Acculab Atilon: Operating Instructions

Download as pdf or txt
Download as pdf or txt
You are on page 1of 59

Operating Instructions

ACCULAB ATILON
Electronic Analytical and Precision Balances

98648-015-55
Contents

3 Warnings and Safety Precautions 44 ISO/GLP-compliant Printout/Record

4 Getting Started 46 Data Interface

5 Installation 47 Troubleshooting Guide

11 Operation 48 Care and Maintenance


11 Display and Operating Elements
12 Basic Weighing Function 49 Recycling
14 Calibration/Adjustment
50 Overview
17 Configuration (Operating Menu) 51 Specifications
17 Functions of the Keys during Configuration 55 Accessories
18 Menu Navigation; Example: Setting the Language 56 C-Marking
19 Parameter Settings: Menu 57 Declaration of Conformity
20 Parameter Settings: Overview 58 Confirmation of an EC Type-Approval Certificate
24 Input: ID Number, Date and Time

26 Application Programs
27 Counting
29 Weighing in Percent
31 Calculation
33 Animal Weighing/Averaging
35 Net-total Formulation
37 Totalizing
39 Density Determination
41 Mass Unit Conversion

2
Warnings and Safety Precautions

Safety – Connect only Acculab accessories, as these are optimally


§ To prevent damage to the equipment, please read these designed for use with your Atilon balance. The operator shall
operating instructions carefully before using the balance. be solely responsible for installation and testing of any modifi-
cations to Acculab equipment, including connection of cables
! Do not use this equipment in hazardous areas. or equipment not supplied by Acculab. On request,
Acculab will be happy to provide information on operating
! The balance may be opened only by trained service specifications (in accordance with the Standards for defined
technicians. immunity to interference).

! Disconnect the balance from power before connecting or $ Do not open the balance housing. If the seal is broken, this
disconnecting peripheral devices. will void all claims under the manufacturer’s warranty.

! If you operate the balance under ambient conditions $ If you have any problems with your balance, contact your
subject to higher safety standards, you must comply with local customer service center.
the applicable installation regulations.
Symbols
! Exposure to excessive electromagnetic interference can The following symbols are used in these instructions:
cause the readout value to change. Once the disturbance § indicates required steps
has ceased, the instrument can be used again in
accordance with its intended purpose. $ indicates steps required only under certain conditions
Make sure that no liquid enters the equipment housing;
use only a slightly moistened cloth to clean the balance. > describes what happens after you have performed a
particular step
Installation
! Make sure the voltage rating printed on the power supply – indicates an item in a list
is identical to your local line voltage.
– Proceed with extreme caution when using pre-wired ! indicates a hazard
RS-232 connecting cables, as the pin assignments may
not be compatible with Acculab equipment. Before
connecting the cable, check all pin assignments against
the cabling diagrams and disconnect any lines that are
assigned differently.

! If there is visible damage to the equipment or power cord,


disconnect the equipment from power and lock it in a
secure place to ensure that it cannot be used for the time
being.

3
Getting Started

Storage and Shipping Conditions Installation


– Do not expose the balance to extreme temperatures, Choose a location that is not subject to the following
moisture, shocks, blows or vibration. negative influences:
– Heat (heater or direct sunlight)
Unpacking the Equipment – Drafts from open windows and doors
§ After unpacking the equipment, please check it immediately – Excessive vibration during weighing
for any external damage. – Excessive moisture

$ If you detect any damage, proceed as directed in the Conditioning the Balance
chapter entitled “Care and Maintenance,” under “Safety Moisture in the air can condense on cold surfaces
Inspection.” whenever the equipment is moved to a substantially
warmer place. To avoid the effects of condensation, allow
$ Save the box and all parts of the packaging for any future the balance to sit for 2 hours, at room temperature, before
transport. Disconnect all cables before packing the balance plugging into AC power.
for shipping.
Seal on Balances Verified for Use in Legal
Equipment Supplied Metrology in the EU*:
– Balance For instruments of accuracy class K, EU legislation
– Weighing pan requires that a control seal be affixed to verified balances.
– Pan support (only for models with a round weighing pan) The control seal consists of a sticker. If the seal is broken,
– AC adapter the verification becomes null and void and the balance
must be re-verified.
Additional equipment supplied with models
ATL-224, -124:
– Sliding-door draft shield
– Drip/breeze ring
– Draft shield base plate

Additional equipment supplied with models


ATL-623, -423, -153:
– Round glass draft shield with cover

* Including the Signatories of the Agreement on the


European Economic Area

4
Installation

Setting Up the Balance


Instruments with sliding-door draft shield:
§ Place components inside the chamber in the following order:
– Draft shield base plate
– Drip/breeze ring
– Pan support
– Weighing pan

Instruments with a round glass draft ring:


§ Position the components listed below in the order given:
– Place the lower lid on the balance with the raised edge facing upwards
– Pan support
– Weighing pan
– Glass draft ring
– Place the upper lid on the draft shield ring with the raised edge facing downwards

Instruments with a round weighing pan


§ Position the components listed below in the order given:
– Pan support
– Weighing pan

Instruments with a rectangular weighing pan:


§ Place the weighing pan on the balance

5
Connecting the Balance to AC Power/Safety Precautions
Use only original Acculab AC adapters. For use within
– Europe: part no. 6971412
– U.S./Canada: part no. 6971413
§ 1) Connect the angle plug to the balance
§ 2) Connect the AC adapter to the wall outlet (mains)

AC Adapter with Country-specific Power Cord


Some models come with separate country-specific power cords for the
AC adapter.

§ 1) Connect the angle plug to the balance


§ 2) Select the power cord for your area and connect it to the AC adapter
§ 3) Plug the power cord into the wall outlet (mains)

Use an original Acclulab AC adapter with a wide input voltage range


$ (100 to 240 V~), order no. 6971966, and replaceable power cord:
6900900 (Europe) 6900905 (AUS)
6900901 (US/CDN) 6900902 (ZA)
6971945 (UK) 6971977 (Argentina)
6971973 (India) 6971978 (China)
6971980 (Denmark) 6971975 (Israel)
6971776 (Italy)

Safety Precautions
Plug-in AC Adapter 6971412/6971413:
The AC adapter rated to Class 2 can be plugged into any wall outlet without
additional safety precautions.

Benchtop AC Adapter 6971966:


The AC adapter rated to Class 1 can be plugged into any wall outlet without
additional safety precautions.

The ground terminal is connected to the


balance housing, which can be additionally grounded for operation.
The data interface is also electrically connected to the balance housing (ground).

6
NOTE: This equipment has been tested and found to If you have a Class B digital device, please read and follow
comply with the limits pursuant to part 15 of FCC Rules. the FCC information given below:
These limits are designed to provide reasonable protection “However, there is no guarantee that interference will not
against harmful interference. This equipment generates, occur in a particular installation. If this equipment does
uses and can radiate radio frequency energy and, if not cause harmful interference to radio or television reception,
installed and used in accordance with these instructions, which can be determined by turning the equipment off and
may cause harmful interference to radio communications. on, the user is encouraged to try to correct the interference
by one or more of the following measures:
For information on the specific limits and class of this
equipment, please refer to the Declaration of Conformity. – Reorient or relocate the receiving antenna.
Depending on the particular class, you are either required
or requested to correct the interference. – Increase the separation between the equipment and
receiver.
If you have a Class A digital device, you need to comply
with the FCC statement as follows: “Operation of this – Connect the equipment into an outlet on a circuit different
equipment in a residential area is likely to cause harmful from that to which the receiver is connected.
interference in which case the user will be required to
correct the interference at his own expense.” – Consult the dealer or an experienced radio/TV technician
for help.”

Before you operate this equipment, check which FCC class


(Class A or Class B) it has according to the Declaration of
Conformity included. Be sure to observe the information of
this Declaration.

7
Connecting Electronic Peripheral Devices
§ Make sure to unplug the balance from AC power before you connect or disconnect
a peripheral device (printer or computer) to or from the interface port.

Warmup Time
To ensure accurate results, the balance must warm up before operation
as follows:
– Models ATL-153 and ATL-822: 2 hours
– All other models: 30 minutes
Only after this time will the instrument have reached the required operating
temperature.

Using Verified Balances in Legal Metrology in the EU*:

$ Make sure to allow the equipment to warm up for at least 24 hours after initial
connection to AC power or after a relatively long power outage.

Operation Outside the Temperature Range


! Operating the balance beyond the temperature range of +10…+30°C
(50°…86°F). Differences from the specifications listed in the chapter on
“Specifications” are possible.

* Including the Signatories of the Agreement on the European Economic Area

8
Below-Balance Weighing
A port for a below-balance weighing hanger
is located on the bottom of the balance.

$ Below-balance weighing is not permitted in legal metrology.

§ Open cover plate on the bottom of the balance.


Important: set the balance on its side to access the cover plate. DO NOT turn the
balance upside-down.

§ Using the built-in hook 1: Attach the sample (e.g., using a suspension wire)
to the hanger.

$ Install a shield for protection against drafts if necessary.

9
Leveling the Balance
Purpose:
– To compensate for unevenness at the place of installation

Always level the balance again any time after it has been moved to a different
location. Only the 2 front feet are adjusted to level the balance.
L R
§ Retract the two rear feet (only on models with a rectangular weighing pan).

R L § Turn the 2 front feet as shown in the diagram until the air bubble is centered
within the circle of the level indicator.

> In most cases this will require several adjustment steps.


R
L
§ On models with a rectangular weighing pan: Lower the 2 rear feet until they
touch the surface on which the balance rests.

Anti-theft Locking Device


§ To secure the balance at the place of installation, fasten a chain or a lock to the
lug located on the rear panel of the balance.

$ Anti-theft locking device (chain and lock): order number LC-1

10
Operation

Overview of Display Elements


12 1
2

11

10 3
9 4
8 5
7 6

Pos. Designation Pos. Designation


1 Weight unit 11 Busy symbol: command is being processed (for example,
2 Menu level indicator “Wait for stability icon); after you turn on the power,
3 Symbol: “GLP printing mode active” A will be displayed until you press a key
4 Symbol: “Printing mode active” 12 Weight value displayed in selected weight unit
5 Symbol: “Application program active” Symbols:
6 Calculated-value indicator (i.e., not a weight value) << Save settings and exit the operating menu
7 Symbol: Gross or net value < One menu level higher
8 Symbols for active application V Scroll through menu items
(W, Z, L, V, R, A, C) > Next item on current menu level
9 Symbol: Calibration/adjustment function ↵ Select a parameter setting
10 Symbols for zero range

Overview of Operating Elements


1
ATILON

ON/OFF
2

8 CLEAR MENUE ENTER


ZERO
TARE
PRINT CAL

3
7 4
6 5

Pos. Designation Pos. Designation


1 On/off 6 Start an application program
2 Level indicator 7 Open the operating menu | Select an application program
3 Start calibration/adjustment routine 8 Clear
4 Data output: This key is generally used to cancel functions:
Press this key to send readout values to the built-in data – Quit application program
interface. – Cancel calibration/adjustment routine |
5 Taring Exit the operating menu

11
Basic Weighing Function

Features Using Verified Balances as Legal Measuring


– Taring the balance Instruments in the EU*:
– Printing weights The type-approval certificate for verification applies only to
non-automatic weighing instruments. For automatic
Preparation operation with or without auxiliary measuring devices, you
§ Switch on the balance: must comply with the regulations applicable to the place of
Press (ON/OFF) installation.

§ Tare the balance, if necessary: § Before using the balance as a legal measuring instrument,
Press (TARE) calibrate and adjust it at the place of use using the built-in
motorized calibration weight; for details, see
$ If necessary, change the configuration settings: “Calibration/Adjustment” in this chapter.
See the chapter entitled “Configuration”
$ The temperature range (°C) indicated on the verification
$ If desired, load the factory settings: label must not be exceeded during operation.
See the chapter entitled “Configuration”

Additional Functions
$ Switching off the balance:
Press (ON/OFF)

$ Balance in standby mode:


the current time is displayed

* Including the Signatories of the Agreement on the


Eruopean Economic Area

12
Example
Simple Weighing

Step Key (or instruction) Display/Printout

Balance in standby mode 10:32:30

1. Switch on the balance (ON/OFF) 0.0 g


Self-test is performed,
followed by automatic
initial tare function.

2. Place container on weighing pan + 11.5 g


(in this example: 11,5 g).

3. Tare the balance (TARE) 0.0 g

4. Place sample in container + 132.0 g


(in this example: 132 g).

5. Print weight (PRINT) N + 132.0 g

13
Calibration and Adjustment

Purpose To block calibration/adjustment:


Calibration is the determination of any difference between – Select Cal.-Adj.-blocked in the menu
the measured value displayed and the true weight (mass) – Close the menu access switch on the back of the balance
of a sample. Adjustment is the correction of this difference,
or its reduction to an allowable level within maximum
permissible error limits.

Using Verified Balancess as Legal Measuring Instruments in


the EU*: Before using your balance as a legal measuring
instrument, internal calibration must be performed at the
place of installation.

Features
Calibration/adjustment can be performed only when:
– there is no load on the balance
– the balance is tared
– the internal signal is stable
– for external calibration, the value displayed for the
calibration weight on the balance does not differ from the
nominal weight value by more than 2%
External Calibration in Verified Balances
If these conditions are not met, an error message is When the balance is used in legal metrology, external
displayed “Err 02”. calibration is blocked as follows:
– The setting of the menu access switch is locked
You can use any of the following weight units in (see “To block calibration/adjustment”)
calibration/adjustment: – The cap over the menu access switch is sealed
CAL.Einh. - gramm, Kilogr. or Pound
(not for verified models) For details on generating an ISO/GLP-compliant printout of
calibration/adjustment results, see page 43.
Following calibration/adjustment, the application program
is cleared.

Internal Calibration/Adjustment
In the operating menu, select Cal.-Adj.-Cal.Int.
before beginning.
The built-in motorized calibration weight is applied and
removed automatically for internal calibration.

§ Select calibration/adjustment: Press (CAL)


> The built-in weight is applied automatically
> The balance is adjusted
> The built-in calibration weight is removed.

* Including the Signatories of the Agreement on the


European Economic Area
14
Internal Calibration/Adjustment (Only on ATL…-I und ATL…-V models)

Set the following parameters:


SETUP - Waage - CAL.Just. - Cal.Int. (menu code 1.1. 9. 4)

The built-in motorized calibration weight is applied and removed automatically for internal calibration.

Step Key (or instruction) Display

1. Unload and tare the balance (TARE) 0.0 g

2. Start calibration (CAL) CAL.Int. g

The built-in weight is CAL.RUN.


applied automaticallyg

3. Calibration/adjustment executed CAL.end g

4. The built-in weight is removed 0.0 g

15
External Calibration

Parameters (changes in factory settings):


SETUP - Waage - CAL.Just. - Cal.Ext. (menu code 1.1. 9. 1)
The required calibration weight is configured at the factory (see “Specifications”)

Step Key (or instruction) Display

1. Unload and tare the balance (TARE) 0.0 g

2. Start calibration. (CAL) Cal.Ext. g

Once you store the zero point - 5000.0 g


the required calibration weight
is prompted (flashing display)

3. Apply the prompted calibration 5000.0 g


weight (in this example: 5000 g)
Weight too light:
a minus sign “–” is shown
Weight too heavy:
a plus sign “+” is shown

The display stops flashing


as soon as the weight value is
within the defined limit.

4. Calibration/adjustment executed; Cal.end g


then the calibration weight
is displayed + 5000.0 g

5. Remove the calibration weight 0.0 g

16
Configuration (Operating Menu)

You can configure the balance; i.e., adapt it to individual requirements.

Functions of the Keys during Configuration

Symbol Key Function


V (MENU)* Scroll through menu items
Press and hold
> (ENTER) One menu level lower
↵ (ENTER) Confirm menu item
(CLEAR) Save settings and exit menu from any position
Press and hold
<< (CLEAR) Save settings and exit menu
< (CLEAR) One menu level higher
Indicates menu level

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

17
Menu Navigation

Example: Setting the Language

Step Key (or instruction) Display

1. Open the menu: (MENU)* (hold) Applic.


In weighing mode: first menu item is shown

2. Scroll upward within the Repeatedly: (MENU)* Input


menu level; after the last ...
menu code, the first languag.
code is displayed again

3. Select menu level Repeatedly: (ENTER) english o


(scrolls to the right)

4. Change setting: (MENU)* Espanol


Scroll until the desired setting is shown

5. Confirm the menu code; (ENTER) Espanol o


“o” indicates the active setting

6. Return to the next higher menu (CLEAR) Lengua


level (from the second level)

$ Set other menu items as desired (MENU)*, (ENTER)

7. Save settings Repeatedly: (CLEAR)


and exit menu

or

$ Exit menu without saving changes (ON/OFF)

> Restart your application 0.0 g

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

18
Parameter Settings: Menu
Level 1 Level 2 Level 3 Menu code

Setup Bal.Scal. Ambient Ambient conditions 1. 1. 1.


Balance parameters app.fil. Application filter 1. 1. 2.
Stab.Rng. Stability range 1. 1. 3.
Taring Taring 1) 1. 1. 5
AutoZer. Auto zero 1. 1. 6
Wt.Unit Basic weight unit 1. 1. 7.
Display Display accuracy1) 1. 1. 8.
Cal./adj. Function of the (CAL) key 1. 1. 9.
Cal.Unit. Weight unit for calibratio 1) 1. 1.11.
Interf. Interface Baud Baud rate 1. 5. 1.
parity Parity 1. 5. 2.
Stopbit Number of stop bits 1. 5. 3.
handshk. Handshake mode 1. 5. 4.
databit Number of data bits 1. 5. 5.
Dat.Rec. Output: SBI (ASCII) or printout 1. 5. 6.
Prnt.Out Print (manual/automatic) 1. 6. 1.
Settings for print function StopAut. Stop automatic printing 1. 6. 2.
Tar./prt. Tare balance after ind. print 1. 6. 4.
Prt.Init. Printout of appl. parameters 1. 6. 5.
format Line format for printout 1. 6. 6.
GLP ISO/GLP-compliant printout 1. 6. 7.
time 12h/24h 1. 6. 8.
date-Format 1. 6. 9.
extras Menu 1. 8. 1.
Additional functions Signal Acoustic signal (beep) 1. 8. 2.
Keys Keypad 1. 8. 3.
ext. Keys. Funktion external Keypad 1. 8. 4.
On Mode Power-on mode 1. 8. 5.
Backlit Display backlighting 1. 8. 6.
reset menu Factory settings 1. 9. 1.
Applic. Weigh 2. 1.
Application programs Unit Toggle wt. unit Disp.Dig. Display accuracy 1) 2. 2. 2.
Count Counting. Resolut. Resolution 2. 3. 1.
Ref.Updt. Autom. ref. sample updating 2. 3. 2.
Percent Weighing in percent Dec.Plcs Decimal places 2. 4. 1.
Net-Tot Net-total formulation Comp.Prt. Printout of components 2. 5. 1.
Total Totalizing Comp.Prt. Printout of components 2. 6. 1.
AnimalW. Animal weighing Activty. Animal activity 2. 7. 1
start 2. 7. 2.
Calc. Calculation Method (operator) 2. 8. 1.
Dec.Plcs Decimal places 2. 8. 2.
Density Density determination Dec.Plcs Decimal places 2. 9. 1.
Input Input ID no., Date, time Input: ID no., date, time 3. 1./2./3.
Info Information Version , Ser. No. , Model Display software ver., serial no., model 4. 1./2./3.
Languag. English (factory setting) 5. 1.
Deutsch German 5. 2.
Franc. French 5. 3.
Ital. Italian 5. 4.
Espanol Spanish 5. 5.
. PyCCKV1V1 Russian 5. 6.
polski Polish 5. 7.
Codes Menu shows codes (not texts) 5. 8.
1
) Setting cannot be changed on verified balances

19
Parameter Settings: Overview
ο = Factory setting √ = User-defined setting

Level 1 Level 2 Level 3 Level 4 Menu code

Setup Bal.Scal. Ambient V.Stable Very stable 1. 1. 1. 1


Balance Ambient ο Stable 1. 1. 1. 2
parameters conditions Unstabl 1. 1. 1. 3
(Filter adaptation) V.Unstbl. Very unstable 1. 1. 1. 4
App.filt. ο Final.Rd. Final readout mode 1. 1. 2. 1
Application filter Filling Filling mode 1. 1. 2. 2
Stab.Rng. 1/4-Dig. (digit) 1. 1. 3. 1
Stability range 1/2-Dig. 1. 1. 3. 2
1-Digit 1. 1. 3. 3
ο 2-Digit 1. 1. 3. 4
4-Digit 1. 1. 3. 5
8- Digit1) 1. 1. 3. 6
Taring 1) W/o Stb W/o stability 1. 1. 5. 1
Taring o W/ StAb After stability 1. 1. 5. 2
AutoZer. Off 1. 1. 6. 1
Auto zero ο On 1. 1. 6. 2
Wt.Unit For list of units, see 1. 1. 7. 1
Basic weight “Toggling between Weight
unit Units” 1. 1. 7. 23
Disp.Dig. ) 1
ο All 1. 1. 8. 1
Display Minus 1 1. 1. 8. 2
accuracy
Cal./adj. ο Cal. ext. External cal./adj.1) 1. 1. 9. 1
Function of the Cal. int. Internal cal./adj.2) 1. 1. 9. 2
(CAL) key Blocked (CAL) key blocked 1. 1. 3. 3
Cal.Unit. Unit1) ο Grams 1. 1.11. 1
for calibration Kilogr. Kilograms 1. 1.11. 2
weight Pounds1) 1. 1.11. 3

1
) Setting cannot be changed on verified balances
2
) Only on ATL-…-I, ATL-…-V models

20
Level 1 Level 2 Level 3 Level 4 Menu code

Setup Interf. Baudrate 600 1. 5. 1. 3


Interface ο 1200 1. 5. 1. 4
2400 1. 5. 1. 5
4800 1. 5. 1. 6
9600 1. 5. 1. 7
19200 1. 5. 1. 8
Parity ο Odd 1. 5. 2. 3
Parity Even 1. 5. 2. 4
None 1. 5. 2. 5
StopBit ο 1 Stop 1. 5. 3. 1
No. of stop bits 2 Stop 1. 5. 3. 2
Handshk. Softw.are 1. 5. 4. 1
Handshake ο Hardw.are 1. 5. 4. 2
mode None 1. 5. 4. 3
DataBit ο 7 Bits 1. 5. 5. 1
No. of data bits 8 Bits 1. 5. 5. 2
Dat.Rec. SBI (ASCII) 1) 1. 5. 6. 1
Communication mode ο Printer (GLP-printout) 1. 5. 6. 2
Prnt.Out Print Man. W/o W/o stability 1. 6. 1. 1
Printing fct. (manual/ ο Man.With W/ stability 1. 6. 1. 2
automatic) Aut. W/o Autom. w/o stability 1. 6. 1. 3
Aut.With. Autom. w/ stability 1. 6. 1. 4
StopAut. Stop ο Off Not possible 1. 6. 2. 1
automatic printing On Use print key (PRINT) 1. 6. 2. 2
aut.Cycl. ο EachVAl (1 display update) 1. 6. 3. 1
Time-dependent
automautomatic printing
Tar./prt. Tare ο off 1. 6. 4. 1
the balance after On 1. 6. 4. 2
individual printout

) Note concerning verified balances as legal measuring instruments in the EU*:


1

In the setting “SBI”, the non-verified display digit is not automatically identified.
Please take the corresponding measures or adjust the settings on the peripheral device.

* Including the signatories of the Agreement on the European Economic Area.

21
Level 1 Level 2 Level 3 Level 4 Menu code

Setup Prnt.Out Prt.Init. Off 1. 6. 5. 1


Printing Printing appli- ο All All parameters 1. 6. 5. 2
cation parameters MainPar. Main parameters 1. 6. 5. 2
Format Line 16 Char. 16 characters (w/o ID) 1. 6. 6. 1
format for printout ο 22 Char. 22 characters (w/ ID) 1. 6. 6. 2
2 Char. 2 characters (w/ ID) 1. 6. 6. 3
(Date/Time and weight value)
GLP Printout ο Off 1. 6. 7. 1
as ISO/GLP- Cal.-adj. Only for calib./adj. 1. 6. 7. 2
compliant Always All printouts 1. 6. 7. 3
time 24 h 24-hour format 1. 6. 8. 1
12 h 12-hour format “AM/PM” 1. 6. 8. 2
date dd.mmm.yy Day/month/year 1. 6. 9. 1
mmm.dd.yy Month/day/year 1. 6. 9. 2
Extras Menu* ο CanEdit Can change settings 1. 8. 1. 1
Additional Rd. Only Read only 1. 8. 1. 2
functions
Signal Off 1. 8. 2. 1
Acoustic signal ο On 1. 8. 2. 2
Keys ο Free 1. 8. 3. 1
Keypad Locked 1. 8. 3. 2
ext.Key ο print key 1. 8. 4. 1
Function of the z/tare key (TARE) 1. 8. 4. 2
external switch cal key (CAL) 1. 8. 4. 3
select key (MENU)* 1. 8. 4. 4
CF key (CLEAR) 1. 8. 4. 5
enter key 1. 8. 4. 9

On Mode ο Off/On Off/on/standby 1. 8. 5. 1


Power-on Standby On/standby 1. 8. 5. 2
mode Auto On Auto on 1. 8. 5. 3
Backlit Off 1. 8. 6. 1
Display ο On 1. 8. 6. 2
backlighting
Reset Menu* Yes Restore defaults 1. 9. 1. 1
Reset menu Factory settings ο No Do not restore defaults 1. 9. 1. 2

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

22
Level 1 Level 2 Level 3 Level 4 Menu code

Applic. Weigh 2. 1.
Applic. Unit Disp.Dig.1) ο All 2. 2. 2. 1
programs Toggle units Display Minus 1 2. 2. 2. 2
accuracy
Counting Resolut. ο Disp.Acc. Display accuracy 2. 3. 1. 1
Resolution 10-fold 10 times > disp. 2. 3. 1. 2
Ref.Updt. ο Off 2. 3. 2. 1
Autom. reference Auto 2. 3. 2. 2
updating
Percent Dec.Plcs None No dec. places 2. 4. 1. 1
Weighing Decimal places ο 1 Dec.Pl. 1 decimal place 2. 4. 1. 2
in percent 2 Dec.Pl. 2 decimal places 2. 4. 1. 3
3 Dec.Pl. 3 decimal places 2. 4. 1. 4
Net-Tot Comp.Prt. Off 2. 5. 1. 1
Net-total Component ο On 2. 5. 1. 2
printout
Total Comp.Prt. Off 2. 6. 1. 1
Totalizing Component ο On 2. 6. 1. 2
printout
AnimalW. Activty. Calm (fluct.: 2% of test obj.) 2. 7. 1. 1
Animal Animal activity ο Active (fluct.: 5% of test obj.) 2. 7. 1. 2
weighing V.Active(fluct.: 20% of test obj.) 2. 7. 1. 3
Start Manual 2. 7. 2. 1
ο Auto. Automatic 2. 7. 2. 2
Calc. Method ο Mul. Multiplier 2. 8. 1. 1
Calculation (operator) Div. Divisor 2. 8. 1. 2
Dec.Plcs None No dec. places 2. 8. 2. 1
Decimal ο 1 Dec.Pl.1 decimal place 2. 8. 2. 2
places 2 Dec.Pl.2 decimal places 2. 8. 2. 3
3 Dec. Pl. 3 decimal places 2. 8. 2. 4
Density Dec.Plcs None No dec. places 2. 9. 1. 1
Density Decimal places ο 1 Dec.Pl.1 decimal place 2. 9. 1. 2
determination

1
) Setting cannot be changed on verified balances

23
Input: ID number, Date and Time

Level 1 Level 2 Level 3 Menu


Code
Input ID No. ID input for ISO/GLP compliant data record; 3. 1.
Input 7 characters max.
Permitted characters: 0 to 9; A to Z;
dash/hyphen; space
DATE Menu item for setting the date 3. 2.
TIME Menu item for setting the time 3. 3.

Depending on the setting you select for the menu item “SETUP-PRNT.OUT-DATE,” the date will be displayed in the
following format:

Format Display: Date


DD.MMM.YY

MMM.DD.YY

Depending on the setting you select for the menu item “SETUP-PRNT.OUT-TIME,” the time will be displayed in the
following format:

Time Display: Time


24-hour format

12-hour format

24
Example: ID No., Date and Time

Step Key (or instruction) Display


1. Open the menu:In weighing mode; (MENU)* hold APPLIC.
first menu item is displayed
2. Select “Input” (MENU)* Input
3. Select input for ID no. twice (ENTER) ID NO.
4. Set or change the ID no. – (MENU)* 3------
hold down key to automatically (MENU)* hold
change the digit(s):
5. Scroll within the 7-digit (ENTER) or 3-aBC12
(CLEAR)
6. Save input when you (ENTER) ID NO.
have reached the last digit of the ID no.
7. “Select “Date” (MENU)*, (ENTER) FEB. 08.
8. Change setting – (MENU)* FEB. 10.
hold down key to automatically change the digit(s) (MENU)* hold
9. Toggle between (ENTER) or FEB. 10.
Day/Month/Year positions (CLEAR)
10. Save setting when you reach the “YEAR” position (ENTER) datE
11. Select “Time” (MENU)*, (ENTER) 10.46.23
12. Change setting – (MENU)* 11.46.23
change the digit(s): (MENU)* hold
13. Toggle between (ENTER) or 11.46.32
Hour/Minute/Second positions (CLEAR)
14. Set seconds to zero (MENU)* 11.47.00
15. Save setting when you have reached the “second” position (ENTER) Time
16. Save all settings repeatedly
and exit the menu (CLEAR)

> Restart your application 0.0 g

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

25
Device Information

Level 1 Level 2 Level 3 Example Menu code

InFormation version Show software version rel.36.03 4. 1.


Ser. No. Show serial number 10801234 4. 2.
To toggle focus between
upper and lower display
sections, press (MENU)*
model Show model designation Atl-8201 4. 3.
(to change focus from upper
to middle to lower display
section and back, press (MENU)*

Display of Menu Items: Text or Codes

Languag. English (factory setting) 5. 1.


Deutsch German 5. 2.
Franc. French 5. 3.
ital.Italian 5. 4.
Espanol Spanish 5. 5.
pyccKV1V1 Russian 5. 6.
polski Polish 5. 7.
codes Menu shows codes (not texts) 5. 8.

Application Programs
Using Verified Balances as Legal Measuring Instruments in the EU*:
All application programs can be selected on balances used as legal measuring instruments.
Calculated values are alternately indicated with the following symbols:
– Percent =%
– Piece count (Counting) = pcs
– Computed value = o, G

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

26
Counting

Display symbol: Z Changing the Reference Sample Quantity


Activate function:
Purpose Press the (MENU)* key
With the Counting program you can determine the number Select the desired reference sample quantity (1 to 100):
of parts that each have approximately equal weight. To do In increments of 1: Press the (MENU)* key briefly
this, a known number of parts (the reference sample quan- In increments of 10:
tity) is weighed first, and the individual piece weight (refer- Press and hold the (MENU)* key.
ence weight) is calculated from this result. The quantity is stored in battery-backed memory.
Thus the number of parts subsequently placed on the bal-
ance can be Reference Sample Updating
determined from their weight. Automatic reference sample updating
optimizes the counting accuracy.
Preparation You can activate or deactivate this function in the menu.
§ Select the Counting application in the menu:
see “Configuration.” Automatic reference sample updating
is performed when the requirements,
§ Set the following parameters: including the specified stability criterion, have been met.

Applic. Application program The abbreviation opt, for “optimizing”, is displayed


briefly with the new reference sample quantity.
Count.
Resolut. Resolution
o Disp.Acc. Display accuracy
10-fold 10-fold higher
Printout: Counting
Ref.Updt. Autom. ref. sample updating nRef + 10 : Reference sample quantity
o Off Display accuracy wRef + 21.14 g : Reference weight
Autom. Automatic Qnt + 500 pcs : Calculated quantity

o = Factory setting

* On some models, the keypad overlay shows the German


word “MENUE“ for MENU.

27
Example : Counting parts of equal weight
Parameter: Applic. - Count. (menu code 2. 3.)

Step Key (or instruction) Display/Data output

1. Place empty container + 22.6 g


on the balance

2. Tare the balance (TARE) 0.0 g


3. Add reference sample quantity
to container (in this example:
20 pcs)

4. Changing the reference sample quantity: (MENU)* ref 10pcs

5. Select reference sample quantity: Repeatedly: (MENU)*


In increments of 1 (1, 2, 3, etc. to 100) Press briefly
In increments of 10 (10, 20, etc. to 100) (MENU)* press and hold ref 20pcs

6. Confirm selected reference sample quantity (ENTER) + 20pcs


and start application. The current reference nRef * 20 pcs
weight remains stored until a new reference wRef 1.07 g
weight remains stored until a new reference
is set or the power supply is interrupted

7. Add desired number of pieces. + 500pcs

8. If desired, print quantity (PRINT) Qnt + 500 pcs

9. Toggle display between 1.07 gG


mean piece weight, weight, Repeatedly: + 535.0 g *
quantity (MENU)* + 500pcs *
*
10. Unload the balance – 21pcs
*
11. Repeat as needed, starting from step 7
12. Delete reference value (CLEAR) 0.0 g

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

28
Weighing in Percent

Display symbol: % Changing the Reference Percentage


Activate function:
Purpose Press the (MENU)* key
This application program allows you to obtain weight read- Select the desired reference (1 to 100):
outs in percent which are in proportion to a reference In increments of 1: Press the (MENU)* key briefly
weight. In increments of 10: Press and hold the (MENU)* key.

Preparation The percentage is stored in battery-backed memory.


§ Select the Weighing in percent application in the menu:
see “Configuration.”

§ Set the following parameters:

Applic. Application program


Printout: Weighing in percent
Percent Weighing in percent pRef 100 : Reference percentage
Dec.Plcs. Decimal places Wxx% 111.6 g : Reference weight net xx% for
selected reference percentage
None No decimal places Prc + 94.9 % : Calculated reference percentage
o 1 Dec. Pl. 1 decimal place
2 Dec. Pl. 2 decimal places
3 Dec. Pl. 3 decimal places

o = Factory setting

* On some models, the keypad overlay shows the German


word “MENUE“ for MENU.

29
Example: Determining residual weight in percent
Parameter settings: Applic. Percent (menu code 2. 4.)
Reference percentage: Ref 100%

Step Key (or instruction) Display/Data output

1. Tare the balance (TARE) 0.0 g

2. Place sample equal to 100%


on the balance
(in this example: 111.6 g)

3. Information:
Enter reference percentage (MENU)* ref 100 %
(Changing the reference: see the previous page)

4. Initialize the balance (MENU)* + 100.0 %


*
The current reference weight pRef 100 %
remains stored until a new Wxx% + 111.6 g
reference is set or the power supply is interrupted

5. Remove sample (e.g., for drying)

6. Place unknown weight on balance + 94.9 %


*
(in this example: 105.9 g)

7. If desired, print percentage (PRINT) Prc + 94.9 %

8. Toggle display between Repeatedly: + 105.9 g


*
weight and percentage (MENU)* + 94.9 %
*
9. Clear display of residual weight (CLEAR) + 105.9 g
and reference percentage

10. If desired, print net residual weight (PRINT) N + 05.9 g

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

30
Calculation

Display symbol: C Setting the Factor or Divisor


Activate function:
Purpose Press the (MENU)* key
With this application program you Select a number of up to 7 digits and, if needed,
can calculate weight value using a multiplier or divisor. This one decimal point (0.000001 to 9999999):
can be used, for example, to determine the weight per unit In increments of 1: Press the (MENU)* key briefly
area, or “gsm” weight (grams per square meter), of paper. To increase the value without pressing repeatedly:
Press and hold the (MENU)* key.

Preparation The selected operator is stored in battery-backed memory.


§ Select the Calculation application in the menu:
see “Configuration.”

§ Set the following parameters:

Applic. Application program


Printout: Calculation
Calc. Calculation
Mul + 1.2634 : Multiplier
Method (Operator) Div + 0.6237 : Divisor
Res + 79.7 o : Result
o Mul. Multiplier
div. Divisor
Dec.Plcs. Decimal places
None No Decimal places
o 1Dec. Pl. 1 Decimal places
2Dec. Pl. 2 Decimal places
3Dec. Pl.. 3 Decimal places

o = Factory setting

* On some models, the keypad overlay shows the German


word “MENUE“ for MENU.

31
Example:
Calculating the weight per unit area of paper: An A4 sheet of paper is used in this example, with surface dimensions of 0.210 m +
0.297 m = 0.06237 m2. To determine the weight per unit area, the total weight is divided by the surface.

Parameter settings:
Applic. Calc.-Method-Div. (menu code 2. 8. 1. 2)

Step Key (or instruction) Display/Data output

1. Tare the balance (TARE) 0.00 g

2. Activate divisor input (MENU)* ______0.

3. Set the divisor (in this example: 0.06237): (ENTER),


Position the decimal point, 5+ (MENU)*, __.00000
2+ (ENTER),
Enter numerals Repeatedly __.06000
(MENU)* or press and hold:
(ENTER), etc. __.06237

4. Store the divisor and initialize the balance


The current divisor remains stored in (ENTER) + 0.0 o
battery-backed memory until the setting Div 0.6237
is changed

5. Weight per unit area: + 79.7 o


*
Place an A4 sheet of paper
on the balance

6. If desired, print result (PRINT) Res + 79.7 o

7. Toggle display between Repeatedly: + 4.97 g *


weight and calculated value (MENU)* + 79.7 o *

8. Unload the balance + 0.0 o


*
9. Repeat as needed, starting from Step 5

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

32
Animal Weighing/Averaging

Display symbol: V Changing the Number of Subweighing


Operations
Purpose Activate function:
Use this program to determine the weights of unstable Press the (MENU)* key
samples (e.g., live animals) or to determine weights under Select the desired number of
unstable ambient conditions. With this program, the bal- measurement (1 to 100):
ance calculates the weight as the average of a defined In increments of 1:
number of individual weighing operations (also referred to Press the(MENU)* key briefly
as “subweighing operations”). In increments of 10:
Press and hold the (MENU)* key.
Preparation
§ Select the Animal weighing application in the menu: see The selected number of measurements is stored in
“Configuration.” battery-backed memory

§ Set the following parameters:


Printout: Animal weighing
Applic. Application program mDef : Number of subweighing
20
operations
AnimalW. Animal weighing
x-Net + 410.1 g : Calculated average
activty. Animal activity
Calm Stable readout
o Active Unstable
V.Active Very unstable readout
Start
manual
o Autom. Automatic

o = Factory setting

* On some models, the keypad overlay shows the German


word “MENUE“ for MENU.

33
Example: Determining animal weight with automatic start and 20 subweighing operations (measurements)

Parameter settings: Applic. AnimalW. (menu code 2. 7.)

Step Key (or instruction) Display/Data output

1. Place animal weighing bowl on the balance 22.6 g

2. Tare the balance (TARE) 0.0 g

3. Change the number of subweighing operations: (MENU)* ref 30

4. Set number of measurements: Repeatedly:


In increments of 1 (1, 2, 3, etc. to 100) (MENU)* Press briefly
In increments of 10 (10, 20, etc. to 100) or (MENU)* and hold ref 20

5. Confirm number of measurements and start (ENTER) + 0.0 g


*
automatic animal weighing.
The number of measurements remains stored
in battery-backed memory until the setting is changed

6. Place first animal in bowl. 888


The balance delays the start of measure 20
ments until the difference between 2 19
measurements meets the criterion ....
1
$ If you selected the manual start mode, press (ENTER)

7. Read off the result + 410.1 g G


*
The result is displayed with the
“*” symbol (= calculated value) mDef 20
and remains displayed until the sample x-Net + 410.1 g
(animal) is removed from the load plate (bowl)

8. Unload the balance + 0.0 g


*
9. Weigh next animal (if desired)

Next weighing series begins automatically

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

34
Net-total Formulation

Display symbol: R Features


– Weigh up to 99 components from “0” to a defined total
Purpose component weight.
With this application program you can weigh in individual
components either by their individual weight or by the total – Store component weights
weight. You can print out both the total weight and the indi- (printout shows “Comp xx” with
vidual weights of the components. – display zeroed automatically after value is stored, and
– automatic printout
Preparation
§ Select the Net-total application in the menu: – Clear component memory following cancellation of the
see “Configuration.” weighing sequence (by pressing) and printout of the total
weight.
§ Set the following parameters:
– Toggling between component weight and total weight by
Applic. Application program pressing and holding (MENU)*.(< 2 sec).
Net-TotL. Net-total formulation
– Printout of the total of the individual component weights
Comp.Prt. Printout of components (T-Comp)
Off
o On

o = Factory setting

Printout: Net-total formulation


Comp 2+ 278.1 g : Second component

T-Comp+2117.5 g : Sum of components

* On some models, the keypad overlay shows the German


word “MENUE“ for MENU.

35
Example: Counting parts into a container

Parameter settings:
Applic.-Net-Tot (menu code 2. 5.)

Step Key (or instruction) Display/Data output

1. Place empty container on the balance.


65.0 g

2. Tare the balance (TARE) 0.0 g

3. Add first component + 120.5 g

4. Store component data (ENTER) + 0.0 g


* NET
Comp 1+ 120.5 g

5. Add next component + 70.5 g


* NET

6. Store component data (ENTER) + 0.0 g


* NET
Comp 2+ 70.5 g

7. Weigh in further components Repeat steps


as desired 5 and 6

8. Fill to desired final value view (MENU)* + 191.0 g


*
the current total weight value

9. Print total weight and clear (CLEAR) + 2117.5 g


the component memory T-Comp+ 2117.5 g

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

36
Totalizing

Display symbol: R Features


– Totalizing memory for up to 99 values
Purpose
With this application program you can add values from suc- – Store component weights (printout shows
cessive, mutually independent weight values to a total that “Comp xx” with automatic printout
exceeds the capacity of the balance.
– Toggle display between the current individual weight value
Preparation and the value in totalizing memory by pressing (MENU)*.
§ Select the Totalizing application in the menu:
see “Configuration”. – Printout of the total of the individual component weights
(S-Comp)
§ Set the following parameters:
– To close the application program and print the total weight:
Applic.Application program press (CLEAR)
Total Totalizing
Comp.Prt Printout of components
Off
o On Printout: Totalizing
Comp 2+ 278.1 g : Second
o = Factory setting component
S-Comp+2117.5 g : Totalizing memory

* On some models, the keypad overlay shows the German


word “MENUE“ for MENU.

37
Example: Totalizing weight values

Parameter settings:
Applic. Total Comp.Prt On (Code 2. 6. 1. 2)

Step Key (or instruction) Display/Data output

1. Tare the balance (TARE) 0.0 g

2. Place sample balance + 380.0 g


(in this example: 380 g)

3. Store value in memory (ENTER) + 380.0 g


*
Comp 1+ 380.0 g

4. Remove sample + 0.0 g


*

5. Place the next sample on the


balance (in this example, 575 g) + 575.0 g
*

6. Store value in memory (ENTER) + 955.0 g


*
+ 575.0 g
*
Comp 2+ 575.0 g

7. View the value in totalizing memory (ENTER) + 955.0 g G*

8. Weigh in further components as desired Repeat steps 5 and 6

9. Print total weight and (CLEAR) 0.0 g


clear the totalizing memory S-Comp+ 2117.5 g

38
Density Determination

Display symbol: W Printout for Density Determination


Wa + 20.0 g : Weight in air
Purpose Wfl + 15.0 g : Weight in liquid
This application program lets you determine the density of Rho 4.0 o : Result: density
solid substances using the buoyancy method. You can have of the sample
results displayed with one decimal place, or no decimal
places: see “Configuration” .
Note: the sample holder and suspension wire used in the
example below are not included with the balance.

Preparation
§ Select the Density Determination application in the menu:
see “Configuration.”

§ Set the following parameters:

Applic. Application program


density Density determination
dec.plcs Decimal places
none No decimal places
o 1dec.pl 1 decimal place

o = Factory setting

39
Example: Determining the density of a solid sample.

Parameter settings:
Applic.-Density-Dec.Plcs-1 Dec. Pl (menu code 2. 9. 1. 2)

Step Key (or instruction) Display/Data output

1. Attach sample holder to suspension wire


2. Tare the balance (TARE) 0.0 g
3. Start application program (ENTER)
4. Confirm “Air” display (ENTER) air ?
5. Determine the weight
of the sample in air:
Place sample on the balance + 20.0 g
*
6. Store value for weight in air (ENTER)
7. Remove sample from the balance water ?
8. Determine weight in liquid:
place sample in holder

9. Confirm “water” display (ENTER) 0.0 g


*
10. Immerse sample in liquid + 15.0 g
*
11. Store value for weight in (ENTER)
liquid, view result, and print + 4.0 o
*
Wa + 20.0 g
Wfl + 15.0 g
Rho 4.0 o
12. Delete result (CLEAR)
13. Repeat as desired, starting from Step 3.

40
Mass Unit Conversion

Purpose Features
With this application program you can change the weight – Set the basic unit and display accuracy in the Setup menu:
value displayed from the basic weight unit to any of 4 see “Configuration.”
application weight units (see table on next page). – Set the application weight units and display accuracies in
the Application menu.
Preparation – These settings are stored in battery-backed memory.
§ Select the Unit application for toggling weight units: – The basic unit is active when the balance is powered up.
see chapter on “Configuration” (Parameter Settings)

§ Set the following parameters:

Applic.ation program
Toggle wt, unit
dec.plcs. Display accuracy
o All All decimal places
minus1 Reduced by 1 place

o = Factory setting

Printout for Wt. Unit Toggling


+ 100.0 g : Weight with 16-character data output format
N + 0.22046 lb : Weight with 22-character data output format

13-Jan-2005 08:35 : Data output format for two-line printout:


N + 3.5275 ozt : Date/time and weight

41
Example: Change display from the basic unit (in this example, grams [g]) to pounds [lb] and then to Troy ounces [ozt].

Set the following parameters: Applic. Unit (Code 2. 2.)

Step Key (or instruction) Display

Preparation:
1 Begin selection of an application weight unit (MENU)* None o

2. Select an application unit; Repeatedly:


in this example, pounds (MENU)* pound
(see table on next page)
3. Confirm the weight unit (pounds) (ENTER) pound o

4. Select the next application (ENTER) KEINE o

weight unit; in this example: Troy ounces Repeatedly:


(see table on next page) (MENU)* troy.oz.
5. Confirm weight unit (Troy ounces) (ENTER) troy.oz. o
6. Select other application units if
desired (max. 4 total) (otherwise,
confirm None by pressing )
7. Save selection (CLEAR) 0.00 g

Conversion:
8. Place sample on balance + 100.00 g

9. Toggle unit for weight value Repeatedly: + 0.22046 lb


(ENTER) + 3.5275 ozt

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

42
The following weight units are available in your Atilon balance (in legal metrology, only units permitted by national law are
available):

Menu item Unit Conversion factor Display symbol


1) UserDef. 1) Grams 1,00000000000 o
2) Grams (factory setting) Grams 1.00000000000 g
3) Kilogr. Kilograms 0.00100000000 kg
4) Carats Carats 5.00000000000 o
5) Pounds Pounds 0.00220462260 lb
6) Ounces Ounces 0.03527396200 oz
7) Troy Oz. Troy ounces 0.03215074700 ozt
8) HKTael Hong Kong taels 0.02671725000 tl
9) Sng.Tael. Singapore taels 0.02645544638 tl
10) Twn.Tael Taiwanese taels 0.02666666000 tl
11) Grains Grains 15.4323583500 GN
12) Peny.Wt. Pennyweights 0.64301493100 dwt
13) Milligr. Milligrams 1000.00000000 mg
14) Pt.P.Lb. Parts per pound 1.12876677120 o
15) Chn.Tael Chinese taels 0.02645547175 tl
16) Mommes mommes 0.26670000000 m
17) Austr.Ct. Austrian carats 5.00000000000 Kt
18) Tola Tola 0.08573333810 o
19) Baht Baht 0.06578947436 b
20) Mesghal Mesghal 0.21700000000 o
21) Tons Tons 0.00000100000 t
22) Lb / Oz 2) Pounds : ounces 0.03527396200 lb oz
23) Newton Newton 0.00980665000 N
1
)= User-defined weight unit; can be loaded in the balance over an optional RS-232 or USB interface using a computer program.
) = The format for display of pounds/ounces cannot be changed: xx:yy.yy x=lb, y=oz
2

! Some weight units may be blocked from use in legal metrology, depending on national verification laws.

43
ISO/GLP-compliant Printout/Record

Features – Line format for printout: include data ID codes


You can have device information, ID texts and date and (22 characters; factory setting):
time printed before (GLP header) and after (GLP footer) Setup Prnt.Out Format 22 Char.
the values of a weighing series. These parameters include: (menu code 1. 6. 6. 2)
– Formats for time:
GLP header: setup prnt.0ut time 24 h
– Date (menu code 1. 6. 8. 1) or
– Time at beginning of measurement setup prnt.oUt time 12 h
– Balance manufacturer (menu code 1. 6. 8. 2)
– Balance model
– Balance serial number – Formats for date:
– Software version number setup prnt.0ut date dd.mmm.yy
– Identification number of the current sampling operation (menu code 1. 6. 9. 1) or
setup prnt.oUt date mmm.dd.yy
GLP footer: (menu code 1. 6. 9. 2)
– Date
– Time at end of measurement ! No ISO/GLP-compliant record
– Field for operator signature is output if any of the following settings are configured:
Setup Prnt.Out Print Aut. W/o or
! Operating the Balance with Printer: Aut.With (menu code 1. 6. 1. 3, 1. 6. 1. 4, ) or
(e.g. YDP03-OCE): Format 16 Char. (menu code 1. 6. 6. 1)
§ Select the following settings on the balance and on the
printer: Function Keys
– Software handshake: Transfer header and first measured value: press (PRINT)
Setup interf. handshk. sftware
(menu code 1. 5. 4. 1) > The header is included with the first printout/data record.

Configuration To output header and reference data automatically when an


§ Setting menu codes for the printout (see “Configuration”): application program is active: press
– ISO/GLP-compliant printout or record only for
calibration/adjustment: Exit the application:
Setup Prnt.Out GLP Cal.-Adj. 1) To send the GLP footer: press (CLEAR)
(menu code 1. 6. 7. 2) or 2) Quit application program: press (CLEAR) again
ISO/GLP-compliant printout or record always on:
Setup Prnt.Out GLP Always on
(menu code 1. 6. 7. 3)

44
The ISO/GLP-compliant printout can contain the following lines:

-------------------- Dotted line


17-Aug-2006 10:15 Date/time (beginning of measurement)
ACCULAB Balance manufacturer
Mod. ATL-8201 Model
Ser. no. 10105355 Balance serial number
Ver. no. 00-36-01 Software version
ID 2690 923 ID
-------------------- Dotted line
L ID Measurement series no.
nRef 10 pcs Counting: reference sample quantity
wRef 21.14 g Counting: reference weight
Qnt + 567 pcs Counting result
-------------------- Dotted line
17-Aug-2006 10:20 Date/time (end of measurement)
Name: Field for operator signature
Blank line
-------------------- Dotted line

ISO/GLP-compliant printout for external calibration/adjustment:

-------------------- Dotted line


17-Aug-2006 10:30 Date/time (beginning of measurement)
ACCULAB Balance manufacturer
Mod. ATL-8201 Model
Ser. no. 10105352 Balance serial number
Ver. no. 00-36-01 Software version
ID 2690 923 ID
-------------------- Dotted line
Cal. Ext. Test Calibration/adjustment mode
Set + 5000.0 g Calibration weight
Diff. + 0.2 g Difference determined in calibration
Cal. Ext. Complete Confirmation of completed calibration procedure
Diff. 0.0 g Difference from target following adjustment
-------------------- Dotted line
17-Aug-2006 10:32 Date/time (end of measurement)
Name: Field for operator signature
Blank line
-------------------- Dotted line

45
Data Interface

Purpose Preparation
Your balance comes equipped with an interface port for You can set these parameters for other devices in the Setup
connection to a computer or other peripheral device. You menu (see the chapter entitled “Configuring the Balance”).
can use a computer to change, start and/or monitor the
functions of the balance and the application programs. You will also find a detailed description of the available
data interface commands in the file “Data Interface
Descriptions for ATL Models”, which you can download
from the Acculab website at www.acculab.com by selecting
“Download Instruction Manuals” from the “Service” menu.

Female Interface Connector: The many and versatile properties of these balances can be
fully utilized for printing out records of the results when you
connect your balance to a Acculab data printer. The record-
1 13
14 25
ing capability for printouts makes it easy for you to work in
compliance with ISO/GLP.

25-contact female connector, RS-232:


Pin 1: Signal ground
Pin 2: Data output (TxD)
Pin 3: Data input (RxD)
Pin 4: Internal ground (GND)
Pin 5: Clear to send (CTS)
Pin 6: Not connected
Pin 7: Internal ground (GND)
Pin 8: Internal ground (GND)
Pin 9: Not connected
Pin 10: Not connected
Pin 11: + 12 V
(power supply for Sartorius printer)
Pin 12: Reset _ Out *) For remote switch
Pin 13: + 5 V
Pin 14: Internal ground (GND)
Pin 15: Universal remote switch
Pin 16: Not connected
Pin 17: Not connected
Pin 18: Not connected
Pin 19: Not connected
Pin 20: Data terminal ready (DTR)
Pin 21: Not connected
Pin 22: Not connected
Pin 23: Not connected
Pin 24: Not connected
Pin 25: +5 V
*) = Hardware restart

46
Troubleshooting Guide

Error codes are shown on the main display for approx. 2 seconds.
The program then returns automatically to the previous mode.

Display Cause Solution


No segments appear No AC power is available Check the AC power supply
on the display
The power supply is not Plug in the power supply
plugged in
high The load exceeds the balance capacity Unload the balance
low or Err 54 Something is touching Move the object that is
the weighing pan touching the weighing pan
err 54, typical Weighing system defect Contact Acculab dealer
App.err. Cannot store data: Increase load
Load on weighing pan too light or no
sample on pan while application is active
dis.err. Data output not compatible Change the configuration in
with output format the operating menu
prt.err. Interface port for printer Reset the menu factory settings,
output is blocked or
Contact Acculab dealer
err 02 Calibration parameter Calibrate only when zero
not met; e.g.: is displayed
– balance not tared – Press [Tare] to tare the balance
– load on weighing pan – Unload the balance
err 10 The [Tare] key is blocked when Press [Clear] to clear the tare
there is data in the second tare memory and release the tare key
memory (net-total); only 1 tare
function can be used at a time
err 11 Tare memory not allowed Press [Tare]
The weight readout Unstable ambient conditions Set up the balance in another area
changes constantly (excessive vibration or draft)
at the place of installation
A foreign object is caught between Remove the foreign object
weighing pan and balance housing
The weight readout The balance was not Calibrate/adjust the balance
is obviously wrong calibrated/adjusted
Balance not tared Tare or zero the balance
before weighing before weighing

47
Care and Maintenance

Repairs
§ Repair work must be performed by trained service technicians. Any attempt by
untrained persons to perform repairs may result in considerable hazards for the
user. If the instrument requires repairs, please contact your Acculab dealer.

Cleaning
§ Unplug the AC adapter from the wall outlet (mains supply). If you have an inter-
face cable connected to the balance port, unplug it from the port.

! Make sure that no liquid enters the balance/scale housing.

! Do not aggressive cleaning agents (solvents or similar agents).

§ After cleaning, wipe down the balance with a soft, dry cloth.

On analytical balances remove and clean the weighing pan as follows:


§ Reach beneath the drip/breeze and lift it carefully, together with the pan support,
to avoid damaging the weighing system.

! Make sure that no liquid enters the balance housing.

Cleaning Stainless Steel Surfaces


Clean all stainless steel parts regularly. Remove the stainless steel weighing pan and
thoroughly clean it separately. Use a damp cloth or sponge to clean stainless steel
parts on the balance. You can use any household cleaning agent that is suitable for
use on stainless steel. Clean stainless steel surfaces only by wiping them down.
Then rinse the equipment thoroughly, making sure to remove all residues.
Afterwards, allow the equipment to dry. If desired, you can apply oil to the cleaned
surfaces as additional protection.

48
Recycling

Safety Inspection Information and Instructions on Disposal and Repairs


If there is any indication that safe Packaging that is no longer required must be disposed of at
operation of the balance is no longer warranted: the local waste disposal facility. The packaging is made of
§ Turn off the power and disconnect the equipment from AC environmentally friendly materials that can be used as sec-
power immediately. ondary raw materials.

> Lock the equipment in a secure place to ensure that it can- The equipment, including accessories and
not be used for the time being. batteries, does not belong in your regular
household waste. The EU legislation
Notify your Acculab dealer. Repair work must be performed requires its Member States to collect elec-
by trained service technicians. trical and electronic equipment and dis-
posed of it separately from other unsorted municipal waste
We recommend having the power with the aim of recycling it.
supply inspected by a certified electrician at regular inter-
vals, according to the following checklist:
– Insulating resistance: > 7 megaohms measured with a For disposal in Germany and in the other Member States of
constant voltage of at least 500 volts at a 500 K-ohm load the European Economic Area (EEA), please contact our ser-
– Leakage current: < 0.05 mA measured with a properly vice technicians on location or your Acculab dealer.
calibrated multimeter
In countries that are not members of the European Econom-
ic Area (EEA) or where no Acculab dealers are located,
please contact your local authorities or a commercial
disposal operator.

Prior to disposal and/or scrapping of the equipment, any


batteries should be removed and disposed of in local
collection boxes.

Equipment contaminated with hazardous materials (ABC


contamination) will not be taken back; neither for repair or
disposal.

49
Overview

Specifications

Built-in motorized All ATL...-I and ATL...-V models


calibration weight
AC power source/power AC adapter 230 V or 115 25 V,
requirements, voltage, frequency +15% to – 20%, 48 – 60 Hz
Power consumption VA maximum 16; typical 8 (STNG6)
Approx. hours of operation with the
YRB05Z rechargeable battery pack
(backlighting on) h 35

50
Specifications for Individual Models

Model ATL-224, ATL-224-I ATL-124, ATL-124-I ATL-84, ATL-84-I


Weighing capacity g 220 120 80
Readability g 0.0001 0.0001 0.0001
Tare range (subtractive) g 220 120 80
Repeatability <± g 0.0001 0.0001 0.0001
(std. deviation)
Linearity <± g 0.0002 0.0002 0.0002
Response time (average) s 2.5
Sensitivity drift within +10 to +30°C <±/K 2 · 10–6
Adaptation By selection of 1 of 4 optimized filter levels; display
to ambient conditions update: 0.1 – 0.4 (depends on filter level selected)
External calibration weight g 200 (E2) 100 (E2) 50 (E2)
Operating temperature range +10 … +30 °C (273 ... 303 K, 50° ... 86 °F)
Net weight, approx.: kg 4.4 | 4.7
Weighing pan size mm 90 d
Weighing chamber height mm 230
Dimensions (W + D + H) mm 230 + 303 + 330

Model ATL-623, ATL-423, ATL-153,


ATL-623-I ATL-423-I ATL-153-I
Weighing capacity g 620 420 150
Readability g 0.001 0.001 0.001
Tare range (subtractive) g 620 420 150
Repeatability <± g 0.001 0.001 0.001
(std. deviation)
Linearity <± g 0.002 0.002 0.003
Response time (average) s 1 1 1.3
Sensitivity drift within +10 to +30°C <±/K 2 · 10–6 2 · 10–6 5 · 10–6
Adaptation By selection of 1 of 4 optimized filter levels; display
to ambient conditions update: 0.05 – 0.4 (depends on filter level selected)
External calibration weight
(of at least accuracy class...) g 500 (E2) 200 (F1) 100 (F1)
Operating temperature range +10 … +30 °C (273 ... 303 K, 50° ... 86 °F)
Net weight, approx: kg 3.2 | 3.6 3.2 | 3.6 2.6 | 3.0
Weighing pan size mm 115 d
Dimensions (W + D + H) mm 230 + 303 + 136

51
Specifications for Individual Models

Model ATL-6202, ATL-4202, ATL-2202, ATL-822,


ATL-6202-I ATL-4202-I ATL-2202-I ATL-822-I
Weighing capacity g 6200 4200 2200 820
Readability g 0.01 0.01 0.01 0.01
Tare range (subtractive) g 6200 4200 2200 820
Repeatability <± g 0.01 0.01 0.01 0.01
(std. deviation)
Linearity <± g 0.02 0.02 0.02 0.03
Response time (average) s 1.1 1.1 1.1 1.0
Sensitivity drift within +10 to +30°C < ±/K 2 · 10–6 2 · 10–6 2 · 10–6 5 · 10–6
Adaptation By selection of 1 of 4 optimized filter levels; display
to ambient conditions update: 0.1–0.4 (depends on filter level selected)
(of at least accuracy class...) g 5000 (E2) 2000 (E2) 2000 (F1) 500 (F2)
Operating temperature range +10 … +30 °C (273 ... 303 K, 50° ... 86 °F)
Net weight, approx.: kg 3.1 | 3.5 3.1 | 3.5 3.1 | 3.5 2.0 | 2.6
Weighing pan size mm 180+180 180+180 180+180 115 d
Dimensions (W + D + H) mm 230+303+91 230+303+87

Model ATL-8201, ATL-6201,


ATL-8201-I ATL-6201-I
Weighing capacity g 8200 6200
Readability g 0.1 0.1
Tare range (subtractive) g 8200 6200
Repeatability (std. deviation) <± g 0.1 0.1
Linearity <± g 0.3 | 0.1 0.3 | 0.1
Response time (average) s 1 1
Sensitivity drift within +10 to +30°C < ±/K 10 · 10–6
Adaptation By selection of 1 of 4 optimized filter levels; display
to ambient conditions update: 0.05 – 0.4 (depends on filter level selected)
External calibration weight
(of at least accuracy class...) g 5000 (F2)
Operating temperature range +10 … +30 °C (273 ... 303 K, 50° ... 86 °F)
Net weight, approx.: kg 2.7 | 3.5
Weighing pan size mm 180+180
Dimensions (W + D + H) mm 230+303+91

52
Verified Models with EC-type Approval: Specifications

Model ATL-224-V ATL-124-V ATL-84-V


Type BD ED 100 BD ED 100 BD ED 100
Accuracy class 1) k k k
Weighing capacity, Max 1) g 220 120 80
Scale interval d 1) g 0.0001 0.0001 0.0001
Tare range (subtractive) < 100% of the maximum capacity
Verification scale interval e 1) g 0.001 0.001 0.001
Minimum capacity, Min 1) g 0.01 0.01 0.01
Response time (average) s 2.5
Range of use according 1) g 0.01 – 220 0.01 – 120 0.01 – 80
Allowable ambient
operating temperature °C +17 to +27 (+63°F to +80°F)
Net weight, approx. kg 4.8
Weighing pan size mm 90 d
Weighing chamber height mm 230
Dimensions (W + D + H) mm 230 + 303 + 330

Model ATL-623-V ATL-423-V ATL-153-V


Type BD ED 200 BD ED 200 BD ED 200
Accuracy class 1) K K K
Weighing capacity, Max 1) g 620 420 150
Scale interval d 1) g 0.001 0.001 0.001
Tare range (subtractive) < 100% of the maximum capacity
Verification scale interval e 1) g 0.01 0.01 0.01
Minimum capacity, Min 1) g 0.02 0.02 0.02
Response time (average) s 1 1 1
Range of use according to CD 1) g 0.02 – 620 0.02 – 420 0.02 – 150
Allowable ambient
operating temperature °C +10 to +30 (+50°F to +86°F)
Net weight, approx. kg 3.6
Weighing pan size mm 115 d
Dimensions (W + D + H) mm 230 + 303 + 136

) CD = Council Directive 90/384/EEC for non-automatic weighing instruments; applicable to the European Economic Area
1

53
Verified Models with EC-type Approval: Specifications

Model ATL-6202-V ATL-4202-V ATL-2202-V


Type BD ED 200 BD ED 200 BD ED 200
Accuracy class 1) K K K
Weighing capacity, Max 1) g 6200 4200 2200
Scale interval d 1) g 0.01 0.01 0.01
Tare range (subtractive) < 100% of the maximum capacity
Verification scale interval e 1) g 0.1 0.1 0.1
Minimum capacity, Min 1) g 0.5 0.5 0.5
Response time (average) s 1.1 1.1 1.1
Range of use according to CD 1) g 0.5 – 6200 0.5 – 4200 0.5 – 2200
Allowable ambient
operating temperature °C +10 to +30 (+50°F to +86°F)
Net weight, approx. kg 3.5
Weighing pan size mm 180 + 180
Dimensions (W + D + H) mm 230 + 303 + 91

Model ATL-822-V ATL-8201-V ATL-6201-V


Type BD ED 200 BD ED 200 BD ED 200
Accuracy class 1) K K K
Weighing capacity, Max 1) g 820 8200 6200
Scale interval d 1) g 0.01 0.1 0.1
Tare range (subtractive) < 100% of the maximum capacity
Verification scale interval e 1) g 0.1 1 1
Minimum capacity, Min 1) g 0.5 5 5
Response time (average) s 1.1 1 1
Range of use according to CD 1) g 0.5 – 820 5 – 8200 5 – 6200
Allowable ambient
operating temperature °C +10 to +30 (+50°F to +86°F)
Net weight, approx. kg 3.5
Weighing pan size mm 180 + 180
Dimensions (W + D + H) mm 230 + 303 + 91

) CD = Council Directive 90/384/EEC for non-automatic weighing instruments; applicable to the European Economic Area
1

54
Accessories

External calibration weights:


For model Accuracy class Weight in grams Order no.
ATL-224 E2 200 YCW5228-00
ATL-124 E2 100 YCW5128-00
ATL-623 E2 500 YCW5528-00
ATL-4202 E2 2000 YCW6228-00
ATL-6202 E2 5000 YCW6528-00
ATL-153 F1 100 YCW5138-00
ATL-423 F1 200 YCW5238-00
ATL-2202 F1 2000 YCW6238-00
ATL-822 F2 500 YCW5548-00
ATL-8201, ATL-6201 F2 5000 YCW6548-00
or ± 25 mg 5000 YSS653-00

Product Order No. Product Order No.


Data printer YDP03-0CE Density determination kit1)
with date, time, statistics – for ED224S, ED124S YDK01LP
evaluation, transaction
counter functions and LCD Industrial AC adapter, model ING2,
protection rating: IP65 in accordance
Remote display1), reflective YRD02Z with EN 60529
(for connection to data interface port) – for 230 V 69 71899
– for 120 V 69 71500
External rechargeable battery pack YRB05Z
With battery-level indicator (LED); can be recharged using Data cable
the AC adapter (charge time for completely discharged – for connecting a computer with
battery pack: 15 hours); see "Specifications" for hours of a USB port YCC01-USBM2
operation. To recharge the battery pack: – for computer connection, 25-pin 7357312
Unplug the AC adapter from the balance and plug it into the – for computer connection, 9-pin 7357314
battery pack
Adapter cable 6965619
SartoConnect1) from D-Sub 25-pin male
data transfer software for direct transmission of weight connector to D-Sub 9-contact
values to another program (e.g., MS Excel) female connector; length: 0.25 m
– with RS-232C connecting cable, length: 1 m (~20 in)
YSC01L Anti-theft locking device LC-1
– with RS-232C connecting cable, length: 5 m (~16 ft)
YSC01L5
– with RS-232C connecting cable, length: 15 m (~50 ft)
YSC01L15

) Not for verified models


1

55
C Marking

The balance complies with the following EC Directives and On request, Sartorius will provide information on the
European Standards: minimum operating specifications (in accordance with
Council Directive 89/336/EEC the Standards listed above for defined immunity to
“Electromagnetic compatibility (EMC)” interference).
73/23/EEC “Electrical equipment designed for
Applicable European Standards: use within certain voltage limits”
Limitation of emissions:
In accordance with product standard EN 61326-1 Class B Applicable European Standards:
(residential area)
EN 60950
Defined immunity to interference: Safety of information technology equipment including
in accordance with product standard EN 61326-1 electrical business equipment
(minimum test requirements,
non-continuous operation) EN 61010
Safety requirements for electrical equipment for measure-
Important Note: ment, control and laboratory use Part 1: General require-
The operator shall be responsible for any modifications ments
to Acculab equipment and must check and, if necessary,
correct these modifications. If you use electrical equipment in installations and under
ambient conditions requiring higher safety standards,
you must comply with the provisions as specified in the
applicable regulations for installation in your country.

56
57
Address label/Dealer’s stamp

Copyright by Acculab.
All rights reserved. No part of this publication
may be reprinted or translated in any form
or by any means without the prior written
permission of Acculab.
The status of the information, specifications
and illustrations in this manual is indicated
by the date given below.
Acculab reserves the right to make changes
to the technology, features, specifications
and design of the equipment without notice.

Status:
June 2007, Acculab

W1A000 · KT
Publication No.: WAT6002-e07062

You might also like