Drägerservice Mode: General Notes 3
Drägerservice Mode: General Notes 3
Drägerservice Mode: General Notes 3
Dräger Medizintechnik
DrägerService Mode
A
B 1 General notes 3
C
3 Calibrations 15
4 Logs 24
For internal use only. Copyright reserved.
Service Mode
Information
Calibrations
Logs
Diagnosis
Back
DrägerService Mode can be used to read out device data and activate/deactivate optional
features.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01
Dräger
Service Mode
Information
Calibrations
Logs
Diagnosis
Back
If additional functions can be selected using the EL display key, it will be shown on the EL
display.
A
B
C
1.3 Exiting the DrägerService Mode
• Exit the DrägerService Mode as follows: Switch off the Caleo or select the „Back“
functions in the respective EL display pages until Caleo is in the operating mode.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01
O2 sensor
Software/Operating hours
Options
Service interval
Back
The EL display shows the serial numbers, the date of manufacture, and the calibration date.
Sensor 1
Sensor 2
R6150000SM1.fm 06.02.01
• If a second O2 sensor is fitted, press the right-hand display key (sensor 2) in order to
switch over to the data of the second O2 sensor.
Service Mode: O2 sensor
A
B Sensor 2
C
Sensor 1
The EL display shows the software version and number of operating hours.
Caleo
Software version:
Operating hours:
should be adjusted within max. 2 minutes. Otherwise the Caleo will return to its normal
operating mode.).
• Exit the DrägerService Mode as follows: Switch off the Caleo.
R6150000SM1.fm 06.02.01
Each module can be addressed individually: The individual modules have different control
functions which can be selected using the control knob. Activated device options will be
tested during the self-test. Device options that are not available/not installed are marked as
"not installed" in the status field of the EL display.
Module Function
O2 module not installed
Monitoring
Closed-loop control
Humidity module not installed
Closed-loop control
Control
Mattress module not installed
installed
Scales module not installed
installed
Interface board not installed
installed
A
B
C
O2 module: not installed
Humidity module: not installed
Mattress module: not installed
Scales module: not installed
Interface: not installed
Deactivated messages are indicated by three double dashes. Switch on the Caleo at the
power switch.
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.
Inspection: -- -- -- -- -- --
Operating hours: -- -- -- -- -- --
MTK: -- -- -- -- -- --
• Use the control knob to select the desired menu item "Inspection".
• Press the control knob to activate the desired menu item.
• If applicable, enter the inspection time using the control knob.
R6150000SM1.fm 06.02.01
− O2 sensor
− Scales
− Height adjustment
− Tilt adjustment
− Back
O2 sensor
Scales
Height adjustment
Tilt adjustment
Back
For internal use only. Copyright reserved.
• Press the return key to jump to the next higher menu level, or
R6150000SM1.fm 06.02.01
Sensor 1
Measured O2 : 0 %
Sensor 2
Measured O2 : 0 %
• Press the control knob to start calibration of the O2 sensors. (Note: If the O2 sensors
are not sufficiently warm it is not possible to carry out a calibration. The EL display
shows the remaining time until the O2 sensors can be calibrated.)
A
B
C
###
###
The message "calibration successful" or "not calibrated" appears on the EL display as soon
as the O2 sensor calibration is finished.
Sensor 1
Measured O2 : 0 %
Status : calibration successful
Sensor 2
Measured O2 : 0 %
Status : not calibrated
Number or calibrations: 00
Current weight: 0 g
Adjustment of scales:
A
B Reference weight: 0000 g
C
Set reference weight
using control knob
Press control knob
to calibrate
The scales are calibrated. The EL display shows the reference weight in grams.
Number of calibrations: 00
Measured weight: 0 g
Old threshold: 0
New threshold: 0
accept
• Turn the control knob to place the cursor on the "new threshold" field and then press the
control knob.
• Use the control knob to adjust the new threshold and then press the control knob.
For internal use only. Copyright reserved.
• Turn the control knob to place the cursor on the "start" field.
• Press the control knob.
A
B
C
Service Mode: Height adjustment calibration
WARNING!
Risk of hand crush
Keep hands away!
CAUTION!
Risk of damage to hoses etc.
Keep hoses etc. away!
###
Old threshold: 0
New threshold: 0
Automatic calibration: start
accept
• Turn the control knob to place the cursor on the "new threshold" field and then press the
control knob.
• Use the control knob to manually adjust a new threshold and then press the control
knob.
• Use the control knob to select the "Accept" field.
For internal use only. Copyright reserved.
• Turn the control knob to place the cursor on the "start" field.
Dräger Medizintechnik
• Press the control knob.
A
B
C
Service Mode: Tilt adjustment calibration
WARNING!
Risk of hand crush
Keep hands away!
CAUTION!
Risk of damage to hoses etc.
Keep hoses etc. away!
###
Error log
Error history
Operation history
Back
The error number displayed in the error log corresponds to the error
description number in the error list.
If an error has occurred, the EL display will show the number of occurrences of this error.
Delete
• The second display key viewed from the left can be used to delete all error events.
• The two display keys on the right can be used to page up/down through the error list.
There will be no symbols on the display keys if there are no error entries up/down the list.
R6150000SM1.fm 06.02.01
Delete
• Delete any error entries from the list using the "Delete" display key.
• Exit the DrägerService Mode as follows: Switch off the Caleo.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01
Delete
• Delete any operation steps from the list using the "Delete" display key.
• Exit the DrägerService Mode as follows: Switch off the Caleo.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01
− Operating voltages
− Keypad
− Sensor PCB
− Actuator PCB
− Output signals
− Digital signals
− Interface
− Back
Back
Key detected:
Air 1 0.0 ˚C ok
Air 2 0.0 ˚C ok
Skin 1 0.0 ˚C ok
Skin 2 0.0 ˚C ok
Humidity 0.0 % ok
O2 1 0.0 % ok
O2 2 0.0 % ok
Mattress 1 0.0 ˚C ok
Mattress 2 0.0 ˚C ok
Mattress 3 0.0 ˚C ok
Mattress 4 0.0 ˚C ok
Heater 1 0.0 ˚C ok
Heater 2 0.0 ˚C ok
Motor current ok
Current limiting ok
Page 2
Fan 1 0.0 ok
Scales 0g ok
A
B
C
Page 1
The EL display shows the components and control signals of the actuators. A feedback is
given whether or not the output signal is present.
To switch „test“ from „off“ to „on“ turn the control knob only one step,
otherwise the status will not be changed.
• Use the control knob to select the respective function (on or off).
Nurse call
Medibus
Modem
Back
The "Nurse call" test is used to test the relay for the external interface.
• Switch on the Caleo at the power switch.
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.
• Use the control knob to select the menu item "Nurse call".
• Press the control knob.
R6150000SM1.fm 06.02.01
A
B
C
To activate/deactivate
press keys below
off on
• If applicable, switch on/off the nurse call using the display keys.
• Exit the DrägerService Mode as follows: Switch off the Caleo.
For internal use only. Copyright reserved.
R6150000SM1.fm 06.02.01
The Medibus test is used to test the RS232 interface; the test result (OK or error) is shown
on the EL display.
• Connect a shorting plug to the Medibus interface.
A
B • Switch on the Caleo at the power switch.
C
• As soon as the device has finished its self-test, press the "menu" and "silence" keys
simultaneously for 4 seconds.
The modem test is used to test the modem interface; the test result (ok or error) is shown on
the EL display.
• Switch on the Caleo at the power switch.
A
B • As soon as the device has finished its self-test, press the "menu" and "silence" keys
C
simultaneously for 4 seconds.