechnik

• If a second O2 sensor is fitted, press the right-hand display key (sensor 2) in order to
switch over to the data of the second O2 sensor.

Service Mode: O2 sensor
A
B Sensor 2
C

PCB serial number:


PCB software version:
Sensor serial number:
Sensor date of manufacture:
Last calibration:
Next calibration:

Sensor 1

Fig. 6: DrägerService Mode, O2 sensor 2

• Exit the DrägerService Mode as follows: Switch off the Caleo.


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 9 of 40


2.2 Software/operating hours
Dräger Medizintechnik

DrägerService Mode "Software/operating hours" shows the software version and the
operating hours on the EL display. The operating hours can be adjusted.
• Switch on the Caleo at the power switch.
A
B
C
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Information".
• Press the control knob.

The EL display shows the "Information" menu.


• Use the control knob to select the menu item "Software/operating hours".
• Press the control knob.

The EL display shows the software version and number of operating hours.

Service Mode: Software/Operating hours

Caleo
Software version:

Current operating hours


Adjust using control knob

Operating hours:

Fig. 7: DrägerService Mode, software version/operating hours

• Press the control knob to activate the operating hours.


• Use the control knob to adjust the desired operating hours.
• Press the control knob to store the desired operating hours (Note: The operating hours
For internal use only. Copyright reserved.

should be adjusted within max. 2 minutes. Otherwise the Caleo will return to its normal
operating mode.).
• Exit the DrägerService Mode as follows: Switch off the Caleo.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 10 of 40


2.3 Options
Dräger Medizintechnik

DrägerService Mode "Options" shows the following options on the EL display:
− O2 module
A − Humidity module
B
C
− Mattress module
− Scales module
− Interface board

Each module can be addressed individually: The individual modules have different control
functions which can be selected using the control knob. Activated device options will be
tested during the self-test. Device options that are not available/not installed are marked as
"not installed" in the status field of the EL display.

Module Function
O2 module not installed
Monitoring
Closed-loop control
Humidity module not installed
Closed-loop control
Control
Mattress module not installed
installed
Scales module not installed
installed
Interface board not installed
installed

• Switch on the Caleo at the power switch.


• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Information".
For internal use only. Copyright reserved.

• Press the control knob.

The EL display shows the "Information" menu.


R6150000SM1.fm 06.02.01

• Use the control knob to select the menu item "Options".

6150.000 Caleo Released 01/2001 DrägerService Mode Page 11 of 40


• Press the control knob.
Dräger Medizintechnik

Service Mode: Options

A
B
C
O2 module: not installed
Humidity module: not installed
Mattress module: not installed
Scales module: not installed
Interface: not installed

Fig. 8: DrägerService Mode, options

• Use the control knob to select the respective menu item.


• Press the control knob to activate the menu item.
• Use the control knob to select the desired function.
• If necessary, switch the Caleo off and on again in order to enable the desired function.
• Exit the DrägerService Mode as follows: Switch off the Caleo.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 12 of 40


2.4 Service interval
Dräger Medizintechnik

DrägerService Mode "Service interval" shows the service interval data on the EL display: A
message is displayed when 80% of the set service interval time has elapsed. The values in
the right-hand column show the remaining time until the next service interval.
− Inspection data
A
B
C
− Operating hours
− MTK (metrological check of skin-temperature measurement)

Deactivated messages are indicated by three double dashes. Switch on the Caleo at the
power switch.
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Information".
• Press the control knob.

The EL display shows the "Information" menu.


• Use the control knob to select the menu item "Service interval".
• Press the control knob.

Service Mode: Service interval

Adjustment of service intervals

Inspection: -- -- -- -- -- --

Operating hours: -- -- -- -- -- --

MTK: -- -- -- -- -- --

Fig. 9: EL display, service intervals


For internal use only. Copyright reserved.

• Use the control knob to select the desired menu item "Inspection".
• Press the control knob to activate the desired menu item.
• If applicable, enter the inspection time using the control knob.
R6150000SM1.fm 06.02.01

• Press the control knob to store the inspection time.

6150.000 Caleo Released 01/2001 DrägerService Mode Page 13 of 40


Dräger Medizintechnik

• If applicable, enter the operating hours using the control knob.


• Press the control knob to activate the menu item.

• If applicable, enter the operating hours using the control knob.
• Press the control knob to store the operating hours.
A

• Use the control knob to select "MTK".


B
C

• Press the control knob to activate the menu item.


• If applicable, use the control knob to enter the time for the metrological check.
• Press the control knob to store the time for the metrological check.
• Exit the DrägerService Mode as follows: Switch off the Caleo.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 14 of 40


3 Calibrations
Dräger Medizintechnik

DrägerService Mode "Calibrations" can be used to calibrate the O2 sensor, the scales, the
height adjustment, and the tilt adjustment.
A
B DrägerService Mode "Calibrations" has the following submenus:
C

− O2 sensor
− Scales
− Height adjustment
− Tilt adjustment
− Back

• Switch on the Caleo at the power switch.


• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Calibrations".
• Press the control knob.

The EL display shows the possible calibrations.

Service Mode: Calibrations

O2 sensor

Scales

Height adjustment
Tilt adjustment
Back
For internal use only. Copyright reserved.

Fig. 10: DrägerService Mode, calibrations

• Press the return key to jump to the next higher menu level, or
R6150000SM1.fm 06.02.01

• Exit the DrägerService Mode as follows: Switch off the Caleo.

6150.000 Caleo Released 01/2001 DrägerService Mode Page 15 of 40


3.1 O2 sensor
Dräger Medizintechnik

DrägerService Mode "O2 sensor" can be used to calibrate O2 sensors to 21 vol.% O2.
• Make sure 21% O2 is available.
A • Switch on the Caleo at the power switch.
B
C
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Calibrations".
• Press the control knob.
• Use the control knob to select the menu item "O2 sensor".
• Press the control knob.

Service Mode: O2 calibration


Last calibration:
Next calibration:

Sensor 1
Measured O2 : 0 %

Sensor 2
Measured O2 : 0 %

Press control knob to calibrate

Fig. 11: DrägerService Mode, O2 calibration 1

• Press the control knob to start calibration of the O2 sensors. (Note: If the O2 sensors
are not sufficiently warm it is not possible to carry out a calibration. The EL display
shows the remaining time until the O2 sensors can be calibrated.)

The O2 sensors are calibrated.


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 16 of 40


ServiceMode:
Service Mode:OO22calibration
calibration
Dräger Medizintechnik

Calibratingsensor
Calibrating sensor
Pleasewait
Please wait

A
B
C
###
###

Fig. 12: DrägerService Mode, O2 calibration 2

The message "calibration successful" or "not calibrated" appears on the EL display as soon
as the O2 sensor calibration is finished.

Service Mode: O2 calibration

Sensor 1
Measured O2 : 0 %
Status : calibration successful

Sensor 2
Measured O2 : 0 %
Status : not calibrated

Fig. 13: DrägerService Mode, O2 calibration 3

• Exit the DrägerService Mode as follows: Switch off the Caleo.


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 17 of 40


3.2 Scales
Dräger Medizintechnik

DrägerService Mode "Scales" is used to calibrate the scales to a specific reference value.
The EL display shows the following information:
− Number of previous calibrations
A
B
C
− Current weight
− Adjusting the scales

• Switch on the Caleo at the power switch.


• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Calibrations".
• Press the control knob.
• Use the control knob to select the menu item "Scales".
• Press the control knob.

The EL display shows the following data:

Service Mode: Scales calibration

Number or calibrations: 00

Current weight: 0 g

Adjustment of scales:

Remove any weight from mattress tray


Press control knob to continue

Fig. 14: DrägerService Mode, calibrating the scales 1


For internal use only. Copyright reserved.

Calibration is done in several steps:


• Remove any weight from the mattress tray before adjusting.
• Press the control knob.
R6150000SM1.fm 06.02.01

The scales are adjusted.

6150.000 Caleo Released 01/2001 DrägerService Mode Page 18 of 40


Service Mode: Scales calibration
Dräger Medizintechnik

Put reference weight
on mattress tray

A
B Reference weight: 0000 g
C
Set reference weight
using control knob
Press control knob
to calibrate

Fig. 15: DrägerService Mode, calibrating the scales 2

• Put a reference weight on the scales.

Use the control knob to enter the reference weight.


• Confirm weight.

The scales are calibrated. The EL display shows the reference weight in grams.

Service Mode: Scales calibration

Number of calibrations: 00

Measured weight: 0 g

Press control knob to exit

Fig. 16: DrägerService Mode, calibrating the scales 3

• Press the control knob to exit the menu.


For internal use only. Copyright reserved.

• Exit the DrägerService Mode as follows: Switch off the Caleo.


R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 19 of 40


3.3 Height adjustment
Dräger Medizintechnik

DrägerService Mode "Height adjustment calibration" is used to calibrate the height
adjustment motor. The EL displays shows the previous height adjustment value, the new
height adjustment value, and the automatic calibration.
• Switch on the Caleo at the power switch.
A
B
C
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Calibrations".
• Press the control knob.
• Use the control knob to select the menu item "Height adjustment".
• Press the control knob.

The EL display shows the following information:

Service Mode: Height adjustment calibration

Start autocalibration using control knob

Old threshold: 0
New threshold: 0

Automatic calibration: start

accept

Fig. 17: DrägerService Mode, height adjustment

3.3.1 Adjusting the manual threshold

• Turn the control knob to place the cursor on the "new threshold" field and then press the
control knob.
• Use the control knob to adjust the new threshold and then press the control knob.
For internal use only. Copyright reserved.

• Use the control knob to select the "Accept" field.


• Press the control knob to confirm.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 20 of 40


3.3.2 Adjusting the automatic threshold
Dräger Medizintechnik

• Turn the control knob to place the cursor on the "start" field.

• Press the control knob.

A
B
C
Service Mode: Height adjustment calibration

Device runs automatically!

WARNING!
Risk of hand crush
Keep hands away!
CAUTION!
Risk of damage to hoses etc.
Keep hoses etc. away!

Press control knob to continue

Fig. 18: DrägerService Mode, height adjustment 2

• Press the control knob.

Calibration is carried out automatically.

Service Mode: Height adjustment calibration

Automatic calibration running

###

Fig. 19: DrägerService Mode, height adjustment 3


For internal use only. Copyright reserved.

• Use the control knob to select the "Accept" field.


• Press the control knob to confirm.
• Exit the DrägerService Mode as follows: Switch off the Caleo.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 21 of 40


3.4 Tilt adjustment
Dräger Medizintechnik

DrägerService Mode "Tilt adjustment calibration" is used to calibrate the tilt adjustment
motor.
• Switch on the Caleo at the power switch.
A
B
C
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Calibrations".
• Press the control knob.
• Use the control knob to select the menu item "Tilt adjustment".
• Press the control knob.

Service Mode: Tilt adjustment calibration

Start autocalibration using control knob

Old threshold: 0
New threshold: 0
Automatic calibration: start

accept

Fig. 20: DrägerService Mode, tilt adjustment motor

3.4.1 Adjusting the manual threshold

• Turn the control knob to place the cursor on the "new threshold" field and then press the
control knob.
• Use the control knob to manually adjust a new threshold and then press the control
knob.
• Use the control knob to select the "Accept" field.
For internal use only. Copyright reserved.

• Press the control knob to confirm.


R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 22 of 40


3.4.2 Adjusting the automatic threshold

• Turn the control knob to place the cursor on the "start" field.
Dräger Medizintechnik

• Press the control knob.

A
B
C
Service Mode: Tilt adjustment calibration

Device runs automatically!

WARNING!
Risk of hand crush
Keep hands away!
CAUTION!
Risk of damage to hoses etc.
Keep hoses etc. away!

Press control knob to continue

Fig. 21: DrägerService Mode, tilt adjustment motor 2

• Press the control knob.

Calibration is carried out automatically.

Service Mode: Tilt adjustment calibration

Automatic calibration running

###

Fig. 22: DrägerService Mode, tilt adjustment motor 3


For internal use only. Copyright reserved.

• Use the control knob to select the "Accept" field.


• Press the control knob.
• Exit the DrägerService Mode as follows: Switch off the Caleo.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 23 of 40


4 Logs
Dräger Medizintechnik

DrägerService Mode "Logs" has the following submenus:
− Error log
A
B
C
− Error history
− Operation history
− Back

Service Mode: Logs

Error log

Error history

Operation history

Back

Fig. 23: DrägerService Mode, log


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 24 of 40


4.1 Error log
Dräger Medizintechnik

DrägerService Mode "Error log" shows recorded error events on the EL display. Error
messages are displayed as plain text. Each error has been assigned an error number and the
number of occurrences. Entries in the error log can be deleted by using the „Delete“ button.
A
B
C

The error number displayed in the error log corresponds to the error
description number in the error list.

• Switch on the Caleo at the power switch.


• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Logs".
• Press the control knob.
• Use the control knob to select the menu item "Error log".
• Press the control knob.

If an error has occurred, the EL display will show the number of occurrences of this error.

Service Mode: Error log

Error Description Number


1 Drift Sensoren to high 0
2 Comm. failure on CAN-Bus 0
3 36 Grad Ref. fehlerhaft 0

5 Air Uebertemp. Test fehlerhaft 0


6 Ungueltige configuration 0
7 40 Grad Ref. fehlerhaft 0
8 Air safety relay blocked 0
9 Air safety relay feedback failure 0

Delete

Fig. 24: DrägerService Mode, error log


For internal use only. Copyright reserved.

• The second display key viewed from the left can be used to delete all error events.
• The two display keys on the right can be used to page up/down through the error list.

There will be no symbols on the display keys if there are no error entries up/down the list.
R6150000SM1.fm 06.02.01

• Exit the DrägerService Mode as follows: Switch off the Caleo.

6150.000 Caleo Released 01/2001 DrägerService Mode Page 25 of 40


4.2 Error history
Dräger Medizintechnik

DrägerService Mode "Error history" shows the most recent error events on the EL display.
The most recent error events (maximum of 30 errors) are stored in a chronological order. The
most recent error event has number 1. The EL displays shows the date and time of
A occurrence and a short description. Error history entries can be deleted from the list using
B the "Delete" display key.
C

• Switch on the Caleo at the power switch.


• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Logs".
• Press the control knob.
• Use the control knob to select the menu item "Error history".
• Press the control knob.

Service Mode: Error history


No. Date Time Description
1 2000-10-01 7:20 Short error description
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00

Delete

Fig. 25: DrägerService Mode, error history

• Delete any error entries from the list using the "Delete" display key.
• Exit the DrägerService Mode as follows: Switch off the Caleo.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 26 of 40


4.3 Operation history
Dräger Medizintechnik

DrägerService Mode "Operation history" shows the most recent operation parameters on the
EL display: The "Operation history log" shows the most recent 30 operation steps in
chronological order. The most recent operation step has number 1. The EL displays shows
A the date and the exact time of the operation. Operation history entries can be deleted from
B the list using the "Delete" display key.
C

• Switch on the Caleo at the power switch.


• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Logs".
• Press the control knob.
• Use the control knob to select the menu item "Operation history".
• Press the control knob.

Service Mode: Operation history


No. Date Time Description
1 2000-09-01 7:20 Short description
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00
0 ---.---.------- 0:00

Delete

Fig. 26: DrägerService Mode, operation history

• Delete any operation steps from the list using the "Delete" display key.
• Exit the DrägerService Mode as follows: Switch off the Caleo.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 27 of 40


5 Diagnosis
Dräger Medizintechnik

DrägerService Mode "Diagnosis" is used to read out operating voltages, interfaces, as well
as analog and digital signals.
A
B DrägerService Mode "Diagnosis" has the following submenus:
C

− Operating voltages
− Keypad
− Sensor PCB
− Actuator PCB
− Output signals
− Digital signals
− Interface
− Back

Service Mode: Diagnosis


Operating voltages
Keypad
Sensor PCB
Actuator PCB
Output signals
Digital signals
Interface

Back

Fig. 27: DrägerService Mode, diagnosis


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 28 of 40


5.1 Operating voltages
Dräger Medizintechnik

DrägerService Mode "Operating voltages" shows the measurable operating voltages and
their status messages on the EL display.
• Switch on the Caleo at the power switch.
A
B
C
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Diagnosis".
• Press the control knob.
• Use the control knob to select the menu item "Operating voltages".
• Press the control knob.

Service Mode: Operating voltages diagnosis


Operating voltage (V) Status
+ 5 V sensor/actuator 0.0 Error
+ 5 V sensor/sensor 4.9 ok
+ 5 V normal 0.0 Error
+ 5 V mattress 0.0 Error
+ 12 V EL display 13.2 ok
+ 24 V 0.0 Error
GoldCap 0.0 Error

Fig. 28: DrägerService Mode, operating voltages

Table 1: Status message display

OK = Operating voltage is OK.


Error = Operating voltage is outside tolerance or not present.

• Exit the DrägerService Mode as follows: Switch off the Caleo.


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 29 of 40


5.2 Keypad
Dräger Medizintechnik

DrägerService Mode "Keypad" is used to test the keys on the control unit.

A 5.2.1 Control keys test


B

This test is used to test the keypad.


C

• Switch on the Caleo at the power switch.


• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Diagnosis".
• Press the control knob.
• Use the control knob to select the menu item "Keypad".
• Press the control knob.

Service Mode: Keypad

Press any key


to test keypad

Press control knob to exit test

Key detected:

Fig. 29: DrägerService Mode, keypad test

• Press the keys one after the other.

The EL display shows the detected key in the center line.


• Exit the DrägerService Mode as follows: Switch off the Caleo.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 30 of 40


5.3 Sensor PCB
Dräger Medizintechnik

DrägerService Mode "Sensor PCB" shows the physical readings including status messages
(OK or error) of the Sensor PCB on the EL display.
• Switch on the Caleo at the power switch.
A
B
C
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Diagnosis".
• Press the control knob.
• Use the control knob to select the menu item "Sensor PCB".
• Press the control knob.

The EL display shows the Sensor PCB data.

Service Mode: Sensor PCB


Signal physic. Status

Air 1 0.0 ˚C ok
Air 2 0.0 ˚C ok
Skin 1 0.0 ˚C ok
Skin 2 0.0 ˚C ok
Humidity 0.0 % ok
O2 1 0.0 % ok
O2 2 0.0 % ok

Fig. 30: DrägerService Mode, Sensor PCB

• Exit the DrägerService Mode as follows: Switch off the Caleo.


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 31 of 40


5.4 Actuator PCB
Dräger Medizintechnik

DrägerService Mode "Actuator PCB" shows the physical readings including status
messages (OK or error) of the Actuator PCB on two pages on the EL display.
• Switch on the Caleo at the power switch.
A
B
C
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Diagnosis".
• Press the control knob.
• Use the control knob to select the menu item "Actuator PCB".
• Press the control knob.

The EL display shows information about the Actuator PCB.

Service Mode: Actuator PCB, page 1


Signal physic. Status

Mattress 1 0.0 ˚C ok
Mattress 2 0.0 ˚C ok
Mattress 3 0.0 ˚C ok
Mattress 4 0.0 ˚C ok
Heater 1 0.0 ˚C ok
Heater 2 0.0 ˚C ok
Motor current ok
Current limiting ok

Page 2

Fig. 31: DrägerService Mode, WT2 Actuator PCB (1)

• Press the "Page 2" key on the EL display.

The EL display shows additional information about the Actuator PCB.


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 32 of 40


Service Mode: Actuator PCB, page 2
Dräger Medizintechnik

Signal physic. Value Status

Fan 1 0.0 ok
Scales 0g ok

A
B
C

Page 1

Fig. 32: DrägerService Mode, WT2 Actuator PCB (2)

• Exit the DrägerService Mode as follows: Switch off the Caleo.


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 33 of 40


5.5 Output signals
Dräger Medizintechnik

DrägerService Mode "Output signals" shows the control signals of the actuators including
status messages (on or off) on the EL display.
• Switch on the Caleo at the power switch.
A
B
C
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Diagnosis".
• Press the control knob.
• Use the control knob to select the menu item "Output signals".
• Press the control knob.

The EL display shows the components and control signals of the actuators. A feedback is
given whether or not the output signal is present.

To switch „test“ from „off“ to „on“ turn the control knob only one step,
otherwise the status will not be changed.

Service Mode: Output signals

Signal Status Test

Main fan off


Air heater off off
Mattress heater off off
Humidity heater off off
O2 valve off off
Alarm LED off off

Fig. 33: DrägerService mode, output signals of actuators


For internal use only. Copyright reserved.

• If applicable, enter the respective component using the control knob.


• Press the control knob.
R6150000SM1.fm 06.02.01

• Use the control knob to select the respective function (on or off).

6150.000 Caleo Released 01/2001 DrägerService Mode Page 34 of 40


• Press the control knob.
Dräger Medizintechnik

• Exit the DrägerService Mode as follows: Switch off the Caleo.



A
B
C
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 35 of 40


5.6 Digital signals
Dräger Medizintechnik

DrägerService Mode "Digital signals" shows the digital signals including status messages (0
or 1) on the EL display.
• Switch on the Caleo at the power switch.
A
B
C
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Diagnosis".
• Press the control knob.
• Use the control knob to select the menu item "Digital signals".
• Press the control knob.

Service Mode: Digital signals

Signal Status Test

Air heater excess temperature 0


Air heater safety relay 1 0
Humidity heater excess temperature 1 0
Mattress heater excess temperature 0
Mattress heater safety relay 1 0

Fig. 34: DrägerService Mode, digital signals

• Exit the DrägerService Mode as follows: Switch off the Caleo.


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 36 of 40


5.7 Interface board
Dräger Medizintechnik

DrägerService Mode "Interface" is used to test the following interfaces:
− Nurse call
A − Medibus
B
C
− Modem
− Back

Service Mode: Interface

Nurse call

Medibus

Modem

Back

Fig. 35: DrägerService Mode, interface

5.7.1 Nurse call

The "Nurse call" test is used to test the relay for the external interface.
• Switch on the Caleo at the power switch.
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Diagnosis".
• Press the control knob.
• Use the control knob to select the menu item "Interface board".
• Press the control knob.
For internal use only. Copyright reserved.

• Use the control knob to select the menu item "Nurse call".
• Press the control knob.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 37 of 40


Service Mode: Nurse call
Dräger Medizintechnik

Nurse call

A
B
C
To activate/deactivate
press keys below

off on

Fig. 36: DrägerService Mode, nurse call

• If applicable, switch on/off the nurse call using the display keys.
• Exit the DrägerService Mode as follows: Switch off the Caleo.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 38 of 40


5.7.2 Medibus
Dräger Medizintechnik

The Medibus test is used to test the RS232 interface; the test result (OK or error) is shown

on the EL display.
• Connect a shorting plug to the Medibus interface.
A
B • Switch on the Caleo at the power switch.
C

• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Diagnosis".
• Press the control knob.
• Use the control knob to select the menu item "Interface board".
• Press the control knob.
• Use the control knob to select the menu item "Medibus".
• Press the control knob.

The EL display shows the test result.

Service Mode: Medibus

Connect shorting plug


to Medibus connection

Test result: error

Fig. 37: DrägerService Mode, Medibus

• Exit the DrägerService Mode as follows: Switch off the Caleo.


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 39 of 40


5.7.3 Modem
Dräger Medizintechnik

The modem test is used to test the modem interface; the test result (ok or error) is shown on

the EL display.
• Switch on the Caleo at the power switch.
A
B • As soon as the device has finished its self-test, press the "menu" and "silence" keys
C
simultaneously for 4 seconds.

The Caleo should now be in DrägerService Mode.


• Use the control knob to select the menu item "Diagnosis".
• Press the control knob.
• Use the control knob to select the menu item "Interface board".
• Press the control knob.
• Use the control knob to select the menu item "Modem".
• Press the control knob.
• Connect a shorting plug to the modem interface.

The EL display shows the test result.

Service Mode: Modem

Connect shorting plug


to modem connection

Test result: error

Fig. 38: DrägerService Mode, modem

• Exit the DrägerService Mode as follows: Switch off the Caleo.


For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01

6150.000 Caleo Released 01/2001 DrägerService Mode Page 40 of 40