Introduction To Spectroscopic Methods of Analysis
Introduction To Spectroscopic Methods of Analysis
SPECTROSCOPIC METHODS OF
ANALYSIS
1
Introduction
2
Introduction
The heat the flame causes electrons in the element to move from their lower
energy ground state to a higher energy excited state. When these excited
electrons fall back to their lower energy levels,
they emit energy in the form of light of a characteristic wavelength which
is the color of the flame we observe.
4
Flame Tests
Features Uses
5
Colorimetry
Used to measure the absorbance of light by a colored solution.
The amount of light absorbed is measured and recorded as the
"absorbance (A)".
The wavelength is selected for maximum absorption for the colored solution.
We can use a "color wheel" to determine the color of light to use.
Features Uses
7
Colorimetry
8
Determining the Concentration of a Solution of CuSO4(aq)
1. Determine (color) of light to use for the colorimetric analysis.
3. Prepare dilute copper sulfate solutions, calculating the concentration of each solution.
4. Use the colorimeter to measure the absorbance of each of these solutions of known concentration.
5. Next we plot these results on a graph and draw a "line of best fit" through the data points.
x
z Electric Field
Magnetic Field
Direction of
propagation
+ Wavelength ()
Electric Field
Amplitude (A)
0
11
- Time or Distance
Wave Properties
The relationship between λ and ν:
Speed of light = Wavelength x Frequency
C = λν
λ = C/ν
ν = C/λ
For electromagnetic waves, the speed of light (C) is a constant (3.00x108 m/s).
Electromagnetic spectrum
12
Definitions
Period (p): the time required for one cycle to pass a fixed point in space.
Frequency (): the number of cycles which pass a fixed point in space per
second.
Wavelength (): The distance between two identical adjacent points in a wave
(usually maxima or minima).
Radiant Power (P): The amount of energy reaching a given area per second.
(Unit in watts)
UV 180 – 380 nm
Mid-IR 2.5 – 50 μm 15
Spectroscopic Measurements
Absorption Methods
Absorption is the measure of the decrease in radiant power.
Quantitative absorption methods require 2 power measurements:
- Po= incident radiant power
- P = transmitted radiant power
Po P
Transmittance Absorbing solution
i) 42.2%
ii) 4.73%
iii) 97.3%
17
Absorption of UV-Vis Radiation
3. The energy from the absorbed UV light is equal to the energy difference
between the higher energy state and the ground state.
18
Basic Principle of UV-Vis Spectroscopy
Beer-Lambert Law
Shows linear relationship between absorbance (A), concentration (C) of
the species measured, sample path length (b) and the absorptivity (a)
Beer-Lambert Law is expressed through this equation:
A=εbC
Where:
A: Absorbance
ε : Absorptivity (L/mol cm)
b: Path length (cm)
C: Concentration (mol/L) 19
Applications of Beer’s Law
Answer:
i) C = A / ε b = 0.64 / 120x0.5 = 0.011 M
ii) T = 10 –A = 10 -0.64 = 0.23
23% of the power of the incident beam is transmitted. 20
Instrumentation
Components of Spectrophotometer
Detector
Light source
Mono-chromator
Sample cell
Recorder
21
Visible Tungsten lamp Glass/quartz Phototube, diode array
Example
Toluene is an important industrial chemical
22
Example
Toluene is an important industrial chemical
24
Atomic Absorption Spectroscopy (AAS)
It uses hollow cathode lamps to emit light with these frequencies which is then
absorbed by the sample containing the metal ion.
The amount of light absorbed by the sample is compared to the amount of light
absorbed by a set of standards of known concentration. 25
Atomic Absorption Spectroscopy (AAS)
Features Uses
26
AAS Instrumentation
27
AAS Instrumentation
28
CALIBRATION CURVE
Calibration Curve:
Used to determine the unknown concentration of an element in a
solution. The instrument is calibrated using several solutions of known
concentrations. The absorbance of each known solution is measured and
then a calibration curve of concentration vs absorbance is plotted.
The sample solution is fed into the instrument, and the absorbance of
the element in this solution is measured. The unknown concentration of the
element is then calculated from the calibration curve
Applications:
Determination of even small amounts of metals (lead, mercury, calcium,
magnesium, etc) as follows:
AAS was used to determine the concentration of lead ions (in ppm) in
fish.
AAS was set up with a lamp that emitted light with a wavelength that is
absorbed by lead atoms.
The graph on the next slide shows the variation of absorbance with
the concentration of lead.
30
Example of a Calibration curve
12 g → mL
? → 10 mL
120 g/2g
32