Acculab Atilon: Operating Instructions
Acculab Atilon: Operating Instructions
Acculab Atilon: Operating Instructions
ACCULAB ATILON
Electronic Analytical and Precision Balances
98648-015-55
Contents
26 Application Programs
27 Counting
29 Weighing in Percent
31 Calculation
33 Animal Weighing/Averaging
35 Net-total Formulation
37 Totalizing
39 Density Determination
41 Mass Unit Conversion
2
Warnings and Safety Precautions
! Disconnect the balance from power before connecting or $ Do not open the balance housing. If the seal is broken, this
disconnecting peripheral devices. will void all claims under the manufacturer’s warranty.
! If you operate the balance under ambient conditions $ If you have any problems with your balance, contact your
subject to higher safety standards, you must comply with local customer service center.
the applicable installation regulations.
Symbols
! Exposure to excessive electromagnetic interference can The following symbols are used in these instructions:
cause the readout value to change. Once the disturbance § indicates required steps
has ceased, the instrument can be used again in
accordance with its intended purpose. $ indicates steps required only under certain conditions
Make sure that no liquid enters the equipment housing;
use only a slightly moistened cloth to clean the balance. > describes what happens after you have performed a
particular step
Installation
! Make sure the voltage rating printed on the power supply – indicates an item in a list
is identical to your local line voltage.
– Proceed with extreme caution when using pre-wired ! indicates a hazard
RS-232 connecting cables, as the pin assignments may
not be compatible with Acculab equipment. Before
connecting the cable, check all pin assignments against
the cabling diagrams and disconnect any lines that are
assigned differently.
3
Getting Started
$ If you detect any damage, proceed as directed in the Conditioning the Balance
chapter entitled “Care and Maintenance,” under “Safety Moisture in the air can condense on cold surfaces
Inspection.” whenever the equipment is moved to a substantially
warmer place. To avoid the effects of condensation, allow
$ Save the box and all parts of the packaging for any future the balance to sit for 2 hours, at room temperature, before
transport. Disconnect all cables before packing the balance plugging into AC power.
for shipping.
Seal on Balances Verified for Use in Legal
Equipment Supplied Metrology in the EU*:
– Balance For instruments of accuracy class K, EU legislation
– Weighing pan requires that a control seal be affixed to verified balances.
– Pan support (only for models with a round weighing pan) The control seal consists of a sticker. If the seal is broken,
– AC adapter the verification becomes null and void and the balance
must be re-verified.
Additional equipment supplied with models
ATL-224, -124:
– Sliding-door draft shield
– Drip/breeze ring
– Draft shield base plate
4
Installation
5
Connecting the Balance to AC Power/Safety Precautions
Use only original Acculab AC adapters. For use within
– Europe: part no. 6971412
– U.S./Canada: part no. 6971413
§ 1) Connect the angle plug to the balance
§ 2) Connect the AC adapter to the wall outlet (mains)
Safety Precautions
Plug-in AC Adapter 6971412/6971413:
The AC adapter rated to Class 2 can be plugged into any wall outlet without
additional safety precautions.
6
NOTE: This equipment has been tested and found to If you have a Class B digital device, please read and follow
comply with the limits pursuant to part 15 of FCC Rules. the FCC information given below:
These limits are designed to provide reasonable protection “However, there is no guarantee that interference will not
against harmful interference. This equipment generates, occur in a particular installation. If this equipment does
uses and can radiate radio frequency energy and, if not cause harmful interference to radio or television reception,
installed and used in accordance with these instructions, which can be determined by turning the equipment off and
may cause harmful interference to radio communications. on, the user is encouraged to try to correct the interference
by one or more of the following measures:
For information on the specific limits and class of this
equipment, please refer to the Declaration of Conformity. – Reorient or relocate the receiving antenna.
Depending on the particular class, you are either required
or requested to correct the interference. – Increase the separation between the equipment and
receiver.
If you have a Class A digital device, you need to comply
with the FCC statement as follows: “Operation of this – Connect the equipment into an outlet on a circuit different
equipment in a residential area is likely to cause harmful from that to which the receiver is connected.
interference in which case the user will be required to
correct the interference at his own expense.” – Consult the dealer or an experienced radio/TV technician
for help.”
7
Connecting Electronic Peripheral Devices
§ Make sure to unplug the balance from AC power before you connect or disconnect
a peripheral device (printer or computer) to or from the interface port.
Warmup Time
To ensure accurate results, the balance must warm up before operation
as follows:
– Models ATL-153 and ATL-822: 2 hours
– All other models: 30 minutes
Only after this time will the instrument have reached the required operating
temperature.
$ Make sure to allow the equipment to warm up for at least 24 hours after initial
connection to AC power or after a relatively long power outage.
8
Below-Balance Weighing
A port for a below-balance weighing hanger
is located on the bottom of the balance.
§ Using the built-in hook 1: Attach the sample (e.g., using a suspension wire)
to the hanger.
9
Leveling the Balance
Purpose:
– To compensate for unevenness at the place of installation
Always level the balance again any time after it has been moved to a different
location. Only the 2 front feet are adjusted to level the balance.
L R
§ Retract the two rear feet (only on models with a rectangular weighing pan).
R L § Turn the 2 front feet as shown in the diagram until the air bubble is centered
within the circle of the level indicator.
10
Operation
11
10 3
9 4
8 5
7 6
ON/OFF
2
3
7 4
6 5
11
Basic Weighing Function
§ Tare the balance, if necessary: § Before using the balance as a legal measuring instrument,
Press (TARE) calibrate and adjust it at the place of use using the built-in
motorized calibration weight; for details, see
$ If necessary, change the configuration settings: “Calibration/Adjustment” in this chapter.
See the chapter entitled “Configuration”
$ The temperature range (°C) indicated on the verification
$ If desired, load the factory settings: label must not be exceeded during operation.
See the chapter entitled “Configuration”
Additional Functions
$ Switching off the balance:
Press (ON/OFF)
12
Example
Simple Weighing
13
Calibration and Adjustment
Features
Calibration/adjustment can be performed only when:
– there is no load on the balance
– the balance is tared
– the internal signal is stable
– for external calibration, the value displayed for the
calibration weight on the balance does not differ from the
nominal weight value by more than 2%
External Calibration in Verified Balances
If these conditions are not met, an error message is When the balance is used in legal metrology, external
displayed “Err 02”. calibration is blocked as follows:
– The setting of the menu access switch is locked
You can use any of the following weight units in (see “To block calibration/adjustment”)
calibration/adjustment: – The cap over the menu access switch is sealed
CAL.Einh. - gramm, Kilogr. or Pound
(not for verified models) For details on generating an ISO/GLP-compliant printout of
calibration/adjustment results, see page 43.
Following calibration/adjustment, the application program
is cleared.
Internal Calibration/Adjustment
In the operating menu, select Cal.-Adj.-Cal.Int.
before beginning.
The built-in motorized calibration weight is applied and
removed automatically for internal calibration.
The built-in motorized calibration weight is applied and removed automatically for internal calibration.
15
External Calibration
16
Configuration (Operating Menu)
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
17
Menu Navigation
or
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
18
Parameter Settings: Menu
Level 1 Level 2 Level 3 Menu code
19
Parameter Settings: Overview
ο = Factory setting √ = User-defined setting
1
) Setting cannot be changed on verified balances
2
) Only on ATL-…-I, ATL-…-V models
20
Level 1 Level 2 Level 3 Level 4 Menu code
In the setting “SBI”, the non-verified display digit is not automatically identified.
Please take the corresponding measures or adjust the settings on the peripheral device.
21
Level 1 Level 2 Level 3 Level 4 Menu code
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
22
Level 1 Level 2 Level 3 Level 4 Menu code
Applic. Weigh 2. 1.
Applic. Unit Disp.Dig.1) ο All 2. 2. 2. 1
programs Toggle units Display Minus 1 2. 2. 2. 2
accuracy
Counting Resolut. ο Disp.Acc. Display accuracy 2. 3. 1. 1
Resolution 10-fold 10 times > disp. 2. 3. 1. 2
Ref.Updt. ο Off 2. 3. 2. 1
Autom. reference Auto 2. 3. 2. 2
updating
Percent Dec.Plcs None No dec. places 2. 4. 1. 1
Weighing Decimal places ο 1 Dec.Pl. 1 decimal place 2. 4. 1. 2
in percent 2 Dec.Pl. 2 decimal places 2. 4. 1. 3
3 Dec.Pl. 3 decimal places 2. 4. 1. 4
Net-Tot Comp.Prt. Off 2. 5. 1. 1
Net-total Component ο On 2. 5. 1. 2
printout
Total Comp.Prt. Off 2. 6. 1. 1
Totalizing Component ο On 2. 6. 1. 2
printout
AnimalW. Activty. Calm (fluct.: 2% of test obj.) 2. 7. 1. 1
Animal Animal activity ο Active (fluct.: 5% of test obj.) 2. 7. 1. 2
weighing V.Active(fluct.: 20% of test obj.) 2. 7. 1. 3
Start Manual 2. 7. 2. 1
ο Auto. Automatic 2. 7. 2. 2
Calc. Method ο Mul. Multiplier 2. 8. 1. 1
Calculation (operator) Div. Divisor 2. 8. 1. 2
Dec.Plcs None No dec. places 2. 8. 2. 1
Decimal ο 1 Dec.Pl.1 decimal place 2. 8. 2. 2
places 2 Dec.Pl.2 decimal places 2. 8. 2. 3
3 Dec. Pl. 3 decimal places 2. 8. 2. 4
Density Dec.Plcs None No dec. places 2. 9. 1. 1
Density Decimal places ο 1 Dec.Pl.1 decimal place 2. 9. 1. 2
determination
1
) Setting cannot be changed on verified balances
23
Input: ID number, Date and Time
Depending on the setting you select for the menu item “SETUP-PRNT.OUT-DATE,” the date will be displayed in the
following format:
MMM.DD.YY
Depending on the setting you select for the menu item “SETUP-PRNT.OUT-TIME,” the time will be displayed in the
following format:
12-hour format
24
Example: ID No., Date and Time
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
25
Device Information
Application Programs
Using Verified Balances as Legal Measuring Instruments in the EU*:
All application programs can be selected on balances used as legal measuring instruments.
Calculated values are alternately indicated with the following symbols:
– Percent =%
– Piece count (Counting) = pcs
– Computed value = o, G
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
26
Counting
o = Factory setting
27
Example : Counting parts of equal weight
Parameter: Applic. - Count. (menu code 2. 3.)
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
28
Weighing in Percent
o = Factory setting
29
Example: Determining residual weight in percent
Parameter settings: Applic. Percent (menu code 2. 4.)
Reference percentage: Ref 100%
3. Information:
Enter reference percentage (MENU)* ref 100 %
(Changing the reference: see the previous page)
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
30
Calculation
o = Factory setting
31
Example:
Calculating the weight per unit area of paper: An A4 sheet of paper is used in this example, with surface dimensions of 0.210 m +
0.297 m = 0.06237 m2. To determine the weight per unit area, the total weight is divided by the surface.
Parameter settings:
Applic. Calc.-Method-Div. (menu code 2. 8. 1. 2)
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
32
Animal Weighing/Averaging
o = Factory setting
33
Example: Determining animal weight with automatic start and 20 subweighing operations (measurements)
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
34
Net-total Formulation
o = Factory setting
35
Example: Counting parts into a container
Parameter settings:
Applic.-Net-Tot (menu code 2. 5.)
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
36
Totalizing
37
Example: Totalizing weight values
Parameter settings:
Applic. Total Comp.Prt On (Code 2. 6. 1. 2)
38
Density Determination
Preparation
§ Select the Density Determination application in the menu:
see “Configuration.”
o = Factory setting
39
Example: Determining the density of a solid sample.
Parameter settings:
Applic.-Density-Dec.Plcs-1 Dec. Pl (menu code 2. 9. 1. 2)
40
Mass Unit Conversion
Purpose Features
With this application program you can change the weight – Set the basic unit and display accuracy in the Setup menu:
value displayed from the basic weight unit to any of 4 see “Configuration.”
application weight units (see table on next page). – Set the application weight units and display accuracies in
the Application menu.
Preparation – These settings are stored in battery-backed memory.
§ Select the Unit application for toggling weight units: – The basic unit is active when the balance is powered up.
see chapter on “Configuration” (Parameter Settings)
Applic.ation program
Toggle wt, unit
dec.plcs. Display accuracy
o All All decimal places
minus1 Reduced by 1 place
o = Factory setting
41
Example: Change display from the basic unit (in this example, grams [g]) to pounds [lb] and then to Troy ounces [ozt].
Preparation:
1 Begin selection of an application weight unit (MENU)* None o
Conversion:
8. Place sample on balance + 100.00 g
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
42
The following weight units are available in your Atilon balance (in legal metrology, only units permitted by national law are
available):
! Some weight units may be blocked from use in legal metrology, depending on national verification laws.
43
ISO/GLP-compliant Printout/Record
44
The ISO/GLP-compliant printout can contain the following lines:
45
Data Interface
Purpose Preparation
Your balance comes equipped with an interface port for You can set these parameters for other devices in the Setup
connection to a computer or other peripheral device. You menu (see the chapter entitled “Configuring the Balance”).
can use a computer to change, start and/or monitor the
functions of the balance and the application programs. You will also find a detailed description of the available
data interface commands in the file “Data Interface
Descriptions for ATL Models”, which you can download
from the Acculab website at www.acculab.com by selecting
“Download Instruction Manuals” from the “Service” menu.
Female Interface Connector: The many and versatile properties of these balances can be
fully utilized for printing out records of the results when you
connect your balance to a Acculab data printer. The record-
1 13
14 25
ing capability for printouts makes it easy for you to work in
compliance with ISO/GLP.
46
Troubleshooting Guide
Error codes are shown on the main display for approx. 2 seconds.
The program then returns automatically to the previous mode.
47
Care and Maintenance
Repairs
§ Repair work must be performed by trained service technicians. Any attempt by
untrained persons to perform repairs may result in considerable hazards for the
user. If the instrument requires repairs, please contact your Acculab dealer.
Cleaning
§ Unplug the AC adapter from the wall outlet (mains supply). If you have an inter-
face cable connected to the balance port, unplug it from the port.
§ After cleaning, wipe down the balance with a soft, dry cloth.
48
Recycling
> Lock the equipment in a secure place to ensure that it can- The equipment, including accessories and
not be used for the time being. batteries, does not belong in your regular
household waste. The EU legislation
Notify your Acculab dealer. Repair work must be performed requires its Member States to collect elec-
by trained service technicians. trical and electronic equipment and dis-
posed of it separately from other unsorted municipal waste
We recommend having the power with the aim of recycling it.
supply inspected by a certified electrician at regular inter-
vals, according to the following checklist:
– Insulating resistance: > 7 megaohms measured with a For disposal in Germany and in the other Member States of
constant voltage of at least 500 volts at a 500 K-ohm load the European Economic Area (EEA), please contact our ser-
– Leakage current: < 0.05 mA measured with a properly vice technicians on location or your Acculab dealer.
calibrated multimeter
In countries that are not members of the European Econom-
ic Area (EEA) or where no Acculab dealers are located,
please contact your local authorities or a commercial
disposal operator.
49
Overview
Specifications
50
Specifications for Individual Models
51
Specifications for Individual Models
52
Verified Models with EC-type Approval: Specifications
) CD = Council Directive 90/384/EEC for non-automatic weighing instruments; applicable to the European Economic Area
1
53
Verified Models with EC-type Approval: Specifications
) CD = Council Directive 90/384/EEC for non-automatic weighing instruments; applicable to the European Economic Area
1
54
Accessories
55
C Marking
The balance complies with the following EC Directives and On request, Sartorius will provide information on the
European Standards: minimum operating specifications (in accordance with
Council Directive 89/336/EEC the Standards listed above for defined immunity to
“Electromagnetic compatibility (EMC)” interference).
73/23/EEC “Electrical equipment designed for
Applicable European Standards: use within certain voltage limits”
Limitation of emissions:
In accordance with product standard EN 61326-1 Class B Applicable European Standards:
(residential area)
EN 60950
Defined immunity to interference: Safety of information technology equipment including
in accordance with product standard EN 61326-1 electrical business equipment
(minimum test requirements,
non-continuous operation) EN 61010
Safety requirements for electrical equipment for measure-
Important Note: ment, control and laboratory use Part 1: General require-
The operator shall be responsible for any modifications ments
to Acculab equipment and must check and, if necessary,
correct these modifications. If you use electrical equipment in installations and under
ambient conditions requiring higher safety standards,
you must comply with the provisions as specified in the
applicable regulations for installation in your country.
56
57
Address label/Dealer’s stamp
Copyright by Acculab.
All rights reserved. No part of this publication
may be reprinted or translated in any form
or by any means without the prior written
permission of Acculab.
The status of the information, specifications
and illustrations in this manual is indicated
by the date given below.
Acculab reserves the right to make changes
to the technology, features, specifications
and design of the equipment without notice.
Status:
June 2007, Acculab
W1A000 · KT
Publication No.: WAT6002-e07062