Model LCT01 Load Cell Tester Manual
Model LCT01 Load Cell Tester Manual
Table of Contents
The Strain Gauge Load Cell ............................................... 3
2
Model LCT01 Load Cell Tester Manual
Consider the example of an axial type load cell, which usually consists
of a hollow or solid cylindrical shaft and four strain gauges mounted
around the circumference. The strain gauges are mounted and con-
nected to form a Wheatstone bridge circuit. The basic relation of Stress
= Load/Area (s = P/A) and Strain = Stress/Young’s module (e = s/E) can
be used to determine the strain under different loads.
3
Model LCT01 Load Cell Tester Manual
The relation between the applied load and the output signal may be
found by calibration: applying known loads and measuring the resultant
output voltage. The resulting curve in most load cells is almost linear,
and this calibration allows for the conversion of the output signal to its
source: the applied load.
Consider the example of a load cell with a rated output of 2 mV/V, and a
maximal load of 10 kg. If the output signal with no load is 0, feeding the
load cell with 10 V excitation voltage will result in an output signal of 20
mV when the applied load is 10 kg. Since the output signal is linearly
proportional to the load applied, a 5 kg load would give an output signal
of 10 mV.
In every load cell the rated output, the available excitation voltage, and
the maximal load are defined in the manufacturer’s specification.
The LCT01 supplies a quick initial evaluation for all types of strain
gauge load cells. Due to its portability and design, the LCT01 is useful
not only in laboratories, but also on-site in hard-to-access locations.
The load cell can be tested without being removed from the installation.
4
Model LCT01 Load Cell Tester Manual
For example, if the output signal displayed by the LCT01 is 26% for a
10 kg load cell, increasing the load by 1 kg (10% of the 10 kg
maximal load) should add 10% to the displayed output signal.
If the load cell is functioning correctly, the expected output signal for a
1 kg load should therefore be 36%.
5
Model LCT01 Load Cell Tester Manual
Battery Installation
If the LCT01 is idle for five minutes it will automatically shut down.
Previously entered load cell characteristics, as well as the last test
results, are maintained in memory until the unit shuts down.
Note: To reuse the previous load cell characteristics press the Scroll
Down button periodically in order to restart the five-minute countdown.
When the batteries need replacing you will see the following message
on the LCD Display:
LOW BATTERY
REPLACE BATTERY !
6
Model LCT01 Load Cell Tester Manual
Follow the procedure below to test your load cell. You will find it useful
to have the load cell specification sheet available.
Attach load cell wires according to the wiring code provided by the
manufacturer (refer to page 2 for the color wiring guide). When the
wires are connected, press the ON/OFF button.
You will see the following screen as the LCT01 performs its self-test.
Note: If the screen is blank or if you see a low battery warning replace
batteries.
The LCT01 will ask if you are testing a four-wire or six-wire load cell.
Six-wire load cells include Sense+ and Sense- (used to detect the
voltage drop along the wire) which run in parallel to the Input+ and
Input- wires.
Press the Scroll Up key to change the default answer. When the display
is correct, press the Enter key.
7
Model LCT01 Load Cell Tester Manual
Load cell output ranges between 1 mV/V and 5 mV/V (most common
being 2 mV/V and 3 mV/V), depending on the model and type of load
cell. The output rate scales the voltage transmitted by the load cell in
the Signal lines which is used in the Output Signal result.
The default Load Cell Output is 2.0 mV/V. Adjust the rated output
according to the load cell specification sheet by pressing the Scroll Up
or Scroll Down key. The value will change in increments of 0.1 mV.
Press the Enter key to accept the displayed output.
Start Test
The LCT01 will now ask you to make sure that the wires are con-
nected. Press the Enter key to begin the test.
WIRES CONNECTED?
PRESS ↵ TO START
Dynamic Test
After pressing the Enter key, the LCT01 will perform the dynamic test.
TEST IN PROGRESS
PLEASE WAIT . . .
When the test is over you will see the following screen:
* END OF TEST *
....................
8
Model LCT01 Load Cell Tester Manual
PRESS TO SCROLL
PRESS ↵ TO START
Press the Scroll Up key to view the test results, or press Enter to run
another test.
Note: The load cell characteristics you set in the first test stay in
memory until the LCT01 is turned off or powers down after five minutes
of inactivity.
If the LCT01 finds results that definitely indicate faulty load cell
function, the Alarm LED will flash, a short beep will sound, and the
atypical result will blink in the relevant results screen.
Note: Absence of an alarm does not indicate that the load cell is
functioning properly. Always compare the LCT01 results with the data
sheet of your load cell.
INPUT = 351Ω
OUTPUT = 412Ω
Input
9
Model LCT01 Load Cell Tester Manual
Ground
Measured Value Resistance between load cell body and all lines
Tested Function The sum of load cell and cable resistance
Result* A result above 10 M Ohms usually indicates
normal function.
Minimum / Maximum Value 0.01 M Ohms / 10 M Ohms
Blinking Display Resistance less than 10 M Ohms
*Note: If the reason for testing the load-cell was a non-stable, or drifting result, it
is recommended to check the resistance between the shield and lines with a
device suitable for checking up to 1000 M Ohms.
Sense Pos.
Sense Neg.
10
Model LCT01 Load Cell Tester Manual
SIGNAL OUTPUT
EQUIVALENT +9%
Signal Output
Example
If a load cell’s maximum load is 100 kg, applying a 10 kg load during a test
should give a 10% change in the result
11
Model LCT01 Load Cell Tester Manual
Load Cell Color Wiring Guide (Manufacturers often change color codes: contact Load Cell Central to be sure)