K090588
K090588
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k090588
C. Measurand:
Vitamin B12, Ferritin, Folate and TSH
D. Type of Test:
Quantitative Chemiluminescent Immunoassay
E. Applicant:
Beckman Coulter, Inc.
G. Regulatory Information:
1. Regulation section:
21CFR Sec.-862.1810 - Vitamin B12 test system.
21CFR Sec.-866.5340 - Ferritin immunological test system.
21CFR Sec.-862.1295 - Folic acid test system.
21CFR Sec.-862.1690 - Thyroid stimulating hormone test system.
21CFR Sec.- 862.2160 - Discrete photometric chemistry analyzer for clinical use.
2. Classification:
II, II, II, II, I respectively
3. Product code:
CDD - Radioassay, Vitamin B12
DBF - Ferritin, Antigen, Antiserum, Control
CGN - Acid, Folic, Radioimmunoassay
JLW - Radioimmunoassay, Thyroid-Stimulating Hormone
JJE - Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
4. Panel:
Chemistry (75) Immunology (91)
H. Intended Use:
1. Intended use(s):
See indication(s) for use below.
The UniCel DxI 800 Access Immunoassay System with laboratory automation
connection is a microcomputer-controlled, random and continuous access
analyzer that includes an external computer. This computer stores the system user
interface (UI) software and allows the operator to interface with and direct the
instrument software. The UniCel DxI 800 System uses enzyme immunoassays
(utilizing paramagnetic particle solid phase and chemiluminescent detection) for
determination of various analytes, such as Vitamin B12, Ferritin, Folate and
hTSH along with other various enzyme immunoassays assays that may be
adaptable to the analyzer depending on the reagent used to induce the enzyme
immunoassay reaction. The UniCel DxI 800 System is an in vitro diagnostic
device for use in the clinical laboratory.
I. Device Description:
The Power Processor is a scalable laboratory automation system (LAS) designed to
streamline pre-analytical processes in the clinical laboratory. A basic Power
Processor System is comprised of a Line Control Computer, Preplink™ Computer,
Inlet Module, Hematology Module, Centrifugation Module, Decapper Module, and
Outlet Module. In the basic configuration, patient sample tubes are loaded onto the
Power Processor system to be sorted to a Hematology Module, or to be centrifuged,
decapped, and sorted to Personality Racks for further processing on other
instruments. Additional modules may be added for aliquot capability, sample
capping, and refrigerated storage.
The Power Processor is an open architecture system that can connect to a variety of
clinical analyzers. Connection modules are extensions of the Power Processor track
system that link with an analyzer’s existing LAS interface. Connection modules
support one of two types of sample transfer methods: onboard or outboard sampling.
Onboard sampling physically transfers the sample tube/rack from the automation
track to the analyzer’s sample load and identification area. With outboard sampling,
the connection unit performs the sample bar code read function, presents the sample
ID to the connected analyzer, and then signals for direct sampling of the open tube by
the connected instrument at an aspiration point on the automation track. The Power
Processor Generic Connection Module is specifically designed to support the
outboard sampling method based on point-in space pipetting technology aligned with
the CSLI guidelines. This method is used to establish connection with Beckman
Coulter’s UniCel DxI 800 Immunoassay System. Power Processor software version
3.5 establishes a dynamic or “smart” connection with the UniCel DxI 800 System to
enable sample routing based on reagent and calibration status.
L. Test Principle:
Paramagnetic particle solid phase and chemiluminescent detection with an automated
sample handling system
1. Analytical performance:
a. Precision/Reproducibility:
Provided in above referenced assay 510(k)’s
e. Analytical specificity:
Provided in above referenced assay 510(k)’s
f. Assay cut-off:
Provided in above referenced assay 510(k)’s
2. Comparison studies:
a. Method comparison with predicate device:
In order to establish equivalence to manually introduced samples to the
analyzer verses automated sample introduction and identification to the
analyzer a method comparison was performed using the above referenced
assays. The method comparison study employed paired sample testing and
Deming regression analysis to assess the data. The assay test menu was
selected by the evaluator and specimen selection was based on the site test
requests. In the study, two blood samples were drawn from each patient. One
sample tube was processed through the laboratory protocol using the Power
Processor System with Generic Connection Module interface to the DxI 800
analyzer, and the other tube was processed on the AutoMate system and
manually loaded on the analyzer.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
4. Clinical cut-off:
Not Applicable
N. Instrument Name:
O. System Descriptions:
1. Modes of Operation:
The Power Processor system can consist of the following interfaces: Instrument
Manager, PrepLink, and Line Control Computer. The LIS sends sample
programming to the PrepLink through the optional Instrument Manager in some
software setups. Sample programming information is used to determine where to
sort sample tubes. The Instrument Manager consolidates multiple sample
programming requests for each sample tube, allowing the Power Processor system
to sort sample tubes to the appropriate racks.
The Line Control software communicates with the PrepLink software to keep
track of sample tubes on the Power Processor system. The Line Control software
sends and receives information from the PrepLink software and the Power
Processor track. This information is contained in the Systems Error Log and the
Operations Error Log (both can be used to evaluate and correct processing
problems). The Line Control Computer Main screen displays a graphical
representation of the automation line.
The Power Processor can be configured with optional hardware to automatically
load centrifuged and decapped sample tubes onto Beckman Coulter’s 800 DxI
System. The Connection Modules operate identically and are designed to support
an onboard sampling mechanism, where the sample tubes are transferred directly
to the connected analyzer sample load area and read by the system barcode reader.
The Power Processor can be configured with optional software and hardware to
allow processing of sample tubes on Connection Instrument. The Power Processor
performs all pre-analytical sample tube preparation, and then sorts the sample
tubes directly to Connection Module where the samples are pipetted for testing.
After the samples are pipetted, the tubes can route to other instruments for
additional testing or to Outlet Racks. Routing decisions are made by the PrepLink,
and are based on the sample programming of each sample tube, the availability of
the Connection instrument, and configuration parameters set by the operator.
2. Software:
The applicant provided software documentation that supports the device was
designed and developed under good software LifeCycle processes.
3. Specimen Identification:
Bar code
On the DxI 800 connection, the Connection Module reads the sample bar code
read prior to releasing the sample from the aspiration point, resulting in the
following sequence from the DxI perspective:
5. Calibration:
Provided in k023764
6. Quality Control:
Provided in k023764
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a
substantial equivalence decision.